Smartphone Slit Lamp Imaging—Usability and Quality Assessment
Abstract
:1. Introduction
2. Methods
2.1. Ethics
2.2. Study Design
2.3. Participants
2.4. Image Acquisition and Grading
2.5. Statistical Analysis
3. Results
4. Discussion
5. Conclusion
Supplementary Materials
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Pearce, J.M. The ophthalmoscope: Helmholtz’s Augenspiegel. Eur. Neurol. 2009, 61, 244–249. [Google Scholar] [CrossRef] [PubMed]
- Helmholtz, H. Beschreibung Eines Augen-Spiegels zur Untersuchung der Netzhaut im Lebenden Auge; Springer: Heidelberg/Berlin, Germany, 1851; p. 45. [Google Scholar]
- Gellrich, M.-M. History of the Slit Lamp. In The Slit Lamp; Springer: Heidelberg/Berlin, Germany, 2014; p. 22. [Google Scholar]
- Gellrich, M.M. A new view of the slit lamp. Br. J. Ophthalmol. 2009, 93, 272–273. [Google Scholar] [CrossRef]
- Kalayoglu, M.V. The Evolution of Slit Lamp Biomicroscopy. 2005. Available online: https://www.ophthalmologyweb.com/Tech-Spotlights/26450-The-Evolution-of-Slit-Lamp-Biomicroscopy/ (accessed on 4 October 2022).
- Haag-Streit Diagnostics. Slit Lamp Imaging Guide: Superior Technology—Reliable Instruments; Haag-Streit Diagnostics: Bern, Switzerland, 2022. [Google Scholar]
- Vogt, A. Atlas der Spaltlampenmikroskopie des Lebenden Auges, 1st ed.; Springer: Heidelberg/Berlin, Germany, 1921; p. 172. [Google Scholar]
- Hirsch, R.J. Seizing the Light: A History of Photography, 2nd ed.; McGraw-Hill: New York, NY, USA, 2000; p. 530. [Google Scholar]
- Schiff-Wertheimer, S.; Loisillier, M.F. Notes on the construction of a photographic slit lamp. Arch. Ophtalmol. Rev. Gen. Ophtalmol. 1958, 18, 833–835. [Google Scholar] [PubMed]
- Franguelli, R. Apparatus for macrophotography in vivo to be used in Zeiss slit-lamp stereomicroscopy. Ann. Ottalmol. Clin. Ocul. 1968, 94, 1582–1586. [Google Scholar] [PubMed]
- Soper, J.W. A camera adapter for slit-lamp photography. Int. Ophthalmol. Clin. 1976, 16, 181–187. [Google Scholar] [PubMed]
- Fong, D.A. An introduction to slit lamp photography. J. Ophthalmic Nurs. Technol. 1984, 3, 101–108. [Google Scholar] [PubMed]
- Gutner, R.K. Slit lamp photodocumentation. Am. J. Optom. Physiol. Opt. 1979, 56, 559–562. [Google Scholar] [CrossRef]
- Hayashi, H. Lens illumination photography with slit lamp microscopic photoapparatus. Ganka 1970, 12, 788–791. [Google Scholar]
- Makes, D.J. Innovations in ophthalmic photography. J. Ophthalmic Nurs. Technol. 1984, 3, 116–117. [Google Scholar]
- Traboulsi, E.I.; Aswad, M.I. Practical slit-lamp photography. Ann. Ophthalmol. 1985, 17, 770. [Google Scholar]
- Miller, D.; Taube, J.; Miller, R.; Gleason, W.; Babyn, P.; Moon, J.; Weiss, J.N. A system for slit-lamp polaroid photography. Ophthalmic Surg. 1981, 12, 328–331. [Google Scholar] [CrossRef] [PubMed]
- Lobel, D. Instant photography with the Polaroid SX-70 Alpha I camera mounted on a slit lamp. Arch. Ophthalmol. 1980, 98, 748. [Google Scholar] [CrossRef] [PubMed]
- Fogla, R.; Rao, S.K. Ophthalmic photography using a digital camera. Indian J. Ophthalmol. 2003, 51, 269–272. [Google Scholar] [PubMed]
- Vedantham, V. Digital ophthalmic photography. Indian J. Ophthalmol. 2004, 52, 83–84. [Google Scholar]
- McLean, C.J.; Tossounis, C.M.; Saleh, G.M. Camera adapter for anterior segment slitlamp photography. J. Cataract Refract. Surg. 2006, 32, 1889–1891. [Google Scholar] [CrossRef] [PubMed]
- Lord, R.K.; Shah, V.A.; San Filippo, A.N.; Krishna, R. Novel uses of smartphones in ophthalmology. Ophthalmology 2010, 117, P1274.E3. [Google Scholar] [CrossRef]
- Chhablani, J.; Kaja, S.; Shah, V.A. Smartphones in ophthalmology. Indian J. Ophthalmol. 2012, 60, 127–131. [Google Scholar] [CrossRef]
- Akkara, J.D.; Anju, K. How-to guide for smartphone slit-lamp imaging. Kerala J. Ophthalmol. 2019, 31, 64–71. [Google Scholar] [CrossRef]
- Hester, C.C.; Feldmann, B.H. Smart Phoneography—How to Take Slit Lamp Photographs with an iPhone. 2021. Available online: https://eyewiki.org/Smart_Phoneography_-_How_to_Take_Slit_Lamp_Photographs_with_an_iPhone (accessed on 4 October 2022).
