1. Introduction
The detection and the measurement of gas concentrations can be carried out using several principles. The most common working principles, sensors, and systems used for gas detection are: electrochemical sensors [
1,
2], systems based on Quartz Crystal Microbalances [
3], sensors based on semiconductors [
4,
5], optical sensors [
6], capacitive sensors [
7], cantilever-based sensors [
8], and colorimetric devices [
9].
Capacitive sensors used as proximity detectors for tracking and control [
10] could be a valid option for gas detection, due to their simple structure, which allows miniaturization, their high sensitivity and low cost. In environmental monitoring, indeed, a sensor’s size, simplicity, and cost are strategic to allow a distributed network of sensing elements in order to monitor and map complex scenarios [
11,
12,
13].
This networking inter-operability asks for simplicity, which gains reliability for standalone functioning; and low cost and size, which allow wide network building.
Starting from the capacity definition:
where is the permittivity in vacuum, is the relativity permittivity, A the electrode area and d the distance between the electrodes (dielectric thickness), it could be possible to detect target molecules by inducing a variation of the dielectric layer in terms of modification of εr or of its thickness. The latter principle is the most interesting for gas detection.
When the sensor system shows meaningful figures (repeatability, sensitivity, and resolution) which are relevant for the specific application, it can be used in fields such as the monitoring of air quality, or in medicine, for example for the analysis of the exhaled breath. This is the application tested in this work, where a commercial sensor platform developed by Microchip (GestIC), normally used to detect, trace, and classify hand movements in space [
14], has been used as an array of chemical sensors.
It is ambitious to propose a gas sensor which is completely novel (especially for the detection of oxygen and carbon dioxide) with sensitivity, resolution, selectivity, and repeatability better than the ones given by the existing technologies (reported in References [
1,
2,
3,
4,
5,
6,
7,
8,
9,
10]). All the mentioned sensors (not pretending to be exhaustive) are complementary in terms of pros and cons, and the best solution can be selected among them for each different application. Thus, the novelty could come from a novel sensing mechanism or from a sensor normally used to detect a different physical quantity, by transforming it into a gas sensor. This is the approach of this study, which finds its motivation in the ambition of exploiting a sensor system that can easily be used in an IoT (Internet of Things) scenario.
2. Materials and Methods
In this experimental set-up, a capacitive sensor, based on a Microchip Gest-IC platform, has been employed as a gas sensor array in order to measure oxygen and carbon dioxide in different concentrations. The device is based on 3D proximity sensing technology [
14], which uses an electric field to detect, trace, and classify hand movements in space. Two overlapped layers comprise the MGC-3030 electrodes system. The first Tx layer shields the ground to allow a propagation of the E-field and grant a better acquisition of the signal. On top of the transmission layer is stacked the Rx layer, which is composed of five electrodes providing spatial resolution. An insulation layer separates the Tx and Rx layers. The five Rx electrodes are placed according to the configuration reported in the
Figure 1.
The responses of the individual electrodes were obtained by interfacing the MGC3030 (I2C serial communication) with an acquisition board based on an ATMEGA 328 microprocessor. A GUI was developed to handle the responses from the five electrodes placed inside the measuring chamber. The interface was developed using MATLAB.
The measuring chamber was designed using Solidworks and it was fabricated using an Ultimaker3 3D printer. It has dimensions of 120 × 85 mm and it is made of acrylonitrile butadiene styrene (ABS).
Anthocyanins extracted from red roses [
15] were used as a sensing material in order to functionalize the sensor surface. The composition of red rose anthocyanin is widely reported in literature. The most significant amount of red rose anthocyanins was characterized as cyanidin 3,5-di-O-glucoside and pelargonidin 3,5-di-O-glucoside using reversed-phase C18 column chromatography. In particular, the cyanidin 3,5-di-O-glucoside is the predominant constituent, representing about 85% of total content [
16].
The results shown in a previous research paper [
15], agreed with already-published results.
To obtain a uniform and reproducible bioactive layer, an adhesive mask was placed above the surface of the Rx electrode and it was used to support manual casting by restricting the functionalization only to the exposed area. A volume of 1 mL of the anthocyanin mixture was poured onto the masked surface and it was allowed to air dry at room temperature. Finally, when the solvent was completely evaporated, the adhesive mask was removed.
The sensor was calibrated using different concentrations of oxygen and carbon dioxide, ranging from 5% to 25%. A flow-meters system was used to control the flow speed and the composition of the gas mixture inside the measuring chamber. The different concentrations were obtained by mixing the gases with inert gas (nitrogen). Nitrogen was also used for sensor recovery after each measurement. The measurement setup is reported in
Figure 2.
In a second phase, the sensor was calibrated to different O
2 and CO
2 concentrations with different humidity levels, from 50% to 90% relative humidity (RH). Each measurement lasted 2 minutes and each recovery phase lasted 5 minutes. Five measurements were executed for each concentration. The output consisted of a voltage signal, which is the result of the electric field perturbation given by the interaction of the sensing material with the gas molecules (see
Figure 3). This perturbation is sensed by the electrodes and transduced in a capacitance shift, which is readable by the voltage output of the electronic circuit interface.
3. Results
In the initial calibration phase, the sensor was not functionalized. The aim of this initial phase was to explore the sensitivity of the system to the target gases without the presence of a chemically active layer. In these conditions, the system did not show the ability to measure different gas concentrations.
In the second calibration phase, the central Rx electrode of the sensor was functionalized using anthocyanins (see the Methods section).
Additionally, in these conditions, the system did not show the ability to measure different gas concentrations.
Thus, in the third calibration phase, two different mixtures were used:
(1) oxygen, nitrogen, and moisture, ranging from 50% to 90% RH.
(2) carbon dioxide, nitrogen, and moisture, ranging from 50% to 90% RH.
The raw data relative to the sensor responses are shown in
Figure 4.
In these conditions, the sensor showed different behaviors for CO
2 and O
2 at different moisture concentrations, as displayed by the response curves in
Figure 5 and
Figure 6. The sensor exhibits a better repeatability with CO
2 concentrations.
Curve fitting models for CO
2 and O
2 are reported below in
Tables S1 and S2 (in the Supplementary Materials). The p1 coefficient always exhibits a far less relevant contribution than p2 and p3, thus all models show a quasi-linear behavior in the concentration range of interest.
To more deeply investigate the sensor response characteristics, sensitivity and resolution were calculated from the curve fitting models at different working points (
Table 1 and
Table 2).
Sensor performances for CO2 and for O2 monitoring are summarized below:
For CO2, resolution ranges from about 0.8% at 50% RH to about 0.006% at 90% RH
For O2, resolution ranges from about 0.2% at 50% RH to about 0.05% at 75% RH
According to the performances listed above, the sensor could be exploited in different fields such as: environmental applications, related to the monitoring of air-quality [
17]; medical applications, related to the study of the exhaled breath composition and to the monitoring of exhaled breath sampling procedures [
18]; and agri-food applications, related to the monitoring of process quality, product packaging, and the assessment of product safety [
19]. Moreover, a proper functionalization of the sensor, using an application-specific chemically active layer, will lead to the maximization of the sensing performances, in terms of sensitivity, resolution, and response speed.
4. Discussion
In this paper, a capacitive proximity sensor (which exploits an electric field’s variations) was used as a chemical sensor. The system was calibrated to oxygen and carbon dioxide, functionalizing the central sensing electrode with a biological sensing material. As shown by the response curves, the sensing system was able to measure both carbon dioxide and oxygen concentrations.
The sensor showed different behaviors to target gases when different moisture concentrations were used. In particular, sensor repeatability was reduced when measuring oxygen concentrations with high moisture levels. To understand this, it is relevant to discuss the role of RH in the sensing mechanism. Potentially, water molecules might act at least at two levels: the first through a direct interaction with gas compounds, and the second by dealing with the anthocyanin sensing layer. In the former case, it is extensively reported by literature that CO
2 can react with water, forming carbonic acid (H
2CO
3), which in turn can release up to two protons (H
+) per molecule, thus contributing to the generation of the signal. However, the sole RH is not enough to induce a significant electric field modification, as demonstrated by the absence of a signal when the non-functionalized Rx electrode is challenged with both carbon dioxide and oxygen gases. The fundamental role exploited by the anthocyanin sensing layer is indeed to enhance such electric field interference, through one of the many chemical reactivity groups typical of the flavonoid classes. The latter level of interaction is between water molecules and the pyrylium ring (C-ring) of the flavylium ion, which is able to alter the chemical structure of anthocyanins by nucleophilic addition at C2 [
20].
On the basis of these results, it will be important to functionalize the system with more sensitive materials to improve sensor performances and allow the measurement of other gases.
Given the ability of the sensor to measure oxygen and carbon dioxide concentrations in the presence of moisture, an interesting future development could also be the application of this device for the analysis of the human exhaled breath, with the aim of including this device in the measuring chain used for breathprinting [
21,
22].
This is a pilot study focused on understanding sensors’ relevance in the detection of the selected gases for different applications. The core of the study was the novelty of the sensors (in gas detection), which shows key features for the current sensor technology: low-cost, low-power, and connectivity. No specific applications have been selected here, even if exhaled breath analysis and environmental monitoring have been cited as most attractive. A broad concentration range of CO2 has been tested in order to investigate possible applications, also including food production and preservation, incubator atmospheres, CO2 entrapment technologies, and monitoring of hazardous sites (mines, etc.). Of course, for a specific application, sensor performances could be improved and tailored in terms of resolution, but the goal of this work was feasibility.
The optimization of the system will allow it to be used in different fields: environmental, medical, and food.
Author Contributions
Conceptualization, A.Z., M.S., and G.P.; methodology, A.Z. and M.S.; software, A.S., and L.V.; validation, A.Z., M.S., and G.P.; formal analysis, M.S., G.P., and L.V.; investigation, A.Z., A.S., C.D.M., and S.G.; resources, M.S. and G.P.; data curation, A.S., C.D.M., and S.G.; writing—original draft preparation, A.Z. and A.S.; writing—review and editing, M.S. and G.P. All authors have read and agreed to the published version of the manuscript.
Funding
This research received no external funding.
Conflicts of Interest
The authors declare no conflict of interest.
References
- Zompanti, A.; Santonico, M.; Vollero, L.; Grasso, S.; Sabatini, A.; Mereu, F.; D’Amico, A.; Pennazza, G. A Gas Sensor with BLE connectivity for Wearable Applications. Proceedings 2018, 2, 765. [Google Scholar] [CrossRef] [Green Version]
- Khan, A.H.; Rao, M.V.; Li, Q. Recent Advances in Electrochemical Sensors for Detecting Toxic Gases: NO2, SO2 and H2S. Sensors 2019, 19, 905. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Wu, Y.-T.; Akoto-Ampaw, P.-J.; Elbaccouch, M.; Hurrey, M.L.; Wallen, S.L.; Grant, C.S. Quartz Crystal Microbalance (QCM) in High-Pressure Carbon Dioxide (CO2): Experimental Aspects of QCM Theory and CO2 Adsorption. Langmuir 2004, 20, 3665–3673. [Google Scholar] [CrossRef] [PubMed]
- Lin, T.; Lv, X.; Hu, Z.; Xu, A.; Feng, C. Semiconductor Metal Oxides as Chemoresistive Sensors for Detecting Volatile Organic Compounds. Sensors 2019, 19, 233. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Zhang, S.; Zhao, Y.; Du, X.; Chu, Y.; Zhang, S.; Huang, J. Gas Sensors Based on Nano/Microstructured Organic Field-Effect Transistors. Small 2019, 15, 1805196. [Google Scholar] [CrossRef]
- Gas Sensing Solution. Available online: https://www.gassensing.co.uk/product/cozir-co2-sensor/, (accessed on 16 July 2019).
- Lang, H.P.; Ramseyer, J.P.; Grange, W.; Braun, T.; Schmid, D.; Hunziker, P.; Jung, C.; Hegner, M.; Gerber, C. An Artificial Nose Based on Microcantilever Array Sensors. J. Physics: Conf. Ser. 2007, 61, 663–667. [Google Scholar] [CrossRef] [Green Version]
- Yoon, I.; Eom, G.; Lee, S.; Kim, B.K.; Kim, S.K.; Lee, H.J. A Capacitive Micromachined Ultrasonic Transducer-Based Resonant Sensor Array for Portable Volatile Organic Compound Detection with Wireless Systems. Sensors 2019, 19, 1401. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Saliu, F.; Della Pergola, R. Carbon dioxide colorimetric indicators for food packaging application: Applicability of anthocyanin and poly-lysine mixtures. Sens. Actuators B Chem. 2018, 258, 1117–1124. [Google Scholar] [CrossRef]
- Braun, A.; Wichert, R.; Kuijper, A.; Fellner, D.W. Capacitive proximity sensing in smart environments. J. Ambient. Intell. Smart Environ. 2015, 7, 483–510. [Google Scholar] [CrossRef] [Green Version]
- Haghi, M.; Stoll, R.; Thurow, K. A Low-Cost, Standalone, and Multi-Tasking Watch for Personalized Environmental Monitoring. IEEE Trans. Biomed. Circuits Syst. 2018, 12, 1144–1154. [Google Scholar] [CrossRef] [PubMed]
- Yi, W.Y.; Lo, K.M.; Mak, T.; Leung, K.S.; Leung, Y.; Meng, M.L. A Survey of Wireless Sensor Network Based Air Pollution Monitoring Systems. Sensors 2015, 15, 31392–31427. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Ye, Z.; Yang, J.; Zhong, N.; Tu, X.; Jia, J.; Wang, J. Tackle environmental challenges in pollution controls using artificial intelligence: A review. Sci. Total Environ. 2019, 699, 134279. [Google Scholar] [CrossRef]
- Microchip. Available online: http://ww1.microchip.com/downloads/en/devicedoc/40001716c.pdf (accessed on 16 July 2019).
- Santonico, M.; Pennazza, G.; Grasso, S.; D’Amico, A.; Bizzarri, M. Design and Test of a Biosensor-Based Multisensorial System: A Proof of Concept Study. Sensors 2013, 13, 16625–16640. [Google Scholar] [CrossRef] [PubMed]
- Lee, J.H.; Lee, H.-J.; Choung, M.-G. Anthocyanin compositions and biological activities from the red petals of Korean edible rose (Rosa hybrida cv. Noblered). Food Chem. 2011, 129, 272–278. [Google Scholar] [CrossRef] [PubMed]
- Jones, A. Indoor air quality and health. Atmospheric Environ. 1999, 33, 4535–4564. [Google Scholar] [CrossRef]
- Popov, T.A. Human exhaled breath analysis. Ann. Allergy Asthma Immunol. 2011, 106, 451–456. [Google Scholar] [CrossRef] [PubMed]
- Knoche, W. Chemical Reactions of CO2 in Water. In Proceedings of the Proceedings in Life Sciences; Springer Science and Business Media LLC, Regensburg, Germany, 17–20 April 1979; pp. 3–11. [Google Scholar]
- Neethirajan, S.; Jayas, D.S.; Sadistap, S. Carbon dioxide (CO2) sensors for the agri-food industry—a review. Food Bioproc. Tech. 2009, 2, 115–121. [Google Scholar] [CrossRef]
- Pennazza, G.; Santonico, M.; Zompanti, A.; Grasso, S.; D’Amico, A. Electronic Interface for a Gas Sensor System Based on 32 MHz QCMs: Design and Calibration. IEEE Sens. J. 2017, 18, 1419–1426. [Google Scholar] [CrossRef]
- De Vincentis, A.; Pennazza, G.; Santonico, M.; Vespasiani-Gentilucci, U.; Galati, G.; Gallo, P.; Picardi, A. Breath-print analysis by e-nose may refine risk stratification for adverse outcomes in cirrhotic patients. Liver Int. 2017, 37, 242–250. [Google Scholar] [CrossRef] [PubMed]
© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).