Next Article in Journal
Design Guidelines for Database-Driven Internet of Things-Enabled Dynamic Spectrum Access
Previous Article in Journal
Biosensing Applications Using Nanostructure-Based Localized Surface Plasmon Resonance Sensors
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Offshore Online Measurements of Total Suspended Solids Using Microscopy Analyzers

Department of Energy Technology, Aalborg University, Niels Bohrs Vej 8, 6700 Esbjerg, Denmark
*
Author to whom correspondence should be addressed.
Sensors 2021, 21(9), 3192; https://doi.org/10.3390/s21093192
Submission received: 6 January 2021 / Revised: 27 April 2021 / Accepted: 28 April 2021 / Published: 4 May 2021
(This article belongs to the Section Sensing and Imaging)

Abstract

:
Accurate online water quality measurements have gained attention during the last decades in the oil and gas industry for improving operational performance and protecting the surrounding environment. One potential solution to extend the reservoirs’ economic life and put less strain on the environment is by re-injecting the produced water, but the injected water quality must be high and consistent to prevent injectivity reduction. This paper evaluates two different online microscopy analyzers that utilize a high-resolution video camera for capturing images of the particles passing their view cell. The calibration procedure for both online microscopy analyzers has been thoroughly validated for steady-state and real-time measurements. The real-time measurements were achieved by post-processing the data captured by the microscopes and applying a trailing moving average window. The performance of measuring the oil-in-water concentration was compared with an online fluorescence-based monitor. The paper addresses the statistical considerations when defining the level of accuracy of the predicted particle size distribution within a defined confidence interval. Both microscopes showed promising results for measuring known particle sizes and oil-in-water concentrations, both in steady-state and real-time.

1. Introduction

Even as a general global goal is to consume energy from renewable energy sources, oil and gas are needed in the transition [1]. Oil production is expected to increase during the next three decades globally, which entails the oil and gas industry to significantly impact the world’s energy consumption in the coming decades [2]. The general trend towards more sustainable energy production also affects the offshore oil and gas industry as discharge legislation becomes stricter [3,4]. With stricter policies, produced water re-injection (PWRI) has gained growing attention to extend the reservoirs’ economic life and decrease produced water (PW) discharge to minimize environmental impacts [5]. However, the re-injected PW and the injected seawater must have a continually high quality to prevent formation damage and unpredicted injectivity reduction [6,7,8,9,10]. Currently, only ∼14.5% of the PW in the Danish sector of the North Sea is re-injected, which is a >50% reduction since 2009 due to reservoir challenges according to the Danish Environmental Protection Agency [3,11]. To increase the PWRI percentage, efficient management is essential to maintain a high PW and seawater quality, which involves proper treatment and accurate monitoring. Monitoring the PW also support protecting the receiving environment when discharging. Accurate information on the amount of oil and particles, sizes, and classification of particles in the injection water (IW) can essentially be used for decision support, reporting, or even advanced control to achieve better operation in the treatment process [6,12]. Therefore, the importance and awareness of accurate online measurements of oil and suspended particles have gained increasing attention within the oil and gas industry [5,6].
The water quality in offshore injection water treatment (IWT) processes is usually assessed in terms of particles’ plugging tendency, also addressed as suspended solids or total suspended solids (TSS) [8,13]. However, the IW quality should be economically viable to achieve and counterbalanced against other solutions, e.g., well-stimulation [12,13]. Currently, offshore IWT processes rely on off-line measurements of TSS concentration following the ISO 11923; thus, in case of a decrease in IW quality, the reaction time is long due to the use of onshore laboratory measurements [14].
Even if at-line execution significantly reduces the reaction time of off-line TSS concentration measurements, it is difficult for operators to investigate where and what is the root cause of the decreased quality when there is no available information related to TSS. Especially as significant water quality decline often occur downstream [8]. Online TSS and oil-in-water (OiW) concentration measurements have not become standardized in the oil and gas industry, despite the long history of measuring particle sizes online [14]. The measurement technique is often not the main reason for the error source in the observed results [15]. Different studies emphasize that the main source of errors often are due to improper installment and calibration [15,16,17].
According to Latif [18] only a few research studies of continuous measurement techniques for online concentration measurements have been published. Latif [18] and Fjords Processing [19] have done a comparison study on different techniques for measuring OiW concentration. However, their focus is on how different variations of parameters such as mixtures, pressures, chemicals, and temperatures can affect the measurements and not on the uncertainty related to the petroleum engineers’ ability to determine an acceptable calibration. Although PWRI is a solution for extending the economic life of oil production, it is beneficial to measure particle sizes to increase IWT processes’ performance. Several different methods for measuring particle sizes exist, each based on several design options from different manufacturers. Besides microscopy, which is based on direct observation, all other techniques are challenged by their property assumptions that only the equivalent diameter of a sphere is measured [12]. The existing morphologies are not taken into account, which complicates the ability to classify particles [12]. Another advantage of microscopy is the manual discrimination of particles captured to evaluate the results. This paper aims to examine two different online microscopy analyzers: Jorin ViPA and Canty InFlow.
Both microscopes are based on the same technique of utilizing a high-resolution video camera to capture images of the particles passing the view cell. Both monitors are based on bright field illumination techniques, presenting a dark image of particles passing with a bright background. The pixel scale of Jorin ViPA is fixed by the manufacturer, and the Canty InFlow is adjusted to a specific pixel scale. A fluorescence-based monitor (Turner TD-4100XDC) is used as a benchmark to evaluate the two online microscopes’ performance to measure OiW concentrations. The fluorescence-based monitor is selected based on a previous work by Hansen et al. [6]. This paper addresses the problems occurring when petroleum engineers must subjectively decide what particles are considered in focus. Furthermore, the calibration procedure will be validated on known solid particle sizes and their ability to measure different OiW concentrations accurately and in real-time. Especially to determine the accurate volumetric concentration in each image is not trivial as the depth of field (DoF) would affect the estimate of particle concentration.

