This section brings together all of the pieces, where the theory is applied to the data and the results interpreted.
5.1. Photomechanical Response
Comparisons of light-induced stress and absorbance in DR1- and DO11-doped PMMA thin films can be used to deduce the mechanisms of the photomechanical response in photoisomerizable samples. The entire area of the sample is illuminated by the pump in a 15 s on and 15 s off cycle. After multiple repetitions, the intensity is increased and the protocol repeated. Typically, the experiment cycles through fifteen different intensities.
As shown in
Figure 2, the photomechanical stress response as a function of time for fixed pump intensity follows a bi-exponential function of the form given by Equations (
2) and (
3). The data are fit to the model with a time offset of
to determine the time constants and amplitudes.
One parameter of interest is the long-time stress response
, where
and
come from the fits to the data to Equations (
2) and (
3) at each intensity. These are plotted in
Figure 6. The long-term stress response of DR1 and DO11 is the same within experimental uncertainties at each intensity.
However, this does not mean that of DR1 is the same as of DO11 and that of DR1 is the same as of DO11, just that the sum is the same. As such, the heating contribution to the photomechanical response may be drastically different in the two materials even though their photomechanical response is the same.
The data for each polarization are fitted to a quadratic polynomial which is in the form of Equation (
1), and it shows that
I is large compared to
over the data range, so the stress response is linear within fractional deviations of 0.02 and 0.13 for DR1 and 0.05 and 0.12 for DO11 with horizontal and vertical polarizations of light, respectively, where the fractional deviation is given by
at the highest measured intensity.
The linear photomechanical constant is the same within experimental uncertainty for both pump polarizations, i.e., , in DR1 and in DO11. This observation leads to two possible hypotheses.
Hypothesis 1. This observation is as expected in DO11, which has only one conformation and so it should be dominated by photothermal heating alone, a process that is polarization-independent. However, since DR1 is known to photoisomerize [1,4], which leads to angular hole burning [1,4,28,29,30], the data suggest that the light-induced change in the molecular orientational distribution function does not translate into a large enough photomechanical response to be detectable within experimental uncertainties relative to the dominating photothermal heating mechanism. In addition, the magnitudes and of DO11 and DR1 agree with each other within experimental uncertainty, strengthening the conclusion that the mechanism of the response is the same in both materials. However, as seen in Figure 6, the horizontal polarization data for DR1 dye is systematically lower than the vertical polarization data and at the edge of experimental uncertainty, allowing for a small contribution from a mechanism that is not found in DO11. This could originate from photoisomerization. Hypothesis 2. Heating and other mechanisms are contributing to the DR1-doped polymers and the magnitude of the heating effect is different in each material so that the individual mechanics contribute differently but in aggregate give the same total response observed in Figure 6. Testing these hypotheses requires that each mechanism be isolated. The sections below describe how one isolates each mechanism by controlling individual experimental parameters such as temperature and intensity while monitoring the stress and transmitted probe intensity.
5.1.1. Heating
Thermally induced stress of clamped DR1 and DO11 thin films is studied by placing the samples in an oven with no light applied and measuring the stress as a function of temperature from 291.15 K to 307.15 K over a time span of 3 min. The result is shown in
Figure 3. The data are fit to Equation (
4) and the fitting parameters are shown in
Table 1.
Equation (
9) has the parameters
n,
and
, which are determined from fits to Equation (
4) to temperature-dependent stress data as shown in
Figure 3.
T is the temperature of the sample prior to illumination and
is the time constant of heating, which can be determined theoretically from the geometry of the material or empirically from the measured stress as a function of time [
32].
The time constant can depend on the intensity of the pump beam, which gives its dependence on intensity.The large difference between DR11- and DO11-doped films in
Figure 3 implies that, for the same amount of absorbed light energy, the photothermal mechanism in DR1-doped polymers will be larger than in DO11-doped polymers and that the other mechanisms, when added to heating, makes the total response similar in the two materials. This supports the second hypothesis.
Determining the cause for this large difference is beyond the scope of this paper but is a worthy goal for future studies. There are potentially multiple causes for this difference including the number of internal degrees of freedom added by the dopant (isomerization in DR1 and tautomerization in DO11), which can be modeled with statistical mechanical approaches, the efficiency of heat transfer to the polymer relative to the other competing processes such as molecular shape changes and plasticization due to the dyes. Combinations of these processes working together could magnify the effects of each.
5.1.2. Hole Burning and Photo-Stress Anisotropy
Table 2 summarizes the parameters determined from fits to the data shown in
Figure 6 for thin films. For comparison, also shown are the values from the literature for dye-doped fibers [
2]. The effect of dye-doping and annealing of the photomechanical effect in fibers has also been studied [
33]. The positive sign of
in both materials and in both thin films and fibers indicates that photothermal heating drives expansion. The ratio of the hole burning to the heating response,
R, vanishes within experimental uncertainty, confirming that photothermal heating dominates in thin films and fibers. In DO11
and
are expected to be zero, because DO11 is nonisomerizable, and the results show this to be true with a high degree of confidence. So, we conclude that angular hole burning and molecular reorientation are absent in DO11. DR1 shows a larger contribution from angular hole burning, and the signs of the coefficients are what is expected but vanish within experimental uncertainty. The sign of
indicates that the length of the sample along the light’s polarization decreases and the positive sign of
shows that the length of the sample increases along its long axis when it is illuminated by light perpendicular to it, refs. [
1,
2], as would be expected. But again, the values are statistically insignificant, so angular reorientation and hole burning are negligible.
To make sure that the response is linear and we are in the low-intensity regime, the ratio
R as a function of intensity is also calculated for both samples.
R is found using Equation (
26).
Figure 7 shows a plot of the relative reorientational fraction
R as a function of intensity, where we have evaluated
and
in Equation (
26) using the values of
,
,
, and
determined from the slopes in
Figure 6 to obtain the points at each intensity
I. The error bars are propagated from the measured uncertainties in
,
,
, and
. The ratio varies more over the intensity range measured for DR1 than for DO11. The ratio
R for DR1 appears to change sign at low intensity, but this sign change is in the noise given the uncertainties. The increase in
R for DR1 is statistically significant, suggesting that hole burning becomes a significant fraction of the response mechanisms at the highest intensities.
While the behavior of the thin films and fibers is similar, the linear photomechanical constants are a factor of three to four times larger in the thin films.
5.2. Evolution of Molecular Orientation
Section 5.1 describes how the photomechanical effect depends on the various mechanisms. Here we focus on measurements that directly determine the orientational order of the dopant molecules in the PMMA polymer host, which can be correlated with the photomechanical response to assess if the orientational order couples mechanically to the polymer and that it contributes to photomechanical actuation.
The transmitted and incident intensities are recorded using a power meter with and without a sample. From these, the optical absorption coefficient of the sample is determined for the two orthogonal polarizations using the Beer–Lambert law [
1,
2]
where
and
and
are the absorbance parallel and perpendicular to the polarization of the pump beam, respectively.
Figure 8 (top) shows the absorbance for vertically and horizontally polarized probe beams in DR1- and DO11-doped PMMA thin films.
Figure 8 (bottom) shows a close-up view of one cycle in a region where transients from the probe lasers being turned on have subsided. For DR1, the vertically polarized pump induces a fast increase in the intensities of both probe polarizations as shown by the decreased absorbance. At later times with the pump on, the intensity of the horizontally polarized probe beam decreases (increased absorbance), while the vertically polarized probe beam shows little change.
This fast behavior can be understood as follows. Irradiating DR1 molecules with vertically polarized light preferentially excites the vertically oriented chromatophores. In the argument here, we assume that DR1 absorbs less light in its excited state and the molecules are one-dimensional, which has been shown to be a good approximation [
34]. The probability of exciting a chromophore that is oriented an angle
relative to the pump polarization is proportional to
, making it most probably to excite vertically oriented molecules at
and improbably in the horizontal direction, where
. The probability of a molecule absorbing vertically polarized probe light is proportional to
, making the probability of absorbing vertically polarized probe light in the presence of the vertically polarized pump beam is proportional to
. Similarly, since the probability of absorbing horizontally polarized light is proportional to
, the probability of absorbing horizontally polarized probe light in the presence of a vertically polarized pump is proportional to
.
Finally, assuming that the sample is isotropic, which should be based on the material’s processing conditions, the ratio of the change in absorbance for the two polarizations for an initially isotropic sample is given by the orientational average ratio
Equation (
31) implies that the drop in the vertically polarized probe is greater than the horizontal one and that they both should drop [
1,
2].
Figure 8 (bottom) shows this to be true for DR1. DO11, on the other hand, also shows a drop in the absorbance for both polarizations, but the one-dimensional approximation does not hold as well for DO11. In both cases, we see a drop in the absorbance due to the pump light exciting the molecules. This effect is labeled “hole burning” because the orientational distribution function takes on a shape like the dimple in an apple. Note that the probe beam intensities are much lower than the pump, so their affect on the excited populations should be minimal.
Another fast process is photothermal heating, which increases the temperature of the material as the absorbed light is converted to heat. Since the trans isomer is of lower energy than the cis isomer, the trans population drops and the cis population increases at higher temperatures, leading to an isotropic decrease in the absorbance of light by virtue of the cis molecule’s slow absorption cross-section. Thus isotropic heating will lower the absorbance of both polarizations equally. The fast component will thus be a combination of both heating and hole burning.
At longer time scales, molecules that undergo photoisomerization will reorient into the horizontal direction, as we would expect; when molecules aligned along the light’s polarization turn into the smaller cis state, they reorient randomly due to diffusion and on average end up in the perpendicular orientation once they decay back to the trans state [
2,
4]. This decreases the vertical absorbance because fewer molecules are aligned with the pump beam and increases the horizontal absorbance where there are more molecules. The horizontal probe sees an increase in absorbance as we would expect. Similarly, the long-time vertical probe sees a decrease in the absorbance, as expected. At an intermediate time between the fast hole-burning process and the slow molecular reorientation process, another feature is observed where both probe polarizations show an increase in absorbance. The region’s time scale includes the photothermal heating mechanism, which can change the relative populations between the cis and trans isomers. Separating the mechanisms to this level of detail is beyond the scope of this work, but it is worth mentioning that the three mechanisms can couple to produce complex behavior, which is difficult to model.
DO11 thin films show a decrease in absorption for both polarizations in response to the pump and an increase when the pump is turned off. The DO11 molecule is known to transition from the keto to the enol state when excited by a photon. Optical absorption of the enol state is weaker, so the absorbance decreases when the pump is on. The DO11 molecule relaxes back to the keto state when the pump is turned off [
26,
35], leading to an increase in the absorbance.
In both DR1 and DO11 molecules, the time dependence of the absorbance when light is turned off is the reverse of the process when the light is turned on. In the case of DR1, where there are multiple exponential processes that contribute amplitudes of opposite signs, the process is left–right anti-asymmetric, while the DO11 process is left–right symmetric. Note that for both cases, flipping the pump-off absorbance data about the horizontal axis and shifting them by 15 s gives the same qualitative shape.
The difference in the absorbance measured by the two orthogonal probe lasers can be used to find the axial orientational order parameter
. For molecules that are approximately one-dimensional, the order parameter is given by [
26,
30]
The order parameter if the long axes of the molecules are fully aligned and when they are randomly aligned. The order parameter can be negative when the molecules lie flat in the plane perpendicular to the symmetry axis.
The order parameter relative to the sample’s long axis, mounted vertically, as a function of time and intensity is shown in
Figure 9 for vertically and horizontally polarized pump light. The top plot references the order parameter to the long axis of the sample while the bottom plot is relative to the pump axis.
Order parameter relative to long axis of sample: We first consider the top plot of
Figure 9, which references the order parameter to the long axis of the thin-film sample. The long axis is always mounted vertically. Turning the pump on and off for both polarizations does not affect the order parameter in DO11—as expected when photo-isomerization is absent. In contrast, for DR1, a
vertically polarized pump decreases the order parameter along the film’s long axis (vertical) and turning the pump off results in an increase in this order parameter back to its pre-pumped state. A
horizontally polarized pump increases the order parameter, which decreases reversibly when the pump is turned off. This behavior is expected if the pump light that is polarized along the trans molecule’s long axis excites the cis state, making a “hole” along the light’s polarization direction by depleting those trans molecules. Subsequently, the number of molecules in the perpendicular orientation increases as the cis molecules relax back to the trans state and take on random orientation but leave a net depletion along the pump’s polarization.
Order parameter relative to pump polarization axis: The bottom part of
Figure 9 shows the order parameter relative to the pump’s polarization axis. For an isotropic sample, the order parameter in a dark sample should vanish and the order parameter in the presence of pump light should be the same for both polarizations if their intensities are the same. The data show that the order parameters are not the same for the two polarizations, so the samples are not isotropic. There are several possible sources of this anisotropy, including the fact that the samples are slightly stretched in the vertical direction when mounted in the sample holder. This is consistent with the observation that a vertical pump polarization has a slight positive order parameter, while the horizontal one has a negative order parameter. In both cases, the presence of the pump leads to a decrease in the order parameter, as expected for molecular reorientation.
Figure 10 shows the same data as in
Figure 9, but as a function of pump power. Because the samples are slightly anisotropic, the behavior is made clearer by considering the change in the order parameter relative to the dark state. This change in the order parameter
is given by
where
is the order parameter when the pump intensity is
I and
is the order parameter of the material prior to the start of the experiment. We call
the total order parameter change because it includes the change in order parameter due to the instantaneous pump power, as well as the accumulated order parameter from all past exposure due to the fact that the dye molecules have not had the chance to relax to the isotropic state during the 14 s period that the pump light is turned off. The low levels of probe light might also add to the cumulative effect. The values of
at
include all the values determined when the pump light is off between measurements where the intensity is changed. If there were no cumulative effects of time as the pump intensity is increased, there would be only one point for each polarization and each material.
Figure 11 shows the order parameter change
as a function of absorbed intensity relative to the sample’s long axis (top) and the pump axis (bottom). For DO11-doped PMMA, the order parameter referenced to the long axis of the sample (top plot) shows no statistically significant change as a function of intensity, while DR1-doped PMMA shows a dramatic effect with the horizontal pump inducing an increase in order parameter and the vertical pump inducing a decrease. The decrease in order parameter for the vertical pump is of slightly larger magnitude than the increase induced by the horizontal pump. These observations are consistent with molecules reorientating away from the light’s polarization and ending up in the plane perpendicular to the pump.
The order parameter referenced to the pump axis (bottom plot) shows a larger difference between the two pump polarizations for DO11 and a similar difference for DR1. But, the trend is the same; the order parameter drops in the direction of the pump polarization axis. The observation of a pump polarization-dependent change in the parameter suggests that the sample might be initially anisotropic. The change in the order parameter appears to saturate at higher intensities, which one would expect for a fixed population of molecules.
Next, we seek to understand the source of the anisotropy. The two obvious candidates are (1) anisotropy due to the slight stretching of the sample along its long axis from mounting strain and (2) long-time molecular reorientation that builds over time if the dyes in the polymer do not have enough time to relax over the 15 s dark cycle when the pump is off. There may also be an effect due to the excitation of molecules due to the ever-present weak probe beams, whose small effects may accumulate over time.
First, we need to determine the change in order parameter due to the current pump cycle. This is best characterized by the short-time order parameter change, defined as the fractional change in the order parameter, which is normalized to the order parameter just before the pump is turned on during each cycle,
(n), given by
where
n labels the cycle number.
Figure 12 shows the short-time fractional order parameter change as a function of intensity relative to the sample’s long axis (top) and relative to the pump’s polarization axis for both vertically and horizontally polarized pumps. DO11 shows the same fractional order parameter change for both pump polarizations, which is consistent with the absence of molecular reorientation. DR1, on the other hand, is consistent with molecular reorientation in that the order parameter change relative to the long axis for a vertically polarized pump is greater than for a horizontally polarized pump and of opposite sign. In the low-intensity regime, the ratio is about −2/3, as predicted by Equation (
31). The ratio becomes smaller at higher intensities, which is an indication that the sample is changing over time due to pump cycling and the effects of the probe beams. This does not preclude an intensity-dependent effect that originates in nonlinearities in the underlying mechanisms. Note that extrapolating the plots to zero intensity suggests clustering around a vanishing fractional order parameter change.
The long-term effects of the light on the sample can be assessed by comparing its properties at a given cycle with the sample prior to the first cycle. To this end, we define the long-term fractional change in the order parameter in the
cycle by referencing it to the order parameter of the sample at the start of the measurement. We start by determining the fractional change in the order parameter when the pump is off after the
cycle, given by
This will determine the change in the dark state of the sample in the presence of only the weak probe lasers. Doing the same for when the pump is on during the
cycle, we obtain
The long-term fractional change in the order parameter as a function of intensity when the pump is off and the pump is on relative to the pristine sample are shown in
Figure 13 and
Figure 14. The data clearly show that the dark state of the material (when pump is off, but the probes are on), over the long-term span of the experiment, evolves over time. The fractional change in the vertically referenced order parameter in DR1 decreases over time (and with increased pump intensity). This change is most likely a combination of the sample’s properties changing over time under the influence of the pump and probe beams, as well as the increased pump intensities, which might produce a longer-lasting change in the material properties. Similarly, when the pump is on, the vertically referenced fractional order parameters decreases over time, but by a greater factor than the dark state. This is attributable to the fact that in the pumped state, the fractional change in the order parameter is influenced by both the long-term changes in the sample and the short-term effects of the pump.
Consider that the fast response shows an increase in the fractional order parameter change over time/intensity from about −0.75 to −2.25, the slow dark state changes from −0.3 to −0.65 and the slow pumped state changes from −0.75 to −1.3. At the lowest intensity and earliest times, there is already evidence of the dark state being anisotropic. At the highest intensity, the fast response shows the largest fractional change in the order parameter, while the long-term response is somewhat lower, but by a statistically significant amount. The long-term change in the fractional order parameter in the dark state is greater than the initial value but still significantly smaller than the pumped state. Thus, the changes in the sample are enough to have a statistically significant effect on the measurements of the pumped state at high intensity, though not a large difference. Changes in the material from long-term light exposure might explain most of the small variations in DO11-doped PMMA thin films. However, the fractional change in the order parameter in DO11 is always similar for both pump polarizations, confirming that angular hole burning and reorientation are insubstantial.
5.4. Photomechanical Time Response
We determine the photomechanical response times by fitting the time-dependent stress as shown in
Figure 2 to a double exponential when the pump is on and a separate double exponential when the pump is off. As a function of intensity, these fits give the intensity dependence of the magnitude and characteristic photomechanical time constants.
Figure A1 and
Figure A2 show the time constants for the photomechanical response, while
Figure A3 and
Figure A4 show the relaxation time constants when the light is turned off for both DR1- and DO11-doped PMMA for both pump polarizations. The insets show the average time constant over the full intensity range for each material at each polarization.
While the fluctuations in the data exceed the error bar range, there is no obvious trend in the intensity dependence. The average of each run for a given sample and pump polarization shows that both the photomechanical response and its relaxation have two characteristic time constants that are about the same for each material. The fast time constant is about 0.5 s for both materials and for both the photomechanical response and the relaxation process. The longer time-scale process shows a different time constant for the pump on and off and range from about 1.5 s for pumped response to 3 s during relaxation.
In contrast, fibers from previous studies display a single time constant for both decay and recovery between about 0.7 s and 1.0 s [
2]. This suggests that the underlying processes in the two sample geometries may be different, which might originate in the surface skin effect, where in thin films, the skin is a much larger fraction of the total volume than in fibers. As such, the thin film measurements reported here most likely are probing a regime where the skin contributes a large fraction of the total response, thus showing two unique processes, while the fiber measurements are mostly probing bulk effects.
The amplitudes of the photomechanical response for the fast and slow processes determined from Equation (
2) are plotted in
Figure 16. The faster response mechanism amplitude
shows a large dependence on the pump power, while the slower response associated with the amplitude
remains approximately unchanged as a function of the intensity.
The fits in
Figure 16 to
and
as a function of intensity to the power series expansions
and
Note that
and
are not the material constants but rather related to the pre-stress
when mounting the samples, which is given by
Then, we can define an intensity-dependent photomechanical response as being given by
and
where
is the long-time photomechanical stress response of process
i to light of intensity
I.
The parameters
and
are determined for DR1- and DO11-doped PMMA for both vertical and horizontal pump intensities from fits to the data to Equation (
37) as shown in the top portion of
Figure 16. The same is performed for the long-term stress response for the process associated with time constant
by fitting the data in the bottom portion of
Figure 16 to the power series expansion given by Equation (
38). In this case, the data are relatively flat, so the fit includes only the linear term.
To summarize, the data given by
Figure 16 are used to obtain the photomechanical coefficients
for each process, which are substituted in Equations (
37) and (
38) to obtain the long-term stress
. These fits are also used to determine the intensity-dependent photomechanical response
for each process.
The strategy for isolating the mechanisms of the photomechanical response is to compare the photothermal heating model’s predictions given by Equation (
9) with the magnitude of the fast and slow response determined from the data. The heating model depends on several types of parameters. First,
,
,
and
n characterize the thermodynamic properties of the material and are determined from the temperature-dependent stress of a clamped sample as shown in
Figure 3. For DO11 and DR1, the data are fit to Equation (
4) as shown in
Figure 3 to obtain these parameters. Next, the physical properties of each sample, such as their thicknesses
w, are directly determined with a micrometer. Properties such as the specific heat capacity of the polymer
c and the material density
are sourced from the literature. We are thus assuming that the dopant dyes do not contribute substantially to these properties because of their low concentration. This assumption may need to be revisited given the large difference in the data between DO11 and DR1 in
Figure 3. Finally, each of the time constants from the fits of the stress response are used to obtain
at each intensity that is used in Equation (
5) to relate the temperature change to the light intensity. Recall that
and
are defined to be the short and long time constants for pump intensity
I. The strategy is to identify which exponential component of the two-exponential fit best agrees with the model’s prediction.
Alternatively, as we saw above, we can eliminate the parameter
in Equation (
9) using Equation (
4), yielding the semi-empirical model given by Equation (
12). In this case, the measured stress
at pump intensity
I determined from the fits for the process of response time
is used in the heating model, de-emphasizing reliance on the phenomenological stress model given by Equation (
4). As we will see below, the small differences between these two approaches are insignificant in determining which process is associated with photothermal heating.
Figure 17 and
Figure 18 show a summary of the theoretical and experimental intensity-dependent photomechanical constant
for the faster process with response time
. The bands are derived from the experimentally determined amplitudes of the fast process and their widths represent the one standard deviation uncertainty band. The vertical and horizonal data markers represent the photothermal heating model’s prediction for the intensity-dependent photomechanical response for the two orthogonal pump polarization directions. Their values are determined using the heating theory given by Equation (
9), plotted in
Figure 17, and Equation (
12), plotted in
Figure 18. The scatter in the plots originate from the scatter in the measured time constants
that are input parameters to the theory. The semi-empirical theory using Equation (
12) has an additional contribution to the uncertainty from
and
, which leads to larger error bars. Similarly, the error bands of the data are coded with vertical and horizontal hash marks to represent the polarization direction.
Figure 19 and
Figure 20 show the slower process with the time constant data and theory using
and
for the analysis.
Figure 17 and
Figure 18 show that the magnitude of the measured fast linear photomechanical response, as represented by the bands, decreases with pump intensity. The samples are pristine at the start of the experiment and characterized in the same physical spot starting with low intensity and then ending at the highest intensity with uniform intensity increases in between. Measurements of the order parameter, as discussed above, show that the material undergoes long-term changes in axial order of the dopant dyes, which we provisionally attribute to long-term molecular reorientation. This change in the fast response with exposure is consistent with the long-term process of molecular reorientation.
Figure 17 and
Figure 18 show that the data and the photothermal heating theory for the fast response agree for DR1 but disagree for DO11. As such, we conclude that the fast process is due to heating in DR1 but not in DO11. This is puzzling unless our assumption that the heat capacities are the same for both materials is wrong. The data suggest that the heat capacity of DO11 would need to be an order of magnitude smaller than in neat PMMA to make the heating theory agree with the data, which seems unlikely.
Figure 19 and
Figure 20 show the experimental data bands and the theory markers for the slower response given by time constant
. For DO11, the photomechanical response of the data and theory seem to marginally agree with some outliers, while for DR1, the data disagree with theory. However, the signs disagree. This suggests that the slow response is not associated with photothermal heating in both materials.
Table 4 summarizes the data. The fast response time of both DR1- and DO11-doped PMMA thin films is the same within experimental uncertainty and is consistent with the time constant being determined from the sample geometry, which determines how heat dissipates. For heating, then, the time constants are expected to be similar, as is
. The fact that the time constants are independent of polarization is also what one would expect from an isotopic sample. This gives us greater confidence that the fast response originates in photothermal heating. The fact that DO11 has a smaller linear response from heating is also consistent with
Figure 3, which shows that heating results in a smaller temperature-dependent force. All of these data together are consistent with photothermal heating being responsible for the fast process.
The slow response in both materials is of negative sign, consistent with sample contraction as one would expect for molecular hole burning and molecular reorientation away from the pump polarization direction. Molecular reorientation would result in a positive response for a perpendicular pump beam, while orientational hole burning would give a negative response, but of smaller magnitude. The data are consistent with hole burning, but the uncertainties are too large to make a definitive conclusion.
The critical intensity
due to the fast heating process is defined by
and quantifies the intensity at which the higher-order term exceeds the linear response. DO11’s critical intensity is smaller than DR1’s value because its linear response is smaller. For both materials and both polarizations, the first nonlinear correction to the photomechanical response
is the same within experimental uncertainty.