Surface characteristics of electrode
Figure 2 shows the X-ray diffraction patterns of the platinum electrode samples prepared by using the two different metal salts. The two patterns in the bottom panel were recorded for the sample prepared with 0.01 M concentration of NaBH
4 and KBH
4 as a reducing agent. The two top patterns were recorded with the sample prepared with 0.08 M concentration of NaBH
4 and KBH
4 as a reducing agent. Initially, both sets of samples were impregnated with 0.01 M concentration of Pt(NH
3)
4Cl
2. The bottom pattern shows the data of the standard JCPDS card 4-802 [
14]. All peaks could be well indexed according to the standard. All four XRD patterns show only Pt peak angles, irrespective of the type and concentration of the reducing agent, i.e. the impregnated Nafion membrane contains [Pt(NH
3)
4]
2+ ions which were reduced completely. A clear XRD pattern was obtained from both strong and weak reducing agents which shows the complete reduction process of metal ions over the Nafion membrane provided with good interfacial contact between the electrode and electrolyte. Lower concentration peaks show larger distortion than in the case of higher concentrations, as well as larger broadening, which is attributed to induced strain between the Pt particles. Smaller particle sizes are expected in the case of broader peaks, in which lower concentrations of NaBH
4 show larger broadening than KBH
4 due to the higher reduction rate of the Na compound. The size of the Pt particles was calculated using Scherrer's equation [
15], evaluating the FWHM of the peak broadening by using the peakfit program. The subtraction of the instrumental broadening is included; those expressions are described elsewhere [
16]. The calculated crystallite size diameter (d) values for different concentrations are shown in
Table 1.
Optical micrographs of the chemically reduced ionic polymer-platinum electrode surfaces are shown in
Fig. 3. The Pt surface of the sample prepared from 0.01 M NaBH
4 shows highly strained and smooth reflecting surfaces, whereas in the case of 0.01 M KBH
4 less strain and larger particle size is observed. XRD crystallite size analyses for the corresponding samples were in agreement. In addition, both sample surfaces have shown a major difference in porosity. The samples prepared with KBH
4 as reducing agent have given more homogeneous pores than NaBH
4. Porous structures were obtained due to the fact of the weak reducing nature of the potassium compound. Similar conditions were followed to prepare Pt electrodes using NaBH
4 and KBH
4 as well, in which the reduction rate differed from one another. The higher the reduction rate the more strain and particle growth is observed, whereas the weak reducing agents show less strain and slow growth rate, resulting in uniform small pores.
Figure 4 shows the SEM images of the Pt particles deposited on the Nafion membrane of different crystallite sizes, which are in agreement with the size calculated from the XRD peak analysis. Energy dispersive X-ray spot analysis shows Pt peaks in M and L energy levels and a small percentage of fluorine and carbon peaks due to the base perfluorated polymer membrane.
Working principles of the sensor
The Pt/Nafion sensor element assembly fixed with a Plexiglas arrangement is shown schematically in
Fig. 1. The sensing performance of the amperometric sensor was conducted for the detection of hydrogen in nitrogen. The following electrochemical processes take place at the electrodes, i.e. oxidation at the anode side,
and reduction at the cathode side [
17].
The reference electrode of the sensor is air, i.e. the hydrogen activity (or combined water and oxygen activities) in air is used as a reference for the above mentioned electrochemical cell. Under short-circuit conditions, reaction
(1) occurs at the sensing electrode, whereas reaction
(2) occurs at the counter electrode. The sensor response current is controlled by the diffusion rate of bulk hydrogen or the oxidation rate of hydrogen. When the oxidation rate of hydrogen is much higher than the diffusion rate, the electrochemical reaction is a diffusion-limited process.
A carbon cloth with numerous micropores of about 0.1 μm aperture was used instead of a capillary hole-type diffusion barrier, which simplifies the structure of the sensing electrode chamber compared to the hole-type arrangement [
18]. The presence of carbon paste coated on carbon cloth as a current collector ensures the gas diffusion controlled process. The fundamental process for sensing hydrogen gas can be described in four steps: (i) the gas (H
2) diffuses through the carbon barrier to the electrode and is then absorbed on the sensing electrode H
2ads, (ii) the electrochemical reaction
(1) occurs, (iii) the protons move toward the counter electrode through the proton-conducting Nafion membrane, (iv) the resulting reaction product
(2) desorbs from the counter electrode and diffuses away. The electronic charge is transferred to or from the electrode in step (ii) which will then be passed through the external circuit [
19]. In amperometry, the observed Faradaic current is due to the electrochemical reaction and the current, dQ/dt, where Q is the charge transferred to the electrode by the chemical reaction.
Because the electrochemical reaction is a diffusion-limited process, the hydrogen gas flow rate and carbon diffusion barrier will influence the resulting response signal.
Figure 5 shows the gas flow rate versus output current for the sensor under the exposure of 4% hydrogen. It can be seen that the response signal increases steeply before the flow rate reaches 40 sccm, then increases steadily, and shows nearly a plateau at 70-80 sccm, the gas diffusion in carbon cloth was limited in this flow rate range. In the present work, all sensor measurements were performed at 80 sccm constant flow rate.
High H2 gas concentration response (NaBH4)
The performance of the hydrogen sensor was observed by employing the Pt | Nafion (0.01 M Pt(NH
3)
4Cl
2 and 0.04 M NaBH
4) assembly for different concentration steps of hydrogen in nitrogen at room temperature. The response curve for 1 min exposure of 1-10% H
2 gas is shown in
Fig. 6.
Different gas concentrations were tested in the ranges beginning at 1% and then going up to 10% hydrogen in steps of one percent hydrogen partial pressure for 1 min each time. The calibration plot of steady current of the sensor versus hydrogen concentration is shown in
Fig. 7. A linear relationship between the output current and the hydrogen concentration was observed for both forward and reverse direction of changes in the concentration from 1 to 10% of hydrogen gas. According to
Fig. 7, the linear regression equation is
y = - 47.66 + 119.79
x, where
y and
x are the response current (μA) of the sensor and the concentrations (%) of hydrogen, respectively. The average maximum sensitivity was observed to be 0.01 μA cm
-2ppm
-1, 0.007 μA cm
-2ppm
-1 and 0.008 μA cm
-2ppm
-1 in the hydrogen partial pressure ranges of 1-2%, 3-7% and 8-10%, respectively. The response time (t
90) to achieve a change in the signal level of 90% was typically less than 5-15 s. Both the response and recovery times decrease with increasing hydrogen concentration. When the samples were tested at low H
2 concentrations (ppm level) a non-linear behaviour was observed.
Low H2 gas concentration response (KBH4)
Optical micrographs of the Pt electrode surfaces prepared with 0.01-0.08 M KBH
4 have shown a more porous nature than with 0.01-0.08 M NaBH
4. Such porous surfaces are suitable for the detection of low H
2 gas concentrations. In order to proof the concept, ppm level concentrations were tested with the samples prepared with 0.01-0.08 M KBH
4. Typical step responses of the sensor element Pt | Nafion (0.01 M Pt(NH
3)
4Cl
2 and 0.01 M KBH
4) to various concentration steps of 100-1000 ppm hydrogen in nitrogen at room temperature are shown in
Fig. 8. Linear output current responses were observed for different samples over this hydrogen concentration range as shown in
Fig. 9. The average maximum sensitivity of the sensor is 0.0026 μA cm
-2ppm
-1 in the range of 100-1000 ppm, which is about 3 times higher than reported [
20]. Between 100 and 1000 ppm hydrogen, the response time is of the order of 25 and 5 s, in both forward and reverse direction, respectively. For practical purposes, the detection of hydrogen at the low ppm level is important. The same sets of samples were exposed to 10-50 ppm hydrogen gas. The response curve for 1 min exposure of 10-50 ppm H
2 gas is shown in
Fig. 10. Linear output current responses were observed for different samples over the hydrogen concentration ranges as shown in
Fig. 11. The average maximum sensitivity of the sensor is 0.0046 μA cm
-2ppm
-1 in the range of 10-50 ppm. A summery of both high and low detection ranges with different sensitivities are listed in
Table 2.
Compared with the sensors prepared by using NaBH4, a linear response was observed for hydrogen gas concentrations at the percentage level, whereas in the case of the sensor prepared by using KBH4, the linear response was extended up to the ppm level.
This behaviour of the different types of reducing agent can be explained by the achievement of the high effective electrode surface area via the porous structure. The electrochemical oxidation of hydrogen takes place along the three-phase boundary sites, where the Pt catalyst, solid electrolyte Nafion and H2 gas come together, which was observed largely in KBH4 samples. The porous structure obtained in KBH4 samples is a major advantage for very low concentration level detection. The small pore size structure provides more three-phase boundary sites for the electrochemical oxidation to occur. Thicknesses of the electrodes were observed in surface profilometry, which depend on concentrations of the reducing agent being used. The sensitivity performance shows a wide range.
However, the sensitivity is proportional to the charging current rather than the electrode thickness. The optimum sensitivity was observed with an electrode prepared using 0.01 M concentration of KBH4 as a reducing agent, the layer thickness of which was about 0.3 μm. Electrodes prepared with higher KBH4 concentrations (i.e., 0.06-0.08 M) have thicker Pt deposits but their active surface areas and hydrogen sensitivities were lower than those of the 0.01 M KBH4 electrode. This may be accounted for by the fact that the thicker Pt deposits have a larger average Pt particle size or isolated Pt particles are formed, which may not fulfil the three-phase boundary conditions for gas sensor applications.
The effective electrode area is sensitively varied by the amount of the reducing agent rather than the amount of [Pt(NH
3)
4]
2+ ions. When the amount of the Pt(NH
3)
4Cl
2 solution was varied within the range 0.001-0.1 M with a fixed value of KBH
4 solution (0.04 M), the charging current values were in the range 0.07-1.2 μA. But, when the reacted amounts of KBH
4 solution were changed within the range 0.01-0.08 M while that of the Pt(NH
3)
4Cl
2 solution was fixed at 0.01 M, the current was observed to be in a wider range of 0.05-0.3 μA. This suggests that, compared to the ion-exchange step, the reduction conditions have greater effects on the microstructure of the resulting electrodes. The response time (t
90) was typically less than 10-30 s.
Figure 12 shows the results of a cyclic exposure of the Pt | Nafion sensor element to 50 ppm hydrogen and no hydrogen. Identical maximum output currents were observed for several cycles changing within one minute from 0 to 50 ppm hydrogen. The diagram shows that the sensor remains stable over a substantial number of cycles. It should be pointed out that, when the hydrogen concentration is above 1000 ppm, with increasing hydrogen concentration, the amount of water and heat generated by the reactions of the sensor increase simultaneously. Hence, the conductivity of the Nafion and the temperature of the sensor are inclined to increase, which likely results in nonlinearity of the response current. A drift in the output current was in the range of 2% per day.