2. Materials and Methods

Two setups were constructed to execute the experiments presented in this paper. The two systems are shown in Figure 1 and Figure 2. The setup shown in Figure 1 is used to calibrate both online microscopes using different known polystyrene particle sizes produced by BS-Partikel [20].
The three different polystyrene particle sizes follow a statistically Gaussian distribution, and the measured means and standard deviations presented in parentheses, are given by the manufacturer:
  • Nominal diameter of 10 µ m : x 3 = 9.8 ( 0.3 ) µ m
  • Nominal diameter of 20 µ m : x 1 = 20.1 ( 0.4 ) µ m
  • Nominal diameter of 40 µ m : x 2 = 40.3 ( 0.9 ) µ m
Figure 2 illustrates a skid-mounted testing platform for validating the two online microscopes’ performance. The platform is equipped with a centrifugal pump (CP), pressure transmitters (Px), flowmeters (Qx), and pneumatic control valves (Vx). The system is configurable to direct the liquid through the sidestream and the mainstream by manipulating the control valves.
By taking advantage of on-line sampling, both microscopes: Jorin Visual Process Analysis (ViPA) B HiFlo (C2) and Canty InFlow VD4912-960 (C3) are installed on a sidestream, making them applicable in most installations regardless of the flow velocity and the pipeline’s dimensions. Although, on-line measurement complicates the sampling procedure as a maldistribution between the run and the branch can occur.
Both microscopes mainly consist of a camera, light source, flow cell, optical lens, and a computer. For C2 a progressive CCD camera (Sony ICX445) with 1292 × 964 pixels captures images at maximum frame rate at full resolution of 31 frames per second (fps) with sensor pixel size of 3.75 × 3.75 µ m . A Comar 10 OA 25 lens is used to output a field of view of 484.50 × 361.50 µ m with a magnification of 10, a numerical aperture of 0.25, and a focal length of 16.9 mm . The manufacturer of C3 retains most of the equipment specifications confidential, however some general specifications of both microscopes are listed in Table 1.
A fluorescence-based monitor (C1) that is sensitive to the aromatic content is installed to measure the OiW concentration. C1 is installed as a benchmark to facilitate the investigation of measuring OiW concentrations by C2 and C3. The accuracy of C1 has been extensively studied by Hansen et al. [6], and will not be evaluated in this paper.
The difficulty of online microscopy analysis is related to the narrow focus area, where only a fraction of the entire flow is directed into the sidestream and through the view cell. A further complication occurs as only a narrow DoF of the passing flow is measured by the microscope; thus, not all TSS that passes the view cell will be observed and are therefore excluded from being measured as shown in Figure 3.
To achieve statistical results, it is necessary to sample a sufficient amount of particles that is constricted by the small DoF. The quantity of captured particles strongly relates to the accuracy of the predicted particle size distribution (PSD). The accuracy of the PSD can be determined within a region of relative error, δ , from a defined confidence level with its represented z-score value, u. The number of sufficient particles, n * , that are required is based on δ and sample standard deviation, s, has been proposed by Masuda and Iinoya [21]. The approach is based upon the assumption that the PSD follows a log-normal distribution.
log ( n * ) = 2 log ( δ ) + log ( ω ) ,
where
ω = u 2 α 2 s 2 ( 2 c 2 s 2 + 1 ) .
α is the non-zero exponential constant that defines the particle distribution, for log-normal distribution α = 2 [22]. c can be calculated as
c = β + α 2 ,
where β is a basis number, for the count basis β = 0 . The number of particles required within a certain range of error changes depending on how the data is scattered. Thus, the number of particles increases as the deviation of the data increases. Simultaneously, the number of particles increases proportionately to the error of magnitude required. On the contrary, the relative error can be estimated based on the number of particles observed.

Experimental Design

This subsection describes the design and objective of all executed experiments on both setups from Figure 1 and Figure 2. The two microscopes were calibrated according to a known particle size from BS-Partikel, followed by a performance validation on known sizes, larger and smaller than the known particle size used for calibration. The two microscopes were installed in series together with the fluorescence-based monitor to observe their performance of measuring OiW concentrations. The following experiments are executed on the setup presented in Figure 1:
  • Experiment I: calibration of both microscopes, by addition of known particle sizes with a mean and standard deviation: x 1 = 20.1 ( 0.4 ) µ m .
  • Experiment II: validation test of the calibration procedure executed in Experiment I, by addition of a known particle size larger and smaller than x 1 : x 2 = 40.3 ( 0.9 ) µ m and x 3 = 9.8 ( 0.3 ) µ m .
The pump speed was fixed at the same value in both Experiment I and II, which generated a flow rate of ∼2 L / min . Most of the presented data in Experiment I are based on data from C2, although the exact same procedure was executed on C3. Both microscopes have three calibration parameters:
  • Threshold value (THV): an 8-bit integer, resulting in a greyscale image with pixel values in the range of 0–255, from black to white for both microscopes.
  • Edge strength/focus rejection value (ESV/FRV): an edge detection method, running a convolution kernel for estimating the gradient at each pixel on the image. The most common method is Sobel, which has been used on C3. Other methods are also available when using C3. C2’s edge detection method is confidential. The ESVs of C2 are in a range of 0–10, where ESV = 10 only includes particles with a large gradient. C3’s FRV ranges from 0–1000, although FRV > 45 did not include any particles in the calibration analysis.
  • Depth of field (DoF): defines the depth of the captured images. DoF normally defines the distance between the closest and farthest particle in the image that appears acceptably sharp. DoF is only necessary to adjust if the user is interested in the sample’s concentration as the DoF influence to field of view volume. To determine the DoF for these microscopes, the DoF is adjusted to present the known concentration based on the accepted measured particles fulfilling THV and ESV/FRV. The DoF is then estimated using
    C = j = 1 k i = 1 n 1 6 π d A i , j 3 A F O V z ¯ k ,
    where C is the known concentration, k is the number of images, n is the number of particles in each jth image, d A i , j is the equivalent area diameter for each ith particle in each j-th image, A F O V is the field of view, and z ¯ is the average DoF. By adjusting z ¯ in (4) of the captured images within a defined time interval, it is possible to adjust the measured concentration from the microscopes to match the known concentration.
Only the first two calibration parameters were analyzed in Experiment I and validated in Experiment II by the addition of the particle sizes: x 2 and x 3 . The DoF value was tuned in Experiment III. The following experiment is executed on the setup presented in Figure 2:
  • Experiment III: a performance evaluation of the microscopes, to measure OiW concentrations in steady-state and real-time measurements, benchmarked according to the known OiW concentration and the measurements obtained from C1. Six nominal OiW concentrations were investigated ( 55 , 100 , 150 , 200 , 250 , 400 ) ppm . The supply tank was filled with 163.59 L of 1 µ m filtered tap water. The solution of oil added are listed in Table 2 for each OiW concentration.
CP was kept at 100% pump speed to ensure that the droplet size distribution does not change over time due to the high shear occurring in CP. V2 was kept at a fixed opening degree 37.5%, and the flow rate through the sidestream was kept constant by manipulating V1.

3. Results

The results are divided into three sections based on Experiment I, II, and III.
Experiment I:
A 2 h recirculation experiment with particle size x 1 dispersed in 1 µ m filtered tap water, was executed with both microscopes and used as calibration data.
Firstly, the THVs were adjusted so that the known particle size x 1 can statistically be represented by the microscopes. A high ESV was selected only to determine the THVs on captured particles in focus. Determining the THV before the ESV is essential as the edge detection method depends on the THV.
Furthermore, for comparison of both microscopes’ performance, particle sizes were based on equivalent area diameter during post-processing of the data. Note that choosing another method of determining the particle diameter will most likely shift the distribution of measured sizes. The measurement in Figure 4 presents the mean results of the first 100 images with particles captured at maximum ESV. The THV was then toggled in a range of 47–81, where at THV 47 and THV 81 , particles no longer fulfill the ESV’s criteria and are no longer counted. The Gaussian distribution in Figure 4 represent the information analyzed by BS-Partikel: μ = 20.1 µ m and σ = 0.4 µ m .
Based on the results in Figure 4, a THV = 62 for C2 was selected. For C3, a THV = 182 was selected. The analysis results with a THV = 62 and maximum ESV = 10 are shown in Figure 5. A probability density function (PDF) is included on the graphs in Figure 5, to visualize the performance of estimating the particle sizes.
Selecting the ESV is more subjective as it is a perception of what appears to be acceptably sharp. Figure 6 presents eleven different particles captured by C2 with a THV = 62 , each with different acceptable ESV between 0–10. Associated information of each particle, presented in Figure 6, is shown in Table 3.
Furthermore, Figure 7 and Table 4 show an artifact of selecting a weak ESV. Even though an ESV of one outputs a measured size within 1 σ from the μ of x 1 , selecting an ESV = 2 , the gradient of the particle is no longer accepted and is considered out-of-focus. Although a smaller, but stronger gradient within the particle, fulfills the requirement and is included as a particle.
Selecting a too high ESV will reduce the counted number of accepted particles in the analysis, or even worse, not represent other types of particles in the process that has a weaker gradient by nature. A further validation of selecting a proper ESV or FRV was executed by analyzing the corresponding results of the first three particles in Table 3 that are within 2 σ from μ of x 1 (ESV: 5, 6, and 7). The results are shown in Figure 8. The same evaluation procedure was executed on C3.
Increasing the ESV from 5 to 7 counted 24% fewer particles in Experiment I. Another phenomenon occurs at ESV = 5 , as the size histogram is skewed left due to the acceptable edge is found closer to the center of the particles as the peripheries are less in focus. Selecting between an ESV = 6 or ESV = 7 is a tradeoff between obtaining less countable particles and reducing the phenomena of underestimating the size of particles that are less in focus. ESV = 6 was chosen for C2, and FRV = 25 was chosen for C3. The calibration results of C3 are shown in Figure 9. The number of particles captured by C3 was ∼5 times larger than the amount captured by C2. This is primarily a result of higher image resolution and larger pixel size than C2 as shown in Table 1.
Experiment II:
A 2 h validation test was executed by adding two additional particle sizes: x 2 and x 3 , along with x 1 . The raw data from C2 and C3, without any classification, are shown in Figure 10.
For each known particle size, 1 ml aqueous surfactant solution containing the polystyrene particles was added. Both microscopes observed a high number of x 3 in the validation test, a result of containing more particles per volume than the others. By truncating the sample for each known particle size with a fixed range of ± 4 µ m from the known μ of x 1 , x 2 , and x 3 , the statistical information can be obtained as shown in Table 5.
Experiment III:
A 9.5 h experiment was carried out with the addition of oil after at least 30 min between each OiW concentration of observation.
Figure 11 shows flow rates of the mainstream (Q1) and sidestream (Q2), and the opening degree of V1. The sidestream was sufficiently kept constant at 3 L / min during the entire experiment by manipulation of V1. Figure 12 shows the time series of measured OiW concentration from C1. The vertical line marks a truncated time series of ∼3.5 h that was left out. The DoF value in C2 and C3 was at that period adjusted to match a OiW concentration of 55 ppm during the first 30 min after the injection of oil into the process by use of (4) and to classify oil droplets. The DoF value in C2 and C3 was estimated to be 43.14 µ m and 94.29 µ m , respectively.
During the last known OiW concentration of 400 ppm , the mixed concentration can no longer be maintained in the setup according to the measurement of C1. Figure 13 presents an error bar of each OiW concentration measured in a duration of 30 min for each online monitor. Each error bar represents the mean, minimum, and maximum OiW concentration measured, where C1 was selected to have a sampling frequency of 0.1 Hz , and C2 and C3 outputting an averaged concentration measurement every minute. Table 6 shows the associated results related to Figure 13.
Figure 14 and Figure 15 present the PSD histogram of the measured OiW concentrations at 55 ppm and 400 ppm for each microscope. Furthermore, the Figure 14 and Figure 15 presents the mean of the log-normal distribution, μ 0 , the standard deviation of the log-normal distribution, σ 0 , and the number of counted particles, X N .
Both OiW concentrations follow a log-normal size distribution measured by both microscopes, with only small changes in the distribution with respect to the concentration. Following the proposed calculation by Masuda and Iinoya [21] in (1)–(3), for obtaining a sufficient statistical representation within 95% confidence level, δ of the PSD can be calculated as shown in Table 7.
A trailing moving average window of 1 min was selected, for measuring the OiW concentration real-time by C2 and C3. Figure 16 and Figure 17 show the results of real-time measurements at the known OiW concentration of 55 ppm from C2 and C3, respectively, together with real-time measurements from C1. The calculated OiW concentration, based on post-processed equivalent volume measurements, are presented for each captured image.
The results presented in Figure 16 and Figure 17 have also been executed at the known OiW concentration of 400 ppm and presented in Figure 18 and Figure 19.

4. Discussion

The process of calibrating the two online microscopes is highly based on the perception of what is considered in focus. To ensure consistent results when calibrating the three main calibration parameters: THV, ESV/FRV, and DoF, the calibrating procedure must be well documented to reduce the difference between each calibration execution. Figure 4 and Figure 8 presented one solution for selecting the THV and ESV/FRV that outputs an estimation that matches the Gaussian distribution of the known particle sizes based on the information from the manufacturer of the produced particles. The skewness observed in Figure 8 when calibrating the ESV could result from selecting a slightly too low THV; thus, a lower ESV may have yielded better results. Although, based on the results shown in Figure 10 the calibration procedure presented in Experiment I revealed only a small deviation from the known PSDs. To further improve online microscopy analyzers, auto-calibration or human-computer interaction (HCI) of the calibration procedure can lower the subjective perception of what is acceptably sharp.
The steady-state values of measuring OiW concentration in Figure 13 showed promising results in relation to the known OiW concentrations. The measurement of C1 confirmed that C2 and C3 are able to measure the mean OiW concentrations by estimating the DoF using (4). The DoF for C3 was close to the theoretical estimation of DoF from the manufacturer seen in Table 1. The DoF for C2 was 63 % lower than the theoretical estimation of DoF given by the manufacturer.
Moreover, C1 also confirmed that the setup was able to keep a well-mixed OiW concentration, apart from high OiW concentrations where a declining tendency of measured concentration was observed. The observed declining tendency may occur due to accumulation within the setup (dead volumes), natural separation in the supply tank, or pipelines and equipment becoming oil-wet.
Regardless of these sources of error, both microscopes were able to measure the mean OiW concentration of 30 min in close relation to the known concentration. A high deviation was observed when outputting OiW concentration every minute in steady-state, especially considering the steady concentration measured by C1 as shown in Figure 12. Figure 20 shows the results of extending the averaging time duration of both microscopes in steady-state. The results in Figure 20, show that increasing the averaging time to 3 min reduces the fluctuation of measured OiW concentration greatly. Meanwhile, the negative impact is that the time resolution of dynamics will be reduced accordingly.
Another solution is to measure the OiW concentration in real-time by incorporating a trailing moving average window. The size of the moving average window is highly based on the plant’s dynamics. Based on the results in Figure 16, Figure 17, Figure 18 and Figure 19, both microscopes are able to measure the the OiW concentration and thus they have the potential to track the transient behavior when it occurs in an IWT process. The fluctuating real-time results obtained with both microscopes, using a trailing moving average window of 1 min , can be questioned to be sufficient to provide qualitative feedback to an operator. However, by extending the moving average window to 3 min , the real-time measurements of OiW concentration from both microscopes are naturally more consistent, as shown in Figure 21. Extending the moving average window is only valuable if the dynamic of the plant is not faster than the moving average window. The higher resolution of C3 caused the microscope to capture twice as many particles compared to C2, thus when using the moving average window, C3 yields a more stable measurement.
One important procedure that has not been addressed in this paper is the classification within both microscopy analyzers. Oil droplets in Experiment III have been classified and withdrawn from the rest of the observed particles. The classification process has not been addressed as the procedure are very different in each microscope. C2 entrust the operator in selecting the classification range of each measured variable. C3 uses machine learning by letting an operator train the classification process by manually defining the captured particles. The different results of PSDs shown in Figure 14 and Figure 15, could be a result of classifying differently. The results obtained by C3 are not representable around 13 µ m if the PSD truly follows a log-normal distribution. Another consequence to the bimodal PSD obtained by C3 could be the installment right after the narrow view cell of C2, which may change the PSD of oil droplets by breakups or coalescence. Another statical consideration was addressed when outputting a PSD, as the PSD should be trustworthy if an operator should draw any conclusion based on the presented PSD. Masuda and Iinoya [21] presented an estimation for predicting the required number of particles to be counted for a certain accuracy. In this paper, δ of the PSDs shown in Figure 14 and Figure 15 were calculated based on the number of particles obtained within 30 min at two different OiW concentrations ( 55 ppm and 400 ppm ) for both microscopes. Based on the results in Table 7, a δ < 5 % was observed within a 95% confidence interval of all the presentable PSDs. If the δ < 5 % is not enough more particles must be counted, i.e., at δ = 1 % roughly 100,000–130,000 for C3 and 90,000–100,000 for C2 must be counted based on the measured OiW concentrations, respectively. Seen from another point of view, if a PSD update every 30 min is too slow, the allowed δ in the presented PSD must be defined to determine the minimum number of particles required in order to achieve a defined level of accuracy.

5. Conclusions

The paper presented an evaluation of two different online microscopy analyzers that utilize a high-resolution video camera for capturing images of the particles passing their view cell. The three main calibration parameters: threshold, edge strength/focus rejection, and DoF, were validated and discussed in order to address the difficulty in determining which particles are considered to be in focus. The main contribution of this paper includes experiments of how the calibration affects the microscopy outputs, which is used to minimize the uncertainty by selecting calibration parameters. The procedure for selecting the three different calibration parameters must be well documented by the manufacturer to increase reproducibility when installed in an offshore oil and gas process. It was discussed whether the manufacturer could accommodate to reduce the use of perception when calibrating the instrument by integrating auto-calibration or human-computer interaction to minimize the uncertainties related to calibrating the microscope. Both microscopes were able to discriminate the particle size distribution of three known particle sizes with high precision, indicating a successful calibration procedure for accurately measuring the particle sizes.
A fluorescence-based monitor was installed as a benchmark to evaluate both online microscopes’ performance to measure oil-in-water (OiW) concentrations, both for steady-state measurements and real-time purposes. The fluorescence-based monitor was selected due to previous work by Hansen et al. [6] showing promising results for measuring OiW concentrations. Both measurements from the microscopy analyzers were reasonable compared to the known OiW concentration and the measured concentration from the fluorescence-based monitor. Real-time measurements from both microscopy analyzers were achievable by post-processing the data captured by the microscopes and applying a trailing moving average window of 1 min . Although selecting the size of the moving average window should be based on the plant’s dynamics.
Lastly, a statical consideration was addressed to determine the minimum number of particles required in order to achieve a defined level of accuracy within a defined confidence interval when outputting the particle size distribution to an operator.

Author Contributions

Conceptualization, D.S.H. and S.J.; methodology, D.S.H.; software, D.S.H. and S.J.; validation, D.S.H.; formal analysis, D.S.H.; investigation, D.S.H. and S.J.; resources, Z.Y.; data curation, D.S.H. and M.V.B.; writing—original draft preparation, D.S.H.; writing—review and editing, D.S.H., S.J., M.V.B., and Z.Y.; visualization, D.S.H.; supervision, Z.Y.; project administration, Z.Y.; funding acquisition, Z.Y. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the DHRTC projects “Injection Water Quality and Control” (Proj-No.: 878040).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.

Acknowledgments

The authors would like to thank the support from the DHRTC and thank the colleague, L. Hansen from AAU, for many valuable discussions and supports. A special thanks go to colleagues at Total Denmark A/S in Esbjerg for valuable supports.

Conflicts of Interest

The authors declare no conflict of interest.

Abbreviations

The following abbreviations are used in this manuscript:
C 1 Turner TD-4100XDC
C 2 Jorin ViPA B HiFlo
C 3 Canty InFlow VD4912-960
CPCentrifugal pump
DoFDepth of field
ESVEdge strength
fpsframes per second
FRVFocus rejection value
HCIHuman-computer interaction
IWInjection water
IWTInjection water treatment
TSSTtotal suspended solids
OiWOil-in-water
P x Pressure transmitters
PDFProbability density function
PSDParticle size distribution
PWProduced water
PWRIProduced water re-injection
Q x Flowmeters
THVThreshold value
V x Pneumatic control valves

References

  1. IPIECA; UNDP; IFC. Mapping the Oil and Gas Industry To the Sustainable Development Goals: An Atlas; Technical Report; The Global Oil and Gas Industry Association for Environmental and Social Issues: London, UK, 2017. [Google Scholar]
  2. EIA. International Energy Outlook 2019: World Energy Projection System Plus. Available online: www.eia.gov/aeo (accessed on 19 November 2020).
  3. Miløstyrelsen. Generel Tilladelse for Total E&P Danmark A/S (TOTAL) til Anvendelse, Udledning og Anden Bortskaffelse af Stoffer og Materialer, Herunder Olie og Kemikalier i Produktions- og Injektionsvand fra Produktionsenhederne Halfdan, Dan, Tyra og Gorm for Perioden 1 Januar 2019–31 December 2020; Technical Report; Total E&P Danmark A/S: Copenhagen, Denmark, 2018. [Google Scholar]
  4. OSPAR Commission. Overview Assessment of the Implementation of OSPAR Recommendation 2001/1 for the Management of Produced Water from Offshore Installations Offshore Industry Series; Technical Report; OSPAR Commission: London, UK, 2010. [Google Scholar]
  5. Blanchard, E. Oil in Water Monitoring is a Key to Production Separation. Offshore 2013, 73, 104–105. [Google Scholar]
  6. Severin Hansen, D.; Jespersen, S.; Bram, M.V.; Yang, Z. Uncertainty Analysis of Fluorescence-Based Oil-In-Water Monitors for Oil and Gas Produced Water. Sensors 2020, 20, 36. [Google Scholar] [CrossRef] [PubMed]
  7. Ostroff, A. Injection Water Problems Identified By Laboratory Analysis. In Proceedings of the Middle East Technical Conference and Exhibition, Manama, Bahrain, 9–12 March 1981; pp. 523–526. [Google Scholar] [CrossRef]
  8. Patton, C.C. Water Quality Control and Its Importance in Waterflooding Operations. J. Pet. Technol. 1988, 40, 1123–1126. [Google Scholar] [CrossRef]
  9. Bennion, D.; Bennion, D.; Thomas, F.; Bietz, R. Injection Water Quality—A Key Factor to Successful Waterflooding. In Proceedings of the 45th Annual Technical Meeting, Calgary, ALB, Canada, 12–15 June 1994; Volume 37, pp. 53–62. [Google Scholar] [CrossRef]
  10. Ogden, B.L. Water Technology: Understanding, Interpreting and Utilizing Water Analysis Data; Southwestern Petroleum Short Course Conference 2008; Southwestern Petroleum Short Course: Lubbock, TX, USA, 2008; p. 12. [Google Scholar]
  11. Larsen, I. Denmark’s Oil and Gas Production—And Subsoil Use 09; Technical Report; Energistyrelsen: Copenhagen, Denmark, 2010. [Google Scholar]
  12. Hansen, D.S.; Bram, M.V.; Lauridsen, S.M.Ø; Yang, Z. Online Quality Measurements of Total Suspended Solids for Offshore Reinjection: A Review Study. Energies 2021, 14, 967. [Google Scholar] [CrossRef]
  13. Bansal, K.; Caudle, D. A New Approach for Injection Water Quality. In Proceedings of the SPE Annual Technical Conference and Exhibition, Washington, DC, USA, 4–7 October 1992; pp. 383–396. [Google Scholar] [CrossRef]
  14. Yang, M. Measurement of Oil in Produced Water. In Produced Water, 1st ed.; Lee, K., Neff, J., Eds.; Springer: New York, NY, USA, 2011; Chapter 2; pp. 57–88. [Google Scholar] [CrossRef]
  15. Merkus, H.G. Particle Size Measurements Fundamentals, Practice, Quality, 1st ed.; Particle Technology Series; Springer: Dordrecht, The Netherlands, 2009; Volume 17, p. 534. [Google Scholar] [CrossRef]
  16. Allen, T. Powder Sampling and Particle Size Determination, 1st ed.; Elsevier: Amsterdam, The Netherlands, 2003; p. 682. [Google Scholar] [CrossRef]
  17. Leschonski, K. Particle Characterization, Present State and possible Future Trends. Part. Part. Syst. Charact. 1986, 3, 99–103. [Google Scholar] [CrossRef]
  18. Latif, Z. Towards Continuous Online Oil-in-Water Measurement for Regulatory Purposes. Ph.D. Thesis, Coventry University, Coventry, UK, 2020. [Google Scholar]
  19. Zhang, J. RPSEA Subsea Produced Water Discharge Sensor Lab Test Results and Recommendations Final Report; Technical Report 3; Cleariew Subsea LLC.: Houston, TX, USA, 2016. [Google Scholar]
  20. Eslahian, K. Collection of Particle Size Standard and Count Control; Certificates; BS-Partikel: Mainz, Germany, 2020. [Google Scholar]
  21. Masuda, H.; Iinoya, K. Theoretical Study on the Scatter of Experimental Data due to the Particle-Size-Distribution. J. Chem. Eng. Jpn. 1971, 4, 60–66. [Google Scholar] [CrossRef] [Green Version]
  22. Jillavenkatesa, A.; Dapkunas, S.J.; Lum, L.S.H. Particle Size Characterization: Recommended Practice Guide, 1st ed.; NIST Recommended Practice Guide; National Institute of Standards and Technology: Gaithersburg, MD, USA, 2001; p. 165. [Google Scholar]
Figure 1. A recirculating test setup for measuring known particle sizes to calibrate both online microscopes.
Figure 1. A recirculating test setup for measuring known particle sizes to calibrate both online microscopes.
Sensors 21 03192 g001
Figure 2. Skid mounted testing platform for validating quality monitors’ performance on a sidestream installation during different flow regimes.
Figure 2. Skid mounted testing platform for validating quality monitors’ performance on a sidestream installation during different flow regimes.
Sensors 21 03192 g002
Figure 3. The depth of field of the measured volume within the view cell.
Figure 3. The depth of field of the measured volume within the view cell.
Sensors 21 03192 g003
Figure 4. Selected particles, all with maximum ESV on C2. The error bars represent the mean, minimum, and maximum particle sizes at different THVs. The Gaussian distribution is calculated based on information from BS-Partikel: µ = 20.1 µ m and = 0.4 µ m .
Figure 4. Selected particles, all with maximum ESV on C2. The error bars represent the mean, minimum, and maximum particle sizes at different THVs. The Gaussian distribution is calculated based on information from BS-Partikel: µ = 20.1 µ m and = 0.4 µ m .
Sensors 21 03192 g004
Figure 5. Included particles with a THV = 62 and ESV = 10, measured by C2.
Figure 5. Included particles with a THV = 62 and ESV = 10, measured by C2.
Sensors 21 03192 g005
Figure 6. Eleven different particles were captured by C2 with a THV = 62, each with different acceptable ESV between 0–10. Associated results based on ESV are shown in Table 3.
Figure 6. Eleven different particles were captured by C2 with a THV = 62, each with different acceptable ESV between 0–10. Associated results based on ESV are shown in Table 3.
Sensors 21 03192 g006
Figure 7. Three images of the same particle with a THV = 62, but with different ESVs: (a) ESV ≤ 1; (b) ESV = 2; (c) ESV ≥ 3. Associated results are shown in Table 4.
Figure 7. Three images of the same particle with a THV = 62, but with different ESVs: (a) ESV ≤ 1; (b) ESV = 2; (c) ESV ≥ 3. Associated results are shown in Table 4.
Sensors 21 03192 g007
Figure 8. Included particles with a THV = 62 and three different ESVs: 5, 6, and 7, respectively, measured by C2.
Figure 8. Included particles with a THV = 62 and three different ESVs: 5, 6, and 7, respectively, measured by C2.
Sensors 21 03192 g008
Figure 9. Included particles with a THV = 182 and a FRV = 25, measured by C3.
Figure 9. Included particles with a THV = 182 and a FRV = 25, measured by C3.
Sensors 21 03192 g009
Figure 10. Three known particle sizes were measured simultaneously by C2 and C3. The mean and sample standard deviation are listed in Table 5.
Figure 10. Three known particle sizes were measured simultaneously by C2 and C3. The mean and sample standard deviation are listed in Table 5.
Sensors 21 03192 g010
Figure 11. Flow rates from Q1, Q2, and the opening the degree of V1 to maintain a constant flow rate through the sidestream of the entire ∼9.5 h experiment.
Figure 11. Flow rates from Q1, Q2, and the opening the degree of V1 to maintain a constant flow rate through the sidestream of the entire ∼9.5 h experiment.
Sensors 21 03192 g011
Figure 12. Time series of measured OiW concentration measured by C1 executed on the setup shown in Figure 2. The vertical line marks the truncated time series of ∼3.5 h.
Figure 12. Time series of measured OiW concentration measured by C1 executed on the setup shown in Figure 2. The vertical line marks the truncated time series of ∼3.5 h.
Sensors 21 03192 g012
Figure 13. Error bars representing the mean, minimum, and maximum OiW concentrations based on the average concentration obtained each minute from C2 and C3, together with error bars based on measurement every 10 s from C1.
Figure 13. Error bars representing the mean, minimum, and maximum OiW concentrations based on the average concentration obtained each minute from C2 and C3, together with error bars based on measurement every 10 s from C1.
Sensors 21 03192 g013
Figure 14. Droplet size distribution obtained with C2 based on the measured OiW concentrations at the known concentration of 55 ppm and 400 ppm, respectively.
Figure 14. Droplet size distribution obtained with C2 based on the measured OiW concentrations at the known concentration of 55 ppm and 400 ppm, respectively.
Sensors 21 03192 g014
Figure 15. Droplet size distribution obtained with C3 based on the measured OiW concentrations at the known concentration of 55 ppm and 400 ppm, respectively.
Figure 15. Droplet size distribution obtained with C3 based on the measured OiW concentrations at the known concentration of 55 ppm and 400 ppm, respectively.
Sensors 21 03192 g015
Figure 16. Real-time OiW concentration measurements calculated using a trailing moving average window of 1 min of the post-processed oil volume ratio in each image captured by C2. The top graph shows the real-time measurement measured by C1 and C2. A 95% confidence interval of the averaging window is shadowed behind the signal. Bottom graph shows the volume concentration of each frame captured.
Figure 16. Real-time OiW concentration measurements calculated using a trailing moving average window of 1 min of the post-processed oil volume ratio in each image captured by C2. The top graph shows the real-time measurement measured by C1 and C2. A 95% confidence interval of the averaging window is shadowed behind the signal. Bottom graph shows the volume concentration of each frame captured.
Sensors 21 03192 g016
Figure 17. Real-time OiW concentration measurements calculated using a trailing moving average window of 1 min of the post-processed oil volume ratio in each image captured by C3. Top graph shows the real-time measurement measured by C1 and C3. A 95% confidence interval of the averaging window is shadowed behind the signal. Bottom graph shows the volume concentration of each frame captured.
Figure 17. Real-time OiW concentration measurements calculated using a trailing moving average window of 1 min of the post-processed oil volume ratio in each image captured by C3. Top graph shows the real-time measurement measured by C1 and C3. A 95% confidence interval of the averaging window is shadowed behind the signal. Bottom graph shows the volume concentration of each frame captured.
Sensors 21 03192 g017
Figure 18. Same procedure as in Figure 16 for real-time OiW concentration measurement based on captured images from C2 at another measurement range of ∼400 ppm.
Figure 18. Same procedure as in Figure 16 for real-time OiW concentration measurement based on captured images from C2 at another measurement range of ∼400 ppm.
Sensors 21 03192 g018
Figure 19. Same procedure as in Figure 17 for real-time OiW concentration measurement based on captured images from C3 at another measurement range of ∼400 ppm.
Figure 19. Same procedure as in Figure 17 for real-time OiW concentration measurement based on captured images from C3 at another measurement range of ∼400 ppm.
Sensors 21 03192 g019
Figure 20. Error bars representing the minimum and maximum OiW concentrations based on different time durations of averaging the OiW concentrations obtained by C2 and C3, respectively. The right error bar illustration shows different time durations according to the greyscale value. The red indicator shows the grand mean of the OiW concentrations obtained within the execution time of 30 min.
Figure 20. Error bars representing the minimum and maximum OiW concentrations based on different time durations of averaging the OiW concentrations obtained by C2 and C3, respectively. The right error bar illustration shows different time durations according to the greyscale value. The red indicator shows the grand mean of the OiW concentrations obtained within the execution time of 30 min.
Sensors 21 03192 g020
Figure 21. Real-time OiW concentration measurements calculated extending the trailing moving average window to 3 min for both C2 and C3.
Figure 21. Real-time OiW concentration measurements calculated extending the trailing moving average window to 3 min for both C2 and C3.
Sensors 21 03192 g021
Table 1. Execution specifications of both microscopes.
Table 1. Execution specifications of both microscopes.
Jorin ViPACanty InFlow
Pixel scale 0.375 µ m / pixel 0.513 µ m / pixel
Pixel dimensions 1292 × 964 pixels 1920 × 1200 pixels
Field of view 484.50 × 361.50 µ m 984.96 × 615.60 µ m
Depth of field *116.50 µ m 93.03 µ m
Frame rate∼30 fps ∼30 fps
Inlet/outlet ports1/4 1/2
Flow velocity *0.03–2.1 m / s 0.25–2.74 m / s
Flow rate *0.05–4 L / min 1.9–20.8 L / min
* Information provided by the manufacturers
Table 2. A list of each wanted OiW concentration by addition of specific amount of oil. V i represent the total amount of oil added, where V i V i 1 represents the amount that should be added to reach the next OiW concentration.
Table 2. A list of each wanted OiW concentration by addition of specific amount of oil. V i represent the total amount of oil added, where V i V i 1 represents the amount that should be added to reach the next OiW concentration.
Wanted [ ppm ] V i [ mL ] V i V i 1 [ mL ]
000
559.009.00
10016.367.36
15024.548.18
20032.728.18
25040.918.18
40065.4624.56
Table 3. Associated results to Figure 6, based on different ESVs.
Table 3. Associated results to Figure 6, based on different ESVs.
Edge StrengthSizeAspect RatioShape Factor
06.530.690.36
116.860.920.86
215.590.920.87
317.180.870.89
417.270.960.97
519.820.990.94
619.560.970.96
719.430.990.97
820.500.960.93
919.560.970.96
1019.970.970.98
Table 4. Associated results to Figure 7.
Table 4. Associated results to Figure 7.
Edge StrengthSizeAspect RatioShape Factor
≤119.700.940.80
21.810.710.96
≥3
Table 5. Associated results to Figure 10.
Table 5. Associated results to Figure 10.
Known μ ( σ ) m ] μ estimat . m ] σ estimat . m ]
9.8(0.3)9.50.6
C320.1(0.4)20.00.6
40.3(0.9)40.90.5
9.8(0.3)9.60.6
C220.1(0.4)20.31.0
40.3(0.9)41.40.6
Table 6. Associated results to Figure 13.
Table 6. Associated results to Figure 13.
Known Conc.55 ppm100 ppm150 ppm200 ppm250 ppm400 ppm
minimum:56.294.7146.1193.9237.6379.0
C1 μ :60.9100.1151.8200.7244.6390.0
maximum:67.7105.4159.5207.7255.8398.4
minimum:31.671.4118.0153.5183.7331.9
C2 μ :49.2106.1168.9215.7263.1393.6
maximum:75.1177.8230.7309.3373.4478.0
minimum:44.984.6136.9172.4231.2362.6
C3 μ :55.1104.0162.4210.0259.3410.6
maximum:72.3136.1189.2239.6289.4451.7
Table 7. Associated results to Figure 14 and Figure 15, calculating the , related to each PSD of each OiW concentration measured with C2 and C3.
Table 7. Associated results to Figure 14 and Figure 15, calculating the , related to each PSD of each OiW concentration measured with C2 and C3.
Known OiW Conc.55 ppm400 ppm
C2 μ 0 :11.713.6
σ 0 :7.38.9
X N :673327154
δ 3.5%1.9%
C3 μ 0 :24.329.0
σ 0 :16.221.4
X N :1106845963
δ :3.0%1.6%
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Hansen, D.S.; Jespersen, S.; Bram, M.V.; Yang, Z. Offshore Online Measurements of Total Suspended Solids Using Microscopy Analyzers. Sensors 2021, 21, 3192. https://doi.org/10.3390/s21093192

AMA Style

Hansen DS, Jespersen S, Bram MV, Yang Z. Offshore Online Measurements of Total Suspended Solids Using Microscopy Analyzers. Sensors. 2021; 21(9):3192. https://doi.org/10.3390/s21093192

Chicago/Turabian Style

Hansen, Dennis Severin, Stefan Jespersen, Mads Valentin Bram, and Zhenyu Yang. 2021. "Offshore Online Measurements of Total Suspended Solids Using Microscopy Analyzers" Sensors 21, no. 9: 3192. https://doi.org/10.3390/s21093192

APA Style

Hansen, D. S., Jespersen, S., Bram, M. V., & Yang, Z. (2021). Offshore Online Measurements of Total Suspended Solids Using Microscopy Analyzers. Sensors, 21(9), 3192. https://doi.org/10.3390/s21093192

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop