4.1. Electrostatic Particle Separation
The PMC-µFC device in
Figure 2 can be operated as a DMA for which according to Equation (3), the number of captured particles increases with the sampling voltage and via its size with electrical particle mobility. For confirmation, we performed FEM using COMSOL Multiphysics showing that at sampling/separation voltages of −4 V, −26 V and −140 V positive-charged NPs of diameters of 5 nm, 50 nm and 2.5 µm, respectively, selectively precipitate on the cantilever with maximum efficiencies of nearly 60% to 80% [
26]. With its ability to detect sampled particle size bins on a cantilever balance, we can consider this device as a MEMS-DMPS with the resonance-frequency shift rate of the PMC as output signal.
The sampling efficiency
ξ(
Ves,
dp), which is necessary for retrieving the size-distribution of mass concentration of an aerosol from a measured voltage dependence of frequency-shift rate can be determined if the distribution of the number of electrical charges per particle
n is known [
31]. Conventionally, for this purpose aerosol samples with Boltzmann and unipolar charge distributions are prepared. A radioactive source as diffusion neutralizer or a unipolar corona charger are used [
32] which, however, are accompanied by a possible excessive mobility overlap between different size bins due to multiple charging in case of unipolar diffusion charging [
31].
Instead, in this study, we dispense with an additional charging unit and sample only the naturally positive-charged aerosol fraction. As a quantitative measure of the charging state of a naturally charged aerosol we adopt the size and charge distributions measured in the range of
dp = 20 nm to 400 nm with a NaCl aerosol [
33], which we consider as a model system for aerosol charge distributions generated by a nebulizer. We assume that this non-equilibrium charge distribution can describe not only NaCl but also other nebulized aerosols with sufficient accuracy even if, in general, the charge distribution may be dependent on the specific aerosol nebulizer configuration and the chemical composition of the nebulized solution. In the following, we describe a method based on
ξ(
Ves,
dp) determined accordingly (cf.
Appendix A and
Appendix B), which we propose for retrieving the size-distribution of nebulized aerosols from measurements with the MEMS-DPMS.
Particle sampling on a cantilever resonator yields a resonance-frequency shift rate of [
13,
25]:
with the particle mass concentration
cm, the air flow rate
Q =
Vs/
ts, the sample volume
Vs, the sampling time
ts, the resonance frequency
f0 and the resonator mass
m0. The frequency shift rate depends via the separation/sampling efficiency
ξ(
Ves,
dp) on the applied voltage
Ves between the cantilever and the µFC wall as well as on the particle’s diameter
dp (cf. Equation (3)). Here we consider ranges of particle size and sampling voltage separated in
u size bins, i.e.,
l = 1, …,
u and voltage steps, i.e.,
k = 1, …,
v with
v ≥
u, respectively, define
as the efficiency for capturing particles of the
l-th size bin at the
k-th sampling voltage and replace Equation (5) by a matrix equation:
The vector of mass concentrations (cm,l) (represented by the u size bins of l = 1, …, u) is multiplied by the efficiency matrix and thus is transformed to a vector of frequency-shift rates ([Δf/Δt]k) (represented by the v voltage values of k = 1, …, v with v ≥ u).
We use FEM to determine
assuming a uniform distribution of particle size across the µFC cross-section area at the position, where the air flow enters the µFC (cf.
Appendix B). For the size range of the polydisperse aerosol we take 20 to 500 nm [
23] and separate it into
u = 6 size bins, which we set according to the 24 bins of a standard FMPS reference instrument as defined in
Table A4, and represent them by mean diameters of
24.1 nm, 43.2 nm, 77.3 nm, 138.0 nm, 245.9 nm, 437.7 nm for
l = 1, …, 6. The elements of the 6 × 6 efficiency matrix in Equation (6) are obtained by averaging the respective elements of the 6 × 24 efficiency matrix in Equation (A1):
As expected from Equation (3) the values of the efficiency matrix in Equation (7) show a trend towards larger values with increasing applied voltage for all size bins (i.e., each column). For each voltage, maximum efficiencies are found for a medium bin size of around 100 nm. The maximum value of
ξ = 36.5% nearly corresponds to the expected fraction of positive-charged particles of ≈ 40% in a nebulized NaCl aerosol (see
Appendix A) indicating that almost all positive-charged particles of
dp = 100 nm were sampled at the highest
Ves. The decrease of efficiency with increasing particle size above 100 nm can be expected according to Equation (3). Due their larger inertia, the trajectories of such larger particles will not be sufficiently bent towards the cantilever to be captured there. The visible efficiency drop for small-size particles reflects the remarkably smaller positive-charged fraction of particles of
dp < 100 nm and can also be related to particle loss by diffusion or Brownian motion to the channel wall.
4.2. Carbon Particle Sizing Using the MEMS-DMPS
We measured the frequency shift rate Δ
f/Δ
t of our MEMS-DMPS at sampling voltages
Ves that varied from –30 V to 0 and from 0 to −25 V with carbon UFPs of a constant mass concentration of 10 µg/m
3 and polydisperse size distribution (see
Figure 4). With the MEMS-DMPS, each Δ
f/Δ
t value was determined from the frequency-shift measured after sampling times of
ts = 9.5 min by averaging over 15 times repeated frequency sweeps. In
Figure 6 these measured values are compared with
according to Equation (A8) from FEM. Reasonable agreement is visible for small sampling voltages
Ves ≤ |−15 V|. Above this range, FEM shows a flatter dependence on
Ves than expected from the experiment. As a possible reason for this, differences in charge distribution between the reference aerosol (NaCl) and carbon may be considered. Furthermore, the capture cross-section, i.e., the maximum distance
rp at which a propagating particle will be captured on the wire, depends on the electrical field around the wire, which is not considered in the modelling. An increase of
rp can be expected at increasing voltage
Ves, which corresponds to a larger cross-section than expected and may lead to the larger frequency-shift rate visible in the experiment.
For retrieving the size distribution of the considered carbon aerosol, Equation (6) was used to transform the given vector of frequency shift rates ([Δ
f/Δ
t]
k) (with
k = 1, …,
v) at
Ves = −5 V, −10 V, −15 V, −20 V, −25 V, and −30 V (FEM values in
Figure A2) back into a vector of mass concentrations
cm,l (
l = 1, …,
u) at
24.1 nm, 43.2 nm, 77.3 nm, 138.0 nm, 245.9 nm, and 437.7 nm. For this, the system of six equations (corresponding to Equation (6)):
was solved using the approximation method of the non-negative least squares in MatLab (
Isqnonneg()). The results are plotted in
Figure 7 superimposed to the size distribution measured using FMPS in the original 24 size bins as well as after compression into the 6 bins of the MEMS-DMPS. The corresponding number concentrations were recalculated from the mass concentrations according to Equation (4). We find good agreement within the entire size range except the smallest bin of
24.1 nm, where a zero concentration value was obtained with the FEM data of the MEMS-DMPS. Here, a mass concentration of 0.16 µg/m
3 (
cn,1 = 8.6 × 10
3 cm
−3) was found by FMPS, which is far below the LOD of the MEMS-DMPS of 0.73 µg/m
3. For the next bin of
43.2 nm, where we have a mass concentration of 0.84 µg/m
3 (
cn,2 = 9.0 × 10
3 cm
−3) according to the FMPS, the MEMS-DMPS yields 0.97 µg/m
3 (
cn,2 = 10.1 × 10
3 cm
−3), which is in quite good agreement of 12% (15%).
From the number concentrations, we can calculate the amount of sampled NPs by the MEMS-DMPS for each considered diameter bin from the sucked-in air volume of vs. ≈3 µL at an air flow rate Q ≈ 0.3 µL/min and a sampling time ts = 9.5 min. We find total particle numbers ranging from ≈ 30,000 particles for the bin of the smallest diameters (l = 2, dp = 43.2 nm) to 366 particles for the bin of the largest diameters (l = 5, dp = 249.9 nm), which cause statistical errors ranging from 0.6% to 5.2%, respectively.
Figure 8 shows a correlation plot of the mass and number concentrations obtained by FEM with the MEMS-DMPS vs. the values measured by FMPS, indicating small maximum deviations of less than 13.6% and 11.1% for mass and number concentration, respectively.