- Dutt, S.; Vadivel, S.S.; Nagarajan, S.; Galagali, A.; Christy, J.S.; Sivaraman, A.; Rao, D.P. A novel approach to anterior segment imaging with smartphones in the COVID-19 era. Indian J. Ophthalmol. 2021, 69, 1257–1262. [Google Scholar] [CrossRef]
- Akkara, J.D. Commentary: Dawn of smartphones in frugal ophthalmic innovation. Indian J. Ophthalmol. 2018, 66, 1619. [Google Scholar] [CrossRef]
- Lee, W.W. Slit Lamp Adapters Turn Smartphones into Clinical Cameras. 2013. Available online: https://www.ophthalmologyweb.com/Featured-Articles/136817-Slit-Lamp-Adapters-turn-Smartphones-into-Clinical-Cameras/ (accessed on 9 October 2022).
- Roy, S.; Pantanowitz, L.; Amin, M.; Seethala, R.R.; Ishtiaque, A.; Yousem, S.A.; Parwani, A.V.; Cucoranu, I.; Hartman, D.J. Smartphone adapters for digital photomicrography. J. Pathol. Inform. 2014, 5, 24. [Google Scholar] [CrossRef] [PubMed]
- Morales-Leon, J.-E.; Díaz-De-León, L.R. Smartphone stereoscopic imaging: Inexpensive 3d telemedicine. Pan-Am. J. Ophthalmol. 2020, 2, 35. [Google Scholar] [CrossRef]
- Elloumi, Y.; Akil, M.; Kehtarnavaz, N. A mobile computer aided system for optic nerve head detection. Comput. Methods Programs Biomed. 2018, 162, 139–148. [Google Scholar] [CrossRef] [PubMed]
- Chan, J.B.; Ho, H.C.; Ngah, N.F.; Hussein, E. DIY—Smartphone Slit-Lamp Adaptor. J. Mob. Technol. Med. 2014, 3, 16–22. [Google Scholar] [CrossRef]
- Ateya, A.; Akkara, J.D.; Kuriakose, A. Custom-made three-dimensional-printed adapter for smartphone slit-lamp photography. Kerala J. Ophthalmol. 2020, 32, 83–86. [Google Scholar] [CrossRef]
- Raju, B.; Raju, N.S.; Akkara, J.D.; Pathengay, A. Do it yourself smartphone fundus camera—DIYretCAM. Indian J. Ophthalmol. 2016, 64, 663–667. [Google Scholar] [CrossRef]
Smartphone Group | Control Group | |
---|---|---|
Group size (n) | Twenty-Six:
| 1 |
Age [years] (median (IQR)) | 32.00 (10.00) | 28.00 (0) |
Gender ratio female:male | 13:13 | 1:0 |
Professional experience [years] (median (IQR)) (min–max range) | 2.58 (6.80) (0–17) | 4.0 ± 0 |
Smartphone experience [years using a smartphone] (median (IQR)) (min–max range) | 100% of participants 13.00 (5.00) (7–21) 52% Apple iOS 48% Google Android | 100% of participants 12.00 (0) (N/A) 100% Google Android |
Personal smartphone competence of (private) photos taken with a smartphone (mean ± SD) | 8.00 (2.00) | N/A |
Number of photos taken with a smartphone [% of participants] | 17%: ≤100/year 50%: >100 and ≤1000/year 33%: >1000/year | N/A |
Number of previous attempts to take a slit lamp photo with a smartphone [% of participants] | 46%: 0 attempts 4%: ≥1 and <2 attempts 50%: ≥2 attempts | N/A |
Anterior Segment | Posterior Segment | Mann–Whitney U test | |
---|---|---|---|
Overall quality (median (IQR)) | 8 (4.00) | 7 (5.00) | p = 0.019 * |
Sharpness (median (IQR)) | 8 (3.75) | 7 (4.50) | p = 0.075 |
Exposure (median (IQR)) | 7 (4.00) | 7 (4.00) | p = 0.964 |
Field of view (median (IQR)) | 8 (4.00) | 6 (5.00) | p < 0.001 * |
Color (median (IQR)) | 8 (5.00) | 7 (4.50) | p = 0.262 |
