Thermal Analysis of Cornea Heated with Terahertz Radiation

1
Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen 518055, China
2
Biomedical Engineering Laboratory for Photoelectric Sensing Technology, Shenzhen 518055, China
*
Authors to whom correspondence should be addressed.
Appl. Sci. 2019, 9(5), 917; https://doi.org/10.3390/app9050917
Submission received: 30 December 2018 / Revised: 15 February 2019 / Accepted: 28 February 2019 / Published: 4 March 2019
(This article belongs to the Special Issue Terahertz Biomedical Imaging and Detection)

Abstract

:

Featured Application

This work may find potential applications in cornea reshaping in the field of ophthalmology, as well as THz radiation safety and regulation for biomedical imaging or therapy.

Abstract

We numerically investigate the thermal effects in a cornea illuminated by terahertz radiation. By modifying the bioheat and Arrhenius equations, we studied the heat-transfer and temperature distributions in the corneal tissue, and evaluated the potential thermal damage. The influence of the beam radius and power density are discussed. We also estimated the effective cornea-collagen shrinkage region, and evaluated the degree of thermal damage in the cornea. We expect this work to open up a novel effective and safe thermal-treatment approach based on THz radiation for cornea reshaping in the field of ophthalmology.

1. Introduction

The cornea is a key component of the eye’s optical system since it has a curved shape and ensures the refractive function [1]. An effective way to correct refractive disorders of the eye, such as myopia, keratoconus, and hyperopia, is to steepen the precise curved shape of the cornea [2]. Reshaping the cornea can be done through heating, which is known as thermokeratoplasty (TKP), since the cornea stroma permanently shrinks when temperature is raised to over 55 °C [3]. TKP techniques include heating with microwaves [2], preheated probes applied to the surface or inserted in the cornea [4], radio-frequency (RF) currents [5], or laser [6]. Although microwave heating combined with surface cooling can produce a significantly deep region of shrinkage, the shrinkage effect is extremely sensitive to the applicator–cornea gap, making it difficult to implement in the clinic [2]. Preheated probes and RF currents are invasive approaches, so accompanying hazards include infection or even destruction of the cornea [4,5]. For laser-based TKP, the shrinkage effect is not deep enough because of the rather limited heating depth of the laser into the cornea [7]. In other words, there exist many problems for conventional heating techniques, demanding the exploration of novel approaches for the effective and safe thermal treatment of cornea reshaping.
Recently, terahertz (THz) technology has emerged as a novel and promising noninvasive candidate in biomedical applications [8]. A typical application is cornea diagnostics in the field of ophthalmology [9]. Since THz waves are extremely sensitive to water content in tissue and are believed to be harmless for biological entities, THz spectroscopy and imaging techniques have been applied in ex vivo or in vivo corneal hydration sensing [9,10,11,12,13,14]. These studies suggest that THz sensing is an ophthalmological diagnostic tool that deserves further clinical tests. Another emerging application is related to THz-induced bioeffects, which were mediated primarily through photothermal mechanisms, i.e., thermal effects. Kristensen et al. [15] derived the analytical solution for the Kirchoff’s heat equation based on a static (steady-state) model, and showed that heating efficiency is about 40% if THz radiation heats the top layer of the water disc. Wilmink et al. [16] investigated the biological effects associated with human dermal fibroblasts exposed to 2.52 THz radiation. Ganesan et al. [17] calculated the terahertz heating effects in realistic tissue (brain and breast tissue). Zilberti et al. [18] analyzed the transient skin heating induced by terahertz radiation. Bottauscio et al. [19] evaluated the thermal response of human tissue exposed to a focused-beam terahertz electromagnetic radiation by solving the coupled electromagnetic–thermal problem. Encouraged by these studies, quite recently, Smolyanskaya et al. [20] investigated heat transfer of the eye cornea in the THz field. However, their model was relatively simple and thus insufficient to understand the transient behavior of heating effects. Questions remain, such as whether there exists thermal damage, and how THz radiation parameters influence the transient heating effect in the cornea.
In this work, we theoretically investigate the thermal effects in the cornea when heated with terahertz radiation. We develop a model to describe transient temperature distributions and to evaluate the potential thermal-damage effects in the cornea heated with THz ratiation. We also discuss the influence of key parameters, including the radius and power density of the THz beam.

