Review of the Mechanical Behavior of Different Implant–Abutment Connections
Abstract
:1. Introduction
2. Materials and Methods
2.1. Search Strategy and Study Selection
2.2. Inclusion and Exclusion Criteria
2.3. Data Extraction
3. Results
3.1. Maintenance of the Preload According to the Type of Connection
3.2. Assessment of Possible Deformations at the Different Interfaces after Repeated Application of the Tightening Torque
3.3. Evaluate The Sealing Capability Of Different Implant Connections against Microleakage
4. Discussion
4.1. Maintenance of the Preload According to the Type of Connection
4.2. Assessment of Possible Deformations at the Different Interfaces after Repeated Application of the Tightening Torque
4.3. Evaluate the Sealing Capability of Different Implant Connections against Microleakage
5. Conclusions
- Maintenance of the preload: Internal connections have a higher preload value than that of the external hexagon design. The conical configuration can spread the load along the fixture and the surrounding bone more homogeneously than both the external hexagon and traditional internal connections.
- Assessment of possible deformations at different interfaces after repeated application of tightening torque: Damage was observed in the threads of the abutment screws, before and after loading, in external and internal implant–abutment connections.
- Evaluation of the sealing capability of different implant connections against microleakage: All connections presented some microgaps and bacterial microleakage. However, the performance of the conical connection systems appeared to be superior to that of other systems.
Author Contributions
Funding
Conflicts of Interest
References
- Goiato, M.C.; Pellizzer, E.P.; da Silva, E.V.; Bonatto Lda, R.; dos Santos, D.M. Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review. Oral Maxillofac. Surg. 2015, 19, 229–242. [Google Scholar] [PubMed]
- Siamos, G.; Winkler, S.; Boberick, K.G. Relationship between implant preload and screw loosening on implant-supported prostheses. J. Oral Implantol. 2002, 28, 67–73. [Google Scholar] [PubMed] [Green Version]
- Branemark, P.I.; Hansson, B.O.; Adell, R.; Breine, U.; Lindstrom, J.; Hallen, O.; Ohman, A. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand. J. Plast. Reconstr. Surg. Suppl. 1977, 16, 1–132. [Google Scholar]
- Adell, R.; Eriksson, B.; Lekholm, U.; Branemark, P.I.; Jemt, T. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int. J. Oral Maxillofac. Implants 1990, 5, 347–359. [Google Scholar] [PubMed]
- Becker, W.; Becker, B.E. Replacement of maxillary and mandibular molars with single endosseous implant restorations: A retrospective study. J Prosthet. Dent. 1995, 74, 51–55. [Google Scholar]
- Jemt, T.; Laney, W.R.; Harris, D.; Henry, P.J.; Krogh, P.H., Jr.; Polizzi, G.; Zarb, G.A.; Herrmann, I. Osseointegrated implants for single tooth replacement: A 1-year report from amulticenter prospective study. Int. J. Oral. Maxillofac. Implants 1991, 6, 29–36. [Google Scholar]
- Pardal-Pelaez, B.; Montero, J. Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review. J. Clin. Exp. Dent. 2017, 9, e1355–e1361. [Google Scholar]
- Binon, P. The external hexagonal interface and screw joint stability: A primer on threaded fasteners in implants dentistry. Quintessence Dent. Technol. 2000, 23, 91–105. [Google Scholar]
- Cranin, N.A.; Klein, M.; Simons, A. Root form implant surgery abutments. In Atlas of Oral Implantology; Thieme Publishing Group: New York, NY, USA, 1993. [Google Scholar]
- Binon, P.P. Implants and components: Entering the new millennium. Int. J. Oral Maxillofac. Implants 2000, 15, 76–94. [Google Scholar]
- Finger, I.M.; Castellon, P.; Block, M.; Elian, N. The evolution of external and internal implant/abutment connections. Pract. Proced. Aesthet. Dent. 2003, 15, 625–632. [Google Scholar]
- Merz, B.R.; Hunenbart, S.; Belser, U.C. Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection. Int. J. Oral. Maxillofac. Implants 2000, 15, 519–526. [Google Scholar] [PubMed]
- Norton, M.R. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin. Oral. Implants Res. 1997, 8, 290–298. [Google Scholar] [PubMed]
- Sutter, R.; Weber, H.P.; Sorensen, J.; Belser, U. The new restorative concept of the ITI dental implant system: Design and engineering. Int. J. Periodontics Restorative Dent. 1993, 13, 409–431. [Google Scholar]
- Schwarz, F.; Hegewald, A.; Becker, J. Impact of implant-abutment connection and positioning of the machined collar/microgap on crestal bone level changes: A systematic review. Clin. Oral. Implants Res. 2014, 25, 417–425. [Google Scholar]
- Park, J.K.; Choi, J.U.; Jeon, Y.C.; Choi, K.S.; Jeong, C.M. Effects of abutment screw coating on implant preload. J Prosthodont. 2010, 19, 458–464. [Google Scholar]
- Albrektsson, T. A multicenter report on osseointegrated oral implants. J. Prosthet. Dent. 1988, 60, 75–84. [Google Scholar]
- Oh, T.J.; Yoon, J.; Misch, C.E.; Wang, H.L. The causes of early implant bone loss: Myth or science? J. Periodontol. 2002, 73, 322–333. [Google Scholar]
- Winkler, S.; Ring, K.; Ring, J.D.; Boberick, K.G. Implant screw mechanics and the settling effect: Overview. J. Oral Implantol. 2003, 29, 242–245. [Google Scholar]
- Bernardes, S.R.; da Gloria Chiarello de Mattos, M.; Hobkirk, J.; Ribeiro, R.F. Loss of preload in screwed implant joints as a function of time and tightening/untightening sequences. Int. J. Oral. Maxillofac. Implants 2014, 29, 89–96. [Google Scholar]
- Piermatti, J.; Yousef, H.; Luke, A.; Mahevich, R.; Weiner, S. An in vitro analysis of implant screw torque loss with external hex and internal connection implant systems. Implant Dent. 2006, 15, 427–435. [Google Scholar]
- Jaarda, M.J.; Razzoog, M.E.; Gratton, D.G. Geometric comparison of five interchangeable implant prosthetic retaining screws. J. Prosthet. Dent. 1995, 74, 373–379. [Google Scholar] [PubMed]
- Papaspyridakos, P.; Chen, C.J.; Chuang, S.K.; Weber, H.P.; Gallucci, G.O. A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients. Int. J. Oral. Maxillofac. Implants 2012, 27, 102–110. [Google Scholar]
- Kazemi, M.; Rohanian, A.; Monzavi, A.; Nazari, M.S. Evaluation of the accuracy and related factors of the mechanical torque-limiting device for dental implants. J. Dent. 2013, 10, 112–118. [Google Scholar]
- Bulaqi, H.A.; Mousavi Mashhadi, M.; Safari, H.; Samandari, M.M.; Geramipanah, F. Dynamic nature of abutment screw retightening: Finite element study of the effect of retightening on the settling effect. J. Prosthet. Dent. 2015, 113, 412–419. [Google Scholar]
- McGlumphy, E.A.; Mendel, D.A.; Holloway, J.A. Implant screw mechanics. Dent. Clin. North Am. 1998, 42, 71–89. [Google Scholar] [PubMed]
- Burguete, R.L.; Johns, R.B.; King, T.; Patterson, E.A. Tightening characteristics for screwed joints in osseointegrated dental implants. J. Prosthet Dent. 1994, 71, 592–599. [Google Scholar] [PubMed]
- Tallarico, M.; Fiorellini, J.; Nakajima, Y.; Omori, Y.; Takahisa, I.; Canullo, L. Mechanical outcomes, microleakage, and marginal accuracy at the Implant-abutment interface of original versus nonoriginal implant abutments: A Systematic REVIEW of in vitro studies. Biomed. Res. Int. 2018. [Google Scholar] [CrossRef] [Green Version]
- Gil, F.J.; Herrero-Climent, M.; Lazaro, P.; Rios, J.V. Implant-abutment connections: Influence of the design on the microgap and their fatigue and fracture behavior of dental implants. J. Mater Sci. Mater. Med. 2014, 25, 1825–1830. [Google Scholar]
- Cooper, L.F.; Tarnow, D.; Froum, S.; Moriarty, J.; De Kok, I.J. Comparison of marginal bone changes with internal conus and external hexagon design implant systems: A prospective, randomized study. Int. J. Periodont. Restor. Dent. 2016, 36, 631–642. [Google Scholar]
- Vigolo, P.; Gracis, S.; Carboncini, F.; Mutinelli, S.; Group ACR. Internal vs. external-connection single implants: A retrospective study in an Italian population treated by certified prosthodontists. Int. J. Oral Maxillofac. Implants 2016, 31, 1385–1396. [Google Scholar]
- Khraisat, A.; Hashimoto, A.; Nomura, S.; Miyakawa, O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J. Prosthet. Dent. 2004, 91, 326–334. [Google Scholar] [PubMed]
- Tsuge, T.; Hagiwara, Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent. Mater. J. 2009, 28, 373–381. [Google Scholar] [PubMed] [Green Version]
- Ricomini Filho, A.P.; Fernandes, F.S.; Straioto, F.G.; da Silva, W.J.; Del Bel Cury, A.A. Preload loss and bacterial penetration on different implant-abutment connection systems. Braz. Dent. J. 2010, 21, 123–129. [Google Scholar]
- Cashman, P.M.; Schneider, R.L.; Schneider, G.B.; Stanford, C.M.; Clancy, J.M.; Qian, F. In vitro analysis of post-fatigue reverse-torque values at the dental abutment/implant interface for a unitarian abutment design. J. Prosthodont. 2011, 20, 503–509. [Google Scholar] [PubMed]
- Ferreira, M.B.; Delben, J.A.; Barao, V.A.; Faverani, L.P.; Dos Santos, P.H.; Assuncao, W.G. Evaluation of torque maintenance of abutment and cylinder screws with Morse taper implants. J. Craniofac. Surg. 2012, 23, e631–e634. [Google Scholar]
- Murmura, G.; Di Iorio, D.; Cicchetti, A.R.; Sinjari, B.; Caputi, S. In vitro analysis of resistance to cyclic load and preload distribution of two implant/abutment screwed connections. J. Oral. Implantol. 2013, 39, 293–301. [Google Scholar]
- Jorge, J.R.; Barao, V.A.; Delben, J.A.; Assuncao, W.G. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling. Int. J. Oral Maxillofac. Implants 2013, 28, 415–422. [Google Scholar]
- Butignon, L.E.; Basilio Mde, A.; Pereira Rde, P.; Arioli Filho, J.N. Influence of three types of abutments on preload values before and after cyclic loading with structural analysis by scanning electron microscopy. Int. J. Oral Maxillofac. Implants 2013, 28, e161–e170. [Google Scholar]
- Shin, H.M.; Huh, J.B.; Yun, M.J.; Jeon, Y.C.; Chang, B.M.; Jeong, C.M. Influence of the implant-abutment connection design and diameter on the screw joint stability. J. Adv. Prosthodont. 2014, 6, 126–132. [Google Scholar]
- Xia, D.; Lin, H.; Yuan, S.; Bai, W.; Zheng, G. Dynamic fatigue performance of implant-abutment assemblies with different tightening torque values. Biomed. Mater. Eng. 2014, 24, 2143–2149. [Google Scholar]
- Al-Otaibi, H.N.; Almutairi, A.; Alfarraj, J.; Algesadi, W. The effect of torque application technique on screw preload of implant-supported prostheses. Int. J. Oral Maxillofac. Implants 2017, 32, 259–263. [Google Scholar] [PubMed]
- Tsuruta, K.; Ayukawa, Y.; Matsuzaki, T.; Kihara, M.; Koyano, K. The influence of implant-abutment connection on the screw loosening and microleakage. Int. J. Implant Dent. 2018, 4, 11. [Google Scholar] [PubMed]
- Al-Otaibi, H.N.; Al-Fouzan, A.F.; Al-Mufleh, T.S.; Labban, N. Effect of different maintenance time of torque application on detorque values of abutment screws in full-arch implant-supported fixed prostheses. Clin. Implant Dent. Relat. Res. 2018, 20, 848–851. [Google Scholar] [PubMed]
- Arshad, M.; Mahgoli, H.; Payaminia, L. Effect of repeated screw joint closing and opening cycles and cyclic loading on abutment screw removal Torque and screw thread morphology: Scanning electron microscopy evaluation. Int. J. Oral Maxillofac. Implants 2018, 33, 31–40. [Google Scholar] [PubMed]
- He, Y.; Fok, A.; Aparicio, C.; Teng, W. Contact analysis of gap formation at dental implant-abutment interface under oblique loading: A numerical-experimental study. Clin. Implant Dent. Relat. Res. 2019, 21, 741–752. [Google Scholar] [PubMed]
- Kim, K.S.; Lim, Y.J. Axial displacements and removal torque changes of five different implant-abutment connections under static vertical loading. Materials Basel 2020, 13, 699. [Google Scholar] [CrossRef] [Green Version]
- Jorneus, L.; Jemt, T.; Carlsson, L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int. J. Oral Maxillofac. Implants 1992, 7, 353–359. [Google Scholar]
- Goheen, K.L.; Vermilyea, S.G.; Vossoughi, J.; Agar, J.R. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int. J. Oral Maxillofac. Implants 1994, 9, 149–155. [Google Scholar]
- Kim, K.S.; Han, J.S.; Lim, Y.J. Settling of abutments into implants and changes in removal torque in five different implant-abutment connections. Part 1: Cyclic loading. Int. J. Oral. Maxillofac. Implants 2014, 29, 1079–1084. [Google Scholar]
- Do Nascimento, C.; Miani, P.K.; Pedrazzi, V.; Goncalves, R.B.; Ribeiro, R.F.; Faria, A.C.; Macedo, A.P.; de Albuquerque, R.F., Jr. Leakage of saliva through the implant-abutment interface: In vitro evaluation of three different implant connections under unloaded and loaded conditions. Int. J. Oral. Maxillofac. Implants 2012, 27, 551–560. [Google Scholar]
- Tripodi, D.; Vantaggiato, G.; Scarano, A.; Perrotti, V.; Piattelli, A.; Izzi, G.; D’Ercole, S. An in vitro investigation concerning the bacterial leakage at implants with internal hexagon and Morse taper implant-abutment connections. Implant Dent. 2012, 21, 335–339. [Google Scholar] [PubMed] [Green Version]
- Mishra, S.K.; Chowdhary, R.; Kumari, S. Microleakage at the different implant abutment interface: A systematic review. J Clin. Diagn. Res. 2017, 11, ZE10–ZE15. [Google Scholar] [PubMed]
- Breeding, L.C.; Dixon, D.L.; Nelson, E.W.; Tietge, J.D. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int. J. Prosthodont. 1993, 6, 435–439. [Google Scholar]
- Bozkaya, D.; Muftu, S. Mechanics of the tapered interference fit in dental implants. J. Biomech. 2003, 36, 1649–1658. [Google Scholar]
- Byrne, D.; Jacobs, S.; O’Connell, B.; Houston, F.; Claffey, N. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies. J. Prosthodont. 2006, 15, 164–171. [Google Scholar]
- Delben, J.A.; Gomes, E.A.; Barao, V.A.; Assuncao, W.G. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int. J. Oral. Maxillofac. Implants 2011, 26, 251–256. [Google Scholar] [PubMed]
- Saboury, A.; Neshandar Asli, H.; Vaziri, S. The effect of repeated torque in small diameter implants with machined and premachined abutments. Clin. Implant Dent. Relat. Res. 2012, 14 (Suppl. 1), e224–e230. [Google Scholar]
- Weiss, E.I.; Kozak, D.; Gross, M.D. Effect of repeated closures on opening torque values in seven abutment-implant systems. J. Prosthet Dent. 2000, 84, 194–199. [Google Scholar]
- Pjetursson, B.E.; Tan, K.; Lang, N.P.; Bragger, U.; Egger, M.; Zwahlen, M. A systematic review of the survival and complication rates of fixed partial dentures (FPDs] after an observation period of at least 5 years. Clin. Oral. Implants Res. 2004, 15, 625–642. [Google Scholar]
- Steinebrunner, L.; Wolfart, S.; Ludwig, K.; Kern, M. Implant-abutment interface design affects fatigue and fracture strength of implants. Clin. Oral Implants Res. 2008, 19, 1276–1284. [Google Scholar]
- Jaarda, M.J.; Razzoog, M.E.; Gratton, D.G. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent. 1994, 3, 17–21. [Google Scholar] [PubMed]
- Schmitt, C.M.; Nogueira-Filho, G.; Tenenbaum, H.C.; Lai, J.Y.; Brito, C.; Doring, H.; Nonhoff, J. Performance of conical abutment (Morse Taper] connection implants: A systematic review. J. Biomed. Mater. Res. A 2014, 102, 552–574. [Google Scholar] [PubMed]
- Coppede, A.R.; Bersani, E.; de Mattos Mda, G.; Rodrigues, R.C.; Sartori, I.A.; Ribeiro, R.F. Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: An in vitro study. Int. J. Prosthodont. 2009, 22, 283–286. [Google Scholar] [PubMed]
- Baj, A.; Bolzoni, A.; Russillo, A.; Lauritano, D.; Palmieri, A.; Cura, F.; Silvestre, F.J.; Gianni, A.B. Cone-morse implant connection system significantly reduces bacterial leakage between implant and abutment: An in vitro study. J. Biol. Regul. Homeost. Agents 2017, 31 (Suppl. 1), 203–208. [Google Scholar]
- Gherlone, E.F.; Cappare, P.; Pasciuta, R.; Grusovin, M.G.; Mancini, N.; Burioni, R. Evaluation of resistance against bacterial microleakage of a new conical implant-abutment connection versus conventional connections: An in vitro study. New Microbiol. 2016, 39, 49–56. [Google Scholar]
Author Year Type of Study | Purpose | Type of Connection Number of Implants Torque Used | Intervention | Evaluation | Outcome of the Study |
---|---|---|---|---|---|
Khraisat A et al. [32] 2004 In Vitro | Effect of lateral cyclic loading with different load positions on abutment screw loosening | EXTERNAL HEXAGON (Branemark) divided into three groups (n = 50) 32 Ncm | A-Cyclic load 50 N 1 × 106 cycles applied centrally. B- cyclic load 50 N 1 × 106 cycles applied eccentrically. C-control unloaded. | Torque gauge Micrometer microscope (1 μm) SEM | RTVs were preserved under eccentric lateral loading compared with centric loading. SEM analysis revealed mild burnishing and scuffing of the screw thread surfaces in all groups. Marked burnishing was observed at the hexagon corners on the compression sides for group B implants. |
Piermatti J et al. [21] 2006 In Vitro | Examine effects of connection design upon screw stability | EXTERNAL HEXAGON (10 BioLok; 10 Nobel Biocare) INTERNAL CONNECTION (10 Zimmer; 10 Astra Tech) (n = 40) 32 Ncm | 10 samples of each system were loaded to 200 N for 1 × 106 cycles. Screws were tightened to manufacturers’ recommendations. Torque audits were done after 250,000, 500,000, 750,000, and 1,000,000 cycles. | Torque meter | The design of the joint (i.e., internal vs. external connections) was not a significant factor in loss of torque. What did appear important was the screw design. Bio-Lok abutment screws (external connection) maintained their tightness in the best of the four systems. |
Tsuge T et al. [33] 2009 In Vitro | Evaluate the effect of eccentric cyclic loading on abutment screw loosening in internal and external hexagon implants with two screw materials | INTERNAL HEXAGON EXTERNAL HEXAGON (n = 64) 4 groups20 Ncm | Load was applied one million times (1.0 × 106 cycles] RTV measured before (initial preload) and after loading (post-loading). Changes in the superstructure and movement of the abutment were inspected using visual and tactile checks every 50,000 cycles. Damage to the surfaces of selected abutment screws was observed using SEM. | Torque gauge Fatigue testing machine SEM | In all the groups, postloading preload was significantly higher than the initial preload. Implant–abutment connection did not have an effect, but the abutment screw material did. Screws damage was probably due to screw tightening and was observed on the flank near the crest. However, no abnormal wear or damage due to micromovement or bending caused by cyclic loading was observed on the abutment screws in all the groups. |
Ricomini Filho AP et al. [34] 2010 In Vitro | Evaluate the preload loss and bacterial penetration through the implant–abutment interface of conical and external hexagon connection systems subjected to thermal cycling and mechanical fatigue. | EXTERNAL HEXAGON MORSE TAPER (n = 6) 32 Ncm 15 Ncm | The assemblies were subjected to a thermal cycling regimen (1000 cycles of 5 and 55 °C) and to mechanical fatigue (1.0 million cycles, 1.0 Hz, 120 N). The assemblies were immersed in tryptic soy + yeast extract broth containing S sanguinis and incubated at 37 °C and 10% CO2 for 72 h. Detorque values were recorded. The bacterial penetration was assessed, and the abutments were observed by SEM. | Electronic torque controller Mechanical fatigue machineSEM | All screw abutment systems showed significantly higher detorque values when subjected to TM, and all conical systems presented bacterial penetration. The results show no relationship between the preload loss and bacterial penetration. SEM micrographs show no bacterial cells on the surface of the external hexagon abutment screw, confirming the microbiological assay. |
Cashman PM et al. [35] 2011 In Vitro | Compare the abutments fatigue resistance to simulated function, using Removal torque Values as an indication of residual preload at the implant–abutment interface. | INTERNAL HEXAGON (Straumann) (n = 40) Four groups: 10 abutments from each manufacturer were evaluated for RTV with and without fatigue loading 35 Ncm | Baseline: torque was carried out twice, 10 min apart. RTV obtained 1 h later. Postfatigue: torque was carried out twice, 10 min apart. RTV postfatigue obtained 1 h later. 10 to 200 N, 15 Hz for 5 × 106 cyclesSEM was carried out at baseline and post fatigue cycling to visualize thread geometry and abutment–implant interface. | Digital torque gauge Moving-magnet linear motor to load specimens SEM | The effect of component manufacturer resulted in a significantly higher RTV in the control group, indicating greater residual preload. There was no significant decrease in postfatigue RTV for either manufacturer compared to baseline. Differences in surface finish and machining tolerances were visualized with SEM. |
Ferreira MB et al. [36] 2012 In Vitro | Evaluate the torque maintenance of retention screws of tapered abutments and cylinders of Morse taper implants submitted to retightening and detorque measurements | MORSE TAPER (n = 12) 30 Ncm | Detorque values were measured by an analogic torque gauge after 3 min of torque insertion. The detorque was measured 10 times for each retention screw of groups I and II, for a total of 120 detorque measurements in each group. | Analogic torque gauge | The abutment and cylinder screws exhibited torque loss after insertion, which indicates the need for retightening during the function of the implant-supported prostheses. |
MurmuraG et al. [37] 2013 In Vitro | Evaluation of the preload distribution in screw-retained implant systems under cyclic load. | INTERNAL HEXAGON (35 Xsign) INTERNAL OCTAGON (35 SSO) (n = 70) 25 Ncm 35 Ncm | Cyclic load between 20 and 200 N for 1 × 106 cycles. After mechanical tests, samples were sectioned along the long axis and analyzed under SEM. Five implants with internal hexagon and five implants with internal octagon were collected for photoelastic analysis. | The loss of preload was evaluated through the presence or absence of abutment mobility SEMPhotoelastic analysis | The design of the abutment connection area affects the functional load transfer to the fixture and connection screw; screw-retained abutment based on an internal octagonal connection is less likely to come loose after cyclic load. |
Jorge VA et al. [38] 2013 In Vitro | Evaluate the role of the implant/abutment system on torque maintenance of titanium retention screws and vertical misfit of screw-retained implant-supported crowns before and after mechanical cycling. | MORSE TAPER EXTERNAL HEXAGON (n = 30) 20 Ncm 30 Ncm | Retention screws received insertion torque and after 3 min, initial detorque was measured. Crowns were retightened and submitted to cyclic loading under 30 degrees of 130 ± 10 N, 2 Hz, 1 × 106 cycles. Final detorque was measured and vertical misfit using a stereomicroscope. | Stereomicroscope Torque gauge Electromechanical equipment for mastication fatigue | All groups presented a significant decrease in torque before and after mechanical cycling. The Morse taper connection promoted the highest torque maintenance. Mechanical cycling reduced the vertical misfit of all groups. |
Butignon LE et al. [39] 2013 In Vitro | Evaluate the effectiveness of 3 types of abutments in the maintenance of screw joint preload before and after cyclic loading and observe possible microdamage in the structure. | EXTERNAL HEXAGON (Neodent) (n = 45) 1. machined titanium (Ti) abutment 2. premachined gold (Au] abutments 3. machined zirconia (ZrO2) abutments 20 Ncm; 32 Ncm | Static vending test using five specimens of each group was conducted to determine load applied in the cyclic loading test. Ten specimens of each group measured the RTV. A cyclic loading (0.5 × 106 cycles, 15 Hz) between 11 and 211 N, angle 30°, was applied. Postload RTV was measured after cyclic loading. SEM was used to detect possible microdamage in the structure of components. | Torque gauge Fatigue test machine SEM | The load application reduced the preload means significantly in all groups. SEM images showed evident structural changes in the mating surface of abutments, which are related to the loss of retained preload in all groups. |
Gil FJ et al. [29] 2014 In Vitro | Evaluate microgap size and fatigue behavior of external and internal connections. | INTERNAL HEXAGON EXTERNAL HEXAGON (Klockner SK2 and Essential) (n = 100) 45 Ncm | Each specimen was sectioned along the longitudinal axis in a total of three slides for SEM.The aim was to find the level of stress at which the sample supported 5 million cycles at 1450 N (15 Hz), which will be considered the fatigue limit. | SEM Servo-hydraulic testing machine | Internal connection presented lower microgap. Very good adaptation between the implant and the screw-retained abutment, preventing infiltration of microorganisms. The fatigue behavior of the external hexagon interface showed superior results compared to internal hexagon interfaces due to better load distribution. |
Shin HM et al. [40] 2014 In Vitro | Evaluate the influence of the implant–abutment connection design and diameter on the screw joint stability | EXTERNAL HEXAGON INTERNAL CONE (n = 35) 30 Ncm | The initial and postload RTV were measured after 100,000 cycles of a 150 N and 10 Hz cyclic load. The rates of the initial and postload removal torque losses were calculated to evaluate the effect of the joint connection design and diameter on the screw joint stability. | Digital torque gauge | External butt joint was more advantageous than the internal cone in terms of the postload removal torque loss. A wide diameter was more advantageous in terms of the torque loss rate. |
Bernardes SR et al. [20] 2014 In Vitro | Determine whether abutment screw tightening and untightening influenced loss of preload in 3 different implant/abutment interfaces or on the implant body | EXTERNAL HEXAGON INTERNAL HEXAGON INTERNAL CONICAL (Neodent) (n = 25) 20 Ncm 32 Ncm | Each sample was submitted to five tightening/untightening sequences, with a pause of 5 min after tightening before the screw was loosened. This resulted in 25 tightening/untightening sequences of each design. | Custom benchtop screw-tightening machine (Prosthetic Dept, UCL, Eastman Dental Institute) | There was no immediate significant loss of preload after screw tightening. Tightening/untightening sequences did not result in any significant loss of preload. Conical implant connections demonstrated greater structural reinforcement within the internal connection. |
Xia D et al. [41] 2014 In Vitro | Evaluate the dynamic fatigue performance of implant–abutment assemblies with different tightening torque values. | INTERNAL CONNECTION (Zimmer Biomet Dental) (n = 30) Three tightening groups: 24, 30, 36 Ncm. | Five specimens of each group were unscrewed and RTV recorded. Another five specimens were subjected to load between 30 and 300 N, 15 Hz for 5 × 106 cycles. RTV postfatigue was recorded if available. Surfaces of specimens were observed with SEM. | Digital torque meter Fatigue testing machine SEM | The specimens that went through fatigue loading had decreased RTV. Insufficient torque will lead to poor fatigue performance of dental-abutment assemblies. Screws should be tightened to the torque recommended by the manufacturer. |
Al-Otabi HN et al. [42] 2017 In Vitro | Examine the effect of different torque application techniques on the removal torque of implant-supported fixed complete dental prostheses. | INTERNAL CONNECTION (Nobel Biocare-Replace) (n = 4) 35 Ncm | The torque experiment consisted of three protocols: 1. Torqueing screws to 35 Ncm once. 2. Torqueing the screws to 35 Ncm and then immediately retorquing the same screws to the same value. 3. Torqueing the same screws to 35 Ncm three consecutive times. Removal torque was recorded. | Digital torque meter | Retightening abutment screws once after the initial torquing could enhance the removal torque of the screw. Care must be taken when retorquing the screws more than once because this may inversely affect the removal torque. |
Tsuruta K et al. [43] 2018 In Vitro | Evaluate three kinds of connection systems from the point of view of microleakage from the gap between the implant and the abutment. | EXTERNAL CONNECTION INTERNAL CONICAL CONNECTION INTERNAL PARALLEL CONNECTION (Nobel Biocare)(n = 21) 35 Ncm | Dye leakage was observed from the abutment screw hole to the outside through the microgap under the excessive compressive and tensile load. Each cycle, one compressive load and one tensile load (10 N each), was applied per 1 s and 2000 cycles of loading were performed. Every 500 cycles, the amount of microleakage was statistically compared. After the completion of 2000-cycle loading, RTV of the abutment screw was measured. | Torque wrench (Nobel Biocare) Universal test machine Spectrophotometer | The amount of microleakage from implant–abutment interface was smaller in conical connection than in internal parallel connection. The increase in the amount of microleakage was observed in all three groups. Removal toque of abutment screw after the cyclic loading showed no statistically significant difference among the groups. |
Al-Otaibi HN et al. [44] 2018 In Vitro | Examine the effect of different maintenance time of torque application on detorque values of implant abutment screw. | INTERNAL HEXAGON (n = 4) 35 Ncm | The abutment screws were subjected to different maintenance time of torque application Protocol A: 35 Ncm “instant” torque application. Protocol B: 35 Ncm torque maintained for 10 s. Protocol C: 35 Ncm torque maintained for 30 s. The procedure was repeated for each protocol five times, in which new sets of screws were used, with a total of 60 new screws. | Digital torque meter | The application of 35 Ncm for different maintenance times of torque application did not appear to affect the detorque value. Maintaining the torque for a prolonged time (10 s or 30 s) was not significantly associated with a higher preload than instant torque application. |
Arshad M et al. [45] 2018 In Vitro | Evaluate the effect of repeated screw joint closing and opening cycles and cycling loading on abutment screw removal torque and screw thread morphology. | INTERNAL CONICAL CONNECTION (Dentium) (n = 30) Three groups 12 and 30 Ncm | Abutments screw were tightened (12 Ncm), removed, and RTV-recorded. This sequence was repeated five times for G1 and 15 times for G2 and G3. The same screws for G1 and G2 and new screws for G3 were tightened to 12 Ncm; this was also followed by screw tightening to 30 Ncm and retightening to 30 Ncm, 15 min later. RTVs were taken after cyclic loading 0.5 × 106, 1 Hz 75 N. Surface topography of one screw in each group was evaluated with SEM and compared with an unused screw. | Electronic torque meter Chewing simulator SD mechatronic SEM | Using a new screw did not significantly increase the value of removal torque. Restricting the amount of screw tightening was more important than replacing the screw. Torque loss values after loading were not shown to be significantly different from each other. |
He Y et al. [46] 2019 In Vitro | Investigate the formation of microgaps and the change in the contact area at the implant–abutment interface of two different connection designs under oblique cyclic loading. | CONICAL CONNECTION EXTERNAL HEXAGON CONNECTION (n = 10)20 Ncm | After loading, the samples were scanned using micro-CT, with silver nitrate as a high contrast penetrant, and the level of leakage was assessed.Three-dimensional finite element (FE) analyses were conducted to reveal the microgap formation process. | Micro CT with silver nitrate Fatigue machine | The conical connection showed more resistance against the formation of microgaps at the implant–abutment interface than the external hexagonal connection, although the minimum load required to bridge the internal implant space was within the range of human biting force. |
Kim KS et al. [47] 2020 In Vitro | Examine the settling of abutments into implants and the removal torque value under static loading. | EXTERNAL CONNECTION INTERNAL HEXAGON INTERNAL OCTAGON (n= 50) 30 Ncm | Ten implant–abutment assemblies were loaded vertically downward with a 700 N load cell at a displacement rate of 1 mm/min. The settling of the abutment was obtained from the change in the total length of the entire implant–abutment unit using an electronic digital micrometer. The postloading RTV was compared to the initial torque value. | Universal testing machine electronic digital micrometer Digital torque gauge | The loss of the preload due to the settling effect can lead to screw loosening during a clinical procedure in the molar region, where masticatory force is relatively greater. |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Vinhas, A.S.; Aroso, C.; Salazar, F.; López-Jarana, P.; Ríos-Santos, J.V.; Herrero-Climent, M. Review of the Mechanical Behavior of Different Implant–Abutment Connections. Int. J. Environ. Res. Public Health 2020, 17, 8685. https://doi.org/10.3390/ijerph17228685
Vinhas AS, Aroso C, Salazar F, López-Jarana P, Ríos-Santos JV, Herrero-Climent M. Review of the Mechanical Behavior of Different Implant–Abutment Connections. International Journal of Environmental Research and Public Health. 2020; 17(22):8685. https://doi.org/10.3390/ijerph17228685
Chicago/Turabian StyleVinhas, Ana Sofia, Carlos Aroso, Filomena Salazar, Paula López-Jarana, José Vicente Ríos-Santos, and Mariano Herrero-Climent. 2020. "Review of the Mechanical Behavior of Different Implant–Abutment Connections" International Journal of Environmental Research and Public Health 17, no. 22: 8685. https://doi.org/10.3390/ijerph17228685
APA StyleVinhas, A. S., Aroso, C., Salazar, F., López-Jarana, P., Ríos-Santos, J. V., & Herrero-Climent, M. (2020). Review of the Mechanical Behavior of Different Implant–Abutment Connections. International Journal of Environmental Research and Public Health, 17(22), 8685. https://doi.org/10.3390/ijerph17228685