Article

Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam

1
Department of Bio-convergence Engineering, Korea University, Seoul 02841, Korea
2
Proton Therapy Center, National Cancer Center, Goyang 10408, Korea
3
Department of Medical Physics, Chinan Biomedical Technology Inc., Zhubei 30268, Taiwan
*
Authors to whom correspondence should be addressed.
Sensors 2018, 18(1), 227; https://doi.org/10.3390/s18010227
Submission received: 24 November 2017 / Revised: 11 January 2018 / Accepted: 13 January 2018 / Published: 15 January 2018
(This article belongs to the Special Issue Optical Sensors based on Micro/Nanofibres)

Abstract

:
This study describes the development of a beam monitoring system for the verification of entrance dose map in pencil beam scanning (PBS) proton therapy based on fiber optic radiation sensors (FORS) and the validation of this system through a feasibility study. The beam monitoring system consisted of 128 optical fibers optically coupled to photo-multiplier tubes. The performance of the beam monitoring system based on FORS was verified by comparing 2D dose maps of square-shaped fields of various sizes, which were obtained using conventional dosimeters such as MatriXX and EBT3 film, with those measured using FORS. The resulting full-width at half maximum and penumbra were compared for PBS proton beams, with a ≤2% difference between each value, indicating that measurements using the conventional dosimetric tool corresponded to measurements based on FORS. For irregularly-shaped fields, a comparison based on the gamma index between 2D dose maps obtained using MatriXX and EBT3 film and the 2D dose map measured by the FORS showed passing rates of 96.9 ± 1.3% and 96.2 ± 1.9%, respectively, confirming that FORS-based measurements for PBS proton therapy agreed well with those measured using the conventional dosimetric tools. These results demonstrate that the developed beam monitoring system based on FORS is good candidate for monitoring the entrance dose map in PBS proton therapy.

1. Introduction

Intensity modulated radiotherapy (IMRT), tomotherapy, and proton therapy are representative examples of novel radiotherapy methods, which can lead to better results when treating cancer than conventional 3D conformal radiotherapy [1,2,3]. In particular, proton therapy can treat cancer cells more effectively due to the characteristics of the Bragg peaks, which can deliver the maximum dose at the desired depth, rather than at a shallow depth [4,5,6]. Two types of beam delivery techniques are currently available for proton therapy: the passive scattering mode and pencil beam scanning (PBS). The passive scattering mode treats tumors in a desired block shape (tumor shape) using small-sized proton beams that scatter the protons, followed by spreading of the beams [7,8]. The PBS mode rapidly scans individually weighted spot beams without spreading small-sized proton beams and irradiates the tumors with the pencil beams. Because this method accurately controls the position and timing of proton beams, it needs sophisticated and precise treatment [9,10,11]. (Figure 1) The PBS mode uses weights and smaller beams than the passive scattering mode, delivering doses with the shape and size of the tumor. The treatment process of PBS mode involves diverse conditions, such as beam position, scanning velocity and the calculation of various complex factors, which results in planning that is less intuitive, all of which may lead to risk of radiation therapy accidents. To reduce the likelihood of this kinds of accidents, in X-ray therapy, beam monitoring systems were developed to monitor the beam delivered to the patient at the final stage of treatment (i.e., in the treatment room). Ion chamber-based integral quality monitoring (IQM, iRT Systems GmbH, Koblenz, Germany) and dolphin (Ion Beam Applications, Louvain-La-Neuve, Belgium) are representative commercially available beam monitoring systems for X-ray therapy. These systems are attached to a gantry and can measure irradiated beams in real-time during patient treatment. Subsequently, these systems determine if the actual irradiated dose is equal to the dose calculated by the treatment planning system (TPS) [12,13,14,15,16]. These systems, however, are expensive and have been developed for X-ray therapy only.
Although the MatriXX 2D ion chamber array (Ion Beam Applications) can be used as a dosimetric tool to measure PBS beams, this system is currently used for quality assurance only. Its thickness, however, makes it unsuitable for use as a beam monitoring system in the treatment room, because it is not easy to attach this device to a proton gantry. Moreover, the center-to-center distance between the ion chambers in the MatriXX is 7.62 mm, making it difficult to precisely and accurately measure PBS 3–4 mm proton beams [17]. Although various new dosimetric tools, including glass dosimeters, films, and TLDs, are both small and thin, they have several disadvantages, including their inability to perform real-time measurements and difficulty in accurately measuring the proton beam from a dosimetric perspective [18,19,20]. Although built-in ion chambers were already being used to verify the dose and position in proton therapy, the resolution of built-in equipment is not good enough to verify the entrance dose map. Therefore, new dosimetric tools are needed for use as beam monitoring systems in proton therapy.
Recent studies have suggested the potential of optical fiber for proton dosimetry [21,22,23,24,25]. Fiber optic radiation sensor (FORS) based on fluorescent light were recently shown to be useful for proton dosimetry because, unlike scintillating fibers, they have no quenching effect. Although there was argument about the physical origins of visible light measured by optical fiber during proton irradiation, it is plausible that the visible light signal, which is proportional to the absorbed dose of proton beam, is not coming from the Cerenkov radiation but from the fluorescence light of the fiber. Darafsheh et al. reported that the signal in fiber optic dosimetry of proton beams is the fluorescence light of the fiber based on the spectral analysis obtained from bare fiber during proton beam irradiation [21]. The previous study also suggests that FORS is independent about energy and dose rate by revealing a constant signal value (or light yield) for various energy and dose rate. In addition, the fluorescence light is linearly related to the irradiated dose in the proton beam [25]. This study describes our development of a FORS based the beam monitoring system for the verification of entrance dose map, involving optical fibers in array forms, and an evaluation of the suitability of FORS as a beam monitoring system for PBS proton beams by comparing this newly developed system with the conventional dosimetric system for PBS proton beams.

