2. Design and Analysis of the Structure
Figure 1a and
Figure 1b show a 3D view and 2D schematic of the proposed D-shaped sensor structure, respectively. The sub-figure
shows the mode field distribution of the core mode, applied for refractive index (RI) measurements. The proposed fibre has an area comprising a
air hole array arranged in a square lattice. The lattice spacing of the array is denoted by
and the basic small air holes have a radius denoted by
. Some basic small air holes are replaced with elliptical holes and big air holes to strengthen the SPR. To ensure the SPR effect successful excitation, a part of the fibre background material is polished with an appropriate distance to the centre to form a D-shaped structure and the flattened plane is coated with two gold nanowires with the thickness and length donated by
and
, respectively. In the central region of the silica, two basic small air holes are omitted for the sake of forming a core mode area, four adjacent downward air holes are replaced with one large and two basic small elliptical holes pointing to the core region to strengthen the SPR effect, the major and minor semi-axis of one big and two small elliptical air holes are represented as
,
and
,
, respectively. During fabrication, tolerance may exist when controlling the elliptical air holes for pointing to the core region, the related effects of the deviation are discussed in the following section. Meanwhile, to further strengthen the intensity of the core mode area when the SPR effect occurs, the upward basic small air hole is replaced with a horizontal ellipse with the major and minor semi-axis represented as
and
, respectively. Moreover, eight basic small air holes on the left and right side of the core mode area are replaced into big air holes with radius
to compress the core region and improve sensor performance. A liquid analyte is introduced on the external part of the fibre.
In the numerical simulation, this work uses finite element method (FEM) to solve an eigenmode. Furthermore, a perfectly matched layer (PML) with a thickness chosen as
μm is placed outside the analyte area to truncate the region beyond the fibre structure and limit the computational area by absorbing all radiation energy [
26,
27].
Figure 2 demonstrates the 2D schematic of the proposed sensor structure in the fabrication process. In the process, we utilized an external container to gather the different types of capillaries and rods in the start-up stage. In this paper, thick-wall, thin-wall, and elliptical-core fibres are applied to form the small hole, big hole and basic elliptical core, respectively. Two solid rods were used to form the core area. An elliptical-core rectangle capillary was utilized in order to form the big elliptical-core beneath the core area. Moreover, ten auxiliary solid rods, as seen in gold in
Figure 2, were designed for fixed position of the proposed structure part. After bringing these capillaries and rods together in a rectangular external container, liquid silica is allowed to infiltrate the space among these structures and the start-up structure is made. Then the auxiliary solid rods are removed and the remaining part is put into a circular external container, after pouring the fused silica, the final sensor structure is thereby made.
Fused silica is applied as the fibre material for its superior properties and performance, with a material RI which can be calculated by the third-order Sellmeier Equation (
1) [
28], where
denotes the wavelength in free-space.
Two gold nanowires, similar to the other metal materials, are used to excite the SPR effect. The dispersion of the gold nanowires can be easily obtained via the Lorentz–Drude model expressed in the following Equation [
29,
30].
where
denotes the dielectric constant with a ultra-high frequency, and
stands for the angular frequency of the light propagating through the proposed fibre.
and
represent the plasma and damping frequency, respectively.
is the weight factor.
and
are the frequency and spectral width of the Lorentz oscillator, respectively, where
and
in angular frequency.
It is generally acknowledged that the SPR-based PCFs work owing to the interactions between electrons on the surface of the plasmonic material along with the evanescent field [
31,
32]. The confinement loss (
) is a good metric representing the loss in propagation, which can be calculated using Equation (
2) [
33,
34], where
denotes the effective RI of the core mode. The unit of
value is dB/cm.
In this article, utilizing two Au nanowires and core regions to excite the SPR effect with variable physical values, the RI value can be measured with a wavelength coverage from 850 to . Furthermore, with a significant red-shift in the curve when the RI value changes between 1.41 and 1.42, the structure achieves an extreme high value.
Figure 3a illustrates the
curve of both the
x- and
y-polarized modes and the real part of the effective RI of the core modes and SPP mode when the analyte RI is 1.40 with a list of structural parameters of the proposed design:
μm,
μm,
μm,
μm,
μm,
μm,
μm,
μm,
nm,
μm,
μm. Furthermore,
Figure 3b–d shows the mode field distributions of the proposed structure simulated with wavelength of 950, 973, 1050 nm, respectively. In
Figure 3a, the core mode along with the SPP mode totally couple together in the area of the polished Au-coated plane when the wavelength is about 973 nm. The value of the
curve is tiny in those wavelength regions where resonance almost does not occur. A reasonable explanation for this phenomenon is that the majority of energy is limited in the core region when the wavelength is in those non-resonant regions, as illustrated in
Figure 3b,d. When SPR is fully excited with a resonant wavelength
, the core mode couples along with the SPP mode and energy distribution changes from the core area to the surface of the flattened Au-coated plane, as illustrated in
Figure 3c. At this moment, the phase-matching condition is achieved, as shown in the intersection between the orange solid curve and crimson dash–dot curve in
Figure 3a. Meanwhile, the
curve reaches a peak. Moreover,
Figure 3a shows that on a dB-scale the
value of the core mode in
y-polarization is already much greater than the core mode in
x-polarization, which is almost fluctuating around 0 dB/cm. This further illustrates the realization of the single-polarization sensing performance. Therefore, we select
y-polarization for operation in this work.
The variation in the analyte RI impacts the effective RI in the region of the excited SPR effect, further changing the mode field distributions and the related
value. Eventually, the
curve starts to red- or blue-shift and the peak value changes. Thus, it is of great significance to find an evaluation scheme to measure the correlation between the changing parameters and the affected sensor performance. In this scenario, the correlation is appropriate to measure the variation of the
curve in conjunction with the analyte RI. Therefore, there are two good metrics: RI sensitivity (
) and amplitude sensitivity (
), which are defined in the equations below [
35],
where
represents the change in the resonant wavelength value, while
stands for the RI variation.
represents the change in the confinement loss and
denotes the initial confinement loss. The unit of
is in nm/RIU and the unit of
is in
.
Another important metric in relation to the performance of the proposed sensor is the figure of merit, focusing on the full width at half maximum (FWHM) value. This is defined by the equation below [
36],
where
denotes the FWHM.
3. Simulation Results and Discussion
Figure 4a shows the
curve of the
y-polarized mode with
chosen as 1.70, 1.75, 1.80, 1.85 μm, and the analyte RI is set as 1.41 and 1.42. From
Figure 4a, with
increasing from 1.70 to 1.85 μm, the
value of the
spectra with analyte RI 1.41 starts to visibly red-shift and the peak value of the
curve gradually increases. The resonance wavelength of the
spectra with analyte RI 1.42 is almost constant and the peak value of the
curve exhibits a noticeable decrease. Furthermore, when making a comparison among the
peak values and related envelopes, it can be further inferred that with an increasing
, the envelope of the
spectra associated with the changing analyte RI appears to blue-shift. A reasonable explanation for this phenomenon is when
increases, the
y-polarized mode region is compressed, causing the intensity of the envelope peak to visibly increase. However, the big elliptical hole, which has its major-axis denoted by
, has a long distance to the gold nanowires. So the variation in
has an insignificant effect to the phase-matching condition, and thus the change has almost no impact on the resonance wavelength and RI sensitivity. In accordance with Equation (
4), the maximum
values achieved were 18,750, 18,850, 19,600, 18,800 nm/RIU with
chosen as 1.70, 1.75, 1.80, 1.85 μm, respectively. The calculated
values of the proposed structure firstly increase and then decrease with
changing from 1.70 to 1.85 μm, reaching a maximum near
μm. In accordance with Equation (
5), the maximum
values achieved were 2945, 2705, 2517.5, and 2311.25
when
was set as 1.70, 1.75, 1.80, 1.85 μm, respectively, and the analyte RI was set as 1.42, demonstrating that the
value of the structure decreases when
rises from 1.70 to 1.85 μm and always has a good sensing performance.
Figure 4b illustrates the
curve of the
y-polarized mode with
chosen as 0.50, 0.60, 0.70 μm, and the analyte RI of 1.41 and 1.42. From
Figure 4b, when the analyte RI is 1.41 and
increases from 0.50 to 0.70 μm, the
value of the
curve has a slight red-shift and the peak of the
curve gradually increases. The resonance wavelength of the
curve with an analyte RI of 1.42 is almost constant and the
CL peak visibly decreases. Furthermore, when making a comparison among the
peak values and related envelopes, it can be found that when the
value increases, the envelope of the
spectra in regard to changing the analyte RI appears to blue-shift slightly. A reasonable explanation for this phenomenon is that when
increases, the
y-polarized mode region is compressed. However,
has a lesser effect on the core mode region compared to
; therefore, the intensity of the
curve slightly increases and the
and
values are almost constant. With the increase in the
peak value with RI set as 1.41 and a decrease in the
intensity when RI is 1.42, the related wavelength of the maximum value of the whole envelope of the
curve gradually changes from 1200 to 1000 nm, which is a blue-shift of the envelope. In accordance with Equation (
4), the maximum
values achieved were 18,750, 19,600, 18,600 nm/RIU with
chosen as 0.50, 0.60, 0.70 μm, respectively. The calculation results show the
value of the proposed structure first increases and then decreases with
changing from 0.50 to 0.70 μm, reaching a maximum near
μm. In accordance with Equation (
5), the maximum
values achieved were 2705, 2517.5, and 2372 RIU
when
was 0.50, 0.60, 0.70 μm, respectively, when the analyte RI was 1.42, illustrating that the
value of the proposed structure diminishes when
increases from 0.50 to 0.70 μm and always has a great sensing performance.
Figure 5a demonstrates the
curve of the
y-polarized mode when
was 0.70, 0.75, 0.80, 0.85 μm, and the analyte RI was chosen as 1.41 and 1.42. From
Figure 5a, when the analyte RI was 1.41 and
increases from 0.70 to 0.85 μm, the
value of the
curve appears to red-shift with a noticeable trend and the
peak value significantly falls. The resonance wavelength of the
spectra with an analyte RI 1.42 has a slight blue-shift and the peak value of the
curve visibly decreases. Furthermore, it is easy to infer from
Figure 5a that the FWHM value of the
curve almost remains intact when
increases. Furthermore, when making a comparison among the
peak values and related envelopes, it can be further deduced that when
increases, the envelope of the
spectra associated with the changing analyte RI appears to visibly red-shift. The main reason considered for this is that when
increases, the channel linking the gold nanowires and the core mode area narrows, changing the distributions of the mode field and influences the
y-polarized mode. In accordance with Equation (
4), the maximum
values achieved were 17,900, 18,900, 19,600, 19,100 nm/RIU with
chosen as 0.70, 0.75, 0.80, 0.85 μm, respectively. The calculation results demonstrate that the
value of the proposed structure first rises and then declines with the variation in
from 0.70 to 0.85 μm, reaching a maximum near
μm. In accordance with Equation (
5), the maximum
values achieved were 835.9, 1348.2, 2226.7, and 3577.143 RIU
when
was set as 0.70, 0.75, 0.80, 0.85 μm, respectively, when the analyte RI was set as 1.42, illustrating that the amplitude sensitivity of the proposed structure significantly boosts when
is changed from 0.70 to 0.85 μm, thus improving its sensing performance.
Figure 5b shows the
curve of the
y-polarized mode with
chosen as 0.20, 0.30, and 0.40 μm, and the analyte RI as 1.41 and 1.42. From
Figure 5b, when the analyte RI was 1.41 and
increased from 0.20 to 0.40 μm, the
value of the
curve has a visible blue-shift and the peak value of the
curve falls significantly. The
value of the
spectra with an analyte RI of 1.42 sees a significant blue-shift and the peak value of the
curve appears to noticeable increase. Moreover, it is evident that the FWHM value of the
spectra decreases when
increases. Furthermore, when making a comparison among the
peak values and related envelopes, it can be concluded that when the
value increases, the envelope of the
spectra in relation to the changing analyte RI appears the visibly red-shift. The main reason considered is that when
increases, the region between the gold nanowires and silica area is compressed, leading to the changing the phase-matching condition, further affecting the resonance wavelength and peak intensity. In accordance with Equation (
4), the maximum
values achieved were 17,200, 19,600, 14,000 nm/RIU with
chosen as 0.20, 0.30, 0.40 μm, respectively. The calculation results demonstrate that the
value of the proposed structure first increases and then decreases with
changing from 0.20 to 0.30 μm, reaching a maximum near
μm. In accordance with Equation (
5), the maximum
values achieved was 488.37, 2226.7, 5874.8 RIU
when
was set as 0.20, 0.30, 0.40 μm, respectively and the analyte RI was 1.42, illustrating that the
value of the structure has a significant leap with
increasing from 0.20 to 0.40 μm, thus optimizing its sensing performance.
Figure 6a,b shows the
curve of the
y-polarized mode with
chosen as 0.80, 0.85, 0.90 μm and
set as 0.40, 0.50, 0.70 μm when the analyte RI is 1.41 and 1.42, respectively. From
Figure 6a,b, when the analyte RI is chosen as 1.41 and 1.42 with
and
increasing, the
value of the
spectra are almost constant and the
peak value changes little. A reasonable explanation for this phenomenon is that the two elliptical small holes are further away from the two gold nanowires and the
y-polarized mode region, so the variation has an almost negligible effect on the mode field distributions, thus having little impact on the
y-polarized mode and sensing performance. In accordance with Equation (
4), the maximum
values achieved were 18,600, 19,300, 19,600 nm/RIU with
set as 0.80, 0.85, 0.90 μm, respectively, and can reach 19,500, 19,600, 19,000 nm/RIU when
is set as 0.40, 0.50, 0.70 μm, respectively. It can be concluded that the
value of the proposed sensor reaches an optimal value near
μm,
μm. In accordance with Equation (
5), the maximum
value can reach about 2315.6, 2260, 2226.7 RIU
when
is set as 0.80, 0.85, 0.90 μm, respectively, and can reach 2253.3, 2226.7, 2151.1 RIU
when
is set as 0.40, 0.50, 0.70 μm, respectively, with an analyte RI chosen as 1.42. It can be deduced that the maximum
value of the sensor has negligible change along with variation in
and
, but the structure can always achieve an outstanding sensing performance.
Another factor which must considered is that in the fabrication process the small elliptical air holes may deviate from pointing towards the core area. In this work, is used to describe the angle of the direction of these elliptical holes and the vertical axis. In ideal condition, .
Figure 7 illustrates the
curve of the
y-polarized mode with
chosen as 35, 45, and
with the analyte RI set as 1.41 and 1.42. From
Figure 7, when the analyte RI is chosen as 1.41 and 1.42 with
increasing, the
value of the
curve almost remains unchanged and the
peak value changes little. A reasonable explanation for this phenomenon is that these two elliptical small holes are far from the gold nanowires and
y-polarized mode region. Therefore, the variation has an insignificant effect on mode field distributions, thus having little impact on the
y-polarized mode and sensing performance. According to Equation (
4), the maximum
values achieved were 19,000, 19,600, 19,400 nm/RIU with
set as 35, 45, and
, respectively, signifying that the RI sensitivity of the proposed structure reaches an optimal value near
. According to Equation (
5), the maximum amplitude sensitivity can reach about 2220, 2226.7, and 2233.3 RIU
when
is set as 35, 45, and
, respectively, when the analyte RI is set as 1.42. The calculation results infer that the maximum
value of the proposed structure changes slightly with variation in
, but the proposed structure always has an excellent sensing performance. In conclusion, the deviation existing in the fabrication process negligible impacts the mode field distributions and sensing performance of the proposed sensor, thus lowering the manufacturing difficulty and increases the fault tolerance in the fabrication process, further reducing the production cost.
Figure 8 demonstrates the
curve of the
y-polarized mode with
chosen as 0.4, 0.5, 0.6,
μm, and analyte RI as 1.41 and 1.42. From
Figure 8, when the analyte RI is chosen as 1.41 and
increases from 0.4 to 0.7 μm, the
value of the
curve exhibits a noticeable blue-shift and the
curve peak value significantly drops. The
value of the
curve with analyte RI set as 1.42 sees a significant blue-shift and the
curve peak value first increases and then decreases. Furthermore, as can be seen in
Figure 7, the FWHM value of the
spectra appears to decrease significantly when
increases. Furthermore, when making a comparison among the
peak values and related envelopes, it can be deduced that when the
value increases, the envelope of the
spectra in association with the changing analyte RI appears to noticeably red-shift and then blue-shift. A reasonable explanation for this phenomenon is that when
increases, the area of the
y-polarized mode is gradually compressed, strengthening the intensity of the
peak value. However, on the other hand, the increasing small air circles narrow the channel linking the two gold nanowires and the
y-polarized mode area, suppressing the SPR effect and changing the mode field distributions, resulting in an intensity decline and a
curve shift. In accordance with Equation (
4), the maximum
values can reach 17,200, 19,600, 12,550, and 9450 nm/RIU when
is set as 0.4, 0.5, 0.6, and 0.7 μm, respectively. The calculation results illustrate that the
value of the proposed structure first increases and then decreases with
changing from 0.4 to 0.7 μm, reaching a maximum near
μm. In accordance with Equation (
5), the maximum
values achieved were 349.9, 2226.7, 4900, and 4130.8
when
was set as 0.4, 0.5, 0.6, and 0.7 μm, respectively, when the analyte RI was set as 1.42. In conclusion, the amplitude sensitivity of the proposed structure first increases significantly and then declines with
increasing from 0.4 to 0.7 μm. Moreover, the sensing performance is significantly improved with the variation.
Figure 9 illustrates the
curve of the
y-polarized mode with
chosen as 1.6, 1.8, 2.0 μm, and analyte RI chosen as 1.41 and 1.42. From
Figure 9, when
increases from 1.6 to 2.0 μm, the
value of the
curve with analyte RI chosen as 1.41 has a visible red-shift and the
peak value has a significant boost. The resonance wavelength of the
spectra with an analyte RI of 1.42 sees a wide-ranging red-shift and the
curve peak value sees a noticeable growth. Furthermore, when making a comparison among the
peak values and related envelopes, it can be found that when the
value increases, the envelope of the
spectra in correlation with the changing analyte RI appears to visibly red-shift and become wider. A reasonable explanation for this phenomenon is when the
is increasing, the area of the
y-polarized mode region is effectively compressed, greatly strengthening the intensity of the peak value. Furthermore, the increase in
smooths the channel between linking the gold nanowires and the
y-polarized area, changing the phase-matching condition, thus resulting in an intensity increase and a
curve shift. In accordance with Equation (
4), the maximum
values achieved were 19,600, 17,750, 5600 nm/RIU with
, respectively, chosen as 1.6, 1.8, 2.0 μm. The calculation results show the
value of the proposed structure witnesses a great fall with
changing from 1.6 to 2.0 μm. In accordance with Equation (
5), the maximum
value achieves about 2226.7, 2915.96, and 1714.3 RIU
when
is, respectively, chosen as 1.6, 1.8, 2.0 μm and analyte RI is set as 1.42. It’s concluded that the amplitude sensitivity of the proposed structure firstly climbs up significantly and then declines greatly with
increasing from 1.6 to 2.0 μm.
Figure 10 shows the
CL curve of
y-polarized mode with
(abbreviated as
t), respectively, chosen as 50, 55, 60, 65, 70 nm, and analyte RI is set as 1.41 and 1.42. From
Figure 10, when analyte RI is chosen as 1.41 and
t increases from 50 to 70 nm, the
value of the
CL curve has a noticeable red shift and the peak value of
CL curve firstly climbs up and then declines slightly. The resonance wavelength of the
CL spectra with analyte RI 1.42 witness a significant red shift and the peak value of
CL curve reduces in a large scale. Moreover, it is clear to be found that the FWHM of the
CL spectra receives a noticeable boost with
increasing when the analyte RI is set as 1.42. Furthermore, when making a comparison among those peak values of
CL curve and related envelopes, it can be inferred that with
t value increasing, the envelope of the
CL spectra in company with changing analyte RI occurs to red-shift and becomes wider visibly. A reasonable explanation for the phenomenon is when
t increases, the SPR effect is further excited, which strengthens the intensity of the peak value and alters mode field distributions, thus influence
y-polarized mode and lead to a red shift of
CL curve. In accordance with Equation (
4), the maximum
value can achieves 9050, 13,350, 19,600, 19,000, and 19,200 nm/RIU when
t is, respectively, set as 50, 55, 60, 65, and 70 nm. The calculation results demonstrate the
value of the proposed structure climbs up significantly at first and then almost remains the same and changes slightly with
t increasing from 50 to 70 nm. In accordance with Equation (
5), the maximum
value achieves about 3020, 3195.3, 2226.7, 1382.1, and 555.3 RIU
when
t is, respectively, set as 50, 55, 60, 65, and 70 nm and analyte RI set as 1.42. In summarize, the
value of the proposed structure firstly rises and then declines dramatically when
t changes from 50 to 70 nm. As a brief conclusion from
Figure 10, the uprising value of thickness of gold coating causes the resulting
CL spectrum to red shift along with a relatively stable
CL curve intensity when analyte RI is 1.41 and a drastically attenuation of
CL peak value when analyte RI is 1.42.
Figure 11 illustrates the
curve of the
y-polarized mode with
(abbreviated as
l), respectively, chosen as 0.8, 1.0, 1.2 μm, and analyte RI is set as 1.41 and 1.42. From
Figure 11, when analyte RI is chosen as 1.41 and
t increases from 0.8 to 1.2 μm, the
value of
CL spectra has a wide-ranging red shift and the peak value of
CL curve climbs up dramatically. The
value of
CL spectra with analyte RI 1.42 witness a wide-ranging red shift and the peak value of
CL curve diminishes significantly. Moreover, it is obviously that the FWHM of the
CL spectra are significantly wider with
l increasing with analyte RI set as 1.42. Furthermore, when making a comparison among those
CL peak values and related envelopes, it can be deduced that with
l value increasing, the envelope of the
CL spectra in regard to changing analyte RI emerges to red-shift. A reasonable explanation for the phenomenon is when
l increases, the SPR effect is further excited, which strengthens the intensity of the peak value and mode field distributions, thus influence
y-polarized mode and lead to a red shift of
CL curve. In accordance with Equation (
4), the maximum
value achieves 11,100, 19,600, 18,500 nm/RIU with
l, respectively, chosen as 0.8, 1.0, 1.2 μm. The calculation results demonstrate the
value of the proposed structure firstly climbs up significantly and then decreases with
l changing from 0.8 to 1.2 μm. In accordance with Equation (
5), the maximum
value achieves about 6110, 2226.7, and 206.95 RIU
when
t is chosen as 0.8, 1.0, and 1.2 μm with analyte RI set as 1.42. In conclusion, the
value of the fiber structure witnesses a dramatic fall with
l changing from 0.8 to 1.2 μm and when
l value is set as an appropriate small value, the proposed structure obtains an superior sensing performance. As a brief conclusion from
Figure 11, the uprising value of length of gold film causes the resulting
CL spectrum to red shift along with a dramatically boost of
CL curve intensity when analyte RI is 1.41 and a significant fall of
CL peak value when analyte RI is 1.42.
Figure 12 demonstrates the
curve of the
y-polarized mode with
chosen as 1.8, 1.9, 2.0,
μm, and analyte RI set as 1.41 and 1.42. From
Figure 12, when
increases from 1.8 to 2.4 μm, the
value of the
spectra with an analyte RI chosen as 1.41 has a visible red-shift and the
curve peak value noticeably falls. The
value of the
spectra with an analyte RI set as 1.42 sees a wide-ranging red-shift and the
curve peak value gradually increasing. Furthermore, when making a comparison among the
peak values and related envelopes, it is deduced that when
increases, the envelope of the
curve in association with the changing analyte RI appears to red-shift. A reasonable explanation for this phenomenon is that with
increasing, the region of the
y-polarized mode becomes compact, impacting the
y-polarized mode and results in a
curve shift. In accordance with Equation (
4), the maximum
values achieved were 16,000, 13,800, 19,600, 14,250 nm/RIU with
set as 1.8, 1.9, 2.0, and
μm, respectively. The calculation results illustrate that the RI sensitivity of the sensor structure first increases and then decreases with a variation in
from 1.8 to 2.4 μm. In accordance with Equation (
5), the maximum
values achieved were 931.4, 1253.7, 2226.7 and 4400 RIU
when
was set as 1.8, 1.9, 2.0, and
μm, respectively, when the analyte RI was set as 1.42, showing that the
value of the sensor structure sharply increases when
changes from 1.8 to 2.4 μm, thus gaining an phenomenal sensing performance.
For the sensor structure in this work, the effects of the structural parameters on the sensing performance are summarized in the following
Table 1.