Next Article in Journal
Mixing Performance of a Serpentine Micromixer with Non-Aligned Inputs
Next Article in Special Issue
On the Interaction between a Nanoparticulate System and the Human Body in Body Area Nanonetworks
Previous Article in Journal
Effect of Laser-Induced Heating on Raman Measurement within a Silicon Microfluidic Channel
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

An Integrated Glucose Sensor with an All-Solid-State Sodium Ion-Selective Electrode for a Minimally Invasive Glucose Monitoring System

Central Research Laboratories, Sysmex Corporation, 4-4-4 Takatsukadai, Nishi-ku, Kobe, Hyogo 651-2271, Japan
*
Author to whom correspondence should be addressed.
Micromachines 2015, 6(7), 831-841; https://doi.org/10.3390/mi6070831
Submission received: 2 May 2015 / Revised: 16 June 2015 / Accepted: 25 June 2015 / Published: 30 June 2015

Abstract

:
We developed a minimally invasive glucose monitoring system that uses a microneedle to permeate the skin surface and a small hydrogel to accumulate interstitial fluid glucose. The measurement of glucose and sodium ion levels in the hydrogel is required for estimating glucose levels in blood; therefore, we developed a small, enzyme-fixed glucose sensor with a high-selectivity, all-solid-state, sodium ion-selective electrode (ISE) integrated into its design. The glucose sensor immobilized glucose oxidase showed a good correlation between the glucose levels in the hydrogels and the reference glucose levels (r > 0.99), and exhibited a good precision (coefficient of variation = 2.9%, 0.6 mg/dL). In the design of the sodium ISEs, we used the insertion material Na0.33MnO2 as the inner contact layer and DD16C5 exhibiting high Na+/K+ selectivity as the ionophore. The developed sodium ISE exhibited high selectivity ( log k N a , K p o t = −2.8) and good potential stability. The sodium ISE could measure 0.4 mM (10−3.4 M) sodium ion levels in the hydrogels containing 268 mM (10−0.57 M) KCl. The small integrated sensor (ϕ < 10 mm) detected glucose and sodium ions in hydrogels simultaneously within 1 min, and it exhibited sufficient performance for use as a minimally invasive glucose monitoring system.

Graphical Abstract

1. Introduction

The incidence of type 2 diabetes mellitus is rapidly increasing [1]. Conventionally, the levels of HbA1c and fasting plasma glucose are used as proxies for detection of early-stage diabetes in routine health check-ups; however, neither method is sufficiently sensitive. The Diabetes Epidemiology Collaborative Analysis of Diagnostic Criteria in Europe and Diabetes Epidemiology Collaborative Analysis of Diagnostic Criteria in Asia studies have shown that the presence of hyperglycemia after a glucose load is associated with elevated risk for cardiovascular disease, even in individuals with early-stage diabetes [2,3]. For detecting hyperglycemia, however, multiple blood sampling is necessary. The availability of a system for estimating postprandial glucose excursion without the need for blood sampling would be beneficial. Therefore, we investigated and developed such a system on the basis of the estimation of the glucose area under the curve (AUC) value using minimally invasive interstitial fluid extraction technology (MIET) [4,5]. In the MIET procedure, a microneedle array (length 300 μm) is stamped on the forearm skin as pretreatment to enhance interstitial fluid (ISF) extraction. Subsequently, a hydrogel patch is placed on the pretreated area to accumulate ISF for a specific period. The ISF extraction level is determined from the glucose AUC during this time and is measured by analyzing the glucose levels in the hydrogel with the simultaneous measurement of sodium ion levels as an internal standard for ISF extraction. The glucose levels in the hydrogel are dependent on both the formation of skin micropores and the blood glucose levels during measurement. In contrast, sodium ion levels in the hydrogel are only dependent on the formation of the skin micropores because the levels show only a narrow variation among individuals, in addition to the low circadian and day-to-day variations in an individual. Therefore, we use sodium ion levels to calibrate differences in ISF glucose extraction [4]. No additional force beyond passive diffusion and osmotic pressure is needed to extract the ISF from the skin. This easy-to-use glucose monitoring system may be a useful tool for not only screening [6] but also enhancing patients’ understanding of the correlation between individual foods and postprandial glucose excursion [7]. In these clinical studies, however, the hydrogel patches after extracting ISF were immersed into pure water for several hours to extract glucose and sodium ions. This was because the glucose was measured by a fluorescence microplate reader, and the sodium ions were measured by a combination of sodium and potassium ion-selective electrodes (ISEs) or by ion chromatography. It would be very useful if we could measure the glucose and sodium ions in a hydrogel patch directly. Therefore, in this study, we developed a small, integrated glucose sensor with a high-selectivity, all-solid-state sodium ISE. The hydrogel consisted of polyvinyl alcohol and a 268 mM (10−0.57 M) KCl solvent for osmotic adjustment, and the hydrogel size was ϕ = 10 mm. Previous clinical studies showed that glucose levels in hydrogel were between 0.6 and 20 mg/dL (0.03–1.1 mM) and the sodium ion levels ranged from 0.4 to 14 mM (from 10−3.4 to 10−1.9 M) [4,5,6]. Although the physiological glucose level of diabetes patients is 40–400 mg/dL (2.2–22 mM) and their physiological level of sodium ions is around 140 mM (10−0.85 M), the levels of glucose and sodium ions in the hydrogel are more than ten times lower than those in blood. This is because small amounts of ISF glucose and sodium ions are extracted through micropores and are diluted by the solvent in the hydrogels. Therefore, high sensitivity is required for our glucose and sodium ion sensors.
Glucose sensors have been widely developed around the world, mostly for self-monitoring blood glucose. Such sensors are meant to measure blood glucose levels, therefore, the sensitivity is not sufficient to measure the glucose levels in the hydrogels of our minimally invasive glucose monitoring. On the other hand, many highly sensitive glucose sensors have been reported thus far: Yao et al. developed an immobilized glucose oxidase (GOD) glucose sensor, and its lower limit was 9 nM [8]. Its operation was based on measuring the increase of the anodic current during the oxidation of hydrogen peroxide (H2O2). H2O2 is produced from the oxidation of glucose by dissolved oxygen in the presence of GOD [9]. Because Ikariyama et al. also developed a ϕ = 10 µm–immobilized GOD glucose sensor [10], a GOD-fixed glucose sensor could possibly be miniaturized and exhibit high sensitivity.
ISEs are often used to analyze ionic components in a liquid sample; however, there were two difficulties facing the development of a sodium ISE for the hydrogel measurement. First, because of the 268 mM KCl in the hydrogel for osmotic adjustment, high Na+/K+ selectivity was necessary for the ionophore of the sodium ion-selective membrane. Because the Na+/K+ selectivity of conventional ionophore Bis12-crown-4 is 100 log k N a , K p o t = −2) [11], a conventional ISE using Bis12-crown-4 is unsuitable for hydrogel measurement. Hence, we decided to use DD16C5 as an ionophore because it shows a high Na+/K+ selectivity (about 1000, ( log k N a , K p o t = −3).
Second, to measure hydrogel samples (ϕ = 10 mm), sensors needed to be miniaturized. Conventional ISEs have internal reference and solution, making them difficult to miniaturize. Hence, for the purposes of miniaturization, all-solid-state ISEs have been widely studied. Although coated-wire electrodes, metallic wire electrodes directly covered with an ion-selective membrane has been studied as all-solid-state ISEs since 1971, they generally show insufficient potential stability. To achieve the potential stability and miniaturization of sodium ISEs, an insertion material Na0.33MnO2 has been studied for all-solid-state ISEs [12,13]. Komaba et al. reported that the ISE using Na0.33MnO2 as an inner contact layer is highly efficient for stabilization [13]. This stabilization of the membrane potential was because the interface resistance at the electrode reduced drastically.
In this study, we developed a highly sensitive GOD-fixed glucose sensor that incorporated a high-selectivity, miniaturized sodium ISE using the insertion material Na0.33MnO2 as an inner contact layer and DD16C5 as an ionophore. We evaluated the response time, Na+/K+ selectivity, accuracy, precision in hydrogel measurement, long stability, and potential drift in solution measurement.

2. Experimental Section

2.1. Construction of a Glucose Sensor

A platinum screen-printed electrode (DRP-550, DropSense, Oviedo, Spain) was used for a glucose sensor. First, the Ag/AgCl ink (BAS, Tokyo, Japan) was pasted onto the reference electrode and dried at 120 °C for 10 min. Then, 3.6 μL of mixed enzyme solution was spread on the working electrode (ϕ = 4 mm). The enzyme solution consisted of 43 mg/mL bovine serum albumin (Sigma-Aldrich, St. Louis, MO, USA), 0.76 U/mL GOD (Wako Pure Chemical Industries, Osaka, Japan), 0.18 U/mL mutarotase (Wako Pure Chemical Industries), and 0.25% glutaraldehyde (Nacalai Tesque, Kyoto, Japan); the dilutions were performed using phosphate-buffered saline. We used mutarotase to enhance the GOD reaction through a rapid conversion of α-D-glucose to β-D-glucose because only β-D-glucose is oxidized and yields hydrogen peroxide by GOD. The mixed-enzyme solution was dried at 25 °C and 30% relative humidity for 24 h.

2.2. Construction of a Sodium ISE

Na0.44MnO2 powder was obtained using a stoichiometric mixture of MnCO3 (Wako Pure Chemical Industries) and Na2CO3 (Wako Pure Chemical Industries). The mixture was successively heated for 24 h at 800 °C under air with intermediate grinding. Na0.44MnO2 powder was aged in a solution of 10−2 M NaNO3 (Wako Pure Chemical Industries). After 40 h, Na0.33MnO2 was obtained by filtration and dried [12].
The Na0.33MnO2 slurry was cast on a carbon screen-printed electrode (DEP-Chip, BioDeviceTechnology, Ishikawa, Japan) and completely dried at room temperature under the conditions of less than 10% relative humidity for at least 1 day. The slurry consisted of 0.04 g Na0.33MnO2, 0.005 g acetylene black as a conductive additive and 0.005 g polyvinylidene difluoride as a binder dispersed in 150 μL of N-methylpyrrolidone solvent. After casting the slurry, the electrode was immersed for at least 1 day in 1 M NaCl solution to condition the electrode.
The tetrahydrofuran (Wako Pure Chemical Industries) solution consisting of DD16C5 (3.0 wt %, Dojindo Molecular Technologies, Kumamoto, Japan) as an ionophore, poly(vinyl chloride) (PVC, 29.0 wt %, degree of polymerization: 1100, Wako Pure Chemical Industries) as a membrane matrix, tris(2-ethylhexyl) phosphate (67.7 wt % Sigma-Aldrich) as a plasticizer, and potassium tetrakis(4-chlorophenyl)borate (KTCBP, 0.33 wt %, Tokyo Chemical Industry, Tokyo, Japan) as a lipophilic anion was further cast on top of the formed Na0.33MnO2 membrane and dried for at least 1 day to form a PVC ion-selective membrane. After drying the membrane, the electrode was immersed for at least 1 day in 1 M sodium chloride solution for conditioning of the electrode.

