Article

Fabrication of a Disposable Electrochemical Immunosensor Based on Nanochannel Array Modified Electrodes and Gated Electrochemical Signals for Sensitive Determination of C-Reactive Protein

1
Shanxi Bethune Hospital, Shanxi Academy of Medical Sciences, Tongji Shanxi Hospital, Third Hospital of Shanxi Medical University, Taiyuan 030032, China
2
Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430030, China
3
Department of Chemistry, Key Laboratory of Surface & Interface Science of Polymer Materials of Zhejiang Province, Zhejiang Sci-Tech University, Hangzhou 310018, China
*
Authors to whom correspondence should be addressed.
Nanomaterials 2022, 12(22), 3981; https://doi.org/10.3390/nano12223981
Submission received: 27 September 2022 / Revised: 27 October 2022 / Accepted: 8 November 2022 / Published: 11 November 2022
(This article belongs to the Special Issue Applications of Nanomaterials for Biological Analysis)

Abstract

:
Sensitive determination of C-reactive protein (CRP) is of great significance because it is an early indicator of inflammation in cardiovascular disease and acute myocardial infarction. A disposable electrode with an integrated three-electrode system (working, reference, and counter electrodes) has great potential in the detection of biomarkers. In this work, an electrochemical immunosensing platform was fabricated on disposable and integrated screen-printed carbon electrode (SPCE) by introducing nanochannel arrays and gated electrochemical signals, which can achieve the sensitive detection of CRP in serum. To introduce active reactive groups for the fabrication of immuno-recognitive interface, vertically-ordered mesoporous silica-nanochannel film (VMSF) with rich amino groups (NH2-VMSF) was rapidly grown by electrochemical assisted self-assembly (EASA). The electrochemically reduced graphene oxide (ErGO) synthesized in situ during the growth of NH2-VMSF was used as a conductive adhesive glue to achieve stable bonding of the nanochannel array (NH2-VMSF/ErGO/SPCE). After the amino group on the outer surface of NH2-VMSF reacted with bifunctional glutaraldehyde (GA/NH2-VMSF/ErGO/SPCE), the converted aldehyde surface was applied for covalent immobilization of the recognitive antibody (Ab) followed with the blocking of the non-specific sites. The fabricated immunosensor, Ab/GA/NH2-VMSF/ErGO/SPCE, enables sensitive detection of CRP in the range from 10 pg/mL to 100 ng/mL with low limit of detection (LOD, 8 pg/mL, S/N = 3). The immunosensor possessed high selectivity and can realize reliable determination of CRP in human serum.

1. Introduction

C-reactive protein (CRP) is an acute protein that rises sharply in the plasma when the body is infected or tissue is damaged [1,2]. CRP plays an important protective role in the body’s natural immune process because it can activate complement and enhance the phagocytic function of phagocytes to remove pathogenic microorganisms that invade the body, as well as damaged, necrotic, and apoptotic tissue cells. More importantly, CRP is designated as an early indicator of inflammation in cardiovascular diseases and acute myocardial infarction [3,4]. For instance, CRP is directly involved in atherosclerosis. It has been demonstrated that large amounts of CRP are deposited in human atherosclerotic plaques. Therefore, the highly sensitive and specific determination of CRP is of great significance for clinical diagnosis [5,6]. Up to now, the methods for the detection of CRP include immunodiffusion, radioimmunosensor, turbidimetric method, enzyme or fluorescence labeled immunosensor, electrochemistry (EC) or electrochemiluminescence (ECL), etc. [3,7,8,9,10,11,12,13]. However, these strategies often have the problems of complex detection processes or insufficient sensitivity. An electrochemical immunosensor is an analytical method that combines immunosensors with electrochemical detection technology. As known, compared with other detection methods (such as optical technology, etc.), electrochemical detection has the advantages of simple operation, rapid detection, high sensitivity, and simple instrumentation [14,15,16,17]. Owing to the high specificity and selectivity of immune recognition and the unique merits of electrochemical detection, electrochemical immunosensors have great application potential in the clinical detection of CRP.
Effective construction of an immune recognition interface with biological recognition is the basis of an electrochemical immunosensor, which is usually fabricated on the surface of the supporting electrode. The most common electrochemical electrodes include noble metal electrodes (e.g., gold or platinum electrodes, etc.), carbon electrodes (e.g., glassy carbon electrode-GCE), etc. [18,19]. Yet these electrodes are very expensive and require a tedious grinding process before use. Compared to these electrodes, disposable electrodes offer significant advantages including high repeatability between each electrode, no need for electrode maintenance, flexibility, and customizable electrode structure [20,21,22,23]. Screen-printed carbon electrodes (SPCEs) are the most commonly used disposable electrodes and have the advantages of simple structure, low cost, easy large-scale production, and flexible design, demonstrating potential applications in electrochemical sensing, photovoltaic power, environmental detection, clinical analysis, food detection, and other fields [24,25,26,27]. Generally, three electrodes are printed on the SPCE substrate, namely the working electrode (WE), the reference electrode (RE), and the auxiliary electrode (AE), leading to a classic electrochemical three-electrode system. Fabrication of the electrochemical immunosensor on SPCE and its application in the detection of CRP is highly desirable.
The accuracy of electrochemical detection is also the key to electrochemical immunosensor. The biggest challenge is the contamination of the electrode surface, which is more prominent in the analysis of complex biological samples (e.g., serum). For example, a large number of proteins contained in serum easily adhered to the electrodes through non-specific adsorption, which reduces the electrochemically active area of the electrode and results in low detection accuracy and stability. Therefore, it is very important to improve the anti-fouling ability of the electrode. Until now, the anti-fouling strategies for electrodes can be commonly divided into two categories. One is by introducing charged hydrophilic polymers [28,29]. The principle of this strategy lies in the reduced non-specific adsorption of proteins by increasing the hydrophilicity of the electrode surface and the exclusion of species with the same charges through electrostatic repulsion. Another strategy is to introduce functional nanomaterials on the electrode surface [30,31]. This strategy is more advantageous due to the rich structure, easy modification, and abundant function of nanomaterials. Amongst, the introduction of vertically-ordered mesoporous silica nanomembrane film (VMSF) on the surface of electrodes has attracted much attention in recent years [32,33,34]. VMSF has ultra-small size (2–3 nm in pore size), high density, and parallel nanochannel array, which ensure the permeability of small molecules to the underlying electrode. In addition, ultra-small nanochannels can effectively exclude substances such as proteins through size sieving [35,36]. Combined with the non-conductivity property of its silica structure, VMSF can effectively prevent the contamination of the underlying conductive electrodes [37,38,39]. In addition, although the nanochannel cannot accommodate bio-recognitive ligands (e.g., protein, DNA, etc.), the outer surface of VMSF, that is, the entrance of nanochannels can be modified by bioligand molecules, leading to significant gated effects [40,41]. For example, the formation of antigen-antibody complexes can significantly alter the diffusion of solution-based electrochemical redox probes in VMSF. Combined with the advantages of easy preparation and low cost of VMSF, VMSF-modified electrodes are promising in signal-gating electrochemical immunosensors.
In this work, a disposable and integrated electrochemical immunosensor was fabricated for sensitive determination of C-reactive protein based on a nanochannel array modified electrode and gated electrochemical signal. Screen-printed electrode (SPCE) was used as the disposable supporting electrode with an integrated three-electrode system. To improve the binding stability of VMSF on SPCE, electrochemically reduced graphene oxide (ErGO) formed in situ during VMSF growth was used as the conductive bonding layer. Using siloxanes containing amino groups as precursors, VMSF containing amino groups (NH2-VMSF) was rapidly grown by the simple electrochemical-assisted self-assembly (EASA) method. When the outer surface of VMSF was derivatized with glutaraldehyde (GA), the aldehyde-based surface enabled covalent immobilization of CRP antibody (Ab). The immuno-electrode obtained after blocking the non-specific sites using bovine serum albumin (BSA) can realize sensitive detection of CRP because the formation of immuno-complexes blocked the nanochannels and reduced the diffusion of solution-phase redox probes (Fe(CN)63−/4−), resulting in a gated electrochemical signal.