2. Materials and Methods

In order to investigate the time-dependent heat transfer and temperature distributions in the cornea illuminated by terahertz radiation, we developed a two-dimensional (2D) axisymmetric model, as illustrated in Figure 1. This model includes the surrounding fluids, like tears and aqueous humor, similar to the models for thermal effects in the cornea under microwaves [2] or RF heating [5]. A continuous-wave terahertz beam of 3 mm radius and 1 THz frequency impinges onto the center of the cornea. Based on the structure of the human eye, the cornea is treated as a homogeneous and curving layer with 0.4∼0.7 mm thickness [21] and 7.5 mm radius [2,5]; the tear film with a thickness of 5∼15 μ m, and the aqueous humor with a thickness of 2 mm [2,5] were modeled as the major water content.
We made use of the bioheat equation to govern the terahertz bioheat effects in the cornea, which can be written as [22]:
ρ C T t = ( k T ) + q .
Here, ρ is the density of the corneal tissue (kg/m 3 ), C is the specific heat of the corneal tissue(J/(kg·K)), T is the corneal tissue’s local temporal temperature (K), and k is the thermal conductivity of the corneal tissue (W/(m·K)). Heat source q is related to absorbed THz radiation (W/m 3 ), which can be given by the Beer–Lambert law [23]:
q ( z , r ) = α ( 1 R f ) I 0 exp ( α | z | ) .
Here, R f is the reflectivity of THz radiation by the corneal tissue, which is taken to be R f = 0 . 02 @1 THz [11], I 0 is the induced localized THz power intensity ( I 0 ( z , r ) = 2 P π w 0 2 exp ( 2 r 2 w 0 2 ) , P is THz laser power in Watts, w 0 is the radius of the THz beam with Gaussian distribution), α is the absorption coefficient of the biotissue, and { z , r } denotes the coordinates. For convenience in the following discussion (i.e., influence of the THz beam radius and power density), P d represents the power density of the THz beam in Watts per square centimeter ( P d = P π w 0 2 ). Heated by THz radiation, the cornea experiences a rise in temperature, and could suffer irreversible thermal damage if temperature rises above a threshold. According to the Arrhenius formulation, the damage degree can be quantified by a parameter, Ω ( t ) , which is expressed as [24]
Ω ( t ) = ln c ( t 0 ) c ( t ) = k B exp Δ S / R T ( t ) exp ( Δ E R · T ( t ) ) d t .
where c ( t 0 ) is the initial concentration of undamaged corneal-tissue molecules, c ( t ) is the concentration of remaining corneal tissue at time t, k B is Boltzmann’s constant, is Planck’s constant, R = 8.313 J/(mol·K) is the universal gas constant, Δ S = 106 kJ/mol and Δ E = 39 J/(mol·K) are the entropy and the enthalpy of collagen denaturation [22,24], respectively. With Equation (3), the spatial distribution of thermal damage can be obtained.
The bioheat equation, i.e., Equation (1), is solved using the finite-element method (COMSOL). The initial temperature and the temperature at the bottom surface of the aqueous humor were set to be 35 C . We recorded the temperature as the damage degree at each time step. Physical and thermal parameters of the water (tear and aqueous humor) and the cornea [5,22,24] that were used in the model are summarized in Table 1.
Additionally, we used an axisymmetry boundary condition at r = 0 in our model. The effects of heat convection in all interfaces (tear–air, cornea–tear, cornea–aqueous humor) of the model were taken into account using thermal-transfer coefficients. According to published reports [5,22,24], convection boundary conditions at tear–air, cornea–tear, cornea–aqueous humor interfaces were assigned as h t e a r a i r = h W a t e r a i r = 20 W/(m 2 ·K), h c o r n e a t e a r = h W a t e r c o r n e a = 500 W/(m 2 ·K), h c o r n e a h u m o r = h W a t e r c o r n e a = 500 W/(m 2 ·K), respectively. Initially, the tissue was assumed to be at a uniform temperature of 35 C , and the ambient temperature of air (above the cornea) was kept at a constant value of 25 C . Temperature at the bottom surface of the aqueous humor was also kept constant at 35 C .

3. Results and Discussion

3.1. Dynamic Temperature Distributions

We took the terahertz beam radius to be w 0 = 3 mm, and power density to be P d = 0.6 W/cm 2 . Figure 2 shows the time-dependent evolution of temperature distribution in the tear, cornea, and aqueous humor. This shows that the maximum temperature is at the top center of the tear film; temperature gradually decays in both the radial (r) and vertical (z) directions. Since the tear film is very thin, only 10 μ m, the temperature difference between the tear film and the top layer of the cornea is negligible. The temperature in all layers increased very quickly when terahertz heating time was shorter than 60 s, and then gradually saturated. With terahertz heating of 180 s, the maximum temperatures in the tear, cornea, and aqueous humor could reach 70.86, 68.38, and 48.6 C , respectively.
Furthermore, based on Figure 2f, we could also obtain the temperature distribution across the surface beneath the terahertz beam (Figure 3). As expected, it is clearly seen that maximum heating occurs just in the center of where the THz beam is incident. This is because the intensity distribution of the THz beam is hypothetically a Gaussian distribution.

3.2. Influence of THz Beam Radius and Power Density

We now discuss the influence of the terahertz beam radius and power density on the heating effects in the cornea. Figure 4 plots the transient-temperature behaviors at the top center of the cornea layer under different beam sizes and power densities. This shows that, for different beam sizes and power densities, temperature rises very quickly for less than 60 s and then saturates; the higher the power density, or the larger the beam size, the higher the saturated temperature. The difference between saturated temperatures for neighboring power densities is a constant, and the constant linearly increases with beam size. These linear behaviors occur because the heat absorbed by the cornea increases linearly with both power density and beam radius.
In order to ensure collagen shrinkage, Fyodorov et al. [3] reported that the minimal cornea temperature should be 55 C , Haw and Manche [25] suggested the maximal allowed temperature is 75 C , and Asbell et al. [26] reported that the best temperature for optimal shrinkage of the cornea collagen is ∼ 65 C . Moreover, Aksan and McGrath [27] showed that to avoid high temperatures and excessive denaturation is crucial so as to prohibit stiffness decrease of the tissue. Therefore, we can take 55∼ 75 C as the safe shrinkage temperature range, and 65∼70 C as the optimal shrinkage-temperature range. With these ranges, we can identify the suitable power density and beam size for achieving collagen shrinkage. For example, for a beam size of w 0 = 3.5 mm (Figure 4e), the suitable power density should be 0.38∼0.6 W/cm 2 , and optimal power density should be 0.5 W/cm 2 .

3.3. Cornea-Shrinkage-Region Estimation

Since the temperature in the cornea can rise to a certain range for collagen shrinkage, terahertz radiation can be used as a novel corneal thermal therapy. Here, we estimate the effective region of cornea shrinkage based on the above spatial and temporal temperature distributions under THz heating with a beam radius of w 0 = 3 mm and power density of P d = 0.6 W/cm 2 . As we discussed earlier, 55∼ 75 C is the safe shrinkage temperature range. Therefore, we can define the region within this temperature range as the shrinkage region. According to the spatial and temporal temperature distributions, we obtained the time-dependent evolution of the corresponding collagen shrinkage regions, as shown in Figure 5. Results show that the tomography of the collagen-shrinkage region evolves from a funnel-like shape into the cylindrical profile. After temperature saturation, the diameter of the collagen-shrinkage region is about 3 mm, which is comparable to that of the microwave-heating approach (diameter is about 1.3∼2.8 mm [5]).
Given the similarity of heating with THz and infrared (IR) lasers (near-surface heating in both cases), we compared our results with typical IR modeling studies and experiments in the literature [7,23]. The comparison of heating the cornea with THz and typical IR lasers is summarized in Table 2. For IR laser-based heating, the shrinkage effect is not deep enough because of the rather limited heating depth of the laser into the cornea (only ∼50%). Therefore, laser irradiation using THz wavelengths may heat deeper into the cornea than IR wavelengths.
Although reshaping the cornea using IR lasers is well-established, there still exist some problems for IR-heating techniques. According to the comparison summarized in Table 2, IR heating techniques cannot produce enough heating depth (only ∼50%) and a good profile of the shrinkage effect (wedge), which are very important for cornea reshaping based on thermal techniques. However, the THz heating technique can produce a significantly deeper region and a better shrinkage profile. Therefore, we expect this result can clarify the potential benefits of using THz radiation compared with using IR lasers in cornea reshaping in ophthalmology.

3.4. Degree of Thermal Damage

Figure 6 shows the spatial and temporal dependence of the damage degree in the cornea under terahertz heating with a beam radius of w 0 = 3 mm and power density P d = 0.6 W/cm 2 . We found that the spatial distribution of the damage degree was similar to the temperature distribution in Figure 2e. The maximal degree of thermal damage ( Ω = 0.17 ) occurred at the top center of the cornea, consistent with the literature [28]. Because Ω = 0.1 corresponds to the onset of collagen denaturation, and Ω > 1 denotes irreversible thermal damage (quantified in terms of collagenous-tissue denaturation) [24,28], we found that collagen denaturation had already happened since Ω = 0.17 in Figure 6. Therefore, one should take appropriate precautions. Fortunately, the degree of thermal damage is a function of both time and temperature [24]. Thus, we can prevent collagen denaturation by limiting the duration of terahertz heating, i.e., t < 720 s so that Ω < 0.1 throughout the cornea.

4. Concluding Remarks

In summary, we investigated terahertz thermal effects in the cornea. By developing a 2D axisymmetric model based on terahertz bioheat effects, we calculated the time-dependent evolution of temperature distribution in the cornea. We also showed that both the power density and the beam radius of the terahertz beam are important for achieving an appropriate temperature range, within which the cornea can experience collagen shrinkage. Based on the evolution of temperature distribution in the cornea, we evaluated the effective region of collagen shrinkage and the degree of thermal damage. Our results suggest that THz radiation could be used as a potential and promising noncontact heat source that is suitable for the thermal treatment of the cornea in the ophthalmology field. Before this expectation can be realized, further research should also include (i) studying the influence of other parameters, such as THz frequency; (ii) experimentally quantifying the relationship between collagen shrinkage of the cornea and temperature; and (iii) conducting in vitro THz exposure to the cornea to verify the obtained results from the developed model in this work.

Author Contributions

Conceptualization, W.L. and Y.L.; methodology, W.L., Y.L., and G.L.; software, R.S., G.W., G.J., and G.L.; investigation and data analysis, all authors; writing—original-draft preparation, W.L. and G.L.; writing—review and editing, all authors; supervision, Y.L., J.L., and G.L.

Funding

This research was funded by the Shenzhen Research Foundation (Grant Nos. GJHZ20160229201047842, JCYJ20150925163313898, JCYJ20160510154531467, and JCYJ20160608153308846); the National Key Research and Development Program of China (2017YFC0803506); and the Youth Innovation Promotion Association, CAS (No. 20160320).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Maurice, D.M. The structure and transparency of the cornea. J. Physiol. 1957, 136, 263–286. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Trembly, B.S.; Keates, R.H. Combined microwave heating and surface cooling of the cornea. IEEE Trans. Biomed. Eng. 1991, 38, 85–91. [Google Scholar] [CrossRef] [PubMed]
  3. Fyodorov, S.N.; Durnev, V.V. Operation of dosaged dissection of corneal circular ligament in cases of myopia of mild degree. Ann. Ophthalmol. 1979, 441, 1885–1890. [Google Scholar]
  4. Neumann, A.C.; Fyodorov, S.; Sanders, D.R. Radial thermokeratoplasty for the correction of hyperopia. Refract. Corneal Surg. 1990, 6, 404–412. [Google Scholar] [PubMed]
  5. Berjano, E.J.; Saiz, J.; Ferrero, J.M. Radio-Frequency Heating of the Cornea: Theoretical Model and In Vitro Experiments. IEEE Trans. Biomed. Eng. 2002, 49, 196–205. [Google Scholar] [CrossRef] [PubMed]
  6. Seiler, T.; Matallana, M.; Bende, T. Laser thermokeratoplasty by means of a pulsed Holmium:YAG Laser for the hyperopic correction. Refrac. Corneal Surg. 1990, 6, 335–339. [Google Scholar]
  7. Thompson, V.; Seiler, T.; Durrie, D.S.; Cavanaugh, T.B. Holmium:YAG laser thermokeratoplasty for hyperopia and astigmatism: An overview. Refract. Corneal Surg. 1993, 9, S134–S137. [Google Scholar] [PubMed]
  8. Smolyanskaya, O.A.; Chernomyrdin, N.V.; Konovko, A.A.; Zaytsev, K.I.; Ozheredov, I.A.; Cherkasova, O.P.; Nazarov, M.M.; Guillet, J.P.; Kozlov, S.A.; Kistenev, Y.V.; et al. Terahertz biophotonics as a tool for studies of dielectric and spectral properties of biological tissues and liquids. Prog. Quantum Electron. 2018, 62, 1–77. [Google Scholar] [CrossRef]
  9. Taylor, Z.D.; Garritano, J.; Sung, S.; Bajwa, N.; Bennett, D.B.; Nowroozi, B.; Tewari, P.; Sayre, J.W.; Hubschman, J.P.; Deng, S.X.; et al. THz and mm-Wave Sensing of Corneal Tissue Water Content: In Vivo Sensing and Imaging Results. IEEE Trans. Terahertz Sci. Technol. 2015, 5, 184–196. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  10. Liu, W.Q.; Lu, Y.F.; Jiao, G.H.; Chen, X.F.; Zhou, Z.S.; She, R.B.; Li, J.Y.; Chen, S.H.; Dong, Y.M.; Lv, J.C. Spectroscopic measurement and terahertz imaging of the cornea using a rapid scanning terahertz time domain spectrometer. Chin. Phys. B 2016, 25, 060702. [Google Scholar] [CrossRef]
  11. Bennett, D.B.; Taylor, Z.; Tewari, P.; Singh, R.S.; Grundfest, M.O.C.W.S.; Sassoon, D.J.; Johnson, R.D.; Hubschman, J.P.; Brown, E.R. Terahertz sensing in corneal tissues. J. Biomed. Opt. 2011, 16, 057003. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  12. Ozheredov, I.; Prokopchuk, M.; Mischenko, M.; Saphonova, T.; Solyankin, P.; Larichev, A.; Angeluts, A.; Balakin, A.; Shkurinov, A. In vivo THz sensing of eye cornea. Laser Phys. Lett. 2018, 15, 055601. [Google Scholar] [CrossRef]
  13. Sung, S.; Selvin, S.; Bajwa, N.; Chantra, S.; Nowroozi, B.; Garritano, J.; Goell, J.; Li, A.D.; Deng, S.X.; Brown, E.R.; et al. Optical System Design for Noncontact, Normal Incidence, THz Imaging of in vivo Human Cornea. IEEE Trans. Terahertz Sci. Technol. 2018, 8, 1–12. [Google Scholar] [CrossRef] [PubMed]
  14. Sung, S.; Selvin, S.; Bajwa, N.; Chantra, S.; Nowroozi, B.; Garritano, J.; Goell, J.; Li, A.D.; Deng, S.X.; Brown, E.R.; et al. THz Imaging System for in vivo Human Cornea. IEEE Trans. Terahertz Sci. Technol. 2018, 8, 27–37. [Google Scholar] [CrossRef] [PubMed]
  15. Kristensen, T.; Withayachumnankul, W.; Jepsen, P.U.; Abbott, D. Modeling terahertz heating effects on water. Opt. Express 2010, 18, 4727–4739. [Google Scholar] [CrossRef] [PubMed]
  16. Wilmink, G.J.; Rivest, B.D.; Roth, C.C.; Ibey, B.L.; Payne, J.A.; Cundin, L.X.; Grundt, J.E.; Peralta, X.; Mixon, D.G.; Roach, W.P. In vitro investigation of the biological effects associated with human dermal fibroblasts exposed to 2.52 THz radiation. Laser Surg. Med. 2011, 43, 152–163. [Google Scholar] [CrossRef] [PubMed]
  17. Ganesan, S.; Yalavarthy, P.K. Modeling of Terahertz Heating Effects in Realistic Tissues. IEEE J. Sel. Top. Quantum Electron. 2013, 19, 8400908. [Google Scholar] [CrossRef]
  18. Zilberti, L.; Arduino, A.; Bottauscio, O.; Chiampi, M. Parametric analysis of transient skin heating induced by terahertz radiation. Bioelectromagnetics 2014, 35, 314–323. [Google Scholar] [CrossRef] [PubMed]
  19. Bottauscio, O.; Chiampi, M.; Zilberti, L. Thermal Analysis of Human Tissues Exposed to Focused Beam THz Radiations. IEEE Trans. Mag. 2015, 51, 7400504. [Google Scholar] [CrossRef]
  20. Smolyanskaya, O.A.; Odlyanitskiy, E.L.; Chivilikchin, S.A.; Schelkanova, I.J.; Kozlov, S.A. Theoretical and experimental investigations of the heat transfer of eye cornea in terahertz field. In Proceedings of the 42nd International Conference on Infrared, Millimeter, and Terahertz Waves (IRMMW-THz), Cancun, Mexico, 27 August–1 September 2017; pp. 1–2. [Google Scholar]
  21. Taylor, Z.D.; Garritano, J.; Sung, S.; Bajwa, N.; Bennett, D.B.; Nowroozi, B.; Tewari, P.; Sayre, J.W.; Hubschman, J.P.; Deng, S.X.; et al. THz and mm-wave sensing of corneal tissue water content: Electromagnetic Modeling and Analysis. IEEE Trans. Terahertz Sci. Technol. 2015, 5, 170–183. [Google Scholar] [CrossRef] [PubMed]
  22. Kim, B.; Jacques, S.L.; Rastegar, S.; Thomsen, S.L.; Motamedi, M. Nonlinear finite-element analysis of the role of dynamic changes in blood perfusion and optical properties in laser coagulation of tissue. IEEE J. Sel. Top. Quantum Electron. 1996, 2, 922–933. [Google Scholar]
  23. Shibib, K.S. Finite element analysis of cornea thermal damage due to pulse incidental far IR laser. Lasers Med. Sci. 2013, 28, 871–877. [Google Scholar] [CrossRef] [PubMed]
  24. Jo, B.; Aksan, A. Prediction of the extent of thermal damage in the cornea during conductive keratoplasty. J. Therm. Biol. 2010, 35, 167–174. [Google Scholar] [CrossRef]
  25. Haw, W.W.; Manche, E.E. Conductive keratoplasty and laser thermal keratoplasty. Int. Ophthalmol. Clin. 2002, 42, 99–106. [Google Scholar] [CrossRef] [PubMed]
  26. Asbell, P.A.; Maloney, R.K.; Davidorf, J.M.; Hersh, P.S.; Mcdonald, M.B.; Manche, E.E. Conductive keratoplasty for the correction of hyperopia. Trans. Am. Ophthalmol. Soc. 2001, 99, 79–87. [Google Scholar] [PubMed]
  27. Aksan, A.; McGrath, J.J. Thermomechanical analysis of soft tissue thermo-therapy. J. Biomech. Eng. 2003, 123, 700–708. [Google Scholar] [CrossRef]
  28. Naoumidi, T.L.; Pallikaris, I.G.; Naoumidi, I.; Astyrakakis, N.I. Conductive Keratoplasty: Histological Study of Human Corneas. Am. J. Ophthalmol. 2005, 140, 984–992. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Schematic diagram of THz thermal treatment of the cornea (out of scale).
Figure 1. Schematic diagram of THz thermal treatment of the cornea (out of scale).
Applsci 09 00917 g001
Figure 2. Time-dependent evolution of temperature distributions under terahertz heating with beam radius w 0 = 3 mm, power density P d = 0.6 W/cm 2 , and exposure time of (a) 10 s; (b) 30 s; (c) 60 s; (d) 180 s; (e) 300 s; and (f) 1200 s.
Figure 2. Time-dependent evolution of temperature distributions under terahertz heating with beam radius w 0 = 3 mm, power density P d = 0.6 W/cm 2 , and exposure time of (a) 10 s; (b) 30 s; (c) 60 s; (d) 180 s; (e) 300 s; and (f) 1200 s.
Applsci 09 00917 g002
Figure 3. Temperature distribution across the surface beneath the THz beam (beam radius w 0 = 3 mm, power density P d = 0.6 W/cm 2 , and exposure time 1200 s).
Figure 3. Temperature distribution across the surface beneath the THz beam (beam radius w 0 = 3 mm, power density P d = 0.6 W/cm 2 , and exposure time 1200 s).
Applsci 09 00917 g003
Figure 4. Time-dependent temperature evolution of at the top center of the cornea layer under different power densities (0.2∼1 W/cm 2 as marked in the figure) and terahertz beam radius of (a) w 0 = 1.5 mm; (b) w 0 = 2.0 mm; (c) w 0 = 2.5 mm; (d) w 0 = 3.0 mm; (e) w 0 = 3.5 mm; and (f) w 0 = 4.0 mm.
Figure 4. Time-dependent temperature evolution of at the top center of the cornea layer under different power densities (0.2∼1 W/cm 2 as marked in the figure) and terahertz beam radius of (a) w 0 = 1.5 mm; (b) w 0 = 2.0 mm; (c) w 0 = 2.5 mm; (d) w 0 = 3.0 mm; (e) w 0 = 3.5 mm; and (f) w 0 = 4.0 mm.
Applsci 09 00917 g004
Figure 5. Time-dependent evolution of collagen-shrinkage tomography under terahertz heating with beam radius w 0 = 3 mm, power density P d = 0.6 W/cm 2 , and exposure time of (a) 10 s; (b) 30 s; (c) 60 s; (d) 180 s; (e) 300 s; and (f) 1200 s. Blue regions indicate nonshrinkage regions, and red regions the shrinkage regions. Shrinkage threshold value was taken to be 55 C .
Figure 5. Time-dependent evolution of collagen-shrinkage tomography under terahertz heating with beam radius w 0 = 3 mm, power density P d = 0.6 W/cm 2 , and exposure time of (a) 10 s; (b) 30 s; (c) 60 s; (d) 180 s; (e) 300 s; and (f) 1200 s. Blue regions indicate nonshrinkage regions, and red regions the shrinkage regions. Shrinkage threshold value was taken to be 55 C .
Applsci 09 00917 g005
Figure 6. Time-dependent degree of damage at the top center of the cornea. The inset shows spatial distribution in the cornea at t = 1200 s. The results were calculated under THz radiation with a beam radius of w 0 = 3 mm and power density of P d = 0.6 W/cm 2 .
Figure 6. Time-dependent degree of damage at the top center of the cornea. The inset shows spatial distribution in the cornea at t = 1200 s. The results were calculated under THz radiation with a beam radius of w 0 = 3 mm and power density of P d = 0.6 W/cm 2 .
Applsci 09 00917 g006
Table 1. Physical and thermal parameters of tear and aqueous humor (modelled as water), and the cornea [5,22,24] used in the model.
Table 1. Physical and thermal parameters of tear and aqueous humor (modelled as water), and the cornea [5,22,24] used in the model.
PropertyMaterialValueMaterialValue
Density ρ (kg/m 3 )Water1000Cornea1060
Thermal conductivity k (W/m/K)Water0.578Cornea0.556
Specific heat C (J/kg/K)Water4180Cornea3830
Absorption coefficient 1 α (1/m)Water24,067Cornea14,400
Convection coefficient h (W/m 2 /K)Water-air20Water-cornea500
Film thickness t ( μ m)Tear film10Cornea600
1 Absorption coefficient is at a frequency of 1 THz.
Table 2. Comparison of heating cornea with THz and typical infrared (IR) lasers.
Table 2. Comparison of heating cornea with THz and typical infrared (IR) lasers.
THz Laser (This Work)Near IR Laser [7]Far IR Laser [23]
MethodTheoryExperimentTheory
Laser wavelength ( μ m)3002.1010.6
Depth 1 ( μ m)600300–400<100
Profile 2 cylinderwedge
Max. temp.68.38 C 56.5 C
1 Depth of shrinkage effect in experiment, depth of temperature ≥55 C in theory; 2 Profile of shrinkage effect in experiment, profile of temperature ≥55 C in theory.

Share and Cite

MDPI and ACS Style

Liu, W.; Lu, Y.; She, R.; Wei, G.; Jiao, G.; Lv, J.; Li, G. Thermal Analysis of Cornea Heated with Terahertz Radiation. Appl. Sci. 2019, 9, 917. https://doi.org/10.3390/app9050917

AMA Style

Liu W, Lu Y, She R, Wei G, Jiao G, Lv J, Li G. Thermal Analysis of Cornea Heated with Terahertz Radiation. Applied Sciences. 2019; 9(5):917. https://doi.org/10.3390/app9050917

Chicago/Turabian Style

Liu, Wenquan, Yuanfu Lu, Rongbin She, Guanglu Wei, Guohua Jiao, Jiancheng Lv, and Guangyuan Li. 2019. "Thermal Analysis of Cornea Heated with Terahertz Radiation" Applied Sciences 9, no. 5: 917. https://doi.org/10.3390/app9050917

APA Style

Liu, W., Lu, Y., She, R., Wei, G., Jiao, G., Lv, J., & Li, G. (2019). Thermal Analysis of Cornea Heated with Terahertz Radiation. Applied Sciences, 9(5), 917. https://doi.org/10.3390/app9050917

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop