Next Article in Journal
Estimate the Unknown Environment with Biosonar Echoes—A Simulation Study
Next Article in Special Issue
Beta-Lactam Antibiotic Discrimination Using a Macromolecular Sensor in Water at Neutral pH
Previous Article in Journal
Attention Fusion for One-Stage Multispectral Pedestrian Detection
Previous Article in Special Issue
Feasibility of Using Electrical Impedance Spectroscopy for Assessing Biological Cell Damage during Freezing and Thawing
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Analytical Parameters of a Novel Glucose Biosensor Based on Grafted PFM as a Covalent Immobilization Technique

by
Margalida Artigues
1,
Joan Gilabert-Porres
2,3,
Robert Texidó
2,3,
Salvador Borrós
2,3,4,
Jordi Abellà
1 and
Sergi Colominas
1,*
1
Electrochemical Methods Laboratory-Analytical and Applied Chemistry Department at Institut Químic de Sarrià, Universitat Ramon Llull, Via Augusta, 390, 08017 Barcelona, Spain
2
Tractivus SL, Via Augusta, 394, 08017 Barcelona, Spain
3
Grup d’Enginyeria de Materials (GEMAT) at Institut Químic de Sarrià, Universitat Ramon Llull, Via Augusta, 390, 08017 Barcelona, Spain
4
CIBER-BBN, Networking Center on Bioengineering, Biomaterials and Nanomedicine, 500018 Zaragoza, Spain
*
Author to whom correspondence should be addressed.
Sensors 2021, 21(12), 4185; https://doi.org/10.3390/s21124185
Submission received: 11 May 2021 / Revised: 31 May 2021 / Accepted: 15 June 2021 / Published: 18 June 2021

Abstract

:
Bioanalytical methods, in particular electrochemical biosensors, are increasingly used in different industrial sectors due to their simplicity, low cost, and fast response. However, to be able to reliably use this type of device, it is necessary to undertake in-depth evaluation of their fundamental analytical parameters. In this work, analytical parameters of an amperometric biosensor based on covalent immobilization of glucose oxidase (GOx) were evaluated. GOx was immobilized using plasma-grafted pentafluorophenyl methacrylate (pgPFM) as an anchor onto a tailored HEMA-co-EGDA hydrogel that coats a titanium dioxide nanotubes array (TiO2NTAs). Finally, chitosan was used to protect the enzyme molecules. The biosensor offered outstanding analytical parameters: repeatability (RSD = 1.7%), reproducibility (RSD = 1.3%), accuracy (deviation = 4.8%), and robustness (RSD = 2.4%). In addition, the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor showed good long-term stability; after 20 days, it retained 89% of its initial sensitivity. Finally, glucose concentrations of different food samples were measured and compared using an official standard method (HPLC). Deviation was lower than 10% in all measured samples. Therefore, the developed biosensor can be considered to be a reliable analytical tool for quantification measurements.

Graphical Abstract

1. Introduction

Biosensors and related bioanalytical tools are increasingly used as complements for standard analytical methods in a large number of applications, such as environmental monitoring [1,2,3], clinical diagnostic [4,5,6], and food analysis [7,8,9]. These devices offer fast and cost-effective measurements that can be made in situ, and are able to provide real-time measurements. Moreover, they are easy to miniaturize and, in most cases, require minimum or no sample preparation [10]. In this context, significant efforts have been recently made to develop wearable and non-invasive biosensors for in vivo monitoring of complex samples [11,12,13]. The combination of new communication technologies, such as smartphones and tablets, with these bioanalytical devices allow the development and acquisition of easy-to-use devices and easy-to-read measurements [14,15].
The main challenge for commercial bioanalytical devices commonly lies in the material used as an interface. This material must be biocompatible or able to generate a bio-friendly environment, have a high surface area, be easy to manufacture, and, for electrochemical biosensors, have intrinsic conductivity [16]. In addition, when the target analyte is a large biomolecule, such as a nucleic acid or protein, additional challenges exist related to surface fouling and non-specific adsorption of molecules at its interface [17]. Fouling is defined as spontaneous accumulation of solids on the biosensor surface [18]. These solids result in blocking, which obstructs the flow of analyte and products from the matrix of the measurement solution to the bioreceptor, and vice versa. Therefore, the biosensor sensitivity decreases, possibly causing interference, reducing the quantification reliability, and usually, negatively affecting the operational lifetime. To minimize these problems, different strategies can be applied, which are classified into passive and active approaches [19]. Active approaches are based on removing already-adsorbed fouling elements, and can be achieved using temperature-responsive materials [20], acoustic waves [21], or mechanical actuation [22]. By comparison, passive approaches are related to the prevention of incoming fouling. Passive techniques usually imply the use of polymers or hydrogels as protective barriers [23,24,25,26,27]. For example, polyhydroxyethyl methacrylate (pHEMA) is an excellent candidate for generating protective barriers against fouling because it is a hydrophilic polymer that exhibits resistance to nonspecific adhesion of proteins [28,29,30].
It is worth mentioning that to effectively manage quality control and perform quality assurance tests, it is critical to monitor specific targets with accuracy, precision, and robustness [31,32]. Unfortunately, the reliability of biosensors is usually limited. Commercial biosensors provide accuracy values around ±20% and require recalibration several times per day using standard solutions [33]. In addition, quantification by direct interpolation to a calibration curve can lead to matrix effect problems. For this reason, analytical methods that do not require a calibration curve for operation are an attractive alternative for this type of sensor. In some cases, the standard additions method is a viable methodology to mitigate matrix effects and improve accuracy in the quantification process [34,35,36].
Furthermore, classical analytical techniques, such as HPLC or spectrophotometry, are usually time consuming, have a high cost, require expert staff, and involve complex sample preparation methods. For this reason, the development of rapid, low-cost, and easy-to-perform methods remains an active research area. Electrochemical biosensors are one of the most attractive options in this field due to their promising characteristics that fulfil these market demands.
In the present work, the analytical parameters of an electrochemical biosensor were evaluated following an electrochemical method for glucose quantification. Measurements were performed using an amperometric configuration based on covalent immobilization of glucose oxidase (GOx) onto an electrochemical platform. This platform was obtained by grafting PFM (pentafluorophenyl methacrylate) onto tailored plasma polymerized HEMA-co-EGDA (ppHEMA-co-EGDA) hydrogel, which coats a titanium dioxide nanotubes array (TiO2NTAs). The biosensor architecture offers several benefits for its analytical performance, such as the high surface area of TiO2NTAs, which increases sensitivity; biocompatibility; covalent immobilization of GOx via amide bond formation, which minimizes leakage of enzymes from the biosensor structure; and the protective film of chitosan, which increases the long-term stability.

2. Materials and Methods

2.1. Materials

Titanium (99.7%), GOx type VII from Aspergillus niger, low molecular weight chitosan, and hydroxyethyl methacrylate (HEMA) were all supplied by Sigma Aldrich. Pentafluorophenyl methacrylate (PFM) was supplied by Apollo Scientific, and Argon 5.0 was supplied by Carburos Metaálicos.

2.2. Biosensor Preparation

First, Ti/TiO2NTAs was obtained by anodic oxidation (applied voltage 30 V for 17 h). The anodization process is explained in more detail in reference [37]. Then, the PFM plasma grafted surface was performed using an initial film deposition of HEMA-co-EGDA. This process is described in more detail in a previous work [38]. Briefly, plasma modification techniques were applied to the Ti/TiO2NTAs electrode to obtain a polymerized film of HEMA-co-EGDA and a grafted deposition of PFM. All the plasma modification were carried out in a stainless-steel vertical plasma reactor (base pressure about 9 × 10−4 mbar). To obtain the ppHEMA-co-EGDA film, the reactor was fed with approximately twice as much monomer (HEMA, 3.6 × 10−2 mbar) as cross-linker (EGDA, 1.2 × 10−2 mbar), and plasma polymerization was performed (20 W, duty cycle 2/22) for 10 min. Then, the ppHEMA-co-EGDA film was activated by Ar plasma and PFM was grafted at 7.7 × 10−2 mbar for 15 min. Thus, Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM electrodes were obtained. Finally, GOx was immobilized by placing 20 µL of a GOx solution (15 mg/500 µL) on the modified PFM/grafted film. Figure 1 shows a schematic representation of the covalent immobilization process applied to attach the glucose oxidase molecules to the TiO2NTAs.
When the biosensors were not in use, they were immersed in 0.1 M PBS (pH 7.0) at 4 °C.

2.3. Electrochemical Measurements

Electrochemical measurements were performed using a Autolab 302N potentiostat/galvanostat and a standard three-electrode configuration (RE Ag/AgCl/3 M KCl and CE Pt foil). The working electrode was mounted in an RDE system (EG&G PARC model 616). The voltage applied to the biosensor was set to −0.4 V vs. Ag/AgCl/3 M KCl reference electrode, and amperometric measurements were performed at 2000 rpm. A schematic representation of the experimental setup is shown in Figure 2.
To perform the measurements, 100 mL of PBS 0.1 M (pH 7.0) was added to the measuring cell and the flowing current was measured continuously.

2.4. Glucose Determination in Food Samples

Glucose was determined using the standard addition method, in a set of different samples, using the developed biosensor. For this purpose, the current was continuously measured by applying a constant potential of −0.4 V vs. ref. Under these conditions, a blank signal corresponding to 100 mL of 0.1 M PBS (pH 7.0) was registered. After 2 min, the required amount of sample was placed without farther preparation into the measuring vessel. Finally, two consecutive glucose additions of a standard solution were made.
HPLC measurements were performed using the procedure described by the AOAC Official Method [39]. Samples for HPLC measurements were prepared as described in our previous work [37,38].

3. Results and Discussion

Glucose biosensors were prepared by covalent immobilization of GOx using PFM as an enzyme anchor via amide bond formation [40,41,42,43]. It was demonstrated in previous work that when PFM plasma was grafted onto an organic film of plasma polymerized HEMA-co-EGDA, the immobilized GOx molecules presented conformational active structures [38]. In previous work, the chemical reaction between PFM groups (plasma polymerized and grafted) was extensively studied, demonstrating the ability of PFM to react with amines, using IRRAS spectroscopy [44]. Thus, plasma grafting allowed high enzymatic activity of GOx on the electrode surface to be maintained.
Glucose measurements were indirectly made by reducing hydrogen peroxide to water at the electrochemical interface (Ti/TiO2NTAs). H2O2 was generated by the enzymatic conversion of glucose to gluconic acid in the presence of oxygen (see Reactions (1) and (2)):
Glucose + O2 → H2O2 + Gluconic Acid
H2O2 + 2 H+ + 2e → 2H2
In previous work, using a similar biosensor architecture, the applied working potential was determined to be −0.4 V vs. the reference electrode (Ag/AgCl/KCl 3M) [37]. In this manner, classical reductive interferents, such as ascorbic acid or citric acid, can be avoided. In the present work, it was assumed that the working potential was not affected by the PFM plasma grafting. This modification technique generates isolated brushes of PFM in the electrochemical interface (TiO2NTAs/Ti), which should not alter the electrochemical capacity of the biosensor following the reduction of H2O2 to H2O [38].
The linear range, limit of detection (LOD), and limit of quantification (LOQ) of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor were reported in a previous work [38]. The obtained values were: linear range from 0.25 to 1.49 mM, sensitivity 9.76 µA·mM−1, LOD 0.10 mM, and LOQ 0.20 mM. In addition, in the present work, the sensitivity of three different biosensors was measured, in addition to LOD and LOQ in the same linear range (from 0.25 to 1.49 mM glucose). The data are shown in Table 1.
Table 1 shows that differences between replicates were obtained (sensitivity, LOD, and LOQ). The average sensitivity was 8.53 ± 2.39 µA·mM−1. The variation of the sensitivity values is attributed to the complex construction process of the biosensor. However, it should be noted that this variation does not affect the sample quantification because it is performed using the standard addition method.
In the present work, we undertake in-depth evaluation of the sensor’s analytical performance.

3.1. Evaluation of the Biosensor Enzymatic Activity

The enzymatic activity of immobilized enzymes is an important parameter for evaluating the development processes of new biosensors. The greater the enzymatic activity, the better the biosensor performance. An indicator of the enzymatic activity exhibited by enzyme molecules is the Michaelis–Menten constant (KM), which is defined as the concentration of substrate necessary for the rate of the enzymatic reaction to reach a value equal to half the maximum rate. The lower the KM value, the greater the ability of the substrate to bind to the active center of the enzyme and, therefore, the greater the catalytic activity of the enzyme [45]. By performing amperometric determinations, it is possible to relate the electrochemical information of the biosensors to the enzyme activity; to do this, the apparent Michaelis−Menten constant (KMapp) is estimated using the Lineweaver−Burk equation (Equation (3)) [46].
1 i ss = 1 i max + k M app i max · 1 C glucose
where iss (µA) is the steady-state response current after the addition of the substrate, imax (µA) is the maximum current under saturated substrate conditions, and Cglucose (mM) is the bulk glucose concentration. Therefore, by plotting the inverse of the substrate concentration vs. the inverse of the corrected intensity, a straight line with slope KMapp/imax and with the intercept equal to 1/imax is obtained. A calibration curve obtained with the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor is used to obtain this information, and is presented in Figure 3.
As shown in Figure 3, the obtained linear regression equation using the data from the calibration curve was 1/iss (µA−1) = −0.083·1/CGlucose (mM−1) − 0.020. Under the conditions of this study, KMapp of Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors was estimated to be 4.20 mM. This determination was repeated using three different Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors to measure its deviation. The resulting average KMapp value was 3.44 ± 0.67 mM with a relative standard deviation (RSD%) equal to 20%. This variation can be associated with the construction process in which several critical steps, such as the TiO2NTAs synthesis or plasma modification steps, can induce slight differences in the GOx immobilization and distribution. The obtained KMapp value was compared with others reported in the literature for similar glucose biosensors (see Table 2).
Table 2 shows that the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor yielded a low KMapp value (3.44 ± 0.67 mM) compared with others reported in the literature. This KMapp value implies a strong ability of substrate binding and high enzymatic activity of the immobilized enzyme molecules. Hence, the biosensors developed in the present work offer a greater affinity towards glucose than those listed in Table 2. This can be related to the enzymatic immobilization process that generates a biocompatible microenvironment for GOx molecules. This specific microenvironment offers adequate pH conditions and low hydrophobicity, and allows diffusion of reagents and products to/from the active center of the enzyme, blocking the passage of interferents. For these reasons, biosensors developed in the present work can maintain excellent catalyst activity once GOx molecules are immobilized.

3.2. Evaluation of Accuracy

Accuracy is the ability of the analytical method to provide values close to a given reference value. One of the simplest approaches to determine accuracy is to calculate the deviation between the obtained results using the tested method and a reference method.
In the present work, glucose concentration of a commercial orange soft drink (test sample) was determined using high-performance liquid chromatography (HPLC) as the reference method and Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor as the tested method.
The standard addition method was used to quantify glucose using the developed biosensor. The amperometric response of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor was continuously monitored by applying −0.4 V vs. ref. at 2000 rpm. The blank signal was registered for 2 min (100 mL of PBS). Then, 250 µL of sample was added to the measuring vessel. Finally, two additions of 0.25 mM glucose standard were made. Figure 4 shows the measured current over time after each addition.
As shown in Figure 4, the registered current after the sample addition increased by approximately 1.3 µA, whereas after each standard injection (0.25 mM glucose) the current increment was about 2.2 µA. These current increments resulted from the reduction of H2O2 to H2O, which occurred due to the enzymatic reaction (see Reactions (1) and (2)). The additions curve was derived by plotting the corrected current vs. the added glucose concentration. The glucose content in the sample was quantified by extrapolation in this addition curve.
Glucose concentration in the test sample was initially determined by HPLC. Measurements were performed using two replicates. Glucose determination using the biosensor was performed using three replicate measurements. The obtained results are shown in Table 3.
Table 3 shows that the obtained glucose concentration in the test sample determined by HPLC was 0.251 ± 0.001 M. The glucose content when the biosensor was used was 0.263 ± 0.005 M. The deviation between the two values was 4.8%. The Food and Drug Administration (FDA) established that a new bioanalytical method is accurate when the deviation between the obtained value and the reference value is below 15% [50]. In this work, the deviation was 4.8%, lower than the limit specified by the FDA. The obtained results demonstrate that Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor was able to determine glucose concentration with sufficient accuracy.

3.3. Evaluation of Precision

Precision indicates the ability of the analytical method to deliver the same results when performing different measurements. To evaluate the precision of the method, repeatability and reproducibility were evaluated. The Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor was used to determine the glucose content in a soft drink (test sample). All determinations were carried out using the standard addition method.
The repeatability of the analytical method was evaluated on the same laboratory session using the same biosensor. Three replicate measurements of the glucose content in the test sample were performed and results are shown in Table 4.
As shown in Table 4, the biosensor yielded an average glucose concentration equal to 0.263 ± 0.005 M with a relative standard deviation (RSD%) lower than the reference value established by the AOAC (1.9%) [51].
In addition, the method reproducibility was evaluated by measuring the glucose content of the test sample during three different laboratory sessions using the same biosensor. The obtained results can be seen in Table 5.
As shown in Table 5, the biosensor offered an average glucose concentration equal to 0.262 ± 0.003 M. In this case, the relative standard deviation (RSD%) of these measurements was lower than the reference value established by AOAC (4.0%) [52].
For the development of bioanalytical methods, the Food and Drug Administration stablished that the precision determined should not exceed 15% of the RSD [53]. The results obtained not only meet this criterion, but also the more restrictive criteria stablished by the AOAC [51,52]. Therefore, it is possible to affirm that Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors can be used to determine the glucose concentration with sufficient precision.

3.4. Evaluation of Robustness

Robustness is a measure of the susceptibility of an analytical method to changes that might occur during routine analysis, and provides an indication of the method’s reliability during normal use. It also shows the ability to reproduce the analytical method under different circumstances without the occurrence of unexpected differences [54].
For the validation of bioanalytical methods, it is not necessary to determine the robustness; however, this can be helpful to anticipate problems that may occur during the measurements [53]. In the present work, robustness was tested to anticipate potential problems related to the construction process of the biosensor that can affect the final glucose quantification. In this context, three different Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors were constructed and then used to quantify the glucose content in the test sample on three different days. The obtained results are shown in Table 6.
Table 6 shows that the average glucose concentration obtained in the robustness test was 0.266 ± 0.006 M with an RSD of 2.4%. Because the RSD does not exceed 15% [53], it is possible to affirm that the quantification method is robust. Consequently, the manufacturing steps of Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors do not have an influence on the glucose quantification process.

3.5. Evaluation of Long-Term Stability

Long-term stability is a pivotal requirement of a biosensor. This parameter defines the period during which the biosensor can be used with sufficient reliability. The long-term stability of the biosensors was evaluated by measuring its performance every few days; the slopes of calibration curves (related to sensitivity) obtained on several days were compared. Figure 5 shows the evolution over time of the sensitivity for three Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors.
Figure 4 shows that sensitivity decreases over time. During the first 20 days following the construction of the biosensors, sensitivity showed slightly deviations for each biosensor. After 20 days, the developed biosensors conserved 89% of their initial sensitivity. However, after 20 days, the decrease in sensitivity became more pronounced. Finally, after 30 days the decrease in sensitivity was greater than 80% compared with the initial value in all cases.
A decrease in sensitivity is caused by enzyme deactivation, which can occur for several reasons. First, it should be noted that hydrogen peroxide generated during the enzymatic reaction can damage the enzyme molecules. H2O2 is a highly oxidizing compound that can react with some of the amino acids present in the structure of glucose oxidase [55,56], causing modifications to its structure and affecting its catalytic activity. Second, pH can also alter the structural conformation of the immobilized enzymes. Even maintaining pH 7.0 using a 0.1 M PBS solution, local significant pH variations may occur inside the nanotubes because of the reduction of H2O2 to H2O. These local pH alterations can affect the inter- and intramolecular bonds that generate the active structure of the enzyme, causing its deactivation [57]. Finally, the degradation of chitosan is another factor to be considered. Chitosan acts as a protective film. As time passes, this film is degraded. Consequently, enzyme molecules are more exposed to the measurement medium. Thus, as chitosan degrades, GOx can more easily lose its biologically active conformation. It is reported that chitosan films are stable for a period of four weeks stored at 4 °C [58], which is in concordance with the long-term stability of the developed biosensors. It must be noted that, in the present work, biosensors were immersed in 0.1 M PBS (pH 7.0) at 4 °C when were not in use.
It should also be noted that the three evaluated Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors showed different sensitivities. This difference is associated with the construction process; the anodization process to obtain the electrochemical interface and surface plasma modification involves several steps that can influence the final active area of the electrochemical interface. Equally, it is possible that the population of enzymes in their biologically active conformation differs slightly between biosensors. Thus, a certain variation between sensitivity values can be obtained. It should be noted that, despite the differences observed between biosensors, the glucose quantification is not compromised because the quantification is carried out by the standard addition method, and not by direct interpolation of a calibration curve. This fact was previously demonstrated in the robustness study (RDS 2.4%).

3.6. Food Sample Quantification

It was demonstrated that the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors developed in the present work met the quality standards required to quantify glucose with reliability. All analytical parameters tested were in good agreement with the highly demanding AOAC and FDA quality standards. To demonstrate the versality of the proposed bioanalytical method, the developed biosensors were used to quantify glucose in six different samples: two soft drinks, one soya sauce, one tomato sauce, one yoghurt, and one sweetened nuts beverage (Spanish horchata). These complex sample matrices were selected because they show classical quantification problems. For example, lactose and galactose present in dairy products are common interferents in glucose quantification by HPLC. Moreover, high protein and fat content can cause unspecific adhesion on the electrode surface that can lead to quantification errors. Soy and tomato sauces present high fat and protein concentrations. Other classical interferents for amperometric glucose quantification are ascorbic and citric acids, which are both usually present in soft drinks [59].
Samples were also analyzed using HPLC and the obtained glucose concentrations were considered to be reference values. Samples were analyzed using two replicate measurements by both techniques, HPLC and amperometric biosensors. Glucose concentrations in each sample and their standard deviations (s), in addition to the deviation between the two methods, are shown in Table 7.
The results obtained by amperometric measurements using the biosensors were in good agreement with HPLC reference values (see Table 7) because the deviation from the reference value was lower than 15% in all cases (the standard established for bioanalytical methods [53]). Therefore, the constructed biosensors showed excellent analytical properties and the proposed quantification method allowed glucose to be determined in complex samples with adequate sensitivity, accuracy, precision, and robustness. Moreover, the described biosensor system offers the advantages of short analysis time, and the capability to be repeated due to long-term stability.
Furthermore, it should be noted that the intrinsic structure of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors developed in this work can be of special interest for clinical diagnostic and in vivo/real time measurements:
  • Electrochemical interfaces based on titanium and titanium (IV) oxide promote the biocompatibility of the system; titanium is a pharmacologically inert metal that does not cause allergic reactions in the immune system and the human body does not reject it. Moreover, TiO2NTAs offers a high surface area and the ability to promote charge transfer processes. Thus, Ti/TiO2NTAs is an excellent electrochemical interface for implantable biosensors.
  • Chitosan acts as a protective barrier for the immobilized enzyme and provides conformational stability to the measurement device. Chitosan is a biocompatible hydrogel obtained from natural sources that prevents enzyme denaturalization because of its high affinity for proteins [60]. Moreover, chitosan preserves its structure under adverse conditions due to its high mechanical stability. For this reason, using chitosan, the long-term stability of the biosensors is improved. The use of chitosan also improves the biosensor’s sensitivity because this polymeric matrix blocks possible interfering macromolecules.
  • Poly-HEMA protects the electrochemical interface from adhesion of molecules that can interfere with the analytical measurement and/or cause damage to the human body, generating a foreign body response [61]. Therefore, the risk of rejection of implantable devices is reduced by using this polymer in the biosensor’ architecture. Furthermore, by using plasma techniques for HEMA polymerization, the obtained films follow the shape of the electrochemical interface [62]. Thus, ppHEMA-co-EGDA films follow the architecture of the TiO2NTAs, and both high specific surface area and high sensitivity are maintained.
  • Covalent immobilization of enzyme molecules provides several benefits. First, it decreases the probability of enzyme leakage from the biosensor architecture. Second, covalent configuration ensures that enzymes are located close enough to the transducer to avoid the loss of the electrons produced from the enzymatic reaction. In addition, the pgPFM surface shows high covalency, which is the ability of the surface to retain the attached molecules after vigorous washing [63]. Hence, the developed biosensors are likely to retain their sensitivity under milder conditions, such as those found in the human bloodstream. Finally, the PFM plasma-grafting modification minimizes enzyme deactivation during the immobilization process [38].
It should also be noted that in our previous work [37] we used polymer entrapment as an enzyme immobilization method in the sensor architecture. In contrast, in the present work the enzyme immobilization process was changed to covalent binding (pgPFM and HEMA-co-EGDA). An advantage of this method is that, because of the stable nature of the bonds formed between the enzyme and matrix, the enzyme is not released into the solution upon use. In addition, when covalent binding is used, the enzyme is strongly bound to the matrix and is also stabilized [64]. Furthermore, GOx molecules adopt biological active conformations (native and molten globule structures) when immobilized in the pgPFM and HEMA-co-EGDA layer. This was demonstrated using QCM-D [38]. For this reason, it was possible to increase the enzymatic capacity of the biosensor and, as a result, a higher sensitivity than that of our previous work [37] (5.46 µA·mM−1) was obtained (9.76 µA·mM−1).
Therefore, Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors showed outstanding analytical properties due to the synergy of the electrochemical interface, the immobilization technique, and the protection matrix. In addition, because of this synergy, the proposed biosensors have significant potential for the development of implantable devices for biomedical applications.

4. Conclusions

The analytical parameters of Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors were evaluated using a commercial soft drink as a test sample. The immobilized enzyme molecules on the biosensor showed a strong ability of substrate binding and high enzymatic activity, as indicated by the KMapp value of 3.44 ± 0.67 mM. In addition, measurements undertaken with this biosensor showed high accuracy, with a deviation from HPLC results equal to 4.8%. This biosensor offered excellent analytical parameters: high repeatability (RSD = 1.7%), high reproducibility (RSD = 1.3%), and high robustness (RSD = 2.4%). Finally, the long-term stability was further examined and after 20 days, the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors retained 89% of their initial sensitivity.
Moreover, glucose quantification was performed in complex samples, in which sample preparation only required dilution. For all of the studied samples, the obtained results were in good agreement with the HPLC reference values. In addition, the proposed biosensor was able to overcome typical sources of interference in glucose quantification, such us galactose, ascorbic acid, fats, and proteins. Therefore, the proposed biosensor can be considered to be an inexpensive, fast, and simple alternative to classical analytical quantification methods.
The exceptional analytical properties of this biosensor can be attributed to its intrinsic architecture. The unique electrochemical platform is a versatile interface that is biocompatible and promotes charge transfer processes, leading to increase sensitivity. The immobilization process, which consists of the formation of an amide bond, minimizes the loss of enzyme molecules. In conjunction with the use of chitosan as a protective barrier, this improves long-term stability. Finally, the presence of polyHEMA, which prevents nonspecific adhesion of proteins, ensures the biosensor has significant versatility. Due to the synergy of these features, the proposed biosensor is a good candidate for future applications in diagnostic, food, and environmental analysis.

Author Contributions

Conceptualization, J.A., S.B. and S.C.; methodology, J.A., J.G.-P., M.A., R.T., S.B. and S.C.; validation, J.A., M.A. and S.C.; formal analysis, J.A., J.G.-P., M.A., R.T., S.B. and S.C.; investigation, J.A., J.G.-P., M.A., R.T., S.B. and S.C.; resources, J.A., S.B. and S.C.; data curation, M.A.; writing—original draft preparation, M.A.; writing—review and editing, J.A., J.G.-P., M.A., R.T., S.B. and S.C.; supervision, J.A., S.B. and S.C.; project administration, J.A. and S.C.; funding acquisition, J.A., S.B. and S.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

The authors wish to thank Francesc Broto from Institut Quimic de Sarria for the HPLC analysis.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Campuzano, S.; Yáñez-Sedeño, P.; Pingarrón, J.M. Electrochemical Affinity Biosensors Based on Selected Nanostructures for Food and Environmental Monitoring. Sensors 2020, 20, 5125. [Google Scholar] [CrossRef]
  2. McConnell, E.M.; Nguyen, J.; Li, Y. Aptamer-Based Biosensors for Environmental Monitoring. Front. Chem. 2020, 8, 434. [Google Scholar] [CrossRef]
  3. Liu, B.; Zhuang, J.; Wei, G. Recent advances in the design of colorimetric sensors for environmental monitoring. Environ. Sci. Nano 2020, 7, 2195–2213. [Google Scholar] [CrossRef]
  4. Wang, H.; He, D.; Wan, K.; Sheng, X.; Cheng, H.; Huang, J.; Zhou, X.; He, X.; Wang, K. In situ multiplex detection of serum exosomal microRNAs using an all-in-one biosensor for breast cancer diagnosis. Analyst 2020, 145, 3289–3296. [Google Scholar] [CrossRef]
  5. Sun, Z.; Wang, L.; Wu, S.; Pan, Y.; Dong, Y.; Zhu, S.; Yang, J.; Yin, Y.; Li, G. An Electrochemical Biosensor Designed by Using Zr-Based Metal–Organic Frameworks for the Detection of Glioblastoma-Derived Exosomes with Practical Application. Anal. Chem. 2020, 92, 3819–3826. [Google Scholar] [CrossRef] [PubMed]
  6. Kou, X.; Tong, L.; Shen, Y.; Zhu, W.; Yin, L.; Huang, S.; Zhu, F.; Chen, G.; Ouyang, G. Smartphone-assisted robust enzymes@MOFs-based paper biosensor for point-of-care detection. Biosens. Bioelectron. 2020, 156, 112095. [Google Scholar] [CrossRef] [PubMed]
  7. Bougadi, E.T.; Kalogianni, D.P. Paper-based DNA biosensor for food authenticity testing. Food Chem. 2020, 322, 126758. [Google Scholar] [CrossRef] [PubMed]
  8. Ye, Y.; Ji, J.; Sun, Z.; Shen, P.; Sun, X. Recent advances in electrochemical biosensors for antioxidant analysis in foodstuff. TrAC Trends Anal. Chem. 2020, 122, 115718. [Google Scholar] [CrossRef]
  9. Zhou, Q.; Tang, D. Recent advances in photoelectrochemical biosensors for analysis of mycotoxins in food. TrAC Trends Anal. Chem. 2020, 124, 115814. [Google Scholar] [CrossRef]
  10. Bettazzi, F.; Marrazza, G.; Minunni, M.; Palchetti, I.; Scarano, S. Biosensors and Related Bioanalytical Tools. In Comprehensive Analytical Chemistry; Elsevier: Amsterdam, The Netherlands, 2017; pp. 1–33. [Google Scholar]
  11. Lee, H.-B.; Meeseepong, M.; Trung, T.Q.; Kim, B.-Y.; Lee, N.-E. A wearable lab-on-a-patch platform with stretchable nanostructured biosensor for non-invasive immunodetection of biomarker in sweat. Biosens. Bioelectron. 2020, 156, 112133. [Google Scholar] [CrossRef]
  12. Su, Y.; Yang, T.; Zhao, X.; Cai, Z.; Chen, G.; Yao, M.; Chen, K.; Bick, M.; Wang, J.; Li, S.; et al. A wireless energy transmission enabled wearable active acetone biosensor for non-invasive prediabetes diagnosis. Nano Energy 2020, 74, 104941. [Google Scholar] [CrossRef]
  13. Karpova, E.V.; Karyakina, E.E.; Karyakin, A.A. Wearable non-invasive monitors of diabetes and hypoxia through continuous analysis of sweat. Talanta 2020, 215, 120922. [Google Scholar] [CrossRef]
  14. Huang, L.; Xiao, W.; Xu, T.; Chen, H.; Jin, Z.; Zhang, Z.; Song, Q.; Tang, Y. Miniaturized Paper-Based Smartphone Biosensor for Differential Diagnosis of Wild-type Pseudorabies Virus Infection versus Vaccination Immunization. Sens. Actuators B Chem. 2021, 327, 128893. [Google Scholar] [CrossRef]
  15. Zhang, M.; Zhang, Y.; Yang, C.; Ma, C.; Tang, J. A smartphone-assisted portable biosensor using laccase-mineral hybrid microflowers for colorimetric determination of epinephrine. Talanta 2021, 224, 121840. [Google Scholar] [CrossRef]
  16. Otero, F.; Magner, E. Biosensors—Recent Advances and Future Challenges in Electrode Materials. Sensors 2020, 20, 3561. [Google Scholar] [CrossRef]
  17. Labib, M.; Sargent, E.H.; Kelley, S.O. Electrochemical Methods for the Analysis of Clinically Relevant Biomolecules. Chem. Rev. 2016, 116, 9001–9090. [Google Scholar] [CrossRef]
  18. Schoenitz, M.; Grundemann, L.; Augustin, W.; Scholl, S. Fouling in microstructured devices: A review. Chem. Commun. 2015, 51, 8213–8228. [Google Scholar] [CrossRef] [PubMed]
  19. Xu, J.; Lee, H. Anti-Biofouling Strategies for Long-Term Continuous Use of Implantable Biosensors. Chemosensors 2020, 8, 66. [Google Scholar] [CrossRef]
  20. Abraham, A.A.; Means, A.K.; Clubb, F.J.; Fei, R.; Locke, A.K.; Gacasan, E.G.; Coté, G.L.; Grunlan, M.A. Foreign Body Reaction to a Subcutaneously Implanted Self-Cleaning, Thermoresponsive Hydrogel Membrane for Glucose Biosensors. ACS Biomater. Sci. Eng. 2018, 4, 4104–4111. [Google Scholar] [CrossRef]
  21. Pan, S.; Zhang, H.; Liu, W.; Wang, Y.; Pang, W.; Duan, X. Biofouling Removal and Protein Detection Using a Hypersonic Resonator. ACS Sens. 2017, 2, 1175–1183. [Google Scholar] [CrossRef] [PubMed]
  22. Park, H.; Raffiee, A.H.; John, S.W.M.; Ardekani, A.M.; Lee, H. Towards smart self-clearing glaucoma drainage device. Microsyst. Nanoeng. 2018, 4, 35. [Google Scholar] [CrossRef] [PubMed]
  23. Wang, Y.; Cui, M.; Jiao, M.; Luo, X. Antifouling and ultrasensitive biosensing interface based on self-assembled peptide and aptamer on macroporous gold for electrochemical detection of immunoglobulin E in serum. Anal. Bioanal. Chem. 2018, 410, 5871–5878. [Google Scholar] [CrossRef]
  24. Yu, B.; Wang, C.; Ju, Y.M.; West, L.; Harmon, J.; Moussy, Y.; Moussy, F. Use of hydrogel coating to improve the performance of implanted glucose sensors. Biosens. Bioelectron. 2008, 23, 1278–1284. [Google Scholar] [CrossRef] [PubMed]
  25. Wu, H.; Lee, C.-J.; Wang, H.; Hu, Y.; Young, M.; Han, Y.; Xu, F.-J.; Cong, H.; Cheng, G. Highly sensitive and stable zwitterionic poly(sulfobetaine-3,4-ethylenedioxythiophene) (PSBEDOT) glucose biosensor. Chem. Sci. 2018, 9, 2540–2546. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  26. Sun, C.; Miao, J.; Yan, J.; Yang, K.; Mao, C.; Ju, J.; Shen, J. Applications of antibiofouling PEG-coating in electrochemical biosensors for determination of glucose in whole blood. Electrochim. Acta 2013, 89, 549–554. [Google Scholar] [CrossRef]
  27. Chan, D.; Chien, J.-C.; Axpe, E.; Blankemeier, L.; Baker, S.; Swaminathan, S.; Piunova, V.; Zubarev, D.Y.; Maikawa, C.; Soh, H.T.; et al. Combinatorial Polyacrylamide Hydrogels for Preventing Biofouling on Implantable Biosensors. bioRxiv 2020. [Google Scholar] [CrossRef]
  28. Tanaka, M.; Mochizuki, A.; Ishii, N.; Motomura, T.; Hatakeyama, T. Study of blood compatibility with poly(2-methoxyethyl acrylate). Relationship between water structure and platelet compatibility in poly(2-methoxyethylacrylate-co-2-hydroxyethylmethacrylate). Biomacromolecules 2002, 3, 36–41. [Google Scholar] [CrossRef]
  29. Iwasaki, Y.; Ishihara, K. Phosphorylcholine-containing polymers for biomedical applications. Anal. Bioanal. Chem. 2005, 381, 534–546. [Google Scholar] [CrossRef]
  30. Morra, M.; Cassinelli, C. Surface field of forces and protein adsorption behavior of poly(hydroxyethylmethacrylate) films deposited from plasma. J. Biomed. Mater. Res. 1995, 29, 39–45. [Google Scholar] [CrossRef] [PubMed]
  31. Malone-Povolny, M.J.; Merricks, E.P.; Wimsey, L.E.; Nichols, T.C.; Schoenfisch, M.H. Long-Term Accurate Continuous Glucose Biosensors via Extended Nitric Oxide Release. ACS Sens. 2019, 4, 3257–3264. [Google Scholar] [CrossRef]
  32. Li, P.; Lee, G.-H.; Kim, S.Y.; Kwon, S.Y.; Kim, H.-R.; Park, S. From Diagnosis to Treatment: Recent Advances in Patient-Friendly Biosensors and Implantable Devices. ACS Nano 2021, 15, 1960–2004. [Google Scholar] [CrossRef]
  33. Li, H.; Dauphin-Ducharme, P.; Ortega, G.; Plaxco, K.W. Calibration-Free Electrochemical Biosensors Supporting Accurate Molecular Measurements Directly in Undiluted Whole Blood. J. Am. Chem. Soc. 2017, 139, 11207–11213. [Google Scholar] [CrossRef] [PubMed]
  34. Tran, T.-T.; Clark, K.; Ma, W.; Mulchandani, A. Detection of a secreted protein biomarker for citrus Huanglongbing using a single-walled carbon nanotubes-based chemiresistive biosensor. Biosens. Bioelectron. 2020, 147, 111766. [Google Scholar] [CrossRef]
  35. Muñoz-San Martín, C.; Pedrero, M.; Gamella, M.; Montero-Calle, A.; Barderas, R.; Campuzano, S.; Pingarrón, J.M. A novel peptide-based electrochemical biosensor for the determination of a metastasis-linked protease in pancreatic cancer cells. Anal. Bioanal. Chem. 2020, 412, 6177–6188. [Google Scholar] [CrossRef]
  36. Bilgi, M.; Ayranci, E. Development of amperometric biosensors using screen-printed carbon electrodes modified with conducting polymer and nanomaterials for the analysis of ethanol, methanol and their mixtures. J. Electroanal. Chem. 2018, 823, 588–592. [Google Scholar] [CrossRef]
  37. Artigues, M.; Abellà, J.; Colominas, S. Analytical Parameters of an Amperometric Glucose Biosensor for Fast Analysis in Food Samples. Sensors 2017, 17, 2620. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  38. Artigues, M.; Oh, S.; Gilabert-Porres, J.; Abellà, J.; Borrós, S.; Colominas, S. Novel grafted electrochemical interface for covalent glucose oxidase immobilization using reactive pentafluorophenyl methacrylate. Colloids Surf. B Biointerfaces 2019, 175. [Google Scholar] [CrossRef] [PubMed]
  39. AOAC Official Method. AOAC 977.20-1977, Separation of Sugars in Honey, Liquid Chromatographic Method. JAOAC 1977, 60, 838. [Google Scholar]
  40. Francesch, L.; Borros, S.; Knoll, W.; Förch, R. Surface reactivity of pulsed-plasma polymerized pentafluorophenyl methacrylate (PFM) toward amines and proteins in solution. Langmuir 2007, 23, 3927–3931. [Google Scholar] [CrossRef]
  41. Duque, L.; Queralto, N.; Francesch, L.; Bumbu, G.G.; Borros, S.; Berger, R.; Förch, R. Reactions of plasma-polymerised pentafluorophenyl methacrylate with simple amines. Plasma Process. Polym. 2010, 7, 915–925. [Google Scholar] [CrossRef]
  42. Duque, L.; Menges, B.; Borros, S.; Förch, R. Immobilization of biomolecules to plasma polymerized pentafluorophenyl methacrylate. Biomacromolecules 2010, 11, 2818–2823. [Google Scholar] [CrossRef]
  43. Montero, L.; Baxamusa, S.H.; Borros, S.; Gleason, K.K. Thin hydrogel films with nanoconfined surface reactivity by photoinitiated chemical vapor deposition. Chem. Mater. 2009, 21, 399–403. [Google Scholar] [CrossRef]
  44. Queralto, N.; Bumbu, G.G.; Francesch, L.; Knoll, W.; Borros, S.; Berger, R.; Förch, R. Investigation into the Chemical Reactivity of Plasma-Deposited Perfluorophenyl Methacrylate Using Infrared Reflection Absorption Spectroscopy and Microcantilever Studies. Plasma Process. Polym. 2007, 4, S790–S793. [Google Scholar] [CrossRef]
  45. Liu, Y.; Zhang, J.; Cheng, Y.; Jiang, S.P. Effect of Carbon Nanotubes on Direct Electron Transfer and Electrocatalytic Activity of Immobilized Glucose Oxidase. ACS Omega 2018, 3, 667–676. [Google Scholar] [CrossRef]
  46. Krishnan, S.K.; Prokhorov, E.; Bahena, D.; Esparza, R.; Meyyappan, M. Chitosan-Covered Pd@Pt Core–Shell Nanocubes for Direct Electron Transfer in Electrochemical Enzymatic Glucose Biosensor. ACS Omega 2017, 2, 1896–1904. [Google Scholar] [CrossRef] [Green Version]
  47. Kang, X.; Mai, Z.; Zou, X.; Cai, P.; Mo, J. A novel glucose biosensor based on immobilization of glucose oxidase in chitosan on a glassy carbon electrode modified with gold-platinum alloy nanoparticles/multiwall carbon nanotubes. Anal. Biochem. 2007, 369, 71–79. [Google Scholar] [CrossRef]
  48. Tang, H.; Chen, J.; Yao, S.; Nie, L.; Deng, G.; Kuang, Y. Amperometric glucose biosensor based on adsorption of glucose oxidase at platinum nanoparticle-modified carbon nanotube electrode. Anal. Biochem. 2004, 331, 89–97. [Google Scholar] [CrossRef]
  49. Zhang, Z.; Xie, Y.; Liu, Z.; Rong, F.; Wang, Y.; Fu, D. Covalently immobilized biosensor based on gold nanoparticles modified TiO2 nanotube arrays. J. Electroanal. Chem. 2011, 650, 241–247. [Google Scholar] [CrossRef]
  50. Center for Drug Evaluation and Research (CDER); Center for Veterinary Medicine (CVM) Bioanalytical Method Validation. Guidance for Industry; The Food and Drug Administration: White Oak Campus, MD, USA, 2018; pp. 1–41. [Google Scholar]
  51. AOAC International. AOAC Peer-Verified Methods Program, Manual on Policies and Procedures; AOAC International: Rockville, MD, USA, 1998. [Google Scholar]
  52. AOAC International. AOAC Definitions and Calculations of Horrat Values from Intralaboratory Data; AOAC International: Rockville, MD, USA, 2004. [Google Scholar]
  53. Tiwari, G.; Tiwari, R. Bioanalytical method validation: An updated review. Pharm. Methods 2010, 1, 25–38. [Google Scholar] [CrossRef]
  54. Vander Heyden, Y.; Nijhuis, A.; Smeyers-Verbeke, J.; Vandeginste, B.G.; Massart, D. Guidance for robustness/ruggedness tests in method validation. J. Pharm. Biomed. Anal. 2001, 24, 723–753. [Google Scholar] [CrossRef]
  55. Drozdz, R.; Naskalski, J.W.; Sznajd, J. Oxidation of amino acids and peptides in reaction with myeloperoxidase, chloride and hydrogen peroxide. Biochim. Biophys. Acta 1988, 957, 47–52. [Google Scholar] [CrossRef]
  56. Harris, J.M.; Reyes, C.; Lopez, G.P. Common Causes of Glucose Oxidase Instability in In Vivo Biosensing: A Brief Review. J. Diabetes Sci. Technol. 2013, 7, 1030–1038. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  57. Dumitraşcu, L.; Stănciuc, N.; Bahrim, G.E.; Ciumac, A.; Aprodu, I. pH and heat-dependent behaviour of glucose oxidase down to single molecule level by combined fluorescence spectroscopy and molecular modelling. J. Sci. Food Agric. 2016, 96, 1906–1914. [Google Scholar] [CrossRef] [PubMed]
  58. Szymańska, E.; Winnicka, K. Stability of Chitosan—A Challenge for Pharmaceutical and Biomedical Applications. Mar. Drugs 2015, 13, 1819–1846. [Google Scholar] [CrossRef]
  59. Sepulveda, J. Challenges in Routine Clinical Chemistry Testing. In Accurate Results in the Clinical Laboratory; Elsevier: Amsterdam, The Netherlands, 2013; pp. 93–129. ISBN 9780124157835. [Google Scholar]
  60. Krajewska, B. Application of chitin- and chitosan-based materials for enzyme immobilizations: A review. Enzym. Microb. Technol. 2004, 35, 126–139. [Google Scholar] [CrossRef]
  61. Baxamusa, S.H.; Montero, L.; Dubach, J.M.; Clark, H.A.; Borros, S.; Gleason, K.K. Protection of sensors for biological applications by photoinitiated chemical vapor deposition of hydrogel thin films. Biomacromolecules 2008, 9, 2857–2862. [Google Scholar] [CrossRef] [PubMed]
  62. Cifuentes, A.; Borrós, S. Comparison of two different plasma surface-modification techniques for the covalent immobilization of protein monolayers. Langmuir 2013, 29, 6645–6651. [Google Scholar] [CrossRef]
  63. Bilek, M.M.; McKenzie, D.R. Plasma modified surfaces for covalent immobilization of functional biomolecules in the absence of chemical linkers: Towards better biosensors and a new generation of medical implants. Biophys. Rev. 2010, 2, 55–65. [Google Scholar] [CrossRef] [Green Version]
  64. Guilbault, G.G.; Kauffmann, J.-M.; Patriarche, G.J. Immobilized Enzyme Electrode as Biosensor. In Protein Immobilization. Fundamentals and Applications; Taylor, R.F., Ed.; Marcel Dekker: New York, NY, USA, 1991; pp. 209–262. [Google Scholar]
Figure 1. Scheme showing the covalent immobilization process. First, a plasma modification technique was used to obtain ppHEMA-co-EGDA/pgPFM film. Then, covalent immobilization of GOx was undertaken via amide bond formation.
Figure 1. Scheme showing the covalent immobilization process. First, a plasma modification technique was used to obtain ppHEMA-co-EGDA/pgPFM film. Then, covalent immobilization of GOx was undertaken via amide bond formation.
Sensors 21 04185 g001
Figure 2. Schematic representation of the experimental setup.
Figure 2. Schematic representation of the experimental setup.
Sensors 21 04185 g002
Figure 3. Representation of the Lineweaver-Burk equation using the data obtained from the calibration curve acquired using a Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor.
Figure 3. Representation of the Lineweaver-Burk equation using the data obtained from the calibration curve acquired using a Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor.
Sensors 21 04185 g003
Figure 4. Current-time plot of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor with an applied potential of -0.4 V when 250 mL of sample and two additions of 0.25 mM glucose injections were made.
Figure 4. Current-time plot of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor with an applied potential of -0.4 V when 250 mL of sample and two additions of 0.25 mM glucose injections were made.
Sensors 21 04185 g004
Figure 5. Evaluation of long-term stability for three Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors.
Figure 5. Evaluation of long-term stability for three Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors.
Sensors 21 04185 g005
Table 1. Values of sensitivity, LOD, and LOQ obtained for several Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors. The linear range was equal for all of the evaluated sensors.
Table 1. Values of sensitivity, LOD, and LOQ obtained for several Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors. The linear range was equal for all of the evaluated sensors.
BiosensorsSensitivity/µA·mM−1LOD/mMLOQ/mM
Individual
Values
Average
[38]9.768.53 ± 2.390.100.20
Present work8.630.120.20
5.150.140.21
10.580.100.19
Table 2. Values of the apparent Michaelis–Menten constant of different glucose biosensors described in the literature and of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor developed in the present work. The value shown in the present work is the average value and its deviation obtained from three different biosensors.
Table 2. Values of the apparent Michaelis–Menten constant of different glucose biosensors described in the literature and of the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor developed in the present work. The value shown in the present work is the average value and its deviation obtained from three different biosensors.
BiosensorKMapp/mMReference
GCE-CNT/GOx5.95–14.50[45]
GCE/CS/CNT/Au-PtNPs/GOx5.20 [47]
CNT/Pt/GOx/Nafion10.11[48]
Ti/TiO2NT/AuNPs/GOx7.2[49]
Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan3.44 ± 0.67Present work
Table 3. Glucose determination using different analytical techniques on the same orange soft drink sample.
Table 3. Glucose determination using different analytical techniques on the same orange soft drink sample.
Method[Glucose] ± s/MRSD/%Deviation/%
HPLC0.251 ± 0.0010.34.8
Biosensor0.263 ± 0.0051.7
Table 4. Study of the system repeatability. Measurements were performed on the same day using the same Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor.
Table 4. Study of the system repeatability. Measurements were performed on the same day using the same Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor.
Replicate[Glucose]/M[Glucose] Average ± s/MRSD/%
10.2600.263 ± 0.0051.7
20.268
30.261
Table 5. Study of the system reproducibility. Measurements were performed on three different days using the same Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor.
Table 5. Study of the system reproducibility. Measurements were performed on three different days using the same Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor.
Day[Glucose]/M[Glucose] Average ± s/MRSD/%
10.2660.262 ± 0.0031.3
20.261
30.260
Table 6. Study of the system robustness. Measurements were performed on three different days using three different Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors.
Table 6. Study of the system robustness. Measurements were performed on three different days using three different Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensors.
DayBiosensor[Glucose]/M[Glucose] Average ± s/MRSD/%
110.2580.266 ± 0.0062.4
220.270
330.268
Table 7. Glucose concentration values obtained using the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor and HPLC values. Standard deviation (s) and deviation of both methods are also shown.
Table 7. Glucose concentration values obtained using the Ti/TiO2NTAs/ppHEMA-co-EGDA/pgPFM/GOx/Chitosan biosensor and HPLC values. Standard deviation (s) and deviation of both methods are also shown.
Sample[Glucose]biosensor ± s/M[Glucose]HPLC ± s/MDeviation %
Orange Soft Drink0.263 ± 0.0050.251 ± 0.0014.8
Lemon Soft Drink0.164 ± 0.0010.150 ± 0.0019.1
Soya Sauce0.087 ± 0.0020.096 ± 0.001−9.7
Tomato Sauce0.559 ± 0.0090.515 ± 0.0058.4
Yoghurt0.188 ± 0.0020.174 ± 0.0018.1
Horchata0.038 ± 0.0020.035 ± 0.0018.5
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Artigues, M.; Gilabert-Porres, J.; Texidó, R.; Borrós, S.; Abellà, J.; Colominas, S. Analytical Parameters of a Novel Glucose Biosensor Based on Grafted PFM as a Covalent Immobilization Technique. Sensors 2021, 21, 4185. https://doi.org/10.3390/s21124185

AMA Style

Artigues M, Gilabert-Porres J, Texidó R, Borrós S, Abellà J, Colominas S. Analytical Parameters of a Novel Glucose Biosensor Based on Grafted PFM as a Covalent Immobilization Technique. Sensors. 2021; 21(12):4185. https://doi.org/10.3390/s21124185

Chicago/Turabian Style

Artigues, Margalida, Joan Gilabert-Porres, Robert Texidó, Salvador Borrós, Jordi Abellà, and Sergi Colominas. 2021. "Analytical Parameters of a Novel Glucose Biosensor Based on Grafted PFM as a Covalent Immobilization Technique" Sensors 21, no. 12: 4185. https://doi.org/10.3390/s21124185

APA Style

Artigues, M., Gilabert-Porres, J., Texidó, R., Borrós, S., Abellà, J., & Colominas, S. (2021). Analytical Parameters of a Novel Glucose Biosensor Based on Grafted PFM as a Covalent Immobilization Technique. Sensors, 21(12), 4185. https://doi.org/10.3390/s21124185

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop