1. Introduction
Impedance spectroscopy is an established measurement method for characterizing electrical properties, i.e., complex conductivity and permittivity of materials and systems [
1]. Impedance measurement systems have been developed for a variety of applications, including material characterization with inductive sensors [
2], chemical parameters investigation [
3], and characterization of biological tissues [
4,
5]. Impedance spectroscopy can be applied to monitor human activity [
6], diagnose muscles [
7], detect and characterize cancer [
8], and monitor human health on a daily basis [
9]. In this case, it is called bioelectrical impedance spectroscopy (BIS) [
5], which is the main focus of this paper. BIS is a non-invasive technique that injects a constant safe current into the biological tissue or human body through a pair of electrodes in a predefined frequency range. Then, a voltage is sensed through the second pair of electrodes reflecting the actual status of the biological tissues, and the bioimpedance is calculated based on the analysis of these two signals [
10].
Unlike other device-under-test (DUT), bioimpedance spectroscopy is injected into human tissues, which requires further attention during system design. It is crucial to design a system that is resilient to the signals generated by the human body while safe enough for human injection. Conventionally, large laboratory equipment has always been used. Nevertheless, such devices are typically bulky and require a permanent infrastructure. Nowadays, those devices are phasing out and being replaced with bioimpedance-embedded solutions. These portable devices are gaining more accessibility owing to their minimal costs, weight, and size. Not only that but many of them can be wearable, which makes them more practical, especially in the era of continuous data collection for machine learning applications. However, this miniaturization of the bioimpedance device is not without drawbacks. As one important part, the system should consider many aspects, including measurement accuracy, lower power consumption, system size, and cost.
In this direction, several studies are carried out. Santoso et al. [
11] developed a BIS system based on a gain phase detector (GPD) chip from analog devices covering the frequency range from 1 kHz to 100 kHz with a maximum deviation of 1.5%. Patil et al. proposed a bioimpedance spectrometer based on a GPD module working in the
dispersion range for total body water monitoring [
10]. In the same context, Anusha et al. [
12] designed a measurement system based on the same technique and validated it in a load range from 1.6
to 1.6 k
in the frequency range from 500 Hz to 1 MHz. On the other hand, a second technique has been investigated based on I-U measurements and fast Fourier transform (FFT) for impedance calculation. Munjal et al. [
13] propose a solution with an extended frequency range until 10 MHz using the under-sampling technique and achieves a maximal standard deviation of 0.5% and 0.55° deviation in amplitude and phase relatively. Based on the same technique, Vergas et al. designed a portable system for detecting abnormalities in the gastrointestinal tract in the range from 10 Hz to 1 MHz with an accuracy of 2% [
14]. A third technique is implemented for BIS systems IQ demodulation measurements. Ibrahim et al. proposed a measurement solution for impedance range from 1–120
for a frequency range from 4 kHz to 120 kHz [
15]. Huynh et al. developed a measurement solution using the IQ demodulation technique to monitor blood pressure based on bio-impedance measurements [
16]. In the same direction, based on the IQ demodulation technique, Hernánder et al. proposed a completed BIS device for body composition estimation with a sophisticated calibration method for better measurement accuracy [
17]. From another perspective, Kusche et al. evaluate the potential of BIS measurement for hand gestures recognition based on novel extracted features [
6].
Based on the research on concurrent BIS systems and the literature study, three primary techniques are mainly used for impedance measurements: gain phase detector, I-U measurement based on fast Fourier transform analysis, and I-Q demodulation measurements. However, a considerable divergence in the literature can be noted. For example, many research papers do not compare the method performance to reference measurement but use the standard deviations between the measurements. While this could be useful to demonstrate the precision of such devices, no information about the accuracy can be extracted. Other papers do not compare these methods fairly as separate excitation, and measurement parts are used in the comparison. All this diversity allows only the comparison on the system level but not on the method level, which is more interesting. Accordingly, this paper focuses on studying and evaluating the three measurement techniques under the same conditions to conclude the advantages and limitations of each one for further development. For that purpose, as a first step, a signal generation unit based on AD9850 commercial direct digital synthesis (DDS) is implemented for all measurement methods, followed by a voltage-controlled current source (VCCS) Tietze cascade structure as they are the shared blocks during the comparison. The excitation signal is typically a stepped sine ranging from 1 kHz to 1 MHz to cover and dispersion. Then each measurement technique is implemented separately and evaluated using three selected DUTs: RC parallel, Cole–Cole model 1, and Cole–Cole model 2. Finally, the comparative study is carried out based on the impedance measurement, phase and magnitude relative deviations, power consumption, and hardware and signal processing complexities.
4. Results and Discussion
The three measurement techniques are implemented on separate PCBs. The PCBs maintain the same designs as the standard electronic blocks to avoid any possible influences on PCB layout. The techniques will be tested on three selected DUTs for this work: R/ /C element, Cole–Cole model 1, and Cole–Cole model 2. In the experimental study, all the DUTs were measured using the commercial impedance analyzer Agilent 4294a as a reference. The measurements are done on 100 frequency points from 1 kHz to 1 MHz for the GPD and FFT-based systems and only for 30 frequency points for the IQ demodulation system, which belongs to the 100 main points. This limitation in the latter method is mainly due to the limited standard values of resistor and capacitors values required to build a local oscillator circuit for the generation of 90° phase shift signal.
4.1. RC Elements
As the first evaluation, a 4.7 k
in parallel with a capacitor of 2.2 nF is taken as DUT.
Figure 9 illustrates the archived results. The GPD-based system in
Figure 9a,b are shown an enormous deviation in the magnitude and phase measurement, which are about 38% and 11° at low frequencies, respectively. This is mainly due to the over-range voltage in the INPA pin in AD8302, which exceeds the maximum allowed value of 0.7 V. This explains the considerable measurement deviation and shows the major limitation of this technique. The GPD-based system achieves a mean deviation of 33.25% and 3.437° for the magnitude and phase, respectively. On the other hand, the IQ demodulation system is showing better results with a mean deviation of magnitude relative deviation equal to 8.593% for the magnitude measurements and 3.394° for phase measurements. The FFT-based system presents the highest accurate results with 1.122% as the mean relative magnitude deviation and 0.972° mean of absolute phase deviation.
4.2. Cole–Cole Model
A Cole–Cole model is one most accessible equivalent models of biological tissue. As shown in
Figure 10a, a 330
resistor is in series to parallel 1.5 k
and 10 nF. These values are selected to simulate the real body composition of extracellular liquid resistance, intracellular liquid resistance, and membrane capacitance, respectively. The output of the instrumentation amplifiers is ensured to be less than 0.7 V for all frequencies to overcome the limitation of the GPD system and ensure the same comparison condition. The GPD-based system performs much better than the previous RC parallel, where the deviation decreased to 9.1% and 3.243° for the magnitude and phase, respectively. The IQ demodulation achieves a 3.299% relative mean deviation of magnitude and 1.803° mean deviation of phase. The IQ-based system shows higher deviation at higher frequencies due to the low voltage output of the low-pass filter. It can be remarked that the output decreases down to a few mV, which is at the proximity of ADC quantization level and noise, which lowers the measurement accuracy at that range.
On the other hand, the UI/FFT-based system shows accurate results, reaching only 1.643% mean relative deviation in magnitude and 0.751° mean relative deviation of phase. Nonetheless, this system shows a higher deviation in higher frequencies in the proximity of the sampling frequency. The used STM32
C has a sampling frequency of 1.8 MHz, and the higher the frequencies, the more quantization deviation is expected [
18]. Nevertheless, this sampling frequency is sufficient to obtain high accuracy, especially in phase measurements until 200 kHz, which is adequate for the targeted application.
A second possible form of the Cole–Cole model (see
Figure 11) is tested and evaluated with the same passive components. All system shows comparable performance to the first Cole–Cole model. The mean relative deviations of the magnitude are 8.159%, 5.145%, and 1.948% for the GPD, IQ, and FFT-based systems, respectively. On the other side, the mean relative deviations of phase 2.605°, 1.661°, and 0.886°, respectively.
As a partial conclusion, the UI/FFT-based system has better accuracy in terms of magnitude and phases compared to GPD and IQ for impedance measurements in the same range as the Cole–Cole model.
4.3. Comparative Study
Three systems for BIS based on different measurement techniques are well implemented and tested under the same conditions. The systems are tested in the frequency range from 1 kHz to 1 MHz using the same signal generator, and VCCS modules explained above. Based on the realized experiments, the accuracy in term phase and magnitude measurements are evaluated.
Table 1 summarizes the achieved results while measuring three different DUTs (RC parallel, Cole–Cole model type 1, and Cole–Cole model type 2). However, the aim is not to compare only the system performance but also the system design challenges and requirements, e.g., signal processing and hardware complexities and power consumption, as shown in
Figure 12. The FFT-based measurement system shows accurate results, with a maximum deviation of 1.94% for relative magnitude and 0.972° for phase deviation to reference measurement from the 4294a Agilent impedance analyzer. In addition, the analog measurement structure is less complex, containing a total of 11 components, including only two integration circuits (ICs). However, this method requires more complex signal processing than other methods, as several aspects must be considered, such as sampling frequency and spectra leakage effect. For that reason, a robust
C is needed for embedded systems like STM32 to ensure accurate measurements for higher frequencies, which is also challenging.
On the other hand, the GPD solution has less signal processing complexity as the AD8302 GPD module provides two relative DC voltages to the magnitude and phase, respectively, with a not complex analog structure containing three main ICs and 25 components in total. The magnitude and phase will be calculated based on predefined equations given by the manufacturer. Nevertheless, the GPD system is highly noise-sensitive, and the DC output can be easily affected. In addition, the main limitation of this solution is that the input sine voltages should be less than ±0.7 V. Otherwise, the system cannot measure the impedance correctly, which leads to massive deviations as the case with the tested RC parallel as mentioned in
Table 1 with a mean deviation of 33.25% for the magnitude.
However, the GPD system shows acceptable results with phase measurement where the mean deviation is around 3°. The results from the IQ demodulation are more accurate than the GPD results, with around 3.2% relative deviation of the magnitude and 1.8° of relative phase deviation for the Cole–Cole model. This method also does not require a high signal processing complexity as the DC filter outputs will be used directly to calculate the real and imaginary parts of the complex impedance. On the other hand, this technique requires a highly complex analog structure due to the convolution process and filtering stage with 30 components, including six ICs. This explains the high power consumption of this technique, which is approximately 290.5 mW compared to the GPD and UI techniques, with power consumption equal to 159.2 mW and 122.4 mW, respectively.
As a perspective, the FFT-based measurement solution has the potential to support and analyze a multifrequency excitation signal compared to the IQ and GPD systems, which can ensure faster measurements for fast real-time application requirements.
5. Conclusions
BIS is a widely utilized measurement technique in a variety of research fields. This paper presents a systematical comparative study of possible BIS measurement methods under well-defined conditions. For that purpose, three bioimpedance measurement methods, namely, the gain phase detector, I/Q demodulation, and FFT-based system, have been investigated, implemented, and compared regarding amplitude accuracy, phase accuracy, hardware and computational complexity, frequency range, and power consumption. For the evaluation of proposed systems, a common signal generator module based on AD9850 and a Tietze cascade structure acting as a voltage-controlled current source have been used as excitation blocks. Then, the developed solutions were tested using the different DUTs (RC parallel and Cole–Cole models) that mimic the typical biological tissue’s behavior in the frequency range from 1 kHz to 1 MHz. The experimental investigations show that the FFT-based method achieves the most accurate results with a mean relative deviation of 1.6% for the magnitude and 0.751° for the Cole–Cole model type 1. The IQ and GPD systems achieve 5.1% and 8.1% mean of relative magnitude deviation and 1.661° and 2.605° mean of phase deviation respectively. In addition, the FFT-based system requires less hardware complexity with less number of components, which explains its low power consumption of 122.4 mW compared to the IQ and GPD solutions with 290.5 mW and 159.2 mW respectively. On the other hand, this measurement method requires the most complex signal processing to avoid spectral leakage and to carry out impedance calculations. This study can be applied to various impedance spectroscopy applications with comparable impedance and frequency ranges even though it primarily targets bioimpedance spectroscopy.