1. Introduction
Since 1960, microsurgery has revolutionized the world of reconstructive surgery, by allowing surgeons to perform sophisticated procedures, with the use of surgical loupes and operating microscopes that magnify the surgical field. Over the last few years, advances in technologies and digital imaging have led to the introduction of robotic systems such as the “Da Vinci” robot, which allowed a new approach to microsurgery and supermicrosurgery.
More recently, the exoscope, a new magnification system that provides a 3D image of the surgical field, has become available. The image on the screen can be magnified as much as with a microscope, therefore microsurgical procedures can be conducted with the aid of this instrument.
The aim of this report is to evaluate the exoscope as a magnification tool in microsurgical vessel anastomosis, both for experienced surgeons and for the training of those with less experience.
In addition, we evaluated this new device and its advantages in terms of image quality, ease of setting, and comfort for the surgeon.
According to this preliminary experience, the high-definition 3D exoscope (VITOM
®) may also be considered for microanastomosis training. Gaining experience in the technique of vascular anastomosis on vessels of a caliber of only a few millimeters requires practice. The basic microsurgical courses usually consist of some theoretical instructions, followed by several hours of practical exercises under the microscope. As an example, microsurgical courses, to be approved by the Italian Society of Microsurgery (SIM), must provide at least twenty hours of hands-on training [
1]. The first microsurgical exercises aim to teach depth perception, stereoscopic vision, and dexterity under the microscope magnification [
2].
The next step is teaching the performance of end-to-end anastomosis using microsurgical instruments, usually on a 2 mm or 3 mm synthetic tube under the microscope, and with an 8–0 or 9–0 suture. Reports on the use of exoscopes for neurosurgical interventions have started to appear [
3], although only a few papers have described the applications for microsurgical vessel anastomoses during free flaps reconstruction [
4,
5].
2. Materials and Methods
2.1. VITOM® 3D 4K Exoscope
The exoscope is a device able to display the images recorded by two compact and light cameras, coupled at the tip of a small crane, on a 3-dimensional 4 k high-resolution video screen. The complete exoscope set also includes a control joystick (Pilot) to zoom, focus, and finely adjust the camera position, and several pairs of three-dimensional eyeglasses (
Figure 1).
2.2. Study Design
Six microsurgeons with various levels of experience participated in the study. All participants filled out a form, which included demographic data (age, gender), hand dominance (right-handed, left-handed, ambidextrous), previous experience (including traditional conventional microsurgery, robotic surgery and simulators, endoscopic surgery and simulators, video games experience), and the level of training (resident, research fellow, junior consultant, senior consultant).
An official representative of the device’s manufacturer explained, in 15 min, how to set up and use the system and its various components.
During the study, each surgeon used the exoscope VITOM 3D as a magnification system for the first time. They were asked to perform three end-to-end anastomoses (3 mm caliber double layer latex vessels) (
Figure 2) and two end-to-side anastomoses (2.8 mm caliber double layer latex vessel to 9 mm caliber single layer latex vessel) on latex vessel models that were fixed on a foam platform and secured with a double-opposing Ackland vascular clamp (
Figure 3).
The microvascular anastomoses were performed with an interrupted suture technique, using a 9.0 nylon thread. At the end of the test, each participant reported their experience on a 4-point Likert scale through the VITOM Quality Assessment Tool [
6], a questionnaire assessing a subjective evaluation of the tool’s qualities. Finally, the time needed to complete each microanastomosis by each surgeon, the number of stitches used, and the quality of the anastomosis, provided an objective evaluation of the performance.
3. Results
The surgeons’ demographic data are reported in
Table 1.
The mean age was 40.6 years, and the range was from 28 years to 65 years old. Four participants were male and two were female.
Five participants were right-handed, one was left-handed.
In terms of previous microsurgical experience, the surgeons were classified as follows: one senior consultant boasted vast microsurgical expertise (category D: over 30 years of practice); two testers were competent (category C: microsurgeons who usually perform this kind of surgery, but have less than 20 years of experience), one participant was considered a beginner (category B: surgeon that attends courses but does not routinely perform microsurgical procedures); two were plastic surgeons in training (residents) and this was their first experience with microsurgical vessel anastomosis (category A: no previous experience).
None of the surgeons involved had previous experience with the exoscope, with robotic surgery, with endoscopic surgery, nor with their training simulators. Fifty percent of the enrolled surgeons (n = 3) had minimal experience with videogames.
All of the surgeons involved in the study completed the five required microsurgical anastomoses.
4. Discussion
The introduction by Nylean in 1954 of the operating microscope into clinical practice [
7,
8] marked a milestone in the history of microsurgery. In 1960, Jacobson and Suarez performed the first vascular anastomosis using an operating microscope [
9].
4.1. Efficiency of The Exoscope as a Magnification Device
The use of the exoscope in neurosurgery has been reported to result in better ergonomics and accessibility of the surgical field.
Belykh et al. anastomosed, with microsurgical techniques and the aid of the exoscope magnification, the small vessels of humans, bovines, and rats [
10].
They concluded that the VITOM 3D 4k was superior to the microscope for clarity and sharpness, the size of surgical field, and depth perception, without lengthening the operating time when compared to the microscope.
Similarly, Ahmad et al., when comparing the 3D exoscope with the conventional microscope during free flap surgery for flap dissection and microvascular anastomosis, concluded that the exoscope is an effective microsurgical tool that may be considered as an alternative to conventional optical magnification [
11].
Piatkowski et al. performed a bilateral breast reconstruction with free Deep Inferior Epigastric Perforator (DIEP) flaps using for internal mammary vessels dissection and microvascular anastomosis the 3D exoscope on one side, and the conventional microscope on the contralateral side. They concluded that the exoscope did not reach the performance of a good quality microscope [
4].
Grammatica et al. used a 3D 4K exoscopic system in a clinical setting for head and neck microsurgical reconstruction. They concluded that the harvesting of free flaps and microanastomosis with the 3D 4K exoscopic system was feasible [
12].
De Virgilio et al. performed a clinical human study of free flap microvascular anastomosis using the VITOM 3D exoscope in 10 consecutive patients undergoing reconstruction after ablative surgery for head and neck carcinoma. Microvascular anastomoses were performed successfully using the exoscope in all patients, without any need for a conventional microscope [
13].
In our study we also found that the instrumentation provided adequate magnification to perform the anastomosis of artificial vessels of 2 mm to 3 mm of caliber. Moreover, its use is easy to learn and the performance of the surgical task improves rapidly, as demonstrated by the reduction in time needed for each anastomosis. A limit of our study in this aspect was the low number of anastomoses performed and the bias of the artificial setting, however the trend to a fast, progressive improvement at each anastomosis can clearly be seen.
4.2. The Exoscope as a Training Device
Laboratories dedicated to teaching technical aspects of the new technologies to surgeons in training are frequently being organized. The exoscope-assisted technology grants a shorter learning curve compared to other magnification systems, and not requiring a change of posture improves the speed with which the anastomosis can be performed.
A preliminary study demonstrated the usefulness of the three-dimensional exoscope when teaching some basic tasks requiring high magnification, although in in this work no microsurgical vessel anastomoses were performed [
6].
In our study, we tested the VITOM 3D as a magnification device for microsurgical anastomosis. We believe that, as a training model, synthetic vessels have several advantages compared to animal tissues [
14], therefore we chose to use small latex tubes. One additional advantage of the exoscope, particularly important in a teaching setting, is the position of the 3D monitor that allows one or more assistants to stand near the surgeon and observe the procedure from a similar angle. (
Figure 4b). In a clinical setting two or more monitors can be used, in different positions, to allow the operators to stand one in front of the other, or in the most comfortable and useful position.
Based on our experience, the exoscope can be considered to be an excellent device to provide magnification during microsurgical courses. Other training models for microsurgical anastomoses that take advantage of modern technologies have been previously proposed [
15]. Using the exoscope from the initial microsurgical training will allow the surgeons to become acquainted with a device that they will later use when practicing in a clinical setting.
4.3. Ergonomics
At present, standard microsurgical reconstruction requires the use of the operating microscope with the surgeon positioned in proximity to the ocular lenses, remaining in a posture uncomfortable for the shoulders, neck, and wrists, especially during long procedures.
In the last few years, more attention has been paid to the physical well-being of surgeons during long interventions [
5].
The wide field of view and deep focusing attributes of the exoscope enable its positioning above the surgical field, and minimize the need for re-positioning and re-focusing during the procedure. Moreover, the freedom of movement of the head and neck of the surgeon without losing the magnified view allows a more relaxed posture, and lets the surgeon change positions and relax their muscles (
Figure 4).
4.4. Cost
No economic evaluation on the use of the VITOM-3D exoscope has been conducted [
6].
A precise evaluation of costs in our setting is difficult due to the method of purchasing goods in public hospitals, although an estimate could be attempted. The cost of a modern, high-quality operating microscope routinely used in microsurgery varies from 130,000 to 400,000 euros [
3] while that of the VITOM
® 3D 4K exoscope averages 150,000 euros.
One advantage of the exoscope in a large university hospital is also related to the possible adjunctive use of some components of this equipment in other kind of surgeries, such as endoscopic procedures, or the entire exoscope in other surgical specialties that may need surgical field magnification or a better view in narrow spaces [
3]. This could result in a faster depreciation of the investment for the healthcare organisation.
5. Conclusions
We tested the use of the exoscope VITOM 3D for microsurgical anastomosis on synthetic vessels as a model for training courses in microsurgery, and to evaluate its qualities as a magnification tool.
According to the surgeons that tested it in this study, these are some of the qualities of the device:
Excellent image quality and depth perception
Possibility to easily switch from the magnified image to the macroscopic view of the anastomotic site
Available space for the assistants to view the operating field, right at the surgeon’s side, demonstrating a clear advantage for education
Easy to use and become acquainted with the apparatus, as shown by the rapid decrease in the time needed to complete each anastomosis
Other device characteristics, as compared to the usual surgical microscope:
Greater freedom of movement
Superior ergonomics which permitted a relaxed posture of the head, neck, and trunk of the surgeon while performing the anastomosis.
The main limit of our study was the small number of surgeons who tested the instrument, and the small number of anastomoses performed. The artificial vessels used were 2 mm and 3 mm wide, and results may be different in living tissues and vessels of 1 mm.
The evaluations of the surgeons involved in this study were comparing the exoscope to their previous, routine practical use of the surgical microscope in clinical practice. A direct comparison of the exoscope to a surgical microscope, side by side, on the same tasks, could be useful in the next study.
Further experience can be gained and described as soon as the device becomes more widely available.
Author Contributions
Conceptualization, V.P. and M.P.; Methodology, V.P. and M.P.; Formal Analysis, M.E.L.M. and F.A.G.; Investigation, V.P., M.E.L.M., F.A.G., A.T., R.C. and M.P.; Resources, M.M.P.; Data Curation, V.P.; Writing-Original Draft Preparation, V.P., M.P., M.M.P.; Writing-Review & Editing, M.P.; Visualization, M.E.L.M. and F.A.G.; Supervision, R.C.; Project Administration, V.P.; Funding Acquisition, A.T. All authors have read and agreed to the published version of the manuscript.
Funding
Zaccanti SpA provided the instrument for the study and support for the printing expenses but had no direct access to the data and no role in the analysis.
Acknowledgments
We would like to thank Giacomo Cannamela, who performed data analysis.
Conflicts of Interest
The authors declare no conflict of interest.
References
- Rules of the Italian Society of Microsurgery for the Accreditation of Basic Microsurgical Courses. Available online: www.microchirurgia.org/formazione-criteri-minimi-accreditamento.htm (accessed on 1 September 2020).
- Masud, D.; Haram, N.; Moustaki, M.; Chow, W.; Saour, S.; Mohanna, P.N. Microsurgery simulation training system and set up: An essential system to complement every training programme. J. Plast. Reconstr. Aesthet. Surg. 2017, 70, 893–900. [Google Scholar] [CrossRef] [PubMed]
- Oertel, J.M.; Burkhardt, B.W. Vitom-3D for exoscopic neurosurgery: Initial experience in cranial and spinal procedures. World Neurosurg. 2017, 105, 153–162. [Google Scholar] [CrossRef] [PubMed]
- Piatkowski, A.A.; Keuter, X.H.A.; Schols, R.M.; van der Hulst, R.R.W.J. Potential of performing a microvascular free flap reconstruction using solely a 3D exoscope instead of a conventional microscope. J. Plast. Reconstr. Aesthet. Surg. 2018, 71, 1664–1678. [Google Scholar] [CrossRef] [PubMed]
- Ichikawa, Y.; Senda, D.; Shingyochi, Y.; Mizuno, H. Potential Advantages of Using Three-Dimensional Exoscope for Microvascular Anastomosis in Free Flap Transfer. Plast. Reconstr. Surg. 2019, 144, 726e–727e. [Google Scholar] [CrossRef] [PubMed]
- De Virgilio, A.; Costantino, A.; Ebm, C.; Conti, V.; Mondello, T.; Di Bari, M.; Spriano, G. High definition three-dimensional exoscope (VITOM 3D) for microsurgery training: A preliminary experience. Eur. Arch. Otorhinolaryngol. 2020, 277, 2589–2599. [Google Scholar] [CrossRef] [PubMed]
- Nylean, C.O. The microscope in aural surgery, its first use and later development. Acta Oto-Laryngol. 1954, 43, 226–240. [Google Scholar] [CrossRef]
- Byvaltsev, V.A.; Akshulakov, S.K.; Polkin, R.A.; Ochkal, S.V.; Stepanov, I.A.; Makhambetov, Y.T.; Preul, M.C. Microvascular Anastomosis Training in Neurosurgery: A Review. Minim. Invasive Surg. 2018, 2018, 6130286. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Jacobson, J.H.; Suarez, E.L. Microsurgery in anastomosis of small vessels. Surg. Forum 1960, 11, 243–245. [Google Scholar]
- Belykh, E.; George, L.; Zhao, X.; Carotenuto, A.; Moreira, L.B.; Yağmurlu, K.; Preul, M.C. Microvascular anastomosis under 3D exoscope or endoscope magnification: A proof-of-concept study. Surg. Neurol. Int. 2018, 9, 115. [Google Scholar] [PubMed]
- Ahmad, F.I.; Mericli, A.F.; DeFazio, M.V.; Chang, E.I.; Hanasono, M.M.; Pederson, W.C.; Selber, J.C. Application of the ORBEYE three-dimensional exoscope for microsurgical procedures. Microsurgery 2020, 40, 468–472. [Google Scholar] [CrossRef] [PubMed]
- Grammatica, A.; Schreiber, A.; Vural, A.; Deganello, A.; Ferrari, M.; Lancini, D.; Nicolai, P. Application of a 3D 4K exoscopic system to head and neck reconstruction: A feasibility study. Eur. J. Plast. Surg. 2019, 42, 611–614. [Google Scholar] [CrossRef]
- De Virgilio, A.; Iocca, O.; Di Maio, P.; Mercante, G.; Mondello, T.; Yiu, P.; Spriano, G. Free flap microvascular anastomosis in head and neck reconstruction using a 4K three-dimensional exoscope system (VITOM 3D). Int. J. Oral Maxillofac. Surg. 2020, 49, 1169–1173. [Google Scholar] [CrossRef] [PubMed]
- Atlan, M.; Lellouch, A.G.; Legagneux, J.; Chaouat, M.; Masquelet, A.C.; Letourneur, D. A New Synthetic Model for Microvascular Anastomosis Training? A Randomized Comparative Study between Silicone and Polyvinyl Alcohol Gelatin Tubes. J. Surg. Educ. 2018, 75, 182–187. [Google Scholar] [CrossRef] [PubMed]
- Huang, T.C.; Sabbagh, M.D.; Adabi, K.; Moran, S.L.; Lu, C.K.; Roh, S.G.; Manrique, O.J. Compact and Economical Microsurgical Training Made Possible with Virtual Reality. Plast. Reconstr. Surg. 2018, 142, 993e–995e. [Google Scholar] [CrossRef] [PubMed]