2.3. Experimental Protocol

2.3.1. Settings for the Sensors

The glucose sensor (working electrode ϕ = 4 mm) and sodium ISE (working electrode ϕ = 2 mm) were positioned to measure the hydrogel (ϕ = 10 mm) (Figure 1). One reference electrode (Ag/AgCl ink) was used for both the glucose sensor and sodium ISE. The chronoamperometric measurements of the glucose sensor and the potentiometric measurements of the sodium ISE were performed using a potentiostat (Als832a, BAS) and a digital multimeter (PC20, Sanwa Electric Instrument, Tokyo, Japan), respectively. For the glucose sensor, the applied potential was set to +0.45 V versus the Ag/AgCl reference electrode. The room temperature was kept at 25 °C during measurements. Both sensors were immersed for 1–2 h in 268 mM KCl solution for conditioning and stabilizing the background currents.
Figure 1. Scheme of the glucose and sodium ion sensors. ISE: ion-selective electrode; W.E.: working electrode; R.E.: reference electrode; C.E.: counter electrode.
Figure 1. Scheme of the glucose and sodium ion sensors. ISE: ion-selective electrode; W.E.: working electrode; R.E.: reference electrode; C.E.: counter electrode.
Micromachines 06 00831 g001

2.3.2. Preparations of Hydrogels

Standard hydrogels were prepared by immersing hydrogels into solutions. Components of these solutions are listed in Table 1. We prepared six types of standard hydrogels, three of which (Gel No. 1, 3, and 6) were used for calibration curve fitting, and the others (Gel No. 2, 4, and 5) were used to evaluate the accuracy and precision of the estimated glucose and sodium ion levels. Other gels (Gel ID: Washing) were used to wash the surfaces of the sensors after the measurement of the levels of Gels No. 1–6. The level range was determined from previous clinical studies [4,5,6], and we set four measurement points between the highest and lowest levels. For each hydrogel, 2.25 mL of immersing solution was prepared. The hydrogels were separated from the adhesive tape of the hydrogel patch used for MIET [4,5] and placed in the immersing solutions for at least 1 day.
Table 1. Composition of the immersion solution for hydrogels.
Table 1. Composition of the immersion solution for hydrogels.
Gel IDNo. 1 *No. 2No. 3 *No. 4No. 5No. 6 *Washing
Glucose (mg/dL)00.612.410200
NaCl (mM) (log[Na+] (M))0.1 (−4.00)0.4 (−3.40)0.5 (−3.30)1 (−3.00)3.5 (−2.46)6.8 (−2.17)0
KCl (mM)268268268268268268268
Note: * hydrogels for calibration.

2.3.3. Hydrogel Measurement Protocol

The standard gel (Table 1, Gel ID: No. 1–6) was taken out of the immersion solution, and placed on the surface of a glucose sensor and sodium ISE. Then, the gel was continuously pressed by a push–pull gauge (AP-30N, Attonic, Aichi, Japan) fixed to a desktop robot arm (V2200, SAN-EI TECH, Chiba, Japan) to stabilize the baseline current of the glucose sensor and the baseline potential of the sodium ISE. After stabilizing the sensors, the gel for measurement was placed on the push–pull gauge, and the robot arm was lowed until the gauge sensed 1 N. The measurement time was 3 min after the attachment of the gel to the sensors. For the gel washing procedure, we used gels that did not contain glucose and sodium ions (Table 1, Gel ID: washing). After 3 min measurement of a standard gel, a washing gel was placed on the surface of the glucose sensor and sodium ISE and was pressed in the same manner as for the measurement of standard gels. Representative response time curves for the glucose sensor and sodium ISE are shown in Figure 2.
Figure 2. Representative response time curves during gel measurements. (a) Glucose sensor and (b) sodium ISE.
Figure 2. Representative response time curves during gel measurements. (a) Glucose sensor and (b) sodium ISE.
Micromachines 06 00831 g002

2.3.4. Data Analysis

The average signals from 40 to 50 s after attaching the gel to the sensors were defined as the output values. The apparent Michaelis–Menten constant K M a p p is generally used to evaluate the biological activity of immobilized enzymes (Equation (1)) [14,15]. In this study, the output current from gel measurement was proportional to the steady–state current Iss (correlation coefficient: r = 0.999). Thus, the glucose sensor calibration curve was determined as follows (Equation (2)):
  I ss = I m a x ×   C g l u ( K M a p p +   C g l u )
where Iss is the steady–state current, Imax is the maximum current measured under glucose saturation, Cglu is the concentration of glucose and K M a p p is the apparent Michaelis–Menten constant [14,15].
I s s = a × O u t p u t   c u r r e n t = I m a x × C g l u ( K M a p p + C g l u ) O u t p u t   c u r r e n t = ( I m a x   / a ) × C g l u ( K M a p p + C g l u )
where a is a constant coefficient.
For the electromotive force (EMF) of sodium ISE, we used the Nicolsky–Eisenman Equation (Equation (3)):
O u t p u t   E M F = 2.303 R T 1 × F l o g 10 ( C N a + k N a , K p o t C K ) + c o n s t = 0.5917 l o g 10 ( C N a + k N a , K p o t C K ) + c o n s t
where CNa is the concentration of sodium ions, CK is the concentration of potassium ions (268 mM), and k N a , K p o t   and const are constant coefficients.
Using the data from gel calibration measurements, the calibration curve coefficients Imax/a, K M a p p (Equation (2)) and k N a , K p o t and const (Equation (3)) were determined using the least-squares method. Estimated glucose levels and sodium ions levels were calculated from Equations (4) and (5):
E s t i m a t e d   C g l u = K M a p p   ×   O u t p u t   c u r r e n t (   I m a x   / a O u t p u t   c u r r e n t )
E s t i m a t e d   C N a = 10 O u t p u t   E M F c o n s t 0.05917 k N a , K p o t × C K

3. Results and Discussion

3.1. Response Time

Figure 3 shows the response time curves during gel measurements, and Figure 4 shows these curves during 1 min after the gel was attached to the sensors. The response times of the developed sodium ISE in hydrogel measurements were t90 = 14 s (corresponding to 90% of the activity change span) or t95 = 27 s (corresponding to 95% of the activity change span). These are almost at the same level of some commercial sodium ISEs: t90 < 30 s (1512A-10C, Horiba, Kyoto, Japan) [16]) and t95 < 30 s (uniPROBE sodium ISE (T.P.S., Brendale, Australia) [17]).
Figure 3. Response time curves during gel measurements. (a) Glucose sensor and (b) sodium ISE.
Figure 3. Response time curves during gel measurements. (a) Glucose sensor and (b) sodium ISE.
Micromachines 06 00831 g003
Figure 4. Response time curves during 60 s after gel was attached to the sensors. (a) Glucose sensor and (b) sodium ISE.
Figure 4. Response time curves during 60 s after gel was attached to the sensors. (a) Glucose sensor and (b) sodium ISE.
Micromachines 06 00831 g004

3.2. Calibration Curve and Na+/K+ Selectivity

Figure 5 shows the calibration curves determined using the data from hydrogel calibration measurement. The calibration curve coefficients were Imax/a = 7214 nA,   K M a p p = 50.24 mg/dL (Equation (2), least-squares method), k N a , K p o t = 0.001692, and const = 104 (Equation (3), Fixed Interference Method [18]) The Na+/K+ selectivity ( log k N a , K p o t = −2.8) was high compared with that of sodium ISE using Bis12-crown-4 ( log k N a , K p o t = −2 [11]). From the result, we confirmed the ionophore DD16C5 in the selective membrane worked well even when it was coated over inner solid insertion membrane. It is sufficient for directly measuring sodium ions in MIET hydrogels including 268 mM (10−0.57 M) KCl solvent.
Figure 5. Calibration curve of (a) glucose sensor and (b) sodium ISE.
Figure 5. Calibration curve of (a) glucose sensor and (b) sodium ISE.
Micromachines 06 00831 g005

3.3. Accuracy, Precision, and Long-Term Stability

A good correlation between estimated and reference values (r > 0.99) was confirmed by measuring three types of standard gels (Figure 6). The estimated values were calculated from the signals obtained from the newly developed glucose sensor and sodium ISE. The reference values are the concentrations of glucose and sodium ions in the measured hydrogels. The coefficient of variation (CV) was less than 5.6% (Table 2). The correlation and coefficient variation remained stable for three months after fabrication (during these three months, glucose sensors were stored at 4 °C and left unused and sodium ISEs were stored at room temperature and used at least three times).
Figure 6. Correlation between estimated and reference values (blank plots: calibration gels). (a) Glucose sensor and (b) sodium ISE.
Figure 6. Correlation between estimated and reference values (blank plots: calibration gels). (a) Glucose sensor and (b) sodium ISE.
Micromachines 06 00831 g006
Table 2. Coefficient of variation of estimated glucose and sodium ion levels.
Table 2. Coefficient of variation of estimated glucose and sodium ion levels.
Gel No.Glucose (mg/dL)CVlog[Na+] (M)CVn
20.62.9%−3.45.6%5
42.44.4%−3.03.3%5

3.4. Reproducibility of the Glucose Sensor

The reproducibility of glucose measurements was evaluated for three developed glucose sensors. The sensors showed good coefficients of variation of estimated glucose levels for measurements of the level of a hydrogel that contains glucose at 0.6 mg/dL (Table 3). Because the glucose was measured by a fluorescence microplate reader, in the previous clinical studies, the hydrogel patches after extracting ISF were immersed into pure water for several hours to extract glucose and sodium ions. The results indicated that the newly developed glucose sensor could be used for measuring MIET hydrogel patches directly without immersing the hydrogel patches into pure water as a substitute for fluorimetry if the sensor could be mass produced in future.
Table 3. Coefficient of variation of estimated glucose levels (three developed glucose sensors).
Table 3. Coefficient of variation of estimated glucose levels (three developed glucose sensors).
Glucose Concentration in the Hydrogels (mg/dL)The Numbers of Measured HydrogelsCV
Sensor 1Sensor 2Sensor 3
0.6n = 54.0%2.7%2.9%

3.5. Potential Drift of the All-Solid-State Sodium ISE

The potential drift of the developed all-solid-state sodium ISE was evaluated in 268 mM KCl solution, using Ag/AgCl (saturated KCl) reference electrode. Compared with the potential drift of Pt/PVC (coat wired electrode, CWE), the developed sodium ISE exhibited almost the same stability as that obtained using a Pt/LixFePO4(insertion)/PVC (Li ionophore) [13] (Figure 7). The good stability performance of carbon/Na0.33MnO2/PVC (DD16C5) was revealed in this study for the first time. The insertion materials layers are ion/electron mixed conductor, it makes ionic and electronic communication possible at the interface between the ion-selective membrane and Pt or carbon electrode. As a result, the electrode impedance is significantly reduced and electrode potential is effectively stabilized [13].

3.6. Study Limitations

In this study, we evaluated the in vitro measurement performance; the next developmental stage is to assess the measurement performance in vivo. We will also evaluate the effect of interference currents generated by chemicals such as ascorbic acid, uric acid, and acetaminophen because our glucose sensor applies a +0.45 V potential for glucose detection.
In addition, the number of calibrations should also be reduced for practical use. To realize no calibration, further studies should require baseline current, potential drift, and sensitivity of the sensors to be more stable.
Figure 7. Long-term potential stability of the developed all-solid-state sodium-ISE (n = 6) compared with Pt/PVC (coat wired electrode, CWE) and Pt/insertion (LixFePO4)/PVC (x = 1.0, 0.5 and 0) in 10 mM LiCl (15 days stability).
Figure 7. Long-term potential stability of the developed all-solid-state sodium-ISE (n = 6) compared with Pt/PVC (coat wired electrode, CWE) and Pt/insertion (LixFePO4)/PVC (x = 1.0, 0.5 and 0) in 10 mM LiCl (15 days stability).
Micromachines 06 00831 g007

4. Conclusions

We developed a small glucose sensor with an all-solid-state sodium ISE integrated into its design. The developed sodium ISE showed high selectivity ( l o g k N a , K p o t = −2.8, FIM) and good potential stability. The performance of the developed glucose sensor with sodium ISE was also sufficient for simultaneously detecting glucose and sodium ions in hydrogel within 1 min, and the size is smaller than the hydrogel (ϕ < 10 mm) for minimally invasive glucose monitoring system.

Acknowledgments

We would like to thank Shinichi Komaba (Tokyo University of Science) and his laboratory members for their assistance and discussion. We would also like to thank Yasuko Kawamoto (Sysmex Corp., Japan) and Yuu Morishita (Sysmex Engineering Corp., Kobe, Japan) for their assistance and precise measurements.

Author Contributions

Junko Kojima researched the data and wrote the manuscript. Samiko Hosoya researched the data. Chihiro Suminaka and Nobuyasu Hori contributed to the discussion. Toshiyuki Sato contributed to the discussion and reviewed all data and the manuscript.

Conflicts of Interest

Junko Kojima, Samiko Hosoya, Chihiro Suminaka, Nobuyasu Hori, and Toshiyuki Sato are employees of Sysmex Corporation. No other potential conflicts of interest relevant to this article were reported.

References

  1. International Diabetes Federation. Diabetes Atlas, 6th ed.; International Diabetes Federation: Brussels, Belgium, 2014. [Google Scholar]
  2. DECODE Study Group. Is the current definition for diabetes relevant to mortality risk from all causes and cardiovascular and noncardiovascular diseases? Diabetes Care 2003, 26, 688–696. [Google Scholar]
  3. Nakagami, T.; DECODA Study Group. Hyperglycaemia and mortality from all causes and from cardiovascular disease in five populations of Asian origin: The DECODA study. Diabetologia 2004, 47, 385–394. [Google Scholar] [CrossRef] [PubMed]
  4. Sato, T.; Okada, S.; Hagino, K.; Asakura, Y.; Kikkawa, Y.; Kojima, J.; Watanabe, T.; Maekawa, Y.; Isobe, K.; Koike, R.; et al. A novel technique of glucose area under the curve monitoring with interstitial fluid. Diabetes Technol. Ther. 2011, 13, 1194–1200. [Google Scholar] [CrossRef] [PubMed]
  5. Sakaguchi, K.; Hirota, Y.; Hashimoto, N.; Ogawa, W.; Sato, T.; Okada, S.; Hagino, K.; Asakura, Y.; Kikkawa, Y.; Kojima, J.; et al. A minimally invasive system for glucose area under the curve measurement using interstitial fluid extraction technology; evaluation of the accuracy and usefulness with oral glucose tolerance tests in subject with and without diabetes. Diabetes Technol. Ther. 2012, 14, 485–491. [Google Scholar] [CrossRef] [PubMed]
  6. Sakamoto, K.; Kubo, F.; Yoshiuchi, K.; Ono, A.; Sato, T.; Tomita, K.; Sakaguchi, K.; Matsuhisa, M.; Kaneto, H.; Maegawa, H.; et al. Usefulness of a novel system for measuring glucose area under the curve while screening for glucose intolerance in outpatients. J. Diabetes Investig. 2013, 4, 552–559. [Google Scholar] [CrossRef] [PubMed]
  7. Kuranuki, S.; Sato, T.; Okada, S.; Hosoya, S.; Seko, A.; Sugihara, K.; Nakamura, T. Evaluation of postprandial glucose excursion using a novel minimally invasive glucose area-under-the-curve monitoring system. J. Healthc. Eng. 2013, 4, 529–540. [Google Scholar] [CrossRef] [PubMed]
  8. Yao, T. A chemically-modified enzyme membrane electrode as an amperometric glucose sensor. Anal. Chim. Acta 1983, 148, 27–33. [Google Scholar] [CrossRef]
  9. Guilbault, G.G.; Lubrano, G.L. An enzyme electrode for the amperometric determination of glucose. Anal. Chim. Acta 1973, 64, 439–445. [Google Scholar] [CrossRef]
  10. Ikariyama, Y.; Yamauchi, S.; Yukiashi, T.; Ushioda, H. Electrochemical fabrication of amperometric microenzyme sensor. J. Electrochem. Soc. 1989, 136, 702–706. [Google Scholar] [CrossRef]
  11. Tamura, H.; Kimura, K.; Shono, T. Coated wire sodium- and potassium-selective electrodes based on bis(crown ether) compounds. Anal. Chem. 1982, 54, 1224. [Google Scholar] [CrossRef]
  12. Sauvage, F.; Baudrn, E.; Tarascon, J.M. Study of the potentiometric response towards sodium ions of Na0.44−xMnO2 for the development of selective sodium ion sensors. Sens. Actuators B Chem. 2007, 120, 638–644. [Google Scholar] [CrossRef]
  13. Komaba, S.; Suzuki, C.; Yabuuchi, N.; Akatsuka, T.; Kanazawa, S.; Hasegawa, T. Redox-active alkali insertion materials as inner contact layer in all-solid-state ion-selective electrodes. ECS Trans. 2013, 50, 279–287. [Google Scholar] [CrossRef]
  14. Kamin, R.A.; Wilson, G.S. Rotating ring-disk enzyme electrode for biocatalysis kinetic studies and characterization of the immobilized enzyme layer. Anal. Chem. 1980, 52, 1198–1205. [Google Scholar] [CrossRef]
  15. Kong, T.; Chen, Y.; Ye, Y.P.; Zhang, K.; Wang, Z.X.; Wang, X.P. An amperometric glucose biosensor based on the immobilization of glucose oxidase on the ZnO nanotubes. Sens. Actuators B Chem. 2009, 138, 344–350. [Google Scholar] [CrossRef]
  16. Horiba, Ltd. Instruction Manuals (Sodium Ion Electrode 1512A-10C). Available online: http://www.horiba.com/application/material-property-characterization/water-analysis/water-quality-electrochemistry-instrumentation/electrodes-accessories/ion-electrodes/details/sodium-ion-electrode-1512a-10c-9563/ (accessed on 27 April 2015).
  17. T.P.S. Pty Ltd. TPS uniPROBE Sodium (Na+) ISE. Available online: http://www.tps.com.au/handbooks/uniPROBE%20Sodium%20Instruction%20Sheet%20v2_2.pdf (accessed on 27 April 2015).
  18. Buck, R.P.; Lindner, E. Recommendations for nomenclature of ionselective electrodes (IUPAC Recommendations 1994). Pure Appl. Chem. 1994, 66, 2527–2536. [Google Scholar] [CrossRef]

Share and Cite

MDPI and ACS Style

Kojima, J.; Hosoya, S.; Suminaka, C.; Hori, N.; Sato, T. An Integrated Glucose Sensor with an All-Solid-State Sodium Ion-Selective Electrode for a Minimally Invasive Glucose Monitoring System. Micromachines 2015, 6, 831-841. https://doi.org/10.3390/mi6070831

AMA Style

Kojima J, Hosoya S, Suminaka C, Hori N, Sato T. An Integrated Glucose Sensor with an All-Solid-State Sodium Ion-Selective Electrode for a Minimally Invasive Glucose Monitoring System. Micromachines. 2015; 6(7):831-841. https://doi.org/10.3390/mi6070831

Chicago/Turabian Style

Kojima, Junko, Samiko Hosoya, Chihiro Suminaka, Nobuyasu Hori, and Toshiyuki Sato. 2015. "An Integrated Glucose Sensor with an All-Solid-State Sodium Ion-Selective Electrode for a Minimally Invasive Glucose Monitoring System" Micromachines 6, no. 7: 831-841. https://doi.org/10.3390/mi6070831

APA Style

Kojima, J., Hosoya, S., Suminaka, C., Hori, N., & Sato, T. (2015). An Integrated Glucose Sensor with an All-Solid-State Sodium Ion-Selective Electrode for a Minimally Invasive Glucose Monitoring System. Micromachines, 6(7), 831-841. https://doi.org/10.3390/mi6070831

Article Metrics

Back to TopTop