2. Materials and Methods

2.1. Chemicals and Materials

Graphene oxide (GO) was synthesized by the conventional Hummers method [42,43]. Tetraethyl orthosilicate (TEOS), 3-aminopropyltriethoxysilane (APTES), hexadecyl trime-thyl ammonium bromide (CTAB), potassium ferricyanide (K3[Fe(CN)6]), sodium phosphate dibasic dodecahy-drate (Na2HPO4•12H2O), bovine whole blood albumin (BSA), glutaraldehyde(GA), and potassium hydrogen phthalate (KHP) were all purchased from Aladdin Chemistry Co., Ltd. (Shanghai, China). Potassium chloride (KCl), sulfuric acid (H2SO4), and ethanol (EtOH) were obtained from Hangzhou Gaojing Chemistry Co., Ltd. (Hangzhou, China). Sodium dihydrogen phosphate dehydrate (NaH2PO4•2H2O) and 3-aminopropyltriethoxysilane (APTES) were purchased from Macklin (Shanghai, China). Hydrochloric acid (HCl) was obtained from Hangzhou Shuanglin Chemical Reagent Co., Ltd. C-reactive protein (CRP), CRP antibody, serum amyloid A (SAA), alpha-fetoprotein (AFP), and procalcitonin (PCT) were bought from Nanjing OkayBio (Nanjing, China). Carcinoembryonic antigen (CEA) was purchased from Beijing KEY-BIO Biotech Co., Ltd. (Beijing, China). A screen-printed carbon electrode (SPCE) with a three-electrode system (DRP-C110-U75) was obtained from Metrohm (Heilissau, Switzerland). Briefly, the carbon electrode was used as the working electrode and the counter electrode (4 mm diameter), and the silver electrode was used as the reference electrode. All chemicals used were of analytical grade and used without further treatment. The ultrapure water (18.2 MΩ•cm) used in the experiment was prepared by Mill-Q Systems (Millipore Company, Billerica, MA, USA).

2.2. Measurements and Instrumentations

X-ray photoelectron spectroscopy (XPS) analysis of GO or ErGO-modified SPCEs was performed using Mg Kα irradiation at 250 W and 14 kV (PHI5300, PE Ltd., Boston, MA, USA). The morphology of NH2-VMSF was investigated by transmission electron microscopy (TEM) using an HT7700 microscope (Hitachi, Tokyo, Japan) with copper mesh as the support material. An accelerating voltage of 100 kV was applied. Before measurement, NH2-VMSF was carefully scraped from the electrode surface and sonicated in ethanol, to obtain the sample dispersion. All electrochemical experiments including cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS), and differential pulse voltammetry (DPV) were performed on an Autolab electrochemical workstation (PGSTAT302N, Metrohm, Heilissau, Switzerland). For EIS measurement, frequencies ranged from 0.1 Hz to 100 kHz with a perturbation amplitude of 5 mV.

2.3. Preparation of NH2-VMSF Modified SPCE

Before use, SPCE was electrochemically polished 10 times by cyclic voltammetry scan in diluted sulfuric acid solution (0.05 M). The potential was from 0.4 V to 1.0 V. Then, the electrode was thoroughly rinsed with ultrapure water and dried with a stream of N2. To modify the electrode, GO dispersion (0.1 mg/mL, 10 μL) was dropped on the WE area of SPCE and dried naturally at room temperature. Then, NH2-VMSF was synthesized in the presence of CTAB micelles (SM) using mixed siloxane precursors containing APTES and TEOS [44]. Specifically, CTAB (1.585 g) and APTES (0.159 g) were added to a mixture containing ethanol (20 mL) and an equal volume of NaNO3 solution (0.1 M, pH = 2.6) under stirring. After the pH of the solution was adjusted to 3.0 with HCl (6 M), TEOS (2.891 g) was added and the obtained mixture was stirred at room temperature for 2.5 h. When the SPCE was placed in the above precursor solution, a current of −350 μA was applied to the working electrode for 10 s. After thorough rinsing with ultrapure water, the electrode containing surfactant SM (SM@NH2-VMSF/ErGO/SPCE) was obtained. After the electrodes were aged at 80 °C for 10 h, SM was removed using 0.1 M HCl/EtOH solution (V:V = 1:1) to obtain electrodes with open nanochannels (NH2-VMSF/ErGO/SPCE). To prove the role of nanochannels, a control electrode containing an amino group but without nanochannels was prepared and denoted as NH2-ErGO/SPCE. Briefly, ErGO/SPCE was prepared in the solution without siloxane precursor under the same conditions. The dried ErGO/SPCE was immersed in an acetone (48 mL) solution containing APTES (1.136 g) for 1 h. Then, NH2-ErGO/SPCE was obtained after successively washing with acetone, ethanol, acetone, and water, respectively.

2.4. Fabrication of the Immuonsensor

To obtain the aldehyde-modified surface, the NH2-VMSF/ErGO/SPCE electrode was immersed in 2% GA (pH = 7.4) solution for 1 h at 37 °C in the dark. After thoroughly washing with phosphate buffer solution (PBS, 0.05 M, pH = 7.0), the electrode was soaked in CRP antibody (50 μg/mL) solution and incubated at 37 °C for 1 h. After the unbound antibody was rinsed with PBS, the obtained electrode was immersed in BSA solution (1 mg/mL) to block the non-specific sites. The resulting immunoelectrode was designated as Ab/NH2-VMSF/ErGO/SPCE. For the incubation, the electrode was inserted into the antigen solution, and then the reaction system was placed in an incubator with constant temperature (37 °C) and humidity (90%) for 1 h.

2.5. Electrochemical Determination of CRP

The Ab/NH2-VMSF/ErGO/SPCE immunosensors were incubated with different concentrations of CRP for 60 min at 37 °C. Using PBS (0.05 M, pH = 7.0) containing Fe(CN)63−/4− (1.25 mM) and KCl (0.1 M) as the detection solution, the electrochemical signals on the immunoelectrodes before and after CRP binding (37 °C, 1 h) were measured. For actual sample analysis, CRP in human serum was determined using a standard addition method. Serum samples were diluted by a factor of 50 before detection. To evaluate the stability in storage, the immunosensor was stored in PBS (0.05 M, pH = 7.0) at 4 °C.

3. Results and Discussion

3.1. Fabrication of Electrochemical Immunosensor on NH2-VMSF Modified SPCE

VMSF has a uniformly sized, high-density nanochannel array. In addition, functional chemical reactive groups can be easily introduced using functional siloxane precursors. However, it is difficult to stably bind VMSF directly to carbon electrodes. Graphene materials exhibit different dimensionals (e.g., 0D grpahene quantum dots [45,46,47,48], 2D graphene sheets [49,50], and 3D graphene foams [51,52,53]) and have been employed as functional elements to improve the sensitivity of (bio)sensors. Herein, electrochemically reduced graphene oxide (ErGO) formed in situ during VMSF growth was used as the conductive bonding layer to improve the bonding stability of VMSF on SPCE. To facilitate the covalent immobilization of the recognitive antibody (Ab), VMSF-containing amino groups (NH2-VMSF) were prepared by the EASA method with amino group-containing siloxane being the precursor. As illustrated in Figure 1, graphene oxide (GO) was first dropped on the surface of SPCE. The mechanism of the EASA method lies in the kinetic control of the sol-gel process of siloxane precursors around surfactant micelle (SM) templates. To grow NH2-VMSF, a negative current was applied on the electrode to increase the pH in situ on the electrode surface through reducing protons, which promoted the hydrolysis and condensation of siloxane precursors around the SM template. During this process, GO drop-coated on the surface of SPCE can be electrochemically reduced in situ, leading to electrochemically reduced graphene oxide (ErGO). On the one hand, ErGO interacted with the underlying carbon electrode through hydrophobic interaction. On the other hand, the oxygen-containing groups on ErGO can undergo a condensation reaction with the silanol groups in NH2-VMSF, realizing covalent bonding. Thus, ErGO acted as the conductive binding layer to improve the binding stability of NH2-VMSF.
When VMSF was grown, the resulting nanochannels were blocked by SM and the electrode was denoted as SM@NH2-VMSF/ErGO/SPCE. After SM was removed, the obtained NH2-VMSF/ErGO/SPCE electrode had an open nanochannel. Since the outer surface of NH2-VMSF contained chemically reactive amino groups, the aldehyde-based surface (GA/NH2-VMSF/ErGO/SPCE) was easily achieved by reacting NH2-VMSF with the bifunctional group reagent glutaraldehyde (GA), which can realize covalent immobilization of the recognitive antibody (Ab) through the reaction between aldehyde group and amino group in Ab. After blocking non-specific sites with BSA, the immunosensor (Ab/GA/NH2-VMSF/ErGO/SPCE) was finally obtained. In comparison with the orientated immobilization strategy that commonly applied biotin and streptavidin, this immobilization of the antibody is simple, fast, and low-cost. Due to the high density of nanochannels in NH2-VMSF, redox probes (Fe(CN)63−/4−) in a solution can efficiently diffuse to the underlying electrode, generating electrochemical signals. When the fabricated immunosensor is incubated with the antigen, the formed antibody-antigen complex with a large size will block the entrance of some nanochannel because of its large size, leading to the decreased mass transport of Fe(CN)63−/4− in the solution to the underlying electrode. As a result, the detected electrochemical signal decreased. This phenomenon is the gated effect. Based on this principle, the constructed immunosensor can realize sensitive electrochemical detection of CRP.

3.2. Characterization of NH2-VMSF and ErGO

The changes in the electrode surface during the growth of NH2-VMSF were investigated using a standard electrochemical probe (Fe(CN)63−). As shown in Figure 2a, a reversible electrochemical redox peak of Fe(CN)63− was observed on the supporting SPCE. When NH2-VMSF growth was completed, no obvious Faraday current was measured on the resulting SM@NH2-VMSF/ErGO/SPCE. This is attributed to the inhibition of probe diffusion to the underlying electrode due to the blocking of the nanochannels by SM. This phenomenon also demonstrated that NH2-VMSF had no cracks or fractured regions, otherwise the redox probes would directly interact with the electrode to generate electrochemical signals. When SM was removed, the opening of the nanochannels resulted in a significant electrochemical signal of Fe(CN)63− on NH2-VMSF/ErGO/SPCE. In addition, the oxidation and reduction peak currents are higher than those obtained on SPCE. This is attributed to the presence of NH2 groups in the nanochannels, which act as cationic sites to enrich negatively charged Fe(CN)63−, resulting in increased peak currents. When NH2-VMSF was removed by sodium hydroxide, the redox peak of the probe could also be detected on ErGO/SPCE. The larger charging current than that of SPCE, proving the increase of electrode active area. Figure 2a also shows CV curves of Fe(CN)63− on ErGO/SPCE, NH2-ErGO/SPCE, and NH2-VMSF/ErGO/SPCE. As seen, NH2-ErGO/SCPE has a higher redox peak current compared with ErGO/SPCE, which is attributed to the fact that the amino group can enrich the negatively charged Fe(CN)63− through electrostatic interaction. However, NH2-VMSF/ErGO/SPCE with grown nanochannels exhibits the highest peak current, proving that the nanochannels rich in amino groups can significantly enrich Fe(CN)63−. In addition, the peak-to-peak difference decreases, indicating the improved reversibility of the electrochemical process of Fe(CN)63−.
The changes in the electrode interface during this process were also demonstrated by electrochemical impedance spectroscopy (EIS) characterization. Figure 2b shows the EIS curves obtained on the above electrodes. Each EIS curve consists of a semicircle in the high- frequency region and a linear part in the low-frequency region, where the former represents an electron transfer-limited process and the latter is a diffusion-limited process. The equivalent diameter of the semicircle in the high-frequency region corresponds to the charge transfer resistance (Rct). As seen, the modification of ErGO on SPCE significantly reduced the Rct of the electrode. When the nanochannels of NH2-VMSF were blocked by SM, no diffusion-limited process of the electrode was observed on the EIS curve in the tested frequency range. In comparison, the following opening of nanochannels after SM removal resulted in a significant reduction in Rct. The function of nanochannels was investigated using the control electrode containing an amino group but without nanochannels, which was prepared by amino functionalization of ErGO/SPCE and denoted as NH2-ErGO/SPCE.
Transmission electron microscopy (TEM) was applied to investigate the morphology of NH2-VMSF. As shown in Figure 3, NH2-VMSF has an ordered silica nanoporous structure. No rupture was observed on NH2-VMSF over larger areas. The high-resolution TEM image (HRTEM, inset in Figure 3) revealed a hexagonal stacked structure with a pore size of about 2–3 nm.
An X-ray photoelectron spectroscopy (XPS) was used to further verify that GO was electrochemically reduced to ErGO in situ during the growth of NH2-VMSF. Figure 4a and 4b show the high-resolution C1s spectra of GO/SPCE and ErGO/SPCE, respectively. As shown, C-C/C=C peak of ErGO/SPCE significantly enhanced in comparison with that of GO/SPCE, indicating the increase of sp2 carbon structure in graphene. This phenomenon proves that GO on the surface of SPCE is reduced to ErGO during the growth of NH2-VMSF. In addition, the appearance of C-O, C=O, O-C=O groups revealed the existence of oxygen-containing groups on ErGO. These oxygen-containing groups will be beneficial to improve the stability of NH2-VMSF because they can form covalent bonds through condensation reactions with silanol groups in NH2-VMSF. Therefore, ErGO can act as a conductive adhesive layer and realize stable modification of NH2-VMSF on SPCE. On the one hand, ErGO can stably bind to SPCE through hydrophobic or π-π interaction. At the same time, NH2-VMSF can be stably bound to ErGO.

3.3. Fabrition of the Immunosensor

Effective immobilization of antibodies is the key to constructing an immunosensing system. Although the nanochannel of NH2-VMSF cannot accommodate the entry of antibodies owing to ultra-small pore size, the outer surface of NH2-VMSF (also the entrance of the nanochannel) can serve as the space for antibody immobilization. As illustrated in Figure 1, the aldehyde group derivatization of the amino group in NH2-VMSF produces an aldehyde group surface that can achieve covalent immobilization of antibodies. Electrochemical methods including cyclic voltammetry (CV) and EIS were also employed to study the changes in the electrode surface during the construction of immunosensor. Figure 5a shows the CV curves of different electrodes in the [Fe(CN)6]4−/3− probe. As demonstrated, the reaction of GA with NH2-VMSF leads to a decrease in the electrochemical signal. This might be ascribed that the reaction between amino group and aldehyde group reduces the number of an amino group, leading to decreased electrostatic interaction towards negatively charged Fe(CN)63−/4− and reduced electrochemical signal. When Ab was covalently immobilized on the aldehyde-modified surface and the non-specific binding sites of the electrode were blocked by BSA, the electrochemical signal of the resulting immunoelectrode (Ab/NH2-VMSF/ErGO/SPCE) significantly reduced. This is attributed to the fact that both the antibody and BSA are proteins, and their non-conductivity leads to an increase in the interfacial resistance of the electrodes. When the immunoelectrode was incubated with CRP, the electrochemical signal of the probe further decreased, proving that the formed antigen-antibody complex further blocked the nanochannel. The EIS measurements shown in Figure 5b also demonstrate the same experimental conclusion. Compared with the typical Randles equivalent circuit model that contains one constant-phase element (CPE), the modified Randles equivalent circuit model with two CPEs is more accurate and has been extensively used to approximate the EIS data of film electrodes [54]. Therefore the Nyquist plots in Figure 5b were fitted to the modified Randles equivalent circuit model [13,55]. As illustrated in inset of Figure 5b, two CPEs, two charge-transfer resistances (Rct), Warburg impedance (W), and the resistance of the electrolyte solution (Rs) are included, where Rct1 and Rct2 correspond to the film resistance and charge-transfer kinetics at electrodes, respectively. Compared with Rct1, Rct2 has great value for quantitative information because of its significant increases with the modification. Derivatization of GA, immobilization of Ab, and blocking of non-specific sites all resulted in a continuous increase in Rct2 (9.59 Ω for NH2-VMSF/ErGO/SPC, 1.08 KΩ for GA/NH2-VMSF/ErGO/SPCE, 1.41 KΩ for Ab/GA/NH2-VMSF/ErGO/SPCE). When the immunoelectrode was incubated with CRP, the Rct2 further increased (3.52 KΩ), indicating the successful capture of CRP on the immuno-recognitive interface.
To further improve the performance of the immunosensor, the time for the covalent immobilization of antibodies during the preparation of the immunoelectrode was optimized. When the antibody reacted with the aldehyde-modified surface at different times, the resulting immunoelectrode was incubated with CRP. The peak currents of each electrode after being bound to CRP were shown in Figure 5c. As shown, the increase in the reaction time resulted in a decrease of the peak current, indicating that the amount of the immobilized antibody and the following bound CRP increased. When the immobilization time was 60 min, the electrochemical signal became stable, proving that the immobilization of antibodies on an aldehyde-based surface reached saturation. In addition, the effect of the binding time of CRP on the immuno-recognitive interface was also investigated. It can be found that the electrochemical signal reached a plateau when the antigen-antibody was 60 min, proving the saturated binding of the antigen (Figure 5d). To investigate the non-specific binding of antigen to the surface, BSA was immobilized on GA/NH2-VMSF/ErGO/SPCE and the obtained electrode was denoted as BSA/GA/NH2-VMSF/ErGO/SPCE. As shown in the inset of Figure 5d, BSA/GA/NH2-VMSF/ErGO/SPCE exhibited almost no change in CV signal in the absence or presence of CRP, indicating no non-specific binding of antigen to the surface.

3.4. Sensitive Determination of CRP Using the Developed Immunosensor

When the antibody on the immune-recognitive specifically binds to CRP, the formed antigen-antibody immunocomplex will hinder the diffusion of the redox probe to the underlying electrode. Based on this gated electrochemical signal, the constructed immunosensor can be used for the electrochemical detection of CRP. Figure 6a shows the differential pulse voltammetry (DPV) curves obtained when the immunosensors incubate with different concentrations of CRP. As seen, the increase in CRP concentration leads to the decreased electrochemical signal of the redox probe. The peak current (I, μA) was linear proportional to the logarithm of the concentration of CRP (logCCRP, ng/mL) in the range of 10 pg/mL to 100 ng/mL (I = −1.31 logCCRP + 6.79, R2 = 0.995) (inset in Figure 6a). The limit of detection (LOD) of 8 pg/mL is obtained at a signal-to-noise ratio of 3. The comparison between the performance of different electrochemical sensors for the determination of CRP was shown in Table 1. The LOD is lower than that obtained from DPV detection by thiol-terminated poly(2-methacryloyloxyethyl phosphorylcholine)/gold nanoparticle modified screen-printed carbon electrode (PMPC-SH/AuNPs-SPCE) [56], ECL detection based on titania nanotubes and platinum nanowire modified indium tin oxide (TiNTs/PtNWs) [57], or DPV detection based on gold nanoparticle [12]. The starting concentration in the linear range of detection was lower than that obtained with DPV detection using branched polyethylenimine functionalized with ferrocene residues (PEI-Fc) [58], but higher than that obtained with EIS detection based on 11-cyanoundecyltrimethoxysilane and PAMAM dendrimers (11-CUTMS/PAMAM) [59] or square wave voltammetry (SWV) detection [60] based on 1-ethyl-3-(3-dimethylamino-propyl) carbodiimide hydro-chloride/N-hydroxy succinimide-activated 3-mercaptoproponic acid modified Au wire substrate (MPA/Au). The LOD is low enough for the determination of CRP levels associated with cardiovascular diseases (<1, 1–3, and >3 μg/mL for low, moderate, and high cardiovascular risks) [9] and acute myocardial infarction (>10 μg/mL) [12] in serum samples.

3.5. Selectivity, Reproducibility, and Stability of the Developed Immunosensor

Biological samples such as serum often contain a large amount of protein. To examine the selectivity of the constructed immunosensor, the electrodes were incubated with different proteins including procalcitonin (PCT), serum amyloid A (SAA), alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA), and CRP. The electrochemical signal of the obtained electrode was measured after the immunosensor was incubated with each protein or its mixture. As shown in Figure 6b, the electrochemical signal of the probe on the electrode significantly changed only when it reacted with CRP or the protein mixture containing CRP, indicating good selectivity of the developed immunosensor. The reproducibility and storage stability of the constructed immunosensor electrodes were investigated. Five electrodes were prepared in parallel in the same batch. The relative standard deviation (RSD) for the detection of CRP (10 ng/mL) was 2.9%, indicating good reproducibility. When the immunosensor was stored in a refrigerator at 4°C, 91% of the original performance for CRP (10 ng/mL) after one week of storage was obtained, suggesting good stability of the immunosensor.

3.6. Detection of CRP in Real Sample

To evaluate the application potential of the developed immunosensor, human serum was selected as a real sample and the concentration of CRP in human serum was determined by the standard addition method. As shown in Table 2, the recoveries of the immunosensors ranged from 97.6% to 102% with low relative standard deviations (less than 4.1%), proving good reliability and accuracy.

4. Conclusions

In this work, an immunosensor for sensitive electrochemical detection of C-reactive protein (CRP) was fabricated based on the modification of a screen-printed carbon electrode (SPCE) by nanochannel array using in situ generated electrochemically reduced graphene (ErGO) as a conductive adhesive layer. With amino-containing siloxanes as precursors to rapidly grow vertically-ordered mesoporous silica nanomembrane film with amino groups (NH2-VMSF). During the growth of NH2-VMSF, graphene oxide on the electrode surface is in situ reduced to ErGO, which increases the stability of NH2-VMSF while providing a conductive surface. The covalent immobilization of the immune-recognitive antibody was achieved after the outer surface of NH2-VMSF was converted into an aldehyde-based surface using a bifunctional group of glutaraldehyde. After the antigen-antibody forms an immune complex, it will hinder the diffusion of the probe in the solution and generate a gated electrochemical signal. Based on this mechanism, sensitive electrochemical detection of CRP can be achieved. The constructed immunosensor possesses good selectivity, reproducibility, and stability. The detection of CRP in human serum has high reliability. Since SPCE is a low-cost and disposable electrode, the developed immunosensor provides an effective strategy for a convenient and sensitive determination of CRP.

Author Contributions

Conceptualization, J.L.; investigation, data curation, N.M. and X.L.; writing—original draft preparation, N.M.; writing—review and editing, supervision, W.W. and J.L. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Natural Science Foundation of China (21904117), and the Zhejiang Provincial Natural Science Foundation of China (LY20B050007).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Chandra, P.; Suman, P.; Airon, H.; Mukherjee, M.; Kumar, P. Prospects and advancements in C-reactive protein detection. World J. Methodol. 2014, 4, 1–5. [Google Scholar] [CrossRef] [PubMed]
  2. Dykun, I.; Clark, D.; Carlo, J.; Lincoff, A.; Menon, V.; Nissen, S.; Nicholls, S.; Puri, R. Longitudinal high-sensitivity C-reactive protein and longer-term cardiovascular outcomes in optimally-treated patients with high-risk vascular disease. Am. J. Cardiol. 2022, 181, 1–8. [Google Scholar] [CrossRef] [PubMed]
  3. Phung, N.; Walter, J.; Jonczyk, R.; Seiler, L.; Scheper, T.; Blume, C. Development of an aptamer-based lateral flow assay for the detection of c-reactive protein using microarray technology as a prescreening platform. ACS Comb. Sci. 2020, 22, 617–629. [Google Scholar] [CrossRef] [PubMed]
  4. Ridker, P. Clinical application of C-reactive protein for cardiovascular disease detection and prevention. Circulation 2003, 107, 363–369. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Wang, Y.; Zhao, X.; Zhou, P.; Liu, C.; Chen, R.; Sheng, Z.; Li, J.; Zhou, J.; Song, L.; Zhao, H.; et al. Impact of postprocedural high-sensitivity C-reactive protein on lipoprotein(a)-associated cardiovascular risk with ST-segment elevation myocardial infarction with percutaneous coronary intervention. Am. J. Cardiol. 2021, 150, 8–14. [Google Scholar] [CrossRef]
  6. Zeller, J.; Bogner, B.; McFadyen, J.; Kiefer, J.; Braig, D.; Pietersz, G.; Krippner, G.; Nero, T.L.; Morton, C.; Shing, K.; et al. Transitional changes in the structure of C-reactive protein create highly pro-inflammatory molecules: Therapeutic implications for cardiovascular diseases. Pharmacol. Ther. 2022, 235, 108165. [Google Scholar] [CrossRef]
  7. Antonio, M.; Lima, T.; Vitorino, R.; Daniel-da-Silva, A. Label-free dynamic light scattering assay for C-reactive protein detection using magnetic nanoparticles. Anal. Chim. Acta 2022, 1222, 340169. [Google Scholar] [CrossRef]
  8. Bayansal, F.; Çetinkara, H. Chemiresistive CuO sensors for label-free C-reactive protein detection. J. Alloys Compd. 2022, 926, 166737. [Google Scholar] [CrossRef]
  9. Hong, D.; Kim, K.; Jo, E.J.; Kim, M. Electrochemiluminescence-incorporated lateral flow immunosensors using Ru(bpy)3(2+)-labeled gold nanoparticles for the full-range detection of physiological C-reactive protein levels. Anal. Chem. 2021, 93, 7925–7932. [Google Scholar] [CrossRef]
  10. Wang, S.; Luo, J.; He, Y.; Chai, Y.; Yuan, R.; Yang, X. Combining porous magnetic Ni@C nanospheres and CaCO3 microcapsule as surface-enhanced raman spectroscopy sensing platform for hypersensitive C-reactive protein detection. ACS Appl. Mater. Interfaces 2018, 10, 33707–33712. [Google Scholar] [CrossRef]
  11. Yang, X.; Zhang, P.; Xing, X.; Lv, Q.; Jiang, Y.; Jiang, Y.; Wang, Z. Improving the fluorescence intensity of lanthanide-doped microspheres via incorporation of lauryl methacrylate: Synthesis and their application in C-reactive protein detection. Colloid Surface A 2022, 653, 130065. [Google Scholar] [CrossRef]
  12. Jampasa, S.; Siangproh, W.; Laocharoensuk, R.; Vilaivan, T.; Chailapakul, O. Electrochemical detection of c-reactive protein based on anthraquinone-labeled antibody using a screen-printed graphene electrode. Talanta 2018, 183, 311–319. [Google Scholar] [CrossRef]
  13. Boonyasit, Y.; Chailapakul, O.; Laiwattanapaisal, W. A folding affinity paper-based electrochemical impedance device for cardiovascular risk assessment. Biosens. Bioelectron. 2019, 130, 389–396. [Google Scholar] [CrossRef]
  14. Baradoke, A.; Hein, R.; Li, X.; Davis, J. Reagentless redox capacitive assaying of C-reactive protein at a polyaniline interface. Anal. Chem. 2020, 92, 3508–3511. [Google Scholar] [CrossRef] [Green Version]
  15. Mahyari, M.; Hooshmand, S.; Sepahvand, H.; Gholami, S.; Rezayan, A.; Zarei, M. Gold nanoparticles anchored onto covalent poly deep eutectic solvent functionalized graphene: An electrochemical aptasensor for the detection of C-reactive protein. Mater. Chem. Phys. 2021, 269, 124730. [Google Scholar] [CrossRef]
  16. Sheen, H.; Panigrahi, B.; Kuo, T.; Hsu, W.; Chung, P.; Xie, Q.; Lin, C.; Chang, Y.; Lin, C.; Fan, Y. Electrochemical biosensor with electrokinetics-assisted molecular trapping for enhancing C-reactive protein detection. Biosens. Bioelectron. 2022, 210, 114338. [Google Scholar] [CrossRef]
  17. Sohrabi, H.; kholafazad Kordasht, H.; Pashazadeh-Panahi, P.; Nezhad-Mokhtari, P.; Hashemzaei, M.; Majidi, M.; Mosafer, J.; Oroojalian, F.; Mokhtarzadeh, A.; de la Guardia, M. Recent advances of electrochemical and optical biosensors for detection of C-reactive protein as a major inflammatory biomarker. Microchem. J. 2020, 158, 105287. [Google Scholar] [CrossRef]
  18. Yan, F.; Luo, T.; Jin, Q.; Zhou, H.; Sailjoi, A.; Dong, G.; Liu, J.; Tang, W. Tailoring molecular permeability of vertically-ordered mesoporous silica-nanochannel films on graphene for selectively enhanced determination of dihydroxybenzene isomers in environmental water samples. J. Hazard. Mater. 2021, 410, 124636. [Google Scholar] [CrossRef]
  19. Yan, F.; Chen, J.; Jin, Q.; Zhou, H.; Sailjoi, A.; Liu, J.; Tang, W. Fast one-step fabrication of a vertically-ordered mesoporous silica-nanochannel film on graphene for direct and sensitive detection of doxorubicin in human whole blood. J.Mater. Chem. C 2020, 8, 7113–7119. [Google Scholar] [CrossRef]
  20. Bhattacharya, G.; Fishlock, S.; Hussain, S.; Choudhury, S.; Xiang, A.; Kandola, B.; Pritam, A.; Soin, N.; Roy, S.; McLaughlin, J. Disposable Paper-Based Biosensors: Optimizing the Electrochemical Properties of Laser-Induced Graphene. ACS Appl. Mater. Interfaces 2022, 14, 31109–31120. [Google Scholar] [CrossRef]
  21. Lin, X.; Zhou, Q.; Huo, X.; Bao, N. Copper tape to improve analytical performance of disposable carbon electrodes in stripping analysis. Microchem. J. 2022, 179, 107428. [Google Scholar] [CrossRef]
  22. Yan, L.; Zhang, C.; Xi, F. Disposable amperometric label-free immunosensor on chitosan–graphene-modified patterned ITO electrodes for prostate specific antigen. Molecules 2022, 27, 5895. [Google Scholar] [CrossRef] [PubMed]
  23. Luo, X.; Zhang, T.; Tang, H.; Liu, J. Novel electrochemical and electrochemiluminescence dual-modality sensing platform for sensitive determination of antimicrobial peptides based on probe encapsulated liposome and nanochannel array electrode. Front. Nutr. 2022, 9, 962736. [Google Scholar] [CrossRef] [PubMed]
  24. Choi, D.; Yang, J.; Hong, S.; Park, J. Molecularly imprinted polymer-based electrochemical impedimetric sensors on screen-printed carbon electrodes for the detection of trace cytokine IL-1beta. Biosens. Bioelectron. 2022, 204, 114073. [Google Scholar] [CrossRef] [PubMed]
  25. Madej, M.; Lipińska, J.; Kapica, R.; Kędzierska-Sar, A.; Frątczak, E.; Kochana, J.; Baś, B.; Tyczkowski, J. Innovative modification of the surface of screen-printed carbon electrodes by nanofilms directly deposited in cold acrylonitrile plasma. Electrochem. Commun. 2022, 137, 107263. [Google Scholar] [CrossRef]
  26. Stoica, B.; Gavrila, A.; Sarbu, A.; Iovu, H.; Brisset, H.; Miron, A.; Iordache, T. Uncovering the behavior of screen-printed carbon electrodes modified with polymers molecularly imprinted with lipopolysaccharide. Electrochem. Commun. 2021, 124, 106965. [Google Scholar] [CrossRef]
  27. Yao, J.; Zhang, Z.; Huang, X.; Guo, Y. Blocker displacement amplification mediated PCR based screen-printed carbon electrode biosensor and lateral flow strip strategy for CYP2C19*2 genotyping. Biosens. Bioelectron. 2022, 207, 114138. [Google Scholar] [CrossRef]
  28. Wang, G.; Han, R.; Su, X.; Li, Y.; Xu, G.; Luo, X. Zwitterionic peptide anchored to conducting polymer PEDOT for the development of antifouling and ultrasensitive electrochemical DNA sensor. Biosens. Bioelectron. 2017, 92, 396–401. [Google Scholar] [CrossRef]
  29. Xu, Z.; Han, R.; Liu, N.; Gao, F.; Luo, X. Electrochemical biosensors for the detection of carcinoembryonic antigen with low fouling and high sensitivity based on copolymerized polydopamine and zwitterionic polymer. Sens. Actuators B Chem. 2020, 319, 128253. [Google Scholar] [CrossRef]
  30. Zhou, P.; Yao, L.; Chen, K.; Su, B. Silica Nanochannel membranes for electrochemical analysis and molecular sieving: A comprehensive review. Crit. Rev. Anal. Chem. 2019, 50, 424–444. [Google Scholar] [CrossRef]
  31. Zhou, H.; Ma, X.; Sailjoi, A.; Zou, Y.; Lin, X.; Yan, F.; Su, B.; Liu, J. Vertical silica nanochannels supported by nanocarbon composite for simultaneous detection of serotonin and melatonin in biological fluids. Sens. Actuators B Chem. 2022, 353, 131101. [Google Scholar] [CrossRef]
  32. Walcarius, A. Electroinduced surfactant self-assembly driven to vertical growth of oriented mesoporous films. Acc. Chem. Res. 2021, 54, 3563–3575. [Google Scholar] [CrossRef]
  33. Ma, K.; Zheng, Y.; An, L.; Liu, J. Ultrasensitive immunosensor for prostate-specific antigen based on enhanced electrochemiluminescence by vertically ordered mesoporous silica-nanochannel film. Front. Chem. 2022, 10, 851178. [Google Scholar] [CrossRef]
  34. Yang, L.; Zhang, T.; Zhou, H.; Yan, F.; Liu, Y. Silica nanochannels boosting Ru(bpy)32+-mediated electrochemical sensor for the detection of guanine in beer and pharmaceutical samples. Front. Nutr. 2022, 9, 987442. [Google Scholar] [CrossRef]
  35. Su, R.; Tang, H.; Xi, F. Sensitive electrochemical detection of p-nitrophenol by pre-activated glassy carbon electrode integrated with silica nanochannel array film. Front. Chem. 2022, 10, 954748. [Google Scholar] [CrossRef]
  36. Wang, K.; Yang, L.; Huang, H.; Lv, N.; Liu, J.; Liu, Y. Nanochannel array on electrochemically polarized screen printed carbon electrode for rapid and sensitive electrochemical determination of clozapine in human whole blood. Molecules 2022, 27, 2739. [Google Scholar] [CrossRef]
  37. Zou, Y.; Zhou, X.; Xie, L.; Tang, H.; Yan, F. Vertically-ordered mesoporous silica films grown on boron nitride-graphene composite modified electrodes for rapid and sensitive detection of carbendazim in real samples. Front. Chem. 2022, 10, 939510. [Google Scholar] [CrossRef]
  38. Zhang, M.; Zou, Y.; Zhou, X.; Yan, F.; Ding, Z. Vertically-ordered mesoporous silica films for electrochemical detection of Hg(II) ion in pharmaceuticals and soil samples. Front. Chem. 2022, 10, 952936. [Google Scholar] [CrossRef]
  39. Zhu, X.; Xuan, L.; Gong, J.; Liu, J.; Wang, X.; Xi, F.; Chen, J. Three-dimensional macroscopic graphene supported vertically-ordered mesoporous silica-nanochannel film for direct and ultrasensitive detection of uric acid in serum. Talanta 2022, 238, 123027. [Google Scholar] [CrossRef]
  40. Gong, J.; Zhang, T.; Chen, P.; Yan, F.; Liu, J. Bipolar silica nanochannel array for dual-mode electrochemiluminescence and electrochemical immunosensing platform. Sens. Actuators B Chem. 2022, 368, 132086. [Google Scholar] [CrossRef]
  41. Gong, J.; Zhang, T.; Luo, T.; Luo, X.; Yan, F.; Tang, W.; Liu, J. Bipolar silica nanochannel array confined electrochemiluminescence for ultrasensitive detection of SARS-CoV-2 antibody. Biosens. Bioelectron. 2022, 215, 114563. [Google Scholar] [CrossRef] [PubMed]
  42. Liu, J.; Guo, S.; Han, L.; Ren, W.; Liu, Y.; Wang, E. Multiple pH-responsive graphene composites by non-covalent modification with chitosan. Talanta 2012, 101, 151–156. [Google Scholar] [CrossRef] [PubMed]
  43. Hummers, W.; Offeman, R. Preparation of graphitic oxide. J. Am. Chem. Soc. 1958, 80, 1339. [Google Scholar] [CrossRef]
  44. Ma, K.; Yang, L.; Liu, J.; Liu, J. Electrochemical sensor nanoarchitectonics for sensitive detection of uric acid in human whole blood based on screen-printed carbon electrode equipped with vertically-ordered mesoporous silica-nanochannel film. Nanomaterials 2022, 12, 1157. [Google Scholar] [CrossRef]
  45. Liu, X.; Chen, Z.; Wang, T.; Jiang, X.; Qu, X.; Duan, W.; Xi, F.; He, Z.; Wu, J. Tissue imprinting on 2D nanoflakes-capped silicon nanowires for lipidomic mass spectrometry imaging and cancer diagnosis. ACS Nano 2022, 16, 6916–6928. [Google Scholar] [CrossRef]
  46. Cui, Y.; Duan, W.; Jin, Y.; Wo, F.; Xi, F.; Wu, J. Ratiometric fluorescent nanohybrid for noninvasive and visual monitoring of sweat glucose. ACS Sens. 2020, 5, 2096–2105. [Google Scholar] [CrossRef]
  47. Wan, Y.; Zhao, J.; Deng, X.; Chen, J.; Xi, F.; Wang, X. Colorimetric and fluorescent dual-modality sensing platform based on fluorescent nanozyme. Front. Chem. 2021, 9, 774486. [Google Scholar] [CrossRef]
  48. Deng, X.; Zhao, J.; Ding, Y.; Tang, H.; Xi, F. Iron and nitrogen co-doped graphene quantum dots as highly active peroxidases for the sensitive detection of l-cysteine. New J. Chem. 2021, 45, 19056–19064. [Google Scholar] [CrossRef]
  49. Chang, Q.; Huang, J.; He, L.; Xi, F. Simple immunosensor for ultrasensitive electrochemical determination of biomarker of the bone metabolism in human serum. Front. Chem. 2022, 10, 940795. [Google Scholar] [CrossRef]
  50. Zhou, H.; Ding, Y.; Su, R.; Lu, D.; Tang, H.; Xi, F. Silica nanochannel array film supported by ß-cyclodextrin-functionalized graphene modified gold film electrode for sensitive and direct electroanalysis of acetaminophen. Front. Chem. 2022, 9, 812086. [Google Scholar] [CrossRef]
  51. Gong, J.; Tang, H.; Wang, M.; Lin, X.; Wang, K.; Liu, J. Novel three-dimensional graphene nanomesh prepared by facile electro-etching for improved electroanalytical performance for small biomolecules. Mater. Design 2022, 215, 110506. [Google Scholar] [CrossRef]
  52. Zhou, H.; Dong, G.; Sailjoi, A.; Liu, J. Facile pretreatment of three-dimensional graphene through electrochemical polarization for improved electrocatalytic performance and simultaneous electrochemical detection of catechol and hydroquinone. Nanomaterials 2022, 12, 65. [Google Scholar] [CrossRef]
  53. Liu, Q.; Zhong, H.; Chen, M.; Zhao, C.; Liu, Y.; Xi, F.; Luo, T. Functional nanostructure-loaded three-dimensional graphene foam as a non-enzymatic electrochemical sensor for reagentless glucose detection. RSC Adv. 2020, 10, 33739–33746. [Google Scholar] [CrossRef]
  54. Shervedani, R.; Hatefi-Mehrjardi, A. Electrochemical characterization of directly immobilized glucose oxidase on gold mercaptosuccinic anhydride self-assembled monolayer. Sens. Actuators. B Chem. 2007, 126, 415–423. [Google Scholar] [CrossRef]
  55. Boonyasit, Y.; Chailapakul, O.; Laiwattanapaisal, W. A multiplexed three-dimensional paper-based electrochemical impedance device for simultaneous label-free affinity sensing of total and glycated haemoglobin: The potential of using a specific single-frequency value for analysis. Anal. Chim. Acta 2016, 936, 1–11. [Google Scholar] [CrossRef]
  56. Pinyorospathum, C.; Chaiyo, S.; Sae-Ung, P.; Hoven, V.; Damsongsang, P.; Siangproh, W.; Chailapakul, O. Disposable paper-based electrochemical sensor using thiol-terminated poly(2-methacryloyloxyethyl phosphorylcholine) for the label-free detection of C-reactive protein. Microchim. Acta 2019, 186, 472. [Google Scholar] [CrossRef]
  57. Rong, Z.; Chen, F.; Jilin, Y.; Yifeng, T. A C-reactive protein immunosensor based on platinum nanowire/titania nanotube composite sensitized electrochemiluminescence. Talanta 2019, 205, 120135. [Google Scholar] [CrossRef]
  58. Kowalczyk, A.; Sek, J.; Kasprzak, A.; Poplawska, M.; Grudzinski, I.; Nowicka, A. Occlusion phenomenon of redox probe by protein as a way of voltammetric detection of non-electroactive C-reactive protein. Biosens. Bioelectron. 2018, 117, 232–239. [Google Scholar] [CrossRef]
  59. Sonuc Karaboga, M.; Sezginturk, M. Determination of C-reactive protein by PAMAM decorated ITO based disposable biosensing system: A new immunosensor design from an old molecule. Talanta 2018, 186, 162–168. [Google Scholar] [CrossRef]
  60. Vilian, A.; Kim, W.; Park, B.; Oh, S.; Kim, T.; Huh, Y.; Hwangbo, C.; Han, Y. Efficient electron-mediated electrochemical biosensor of gold wire for the rapid detection of C-reactive protein: A predictive strategy for heart failure. Biosens. Bioelectron. 2019, 142, 111549. [Google Scholar] [CrossRef]
Figure 1. Schematic illustration for fabrication of electrochemical immunosensor on NH2-VMSF modified SPCE and the detection of CRP based on gated electrochemical signal.
Figure 1. Schematic illustration for fabrication of electrochemical immunosensor on NH2-VMSF modified SPCE and the detection of CRP based on gated electrochemical signal.
Nanomaterials 12 03981 g001
Figure 2. (a) Cyclic voltammetry (CV) curves obtained on different electrodes in 0.05 M KHP containing 0.5 mM Fe(CN)63−. (b) EIS curves obtained on different electrodes. The electrolyte solution is 2.5 mM Fe(CN)63−/4− containing 0.1 M KCl, The inset is the magnified EIS plots on SPCE and ErGO/SPCE.
Figure 2. (a) Cyclic voltammetry (CV) curves obtained on different electrodes in 0.05 M KHP containing 0.5 mM Fe(CN)63−. (b) EIS curves obtained on different electrodes. The electrolyte solution is 2.5 mM Fe(CN)63−/4− containing 0.1 M KCl, The inset is the magnified EIS plots on SPCE and ErGO/SPCE.
Nanomaterials 12 03981 g002
Figure 3. Top-view TEM image of NH2-VMSF at different magnification.
Figure 3. Top-view TEM image of NH2-VMSF at different magnification.
Nanomaterials 12 03981 g003
Figure 4. High-resolution C 1 s XPS spectra of GO/SPCE (a) or ErGO/SPCE (b) electrode.
Figure 4. High-resolution C 1 s XPS spectra of GO/SPCE (a) or ErGO/SPCE (b) electrode.
Nanomaterials 12 03981 g004
Figure 5. (a) CV curves obtained on different electrodes obtained in the fabrication of the immunosensor in 0.1 M KCl containing 1.25 mM Fe(CN)63−. (b) EIS curves obtained on different electrodes obtained in the fabrication of the immunosensor. The electrolyte solution is 2.5 mM Fe(CN)63−/4− containing 0.1 M KCl. Inset is the illustration for the equivalent circuit. (c) Effect of incubation time between Ab on the aldehyde surface on the current response of the proposed immunosensor. (d) Effect of incubation time between the immuno-recognitive surface and the detected CRP on the current response of the proposed immunosensor. The error bars represent the relative standard deviation (RSD) of three measurements. Inset in (d) is CV curve obtained on different electrodes. The electrolyte solution is 2.5 mM Fe(CN)63−/4− containing 0.1 M KCl.
Figure 5. (a) CV curves obtained on different electrodes obtained in the fabrication of the immunosensor in 0.1 M KCl containing 1.25 mM Fe(CN)63−. (b) EIS curves obtained on different electrodes obtained in the fabrication of the immunosensor. The electrolyte solution is 2.5 mM Fe(CN)63−/4− containing 0.1 M KCl. Inset is the illustration for the equivalent circuit. (c) Effect of incubation time between Ab on the aldehyde surface on the current response of the proposed immunosensor. (d) Effect of incubation time between the immuno-recognitive surface and the detected CRP on the current response of the proposed immunosensor. The error bars represent the relative standard deviation (RSD) of three measurements. Inset in (d) is CV curve obtained on different electrodes. The electrolyte solution is 2.5 mM Fe(CN)63−/4− containing 0.1 M KCl.
Nanomaterials 12 03981 g005
Figure 6. (a) Differential pulse voltametric curves of the immunosensor to various concentrations of CRP. Inset was the corresponding calibration curve for the detection of CRP. (b) The ratio of peak current when the immunosensor was incubated with different protein or their mixture. I0 and I were the peak current before and after the incubation. The concentration of each protein was 1 ng/mL. The mixture contained CRP (1 ng/mL) and all other proteins (each concentration was 1 ng/mL). The immunosensors were incubated with different protein or their mixture for 60 min at 37 °C. The error bars represent the relative standard deviation (RSD) of three measurements.
Figure 6. (a) Differential pulse voltametric curves of the immunosensor to various concentrations of CRP. Inset was the corresponding calibration curve for the detection of CRP. (b) The ratio of peak current when the immunosensor was incubated with different protein or their mixture. I0 and I were the peak current before and after the incubation. The concentration of each protein was 1 ng/mL. The mixture contained CRP (1 ng/mL) and all other proteins (each concentration was 1 ng/mL). The immunosensors were incubated with different protein or their mixture for 60 min at 37 °C. The error bars represent the relative standard deviation (RSD) of three measurements.
Nanomaterials 12 03981 g006
Table 1. Comparison between the performance of different electrochemical sensors for the determination of CRP.
Table 1. Comparison between the performance of different electrochemical sensors for the determination of CRP.
Electrode MaterialDetection MethodConstruction MethodLinear Range
(ng/mL)
LOD
(ng/mL)
Ref.
PMPC-SH/AuNPs-SPCEDPVLabel-free5–50001.6[56]
TiNTs/PtNWs/ITOECLLabel-free0.05–6.250.011[57]
AuNPs/SPGEDPVLabel-based10–15,0001.5[12]
GCE/PEI-Fc/Ab/CRPDPVLabel-free1–5×1040.5[58]
11-CUTMS/PAMAMEISLabel-free2.1 × 10−5–6.148 × 10−33.4 × 10−7[59]
MPA/Au substrateSWVLabel-free5 × 10−6–2.2 × 10−42.25 × 10−6[60]
Ab/GA/NH2-VMSF/ErGO/SPCEDPVLabel-free0.01–1008 × 10−3This work
Table 2. Determination of CRP in human serum samples.
Table 2. Determination of CRP in human serum samples.
SampleSpiked
(ng/mL)
Found
(ng/mL)
Recovery (%)RSD (%)
Human serum a0.1000.1021024.1
1.000.97697.63.3
10099.199.13.8
a Samples were diluted by a factor of 50. The indicated concentration is the concentration of CRP after dilution. The error bars represent the relative standard deviation (RSD) of three measurements.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Ma, N.; Luo, X.; Wu, W.; Liu, J. Fabrication of a Disposable Electrochemical Immunosensor Based on Nanochannel Array Modified Electrodes and Gated Electrochemical Signals for Sensitive Determination of C-Reactive Protein. Nanomaterials 2022, 12, 3981. https://doi.org/10.3390/nano12223981

AMA Style

Ma N, Luo X, Wu W, Liu J. Fabrication of a Disposable Electrochemical Immunosensor Based on Nanochannel Array Modified Electrodes and Gated Electrochemical Signals for Sensitive Determination of C-Reactive Protein. Nanomaterials. 2022; 12(22):3981. https://doi.org/10.3390/nano12223981

Chicago/Turabian Style

Ma, Ning, Xuan Luo, Weidong Wu, and Jiyang Liu. 2022. "Fabrication of a Disposable Electrochemical Immunosensor Based on Nanochannel Array Modified Electrodes and Gated Electrochemical Signals for Sensitive Determination of C-Reactive Protein" Nanomaterials 12, no. 22: 3981. https://doi.org/10.3390/nano12223981

APA Style

Ma, N., Luo, X., Wu, W., & Liu, J. (2022). Fabrication of a Disposable Electrochemical Immunosensor Based on Nanochannel Array Modified Electrodes and Gated Electrochemical Signals for Sensitive Determination of C-Reactive Protein. Nanomaterials, 12(22), 3981. https://doi.org/10.3390/nano12223981

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.