Pulmonary Vein Isolation with Pulsed Field Ablation and Size-Adjustable Cryo-Balloon: A Comparative Procedural Analysis of First-Time Use
Abstract
:1. Introduction
2. Materials and Methods
2.1. Study Population
2.2. Periprocedural Management, Ablation Procedure and Postprocedural Management
2.2.1. Pulsed Field Ablation Protocol
2.2.2. Size-Adjustable Cryo-Balloon Ablation Protocol
2.3. Statistical Analysis
3. Results
3.1. Patient Characteristics
3.2. Procedural Characteristics
3.3. Safety
4. Discussion
4.1. Efficiency
4.2. Safety
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Boersma, L.; Andrade, J.G.; Betts, T.; Duytschaever, M.; Pürerfellner, H.; Santoro, F.; Tzeis, S.; Verma, A. Progress in Atrial Fibrillation Ablation during 25 Years of Europace Journal. Europace 2023, 25, euad244. [Google Scholar] [CrossRef] [PubMed]
- Chauhan, V.S.; Krahn, A.D.; Klein, G.J.; Skanes, A.C.; Yee, R. Supraventricular tachycardia. Med. Clin. N. Am. 2001, 85, 192–223. [Google Scholar]
- Sagris, M.; Vardas, E.P.; Theofilis, P.; Antonopoulos, A.S.; Oikonomou, E.; Tousoulis, D. Atrial Fibrillation: Pathogenesis, Predisposing Factors, and Genetics. Int. J. Mol. Sci. 2021, 23, 6. [Google Scholar] [CrossRef] [PubMed]
- Hindricks, G.; Potpara, T.; Dagres, N.; Arbelo, E.; Bax, J.J.; Blomström-Lundqvist, C.; Boriani, G.; Castella, M.; Dan, G.-A.; Dilaveris, P.E.; et al. 2020 ESC Guidelines for the Diagnosis and Management of Atrial Fibrillation Developed in Collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur. Heart J. 2021, 42, 373–498. [Google Scholar] [CrossRef] [PubMed]
- Joglar, J.A.; Chung, M.K.; Armbruster, A.L.; Benjamin, E.J.; Chyou, J.Y.; Cronin, E.M.; Deswal, A.; Eckhardt, L.L.; Goldberger, Z.D.; Gopinathannair, R.; et al. 2023 ACC/AHA/ACCP/HRS Guideline for the Diagnosis and Management of Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 2024, 149, e1–e156. [Google Scholar] [CrossRef] [PubMed]
- Tzeis, S.; Gerstenfeld, E.P.; Kalman, J.; Saad, E.B.; Sepehri Shamloo, A.; Andrade, J.G.; Barbhaiya, C.R.; Baykaner, T.; Boveda, S.; Calkins, H.; et al. 2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation. Europace 2024, 26, euae043. [Google Scholar] [CrossRef] [PubMed]
- Parameswaran, R.; Al-Kaisey, A.M.; Kalman, J.M. Catheter Ablation for Atrial Fibrillation: Current Indications and Evolving Technologies. Nat. Rev. Cardiol. 2021, 18, 210–225. [Google Scholar] [CrossRef]
- Haïssaguerre, M.; Jaïs, P.; Shah, D.C.; Takahashi, A.; Hocini, M.; Quiniou, G.; Garrigue, S.; Le Mouroux, A.; Le Métayer, P.; Clémenty, J. Spontaneous Initiation of Atrial Fibrillation by Ectopic Beats Originating in the Pulmonary Veins. N. Engl. J. Med. 1998, 339, 659–666. [Google Scholar] [CrossRef] [PubMed]
- Reddy, V.Y.; Gerstenfeld, E.P.; Natale, A.; Whang, W.; Cuoco, F.A.; Patel, C.; Mountantonakis, S.E.; Gibson, D.N.; Harding, J.D.; Ellis, C.R.; et al. Pulsed Field or Conventional Thermal Ablation for Paroxysmal Atrial Fibrillation. N. Engl. J. Med. 2023, 389, 1660–1671. [Google Scholar] [CrossRef]
- Teumer, Y.; Hilgarth, F.; Katov, L.; Melnic, R.; Rottbauer, W.; Bothner, C.; Weinmann, K. Pulmonary Vein Isolation with a Novel Size-Adjustable Cryo-Balloon Catheter: A Tailored Ablation Protocol. J. Clin. Med. 2024, 13, 2262. [Google Scholar] [CrossRef]
- Weinmann, K.; Heudorfer, R.; Lenz, A.; Aktolga, D.; Rattka, M.; Bothner, C.; Pott, A.; Öchsner, W.; Rottbauer, W.; Dahme, T. Safety of Conscious Sedation in Electroanatomical Mapping Procedures and Cryoballoon Pulmonary Vein Isolation. Heart Vessel. 2021, 36, 561–567. [Google Scholar] [CrossRef] [PubMed]
- Katov, L.; Teumer, Y.; Lederbogen, K.; Melnic, R.; Rottbauer, W.; Bothner, C.; Weinmann-Emhardt, K. Transesophageal Echocardiography Improves Precision in Transseptal Puncture Compared to Fluoroscopy in Left Atrial Electrophysiological Procedures. J. Clin. Med. 2024, 13, 2476. [Google Scholar] [CrossRef] [PubMed]
- Lemoine, M.D.; Fink, T.; Mencke, C.; Schleberger, R.; My, I.; Obergassel, J.; Bergau, L.; Sciacca, V.; Rottner, L.; Moser, J.; et al. Pulsed-Field Ablation-Based Pulmonary Vein Isolation: Acute Safety, Efficacy and Short-Term Follow-up in a Multi-Center Real World Scenario. Clin. Res. Cardiol. 2023, 112, 795–806. [Google Scholar] [CrossRef] [PubMed]
- Badertscher, P.; Weidlich, S.; Knecht, S.; Stauffer, N.; Krisai, P.; Voellmin, G.; Osswald, S.; Sticherling, C.; Kühne, M. Efficacy and Safety of Pulmonary Vein Isolation with Pulsed Field Ablation vs. Novel Cryoballoon Ablation System for Atrial Fibrillation. Europace 2023, 25, euad329. [Google Scholar] [CrossRef]
- Della Rocca, D.G.; Marcon, L.; Magnocavallo, M.; Menè, R.; Pannone, L.; Mohanty, S.; Sousonis, V.; Sorgente, A.; Almorad, A.; Bisignani, A.; et al. Pulsed Electric Field, Cryoballoon, and Radiofrequency for Paroxysmal Atrial Fibrillation Ablation: A Propensity Score-Matched Comparison. Europace 2023, 26, euae016. [Google Scholar] [CrossRef]
- Ekanem, E.; Reddy, V.Y.; Schmidt, B.; Reichlin, T.; Neven, K.; Metzner, A.; Hansen, J.; Blaauw, Y.; Maury, P.; Arentz, T.; et al. Multi-National Survey on the Methods, Efficacy, and Safety on the Post-Approval Clinical Use of Pulsed Field Ablation (MANIFEST-PF). Europace 2022, 24, 1256–1266. [Google Scholar] [CrossRef]
- Turagam, M.K.; Neuzil, P.; Schmidt, B.; Reichlin, T.; Neven, K.; Metzner, A.; Hansen, J.; Blaauw, Y.; Maury, P.; Arentz, T.; et al. Safety and Effectiveness of Pulsed Field Ablation to Treat Atrial Fibrillation: One-Year Outcomes from the MANIFEST-PF Registry. Circulation 2023, 148, 35–46. [Google Scholar] [CrossRef]
- Metzner, A.; Fiala, M.; Vijgen, J.; Ouss, A.; Gunawardene, M.; Hansen, J.; Kautzner, J.; Schmidt, B.; Duytschaever, M.; Reichlin, T.; et al. Long-Term Outcomes of the Pentaspline Pulsed-Field Ablation Catheter for the Treatment of Paroxysmal Atrial Fibrillation: Results of the Prospective, Multicentre FARA-Freedom Study. Europace 2024, 26, euae053. [Google Scholar] [CrossRef]
- Ellenbogen, K.A.; Mittal, S.; Varma, N.; Aryana, A.; Marrouche, N.; Anić, A.; Nair, D.; Champagne, J.; Iacopino, S.; De Asmundis, C.; et al. One-year Outcomes of Pulmonary Vein Isolation with a Novel Cryoballoon: Primary Results of the FROZEN AF Trial. J. Cardiovasc. Electrophysiol. 2024, 35, 832–842. [Google Scholar] [CrossRef]
- Tanese, N.; Almorad, A.; Pannone, L.; Defaye, P.; Jacob, S.; Kilani, M.B.; Chierchia, G.; Venier, S.; Cardin, C.; Jacon, P.; et al. Outcomes after Cryoballoon Ablation of Paroxysmal Atrial Fibrillation with the PolarX or the Arctic Front Advance Pro: A Prospective Multicentre Experience. Europace 2023, 25, 873–879. [Google Scholar] [CrossRef]
- Howard, B.; Verma, A.; Tzou, W.S.; Mattison, L.; Kos, B.; Miklavčič, D.; Onal, B.; Stewart, M.T.; Sigg, D.C. Effects of Electrode-Tissue Proximity on Cardiac Lesion Formation Using Pulsed Field Ablation. Circ. Arrhythmia Electrophysiol. 2022, 15, e011110. [Google Scholar] [CrossRef]
- Moser, F.; Rottner, L.; Moser, J.; Schleberger, R.; Lemoine, M.; Münkler, P.; Dinshaw, L.; Kirchhof, P.; Reissmann, B.; Ouyang, F.; et al. The Established and the Challenger: A Direct Comparison of Current Cryoballoon Technologies for Pulmonary Vein Isolation. J. Cardiovasc. Electrophysiol. 2022, 33, 48–54. [Google Scholar] [CrossRef]
- Heeger, C.-H.; Schuette, C.; Seitelberger, V.; Wissner, E.; Rillig, A.; Mathew, S.; Reissmann, B.; Lemes, C.; Maurer, T.; Fink, T.; et al. Time-to-Effect Guided Pulmonary Vein Isolation Utilizing the Third-Generation versus Second Generation Cryoballoon: One Year Clinical Success. Cardiol. J. 2019, 26, 368–374. [Google Scholar] [CrossRef] [PubMed]
- Bradley, C.J.; Haines, D.E. Pulsed Field Ablation for Pulmonary Vein Isolation in the Treatment of Atrial Fibrillation. J. Cardiovasc. Electrophysiol. 2020, 31, 2136–2147. [Google Scholar] [CrossRef] [PubMed]
- Heeger, C.-H.; Pott, A.; Sohns, C.; Riesinger, L.; Sommer, P.; Gasperetti, A.; Tondo, C.; Fassini, G.; Moser, F.; Lucas, P.; et al. Novel Cryoballoon Ablation System for Pulmonary Vein Isolation: Multicenter Assessment of Efficacy and Safety—ANTARCTICA Study. Europace 2022, 24, 1917–1925. [Google Scholar] [CrossRef]
- Hoffmann, E.; Straube, F.; Wegscheider, K.; Kuniss, M.; Andresen, D.; Wu, L.-Q.; Tebbenjohanns, J.; Noelker, G.; Tilz, R.R.; Chun, J.K.R.; et al. Outcomes of Cryoballoon or Radiofrequency Ablation in Symptomatic Paroxysmal or Persistent Atrial Fibrillation. Europace 2019, 21, 1313–1324. [Google Scholar] [CrossRef]
- Heeger, C.-H.; Popescu, S.Ș.; Sohns, C.; Pott, A.; Metzner, A.; Inaba, O.; Straube, F.; Kuniss, M.; Aryana, A.; Miyazaki, S.; et al. Impact of Cryoballoon Application Abortion Due to Phrenic Nerve Injury on Reconnection Rates: A YETI Subgroup Analysis. Europace 2023, 25, 374–381. [Google Scholar] [CrossRef] [PubMed]
- Duytschaever, M.; De Potter, T.; Grimaldi, M.; Anic, A.; Vijgen, J.; Neuzil, P.; Van Herendael, H.; Verma, A.; Skanes, A.; Scherr, D.; et al. Paroxysmal Atrial Fibrillation Ablation Using a Novel Variable-Loop Biphasic Pulsed Field Ablation Catheter Integrated with a 3-Dimensional Mapping System: 1-Year Outcomes of the Multicenter inspIRE Study. Circ. Arrhythmia Electrophysiol. 2023, 16, e011780. [Google Scholar] [CrossRef]
- Verma, A.; Haines, D.E.; Boersma, L.V.; Sood, N.; Natale, A.; Marchlinski, F.E.; Calkins, H.; Sanders, P.; Packer, D.L.; Kuck, K.-H.; et al. Pulsed Field Ablation for the Treatment of Atrial Fibrillation: PULSED AF Pivotal Trial. Circulation 2023, 147, 1422–1432. [Google Scholar] [CrossRef]
- Tilz, R.R.; Vogler, J.; Kirstein, B.; Eitel, C.; Phan, H.-L.; Popescu, S.Ș.; Hatahet, S.; Große, N.; Subin, B.; Lopez, L.D.; et al. Pulsed Field Ablation-Based Pulmonary Vein Isolation Using a Simplified Single-Access Single-Catheter Approach—The Fast and Furious PFA Study. Circ. J. 2023, 87, 1722–1726. [Google Scholar] [CrossRef]
- Mansour, M.; Gerstenfeld, E.P.; Patel, C.; Natale, A.; Whang, W.; Cuoco, F.A.; Mountantonakis, S.E.; Gibson, D.N.; Harding, J.D.; Holland, S.K.; et al. Pulmonary Vein Narrowing after Pulsed Field versus Thermal Ablation. Europace 2024, 26, euae038. [Google Scholar] [CrossRef]
- Heeger, C.-H.; Sohns, C.; Pott, A.; Metzner, A.; Inaba, O.; Straube, F.; Kuniss, M.; Aryana, A.; Miyazaki, S.; Cay, S.; et al. Phrenic Nerve Injury during Cryoballoon-Based Pulmonary Vein Isolation: Results of the Worldwide YETI Registry. Circ. Arrhythmia Electrophysiol. 2022, 15, e010516. [Google Scholar] [CrossRef]
System Characteristics | Pulsed Field Ablation | Size-Adjustable Cryo-Balloon |
---|---|---|
Mechanism of ablation | Selective cell electroporation | Unselective freeze |
Energy form | Non-thermal | Thermal |
Device diameter | 31 mm or 35 mm 1 | 28 mm–31 mm 2 |
Device configuration during energy delivery | ‘basket’ and ‘flower’, as well as every position in between (if necessary) | Oval |
Energy delivery | At least 8 applications per vein with duration 2.5 s each | At least 1 application per vein with duration 180 s (if TTI < 30 s 120 s) |
Signals monitoring in the pulmonary veins | 20 electrodes | 8 electrodes |
Baseline Characteristics | All Patients (n = 100) | Pulsed Field Ablation (n = 50) | Size-Adjustable Cryo-Balloon (n = 50) | p-Value |
---|---|---|---|---|
Age [years], median (IQR) | 71 (64; 78) | 70 (62; 77) | 72 (65; 79) | 0.546 |
Female, n (%) | 46 (46.0) | 22 (44.0) | 24 (48.0) | 0.688 |
BMI [kg/m2], median (IQR) | 27.3 (24.2; 30.4) | 27.6 (27.6; 30.4) | 26.7 (23.9; 30.8) | 0.804 |
CHA2DS2-VASc score, median (IQR) | 4 (2; 5) | 4 (2; 5) | 4 (2; 4) | 0.283 |
LA diameter [mm], median (IQR) | 45 (40; 50) | 45 (40; 50) | 45 (42; 50) | 0.095 |
LAVI [mL/m2], median (IQR) | 39.8 (29.8; 49.4) | 39.4 (29.7; 49.9) | 37.0 (27.6; 48.2) | 0.382 |
LVEF (%), median (IQR) | 59 (48.0; 65.0) | 59 (48.0; 65.0) | 59 (46.0; 65.0) | 0.670 |
Hypertension, n (%) | 77 (77.0) | 40 (80.0) | 37 (74.0) | 0.476 |
Diabetes mellitus, n (%) | 16 (16.0) | 11 (22.0) | 5 (10.0) | 0.102 |
Hyperlipoproteinemia, n (%) | 79 (79.0) | 38 (76.0) | 41 (82.0) | 0.461 |
Coronary artery disease, n (%) | 49 (49.0) | 24 (48.0) | 25 (50.0) | 0.841 |
OSA, n (%) | 6 (6.0) | 2 (4.0) | 4 (8.0) | 0.400 |
Prior Stroke/ TIA, n (%) | 10 (10.0) | 4 (8.0) | 6 (12.0) | 0.452 |
Ablation Parameters | Pulsed Field Ablation (n = 50) | Size-Adjustable Cryo-Balloon (n = 50) | p-Value |
---|---|---|---|
Acute PVI, n (%) | 50 (100) | 50 (100) | 1.000 |
FPI LSPV, n (%) | 44 (95.6) | 41 (91.1) | 0.434 |
FPI LIPV, n (%) | 46 (100) | 40 (88.9) | 0.026 |
FPI RIPV, n (%) | 50 (100) | 43 (86.0) | 0.012 |
FPI RSPV, n (%) | 50 (100) | 43 (86.0) | 0.012 |
FPI LCPV, n (%) | 4 (100) | 3 (60.0) | 0.444 |
FPI RMPV, n (%) | N/A | 1 (100) | N/A |
ED per patient, median (IQR) | 32.0 (32.0; 32.0) | 6.0 (6.0; 8.0) | N/A |
Procedure Characteristics | Pulsed Field Ablation (n = 50) | Size-Adjustable Cryo-Balloon (n = 50) | p-Value |
---|---|---|---|
Procedure duration [min], median (IQR) | 59.0 (47.7; 68.5) | 76.5 (64.0; 95.5) | < 0.001 |
Dwell time [min], median (IQR) | 32.5 (25.0; 44.2) | 41.0 (32.0; 55.0) | 0.002 |
Fluoroscopy time [min], median (IQR) | 18.5 (16.0; 23.2) | 17.7 (12.5; 22.0) | 0.188 |
Fluoroscopy dose [cGy·cm2], median (IQR) | 1170.4 (831.2; 1852.6) | 2364.8 (1376.8; 4566.9) | < 0.001 |
Contrast agent [mL], median (IQR) | 30.0 (30.0; 90.0) | 90.0 (30.0; 140.0) | 0.003 |
PFA Procedure Characteristics | First Examinations | Last Examinations | p-Value |
---|---|---|---|
Procedure duration [min], median (IQR) | 64.0 (57.0; 74.0) | 58.0 (38.0; 75.0) | 0.112 |
Dwell time [min], median (IQR) | 40.0 (32.0; 50.0) | 34.0 (22.0; 45.0) | 0.268 |
Fluoroscopy time [min], median (IQR) | 22.3 (18.6; 27.0) | 18.0 (13.1; 23.6) | 0.023 |
Fluoroscopy dose [cGy*cm2], median (IQR) | 1630.0 (973.0; 2062.0) | 882.0 (611.0; 1144.4) | 0.085 |
Contrast agent [mL], median (IQR) | 30.0 (30.0; 90.0) | 30.0 (30.0; 42.5) | 0.906 |
Size-Adjustable Cryo-Balloon Procedure Characteristics | First Examinations | Last Examinations | p-Value |
---|---|---|---|
Procedure duration [mins], median (IQR) | 80.0 (75.0; 110.0) | 80.0 (55.0; 95.0) | 0.345 |
Dwell time [mins], median (IQR) | 45.5 (38.0; 63.7) | 41.0 (28.0; 56.0) | 0.505 |
Fluoroscopy time [min], median (IQR) | 21.1 (19.5; 29.8) | 17.8 (12.7; 22.1) | 0.137 |
Fluoroscopy dose [cGy*cm2], median (IQR) | 2577.6 (1869.0; 6570.6) | 1735.2 (1301.0; 4613.0) | 0.367 |
Contrast agent [mL], median (IQR) | 140.0 (90.0; 140.0) | 40.0 (35.0; 90.0) | 0.013 |
Complications | Pulsed Field Ablation | Size-Adjustable Cryo-Balloon | p-Value |
---|---|---|---|
Transient phrenic nerve palsy, n (%) | 0 (0) | 2 (4.0) | |
Persistent phrenic nerve palsy, n (%) | 0 (0) | 0 (0) | |
Pericarditis, n (%) | 0 (0) | 1 (2.0) | |
Pericardial effusion/ tamponade, n (%) | 0 (0) | 0 (0) | 0.213 |
Vascular access complications, n (%) | 0 (0) | 0 (0) | |
Air embolism, n (%) | 0 (0) | 0 (0) | |
Strokes/transient ischemic attacks, n (%) | 0 (0) | 0 (0) |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Katov, L.; Teumer, Y.; Bothner, C.; Rottbauer, W.; Weinmann-Emhardt, K. Pulmonary Vein Isolation with Pulsed Field Ablation and Size-Adjustable Cryo-Balloon: A Comparative Procedural Analysis of First-Time Use. J. Clin. Med. 2024, 13, 3113. https://doi.org/10.3390/jcm13113113
Katov L, Teumer Y, Bothner C, Rottbauer W, Weinmann-Emhardt K. Pulmonary Vein Isolation with Pulsed Field Ablation and Size-Adjustable Cryo-Balloon: A Comparative Procedural Analysis of First-Time Use. Journal of Clinical Medicine. 2024; 13(11):3113. https://doi.org/10.3390/jcm13113113
Chicago/Turabian StyleKatov, Lyuboslav, Yannick Teumer, Carlo Bothner, Wolfgang Rottbauer, and Karolina Weinmann-Emhardt. 2024. "Pulmonary Vein Isolation with Pulsed Field Ablation and Size-Adjustable Cryo-Balloon: A Comparative Procedural Analysis of First-Time Use" Journal of Clinical Medicine 13, no. 11: 3113. https://doi.org/10.3390/jcm13113113
APA StyleKatov, L., Teumer, Y., Bothner, C., Rottbauer, W., & Weinmann-Emhardt, K. (2024). Pulmonary Vein Isolation with Pulsed Field Ablation and Size-Adjustable Cryo-Balloon: A Comparative Procedural Analysis of First-Time Use. Journal of Clinical Medicine, 13(11), 3113. https://doi.org/10.3390/jcm13113113