Implementation of Constrained Swept Synthetic Aperture Using a Mechanical Fixture
Abstract
:1. Introduction
2. Materials and Methods
2.1. Pulse Sequencing
2.2. Simulation
- Reference image—Multistatic acquisition (each transmit/receive element pair) from a single array position generates a synthetically focused baseline image without the SSA technique.
- Subsample estimation sweep—Simulations of plane wave transmission and single-element diverging transmission (from the center of the array) were produced from fine displacements up to a single element pitch in 0.03 mm steps.
- Large displacement estimation sweep—Simulations of plane wave transmission and single-element diverging transmission were produced from larger displacements as would be found in a 5 cm SSA sweep in steps of 0.1 mm.
- Equivalent image—Full synthetic aperture acquisition from a fully populated 7 cm long array produces a similar image to the 5 cm SSA approach when the active transmit aperture is limited to 5 cm.
2.3. Phantom Acquisition
2.3.1. Translation Stage
2.3.2. Slide Device
2.3.3. Data Acquisition
- The imaging target was the ATS 549 tissue mimicking phantom (CIRS, Norfolk, VA, USA), and the transducer was attached to the translation stage as described above. The transducer was swept over 5 cm in 0.033 mm steps, alternating between plane wave transmission and left/right diverging transmissions (virtual source and 11 elements with f-number −0.75) in that order to mimic a continuous sweep with spacing of 0.1 mm between repetitions.
- The transducer was used with the linear slide device on the ATS 549 phantom. The phantom was positioned such that point targets were located below lesion targets, and both were visible in the image. The transducer was coupled with gel, and the device was held steady on the phantom surface by hand. The transducer was swept along the track covering approximately 5 cm in 1 s, although the temporal dynamics and extent varied from sweep to sweep. During the sweep, transmissions were alternated as described in Figure 2b at an overall pulse repetition frequency of 500 Hz.
- The manual sweep was repeated with an ex vivo tissue sample between the transducer and an ATS 539 phantom. The phantom was positioned so that lesion targets were located below point targets, and both were visible in the image. Three different store-bought meats were used to study whether the intervening layer affected motion tracking. Pork chop (12 mm thick), beef loin (10 mm thick) and chicken breast (9–19 mm thick) were used as varying models. Samples were prepared to be wider than the slide device so that they rested entirely on top of the sample. The degassed samples were placed on top of plastic wrap, and all layers were coupled with gel. The same sweep sequence was used as above.
2.4. Motion Estimation Methods
2.4.1. Speckle Tracking
2.4.2. Channel Correlation
2.4.3. Subsample and Multi-Lag Estimation
2.5. Image Reconstruction and Analysis
3. Results
3.1. Simulation
3.1.1. Displacement Estimation
3.1.2. Imaging Results
3.2. Phantom
3.2.1. Translation Stage
3.2.2. Linear Fixture
4. Discussion
5. Conclusions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
Abbreviations
FWHM | Full width at half maximum |
gCNR | Generalized contrast-to-noise ratio |
HCC | Hepatocellular carcinoma |
PSF | Point spread function |
RF | Radio frequency |
ROI | Region of interest |
SSA | Swept synthetic aperture |
References
- Royer, D.F. Seeing with Sound: How Ultrasound Is Changing the Way We Look at Anatomy. In Advances in Experimental Medicine and Biology; Springer International Publishing: Berlin/Heidelberg, Germany, 2019; pp. 47–56. [Google Scholar] [CrossRef]
- Miller, D.L.; Abo, A.; Abramowicz, J.S.; Bigelow, T.A.; Dalecki, D.; Dickman, E.; Donlon, J.; Harris, G.; Nomura, J. Diagnostic Ultrasound Safety Review for Point-of-Care Ultrasound Practitioners. J. Ultrasound Med. 2020, 39, 1069–1084. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Singal, A.G.; El-Serag, H.B. Rational HCC screening approaches for patients with NAFLD. J. Hepatol. 2022, 76, 195–201. [Google Scholar] [CrossRef] [PubMed]
- European Society of Radiology (ESR). ESR statement on portable ultrasound devices. Insights Imaging 2019, 10, 89. [Google Scholar] [CrossRef] [Green Version]
- Smith, S.W.; Wagner, R.F.; Sandrik, J.F.M.F.; Lopez, H. Low contrast detectability and contrast/detail analysis in medical ultrasound. IEEE Trans. Sonics Ultrason. 1983, 3, 164–173. [Google Scholar] [CrossRef]
- Tzartzeva, K.; Obi, J.; Rich, N.E.; Parikh, N.D.; Marrero, J.A.; Yopp, A.; Waljee, A.K.; Singal, A.G. Surveillance Imaging and Alpha Fetoprotein for Early Detection of Hepatocellular Carcinoma in Patients With Cirrhosis: A Meta-analysis. Gastroenterology 2018, 154, 1706–1718.e1. [Google Scholar] [CrossRef] [Green Version]
- Dong, Y.; Teufel, A.; Wang, W.P.; Dietrich, C.F. Current Opinion about Hepatocellular Carcinoma <10 mm. Digestion 2021, 102, 335–341. [Google Scholar] [CrossRef]
- Eisenbrey, J.R.; Gabriel, H.; Savsani, E.; Lyshchik, A. Contrast-enhanced ultrasound (CEUS) in HCC diagnosis and assessment of tumor response to locoregional therapies. Abdom. Radiol. 2021, 46, 3579–3595. [Google Scholar] [CrossRef]
- American College of Radiology. US LI-RADS v2017 CORE. Available online: https://www.acr.org/-/media/ACR/Files/RADS/LI-RADS/LI-RADS-US-Algorithm-Portrait-2017.pdf (accessed on 10 April 2023).
- Bottenus, N.; Pinton, G.; Trahey, G. The Impact of Acoustic Clutter on Large Array Abdominal Imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2020, 67, 703–714. [Google Scholar] [CrossRef]
- Bottenus, N.; Long, W.; Morgan, M.; Trahey, G. Evaluation of Large-Aperture Imaging Through the ex Vivo Human Abdominal Wall. Ultrasound Med. Biol. 2018, 44, 687–701. [Google Scholar] [CrossRef]
- Bottenus, N.; Long, W.; Zhang, H.K.; Jakovljevic, M.; Bradway, D.P.; Boctor, E.M.; Trahey, G.E. Feasibility of Swept Synthetic Aperture Ultrasound Imaging. IEEE Trans. Med. Imaging 2016, 35, 1676–1685. [Google Scholar] [CrossRef]
- Peralta, L.; Gomez, A.; Luan, Y.; Kim, B.; Hajnal, J.V.; Eckersley, R.J. Coherent Multi-Transducer Ultrasound Imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2019, 66, 1316–1330. [Google Scholar] [CrossRef] [Green Version]
- Peralta, L.; Ramalli, A.; Reinwald, M.; Eckersley, R.J.; Hajnal, J.V. Impact of Aperture, Depth, and Acoustic Clutter on the Performance of Coherent Multi-Transducer Ultrasound Imaging. Appl. Sci. 2020, 10, 7655. [Google Scholar] [CrossRef]
- Foiret, J.; Cai, X.; Bendjador, H.; Park, E.Y.; Kamaya, A.; Ferrara, K.W. Improving plane wave ultrasound imaging through real-time beamformation across multiple arrays. Sci. Rep. 2022, 12, 13386. [Google Scholar] [CrossRef]
- Wodnicki, R.; Kang, H.; Li, D.; Stephens, D.N.; Jung, H.; Sun, Y.; Chen, R.; Jiang, L.M.; Cabrera-Munoz, N.E.; Foiret, J.; et al. Highly Integrated Multiplexing and Buffering Electronics for Large Aperture Ultrasonic Arrays. BME Front. 2022, 2022, 9870386. [Google Scholar] [CrossRef]
- Zhang, H.K.; Cheng, A.; Bottenus, N.; Guo, X.; Trahey, G.E.; Boctor, E.M. Synthetic tracked aperture ultrasound imaging: Design, simulation, and experimental evaluation. J. Med. Imaging 2016, 3, 027001. [Google Scholar] [CrossRef] [Green Version]
- Ylitalo, J.; Ermert, H. Ultrasound synthetic aperture imaging: Monostatic approach. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1994, 41, 333–339. [Google Scholar] [CrossRef]
- Weng, L.; Tirumalai, A.; Lowery, C. US extended-field-of-view imaging technology. Radiology 1997, 203, 877–880. [Google Scholar] [CrossRef]
- Prager, R.W.; Ijaz, U.Z.; Gee, A.H.; Treece, G.M.; Wells, P.N.T. Three-dimensional ultrasound imaging. Proc. Inst. Mech. Eng. Part H J. Eng. Med. 2010, 224, 193–223. [Google Scholar] [CrossRef]
- Prager, R.W.; Gee, A.; Berman, L. Stradx: Real-time acquisition and visualization of freehand three-dimensional ultrasound. Med. Image Anal. 1999, 3, 129–140. [Google Scholar] [CrossRef]
- Chung, S.W.; Shih, C.C.; Huang, C.C. Freehand three-dimensional ultrasound imaging of carotid artery using motion tracking technology. Ultrasonics 2017, 74, 11–20. [Google Scholar] [CrossRef]
- Zheng, S.; Huang, Q.; Jin, L.; Wei, G. Real-time extended-field-of-view ultrasound based on a standard PC. Appl. Acoust. 2012, 73, 423–432. [Google Scholar] [CrossRef]
- Prevost, R.; Salehi, M.; Jagoda, S.; Kumar, N.; Sprung, J.; Ladikos, A.; Bauer, R.; Zettinig, O.; Wein, W. 3D freehand ultrasound without external tracking using deep learning. Med. Image Anal. 2018, 48, 187–202. [Google Scholar] [CrossRef] [PubMed]
- Karaman, M.; Bilge, H.; O’Donnell, M. Adaptive multi-element synthetic aperture imaging with motion and phase aberration correction. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1998, 45, 1077–1087. [Google Scholar] [CrossRef] [PubMed]
- Bottenus, N. Estimation of transducer translation using channel-domain correlation. In Proceedings of the 2019 IEEE International Ultrasonics Symposium (IUS), Glasgow, Scotland, 6–9 October 2019; pp. 1009–1012. [Google Scholar] [CrossRef]
- O’Donnell, M.; Silverstein, S. Optimum displacement for compound image generation in medical ultrasound. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1987, 35, 470–476. [Google Scholar] [CrossRef] [PubMed]
- Walker, W.F.; Trahey, G.E. Fundamental limit on delay estimation using partially correlated speckle signals. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1995, 42, 301–308. [Google Scholar] [CrossRef]
- Karaman, M.; O’Donnell, M. Synthetic aperture imaging for small scale systems. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1995, 42, 429–442. [Google Scholar] [CrossRef]
- Jensen, J.A.; Svendsen, N.B. Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1992, 39, 262–267. [Google Scholar] [CrossRef] [Green Version]
- Jensen, J.A. Field: A Program for Simulating Ultrasound Systems. Med. Biol. Eng. Comput. 1996, 34, 351–353. [Google Scholar]
- Kasai, C.; Namekawa, K.; Koyano, A.; Omoto, R. Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique. IEEE Trans. Sonics Ultrason. 1985, 32, 458–464. [Google Scholar] [CrossRef]
- Loupas, T.; Powers, J.; Gill, R.W. An Axial Velocity Estimator for Ultrasound Blood Flow Imaging, Based on a Full Evaluation of the Doppler Equation by Means of a Two-Dimensional Autocorrelation Approach. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1995, 42, 672–688. [Google Scholar] [CrossRef]
- Trahey, G.E.; Allison, J.W.; von Ramm, O.T. Angle independent ultrasonic detection of blood flow. IEEE Trans. Biomed. Eng. 1987, 34, 965–967. [Google Scholar] [CrossRef]
- Jensen, J.A.; Munk, P. A new method for estimation of velocity vectors. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1998, 45, 837–851. [Google Scholar] [CrossRef]
- Anderson, M.E. Multi-dimensional velocity estimation with ultrasound using spatial quadrature. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1998, 45, 852–861. [Google Scholar] [CrossRef]
- Trahey, G.; Smith, S.; von Ramm, O. Speckle Pattern Correlation with Lateral Aperture Translation: Experimental Results and Implications for Spatial Compounding. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1986, 33, 257–264. [Google Scholar] [CrossRef]
- Céspedes, I.; Huang, Y.; Ophir, J.; Spratt, S. Methods for Estimation of Subsample Time Delays of Digitized Echo Signals. Ultrason. Imaging 1995, 17, 142–171. [Google Scholar] [CrossRef]
- VanDecar, J.; Crosson, R. Determination of teleseismic relative phase arrival times using multi-channel cross-correlation and least squares. Bull. Seismol. Soc. Am. 1990, 80, 150–169. [Google Scholar]
- Montaldo, G.; Tanter, M.; Bercoff, J.; Benech, N.; Fink, M. Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2009, 56, 489–506. [Google Scholar] [CrossRef]
- Rodriguez-Molares, A.; Marius, O.; Rindal, H.; Jan, D. The Generalized Contrast-to-Noise ratio: A formal definition for lesion detectability. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2019, 67, 745–759. [Google Scholar] [CrossRef] [Green Version]
- Bottenus, N.; Byram, B.; Hyun, D. Histogram matching for visual ultrasound image comparison. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2020, 68, 1487–1495. [Google Scholar] [CrossRef]
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2023 by the author. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Bottenus, N. Implementation of Constrained Swept Synthetic Aperture Using a Mechanical Fixture. Appl. Sci. 2023, 13, 4797. https://doi.org/10.3390/app13084797
Bottenus N. Implementation of Constrained Swept Synthetic Aperture Using a Mechanical Fixture. Applied Sciences. 2023; 13(8):4797. https://doi.org/10.3390/app13084797
Chicago/Turabian StyleBottenus, Nick. 2023. "Implementation of Constrained Swept Synthetic Aperture Using a Mechanical Fixture" Applied Sciences 13, no. 8: 4797. https://doi.org/10.3390/app13084797
APA StyleBottenus, N. (2023). Implementation of Constrained Swept Synthetic Aperture Using a Mechanical Fixture. Applied Sciences, 13(8), 4797. https://doi.org/10.3390/app13084797