2. Materials and Methods

2.1. System Configuration

The two-dimensional beam monitoring system was based on FORS (see Figure 2). The plastic optical fibers (POF: BCF-98, SAINT-GOBAIN CRYSTALS, Nemours, France) used in this system were square in shape, 1 mm in size, and transmitted only fluorescent light to photomultiplier tube (PMT). The POFs chosen in this study are step index optical fibers with a core/cladding structure. The outer length of the square shaped POFs is 1.0 mm and the cladding thickness is 0.04 mm, and the core thickness is 0.96 mm. The core of the POFs was a polystyrene material of refractive index ncore = 1.60. Plastic optical fibers (an acrylic material with refractive index nclad = 1.49) were used as cladding. Each FORS consisted of two layers of optical fibers inserted into an enclosure frame made of acrylic materials of 3 mm total thickness. 128 optical fibers (64 fibers for x-coordinate, 64 fibers for y-coordinate) were arranged in the acrylic enclosure with center-to-center spacing of 2 mm (i.e., the edge-to-edge distance is 1 mm). The first layer was composed of 64 POFs arranged along the x-coordinate, and the second layer of 64 POFs arranged along the y-coordinate; thus, the same coordinate was rotated 90°. Although one dose plane is ~1 mm below the other, it seems to be OK to verify the entrance dose map at the treatment room. In our experiment, the optical fiber array was always installed in the plane perpendicular to the beam direction.
The POFs of the layers were optically coupled to the photomultiplier tubes (PMTs) of each channel (H12428-203, Hamamatsu Photonics, Hamamatsu, Japan). Each channel of PMT is connected to the electronic circuit which is designed to convert current signal from PMT into the voltage. The voltage generated in electronic circuit was amplified using pre-amp and transmitted to the NI-DAQ system (National Instruments, Austin, TX, USA) through the D-sub cable. Signals were obtained through the NI-DAQ 9188 and NI-9205 systems using self-developed software.
In our experimental system, PMT signals were processed using a data acquisition system with a 1 kHz acquisition frequency, meaning that one 2D-data set from 128 channels was acquired every 1 ms. Based on the fact that the beam current at nozzle is ~5 nA for maximum energy in PBS mode, the developed system can measure the dose with dose rate up to ~0.2 MU/ms. Therefore, if the dose rate is higher than this limit, the individual spots cannot be resolved and the higher speed DAQ system should be applied. To reduce the optical crosstalk, the optical fiber was inserted in “enclosure frame” and firmly attached to PMT tube. And then, noise signal from optical crosstalk was eliminated based on the comparison with reference measurement (i.e., ion chamber measurement).

2.2. PBS Beam Measurement

The beam monitoring system based on FORS was evaluated by obtaining the spot profile created by irradiating the beams of a square-shaped field on the FORS. Gaussian fitting was applied to determine the center (beam position) of the PBS proton beams. The measurement using high resolution detector showed that PBS beam has a Gaussian shape with different width for different energy. Because optical fiber array has a limited resolution (i.e., 2 mm), one had to fit the measured value to make the dose map. The measured Gaussian beam shape of proton pencil beam, which was also inserted to TPS, was used to fit our data. As seen in Figure 3, several fibers record a signal and each fiber corresponds to each x or y-coordinates. The signal of proton beam in each optical fiber is proportional to the length exposed to proton beam. The results of this system were compared with data obtained using MatriXX and EBT3 film, both of which are conventional dosimetric tools. Square PBS beams of different sizes were irradiated, and high energy (227.1 MeV) and low energy (122.6 MeV) beams were irradiated to determine the effect of irradiation on energy.
Five irregularly-shaped fields of various sizes were irradiated with PBS beams. Gaussian fitting was applied to the measured profiles of these irregularly-shaped fields for two energies (122.6 and 227.1 MeV), resulting in the center value for the PBS beam. The beam modeling value of the TPS was applied to the center value. As in the irradiation of square-shaped fields, 2D dose distribution obtained using FORS was compared with the data measured by MatriXX and EBT3 films.

2.3. Comparison Methods

To verify the performance of FORS, its determinations of full width half maximum (FWHM: 50–50%) and penumbra (20–80%) at the isocenter position were compared with those measured by MatriXX and EBT3 film. Doses to an irregularly-shaped field were determined using the gamma evaluation method, a quantitative 2D dose verification method [26]. The gamma evaluation method considers both the dose difference (DD) and distance-to-agreement (DTA) and confirms whether these tolerances are satisfied for each measured point. This method has been shown to detect a 3% DD in low-depth gradient areas and 3 mm DTA in high-depth gradient areas, making it sensitive to both components. A gamma index (GI) of less than 1, which corresponds to DD < 3% and DTA < 3 mm, represents good agreement between the values calculated by TPS and those measured by each dosimetry tool. GI was evaluated using commercially available Omnipro IMRT software for MatriXX measurements and commercially available RIT113 software for EBT3 film and FORS measurements.

3. Results

Figure 4 shows a 2D dose distribution of square-shaped PBS beams (proton energy: 122.6, 227.1 MeV) measured using MatriXX, EBT3 film, and FORS. Table 1 shows the FWHM and penumbra (20–80%) values on horizontal and vertical lines in the 2D dose distributions measured by MatriXX, EBT3 film, and FORS, as well as differences between the values measured using both MatriXX and EBT3 film and those measured using FORS. The differences in FWHM values (i.e., the width between 50% of the maximum dose) between MatriXX and FORS based values and between EBT3 film and FORS based measurements were 1.2 ± 0.5% (range, 0.4–1.8%) and 0.8 ± 0.6% (range, 0.2–1.6%), respectively (Table 1). The differences in the penumbra values (the width of 20–80% of the maximum dose) between the MatriXX and FORS based values and between the EBT3 film and FORS based measurements were 1.3 ± 0.5% (range, 0.3–2.0%) and 1.5 ± 0.5% (range, 0.0–2.0%), respectively (Table 1). The relationships between FWHM and penumbra with regard to various sizes of square-shaped fields show that the FORS system performed similarly to the conventional dosimetric tools.
Figure 5 shows an example of 2D dose distribution for PBS beams (energy: 122.6, 227.1 MeV) from irregularly-shaped field 4 in Table 2, which had been measured using MatriXX, EBT3 film and FORS. The GI passing rates (3%, 3 mm threshold level: 30%) of the 2D dose maps obtained using the values obtained for MatriXX and FORS were 96.8 ± 1.1% for the 122.6 MeV PBS beam and 97.0 ± 1.4% for the 227.1 MeV PBS beam. Similarly, the GI passing rates of the FORS relative to the EBT3 film measured values were 96.7 ± 1.0% for the 122.6 MeV PBS beam and 95.7 ± 2.3% for the 227.1 MeV PBS beam. Figure 6 shows a gamma image result obtained by applying the gamma evaluation method to FORS measured values relative to the MatriXX and EBT3 film measured values of irregularly-shaped field 5 in Table 2.

4. Discussion

The findings presented here indicate that a beam monitoring system based on fluorescence light can simultaneously verify the 2D dose distribution, including FWHM, penumbra, and gamma evaluation, in patients undergoing PBS proton therapy. The differences among patients treated with MatriXX, EBT3 Film, and FORS plans for the FWHM, penumbra, and failure rate in gamma index were small, indicating that the 2D dose distributions calculated using FORS were in good agreement with conventional dosimetric tools.
Beam monitoring is important in PBS proton therapy, because the dose map will change if some beams do not irradiate the planned position. A patient-specific QA method (pre-treatment QA) is used to validate the 2D dose distribution by comparing it with the calculated dose distribution obtained from the TPS. Although this method can detect systematic errors that occur during beam irradiation, it cannot measure the beams during patient treatment; rather it measures the beam during QA. However, measurement of pre-treatment QA cannot detect any unexpected random errors that result in irradiation with unwanted beams. One of the main advantages of the beam monitoring system based on FORS is its ability to evaluate beam delivery in real-time while a patient is being treated. This requires the dosimetric system to be physically thin and cause minimal proton beam interference, as the system should be located between the proton gantry and the patient. The system described in this report is therefore very promising, as it can be easily attached to the gantry, enabling proton beams to be measured during treatment without a large interference.
Although the FORS induced perturbation can be small, it is important to know how much of an influence the device has on the spot size, particularly for low energy. To examine the perturbation due to the device, the spot sizes (one sigma) were measured for various energies with/without the device (i.e., FORS) in air at isocenter with EBT3 film. Compared to original spot sizes, the increased spot sizes due to the device ranged from 0.20 mm to 0.28 mm for various energies (E = 102 MeV to 197 MeV). For low beam energy (E = 102.7 MeV) and very low beam energy (E = 102.7 MeV with 7cm range shifter), the increased spot sizes due to the device were 0.24 mm and 0.23 mm, respectively, which mean ~2% of increase in spot sizes. These results suggest that the increase in spot size due to FORS device is not significant compared to original spot size. To use the device during treatment, however, the perturbation should be resolved by taking the matter into account in the treatment planning stage. For example, although the range shifter or compensator in proton therapy can also cause the perturbation of incoming proton beam, the treatment planning system can be commissioned to include the perturbation from this device.
There are some limits in current study. First, the 2D dose maps from clinical cases should be compared to dose maps measured with developed FORS system. Second, high-speed DAQ system should be applied to resolve the high speed individual proton beam spots.

5. Conclusions

The results of this study suggest that the developed system, which addresses the drawbacks of conventional systems, has good real-time monitoring capabilities for pencil beam scanning proton therapy. Because it can be easily attached to a proton gantry without large beam interference, it may be useful to verify the entrance dose map in the treatment room.

Acknowledgments

This work was supported by the National Nuclear R&D Program through the National Research Foundation of Korea (NRF), funded by the Ministry of Education, Science and Technology (NRF-2015M2A2A7A02045275) and (NRF-2017M2A2A7081416).

Author Contributions

Dongho Shin and Myonggeun Yoon conceived of the key idea and designed the experiments. Jaeman Son performed the experiments and wrote the paper. Se Byeong Lee and Youngkyung Lim irradiated the proton beam under various conditions. Sung Yong Park and Kwanho Cho distributed medical opinion. All authors made contributions to writing and revising the manuscript.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Langer, M.; Leong, J. Optimization of beam weights under dose-volume restrictions. Int. J. Radiat. Oncol. 1987, 13, 1255–1260. [Google Scholar] [CrossRef]
  2. Mock, U.; Georg, D.; Bogner, J.; Auberger, T.; Pötter, R. Treatment planning comparison of conventional, 3D conformal, and intensity-modulated photon (IMRT) and proton therapy for paranasal sinus carcinoma. Int. J. Radiat. Oncol. 2004, 58, 147–154. [Google Scholar] [CrossRef]
  3. Yu, C.X. Intensity-modulated arc therapy with dynamic multileaf collimation: An alternative to tomotherapy. Phys. Med. Biol. 1995, 40, 1435. [Google Scholar] [CrossRef] [PubMed]
  4. Levin, W.; Kooy, H.; Loeffler, J.; DeLaney, T. Proton beam therapy. Br. J. Cancer 2005, 93, 849. [Google Scholar] [CrossRef] [PubMed]
  5. Paganetti, H.; Niemierko, A.; Ancukiewicz, M.; Gerweck, L.E.; Goitein, M.; Loeffler, J.S.; Suit, H.D. Relative biological effectiveness (RBE) values for proton beam therapy. Int. J. Radiat. Oncol. 2002, 53, 407–421. [Google Scholar] [CrossRef]
  6. Pedroni, E.; Bacher, R.; Blattmann, H.; Böhringer, T.; Coray, A.; Lomax, A.; Lin, S.; Munkel, G.; Scheib, S.; Schneider, U. The 200-MeV proton therapy project at the Paul Scherrer Institute: Conceptual design and practical realization. Med. Phys. 1995, 22, 37–53. [Google Scholar] [CrossRef] [PubMed]
  7. Engelsman, M.; Lu, H.M.; Herrup, D.; Bussiere, M.; Kooy, H. Commissioning a passive-scattering proton therapy nozzle for accurate SOBP delivery. Med. Phys. 2009, 36, 2172–2180. [Google Scholar] [CrossRef] [PubMed]
  8. Smith, A.R. Proton therapy. Phys. Med. Biol. 2006, 51, R491. [Google Scholar] [CrossRef] [PubMed]
  9. Gillin, M.T.; Sahoo, N.; Bues, M.; Ciangaru, G.; Sawakuchi, G.; Poenisch, F.; Arjomandy, B.; Martin, C.; Titt, U.; Suzuki, K. Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas MD Anderson Cancer Center, Proton Therapy Center, Houston. Med. Phys. 2010, 37, 154–163. [Google Scholar] [CrossRef] [PubMed]
  10. Kanai, T.; Kawachi, K.; Kumamoto, Y.; Ogawa, H.; Yamada, T.; Matsuzawa, H.; Inada, T. Spot scanning system for proton radiotherapy. Med. Phys. 1980, 7, 365–369. [Google Scholar] [CrossRef] [PubMed]
  11. Lomax, A. Intensity modulation methods for proton radiotherapy. Phys. Med. Biol. 1999, 44, 185. [Google Scholar] [CrossRef] [PubMed]
  12. Goulet, M.; Gingras, L.; Beaulieu, L. Real-time verification of multileaf collimator-driven radiotherapy using a novel optical attenuation-based fluence monitor. Med. Phys. 2011, 38, 1459–1467. [Google Scholar] [CrossRef] [PubMed]
  13. Lin, L.; Qian, J.; Gonzales, R.; Keck, J.; Armour, E.; Wong, J. Su-e-t-483: In vivo dosimetry of conventional and rotational intensity modulated radiotherapy using integral quality monitor (IQM). Med. Phys. 2015, 42, 3445–3446. [Google Scholar] [CrossRef]
  14. Talamonti, C.; Arilli, C.; Marrazzo, L.; Calusi, S.; Galeotti, C.; Wandel, Y.; Meattini, I.; Scoccianti, S.; Pallotta, S. A new approach in patient QA: Beam delivery check and in-vivo dosimetry during radiotherapy treatment. Phys. Medica 2016, 32, 237–238. [Google Scholar] [CrossRef]
  15. Thoelking, J.; Sekar, Y.; Fleckenstein, J.; Lohr, F.; Wenz, F.; Wertz, H. Characterization of a new transmission detector for patient individualized online plan verification and its influence on 6MV X-ray beam characteristics. Z. Med. Phys. 2016, 26, 200–208. [Google Scholar] [CrossRef] [PubMed]
  16. Thoelking, J.; Fleckenstein, J.; Sekar, Y.; Boggula, R.; Lohr, F.; Wenz, F.; Wertz, H. Patient-specific online dose verification based on transmission detector measurements. Radiother. Oncol. 2016, 119, 351–356. [Google Scholar] [CrossRef] [PubMed]
  17. Arjomandy, B.; Sahoo, N.; Ding, X.; Gillin, M. Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance. Med. Phys. 2008, 35, 3889–3894. [Google Scholar] [CrossRef] [PubMed]
  18. Rah, J.-E.; Oh, D.H.; Shin, D.; Kim, D.-H.; Ji, Y.H.; Kim, J.W.; Park, S.Y. Dosimetric evaluation of a glass dosimeter for proton beam measurements. Appl. Radiat. Isot. 2012, 70, 1616–1623. [Google Scholar] [CrossRef] [PubMed]
  19. Reinhardt, S.; Hillbrand, M.; Wilkens, J.; Assmann, W. Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. Med. Phys. 2012, 39, 5257–5262. [Google Scholar] [CrossRef] [PubMed]
  20. Zullo, J.R.; Kudchadker, R.J.; Zhu, X.R.; Sahoo, N.; Gillin, M.T. LiF TLD-100 as a dosimeter in high energy proton beam therapy—Can it yield accurate results? Med. Dosim. 2010, 35, 63–66. [Google Scholar] [CrossRef] [PubMed]
  21. Darafsheh, A.; Taleei, R.; Kassaee, A.; Finlay, J.C. The visible signal responsible for proton therapy dosimetry using bare optical fibers is not Čerenkov radiation. Med. Phys. 2016, 43, 5973–5980. [Google Scholar] [CrossRef] [PubMed]
  22. Jang, K.W.; Yoo, W.J.; Shin, S.H.; Shin, D.; Lee, B. Fiber-optic cerenkov radiation sensor for proton therapy dosimetry. Opt. Express 2012, 20, 13907–13914. [Google Scholar] [CrossRef] [PubMed]
  23. Darafsheh, A.; Taleei, R.; Kassaee, A.; Finlay, J.C. Proton therapy dosimetry using the scintillation of the silica fibers. Opt. Lett. 2017, 42, 847–850. [Google Scholar] [CrossRef] [PubMed]
  24. Darafsheh, A.; Zhang, R.; Kanick, S.C.; Pogue, B.W.; Finlay, J.C. Spectroscopic separation of čerenkov radiation in high-resolution radiation fiber dosimeters. J. Biomed. Opt. 2015, 20, 095001. [Google Scholar] [CrossRef] [PubMed]
  25. Son, J.; Kim, M.; Shin, D.; Hwang, U.; Lee, S.; Lim, Y.; Park, J.; Park, S.; Cho, K.; Kim, D.; et al. Development of a novel proton dosimetry system using an array of fiber-optic cerenkov radiation sensors. Radiother. Oncol. 2015, 117, 501–504. [Google Scholar] [CrossRef] [PubMed]
  26. Low, D.A.; Harms, W.B.; Mutic, S.; Purdy, J.A. A technique for the quantitative evaluation of dose distributions. Med. Phys. 1998, 25, 656–661. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Proton beam delivery techniques: Passive scattering mode (upper panel) and Pencil beam scanning mode (lower panel).
Figure 1. Proton beam delivery techniques: Passive scattering mode (upper panel) and Pencil beam scanning mode (lower panel).
Sensors 18 00227 g001
Figure 2. (a) Configuration of the developed FORS based proton beam dosimetry system; (b) The inside view of the developed proton beam dosimetry system; (c) Schematic image of the side-view of the FORS; (d) Schematic image of cross-section of the optical fiber.
Figure 2. (a) Configuration of the developed FORS based proton beam dosimetry system; (b) The inside view of the developed proton beam dosimetry system; (c) Schematic image of the side-view of the FORS; (d) Schematic image of cross-section of the optical fiber.
Sensors 18 00227 g002
Figure 3. Schematic image of position localization using the FORS.
Figure 3. Schematic image of position localization using the FORS.
Sensors 18 00227 g003
Figure 4. An example of measured 2D dose distributions using various dosimetric tools (Case 3 of the square-shaped field in Table 1). (a) MatriXX; (b) EBT3 film; (c) FORS.
Figure 4. An example of measured 2D dose distributions using various dosimetric tools (Case 3 of the square-shaped field in Table 1). (a) MatriXX; (b) EBT3 film; (c) FORS.
Sensors 18 00227 g004
Figure 5. An example of measured 2D dose distribution using various dosimetric tools for the irregular-shaped field. (Field 4 in Table 2) (a) MatriXX; (b) EBT3 film; (c) FORS.
Figure 5. An example of measured 2D dose distribution using various dosimetric tools for the irregular-shaped field. (Field 4 in Table 2) (a) MatriXX; (b) EBT3 film; (c) FORS.
Sensors 18 00227 g005
Figure 6. Example of dose comparison using the gamma evaluation method for an irregularly-shaped field. (Field 5 in Table 2). (a) MatriXX vs. FOR; (b) EBT3 film vs. FORS. The number denotes the gamma index.
Figure 6. Example of dose comparison using the gamma evaluation method for an irregularly-shaped field. (Field 5 in Table 2). (a) MatriXX vs. FOR; (b) EBT3 film vs. FORS. The number denotes the gamma index.
Sensors 18 00227 g006aSensors 18 00227 g006b
Table 1. Profiles for square-shaped fields, including FWHM and penumbra, using MatriXX, EBT3 film, and FORS.
Table 1. Profiles for square-shaped fields, including FWHM and penumbra, using MatriXX, EBT3 film, and FORS.
Horizontal Profile (x-Axis)Vertical Profile (y-Axis)
Energy Measured Data (mm)%Diff.Measured Data (mm)%Diff.
CaseMatriXXEBT3 FilmFORSMatriXX vs. FORSEBT3 Film vs. FORSMatriXXEBT3 FilmFORSMatriXX vs. FORSEBT3 Film vs. FORS
FWHM (mm)122.6 MeV#144.344.845.11.80.644.344.445.01.71.3
#263.864.064.91.81.465.064.164.21.10.2
#3104.8103.7103.90.90.2104.4104.4106.01.51.5
227.1 MeV#143.644.144.21.40.244.143.944.61.11.6
#266.466.465.31.71.663.463.864.11.20.5
#3103.6104.7104.10.50.5104.0104.0104.40.40.3
Penumbra (mm)122.6 MeV#116.216.016.20.41.316.315.716.01.52.0
#216.916.716.71.50.016.516.016.31.51.9
#316.816.316.51.51.416.916.416.71.42.0
227.1 MeV#19.89.59.61.61.810.09.79.82.01.1
#29.79.49.51.81.59.99.69.80.91.8
#39.79.59.70.31.810.09.89.91.01.5
Table 2. Mean passing rates in gamma indices based on the gamma evaluation method for irregularly-shaped fields using MatriXX, EBT3 film, and FORS (units: %).
Table 2. Mean passing rates in gamma indices based on the gamma evaluation method for irregularly-shaped fields using MatriXX, EBT3 film, and FORS (units: %).
EnergyToolField 1Field 2Field 3Field 4Field 5Mean
122.6 MeVMatriXX vs. FORS98.497.196.497.195.196.8 ± 1.1
EBT3 Film vs. FORS97.495.795.397.697.396.7 ± 1.0
227.1 MeVMatriXX vs. FORS96.299.696.297.595.797.0 ± 1.4
EBT3 Film vs. FORS96.996.591.196.397.795.7 ± 2.3

Share and Cite

MDPI and ACS Style

Son, J.; Lee, S.B.; Lim, Y.; Park, S.Y.; Cho, K.; Yoon, M.; Shin, D. Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam. Sensors 2018, 18, 227. https://doi.org/10.3390/s18010227

AMA Style

Son J, Lee SB, Lim Y, Park SY, Cho K, Yoon M, Shin D. Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam. Sensors. 2018; 18(1):227. https://doi.org/10.3390/s18010227

Chicago/Turabian Style

Son, Jaeman, Se Byeong Lee, Youngkyung Lim, Sung Yong Park, Kwanho Cho, Myonggeun Yoon, and Dongho Shin. 2018. "Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam" Sensors 18, no. 1: 227. https://doi.org/10.3390/s18010227

APA Style

Son, J., Lee, S. B., Lim, Y., Park, S. Y., Cho, K., Yoon, M., & Shin, D. (2018). Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam. Sensors, 18(1), 227. https://doi.org/10.3390/s18010227

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop