Fast Multi-Distance Time-Domain NIRS and DCS System for Clinical Applications
Abstract
:1. Introduction
2. Materials and Methods
2.1. Device Description
2.2. Characterization Measurements
2.2.1. BIP Protocol
- The relevant parameters of the source component: illuminated area on the sample surface, maximum power delivered to the sample.
- The responsivity of the detection system, defined as the ratio between the detected photon count rate for a given input radiance L and the radiance itself:
- Differential nonlinearity (DNL) of the timing electronics is defined as the nonuniformity of the time channel width of the TCSPC board. To assess the DNL, a 5-min measurement with 500 ms integration time was taken in response to a continuous light signal. The histograms for each repetition were then summed to obtain the DNL with a good signal-to-noise ratio.
- Instrument Response Function (IRF), both as a whole (shape) and through its relevant parameters (signal-dependent background, center of mass and Full-Width-at-Half-Maximum (FWHM)). The IRF for each source-detector pair was measured by facing the injection and detection fibers, with a thin layer of diffusive material and ND filters in between. The signal-dependent background was characterized by calculating the afterpulsing ratio RAP:
2.2.2. MEDPHOT Protocol
- Linearity and accuracy: The retrieved optical parameters for all phantoms were compared to the ones measured on the same set of phantoms by a state-of-the-art laboratory TD-NIRS system by the same manufacturer. The goal of this measurement was to assess the ability of the TD NIRS module to follow variations without distortions (linearity) and to retrieve accurate values of the optical parameters (accuracy). Ten repeated measurements were taken on each phantom, with an integration time of 500 ms per wavelength.
- Noise: The B2 phantom (µa ≅ 0.07 cm−1, µs′ ≅ 10 cm−1) was measured, repeatedly varying the photon counts to assess their effect on the retrieved DTOFs. For each measurement, the integration time was set so as to meet the target photon count, and 10 repeated measurements were taken. Only the detection channel at 2.5 cm source-detector separation was used. The coefficient of variation (CV) of the optical parameters was then calculated as the ratio between the standard deviation and the mean value over the 10 repetitions.
- Stability: A 14 h measurement with an integration time of 1 s was carried out on a liquid phantom to assess the behavior of the TD NIRS module over time. The phantom was composed of distilled water, Intralipid and black India ink, and was crafted to mimic optical properties typical of biological tissue (µa ≅ 0.1 cm−1, µs′ ≅ 10 cm−1) [28]. The extracted quantities were the optical parameters and the photon counts of the DTOFs.
- Reproducibility: Over the course of several days, several measurements were carried out on the same liquid phantom to assess the ability of the device to reproduce the same results in the same experimental conditions. The phantom used was a non-degradable water-based solution of polydisperse microparticles (HemoPhotonics S. L., Barcelona, Spain). Each measurement consisted of 30 repetitions with an integration time of 1 s, and the optical parameters obtained on different days were compared.
2.2.3. nEUROPt Protocol
- Depth sensitivity: A solid phantom with a movable absorbing inclusion was employed. The source and detection fibers were placed at a 2.5 cm distance so that the inclusion was at the midpoint between them. The inclusion depth (z coordinate) was then varied from 0 to 30 mm in steps of 2 mm, and for each depth, 10 repetitions were acquired with an integration time of 500 ms per wavelength. The resulting DTOFs were divided into time gates 500 ps long, and, for each gate k and depth z, the contrast Ck was then calculated as follows:
- Longitudinal resolution: the same phantom used in the previous measurement was employed, this time using both source-detector separations, 1.5 cm and 2.5 cm, respectively. The depth of the inclusion was fixed at 15 mm and its position along the source-detector axis (x coordinate) was varied from -40 mm to 40 mm in steps of 1 mm, x = 0 being the midpoint between source and short-distance detection. For each position, 10 repetitions were acquired with an integration time of 500 ms per wavelength, and the corresponding contrast Ck(x) was calculated using Equation (4) with the x coordinate instead of the z coordinate.
- Depth selectivity: The capability of the device to separate absorption changes occurring in shallow vs. deep compartments was evaluated using bilayer liquid phantoms made of distilled water, Intralipid and black India ink. The thickness of the upper layer was fixed at 10 mm. First, a measurement was taken with the upper and lower compartments containing the same phantom (nominal optical parameters µa0 = 0.15 cm−1, µs0′ = 10 cm−1). Then, for a given absorption change Δµa, the absorption coefficient of the upper layer was increased by Δµa while keeping that of the lower layer fixed to µa0. The contrast for the absorption change in the upper layer Ck,UP(Δµa) was calculated as follows:
2.2.4. DCS Module Characterization
- Measurements at variable viscosity levels: the value of the Brownian diffusion coefficient DB for a phantom depends on the phantom properties according to the Einstein relation [29]:
- Stability: the same phantom and procedure used for evaluating the stability of the TD NIRS module were employed to assess the stability of the DCS component. The quantities measured were the Brownian diffusion coefficient DB and the coherence parameter β over the 14-h measurement.
- Reproducibility: the same phantom and procedure used in assessing the reproducibility of the TD NIRS module were employed to evaluate the reproducibility of the DCS component. The quantities measured were the Brownian diffusion coefficient DB and the coherence parameter β, calculating their CV over the days.
- DCS noise level: the noise level of the DCS module was assessed following the procedure described by Cortese et al. [30]. Briefly, a liquid phantom (µa ≅ 0.05 cm−1, µs′ ≅ 7 cm−1) was measured several times at different count rates. Each measurement consisted of 100 iterations of 1 s integration time, using all four detectors at a single source-detector separation ρ = 2.5 cm. Data were analyzed as described in [30] to obtain the dependence of the CV of DB on measurement duration, number of detection channels and count rate.
- To test the consistency of the instrument, a liquid phantom made of distilled water, Intralipid and black India ink (µa ≅ 0.1 cm−1, µs′ ≅ 10 cm−1) was measured five consecutive times, removing and repositioning the probe in between measurements. Each measurement consisted of 30 repetitions at 1 s integration time, and the inter-measurement CV of the following parameters was evaluated: µa, µs′, DB, and β.
2.3. In-Vivo Measurements
2.3.1. Stepped Vascular Occlusion
2.3.2. Pulsatility Measurements
3. Results
3.1. Characterization Measurements Results
3.1.1. BIP Protocol Results
3.1.2. MEDPHOT Protocol Results
3.1.3. nEUROPt Protocol Results
3.1.4. DCS Module Characterization Results
3.2. In-Vivo Measurements Results
3.2.1. Stepped Vascular Occlusion Results
3.2.2. Pulsatility Measurements Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Durduran, T.; Choe, R.; Baker, W.B.; Yodh, A.G. Diffuse Optics for Tissue Monitoring and Tomography. Rep. Prog. Phys. 2010, 73, 076701. [Google Scholar] [CrossRef] [PubMed]
- Grassi, B.; Quaresima, V. Near-Infrared Spectroscopy and Skeletal Muscle Oxidative Function in Vivo in Health and Disease: A Review from an Exercise Physiology Perspective. J. Biomed. Opt. 2016, 21, 091313. [Google Scholar] [CrossRef] [PubMed]
- Fantini, S.; Sassaroli, A. Near-Infrared Optical Mammography for Breast Cancer Detection with Intrinsic Contrast. Ann. Biomed. Eng. 2012, 40, 398–407. [Google Scholar] [CrossRef] [PubMed]
- Boas, D.A.; Yodh, A.G. Spatially Varying Dynamical Properties of Turbid Media Probed with Diffusing Temporal Light Correlation. J. Opt. Soc. Am. A 1997, 14, 192–215. [Google Scholar] [CrossRef]
- Amendola, C.; Buttafava, M.; Carteano, T.; Contini, L.; Cortese, L.; Durduran, T.; Frabasile, L.; Guadagno, C.N.; Karadeniz, U.; Lacerenza, M.; et al. Assessment of Power Spectral Density of Microvascular Hemodynamics in Skeletal Muscles at Very Low and Low-Frequency via near-Infrared Diffuse Optical Spectroscopies. Biomed. Opt. Express 2023, 14, 5994–6015. [Google Scholar] [CrossRef]
- Quaresima, V.; Farzam, P.; Anderson, P.; Farzam, P.Y.; Wiese, D.; Carp, S.A.; Ferrari, M.; Franceschini, M.A. Diffuse Correlation Spectroscopy and Frequency-Domain near-Infrared Spectroscopy for Measuring Microvascular Blood Flow in Dynamically Exercising Human Muscles. J. Appl. Physiol. 2019, 127, 1328–1337. [Google Scholar] [CrossRef]
- Ma, K.F.; Kleiss, S.F.; Schuurmann, R.C.L.; Bokkers, R.P.H.; Ünlü, Ç.; De Vries, J.-P.P.M. A Systematic Review of Diagnostic Techniques to Determine Tissue Perfusion in Patients with Peripheral Arterial Disease. Expert Rev. Med. Devices 2019, 16, 697–710. [Google Scholar] [CrossRef]
- Durduran, T.; Yodh, A.G. Diffuse Correlation Spectroscopy for Non-Invasive, Micro-Vascular Cerebral Blood Flow Measurement. NeuroImage 2014, 85, 51–63. [Google Scholar] [CrossRef]
- Busch, D.R.; Balu, R.; Baker, W.B.; Guo, W.; He, L.; Diop, M.; Milej, D.; Kavuri, V.; Amendolia, O.; Lawrence, K.S.; et al. Detection of Brain Hypoxia Based on Noninvasive Optical Monitoring of Cerebral Blood Flow with Diffuse Correlation Spectroscopy. Neurocrit. Care 2019, 30, 72–80. [Google Scholar] [CrossRef]
- Buckley, E.M.; Parthasarathy, A.B.; Grant, P.E.; Yodh, A.G.; Franceschini, M.A. Diffuse Correlation Spectroscopy for Measurement of Cerebral Blood Flow: Future Prospects. Neurophotonics 2014, 1, 011009. [Google Scholar] [CrossRef]
- Wang, Q.; Pan, M.; Kreiss, L.; Samaei, S.; Carp, S.A.; Johansson, J.D.; Zhang, Y.; Wu, M.; Horstmeyer, R.; Diop, M.; et al. A Comprehensive Overview of Diffuse Correlation Spectroscopy: Theoretical Framework, Recent Advances in Hardware, Analysis, and Applications. NeuroImage 2024, 298, 120793. [Google Scholar] [CrossRef] [PubMed]
- Ruesch, A.; Yang, J.; Schmitt, S.; Acharya, D.; Smith, M.A.; Kainerstorfer, J.M. Estimating Intracranial Pressure Using Pulsatile Cerebral Blood Flow Measured with Diffuse Correlation Spectroscopy. Biomed. Opt. Express 2020, 11, 1462–1476. [Google Scholar] [CrossRef] [PubMed]
- Fischer, J.B.; Ghouse, A.; Tagliabue, S.; Maruccia, F.; Rey-Perez, A.; Báguena, M.; Cano, P.; Zucca, R.; Weigel, U.M.; Sahuquillo, J.; et al. Non-Invasive Estimation of Intracranial Pressure by Diffuse Optics: A Proof-of-Concept Study. J. Neurotrauma 2020, 37, 2569–2579. [Google Scholar] [CrossRef] [PubMed]
- Otic, N.; Sunwoo, J.; Huang, Y.; Martin, A.; Robinson, M.B.; Zimmermann, B.; Carp, S.; Inder, T.; El-Dib, M.; Franceschini, M.A.; et al. Multi-Wavelength Multi-Distance Diffuse Correlation Spectroscopy System for Assessment of Premature Infants’ Cerebral Hemodynamics. Biomed. Opt. Express 2024, 15, 1959–1975. [Google Scholar] [CrossRef] [PubMed]
- Amendola, C.; Lacerenza, M.; Buttafava, M.; Tosi, A.; Spinelli, L.; Contini, D.; Torricelli, A. A Compact Multi-Distance DCS and Time Domain NIRS Hybrid System for Hemodynamic and Metabolic Measurements. Sensors 2021, 21, 870. [Google Scholar] [CrossRef]
- Milej, D.; Abdalmalak, A.; Rajaram, A.; Lawrence, K.S. Direct Assessment of Extracerebral Signal Contamination on Optical Measurements of Cerebral Blood Flow, Oxygenation, and Metabolism. Neurophotonics 2020, 7, 045002. [Google Scholar] [CrossRef]
- Carp, S.A.; Farzam, P.; Redes, N.; Hueber, D.M.; Franceschini, M.A. Combined Multi-Distance Frequency Domain and Diffuse Correlation Spectroscopy System with Simultaneous Data Acquisition and Real-Time Analysis. Biomed. Opt. Express 2017, 8, 3993–4006. [Google Scholar] [CrossRef]
- Gómez, C.A.; Brochard, L.; Goligher, E.C.; Rozenberg, D.; Reid, W.D.; Roblyer, D. Combined Frequency Domain Near-Infrared Spectroscopy and Diffuse Correlation Spectroscopy System for Comprehensive Metabolic Monitoring of Inspiratory Muscles during Loading. J. Biomed. Opt. 2024, 29, 035002. [Google Scholar] [CrossRef]
- Boas, D.A.; Franceschini, M.A. Haemoglobin Oxygen Saturation as a Biomarker: The Problem and a Solution. Philos. Trans. R. Soc. Math. Phys. Eng. Sci. 2011, 369, 4407–4424. [Google Scholar] [CrossRef]
- Martelli, F.; Sassaroli, A.; Yamada, Y.; Zaccanti, G. Method for Measuring the Diffusion Coefficient of Homogeneous and Layered Media. Opt. Lett. 2000, 25, 1508–1510. [Google Scholar] [CrossRef]
- Gagnon, L.; Desjardins, M.; Jehanne-Lacasse, J.; Bherer, L.; Lesage, F. Investigation of Diffuse Correlation Spectroscopy in Multi-Layered Media Including the Human Head. Opt. Express 2008, 16, 15514–15530. [Google Scholar] [CrossRef] [PubMed]
- Wahl, M.; Gregor, I.; Patting, M.; Enderlein, J. Fast Calculation of Fluorescence Correlation Data with Asynchronous Time-Correlated Single-Photon Counting. Opt. Express 2003, 11, 3583–3591. [Google Scholar] [CrossRef] [PubMed]
- Eid, J.S.; Müller, J.D.; Gratton, E. Data Acquisition Card for Fluctuation Correlation Spectroscopy Allowing Full Access to the Detected Photon Sequence. Rev. Sci. Instrum. 2000, 71, 361–368. [Google Scholar] [CrossRef]
- Lacerenza, M.; Buttafava, M.; Renna, M.; Mora, A.D.; Spinelli, L.; Zappa, F.; Pifferi, A.; Torricelli, A.; Tosi, A.; Contini, D. Wearable and Wireless Time-Domain near-Infrared Spectroscopy System for Brain and Muscle Hemodynamic Monitoring. Biomed. Opt. Express 2020, 11, 5934–5949. [Google Scholar] [CrossRef] [PubMed]
- Wabnitz, H.; Taubert, D.R.; Mazurenka, M.; Steinkellner, O.; Jelzow, A.; Macdonald, R.; Milej, D.; Sawosz, P.; Kacprzak, M.; Liebert, A.; et al. Performance Assessment of Time-Domain Optical Brain Imagers, Part 1: Basic Instrumental Performance Protocol. J. Biomed. Opt. 2014, 19, 086010. [Google Scholar] [CrossRef]
- Pifferi, A.; Torricelli, A.; Bassi, A.; Taroni, P.; Cubeddu, R.; Wabnitz, H.; Grosenick, D.; Möller, M.; Macdonald, R.; Swartling, J.; et al. Performance Assessment of Photon Migration Instruments: The MEDPHOT Protocol. Appl. Opt. 2005, 44, 2104–2114. [Google Scholar] [CrossRef]
- Wabnitz, H.; Jelzow, A.; Mazurenka, M.; Steinkellner, O.; Macdonald, R.; Milej, D.; Żołek, N.; Kacprzak, M.; Sawosz, P.; Maniewski, R.; et al. Performance Assessment of Time-Domain Optical Brain Imagers, Part 2: nEUROPt Protocol. J. Biomed. Opt. 2014, 19, 086012. [Google Scholar] [CrossRef]
- Cortese, L.; Lo Presti, G.; Pagliazzi, M.; Contini, D.; Mora, A.D.; Pifferi, A.; Sekar, S.K.V.; Spinelli, L.; Taroni, P.; Zanoletti, M.; et al. Liquid Phantoms for Near-Infrared and Diffuse Correlation Spectroscopies with Tunable Optical and Dynamic Properties. Biomed. Opt. Express 2018, 9, 2068–2080. [Google Scholar] [CrossRef]
- Einstein, A. Über Die von Der Molekularkinetischen Theorie dWärme Geforderte Bewegung von in Ruhenden Flüssigkeiten Suspendierten Teilchen. Ann. Phys. 1905, 322, 549–560. [Google Scholar] [CrossRef]
- Cortese, L.; Lo Presti, G.; Pagliazzi, M.; Contini, D.; Mora, A.D.; Dehghani, H.; Ferri, F.; Fischer, J.B.; Giovannella, M.; Martelli, F.; et al. Recipes for Diffuse Correlation Spectroscopy Instrument Design Using Commonly Utilized Hardware Based on Targets for Signal-to-Noise Ratio and Precision. Biomed. Opt. Express 2021, 12, 3265–3281. [Google Scholar] [CrossRef]
- Zhou, C. In-Vivo Optical Imaging and Spectroscopy of Cerebral Hemodynamics; University of Pennsylvania: Philadelphia, PA, USA, 2007. [Google Scholar]
Synthetic Descriptor | Value | |
---|---|---|
685 nm | 830 nm | |
Illuminated area on sample surface | 2.89 mm2 | 2.89 mm2 |
Maximum power delivered to sample | 3.32 mW | 2.58 mW |
Responsivity | 1.44 × 10−8 m2 sr | 0.90 × 10−8 m2 sr |
Afterpulsing ratio | 2.4% | 5.0% |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Nabacino, M.; Amendola, C.; Contini, D.; Re, R.; Spinelli, L.; Torricelli, A. Fast Multi-Distance Time-Domain NIRS and DCS System for Clinical Applications. Sensors 2024, 24, 7375. https://doi.org/10.3390/s24227375
Nabacino M, Amendola C, Contini D, Re R, Spinelli L, Torricelli A. Fast Multi-Distance Time-Domain NIRS and DCS System for Clinical Applications. Sensors. 2024; 24(22):7375. https://doi.org/10.3390/s24227375
Chicago/Turabian StyleNabacino, Marco, Caterina Amendola, Davide Contini, Rebecca Re, Lorenzo Spinelli, and Alessandro Torricelli. 2024. "Fast Multi-Distance Time-Domain NIRS and DCS System for Clinical Applications" Sensors 24, no. 22: 7375. https://doi.org/10.3390/s24227375
APA StyleNabacino, M., Amendola, C., Contini, D., Re, R., Spinelli, L., & Torricelli, A. (2024). Fast Multi-Distance Time-Domain NIRS and DCS System for Clinical Applications. Sensors, 24(22), 7375. https://doi.org/10.3390/s24227375