1. Introduction
Malignant tumors with high infiltration levels rapidly metastasize and have a high risk of recurrence [
1,
2,
3]. Surgeries aim for the complete resection of malignant tumors [
1,
2,
3]. However, tumors may be distributed in blood vessels [
4].
The colors of the tumor and blood vessels are similar, making it difficult to visually observe the boundaries [
5]. Therefore, the state of tumor removal can be observed using a fluorescence surgical microscope by injecting fluorescence sodium (Fluorescite injection 10% [500 mg] @ Korea Alcon. Co., Ltd., Seoul, Republic of Korea) into the body [
6,
7,
8,
9]. In this study, a light emitting diode (LED), suitable for use as a fluorescent contrast agent, was irradiated in vitro. When an LED with a 405 nm excitation wavelength (λ
ext) was irradiated on the tumor, a 530–560 nm fluorescence emission wavelength (λ
em) was generated by a chemical reaction of the fluorescent contrast agent present in the tumor [
6]. The color (yellow), corresponding to the spectrum of fluorescence emission, was photographed using a camera. Finally, the captured images were observed using an external monitor.
The fluorescent contrast medium must have sufficient LED beam strength (>0.5 mW) and beam width (30–60°). If the beam width is narrow, the lesion observation field becomes narrow [
10]. For observation with a clear image quality, the working distance (WD) between the LED (or camera) and the lesion must be 15–30 cm, and the range (θ
D) of the left and right movement of the LED (or camera) must also correspond to 30–60° [
10]. However, a WD of 15–30 cm consumes LED beam energy for delivery to the lesion, resulting in LED power loss [
11,
12]. Therefore, the fluorescence emission of small lesions may be weak. In addition, surgical microscopes are large and heavy, with high maintenance costs [
13]. Therefore, it is necessary to develop a handheld lightweight microscope, with a free radius of movement of the WD and cameras (or LEDs) to make lesion observation easier. The beam energy loss generated by WD is small. Commercial LEDs are utilized to minimize the size and cost. However, it is impossible to secure the lesion observation field due to the narrow beam width (θ
D = 10°) of LEDs [
14]. That is, the beam width of LEDs is wide, and the observation field is irradiated more uniformly as the beam width increases. Therefore, lesions > 10 cm in size can be observed. The number of LEDs can be increased to increase the beam width and irradiation intensity. However, LED irradiation methods lead to the formation of shadows within the lesion, and the costs increase with a rising number of LEDs.
Shadows interfere with the visualization of the lesion. Therefore, it is necessary to increase the beam strength and width by using only a small number of LEDs. As WD increases, illuminance decreases and the intensity of the light is not uniform, even within the beam width [
14]. Blue LEDs are highly utilized due to their small volume and low cost, but because the beam width (θ
D = 10°) is relatively narrow (diameter of survey area = 26 mm), it is difficult to secure a lesion viewing angle. As this leads to a decrease in the accuracy of lesion observation and diagnosis using a surgical microscope, research is needed to design an LED arrangement that expands the beam width of light and minimizes shadow formation in the lesion.
LED arrangements in the form of rings and array matrices have been studied and developed [
15,
16]. These arrangements are suitable for fluorescence emission-guided sample monitoring in the operating room because of their uniform light intensities and sufficient observation fields of the object. The ring-type LED was designed with a wide-wavelength band light source by placing RGB LEDs at the same ratios as those for application in machine vision systems. However, considering the nature of the wavelength of LEDS with applications in surgical microscopes, the fluorescence emission-guided LED irradiation wavelength is a single wavelength, and is therefore not suitable for sample observation [
15]. LEDs arranged in an array matrix exceed the power level required for surgery [
16]. These have the disadvantage of being burdened by applications in operating rooms because of their unnecessary energy consumption and high manufacturing costs, and it is difficult to adjust the radius angle and WD of the light source.
This paper proposes a method for designing ring-type asymmetric LED beam irradiation with a wide beamwidth, based on the uniform beam irradiation intensity of an LED, to secure a field of view so that lesions with large areas can be sufficiently observed through fluorescence emission. The proposed ring-type asymmetric LED beam irradiation method can optimize the number of LEDs, minimize beam irradiation power loss at a fixed WD, and secure a sufficient observation field and beam irradiation strength that can induce fluorescence emission.
2. Light Source Irradiation Characteristics Analysis and LED Arrangement Method
As shown in
Figure 1a, LED beam irradiation methods are generally classified as symmetric or asymmetric beam irradiation. As shown in
Figure 1b, there is a difference in the incident angles (θ
ext) on the irradiation area (t) of the target [
17].
A typical LED has an incident angle (θext) of 0°. Therefore, when light radiates from an LED, the beamwidth of the light is constant (θD = 10°), and the light spreads evenly in all directions. These methods are classified as symmetric beam irradiation techniques. If the LED is tilted away from 0°, the beam spreads out. Therefore, the beamwidth is no longer 10°. This is known as the asymmetric method.
The symmetric method has a relatively small beamwidth but a strong beam intensity. In contrast, the asymmetric method has a wide beamwidth but a weak beam intensity. The common point between the two methods is that the beam intensity is strong when the beam is perpendicular to the irradiation area of the LED. The longer the working distance, the weaker the beam intensity becomes. However, when the beam deviates from the perpendicular direction, its intensity (Ph ≠ Pl) weakens.
The dual asymmetric method can widen the beam and increase its intensity, as shown in
Figure 1a. However, as the intensity of the beam becomes stronger only in the area where the LED beamwidths overlap, a shaded area forms, obstructing the view of the lesion. The multi-asymmetric method widens the beam and uniformly increases its intensity. Although they can provide the most optimal effect, dual- and multi-asymmetric methods increase the number of LEDs, resulting in increased costs and complicated structures.
A ring-type array can increase the number of LEDs and widen the beam width, as shown in
Figure 2. In addition, the power and brightness of the beam can be increased by using more LEDs. Because of these functional features, a ring-type array design can be used to effectively observe the target, as shown in
Figure 1b. A structural feature of the ring-type array is that it is possible to install a camera in the center hole (h), which can reduce the formation of shadows in the monitored image [
18].
LEDs (Thorlabs LED405E) with a beamwidth (θ
D) of 10° are attached in a ring-type array structure, maintaining a constant distance (D
r: 4.32 cm) between the LEDs, as shown in
Figure 2. As shown in
Figure 1b, when collectively irradiating the target (lesion), there might be areas where the light is unable to reach uniformly. Therefore, shadow generation must be eliminated to secure the field of view for fluorescence emission observations. The symmetric irradiation method has limitations in securing a uniform observation field of the lesion, due to the formation of shadows. In addition, it is uneconomical because the standards of the light source, the manufacturing cost, and the power consumption increase with an increase in the number of LEDs. Therefore, it is necessary to propose an LED beam irradiation method in which the entire lesion can emit fluorescence using only a small number of LEDs without shadow formation. An asymmetric beam irradiation method rotates the incident angle (θ
ext > 0°) so that the irradiation area of light becomes wider than the symmetric irradiation area, as shown in
Figure 1. The single asymmetric irradiation method has different excitation powers (
Ph ≠
Pl) in both directions at the center point (d
g) of the irradiation area (t); therefore, the fluorescence is not uniform within the irradiation area (t), and the light source generates a different power level for each position of the beam width. To compensate for this phenomenon, we propose a multi-asymmetric beam irradiation method. The multi-asymmetric beam irradiation method increases the number of LEDs, as shown in
Figure 2. The LEDs are arranged with a relatively weak power overlap. Therefore, the proposed method can induce uniform fluorescence (
Ph =
Pl) and is designed to enable fluorescence emission in a wider area. It has the advantage that fewer LEDs are used, compared to a symmetric beam irradiation method under an identical WD, light source size, and irradiation power condition.
When performing LED beam irradiation, the variables of LED beam irradiation are the number of LEDs, the arrangement interval (D
r), the incident angle (θ
ext), and the WD. To ensure sufficient resolution, even at the maximum WD, the maximum WD of the multi-asymmetric beam irradiation method was determined to be 30 cm. To obtain the optimal angle of incidence for LED irradiation, it was necessary to observe the change in the irradiation area of the LED as the angle of incidence changes, while the other performance variables are fixed. Therefore, the number of LEDs and the arrangement spacing should be identical, as shown in
Figure 2. The LED irradiation area (t) increases in proportion with the incident angle (θ
ext), as shown in
Figure 1. Shadows form when the incident angle (θ
ext) is more than 15.37°.
Figure 3a shows that, if the angle of incidence (θ
ext) increases for the irradiation sections of LED
n and LED
n+4, the width of the irradiation area from LED
n to LED
n+4 increases proportionally, and the shadow phenomenon occurs between each irradiation area (t). When observing a lesion during surgery, shadows always occur at the center of the lesion. Therefore, the shaded area is not irradiated with an LED (at the excitation wavelength). Therefore, fluorescence emission (at the emission wavelength) cannot occur in shaded areas, and lesions cannot be observed in these areas. Additionally, if a lesion is observed with the brightness of the LED, the shaded area is not irradiated by the LED, which obstructs the view during the lesion observation process. Quantification of the irradiated area is required to evenly irradiate the entire area of the lesion with an LED. As shown in
Figure 3b, if eight LEDs simultaneously irradiate the entire area (A
ext) of the lesion, depending on the incident angle (θ
ext) given in
Figure 3a, the single LED irradiation areas (A
s) might overlap, and if not, shaded areas might occur. As the result, with the boundaries of the overlapping regions canceled out, each irradiation area (A
s) remains in the form of the sum of areas (A
ext). Each irradiation area (A
s) has an elliptical shape. This is because LED irradiation is asymmetric. Therefore, it is necessary to obtain the value of the irradiation area (A
ext) of the LED, to obtain results for the entire area of the lesion to be irradiated.
As a condition for the test, a minimum incident angle (θ
ext), capable of irradiating the entire area of the phantom, must be established. As shown in
Figure 4a, during the design process of the experiment, the fixing frame of the LED was manufactured using a 3D printer. The incident angle (θ
ext) of the LED is 15.37°, as shown in
Figure 4b.
3. Experimental Results
To test the light source performance of the proposed method, a phantom made of liquid latex rubber was injected with a fluorescent contrast agent of sodium fluorescein (Alcon, Seoul, Republic of Korea, 5 mL of Fluorescite injection 10%) at a concentration of 0.02 mM [
19]. This phantom was used to represent lesions. The size of the phantom was approximately 10 cm, which is similar to that of colon cancer lesions. Referring to
Figure 2, the power reaching the phantom (P
r) and the diameter of the irradiance area were measured. The incident angle and beam irradiation power (P
ext) of all LEDs were 15.37° and 10 mW, respectively, at an irradiation wavelength of 405 nm, and the total number of LEDs was eight.
The LEDs were arranged in a ring-array type structure. The diameter of the beam irradiation area for the guided fluorescence emission was 9.5 cm. This is sufficient to irradiate most types of tumor tissue. To set up the experimental conditions, the light source was fixed to a clamp (Manfrotto 244N, Manfrotto, Veneto, Italy), as shown in
Figure 5, and the height of the clamp was 30 cm. Thus, the working distance between the LED and target was 30 cm.
The bias voltage of the LED for irradiating the light source was 5.0 V. Light power (P
r) and light brightness were measured using a power meter (Thorlabs S121C) and a lux meter (Thorlabs S142C, Sincon ST-126), respectively. The irradiation area of the light source was measured, as shown in
Figure 5. The measurement results are presented in
Table 1.
The measured power (P
r) of the light source was 4.63 mW at an irradiation wavelength of 405 nm, and this power satisfied the conditions for fluorescence emission (>0.5 mW). These fluorescence results are shown in
Figure 6 and were observed using an NIR camera (Lt-225c, Lumenera camera, TELEDYNE, Inc., Thousand Oaks, CA, USA) and a phantom to induce fluorescence. With a WD of 30 cm, the LED supply bias voltage was adjusted from 3.0 to 5.0 V, and the phantom area of interest (ROI) in the NIR camera image was converted into an average RGB value.
The irradiance presented in
Figure 6 can be calculated using the average RGB value, which means that the higher the brightness value, the better the fluorescence emission. The proposed light source emits fluorescence at the maximum WD (30 cm).
The irradiation area of the LED decreased proportionally with a decrease in the WD. Therefore, a WD of at least 15 cm was used to irradiate the entire phantom. This phenomenon was observed at the maximum LED beam irradiation intensity and an external voltage of 3.5 V or higher. As shown in
Figure 6, the light source has excellent light power, brightness, and fluorescence emission. The phantom area (A
ext) where fluorescence emission occurred under the irradiation of the light source was measured to be 9.5 cm.
4. Discussion
A 405 nm LED with a beam width (θ
ext) of 10° and an optical source power of 10 mW was used to analyze fluorescence emission. The experiment was conducted after injecting a fluorescent contrast agent into the phantom [
19]. Most surgeries using 405 nm fluorescein sodium are craniotomies [
20]. Compared to the incision size of the craniotomy, the irradiation area of the light source was suitable (diameter of 9.5 cm). Therefore, the fluorescence emission-guided multi-asymmetric beam irradiation method using sodium fluorescein is suitable for observing tumor removal during surgery. In a symmetric LED irradiation method, direct irradiation must be performed in a straight line using a beam with a 0° reference (
Figure 1). Therefore, only the LED-irradiated position of the lesion becomes stronger with increasing LED intensity. In addition, because the beams are not irradiated at intervals between LEDs, a shadowed area may form.
Table 2 summarizes the differences between the proposed method and other research results. It is difficult to obtain an absolute comparison with other studies ([
15,
16,
21,
22]) because the WD, LED quantity, and irradiation power were all different. The power of the LED depends on the number of LEDs, and the WD depends on the research objectives. Therefore, we were interested in the received power and irradiation area, to determine whether the fluorescence emission was sufficiently wide (wide irradiation area) and strongly irradiated for a WD of 30 cm, when applied to a surgical field using a small number of LEDs. We compared our results with those in references [
15,
16,
21,
22]. The number of LEDs was small compared to those in [
15,
16,
21,
22], and the WD was longer. The beam irradiation power was relatively low. The received power and irradiation area were the largest.
The prior studies share a common design feature; a dome-type array structure [
21,
22]. The dome-type array structures were connected by using 90 and 130 LEDs, respectively [
21,
22]. The reason for increasing the number of LEDs is that the brightness of the light source increases proportionally with the number of LEDs. Thus, even if the WD increases, the irradiation brightness of the LEDs and the irradiation beam width increase proportionally, and a large area can be irradiated uniformly. However, the optimal WD range at a surgical site is 15–30 cm [
10]. Therefore, an excessive WD increase is unnecessary. If the WD increases, the intensity of the irradiation beam should be increased by increasing the number of LEDs, so that fluorescence can be sufficiently expressed. However, an increase in the number of LEDs and the intensity of beam irradiation increases the energy consumption, which causes heat generation in the LED module and mechanical damage to the module [
22]. The studies adjusted the excitation angle range from 0 to 65° to increase the beam irradiation effect [
21,
22]. In particular, the incident angles were adjusted differently for each LED layer (θ
0 = 47.5°, θ
1 = 42.0°, θ
2 = 36.6°, θ
3 = 31.0°, θ
4 = 22.9°) to enable a wide area of beam irradiation [
22]. However, the actual irradiated area is not the only relevant factor for observing tumors during fluorescence-induced surgery. Some other factors include the angle of incidence, the WD, and the arrangement, direction, and power of LED.
For a surgical fluorescence induction microscope, with eight LEDs and a WD of 30 cm, the LED beam irradiation was strong enough to obtain an optimal received power (>0.05 mW) for fluorescence expression. This study analyzed an asymmetric LED beam irradiation method to irradiate a wide lesion with uniform intensity. This resulted in an even expression, making it is easier to observe the entire lesion. This was achieved by using a ring-type array design for LED beam irradiation.