3. Results and Discussion
As already described above, the execution of a complete single-cell annual test at a constant load of 0.3 A/cm
2 was one of the main tasks in the context of the project QUALIFIX. The first annual test was conducted with an MEA with Fumapem
® AM-40 as the membrane and under reformate and air supply. This MEA is hereinafter referred to as FMT-A.
Figure 1 shows the voltage gradient of this single cell (grey). Within the first 1000 operating hours, the voltage increased with at an average rate of 10 µV/h. One possible cause for the voltage exaltation after start-up could be that the full potential of the MEA was not reached, as it has already been observed several times [
6,
22]. After 1000 h of operation, as foreseen in the test protocol, the single cell was again completely characterized. The degradation increased in the following 1000 h of operation, so that a degradation rate of −3.5 µV/h was recorded over the total period of 2000 operating hours. In order to reduce the influence of the characterization on the test specimen degradation, the characterization of the MEA was shortened. Especially, the CV and LSV measurements burdened the MEA significantly more than the EIS or polarization curves under the given conditions. For all subsequent characterizations, with exception of the final characterization at the end of the test, the CV and LSV measurements were omitted; this resulted also in a shortened characterization after 2000 h of operating time for the FMT-A. From the ensuing voltage gradient, a rapid loss of voltage, starting directly after characterization, became clear. The average degradation rate from 2000 h of operation until EoL was approximately −80 µV/h. Due to the high degradation, the MEA reached the EoL threshold (EoL = initial voltage value – 10%) after approximately 700 additional operating hours. The rapid loss of voltage results from irreversible damage to the MEA which made further operation of the test specimen impossible. After 2600 h of operation, the intended annual test with FMT-A was completed with a full electrochemical characterization.
Due to the early EoL of the annual test of MEA FMT-A, a second annual test was started. For this purpose, an MEA with a Fumapem
® AP-40 membrane was used. This MEA bears the name FMT-B in this publication. The associated voltage gradient is also shown in
Figure 1 (green). It should be mentioned that the number of MEAs presented here in the comparison study cannot be judged as a statistical one. On the other hand, the measurements of performance as an initial hurdle suggest a systematic difference between Fumapem
® AM and Fumapem
® AP. It can be concluded that the interface of the electrode in the Fumapem
® AP might have the more advantageous structure in terms of performance as shown in
Figure 1 and
Figure 2d. What is noticeable in the direct comparison of FMT-A and -B, besides the short running time and the significantly higher degradation, is the considerably higher initial voltage value of the FMT-B, despite the same platinum loading of this MEA compared with FMT-A.
As can be clearly seen, the voltage of FMT-A increased even further after the start of the test and reaches a maximum of 0.646 V at a constant current density of 0.3 A/cm2 within approximately 500 h of operation. The voltage increase could be due to the insufficient activation of the MEA during the start-up as already detected with MEA FMT-A. In comparison to FMT-A, the test procedure for FMT-B was adapted in some places in order to protect the MEA even more and to detect a material failure as early as possible. The characterization of the single cell, which is performed during the annual test every 1000 h of operation, was shortened. To keep the influence of the electrochemical characterization on the MEA degradation as small as possible, the characterization was reduced to polarization curves with dry reformate and air including the corresponding electrochemical impedance spectroscopy (EIS).
Between 7000 and 8000 h of operation technical problems occurred which were due to the experimental setup and not to the MEA which showed a fairly stable behavior as can be seen in
Figure 1. The significant drop in the voltage after 7100 h was due to the anode-side gas supply, where there was a narrowing of the gas supply line which resulted in a lower gas stream. This could be solved by a bypass and the MEA FMT-B was able to recover almost completely from this incident. The MEA FMT-B can be operated for over 9500 h under reformate supply without reaching EoL.
For FMT-C, a significantly higher voltage of 700 mV at the start of the test was noticeable. The reason for this was probably the use of pure hydrogen which, compared to FMT-A and -B, eliminates dilution by the gases N
2 and CO
2 and the effect of CO. The first significant voltage drop in the cell voltage after approximately 650 h was due to a failure of the test station. As a result, the MEA lingered at approximately 70 h at OCV which caused irreparable damage and is reflected in the voltage drop to 641 mV. The next recorded voltage reduction at approximately 1250 h was caused by a faulty supply of hydrogen. This fuel starvation also caused damage [
23], causing the voltage to drop to 620 mV. From this incident, there was a slight regeneration [
23] over the period up to approximately 1600 h to 625 mV but subsiding from here with a degradation rate of 23.3 μV/h to 604 mV after approximately 2450 h of operation. The subsequent increase to 612 mV can be attributed to an increase in the cathode stoichiometry from 2 to 3.5. This increase continued up to 614 mV until approximately 2700 h. In the following 300 h, the data are missing due to the fault of the test station but does not result in a test abort. After this time, the voltage dropped to 607 mV which is a degradation rate of 22.7 μV/h and very close to the rate at a stoichiometry of two. It can be concluded that the increase in the cathode flow had neither a positive nor negative effect on the MEA, since with increased flow, more phosphoric acid can also be discharged; this would have been an expected negative consequence. After a controlled shutdown for operational reasons, the test was continued again with a cathode stoichiometry of 2 (from 3000 h on). After a break-in phase, a significantly lower voltage was noticeable which stabilizes at 592 mV. If the abovementioned degradation rate were to continue from 2450 h (cathode stoichiometry = 2), the result would arithmetically be a value of 589 mV. It can be concluded that the degradation was constant from the period of 1600 h to 3100 h despite various influences and was caused by the same process. The last voltage drop, at about 3550 h, was due to the second undersupply of hydrogen. However, in this case, a larger amount to damage occurred which cannot be recovered from as shown by the drop in voltage to 540 mV at the end of the test after approximately 3840 h at a significantly increased degradation rate of 83.3 μV/h.
During the course of the test of the FMT-A, until 2000 h of operation, the performance steadily increased (see
Figure 2a). The polarization curve at EoL after 2600 h of operation shows a distinctly reduced performance than the previous polarization curves which also corresponds to the significantly higher degradation rate (from 2000 h, see
Figure 1) and a bend in the curve due to the higher stoichiometry, up to 0.06 A/cm
2, because of a mass flow controller which reinforces the H
2 crossover effect. The attrition of the OCVs, which is already recognizable after 1000 h, is striking. At the beginning of test, the OCV exhibited a value of 950 mV, but then it fell steadily. At 2600 h, an OCV of only 650 mV was reached. The open circuit voltage of the MEA was directly affected by the mixed potential losses in the current-free state. By diffusion of hydrogen and oxygen through the membrane, the direct reaction of the reactants without electron transport via load circuit occurred. The result was a reduction in the OCV. In practice, the flow of hydrogen was dominant due to the small molecular size. The illustrated decrease in the OCVs therefore indicates an increase in hydrogen crossover.
As previously described, the characterization for the MEA FMT-B was adjusted. Therefore, the polarization curves were performed up to a minimum voltage of 0.5 V, except the begin and end of tests; during the extended characterization (at 0, 9000, and 9500 h) the voltage threshold during the polarization curves went down to 0.4 V. In
Figure 2b, these polarization curves are presented. During the first 2000 h of the long-term test, the FMT-B, like FMT-A, showed an increase in performance and, thus, represents an activation of the MEA. Subsequently, this MEA showed uniform and low-performance consumption which is also reflected in the low degradation rate (see
Figure 1). In addition, there was hardly any decrease in the OCV values during the extended long-term test (see
Figure 2b).
In
Figure 2c, the polarization curves also showed no significant loss in the OCV for FMT-C. However, the fuel starvations left a clearly damaged MEA; this effect was further intensified by natural degradation.
In
Figure 2d, the polarization curves for all three MEAs are compared. Both FMT-B and -C show a much better performance than FMT-A. It should be noted here, again, that FMT-A and -B were operating under reformate and FMT-C under pure but humidified hydrogen.
It is also important to note that, while the MEA FMT-A showed only 2600 h between the two polarization curves, the time difference for FMT-B adds up to 9000 h. Although a slight deterioration in performance was observed for FMT-B, the degradation for FMT-A was significantly higher. These observations confirmed the selection of the correct adaptations for FMT-B.
In addition to the polarization curves, electrochemical impedance spectra were recorded, and the results are shown in
Figure 3 and
Figure 4. The values presented in
Figure 4 were derived from the impedance measurements plots shown in
Figure 3. The ohmic resistance was determined from the intersection in the high-frequency determined area [
24]; in combination with the diameters of the semi-circles, the total resistance [
25] was obtained. The results show that the ohmic resistance of FMT-A decreased by 0.1 Ω/cm
2 over the test duration of 2600 h (see
Figure 4a,b). The noticeable decrease in ohmic resistance indicates a highly diluted membrane. In the case of MEA FMT-B, the decrease in ohmic resistance (from 0.58 to 0.53 Ω/cm
2) was only half as high compared to FMT-A over the complete test duration of >9000 h. This shows that despite the three times shorter test duration, the decrease in ohmic resistance was higher for FMT-A. This also shows the improvement in the membrane quality between FMT-A (Fumapem
® AM-40) and FMT-B (Fumapem
® AP-40) very clearly. One of the possible explanations is the poorer retention of phosphoric acid in the interface of the MEA using Fumapem
® AM (FMT-A) which is observed during the long-time experiments. This is shown by the similar phosphoric acid contents of the EoL-MEAs as presented in
Figure 5. However, the similarity of the EoL content was achieved after approximately 2500 h (AM) versus 10,000 h (AP).
The data for FMT-A under hydrogen at EoL are missing in
Figure 4, because the deviation of the values was too large, hence measurement was not possible. One assumption is that, in the case of pure hydrogen, more gas passes through the damaged membrane in comparison to reformate gas, where the hydrogen is diluted. If one considers the change from reformate to hydrogen for FMT-B during the BoL examination (
Figure 4), a decrease in the ohmic resistance under hydrogen operation is shown. One reason for the decrease could be the reduced volume flow in the case of hydrogen compared to the reformate supply. Due to the reduced volume flow when using hydrogen, less gas is available which can transport water from the MEA. This can therefore lead to a dilution of the electrolyte which, in turn, improves the conductivity as already shown by Liu [
26]. However, this explanation cannot be cited for the EoT observations shown in
Figure 4b, since the ohmic resistance in this case rose when reformate was changed to hydrogen. This shows the complex reactions caused by the reformate operation and the degradation after more than 9000 h of continuous operation; these observations will have to be examined more closely and further in the future.
The determination of the ohmic resistance for the two MEAs FMT-A and FMT-B indicates a distinctly different degradation behavior of the two membrane types. For this reason, the H
2 crossover and the internal short circuit resistance were also determined with the help of LSV measurements, and the results are presented in
Figure 6. Due to the different test protocols, such a measurement was not carried out with FMT-C.
In the case of FMT-A (see
Figure 6), the hydrogen crossover increased significantly over the 2600 hours of operation. A crossover current of approximately 4 mA/cm
2 was detected at the start of the test. After 1000 h of operation, the current increased to approximately 30 mA/cm
2 and reached a maximum of 200 mA/cm
2 at 2600 h. At the same time, the internal resistance of the MEA decreased similarly. After 2600 h, the internal short circuit resistance fell from the initial 10 Ω to nearly 0 Ω. All the results of electrochemical characterization indicate possible damages of the membrane which could be the cause of the cell failure. In terms of hydrogen crossover, the MEA FMT-B showed quite a different behavior in direct comparison with the FMT-A: FMT-B showed a very low and also stable hydrogen crossover current in the course of the experiment, while the FMT-A, as described above, had a similar starting value but increased significantly over the course of the considerably shorter test. In addition, the internal short circuit resistance of FMT-B decreased by 50%. While the constant hydrogen crossover is an indicator of a defect-free membrane, the reduction of the internal short circuit resistance is a sign of a possible thinning of the membrane [
27].
Due to the assumptions of a destroyed membrane in the case of MEA FMT-A and a thinned membrane within FMT-B based on the electrochemical investigations presented above, the MEAs were examined by post-mortem analysis with help of µ-CT investigations and compared with the respective untested reference MEA from the same batches. In
Figure 7a, the ante-mortem test results of the reference MEA to FMT-A (hereinafter referred to as BoL) are shown, and
Figure 7b represents the results of the post-mortem analysis of the FMT-A after 2600 h under constant load operation. Because the BoL-MEA was not hot pressed, the MEA was delaminated, while the tested MEA FMT-A could be scanned as one because of the compression force within the Baltic CCU.
In a direct comparison, it is clear to see how severe the MEA was degraded after 2600 h of operation at a constant load. The membrane, which is well recognizable in the reference MEA, showed uniform layer thicknesses and no defects as can be seen clearly in
Figure 7a, is almost completely destroyed after operation. One reason for the destruction of the PBI membrane might be oxidative degradation due to the incomplete reduction of O
2 on the cathode (C
+) which leads to the generation of hydroperoxyl and hydroxyl radicals [
28,
29]. The catalyst layers, anodic and cathodic, are directly adjacent to each other and, thus, provide the reason for the high hydrogen crossover (
Figure 6), the decrease of ohmic resistance (
Figure 3 and
Figure 4), and the reduction of the internal short circuit resistance to 0 Ω. What is striking, when looking at the catalyst layer, is that this layer is covered by many fine hairline cracks (
Figure 7a,b). This is also the case for the post-mortem observation; there is hardly a further expanding of the drying cracks, as it can be seen within the two 2D zoomed-in images of both MEAs. The layer thickness analysis also clearly shows (see
Table 3) that the other layers hardly show any changes, a slight overall decrease on the anode (A
–) side, but the failure of the MEA FMT-A is almost entirely attributable to the membrane.
Figure 8 shows the µ-CT studies of the MEA FMT-B (
Figure 8b) which exceeded the long-term test. A direct comparison with an untested MEA from the same batch was also mentioned here (
Figure 8a). A slight delamination could be also observed for the reference MEA, but it was possible to scan the MEA as one. The MEA FMT-B was delaminated while removing it from the CCU.
Also in the case of the FMT-B, with the help of this imaging method, a statement made by the electrochemical measurements could be substantiated. The thinning of the membrane after more than one year of running time could be confirmed by µ-CT investigations as shown in
Figure 8b, and
Table 4 confirms that the membrane shrunk from just under 60 µm to 21 µm and, thus, has only a third of the starting thickness. Furthermore, it can be observed that the drying cracks in the catalyst layer widened significantly in the course of the experiment as shown in
Figure 8a,b). The chosen test conditions seem to indicate structural changes in the CL [
30] for both MEA types (see
Figure 6 and
Figure 7). Similar observations were made by Pokhrel at al. [
31]; the slight drying cracks, due to the manufacturing process, within the BoL-CL also widened significantly, even more so, the CL exhibit an island structure. This could be due to the dehydration of the catalyst layer or bulk carbon corrosion [
32]. But, as mentioned above in the
Section 2, the catalyst loading was chosen to be significantly higher (e.g., on the cathode 1.6–1.7 mg/cm
2 where values of 1 mg/cm
2 are common [
33]) to avoid significant influences by the catalyst degradation. This can be shown for FMT-B, since it was possible to quantify the loss of Pt ESA by cyclovoltammetry. The values for FMT-B showed a Pt ESA loss of 34% which is an acceptable value considering the very long test duration of 9800 h. Measurements with MEAs of similar performance show a loss of 12%–20% already after 500 h [
11].
For FMT-C, similar statements can be made as for FMT-B regarding the membrane (
Figure 9). Due to the circumstances of the measurement, there is no BoL Image of the MEA FMT-C. Assuming, however, that the thickness was initially at 60 microns as for FMT-B, the EoT value of 40 microns (see
Table 5) is comparable to the rate of thinning over time for FMT-B. Striking is the strong decrease of the CL layer on the anode and the adhesion of this layer to the membrane. These facts support the abovementioned assumption that the process of anode starvation was responsible for a main part of the degradation as observed by Yezerska et al. [
23].
Figure 5 shows the H
3PO
4 contents of the tested MEA before (solid pillar) and after (striped pillar) the two long-term tests at a constant load of 0.3 A/cm
2. Although the MEAs were equipped with different membranes, the phosphoric acid content was similar before and after the respective tests.
Figure 10 shows an alternative method for determining the phosphoric acid content of FMT-C. Since, as mentioned above, no such measurement could be performed for this MEA. The data shown here from a dynamic mechanical analysis (DMA) performed at FUMATECH show an example of an MEA/membrane of the same membrane chemistry after 1000 h (FMT-X) operation compared with FMT-C after EoT (3850 h). The significantly increased modulus of elasticity for FMT-C shows a higher strength of the membrane. From this, one can deduce an increased rigidity of the membrane which is an indication of the loss of phosphoric acid. This fact reinforces the suspicion that the membrane, as confirmed by FMT-B by determining the H
3PO
4 content, thinned due to the phosphoric acid loss. However, a direct comparison between FMT-B and -C shows that this fact does not lead to a failure of the MEA which speaks in favor of the AP membrane chemistry.
The different methods for determining the phosphoric acid content demonstrate that FUMATECH has managed to ensure a uniform doping for the membranes over different batches and MEA types. This also shows that, in this case, the H
3PO
4 content can be removed from the degradation effect analysis, although phosphoric acid loss is one of the driving degradation paths [
34]. However, in a direct comparison of the BoL- and EoL-H
3PO
4 contents of FMT-A and -B, it is noticeable that in both cases, the phosphoric acid concentration within the MEA was reduced by almost 50%. This observation agrees with the hypothesis made by Oono [
35] that a low thinning rate of the membrane is associated with a low phosphoric loss rate because of the high H
3PO
4 loss is in conjunction with the destruction of the membrane of FMT-A and the thinning of the membrane by two-thirds in the case of the MEA FMT-B and one-third for FMT-C (see also Reference [
6]).