1. Introduction
Indoor air quality has attracted significant attention for improving the comfort and health of indoor occupants. Poor indoor air quality is becoming a serious health concern as it can lead to health problems, impaired learning, and reduced productivity [
1]. Indoor air mainly contains Volatile Organic Compounds (VOCs), airborne particles, microorganisms, household odours, and gases. VOCs constitute a considerable portion of indoor airborne pollutants [
2,
3,
4]. Among these VOCs, Benzene, Toluene, Ethylbenzene, and Xylenes (BTEX) pose an alarming threat to human health. They are aromatic hydrocarbons and its common indoor emission sources are coal burning, cigarette smoking, combustion, use of cleaning products, laser printing, floor adhesives, paint, wood panelling, and evaporation of fuel from an adjoining garage [
5,
6,
7]. The infiltration of outside air can also contribute to high BTEX levels, particularly near industrial sources or high road traffic. Long-term exposure to airborne BTEX causes fatigue, respiratory, kidney, nervous system, mental, and cardiovascular-related problems [
8,
9,
10,
11]. Among BTEX, benzene is carcinogenic, as classified by the International Agency for Research on Cancer [
12]. There are strict regulations about the exposure limit of BTEX in indoor spaces. For instance, the National Institute of Occupational Safety and Health (NIOSH) and the Occupational Safety and Health Administration (OSHA) have set an exposure limit of 0.1 ppm and 1 ppm, respectively, for benzene in indoor spaces. For toluene, the exposure limit established by NIOSH and OSHA is 100 ppm and 200 ppm, respectively [
13,
14].
A highly sensitive gas detector having excellent reliability and ease of integration for instance with GC is therefore needed for monitoring of BTEX in indoor spaces. Sensors coupled with micro gas chromatography (μGC), usually called BTEX analysers, can detect each component in the BTEX mixture [
15,
16,
17]. Optical gas sensors have demonstrated high sensitivity and selectivity for detection of different airborne gases. They have a quick time response with minimal drift and poisoning issues in comparison to other sensing methods for instance electrochemical sensors and metal-oxide semiconductor gas sensors [
14]. BTEX molecules absorb strongly in the wavelength range of 240 nm to 275 nm [
18]. Thus, absorption spectrometry offers a sensitive, rapid, and non-destructive approach for airborne BTEX detection. This optical detection method is self-referenced and reliable with a minimal cross-response to other molecules. Such optical sensors are based on directly measuring the inherent physical property of a molecule, i.e., absorbance of a molecule at a specific wavelength, which effectively acts as a fingerprint. Recent advancements in optoelectronics and optofluidics have led to the development of LEDs, photodiodes, and hollow-core waveguides (HCW), for the UV-range which can be exploited to develop a portable deep UV absorbance detector.
LEDs provide a portable and quasi-monochromatic UV emission source with good spatial stability, low power input, and lower cost compared to conventional UV emission sources, like mercury lamps. Aside from their cost and bulk size, it is not possible to achieve the target wavelength range for BTEX with Deuterium and Xenon lamps without costly optical filters. For the detection part, photodiodes offer a portable and an efficient UV photodetector: They exhibit good photocurrent linearity, visible-blindness, low-dark current, high quantum efficiency, low input power and noise, portability, and low cost [
19]. In addition to a UV source and photodetector, a gas cell with a high length-to-volume is required to have a compact instrument compatible with GC for the detection of BTEX mixture at ppb or sub-ppm level. In view of these constraints, Hollow-Core Waveguides (HCW) is an excellent candidate that can be applied as a miniature gas cell with an efficient radiation transmission. HCWs transmit UV radiation in a leaky-mode through the reflection of the emitted radiation inside the co-axial hollow core. Aluminium is considered to be an efficient coating material for deep-UV radiation compared to other metals, for instance silver, and it has good chemical compatibility with airborne BTEX molecules [
20].
BTEX components have overlapping spectra and the detection of individual component with high sensitivity in BTEX mixture is challenging. However, an absorbance detector compatible with μGC can separate, identify, and detect each molecule of BTEX mixture. Such a photometer needs a gas cell with a microliter inner volume and larger length having minimum dead volume to detect each BTEX molecule with an individual peak. Previous research on BTEX detection was primarily focussed on the sensitivity of the sensor and mostly an individual BTEX molecule was tested. Eckhardt et al. [
21,
22] developed a detector with an aluminium-coated HCW (length, 1 m and diameter, 1 mm) as a gas cell, deuterium lamp (30 W) as UV source, and CCD-based spectrometer as a UV detector. The setup was coupled with GC and tested for detection of aromatics compounds like ethylbenzene and bromobenzene. Its main limitations are the high volume of the gas cell (785 μL) and the overall size of the spectrophotometer (emission source and detector). Bui et al. [
23] employed deep-UV LED (λ
peak = 260 nm) and photodiodes for the direct detection of BTEX with good sensitivity (152 μA·U-ppm
−1; limit of detection of 0.66 ppm for toluene). An aluminium-based gas cell (length, 40 cm and inner diameter, 2 mm) was used with an inner volume of 1256 μL. The higher inner volume of the gas cell, which limits its use for GC application. Khan et al. [
24] demonstrated a simple toluene detector using UV-LED, HCWs, and spectrometer. Two types of HCWs were investigated: Glass-coated HCW and aluminium-based HCW. A limit of detection of 8.15 ppm was reported with aluminium-coated glass HCW with an inner volume of 275 μL. The sensitivity and high dead volume are its main drawbacks, which limit its application for GC analysis.
In this paper, the development of deep-UV absorbance detector compatible with μGC for detection of the individual BTEX components is reported. The design is based on using deep-UV LED, photodiodes, and HCW with a microliter volume gas cell with negligible dead-volume. The design allows direct measurement of absorbance without mono-chromator and ease of coupling with μGC. To the best of our knowledge, a sensitive and portable deep-UV absorbance detector based on deep-UV LED, photodiodes, and HCW with microliter volume having good compatibility with μGC has not been reported in the literature.
3. Results and Discussion
The absorbance detector is designed to directly measure the absorbance of the gas components as a voltage, which is recorded with a multimeter. The emitted UV radiation of the LED is sensitive towards the input current and temperature of the LED. To minimize the fluctuation of the signal, a referencing technique is applied. The UV light emitted from the LED is split into two beams using a fibre-coupled beam splitter, one beam
Io acts as a reference, while the other serves as a measurement beam
I, passes through the gas cell, as represented in
Figure 1b. The two intensities (i.e.,
Io and
I) are recorded using photodiodes, which generate photocurrents
io and
i, respectively. The photocurrents generated are proportional to the intensity of the incident UV light. A log-ratio amplifier processes the two photocurrents to compute the absorbance by taking a logarithmic ratio of
io and
i, and gives a single output in volts. The output voltage represents the absorbance according to the Beer-Lambert equation which can be expressed as
where
Av is the absorbance in volts,
io and
i represents the photocurrents from the reference photodiode and test photodiode, respectively.
The absorption spectrum of the target BTEX molecules, were initially investigated and compared with the emission spectrum of the LED. The absorption data of the molecules were collected from high-resolution transmission molecular absorption database (HITRAN) [
18] and were mapped with the emission spectrum of the LED as shown in
Figure 4a. The emission spectrum of the LED was recorded using the mini-spectrometer (C10082CH, Hamamatsu, Hamamatsu city, Japan) without passing through the gas cell. The LED spectrum covers a significant portion of the absorption spectra of the BTEX components for a wavelength range of 245 nm to 275 nm with a peak at 260 nm. The maximum peak absorption cross-sections σ of benzene, toluene, ethylbenzene, m-xylene, o-xylene, and p-xylene with the corresponding peak wavelength are summarized in
Table 1. For comparison, the integrated absorbance for wavelength range (245 nm to 275 nm) was calculated by using area integral of the absorption spectrum as given in
Figure 4b. This data is important during detection of the BTEX mixture, as the detection of the components is dictated by their corresponding absorbance. p-xylene shows the highest absorption, while benzene shows the lowest absorption, implying that detection of p-xylene is easier compared to other components.
3.1. Detection of Toluene
The absorbance detector was first tested with a direct injection of different concentrations of toluene in N
2. The experiments were carried out by first injecting pure N
2, followed by injection of the toluene-N
2 mixture for 5 min. When the toluene molecules were injected into the gas cell, the output voltage recorded increased due to the absorbance of toluene molecules. For instance, the recorded signal for the toluene-N
2 mixture of 5 ppm and 20 ppm is shown in
Figure 5. Each concentration was injected three times with pure N
2 injections in-between. A significant rise in the signal was observed, as the toluene concentrations were injected into the cell. High spikes were observed in the absorbance signal at the instant when the toluene cylinder value was turned ON. The gas dynamics inside the capillary tube at the instant of valve opening could be the reason for these instant spikes. No such spikes were observed in case of BTEX detection with μGC, which is discussed in the following section.
The experimental device was then tested with direct injection of a toluene-N
2 concentration of 1.5 ppm to 50 ppm. A calibration curve for concentrations (1.5 ppm to 50 ppm) of toluene is shown in
Figure 6. Each data point on the calibration curve represents an average of three observations while each observation is an average of absorbance values for five mins. The design demonstrated good linearity (R
2 = 0.99) for the investigated concentration range 1.5–50 ppm with a sensitivity of 1.76 × 10
−4 A·U-ppm
−1. A limit of detection of 196 ppb was derived from the curve by using a concentration corresponding to a signal with value three times higher than the baseline noise. The baseline noise was calculated by using a standard deviation of the blank obtained with pure nitrogen flow for a time duration of 250 s (
Figure S3) and repeated three times.
3.2. Detection of BTEX with μGC
The detector was then tested for detection of BTEX mixtures by coupling it with the μGC setup developed by our research group. The μGC process for separation and detection of the BTEX mixture was carried out in two modes, i.e., the sampling mode and analysis mode, as described in the
Section 2.1. The separation and detection of BTEX were conducted in the analysis mode. In the sampling mode, the BTEX species were sampled into a loop of fixed volume (200 μL) at a flow rate of 30–50 mL/min (P = 1 bar, T = 25 °C) for 15 s. The carrier gas N
2 was passing through the column and detector at a flow rate of 2.45 mL/min. The transition from sampling mode to analysis mode was carried out with the 6-ports valve. During the analysis mode, the 6-ports valve was turned ON and the BTEX components were injected into the μGC. The separated species were then injected into the absorbance detector as shown in
Figure 3. In the GC analysis, the 6-ports valve switches very rapidly between sampling and analysis modes, whereas the pressure was maintained constant by a forward pressure controller. There was no valve opening/closing involved, therefore no spikes were observed at the instant of injection of BTEX.
A total BTEX-N
2 mixture of 60 ppm with each compound of 10 ppm was injected into GC. The μGC column’s temperature was maintained at 60 °C. A chromatogram of 700 s was recorded in which all the peaks were separated, corresponding to benzene, toluene, ethylbenzene, (m-,p-) xylenes, and o-xylene in temporal sequence, as shown in
Figure 7 (red line). Each peak represents an individual BTEX molecule. The elution sequence of BTEX molecule is performed according to the boiling points of BTEX components. The boiling points of m-xylene and p-xylene are very close (i.e., 139.1 °C and 138.4 °C, respectively) so that they exhibit coelution. (m-,p-) xylenes have the highest peak followed by o-xylene, toluene, ethylbenzene, and benzene, which is in accordance with the absorbance value of the components for the wavelength range of 245 nm to 275 nm (see
Figure 4b). The device was then tested for total BTEX-N
2 mixture of 30 ppm and 15 ppm, and the concentration of the individual molecule was 5 ppm (orange line) and 2.5 ppm (yellow line), respectively. The chromatograms for the separated BTEX species are shown in
Figure 7. All the BTEX components were detected for all three concentrations injected. However, a weak signal was obtained for 2.5 ppm with no clear signal for benzene and ethylbenzene. This is due to the relatively lower absorption coefficient of benzene and ethylbenzene.
An area of each peak was calculated for the BTEX molecule and plotted for the three concentrations (10 ppm, 5 ppm, 2.5 ppm) as shown in
Figure 8. The detector demonstrated good linearity for all the BTEX components with R
2 ~ 0.99. It should be noted that only three data points were used for calibration due to the limitation of the BTEX cylinder and MFCs employed in the concentration generation setup. A sensitivity of 1.91 × 10
−4 A·U-ppm
−1, 6.55 × 10
−4 A·U-ppm
−1, 1.34 × 10
−4 A·U-ppm
−1, 9.58 × 10
−4 A·U-ppm
−1, and 5.70 × 10
−4 A·U-ppm
−1 was calculated for benzene, toluene, ethylbenzene, (m-,p-) xylenes, and o-xylene, respectively. The pattern of sensitivities is according to the absorbance cross-section values of the species.
Furthermore, experiments were conducted at a different column temperature of 40 °C, 60 °C, and 80 °C as illustrated in
Figure 9. At the lowest temperature, i.e., 40 °C, all the BTEX were distinctively separated at a cost of extended time for the chromatogram (see
Figure 9a). At the highest temperature, i.e., 80 °C, a chromatogram was obtained in a shorter time with lower values for peaks resolution. The peak for ethylbenzene is not discernible for the case of 80 °C, as observed in
Figure 9c. Therefore, the optimal temperature was found to be 60 °C, as already reported by Nasreddine et al. [
27], where they used a photo-ionisation detection module. The peaks-resolution values for BTEX components at the three studied temperatures are summarized in
Table 2. The degree of separation between the two peaks (
a and
b) was quantified by using a resolution,
R, as given by [
28],
where
tr and
w represent the retention time and baseline width (full width at half maximum—FWHM), respectively. At lower temperature (i.e., 40 °C), peak resolutions have higher values (about two times higher) compared to those obtained with the column temperature of 60 °C.
During testing all the gases were flushed into the systems from cylinders with a negligible level of humidity. Although the detector signal is not affected by humidity in the air. However, humidity could absorb into the reflective surface of the HCW and decrease the transmission of UV radiation with time. Nafion membrane at the inlet or quartz lining of the aluminium coating are usually used to avoid the negative effects of humidity on the sensing system [
20].
4. Conclusions
A novel, self-referenced deep-UV absorbance detector was developed for the detection of BTEX components. The absorbance detector composed of portable devices: UV-LED, HCW as a gas cell, and photodiodes. The detector directly measures the absorbance values in volts. The design is based on using a log-ratio amplifier, which used the photocurrents from a reference and measurement photodiodes to compute the absorbance in volts according to Beer–Lambert law. The design was tested with direct injection of toluene and with a mixture of BTEX species. For toluene, the detector response increases linearly with the gaseous toluene concentration in the range 1.5–50 ppm, i.e., toluene levels consistent with the exposure limits of 100–200 ppm given by NIOSH and OSHA. A corresponding sensitivity and limit of detection of 1.76 × 10−4 A·U-ppm−1 and 196 ppb, respectively, were obtained for toluene. The absorbance detector was also coupled with μGC set up, enabling the separation of BTEX components prior to their detection by UV absorbance detector. All the BTEX components were separated and detected with an individual peak correspond to each BTEX molecule for a concentration range of 10 ppm to 2.5 ppm. The design was also tested at a different column temperature of 40, 60, and 80 °C. The proposed detection device has low volume with minimal dead-volume and directly measures the absorbance with a stable signal. The design has tremendous potential to be employed for the chromatographic studies of different analytes in media like gas or liquid.