Next Article in Journal
Repeat Local Treatment of Recurrent Colorectal Liver Metastases, the Role of Neoadjuvant Chemotherapy: An Amsterdam Colorectal Liver Met Registry (AmCORE) Based Study
Next Article in Special Issue
Elucidating the Role of Extracellular Vesicles in Pancreatic Cancer
Previous Article in Journal
Delineating the Switch between Senescence and Apoptosis in Cervical Cancer Cells under Ciclopirox Treatment
Previous Article in Special Issue
Cell-Main Spectra Profile Screening Technique in Simulation of Circulating Tumour Cells Using MALDI-TOF Mass Spectrometry
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Systematic Review

The Evolving Role of Radiofrequency Guided Localisation in Breast Surgery: A Systematic Review

London Breast Institute, Princess Grace Hospital, Hospital Corporation of America (HCA), 42-52 Nottingham Place, London W1U 5NY, UK
*
Author to whom correspondence should be addressed.
Cancers 2021, 13(19), 4996; https://doi.org/10.3390/cancers13194996
Submission received: 20 June 2021 / Revised: 17 September 2021 / Accepted: 2 October 2021 / Published: 5 October 2021
(This article belongs to the Special Issue Emerging Technologies in Cancer Diagnostics and Therapeutics)

Abstract

:

Simple Summary

A large portion of breast lesions are not palpable and need to be marked before surgery in order to aid complete surgical removal. Currently, this is accomplished by placing a hook in the lesion with radiological guidance, which is in turn attached to a wire through the skin. This is a well-understood technique and is called wire-guided localisation (WGL). It has drawbacks, including being uncomfortable for the patient, and the need for the wire to be placed within less than a day of the surgery. This is why certain wireless techniques have been developed to replace WGL. LOCalizer™ is one such technique, which uses radio-frequency identification tags. In this study, we have systematically reviewed the literature regarding LOCalizer™, and confirmed that it is a valid alternative to WGL. We have also highlighted its limitations and suggested potential technical refinements to improve its clinical performance.

Abstract

Wire-guided localisation (WGL) has been the gold-standard for localising non-palpable breast lesions before excision. Due to its drawbacks, various wireless alternatives have been developed, including LOCalizer™, which is based on radio-frequency identification (RFID) technology. In this systematic review, we consulted EMBASE, Medline and PubMed databases using appropriate search terms regarding the use of RFID technology in the localisation of occult breast lesions. Retrospective and prospective studies were included if they quoted the number of patients, rate of successful placement, retrieval rate, margin positivity rate and the re-excision rate. In addition, studies comparing RFID to WGL were also included and analysed separately. Seven studies were included in this systematic review spanning 1151 patients and 1344 tags. The pooled deployment rate was 99.1% and retrieval rate was 100%. Re-excision rate was 13.9%. One complication was identified. Two studies compared RFID with WGL (128 vs. 282 patients respectively). For both techniques the re-excision rate was 15.6% (20/128 vs. 44/282 respectively, p value is 0.995). Based on our review, LOCalizer™ is safe and non-inferior to WGL in terms of successful localisation and re-excision rates. However, further research is required to assess the cost effectiveness of this approach and its impact on the aesthetic outcome compared with WGL and other wire free technologies to better inform decision making in service planning and provision.

1. Introduction

Non-palpable lesions form a large plurality of breast cancer cases treated every year. Recognition of the peculiar diagnostic challenge posed provided impetus to the development of mammography in the 1950s, which has made breast cancer screening possible [1,2]. Breast screening does lead to detection of non-palpable lesions at an earlier, more treatable stage, which would require preoperative localisation of the detected lesions [3].
The current gold standard, which has been prevalent since the 1970s, is the placement of hooks at the site of the lesion marked by wires protruding from the skin. This is commonly termed as wire-guided localisation (WGL) of the occult breast lesion [4,5,6].
The professional community at large has significant experience and familiarity with this technique, which is not to say that WGL is not without its drawbacks. It is uncomfortable for the patient. It carries a risk of needle-stick injury for both the surgeon and the radiologist. Dislodgement, fracturing and migration of the wire are all well-known possible issues which could arise. However, the most salient reason to find an alternative to WGL is the onerous scheduling requirement imposed on health systems by closely coupling radiological appointments with surgery slots [4,7,8].
Several wireless localisation techniques have been developed recently for the localisation of occult breast lesions. While they are still awaiting wider acceptance, their potential benefits in terms of patient satisfaction and reduction of strain on hospital scheduling systems could not be ignored.
We have recently investigated the clinical performance of currently available radiation-free wireless localisation technologies, specifically SAVI SCOUT (Merit Medical, Alieso Viejo, CA, USA) [9] and Magseed (Endomag Limited, Cambridge, UK) [10]. We have also systematically reviewed the literature regarding the currently available wireless localisation systems. SAVI SCOUT is based on reflectors using radar technology [11]. Magseed utilises ferromagnetic markers which are located by handheld proprietary magnetometers [12].
A further wireless localisation system utilises glass-enclosed radio-frequency identification (RFID) tags (LOCalizer™, Hologic Inc., Santa Clara, CA, USA). In this article, we present the findings of our systematic review of the literature regarding LOCalizer™, which shall guide our discussion regarding the merits and issues pertaining to the system, especially in view of the available alternatives. We focused in this review on three aspects: successful deployment of the tags, successful retrieval of the tags and re-excision rates.

2. Materials and Methods

2.1. Inclusion and Exclusion Criteria

To be included in this systematic review, a study needed to evaluate and report the findings on the use of RFID tags technology to localise non-palpable breast lesions in the abstract.
Only peer-reviewed published articles were included. Retrospective and prospective cohort studies were included. Publications needed to summarise findings when exploring the use of RFID technology to localise non-palpable breast lesions in the abstract. In the full text, the following raw data had to be included: total number of patients undergoing RFID tag localisation, successful placement/localisation of the RFID tags, successful identification/retrieval of the RFID tags and re-excision rates/margin positivity for cancer cases.
Publications comparing RFID tags to other localisation methods were also included. Data regarding other methods were ignored for the purposes of our calculations, except where relevant to the smaller pooled analysis. When available, data regarding re-excision rates were also included. If the publication detailed only margin positivity, this was assumed to indicate re-excision.

2.2. Data Sources and Searches

A computer-aided literature search using the EMBASE, Medline and PubMed databases was performed to identify relevant articles for inclusion in the study up to the 4 March 2021 with no lower limit. The following string was used for searching the aforementioned databases: ((breast AND radiofrequency) NOT ablation).
All the titles and abstracts of the studies resulting from the searches were reviewed and articles that were irrelevant were excluded. References of assessed full-text publications were also screened for any relevant publications, as well as previously published reviews. Attempts were made to complete the missing data by asking study authors for the necessary data via electronic mail.

2.3. Data Management

The authors extracted and combined data to calculate the overall rates of successful placement and retrieval from data sets of included studies. Some studies included patients who had multiple RFID tags placed for localisation. When no extra data were provided, it was assumed that the number of patients was equal to the number of tags placed.
Mean values were calculated by combining data sets from each included study to give overall rates for successful placement/localisation, retrieval and re-excision. Retrieval rate was calculated using tags successfully removed, whether they were placed accurately or not. Re-excision rate was computed using only cases which had malignancy in their preoperative biopsy or postoperative pathology and had successful tag placement.
A separate analysis was performed on the studies that directly compared the WGL with RFID localisation.

3. Results

3.1. Literature Search Results and Characteristics of the Included Studies

A total of 814 records were initially identified (239 from PubMed; 353 from EMBASE; 222 from Medline). After removing duplicates, 386 publications were initially assessed for inclusion. All the titles and abstracts were reviewed and articles that were irrelevant, such as studies on radiofrequency ultrasound and radiofrequency spectroscopy were excluded. Conference abstracts, reviews, studies reporting on RFID tags use in localisation of axillary lymph nodes only but not breast lesions were also excluded.
Full texts were then examined for the 10 abstracts, which met the inclusion criteria for our review [10,13,14,15,16,17,18,19,20,21]. One study (Tayeh et al.) [10] was excluded since its results were part of a larger study that is already included in our analysis [19]. Therefore, nine studies were used in the final analysis. Seven cohort studies used to calculate the rates of successful placement, successful retrieval and re-excision in the final pooled analysis [13,14,15,16,17,18,19].
Two studies compared RFID tags to WGL [20,21]. These were included in a smaller pooled analysis, which selectively investigated re-excision rates in the use of RFID tags in direct comparison to WGL. The results of these two studies were not included in the larger pooled analysis as the cases were from the same authors, institution or study period of two of the seven studies that were included in the larger pooled analysis (Figure 1).

3.2. Results of Pooled Analysis

Across the nine studies included in the overall pooled analysis, 1344 RFID tags were inserted in 1151 patients. Of these, 1332 were successfully placed and 1343 were successfully retrieved using RFID. This gives a successful deployment rate of 99.1% and a successful retrieval rate of 100%. There were 836 malignant cases with 117 cases requiring re-excision. The re-excision rate was therefore 13.9%. (Table 1) (Figure 2).
Across the two studies directly comparing RFID to wire-guided localisation, 128 RFID tag and 282 wires were inserted to localise malignant lesions. Of these, 20 RFID cases required re-excision compared to 44 wire guided localisation cases. This gives a similar re-excision rate of 15.6 for both techniques. (Chi-squared test, p = 0.9954) (Table 2)
Five studies reported no complications [14,15,17,18,19,22]. One study reported a postoperative haematoma which was managed conservatively [13].
Two studies reported on patient, surgeon and radiologist feedback [15,19]. Both studies reported positive feedback. DiNome et al. reported that most patients in their study agreed or strongly agreed (94%) that the procedure went smoothly and agreed or strongly agreed (78%) that the procedure was easier than expected. Radiologists and surgeons suggested that the RFID tag was as fast and reliable as the wire-localised procedure. Surgeons also generally agreed or strongly agreed that the distance gauge was helpful in guiding the surgical dissection [15]. Wazir and colleagues reported patient feedback obtained from seven patients in their study. They reported a mean satisfaction score of 9.9 out of 10 (range = 9–10) using a linear visual analogue scale. In the same study, both radiologists and surgeons rated the LOCalizer™ technique as better compared to wire-guided localization [15].

4. Discussion

The limitations of WGL highlighted earlier have given impetus to the development of alternative localisation techniques [23]. Early attempts towards development of an alternative to WGL involved the use of radio-active titanium seeds, which was termed radioactive seed localisation (RSL) [24,25,26]. This method was found to have a favourable learning curve and a low margin positivity rate [27,28]. However, wider acceptance of this modality was curtailed by the regulatory requirements for the handling of radioactive materials. Furthermore, in most jurisdictions, the radio-active markers could be safely retained for only five days, thus requiring coupling of surgical and radiological sessions [29,30].
The limitations of WGL and RSL have inspired the evolution of three wire-free and radiation-free localisation systems which vary on the basis of the underlying principles and technologies [31].
Magseed is based on the detection of a ferromagnetic seed by a proprietary handheld magnetometer called Sentimag which was originally developed for sentinel lymph node and occult breast lesion localisation in the MagSNOLL trial [32,33]. For the purpose of occult breast lesion localisation, a 5 mm seed is implanted using an 18 G introducer. This system was approved by the Food and Drug Administration (FDA) for implantation in 2016 [34,35]. This system compares favourably to WGL in terms of patient acceptability and margin positivity [36,37,38]. MAgnetic MArker LOCalisation (MaMaLoc) (Sirius Medical, Eindhoven, The Netherlands) is another technique based on magnetometry with limited clinical data [39,40,41].
However, magnetometry-centred systems do have some inherent limitations. The markers have been noted to leave significant void signals (>4 cm) in magnetic resonance imaging (MRI) scans, thus limiting their utility in patients who require MRI for their surveillance and diagnosis. Furthermore, the operative field needs to be cleared of ferromagnetic instruments during the use of the magnetometer as they would interfere with the localisation of the seed [12].
SAVI SCOUT is another FDA-approved breast lesion localisation system. This system makes use of proprietary reflectors which reflect electromagnetic signals emitted from and detected by a handheld detector [42,43,44,45]. It has been found to be at least non-inferior to WGL in terms of successful localisation and margin positivity rates. In addition, the device was reported to have minimal MRI void artefacts (<5 mm), in addition to good surgeon, patient and radiologist acceptability [46,47,48].
The LOCalizer™ system is based on ubiquitous RFID technology. It consists of a RFID tag with a unique identification number preloaded in a needle applicator, a surgical probe with 8 mm tip and a portable handheld reader. The tracer chip is encased in glass and is deployed using a 12 G introducer. The surgical probe is the size of a pencil, the reader is portable and displays the distance to the tag in millimetres along with the tag’s ID number (Figure 2). The device is certified for long term implantation. Arguably, using a widespread and well-understood technology would make LOCalizer™ a robust choice.
Our review of the literature identified 1344 devices implanted in 1151 patients. The pooled successful deployment and retrieval rates were 99% and 99.1%, respectively. We found two studies that compared LOCalizer™ with WGL, which found that both modalities had comparable positive margin rates and rates of re-excision. Therefore, it would be reasonable to assert non-inferiority of this modality compared to the current gold standard of WGL in terms of successful localisation and margin positivity rates with the added advantage of flexible scheduling. However, none of the primary studies included in our review was a randomised controlled trial. Moreover, every tag has a unique identification number which is particularly valuable when multiple markers are deployed for localisation.
However, the LOCalizer™ system has certain limitations worth highlighting. First, the tags cause void artefacts in MRI scans, albeit smaller than Magseed, which can have implications for post-deployment imaging in certain patients especially in the context of neoadjuvant systemic therapy (NST) [49]. Second, the detection range of the system is limited to 6 cm only, potentially causing issues in larger breasts and deeper lesions [50]. Third, the current LOCalizer™ introducer needle is wider than that of Magseed and Savi Scout. Therefore, deployment can be difficult, particularly in dense breast tissue [14,50]. Furthermore, the wide bore introducer needle (12 G, or 2.77 mm) often requires a skin incision for insertion and can create a wide track along which the glass-encased tag could migrate (Figure 3). Although studies have not reported significant migration before surgery, migration has been reported to occur during excision of the surgical specimen. Dauphine et al. observed migration in 3 (15%) of 20 cases, as the lesion was being retracted with fingers to make the final cut along the deep surface of the specimen [14]. Lamb et al. have also implicated the wide gauge introducer needle in the loss of titanium clip markers during tag deployment in some cases [16]. Finally, patients with cardiac pacemakers and defibrillators should be excluded as a precautionary measure since radiofrequency signals may interfere with the function of these devices [14].
Therefore, the LOCalizer™ technology can be further optimised in its design to improve clinical performance. We previously suggested that the bore of the introducer needle should be reduced to 16 gauge in order to facilitate deployment within dense tissue and reduce the probability of tag migration along the introducer needle track [19,21]. We also recommend that removal of the glass casing of the RFID tag should be considered since it contributes to its migratory potential and MRI void artefacts [14]. Furthermore, retention of glass fragments can occur in the case of tag fracture.
In addition to this review, we previously reported the results of a pooled analysis of all published studies that investigated the use of SAVI SCOUT [11] and Magseed [12] and reported an overall successful localisation rate of 99.64% and 99.86%, respectively. We found no significant difference in re-excision rates when comparing Magseed with WGL [10,12]. However, our pooled analysis of SAVI SCOUT performance reported that the use of the device was associated with a significant reduction in the re-excision rates compared with WGL in subgroup analysis [9,11,46]. However, this requires further validation in adequately powered prospective studies.
The strength of our review stems from the fact that it represents the first pooled analysis of all published studies reporting experience with the use of RFID technology for the localisation of more than 1000 non-palpable breast lesions thus providing an overview of its clinical performance. Our findings confirm the safety of this wire-free approach and will be helpful for the design of future prospective studies. We have also proposed certain technical modifications of the deployment system and the tag in order to optimise their clinical performance.
Our study has several limitations. Although we included nine studies in our systematic review, most of the data arose from a single retrospective analysis with paucity of data from prospective trials [16]. None of the studies was a prospective randomized trial. The number of patients in most of the primary studies was relatively low (<100). Only two studies included more than 100 patients [16,17]. Only one small study compared RFID with other wireless, radiation free technologies [20].
Furthermore, most of the cases included in our analysis had the radio-frequency tag deployed on the day of surgery, thus making it difficult to draw conclusions regarding the clinical performance of the technology with regards to the intended decoupling of surgical and radiological scheduling. This will also compromise the evaluation of tag migration over time which is an important consideration.
Moreover, the RFID tags were deployed using mammographic guidance to localise lesions visible on ultrasonography in a large number of cases [16], thus prolonging the duration of the localisation procedure and increasing radiation exposure.
The retrospective and heterogeneous nature of most primary studies in addition to lack of methodology standardisation precluded a meaningful analysis of important secondary outcomes such as the depth of the lesion from the skin surface, the weight of the surgical specimen, duration of procedure, the size of MRI void signals and aesthetic considerations.
Finally, none of the studies included have evaluated the cost-effectiveness of using this technology.

5. Conclusions

Our systematic review includes 1151 patients and 1344 RFID tags. The pooled deployment rate was 99.1% and retrieval rate was 100%. Re-excision rate was 13.9%. In studies comparing RFID with WGL, the re-excision rate was comparable.
In summary, our review confirms that the RFID-based LOCalizer™ system is a valid safe alternative to WGL in terms of successful localisation and margin positivity with the added advantage of decoupling surgery and radiology scheduling. However, the system requires technical refinements to optimise its clinical performance, and its impact on aesthetic outcome and healthcare economics should be evaluated in future prospective research.

Author Contributions

K.M. supervised the project and edited the manuscript. S.T. initiated the project, performed literature search, data collection and analysis and edited the manuscript. U.W. drafted the manuscript and performed the literature review. All authors have read and agreed to the published version of the manuscript.

Funding

The Breast Cancer Hope Foundation has provided an educational funding supporting the publication.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Publicly available datasets were analyzed in this study.

Conflicts of Interest

K.M. has received honoraria for providing academic and clinical advice to Merit Medical. S.T. and U.W. declared no conflict of interest.

References

  1. Egan, R.L. Experience with mammography in a tumor institution. Evaluation of 1000 studies. Radiology 1960, 75, 894–900. [Google Scholar] [CrossRef] [PubMed]
  2. Egan, R.L. Mammography and diseases of the breast. CA Cancer J. Clin. 1968, 18, 279–283. [Google Scholar] [CrossRef] [PubMed]
  3. Cady, B.; Stone, M.D.; Schuler, J.G.; Thakur, R.; Wanner, M.A.; Lavin, P.T. The new era in breast cancer. Invasion, size, and nodal involvement dramatically decreasing as a result of mammographic screening. Arch. Surg. 1996, 131, 301–308. [Google Scholar] [CrossRef] [PubMed]
  4. Cheang, E.; Ha, R.; Thornton, C.M.; Mango, V.L. Innovations in image-guided preoperative breast lesion localization. Br. J. Radiol. 2018, 91, 20170740. [Google Scholar] [CrossRef]
  5. Frank, H.A.; Hall, F.M.; Steer, M.L. Preoperative localization of nonpalpable breast lesions demonstrated by mammography. N. Engl. J. Med. 1976, 295, 259–260. [Google Scholar] [CrossRef]
  6. Blichert-Toft, M.; Dyreborg, U.; Bogh, L.; Kiaer, H. Nonpalpable breast lesions: Mammographic wire-guided biopsy and radiologic-histologic correlation. World J. Surg. 1982, 6, 119–125. [Google Scholar] [CrossRef] [PubMed]
  7. Chan, B.K.; Wiseberg-Firtell, J.A.; Jois, R.H.; Jensen, K.; Audisio, R.A. Localization techniques for guided surgical excision of non-palpable breast lesions. Cochrane Database Syst. Rev. 2015, 12, CD009206. [Google Scholar] [CrossRef]
  8. Tardioli, S.; Ballesio, L.; Gigli, S.; Di Pastena, F.; D’Orazi, V.; Giraldi, G.; Monti, M.; Amabile, M.I.; Pasta, V. Wire-guided Localization in Non-palpable Breast Cancer: Results from Monocentric Experience. Anticancer Res. 2016, 36, 2423–2427. [Google Scholar] [PubMed]
  9. Tayeh, S.; Muktar, S.; Heeney, J.; Michell, M.J.; Perry, N.; Suaris, T.; Evans, D.; Malhotra, A.; Mokbel, K. Reflector-guided Localization of Non-palpable Breast Lesions: The First Reported European Evaluation of the SAVI SCOUT(R) System. Anticancer Res. 2020, 40, 3915–3924. [Google Scholar] [CrossRef]
  10. Tayeh, S.; Gera, R.; Perry, N.; Michell, M.; Malhotra, A.; Mokbel, K. The Use of Magnetic Seeds and Radiofrequency Identifier Tags in Breast Surgery for Non-palpable Lesions. Anticancer Res. 2020, 40, 315–321. [Google Scholar] [CrossRef]
  11. Kasem, I.; Mokbel, K. Savi Scout(R) Radar Localisation of Non-palpable Breast Lesions: Systematic Review and Pooled Analysis of 842 Cases. Anticancer Res. 2020, 40, 3633–3643. [Google Scholar] [CrossRef] [PubMed]
  12. Gera, R.; Tayeh, S.; Al-Reefy, S.; Mokbel, K. Evolving Role of Magseed in Wireless Localization of Breast Lesions: Systematic Review and Pooled Analysis of 1559 Procedures. Anticancer Res. 2020, 40, 1809–1815. [Google Scholar] [CrossRef] [PubMed]
  13. Cullinane, C.M.; Byrne, J.; Akmenkalne, L.; DP, O.L.; Connors, A.M.; Corrigan, M.A.; Redmond, H.P.; Kelly, L.; O’Sullivan, M.J. The LOCalizer Radiofrequency Identification System: An Effective New Technology for Localizing Non-Palpable Breast Lesions for Surgery. Surg. Innov. 2021, 28, 473–478. [Google Scholar] [CrossRef]
  14. Dauphine, C.; Reicher, J.J.; Reicher, M.A.; Gondusky, C.; Khalkhali, I.; Kim, M. A prospective clinical study to evaluate the safety and performance of wireless localization of nonpalpable breast lesions using radiofrequency identification technology. AJR Am. J. Roentgenol. 2015, 204, W720–W723. [Google Scholar] [CrossRef] [Green Version]
  15. DiNome, M.L.; Kusske, A.M.; Attai, D.J.; Fischer, C.P.; Hoyt, A.C. Microchipping the breast: An effective new technology for localizing non-palpable breast lesions for surgery. Breast Cancer Res. Treat. 2019, 175, 165–170. [Google Scholar] [CrossRef]
  16. Lamb, L.R.; Gilman, L.; Specht, M.; D’Alessandro, H.A.; Miles, R.C.; Lehman, C.D. Retrospective Review of Preoperative Radiofrequency Tag Localization of Breast Lesions in 848 Patients. AJR Am. J. Roentgenol. 2021, 217, 605–612. [Google Scholar] [CrossRef]
  17. Lowes, S.; Bell, A.; Milligan, R.; Amonkar, S.; Leaver, A. Use of Hologic LOCalizer radiofrequency identification (RFID) tags to localise impalpable breast lesions and axillary nodes: Experience of the first 150 cases in a UK breast unit. Clin. Radiol. 2020, 75, 942–949. [Google Scholar] [CrossRef]
  18. Malter, W.; Holtschmidt, J.; Thangarajah, F.; Mallmann, P.; Krug, B.; Warm, M.; Eichler, C. First Reported Use of the Faxitron LOCalizer Radiofrequency Identification (RFID) System in Europe—A Feasibility Trial, Surgical Guide and Review for Non-palpable Breast Lesions. In Vivo 2019, 33, 1559–1564. [Google Scholar] [CrossRef] [Green Version]
  19. Wazir, U.; Tayeh, S.; Perry, N.; Michell, M.; Malhotra, A.; Mokbel, K. Wireless Breast Localization Using Radio-frequency Identification Tags: The First Reported European Experience in Breast Cancer. In Vivo 2020, 34, 233–238. [Google Scholar] [CrossRef]
  20. Lee, M.K.; Sanaiha, Y.; Kusske, A.M.; Thompson, C.K.; Attai, D.J.; Baker, J.L.; Fischer, C.P.; DiNome, M.L. A comparison of two non-radioactive alternatives to wire for the localization of non-palpable breast cancers. Breast Cancer Res. Treat. 2020, 182, 299–303. [Google Scholar] [CrossRef] [PubMed]
  21. McGugin, C.; Spivey, T.; Coopey, S.; Smith, B.; Kelly, B.; Gadd, M.; Hughes, K.; Dontchos, B.; Specht, M. Radiofrequency identification tag localization is comparable to wire localization for non-palpable breast lesions. Breast Cancer Res. Treat. 2019, 177, 735–739. [Google Scholar] [CrossRef] [PubMed]
  22. Malter, W.; Eichler, C.; Hanstein, B.; Mallmann, P.; Holtschmidt, J. First Reported Use of Radiofrequency Identification (RFID) Technique for Targeted Excision of Suspicious Axillary Lymph Nodes in Early Stage Breast Cancer—Evaluation of Feasibility and Review of Current Recommendations. In Vivo 2020, 34, 1207–1213. [Google Scholar] [CrossRef]
  23. Lovrics, P.J.; Cornacchi, S.D.; Vora, R.; Goldsmith, C.H.; Kahnamoui, K. Systematic review of radioguided surgery for non-palpable breast cancer. Eur. J. Surg. Oncol. 2011, 37, 388–397. [Google Scholar] [CrossRef]
  24. Dauway, E.; Saunders, R.; Friedland, J. Innovative diagnostics for breast cancer: New frontiers for the new millennium using radioactive seed localization. In 85th Annual American College of Surgeons Clinic Congress; American College of Surgeons: Chicago, IL, USA, 1999; Volume 50. [Google Scholar]
  25. Gray, R.J.; Salud, C.; Nguyen, K.; Dauway, E.; Friedland, J.; Berman, C.; Peltz, E.; Whitehead, G.; Cox, C.E. Randomized prospective evaluation of a novel technique for biopsy or lumpectomy of nonpalpable breast lesions: Radioactive seed versus wire localization. Ann. Surg. Oncol. 2001, 8, 711–715. [Google Scholar] [CrossRef] [PubMed]
  26. Zhang, Y.; Seely, J.; Cordeiro, E.; Hefler, J.; Thavorn, K.; Mahajan, M.; Domina, S.; Aro, J.; Ibrahim, A.M.; Arnaout, A.; et al. Radioactive Seed Localization Versus Wire-Guided Localization for Nonpalpable Breast Cancer: A Cost and Operating Room Efficiency Analysis. Ann. Surg. Oncol. 2017, 24, 3567–3573. [Google Scholar] [CrossRef]
  27. Velazco, C.S.; Wasif, N.; Pockaj, B.A.; Gray, R.J. Radioactive seed localization for breast conservation surgery: Low positive margin rate with no learning curve. Am. J. Surg. 2017, 214, 1091–1093. [Google Scholar] [CrossRef]
  28. Wang, G.L.; Tsikouras, P.; Zuo, H.Q.; Huang, M.Q.; Peng, L.; Bothou, A.; Zervoudis, S.; Tobias Teichmann, A. Radioactive seed localization and wire guided localization in breast cancer: A systematic review and meta-analysis. J. BUON 2019, 24, 48–60. [Google Scholar] [PubMed]
  29. Jakub, J.; Gray, R. Starting a Radioactive Seed Localization Program. Ann. Surg. Oncol. 2015, 22, 3197–3202. [Google Scholar] [CrossRef]
  30. Goudreau, S.H.; Joseph, J.P.; Seiler, S.J. Preoperative Radioactive Seed Localization for Nonpalpable Breast Lesions: Technique, Pitfalls, and Solutions. Radiographics 2015, 35, 1319–1334. [Google Scholar] [CrossRef] [Green Version]
  31. Kiruparan, N.; Kiruparan, P.; Debnath, D. Use of wire guided localisation and radio-guided occult lesion localisation for non-palpable breast lesions: A systematic literature review and meta-analysis of current evidence. Asian J. Surg. 2021. [Google Scholar] [CrossRef]
  32. Ahmed, M.; Anninga, B.; Goyal, S.; Young, P.; Pankhurst, Q.A.; Douek, M.; Mag, S.T.G. Magnetic sentinel node and occult lesion localization in breast cancer (MagSNOLL Trial). Br. J. Surg. 2015, 102, 646–652. [Google Scholar] [CrossRef] [PubMed]
  33. Pieszko, K.; Wichtowski, M.; Cieciorowski, M.; Jamont, R.; Murawa, D. Evaluation of the nonradioactive inducible magnetic seed system Magseed for preoperative localization of nonpalpable breast lesions—Initial clinical experience. Contemp. Oncol. 2020, 24, 51–54. [Google Scholar] [CrossRef] [PubMed]
  34. Hayes, M.K. Update on Preoperative Breast Localization. Radiol. Clin. North. Am. 2017, 55, 591–603. [Google Scholar] [CrossRef] [PubMed]
  35. Singh, P.; Scoggins, M.E.; Sahin, A.A.; Hwang, R.F.; Kuerer, H.M.; Caudle, A.S.; Mittendorf, E.A.; Thompson, A.M.; Bedrosian, I.; Teshome, M.; et al. Effectiveness and Safety of Magseed-localization for Excision of Breast Lesions: A Prospective, Phase IV Trial. Ann. Surg. Open 2020, 1, e008. [Google Scholar] [CrossRef]
  36. Zacharioudakis, K.; Down, S.; Bholah, Z.; Lee, S.; Khan, T.; Maxwell, A.J.; Howe, M.; Harvey, J. Is the future magnetic? Magseed localisation for non palpable breast cancer. A multi-centre non randomised control study. Eur. J. Surg. Oncol. 2019, 45, 2016–2021. [Google Scholar] [CrossRef] [Green Version]
  37. Micha, A.E.; Sinnett, V.; Downey, K.; Allen, S.; Bishop, B.; Hector, L.R.; Patrick, E.P.; Edmonds, R.; Barry, P.A.; Krupa, K.D.C.; et al. Patient and clinician satisfaction and clinical outcomes of Magseed compared with wire-guided localisation for impalpable breast lesions. Breast Cancer 2021, 28, 196–205. [Google Scholar] [CrossRef]
  38. Harvey, J.R.; Lim, Y.; Murphy, J.; Howe, M.; Morris, J.; Goyal, A.; Maxwell, A.J. Safety and feasibility of breast lesion localization using magnetic seeds (Magseed): A multi-centre, open-label cohort study. Breast Cancer Res. Treat. 2018, 169, 531–536. [Google Scholar] [CrossRef] [Green Version]
  39. Schermers, B.; van der Hage, J.A.; Loo, C.E.; Vrancken Peeters, M.; Winter-Warnars, H.A.O.; van Duijnhoven, F.; Ten Haken, B.; Muller, S.H.; Ruers, T.J.M. Feasibility of magnetic marker localisation for non-palpable breast cancer. Breast 2017, 33, 50–56. [Google Scholar] [CrossRef]
  40. Struik, G.M.; Schermers, B.; Mares, I.; Lont, H.E.; Bradshaw, J.W.; Ten Haken, B.; Ruers, T.J.M.; Mourik, J.E.M.; Birnie, E.; Klem, T. Randomized controlled trial comparing magnetic marker localization (MaMaLoc) with wire-guided localization in the treatment of early-stage breast cancer. Breast J. 2021, 27, 638–650. [Google Scholar] [CrossRef]
  41. Yeung, S.T.S.; Farrah, K. Magnetic Seed Localization for Soft Tissue Lesions in Breast Patients: Clinical Effectiveness, Cost-Effectiveness, and Guidelines; Canadian Agency for Drugs and Technologies in Health: Ottawa, ON, Canada, 2019. [Google Scholar]
  42. Cox, C.E.; Russell, S.; Prowler, V.; Carter, E.; Beard, A.; Mehindru, A.; Blumencranz, P.; Allen, K.; Portillo, M.; Whitworth, P.; et al. A Prospective, Single Arm, Multi-site, Clinical Evaluation of a Nonradioactive Surgical Guidance Technology for the Location of Nonpalpable Breast Lesions during Excision. Ann. Surg. Oncol. 2016, 23, 3168–3174. [Google Scholar] [CrossRef]
  43. Cox, C.E.; Garcia-Henriquez, N.; Glancy, M.J.; Whitworth, P.; Cox, J.M.; Themar-Geck, M.; Prati, R.; Jung, M.; Russell, S.; Appleton, K.; et al. Pilot Study of a New Nonradioactive Surgical Guidance Technology for Locating Nonpalpable Breast Lesions. Ann. Surg. Oncol. 2016, 23, 1824–1830. [Google Scholar] [CrossRef]
  44. Falcon, S.; Weinfurtner, R.J.; Mooney, B.; Niell, B.L. SAVI SCOUT(R) localization of breast lesions as a practical alternative to wires: Outcomes and suggestions for trouble-shooting. Clin. Imaging 2018, 52, 280–286. [Google Scholar] [CrossRef] [PubMed]
  45. Cornella, K.N.; Palafox, B.A.; Razavi, M.K.; Loh, C.T.; Markle, K.M.; Openshaw, L.E. SAVI SCOUT as a Novel Localization and Surgical Navigation System for More Accurate Localization and Resection of Pulmonary Nodules. Surg. Innov. 2019, 26, 469–472. [Google Scholar] [CrossRef]
  46. Wazir, U.; Kasem, I.; Michell, M.J.; Suaris, T.; Evans, D.; Malhotra, A.; Mokbel, K. Reflector-Guided Localisation of Non-Palpable Breast Lesions: A Prospective Evaluation of the SAVI SCOUT® System. Cancers 2021, 13, 2409. [Google Scholar] [CrossRef]
  47. Srour, M.K.; Kim, S.; Amersi, F.; Giuliano, A.E.; Chung, A. Comparison of wire localization, radioactive seed, and Savi scout((R)) radar for management of surgical breast disease. Breast J. 2019, 26, 406–413. [Google Scholar] [CrossRef]
  48. Kalambo, M.; Parikh, J.R. Implementing the SAVI SCOUT System in Community Radiology Practice. J. Am. Coll. Radiol. 2017, 14, 1234–1238. [Google Scholar] [CrossRef] [PubMed]
  49. Reicher, J.; Reicher, D.; Reicher, M. Use of radio frequency identification (RFID) tags in bedside monitoring of endotracheal tube position. J. Clin. Monit. Comput. 2007, 21, 155–158. [Google Scholar] [CrossRef] [PubMed]
  50. LOCalizer™ Wire-free Guidance System. Available online: https://www.hologic.com/hologic-products/breast-health-solutions/localizer-wire-free-guidance-system (accessed on 17 June 2021).
Figure 1. PRISMA flow-diagram illustrating the inclusion and exclusion of studies reviewed for this study.
Figure 1. PRISMA flow-diagram illustrating the inclusion and exclusion of studies reviewed for this study.
Cancers 13 04996 g001
Figure 2. LOCalizer™ handheld reader device being used intra-operatively to locate the RFID tag implanted at the tumour site in the axillary tail of the left breast during surgery. RFID: radio-frequency identification.
Figure 2. LOCalizer™ handheld reader device being used intra-operatively to locate the RFID tag implanted at the tumour site in the axillary tail of the left breast during surgery. RFID: radio-frequency identification.
Cancers 13 04996 g002
Figure 3. (A) A control craniocaudal mammography film showing the radio-frequency tag within the tumour. (B) A control oblique mammography film showing the radio-frequency tag within the tumour. (C) A specimen radiograph demonstrating that the radio-frequency tag had migrated by 1 cm along the needle introducer track. RCC: right craniocaudal; RML: right medialateral.
Figure 3. (A) A control craniocaudal mammography film showing the radio-frequency tag within the tumour. (B) A control oblique mammography film showing the radio-frequency tag within the tumour. (C) A specimen radiograph demonstrating that the radio-frequency tag had migrated by 1 cm along the needle introducer track. RCC: right craniocaudal; RML: right medialateral.
Cancers 13 04996 g003
Table 1. Details of studies included in pooled analysis. CI: confidence intervals; RFID: radiofrequency identification.
Table 1. Details of studies included in pooled analysis. CI: confidence intervals; RFID: radiofrequency identification.
ReferencesAuthorCountryYear of PublicationNumber of PatientsNumber of Patients with Breast CancerNumber of RFID Tags DeployedNumber of RFID Tags Successfully Deployed (Successful Deployment Rate)Number of RFID Tags Successfully Retrieved Surgically (Retrieval Rate)Number of Patients Requiring Re-Excision/Number of Patients with Breast Cancer (Re-Excision Rate)
[9]CullinaneIreland202069636969 (100%)69 (100%)12/63 (19%)
[10]DauphineUSA201520152020 (100%)20 (100%)4/15 (26.7%)
[11]DiNomeUSA2019.50335050 (100%)50 (100%)2/33 (6.1%)
[12]LambUSA202084856810131004 (99.1%)1012 (100%) *86/568 (15.1%)
[13]LowesUK2020150150177174 (98%)177/177 (100%)13/150 (8.7%)
[14]MalterGermany20194044 (100%)4 (100%)All benign
[15]WazirUK20201071111 (100%)11 (100%)0/7 patients (0%)
Total 115183613441332/1344(99.11%)
95% CI: 98.45–99.54
1343/1343 (100%)
95% CI: 99.73–100.00
117/836 (13.9%)
95% CI: 11.71–16.53
* One patient with one tag did not have surgery due to metastatic disease (information from the author’s personal communication).
Table 2. Details of studies included in pooled analysis comparing re-excision rate between radio-frequency identification (RFID) tag localisation and wire-guided localisation (WGL). (Chi-squared test, p = 0.9954).
Table 2. Details of studies included in pooled analysis comparing re-excision rate between radio-frequency identification (RFID) tag localisation and wire-guided localisation (WGL). (Chi-squared test, p = 0.9954).
ReferencesAuthorYear of PublicationNumber of Patients with Breast Cancer Undergoing RFID Tag LocalisationNumber of Patients with Breast Cancer Undergoing RFID Tag Localisation Requiring Re-Excision (Re-Excision Rate)Number of Patients with Breast Cancer Undergoing WGLNumber of Patients Undergoing WGL Requiring Re-Excision (Re-Excision Rate)
[16]Lee2020332 (6.1%)505 (10%)
[17]McGugin20199518 (18.9%)23239 (16.8)
Total 12820 (15.6%)
95% CI: 11.57–20.37
28244 (15.6%)
95% CI: 9.81–23.10
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Tayeh, S.; Wazir, U.; Mokbel, K. The Evolving Role of Radiofrequency Guided Localisation in Breast Surgery: A Systematic Review. Cancers 2021, 13, 4996. https://doi.org/10.3390/cancers13194996

AMA Style

Tayeh S, Wazir U, Mokbel K. The Evolving Role of Radiofrequency Guided Localisation in Breast Surgery: A Systematic Review. Cancers. 2021; 13(19):4996. https://doi.org/10.3390/cancers13194996

Chicago/Turabian Style

Tayeh, Salim, Umar Wazir, and Kefah Mokbel. 2021. "The Evolving Role of Radiofrequency Guided Localisation in Breast Surgery: A Systematic Review" Cancers 13, no. 19: 4996. https://doi.org/10.3390/cancers13194996

APA Style

Tayeh, S., Wazir, U., & Mokbel, K. (2021). The Evolving Role of Radiofrequency Guided Localisation in Breast Surgery: A Systematic Review. Cancers, 13(19), 4996. https://doi.org/10.3390/cancers13194996

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop