1. Introduction
Parkinson’s Disease (PD) is a neurodegenerative disorder caused by a dopaminergic neural loss in the substantia nigra. Dopamine has an inhibitory effect on the excitatory signals that are propagated from the nigrostriatal pathway to the motor cortex. Consequently, the decrease of dopamine transmission between neurons will lead the motor pathway to remain in an excited state. The cardinal symptoms of PD include the bradykinesia, resting tremor, postural instability, and rigidity [
1].
Nowadays, the cure for PD is still unknown and dopamine replacement therapy with Levodopa (L-dopa) remains the major medical approach to control the PD symptoms. However, over the years, the use of L-dopa may lead to highly disabling fluctuations and dyskinesias [
2]. More recently, deep brain stimulation (DBS) surgery has emerged as the second therapy in those patients, leading to the improvement of dyskinesia, as well as of dopamine-responsive motor symptoms [
3]. The DBS surgical procedure is based on the implantation of stimulation electrodes at high frequencies (130 Hz) with a pulse length at 60
in the basal ganglia structures (subthalamic nucleus (STN) or internal globulus pallidum (GPi)), aiming to promote the functional inhibition of the excited motor control pathways [
3]. The stereotactic target of stimulation is first defined based on pre-operative medical imaging. Then, an electrophysiological exploration is performed with a tetrapolar electrode, while varying the stimulation parameters (current intensity, voltage, depth and anatomical position) and testing symptoms and side effects to determine the final electrode configuration [
3].
To assess the improvement of patient response to a therapeutic approach, the severity of the PD symptoms may be evaluated. Rigidity is present in 90–99% of patients [
4]. To assess the rigidity improvement, the clinician usually imposes a passive flexion to a patient’s joint and attributes it a subjective score, translating the severity of the muscles rigidity, ranging from 0% to 80% of improvement [
3]. The wrist rigidity is a trustworthy functional feature since the cogwheel rigidity, which is defined as a PD-typical muscular stiffness throughout the range of wrist passive movement, can be easily perceived [
5]. Generally, a second neurologist performs an intra-operatory blind test to confirm the first evaluation. Nevertheless, this assessment is subjective and depends on the experience and perception of each neurologist. Therefore, it is crucial to develop quantitative methods for assessing rigidity.
Table 1 presents a summary of the literature in this issue.
Patrick et al. initiated the first attempts to evaluate the rigidity of the elbow joint as a measure of mechanical impedance. For this purpose, they developed a custom-made device which includes two pads placed where the physician will force the joint flexion, measuring the resultant force applied on them, and a gyroscope sensor which measured the angular velocity [
6]. Later, Shapiro et al., showed the effectiveness of DBS stimulation by measuring the resistive torque of the elbow joint movement. The proposed methodology allowed the authors to discriminate PD patients from healthy controls [
7]. Following these studies, other biomechanical variables have been proposed for rigidity quantification, including the peak torque [
8], impulse [
11], difference of bias [
9], elastic stiffness [
11], damping constant [
11], and surface electromyography of the biceps and triceps branchii [
9,
10]. Despite the advances made, the proposed systems were bulky and some of them provided values of features that could hinder the intuitive understanding of the physician about the rigidity severity. Further, these methodologies relied on relatively heavy processing which might lead to delays in the output, hampering its immediate feedback to physicians when searching for the best DBS electrode positioning and parameters during surgery. Therefore, they were not considered suitable for intra-operative conditions, such as DBS.
In 2014, Kwon et al. were the first to evaluate the wrist rigidity during intraoperative DBS, measuring several biomechanical properties, such as mechanical impedance, impulse, elastic stiffness and damping constant. The rigidity evaluation device consisted of a potentiometer to measure the joint angle, a load cell and an accelerometer to measure inertia [
11]. Later, Shah et al. developed a method to measure rigidity during DBS using an accelerometer sensor. The authors concluded that a change in rigidity modifies the speed of passive movements [
14]. Van den Noort et al. conducted a proof-of-principle study in which the patient wrist rigidity was assessed using a custom-made system, the PowerGlobe, which included inertial and force sensors [
16]. Similarly, Angeles et al. used inertial motion, muscle activity and force sensors to measure both elbow and wrist rigidity to assess the impact of DBS therapy on the rigidity symptom [
17]. Di Biase et al., on the other hand, measured elbow and wrist rigidity and their changes during DBS procedure, using magnetic-inertial wearable sensors [
18]. More recently, Perera et al. developed a new system to measure the rigidity of the metacarpophalangeal joint, consisting of a miniature motor to flex the third digit of the hand. With the developed device, the authors studied the impact of the DBS and performed comparisons between PD patients and healthy individuals [
20]. Despite these advances in quantifying rigidity reduction during DBS, none of these systems has been able to provide real-time feedback of the patient’s rigidity to the examiner during DBS surgery procedure, supporting the decision-making towards the final placement of the DBS electrode.
Our group has developed a real-time feedback system to evaluate the patient’s wrist rigidity in situ. To our knowledge, there is no other group following this line of research. In [
13], we presented the iHandU system: a novel, comfortable and wireless system, designed to classify the wrist rigidity during the DBS, by deriving features from angular velocity values of wrist flexion and using a polynomial mathematical model to classify signals, stored in a computer. This system was able to provide in real-time the wrist rigidity reduction the DBS procedure to clinicians, hence supporting the physician decision. We showed that the iHandU classified correctly 83.9% of the evaluated signals against the blind agreement of two experts. The main difficulties were found in distinguishing intermediate levels of wrist rigidity. Also, we suggested that the wearable device should be improved and adapted to the operating room environment. In [
15] and [
19], we initiated efforts to improve the iHandU system at three different levels: wrist rigidity classification model, system’s hardware and system’s software. This work gave rise to three new versions along the years: the iHandU 2.0, described in [
15], the iHandU 3.0, described in [
19] and our more recent version, the iHandU 4.0, which was developed to increase system comfort at the patient and clinician levels, as well as to enable faster signal processing and an easier methodology to record surgery-related information.
This publication aims to review all the work done so far, outlining the iHandU system evaluation over these years, with a special focus on the wrist rigidity models developed to enhance the iHandU performance and drawing the lessons learned from each iteration. We also introduce the iHandU 4.0, which is currently being use in all DBS surgeries performed at São João Hospital in Porto, Portugal.
This paper is organized as the follows. In
Section 2 (iHandU 1.0-Proof-of-Concept) we describe the methodology, results and limitations of the first iHandU version. In
Section 3 (iHandU System Evaluation), we describe the following versions of iHandU system in terms of hardware, software and wrist rigidity classification models, while in
Section 4 (Wrist Rigidity Classification Models), we address the classification models developed over time in greater detail. In the last section (Conclusions and Future Directions), we emphasize the conclusions of this research line and what we have learned throughout each iHandU version, as well as what we intend to achieve in the near-future.
2. iHandU 1.0-Proof-of-Concept
The general architecture of the iHandU system includes a Motion Mote (MoMo), consisting in a Bluetooth-enabled Inertial Measurement Unit (IMU) embedded into a specially designed textile band, that is worn on the palm of the patient’s hand. This sensor communicates (via Bluetooth v2.0) with an external device for data analysis and presentation.
Figure 1 shows the existing setup of the motion sensor and angular velocity signals, acquired from different rigidity states. This device was developed a few years ago by our research group and has a gyroscope (± 2000°/s) an accelerometer (2, 4, or 8 g) and a magnetometer (with compass heading accuracy of 1° to 2°) with a battery for up to 8 hours of continuous working. A dedicated microcontroller unit (MCU) collects data from all the sensors at a sampling rate of 42 Hz and then transmits the data via Bluetooth to an external device wirelessly [
13]. In this first proof-of-concept version, the MoMo was transmitting data to an Intel Core i7-4600 CPU @ 2.70 GHz computer, which allowed data visualization and computation of the reduction of wrist rigidity during DBS surgery, through a National Instruments LabView 2014 script [
13]. This data processing was performed using only the data of the y-axis from gyroscope sensor, due to its high level of pronunciation during the wrist passive flexion. Such a configuration was designed with the aim to not interfere in the normal passive wrist flexion movement nor with the surgical procedure.
2.1. Subjects and Experiments
The system was used during the bilateral subthalamic nucleus (STN) DBS of parkinsonian patients, while an examiner was evaluating the wrist rigidity improvement for different stimulation settings. The subjects had their medication withdrawn 12 h before the procedure and were under local anaesthesia during the surgical procedure. A stereotactic target for stimulation and trajectory for the electrode were previously defined, based on medical imaging. Then, electrophysiological exploration was performed during the surgical procedure to determine the optimal electrode placement. The stimulation frequency was set at 130 Hz, while voltage, depth and placement were varied. To train the rigidity classification model, data was first acquired from a total of 6 patients (48 signals); then, 4 patients were considered (156 signals) to test them [
13]. The wrist rigidity improvement classification derived by our system was compared to the evaluation of at least two experts’ agreement. A classification was considered accurate if it was within a range of 5% away from their classification. Patient monitoring was authorized by the local ethics committee and all patients signed an informed consent form.
2.2. Signal Processing
The data acquired from the y-gyroscope sensor (
y) was first converted to angular velocity based on the gyroscope specifications as follows:
Then, a 4-sample moving average low-pass filter was applied for non-movement noise reduction. To keep only samples corresponding to wrist flexion movements, only the negative arcades of
were considered. Note that the MoMo orientation makes the flexion movement to be in the opposite direction of the positive rotation around the
y-axis. For every 300 samples of angular velocity, two angular velocity features were extracted: the average angular velocity,
, and the average peak value,
. Absolute peaks were computed as the highest values between two valleys of the signal, within a margin of 0.2 °s
−1. Although low rigidity and
values have a direct correlation to wrist rigidity states (see
Table 2), this feature, by itself, did not guarantee accurate discrimination of rigid
versus non-rigid, since signals with widely different shapes can present similar peak values. Elongated signal arcades, few peaks in a certain period of time or unexpected plateaus during the flexion movement can by other hand correspond to some residual rigidity even in signals with high amplitude and must be taken in consideration. Such information can be yielded by
, since the average of the signal decreases for non-smooth signals. Therefore, we proposed to combine these two features by deriving a signal descriptor [
13], defined as:
The signal descriptor aimed at discriminating between the stimulation settings that improve the patient’s condition and those who do not. We then clustered the training dataset into the specified classes of rigidity improvement: 10%, 20%, 30%, 40%, 50%, 60%, 70%, and 80%, according to the clinician’s classification of the wrist rigidity improvement. Jarque–Bera tests [
21] confirmed the data normality profile and descriptive power was assessed using two-tailed t-tests [
13].
2.3. Classification Model
The signal descriptor defined by the Equation (1) was used to build the wrist rigidity model, where the rigidity during passive wrist flexion was labeled according to the expert’s assessment (medical label) following a discrete decimal scale ranging from 0% to 80% of rigidity improvement. Then, we computed the average values of the signal descriptors belonging to the same medical label classes. Finally, we determined the polynomial mathematical model that best approximates the modeled relationship between the percentage of rigidity improvement (rigidity label) and the signal descriptor. Approximations of higher degrees were discarded since they can lead to overfitting and can be less responsive to widely different incoming signals. Standard machine learning techniques were also excluded due to their heavy computational cost, which would limit the implementation of the classification model with local signal processing in “wearable-embedded” signal processing hardware. We used a custom-made program running on MathWorks
® MATLAB to analyze the training set. The training error was estimated using the leave-one-out cross-validation with 5000 iterations. The classification model was tested in four patients, during DBS surgery [
13].
2.4. First Model Usability Results
We designed a system to quantitatively assess the wrist rigidity during DBS procedure, capable to send the result to clinicians in real-time, helping in the interactive determination of the optimal stimulation setting.
From the statistical analysis’ results summarized in
Table 2, we concluded that
had more discriminative power between rigid and non-rigid states, since their p-values are lower than the p-values of its counterparts. The polynomial model derived for rigidity classification is presented in
Figure 2 and had a training error of
. This error range was considered acceptable, since we modeled a discrete scale using a continuous function. The classification model allowed a correct classification of 131 out of 156 classifications of rigidity, corresponding to an accuracy of 83.9% [
13].
The main difficulties were found when distinguishing intermediate levels of wrist rigidity, whose correlation with the classification model were lower. This could be related to the nonlinear dependence of the rigidity with the angular velocity amplitude. Therefore, we hypothesized that this model could be improved by considering other angular velocity features, in addition to the existing ones, such as the cogwheel rigidity of wrist and the patient’s rigidity baseline, i.e., the Unified Parkinson’s Disease Rating Scale (UPDRS) scores before starting stimulation.
The premise of this analysis is that the assessment is being performed correctly, ensuring three main conditions: (1) It is expected that the imposing flexion should be the largest amplitude that occurs on the y direction, avoiding rotational movements; (2) The force applied by the clinician does not vary greatly over the course of DBS surgery and between surgeries as well; (3) Patients can not actively try to counter or help in bending the wrist, in order to avoid undesired rigidity profiles. The last requirement can be partially fulfilled in our system, since we can label these signals as “invalid” after indication of the clinicians. The first and second requirements, on the other hand, cannot be secured in our system, which presents another limitation. Therefore, we also concluded that the angular velocity acquired from other gyroscope axes (besides the y axis) should be explored, as well as the possibility of integrating force sensors into our system. Another limitation of the system arises due the angular velocity features of the training set which were extracted from every 300 samples, while the system classified every train of 200 samples (~4 s). This might have influenced the classification as it might lead to a low average angular velocity. Therefore, we suggested to decrease the number of samples to extract the angular velocity features.
In addition to the limitations on the classification model, the first version also had limitations on the hardware level. The iHandU 1.0 required connection to a computer for signal processing and data visualization. Consequently, to provide the wrist rigidity improvement to clinicians at each stimulation, i.e., in real-time, it was necessary to integrate the computer into the operating room environment. This was a disadvantage due to the space that a computer occupies. Therefore, we suggested that a smaller external device (but with equivalent processing power) should be integrated to fulfill that role, improving usability in intra-operatory conditions. Also, we agreed that the wearable device could be improved in term of patient’s comfort and clinicians’ usability, since this version was large, thicker and difficult to fix to the patient’s hand.
In this section we have presented a first proof-of-concept of the iHandU system, reported in [
13], which showed good results. After this first iteration, and bearing in mind the limitations presented, the iHandU system has been improved at different levels: (1) design, implementation and test of four different new wrist rigidity classification models (
Section 3.1 and
Section 4); (2) wearable device, by adapting it for operating room environment (
Section 3.2); (3) smartphone application, user-friendly, capable to store, process and show real-time results during intra-operatory assessment (
Section 3.3). This was an iterative process where several versions were developed and validated in real DBS surgeries to improve the iHandU usability and functionality in the operating room. The following sections present and discuss these enhancements, showing their relevance and impact, such the methodologies used and validation results. It is of major importance to mention that after this technology proof-of-concept, a patent was filled covering the use of these type of devices to assess rigidity in DBS surgeries, which is currently covered in Europe Union, United States, People’s Republic of China, Canada, Japan and Republic of Korea, with the number WO2016166702A1.
4. Wrist Rigidity Classification Models
So far, four iHandU system versions have been developed. In all versions, a rigidity classification model has been developed to assess the wrist rigidity reduction during the DBS procedure, which all of them shared a similar framework of rigidity classification (
Figure 6). This framework comprises the following steps: (1) signal is acquired during the DBS surgery for a group of patients, i.e., the training set; (2) features are extracted from the signal, reflecting different rigidity levels; (3) all features are combined in one function, designated as signal descriptor; (4) the average values of the signal descriptors belonging to the same medical label classes of rigidity improvement, ranging from 0 to 80%, are computed; (5) the second-order polynomial function that best model the relationship between the percentage of rigidity improvement (Rigidity Label) and the signal descriptor is estimated; (6) the previous function is used to classify the rigidity improvement of a second group of patients during the DBS surgery, i.e., the test set. A classification is considered accurate if it is within a range of 5% away from the clinician’s classification.
In the first model, the angular velocity during the passive wrist flexion was recorded using the
y-axis of the gyroscope sensor, from which two features were computed for each 300 samples: the average angular velocity value and the angular velocity peak. The number of samples was defined so as to ensure the presence of a minimum of two to three arches in signal processing. Then, these features were combined using Equation (1). As a result, the polynomial function present in
Figure 2 was estimated. This classification model labeled correctly 83.9% of the evaluated signals against the blind agreement of two specialists. The main limitations were found when classifying intermediate levels of wrist rigidity. Our research group has been trying to overcome some limitations of the first model since 2015. Four models were developed so far: The (A) multi-model (MM), the (B) cogwheel rigidity model (CRM), the (C) three-gyroscope-axis model (3GAM) and the (D) reduced model (RM).
(A) Multi-model: consists of two rigidity models, specific for each baseline wrist rigidity: low or high rigidity states, measured before stimulation. The most appropriate model can then be selected at start-off of the surgery, according to the physician’s initial assessment [
15].
(B) Cogwheel rigidity model: The cogwheel rigidity was included in the signal descriptor as a third angular velocity feature [
19]. To detect cogwheel rigidity from the angular velocity signal, we used a methodology developed by Costa et al., [
13].
(C) Three-gyroscope-axis model: The angular velocity of the three Cartesian axes was considered instead of only the
y-axis [
19] and, finally.
(D) Reduced Model: the number of samples to extract the angular velocity features (angular velocity mean value and angular velocity mean peak) was decreased from 300 to 200 [
19].
Since the models were not developed at the same time, different datasets, as well as different iHandU versions were used.
Table 4 summarizes the dataset characterization and
Table 5 correlates the type of data, features extracted, system version and dataset considered to each wrist rigidity model. The experiment considered in all models during the surgical procedure was similar to that described above.
4.1. Signal Processing
In all models, data was acquired and processed in a smartphone. Only the 3-gyroscope-axis model considered data from the 3-axis of the gyroscope sensor. In all other models, only the data of the
y-axis of the gyroscope sensor was considered. Signal processing was similar to the performed for the first model [
13]: First, the raw data was converted to angular velocity, according to the previously presented Equation (2). Then, only its negative values were considered, restricting the analysis to the flexion wrist movement. In the Multi-model, features were extracted for every 300 samples, while in the remaining ones, the features were extracted for every 200 samples. For Multi-model, 3-gyroscope-axis model and Reduced model, two features were considered: average angular velocity and angular velocity peak value. In the 3-GAM, these features were extracted in all Cartesian axes. In the Cogwheel rigidity model, on the other hand, a third feature was considered: the cogwheel artefact number,
. The angular velocity features were combined by several signal descriptors, aiming to describe the decrease in rigidity.
Table 6 summarizes the signal descriptors considered in each wrist rigidity model.
4.2. Classification Model
Each signal descriptors present in
Table 6 were used to build a wrist rigidity model where the rigidity during the passive wrist flexion was labeled according to the medical expert’s assessment agreement, following a discrete decimal scale, ranging between 0 and 80% of rigidity improvement. We then computed the average values of the signal descriptors belonging to the same medical label classes. Contrary to what was done in the first model, now we only considered the classes 0%, 40%, 50%, 60%, 70%, and 80% of improvement, since the rigidity improvement between 0% and 40% was considered more difficult to discriminate by the clinicians. Finally, the polynomial models that best approximate the relationship between the percentage of rigidity improvement and the signal descriptor were derived. We used a custom-made program running on MathWorks
® MATLAB to analyze the training set of each rigidity model. In the Multi-model, the training set was separated in two different clusters: low baseline rigidity (98 signals), which corresponds to UPDRS scores of 1 or 2, and high baseline rigidity (139 signals), in which the patients had an UPDRS of 3. Patients with UPDRS 4 were not found for this work. Each of these subsets were used to obtain the classification model. In all rigidity models, the training errors were estimated using the Leave-one-out cross-validation method with 5000 iterations. The signal descriptors with the lowest training error of each model were tested during DBS surgery.
4.3. Results and Discussion
The signal descriptors
and
(low and high baseline rigidity models, respectively),
,
and
which presented the larger accuracy and the lowest test error (maximum: 3.6 ± 4.0%), leading to the best system performance, as presented in
Figure 7, including the accuracy, training and test errors obtained. The remaining signal descriptors of
Table 6 were discarded since they exhibited higher training errors.
From the analysis of this figure, we can conclude a high fitting of the mean values of the signal descriptors in each step of improvement, for all wrist rigidity models. The worse fitting was the one of the Multi-Model of patients initially less rigid (
Figure 7a). This could be related to a higher difficulty by the examiner to identify small differences when the rigidity baseline is low. For all rigidity models, approximately the same accuracy was obtained
—except for the 3-gyroscope-axis that decreased the system performance
supports the reliability of this approach to sustain the examiner’s assessment in intra-operatory conditions, by providing real-time feedback. It also is the same result as the first model, suggesting, since these models were trained and tested for a larger number of patients, that a larger dataset does not compromise the system performance. At the same time, the training error was smaller when we considered a larger dataset.
The classification error, i.e., the test error, was similar for all models, supporting the idea that the polynomial regression is a good representative model, even when we are comparing a continuous scale to a decimal one. On the other hand, the classification error did not decrease considering the Reduced model compared to the others “single-axis” models, as we expected. However, the decrease of the number of samples to extract the angular velocity features does not compromise system performance. This is a valuable result, since it suggests less time for each intraoperative evaluation, reducing the overall procedure duration. This may bring benefits to the patients, who are having awake brain surgery, decreasing the global risk of surgical infection.
Further, contrary to what was expected, a multi-model approach did not improve the system performance. This proves the limitation of kinematic measures in further distinguishing different wrist rigidity levels since both models lead to similar results, even though differences were evident. Hence, we expected that the cogwheel rigidity, which is considered one major feature in PD rigidity, that highly distinguishes UPDRS sub-scores of 2 and 3, could increase the system performance. However, our results suggest that the cogwheel rigidity does not increase the system’s performance when compared with the first model. In fact, with the methodology used, it is possible to detect cogwheel rigidity even for 80% of rigidity improvement and similar average of these artefacts can be found for intermediate levels of rigidity (see
Table 7). This suggests that this angular velocity feature may not play a role in intermediate improvements.
Finally, from the angular velocity of x and z-axis we may conclude the z-axis does not vary with the expert’s classification agreement. The x-angular velocity, on the other hand, varied with the wrist rigidity reduction but this gain did not cause any improvement to our system. Therefore, the decrease in system performance may be related to the x-axis associated noise.
5. Conclusions and Future Directions
This work reported the iHandU system evolution, from its first proof-of-concept until the most recent version that is being used nowadays. Also, we described the main lessons learned in each iteration, enabling the system improvement and adding relevant scientific knowledge related with wrist rigidity assessment during DBS surgeries, a technique that is being used worldwide. The collected feedback during the system evolution made possible to achieve a more robust, user-friendly and customized system, but also new features and improvements that will be made in the next iteration named iHandU 5.0, furthered explained.
The iHandU is proving its capability to support clinicians showing very good results, as shown before. This system evolution results from a high level of interaction between medicine and engineering area resulting in a persuasive system focused on personalized health with direct impact on healthcare procedures. The evolution of iHandU 1.0 to iHandU 2.0 with the introduction of a Smartphone was crucial to improve the system reliability enabling its adaptation to the clinical environment without any interference with the normal surgical procedure. Another adaptation that has recently improved the clinical assessment was the new textile band (textile band v2) that is not only contributing for a higher comfort of the patient, but mainly in the wrist rigidity assessment method allowing clinicians to easily hold the patients hand without feeling the interference of a wearable device. This contributed for a more standard evaluation assessment and for iHandU it represents a huge evolution in the system adaptation to the clinical environment requirements. Regarding the wrist rigidity classification, it was of the utmost importance to understand and explore if it was possible to obtain better results using more parameters. As stated in previous publications [
13,
15,
19] and in this work, although an effort was made with the development and test of new models it was still not possible to achieve a significant improvement due to the high accuracy already presented by the first model. This fact allows us to conclude that our results showed that y-gyroscope-axis remains the best way to classify the wrist rigidity reduction and also the fact that in all iterations, the y-gyroscope axis showed an accuracy of ~80%, proving the robustness of our model. On the other hand, although the y-Reduced Model (only 200 samples for wrist rigidity improvement classification) does not improve the model overall accuracy, it can be considered as our best result, since it optimizes the signal processing methodology.
In this work we introduced the last version, the iHandU 4.0, currently used in DBS surgeries at São João Hospital in Portugal. This version includes: (1) the textile band v2; (2) the MoMo v2 with a smaller design (i.e., a smaller formfactor); (3) the y-reduced model and (4) the iHandU 4.0 Andorid App, in which a voice recording button was added to record all information related with the DBS procedure.
The authors have in mind that other biomechanical properties could eventually be explored to improve our classification system, derived from the resistive torque, as in [
7,
8,
9,
11,
12]. However, as was described previously, most setups designed for rigidity quantification are bulky and complex, hence, inappropriate to take into the operating room. Additionally, some of these are not easily available and clinically not applicable [
11]. Others still depend on the velocity of the imposed movement [
14]. Intermediate improvements of wrist rigidity are often related to false improvements: a decrease in wrist rigidity means less resistive force to imposed movement and, if an expert applies about the same force for each assessment, then higher values of angular velocity will be achieved; nonetheless, if the expert, when feeling less resistance, imposes less force, then no significant change in velocity will occur, thus, no improvement or less than expected is observed. Therefore, the addition, to our system, of force sensors capable of acquiring the force that the examiner imposes in every trial, may lead to the correction of false improvements, therefore enhancing the system’s performance. Adding these sensors, the authors believe that the development of a new model can enable to perceive whether the patient actively participates in the passive flexion, leading to the false detection of improvements.
To explore these new possibilities and understand how to improve even more the iHandU system in the main three areas (hardware, software and wrist rigidity classification), our research group is already working in the iHandU 5.0 version that will integrate the force sensors integrated in an improved textile band and a new wrist rigidity classification model containing information about the clinician force. Also, this version will contemplate a new redesigned hardware device with Bluetooth Low Energy to reduce power consumption and with a much small form-factor. These new hardware developments are showing that we are able to reduce the device consumption around four times and reduce the size by almost five times when comparing with the existing one. The synergy between the textile band and the hardware is also under development and in a near future the wearable device will contemplate conductive yarns and new sensing features to contribute for an improvement of the rigidity model.
Prospectively, we also intend to design an “intelligent” system, able to refit the polynomial function in each wrist rigidity assessment, since the system classifies every train of 200 samples (4 s), the signal descriptor and the medical label can be added to the training set. It is also important to perform a system validation with a larger number of patients across different clinical centers, to support its use as a standard tool to assist DBS surgery. For this purpose, some connections with international centers is being made to perform a designed study protocol, in which we propose the system to be first used by a top-level experience neurologist, to find the best polynomial model for signal classification, and therefore, be used, in the future, as a learning tool for less experienced DBS specializing neurologists. Also, the iHandU’s usability may be extended to diagnostic and follow-up of PD patients, not only focused on rigidity evaluation, but also other cardinal symptoms, such as tremor and bradykinesia, which are already under development.
Finally, we foresee the use of iHandU as a supporting tool to evaluate the effectiveness of therapeutic drugs during pharmacological clinical trials procedure to enable the quantification of drug effects, as well as in the clinical response to acute Levodopa challenge test (LDCT), to predict the efficacy of levodopa in the long-term treatment of PD.
Until now, the iHandU system has been used to classify the wrist rigidity improvements of more than 100 patients, corresponding to more than 1600 classifications (107 minutes of data acquisition). A database of wrist rigidity evaluation is being built to be shared as the first database (to the best of our knowledge) of patient inertial data during DBS surgery with a medical label.