3.1. Variations of Time Constant with Respect to Air Cavity in Reference Fluid Syringe
According to previous studies [
22,
23,
26,
29], air compliance was widely used to stabilize fluidic instability resulting from syringe pumps. In this study, a reference fluid was injected at a constant flow rate with a syringe pump. However, the air cavity in the reference fluid syringe might have an influence on the dynamic variation of the interface in co-flowing channels. For this reason, it was necessary to evaluate the contributions of the air cavity in the reference fluid syringe to time constants of the interface in co-flowing channels. The air cavity in the reference fluid syringe was set to
Vair, R = 0, 0.1, and 0.2 mL. To separate the effect of the air cavity in the test fluid syringe, such a cavity was set to zero (
Vair, T = 0). Blood samples (normal RBCs suspended in 1× PBS, Hct = 50%) and glycerin (20%) were prepared as test fluids.
To model the contribution of the air cavity in the syringe to interface, it was required to derive a governing equation for two fluids flowing in a co-flowing channel. As shown in
Figure 2A, a fluidic circuit model for two fluids (reference fluid and test fluid) flowing in a co-flowing channel was constructed with discrete circuit elements (i.e., flow rate elements:
QR,
QT, resistance elements:
RR,
RT, and compliance element:
CT). Here,
CT denotes the compliance element that was combined with flexible tubing, a microfluidic channel, and an air cavity in the syringe. Additionally, ground (▼) represents pressure set to zero. To keep the mathematical model simple, the interface in the co-flowing channel was modeled as a virtual wall. Different boundary conditions between the real physical model and the mathematical model were compensated by adding a correction factor (
Cf) into the governing equation [
39]. Thus, both fluids in the co-flowing channel were modeled independently with discrete circuit elements. The governing equation on interface (
α) for both fluids flowing in the co-flowing channel is expressed as follows:
Here,
is given by
, where
denotes the channel length of the co-flowing channel. Subscript T means test fluid. Instead of subscript T, subscript B is also used for representing blood sample. According to a numerical simulation using CFD-ACE+ (Ver. 2019, ESI Group, Paris, France), the correction factor could be approximately expressed as
Cf = 6.6908
α4 – 13.382
α3 + 10.81 α
2 – 4.1196 α + 1.6206 (
R2 = 0.9922, 0.1 <
α < 0.9) (
Figure A1,
Appendix A). Because of the nonlinear terms in the left member of Equation (1), the differential equation was difficult to solve substantially. Based on an approximate procedure [
39], two approximate coefficients (
F1,
F2) were obtained as
F1 = 1.112,
F2 = 1.129, respectively. Consequently, 1/(1-α) was converted into
β and Equation (1) was transformed into a linear differential equation as follows:
In Equation (2), the time constant (λ) is expressed as . The compliance element (CT) presents a linear relation with the time constant (λ) and includes the effect of the air cavity in the syringe. Thus, the contribution of the air cavity could be obtained quantitatively by measuring the time constant (λ) with transient behaviors of β.
As shown in
Figure 2B-a, temporal variations of
α and
β were obtained with respect to the blood sample (normal RBCs suspended in 1× PBS, Hct = 50%). Here, the air cavity in the reference fluid syringe was set to 0.2 mL (
Vair, R = 0.2 mL). Based on Equation (2), the temporal variations of
β were represented as shown in
Figure 2B-b. When sequentially turning syringe pumps on and off, two time constants (
λoff,
λon) could be obtained by analyzing transient variations of
β. First, under the turn-off operation of a syringe pump, temporal variations of
βoff were extracted for 60 s. Based on an exponential model (i.e.,
βoff =
β0 +
β1 exp (-
t/
λoff)),
λoff was obtained by conducting nonlinear regression analysis with Matlab 2019. Second, under the turn-on operation of a syringe pump,
βon converged in a shorter time interval than for
βoff. Temporal variations of
βon were extracted for 20 s. Similarly, based on an exponential model (i.e.,
βon =
β0 +
β1 exp (-
t/
λon)),
λon was obtained by conducting non-linear regression analysis.
Figure 2C-a shows variations of
λoff and
λon with respect to
Vair, R and glycerin (20%). All experimental data were expressed as mean ± standard deviation. The error bar represented single standard deviation. Note that
λoff was much longer than
λon within 0.2 mL of the air cavity. Additionally,
λoff decreased substantially when the cavity volume increased from 0 to 0.1 mL. Above
Vair, R = 0.1 mL, it decreased slightly.
Figure 2C-b shows variations of
λoff and
λon with respect to
Vair, R and the blood sample. Similar to the glycerin solution,
λoff decreased considerably when the air cavity increased from 0 to 0.1 mL. The air cavity in the reference fluid syringe (~0.1 mL) contributed to decreasing the time constant (
λoff) significantly. However,
λon did not present distinctive variations with respect to the air cavity in the reference fluid syringe. Additionally, two time constants remained unchanged above
Vair, R = 0.1 mL. According to discrete fluidic circuit analysis, air compliance (
C) plays a role in regulating the alternating component of the flow rate. In this study, flow rate of the reference fluid remained unchanged over time. It was modeled as direct component of flow rate. Thus, air cavity secured in reference syringe did not contribute to the changing time constant. However, air cavity with 0.1 mL decreased time constant substantially. Taking into account the fact that air compliance caused the time constant to increase generally, the result showed different trends. Above a 0.1 mL air cavity, the time constant varied slightly. The constant value of the time constant was obtained through fluid viscosity and the compliance effect of the tubing and PDMS device. According to these experimental results, the air cavity in the reference fluid syringe (~0.1 mL) contributed to decreasing
λoff greatly. Note that
λoff decreased more significantly than
λon. Above an air cavity volume of 0.1 mL, the time constants did not present substantial variation.
3.2. Valuations of Time Constant with Respect to Hematocrit and Air Cavity in Blood Sample Syringe
First, to evaluate the contribution of hematocrit to the time constant, the blood sample (Hct = 30%, 40%, and 50%) was prepared by adding normal RBCs into 1× PBS. As shown in
Figure 3A-a, variations of
λoff and
λon were obtained with respect to Hct. To evaluate the contribution of the air cavity in the reference fluid syringe, such a cavity was set to
Vair, R = 0 and 0.1 mL. The air cavity in the blood sample syringe was set to
Vair, B = 0. In contrast with
λon,
λoff increased largely with respect to Hct. In addition, the air cavity in the reference fluid syringe contributed to decreasing the time constant substantially. As shown in
Figure 3A-b, a scatter plot was constructed by plotting
λon on
Y-axis and
λoff on
X-axis. According to linear regression analysis, the following linear regression formula was obtained:
λon = 0.2815
λoff + 1.4967 (
R2 = 0.8282). The high regression coefficient (
R2) denotes that
λon and
λoff showed a strong linear relationship. From these results,
λoff was selected as the representative time constant throughout this study.
Second, to evaluate the effect of the air cavity in the blood sample syringe (
Vair, B) on the time constant (
λoff), such cavity was set to
Vair, B = 0, and 0.1 mL. Additionally, to stabilize the fluidic instability resulting from the syringe pump, the air cavity in the reference fluid syringe was set to
Vair, R = 0.1 mL. As shown in
Figure 3B, variations of
λoff were obtained with respect to Hct = 30%, 40%, and 50% and
Vair, B = 0, and 0.1 mL. The air cavity in the blood sample syringe contributed to increasing the time constant substantially. Theoretically, the size of syringe pump did not contribute to the varying time constant. According to the previous study [
22], the time constant tended to increase linearly with respect to air cavity volume. In other words, air cavity secured in each syringe varied dynamic behaviors of
β in coflowing channels (i.e., time constant). Thus, it is necessary to fix air cavity secured in each syringe. Note that, interestingly, the air cavity in the reference fluid syringe contributed to decreasing
λoff, as shown in
Figure 3A-a. From these results, we inferred that the air cavity increased or decreased the time constant depending on whether it existed in the reference fluid syringe or the blood sample syringe.
Third, to compare the time constant with temporal variations of
β, the time constant was additionally obtained with temporal variations of the average velocity of the blood flow in the test fluid channel (
<U>μPIV). A blood sample (Hct = 50%) was prepared as the test fluid by adding normal RBCs into 1× PBS.
Figure 3C-a shows
λoff of
<U>μPIV and
λoff of
β with respect to
Vair, R = 0 and 0.1 mL. Here, the air cavity in the blood sample syringe was set to zero. Consequently,
λoff tended to decrease with respect to
Vair, R. Both
<U>μPIV and
β exhibited a similar trend of
λoff with respect to
Vair, R. Figure 3C-b shows a comparison of
λoff obtained from
<U>μPIV and
β with respect to
Vair, B = 0 and 0.1 mL. Here, the air cavity in the reference fluid syringe was set to 0.1 mL. Consequently,
λoff tended to increase with respect to
Vair, B. Both
<U>μPIV and
β exhibited increase in
λoff significantly with respect to
Vair,B. The time constant obtained with
β presented a very similar trend with respect to the air cavity compared with the time constant obtained with
<U>μPIV. As quantification of
<U>μPIV required an expensive high-speed camera and much time for the micro-PIV procedure, the quantification of
β could be considered more effective.
Finally, to evaluate the contribution of the air cavity in the blood sample syringe to blood viscosity (
μB), the value of
μB was obtained with respect to
Vair, B = 0 and 0.1 mL. The blood viscosity was quantified under constant flow rate; both fluids were infused at the same flow rate (i.e.,
QB =
QR). By setting
in Equation (2), a formula of blood viscosity was derived as follows:
As shown in
Figure 4A,
μB tended to increase with respect to Hct. As expected, the air cavity in the blood sample syringe did not contribute to varying blood viscosity. In addition, it was inferred that the air cavity in the reference fluid syringe (~0.1 mL) was sufficient to maintain a constant flow rate, even at
Vair, B = 0.
To compare with the blood viscosity obtained with the present method (i.e., co-flowing method), the blood viscosity of the same blood sample was also obtained with a previous method (i.e., flow-switching method) [
40]. The previous method produced a higher value of blood viscosity than the present method. The Fåhræus–Lindqvist effect indicated that blood viscosity varied with respect to channel diameter. In other words, blood viscosity tended to decrease at a smaller channel due to the existence of a cell-free layer. However, blood viscosity remained constant for wider channel with above 300~500 μm. Here, the contribution of a cell-free layer was negligible because it was much smaller than the channel size. As a rectangular channel (width =
W, and depth =
h) was filled with a blood sample, an equivalent circular diameter (
d) was estimated as
with mass conservation. For the previous method (i.e., switching flow method), a single fluidic channel was filled with blood sample completely when reversal flow in junction occurred. Then, equivalent diameter was estimated as
d = 358 μm. However, for the present method (i.e., co-flowing method), the corresponding interface of each hematocrit was obtained as
α = 0.65 ± 0.01 for Hct = 30%, α = 0.67 ± 0.01 for Hct = 40%, and α = 0.68 ± 0.01 for Hct = 50%. The equivalent diameter was then estimated as
d = 288~294 μm. According to the previous study [
41], for channel diameter with below
d = 400 μm, blood viscosity tended to decrease gradually with respect to equivalent diameter. Because the present method had smaller equivalent diameter than the previous method, it was reasonable that blood viscosity obtained by the present method was underestimated substantially when compared with blood viscosity obtained by the previous method. To obtain a linear relationship between both methods, as shown in the inset of
Figure 4B, a scatter plot was constructed by plotting the viscosity obtained by the present method (i.e., co-flowing method with
μB, CFM) on
Y-axis and the viscosity obtained by previous method (i.e., flow-switching method:
μB, FSM) on
X-axis. According to regression analysis, a linear regression formula was obtained:
μB, CFM = 0.369
μB, FSM + 1.2049 (
R2 = 0.9485). The high value of the regression coefficient (
R2) means that the co-flowing method (i.e., the present method) could be used effectively to monitor blood viscosity compared with the flow-switching method (i.e., the previous method).
3.3. Quantitative Evaluations of Image Intensity, Blood Velocity, and Interface with Respect to Diluent
To evaluate variations of mechanical properties of a blood sample at constant blood flow rate, a blood sample (Hct = 50%) was prepared by adding normal RBCs into two different diluents, namely 1× PBS and dextran solution (10 mg/mL). Here, the dextran solution was used as a diluent to enhance the RBC aggregation in the blood sample. The contribution of the dextran solution to the mechanical properties of the blood sample was evaluated by measuring image intensity (<I>), average velocity (<U>μPIV), and interface (α =1 - β−1) with respect to the blood flow rate (or shear rate). Using two syringe pumps, both fluids were injected at the same flow rate (QR = QB = Qsp). The air cavity in each syringe was set to 0.1 mL (i.e., Vair, R = Vair, B = 0.1 mL).
As shown in
Figure 5A, the variation of image intensity (
<I>) was obtained with respect to
Qsp and the diluent. The right side panel in the figure shows microscopic images captured at specific flow rates (
Qsp): (a)
Qsp = 0.075 mL/h, (b)
Qsp = 0.2 mL/h, (c)
Qsp = 0.6 mL/h, (d)
Qsp = 1 mL/h, and (e)
Qsp = 5 mL/h. For the dextran solution as diluent,
<I> decreased gradually up to
Qsp = 0.4 mL/h. RBC aggregation caused to increase
<I> at a lower flow rate. However, when the flow rate increased, RBCs tended to disaggregate. Above
Qsp = 0.6 mL/,
<I> tended to increase gradually with respect to
Qsp. According to a previous study, the orientation and deformability of RBCs contribute to increasing image intensity [
42]. Given that RBCs in 1× PBS did not include RBC aggregation,
<I> did not increase, even at lower flow rates. The value of
<I> tended to increase gradually by increasing the flow rate.
While measuring blood viscosity accurately, it is necessary to evaluate the effect of flow rate on interface (
α =
WB/W) in the coflowing channel. As shown in
Figure 1A-c, blood-filled width (
WB) could be obtained accurately by conducting image processing. However, the channel width was assumed as
W = 1000 μm. Maximum flow rate was estimated as 2 mL/h when test fluid and reference fluid were set to the same flow rate of 1 mL/h. While infusing the blood sample into single microfluidic channel, the deformed channel width was quantified by increasing flow rate. Variation of
W was obtained by varying flow rate (
QB = 0.05, 0.1, 0.2, 0.4, 0.6, 0.8, 1, 2, 3, 4, and 5 mL/h). As shown in
Figure A2 (
Appendix A), the channel width of the corresponding flow rate was quantified as
W = 1009 ± 0.2 μm (
QB = 1 mL/h),
W = 1012.8 ± 2.1 μm (
QB = 2 mL/h), and
W = 1017.5 ± 1.7 μm (
QB = 4 mL/h). From the results, variation of channel width was estimated as less than 2% under the maximum flow rate of 2 mL/h.
Variations of
<U>μPIV with respect to
Qsp and diluent were obtained, as shown in
Figure 5B. The value of
<U>μPIV tended to increase linearly with respect to
Qsp. According to linear regression analysis, a linear regression formula for each diluent was obtained:
<U>μPIV = 2.6287
Qsp (
R2 = 0.9971) for dextran solution (10 mg/mL) and
<U>μPIV = 2.0732
Qsp (
R2 = 0.9956) for 1× PBS. These results indicated that RBCs suspended in dextran solution reached a higher value of
<U>μPIV (~26.8%) compared with RBCs suspended in 1× PBS.
Finally, to evaluate variations of interface (α) with flow rate and diluent, variations of
α and
μB were obtained with respect to shear rate and diluent. For a rectangular channel (width =
W, and depth =
h) with low aspect ratio [
8], a shear rate for each flow rate (
Qsp) is given approximately by
. Using Equation (3), the blood viscosity of the blood sample was obtained in terms of the shear rate. As shown in
Figure 5C, the interface (
α) of RBCs suspended in the dextran solution reached a higher value of interface compared with RBCs suspended in 1× PBS. The blood viscosity decreased gradually with respect to the shear rate. The blood sample behaved as a non-Newtonian fluid (or shear-thinning fluid). Furthermore, a dextran solution (10 mg/mL) as diluent contributed to increasing the blood viscosity significantly compared with 1× PBS. When compared with previous results [
8], our results showed consistent trends with respect to diluents.
3.4. Variations of Red Blood Cells (RBC) Aggregation, Viscosity, and Viscoelasticity with Respect to Diluent and Air Cavity in Syringe
To quantify three mechanical properties of blood sample (RBC aggregation, viscosity, and viscoelasticity) with respect to air cavity (or air compliance), variations of
<I>,
<U>μPIV, and
β were simultaneously obtained with respect to diluent and air cavity in each syringe. A blood sample (Hct = 50%) was prepared by adding normal RBCs into four different diluents, namely 1× PBS, two dextran solutions (5, and 10 mg/mL), and plasma. The air cavity in the reference fluid syringe was fixed at
Vair, R = 0.1 mL. Additionally, the air cavity in the blood sample syringe varied from
Vair, B = 0 to
Vair, B = 0.1 mL. Based on experimental results shown in
Figure 5A, the flow rate of each fluid was reset to
Q0 = 0.5 mL/h for measuring RBC aggregation effectively.
First, as shown in
Figure 6A, temporal variations of
<I> and
<U>μPIV were obtained with respect to diluents. Here, the air cavity in the blood sample syringe was set to
Vair, R = 0. When the syringe pump was turned off periodically,
<U>μPIV decreased suddenly over time. RBC aggregation increased
<I> gradually over time. Given that 1× PBS did not stimulate RBC aggregation,
<I> of 1× PBS remain unchanged over time. However, two dextran solutions contributed to increasing
<I> over time substantially. Given that plasma proteins contributed to RBC aggregation [
43,
44],
<I> of plasma increased gradually over time.
To evaluate the effect of air compliance on RBC aggregation, the air cavity in the blood sample syringe was varied from
Vair, B = 0 to
Vair, B = 0.1 mL. As shown in
Figure 6B, temporal variations of
<I> and
<U>μPIV were obtained with respect to diluent. Even when turning off the syringe pump,
<U>μPIV tended to decrease gradually over time. For this reason, except for a higher concentration of dextran solution (10 mg/mL),
<I> did not show substantial increase over time. From these results, the air cavity (~0.1 mL) in the blood sample syringe delayed the transient behaviors of blood velocity considerably. Thus, it was inferred that air compliance hindered the quantification of RBC aggregation. To quantify RBC aggregation with
<I>, it was necessary to define an RBC aggregation index. From
Figure 6A, temporal variations of
<I> and
<U>μPIV were redrawn from
t = 0 to
t = 360 s. As shown in
Figure 6C, after
t = 120 s (i.e., turn-off operation of the syringe pump),
<U>μPIV decreased largely over time. Note also that
<I> tended to increase gradually over time. Here, a specific time instant and minimum value of
<I> were denoted as
t =
t0 and <
I (t = t0) >, respectively. The RBC aggregation index was then obtained by analyzing
<I> from
t =
t0 to
t =
t0 +
ts. According to a previous study [
45], an RBC aggregation index (
AIRBC) can be defined as follows:
Based on the temporal variations of
<I> shown in
Figure 6A,B,
AIRBC was quantified at periodic intervals (
T = 240 s).
Figure 6D shows variations of
AIRBC with respect to diluent and
Vair, B. The inset of
Figure 6D shows temporal variations of
AIRBC with respect to dextran solution (10 mg/mL) and
Vair, B = 0. The RBC aggregation index was quantified with repetitive tests (n = 8) and expressed as mean ± standard deviation. Under no air cavity in the blood sample syringe, the RBC aggregation index for each diluent was obtained as
AIRBC = 0.003 ± 0.001 for 1× PBS,
AIRBC = 0.025 ± 0.001 for dextran solution (5 mg/mL),
AIRBC = 0.071 ± 0.008 for dextran solution (10 mg/mL), and
AIRBC = 0.021 ± 0.003 for plasma. The dextran solutions and plasma contributed to increasing the RBC aggregation index significantly compared with 1× PBS. Additionally,
AIRBC tended to increase significantly at higher concentration of dextran solution. When the air cavity in the blood sample syringe was reset to 0.1 mL, the RBC aggregation index for the two dextran solutions was obtained as
AIRBC = 0.004 ± 0.001 for the first dextran solution (5 mg/mL) and
AIRBC = 0.008 ± 0.003 for the second dextran solution (10 mg/mL). Given that the air cavity (or air compliance) tended to delay the transient behavior of the blood velocity,
AIRBC decreased considerably. From these results, the air cavity (~0.1 mL) in the blood sample syringe hindered the quantification of the RBC aggregation substantially.
Second, as shown in
Figure 7A-a, the temporal variations of
β were obtained with respect to diluent. Air cavities of each syringe were set to
Vair, R = 0.1 mL and
Vair, B = 0, respectively. Among the values of
β obtained in
Figure 7A-a, to represent how the blood viscosity (
μB) and the time constant (
λoff) were quantified over a single period, temporal variations of
β were redrawn at specific durations ranging from
t = 240 s to
t = 500 s.
As shown in
Figure 7A-b, the value of
μB was obtained with the Equation (3) under turn-on operation of the syringe pump (
t < 0.5
T). Afterward,
λoff was estimated with regression analysis (
βoff =
β0 +
β1 exp (-
t/
λoff) under turn-off operation of the syringe pump (0.5
T <
t <
T). As shown in
Figure 7A-c, variations of
μB were obtained at intervals of 240 s with respect to diluent. The value of
μB for each diluent was obtained as
μB = 2.95 ± 0.12 cP for 1× PBS,
μB = 3.53 ± 0.15 cP for the first dextran solution (5 mg/mL),
μB = 5.89 ± 0.28 cP for the second dextran solution (10 mg/mL), and
μB = 4.59 ± 0.14 cP for plasma. Under the turn-off operation of the syringe pump, the time constant (
λoff) was obtained with respect to diluent. Using a linear Maxwell model (i.e.,
λoff =
μB/
GB),
GB was obtained by dividing
μB by
λoff. As shown in
Figure 7B-b, variations of
λoff were represented with respect to diluent and
Vair, B = 0. Additionally,
Figure 7A-d shows variations of elasticity (
GB) with respect to diluent. The value of
GB for each diluent was obtained as
GB = 0.5 ± 0.02 mPa for 1× PBS,
GB = 0.62 ± 0.03 mPa for the first dextran solution (5 mg/mL),
GB = 0.79 ± 0.03 mPa for the second dextran solution (10 mg/mL), and
GB = 0.76 ± 0.05 mPa for plasma. From these results, blood viscoelasticity (viscosity, elasticity) was quantified consistently with respect to diluent under the air cavity in each fluid syringe (
Vair, R = 0.1 mL, and
Vair, B = 0). The dextran solution as diluent contributed to increasing viscosity and elasticity substantially compared with 1× PBS. The plasma reached a higher value of viscosity and elasticity compared with 1× PBS. In other words, the plasma proteins led to increased viscosity and elasticity.
To quantify the effect of the air cavity in the blood sample syringe on
β, the volume of such air cavities was varied from 0 to 0.1 mL. Additionally, the volume of the air cavity in the reference fluid syringe was set to 0.1 mL. As shown in
Figure 7B-a, temporal variations of
β were obtained with respect to diluent. The air cavity in the blood sample syringe delayed the transient behavior of
β substantially. During turn-on and turn-off operation of the syringe pump,
β did not reach a constant value within a specific duration. Given that Equation (3) as blood viscosity was effective for blood viscosity only for constant values of
β, it was impossible to obtain the blood viscosity with no information on flow rate (or velocity) at a specific time instant.
Figure 7B-b shows variations of
λoff with respect to diluent and
Vair, B. Note that the case
Vair, B = 0.1 mL contributed to increasing
λoff significantly compared with
Vair, B = 0. When setting
Vair, B = 0.1 mL,
λoff was increased largely at a higher concentration of dextran solution.
As shown in
Figure 7B-b, while increasing air cavity was secured in the blood syringe from 0 to 0.1 mL, the corresponding time constant of each diluent increased about
Δλoff = 13.3 s (1× PBS),
Δλoff = 18.4 s (dextran sol. 5 mg/mL),
Δλoff = 27.8 s (dextran sol. 10 mg/mL), and
Δλoff = 10 s (plasma). As transient time increased largely within a half period,
β did not arrive to constant value under periodic on-off operation of syringe pump. As a solution, it is necessary to increase period of blood flow rate. Taking into account the fact that time constant increased about 10~27.8 s for each diluent, half period of blood flow rate should increase at least 27.8 s. When the period of blood flow rate changes from
T = 240 s to
T = 300 s, it will be inferred that
β exhibits similar trends as shown in
Figure 7A-a. Thus, blood viscosity and RBC aggregation will be obtained without additional information on temporal variations of blood velocity.
To quantify blood viscosity under varying blood flows, it was necessary to obtain
β and
<U>μPIV over time.
Figure 8A shows temporal variations of
β and
<U>μPIV with respect to two diluents, namely 1× PBS and dextran solution (10 mg/mL). As shown in
Figure 5B, the relationship between
<U>μPIV and
Qsp was obtained in advance as a linear regression formula with respect to each diluent, and the average velocity of the blood sample (
<U>μPIV) was converted into the blood flow rate with a regular formula.
Given that the flow rate of the blood sample varied over time, the formula of blood viscosity was corrected as
μB =
μR × (
β − 1) ×
F2 × (
QR/
QB) by adding a flow rate term into Equation (3).
Figure 8B shows temporal variations of
and
μB with respect to diluent during a single period of 240 s. Note that
μB presented large scattering at
Qsp < 0.1 mL/h. Thus, the minimum value of
Qsp was set to 0.1 mL/h. As shown in
Figure 8C, variations of
μB were obtained with respect to
during the turn-on and turn-off operation of the syringe pump. For 1× PBS as diluent,
μB remained constant with respect to the shear rate. However, for the dextran solution as diluent,
μB decreased gradually with respect to the shear rate. As shown in
Figure 8D-a, variations of
μB and
GB were obtained with respect to the syringe operation (i.e., turn-on, turn-off). Given that
μB remained unchanged over the shear rate,
GB was calculated by dividing
μB by the time constant (i.e.,
GB =
μB/
λoff for turn-off operation, and
GB =
μB/
λon for turn-on operation). The value of
μB for each operation was obtained as
μB =3.38 ± 0.19 cP for the turn-off operation, and
μB = 3.52 ± 0.12 cP for the turn-on operation. Additionally, the value of
GB for each operation was obtained as
GB = 0.18 ± 0.02 cP for the turn-off operation, and
GB = 0.21 ± 0.05 cP for the turn-on operation. From these results,
μB and
GB remained unchanged irrespective of the syringe operation. Compared with the results obtained at
Vair, B = 0 as showed in
Figure 7A-c, a 0.1-mL air cavity (~0.1 mL) in the blood sample syringe caused to overestimate
μB. However, it caused
GB to be underestimated.
Figure 8D-b shows variations of
and
n with respect to the first dextran solution (10 mg/mL) and syringe operation. The constant
decreased significantly by switching the syringe pump from turn-off operation to turn-on operation. In addition, the turn-on operation caused to increase the index
n substantially compared with the turn-off operation.
In this study, three mechanical properties of blood sample (viscosity, RBC aggregation, and time constant) were quantified with methods suggested in previous studies. First, cone-and-plate viscometer as conventional method has been used to measure blood viscosity. According to the quantitative comparison between conventional viscometer and microfluidic viscometer [
40,
41,
46,
47], the previous studies indicated that blood viscosity could be measured consistently in a microfluidic environment. Based on the previous study, as shown in
Figure 4A, co-flow method (present method) and flow switching method (previous method) were used to obtain blood viscosity with respect to hematocrit. The present method underestimated blood viscosity when compared with the previous method. However, as shown in
Figure 4B, both methods exhibited a high degree of linear relationship (i.e.,
R2 = 0.9485). Second, RBC aggregation as a conventional method has been quantified by analyzing light intensity [
48] or electric impedance [
49] of blood sample flowing in slit channel. According to quantitative comparison study [
50,
51], microscopic image intensity exhibited variations of RBC aggregation in microfluidic channel sufficiently. Thus, without quantitative comparison study, variations of RBC aggregation were quantified by analyzing image intensity of blood flows in the test channel under turn-off blood flows. Finally, under transient flow conditions, time constant has been obtained by analyzing temporal variations of physical parameters including blood velocity, flow rate, and pressure. Based on Equation (2), time constant (
l) was obtained by analyzing temporal variations of
β. Furthermore, to compare with time constant obtained from information of
β, time constant was quantified additionally by analyzing temporal variations of blood velocity (
<U>). As shown in
Figure 3C, both time constants exhibited consistent variations with respect to air cavity in reference syringe.
From these experimental results, it leads to the conclusion that the air cavity in the blood sample syringe made the RBC aggregation and blood viscoelasticity vary substantially. The RBC aggregation index decreased largely, even for a 0.1-mL air cavity in the blood sample syringe because of a longer transient behavior of blood flows. Thus, to measure RBC aggregation and viscoelasticity of blood samples consistently, a 0.1-mL air cavity must be secured in the reference fluid syringe as a minimum condition. Additionally, the air cavity in the blood sample syringe must be minimized as much as possible.