Clinical interpretability (median (IQR)) | 8 (5.00) | 7 (5.00) | p = 0.007 * |
Artifacts | 8 (4.75) | 6 (4.00) | p < 0.001 * |
Take 1 | Take 2 | Take 3 | Kendall’s W | |
---|---|---|---|---|
Anterior segment | ||||
Quality overall (median (IQR)) | 6.83 (3.06) | 7.17 (3.92) | 7.33 (3.04) | p = 0.060 |
Quality sharpness (median (IQR)) | 8.00 (3.00) | 7.00 (3.50) | 8.00 (3.00) | p = 0.450 |
Quality exposure (median (IQR)) | 6.50 (3.38) | 7.00 (4.50) | 7.00 (3.75) | p = 0.108 |
Quality field of view (median (IQR)) | 7.50 (4.13) | 8.50 (2.50) | 8.25 (2.38) | p = 0.059 |
Quality color (median (IQR)) | 7.50 (3.25) | 7.50 (2.50) | 7.25 (3.38) | p = 0.347 |
Quality clinical interpretability (median (IQR)) | 7.25 (3.88) | 8.00 (3.50) | 7.75 (3.88) | p = 0.350 |
Quality artifacts (median (IQR)) | 7.25 (3.38) | 7.50 (3.50) | 7.50 (3.75) | p = 0.264 |
Acquisition time [sec] (median (IQR)) | 143 (67.00) | 129 (68.50) | 127 (56.75) | p = 0.558 |
Subjective difficulty image acquisition (median (IQR)) | 8.00 (3.00) | 9.00 (2.00) | 9.00 (2.00) | p < 0.001 * |
Posterior segment | ||||
Overall quality (median (IQR)) | 6.00 (5.13) | 6.75 (3.00) | 7.50 (5.50) | p = 0.094 |
Sharpness (median (IQR)) | 6.75 (5.13) | 7.25 (3.13) | 8.25 (4.13) | p = 0.015 |
Exposure (median (IQR)) | 6.25 (3.88) | 7.25 (3.13) | 7.00 (4.13) | p = 0.788 |
Field of view (median (IQR)) | 5.00 (4.38) | 5.75 (3.38) | 6.75 (3.75) | p = 0.284 |
Color (median (IQR)) | 6.50 (3.38) | 7.00 (2.88) | 7.00 (4.13) | p = 0.351 |
Clinical interpretability (median (IQR)) | 5.75 (5.00) | 6.50 (3.00) | 7.00 (5.13) | p = 0.097 |
Artifacts (median (IQR)) | 5.25 (2.88) | 6.00 (3.75) | 6.25 (4.63) | p = 0.149 |
Acquisition time [sec] (median (IQR)) | 170 (117.00) | 141 (80.00) | 143 (78.75) | p = 0.622 |
Subjective difficulty image acquisition (median (IQR)) | 6.00 (2.50) | 7.00 (2.00) | 8.00 (2.00) | p = 0.002 * |
Subjective difficulty device installation (median (IQR)) | 9.00 (1.75) | 9.50 (1.00) | 10.00 (1.00) | p = 0.004 * |
Smartphone Group | Control Group | |
---|---|---|
Acquisition time [sec] anterior (median (IQR)) | 133.67 (55.33) | 192.00 (20.50) |
Acquisition time [sec] posterior (median (IQR)) | 153.17 (95.83) | 140.00 (19.00) |
Disturbance of adapter on slit lamp (median (IQR)) | 8.00 (1.00) | N/A |
Disturbance of adapter on the phone (median (IQR)) | 6.00 (4.00) | N/A |
Complexity of the adapter set (median (IQR)) | 9.00 (2.00) | N/A |
Anterior Segment | Posterior Segment | |||
---|---|---|---|---|
Image quality vs. acquisition time | r = −0.371 | p = 0.068 | r = −0.172 | p = 0.444 |
Image quality vs. work experience | r = 0.291 | p = 0.167 | r = 0.476 | p = 0.025 * |
Image quality vs. smartphone use | r = −0.008 | p = 0.969 | r = 0.145 | p = 0.508 |
Image quality vs. personal smartphone confidence | r = 0.097 | p = 0.644 | r = −0.087 | p = 0.694 |
Acquisition time vs. work experience | r = 0.202 | p = 0.356 | r = 0.252 | p = 0.284 |
Acquisition time vs. smartphone use | r = 0.090 | p = 0.675 | r = 0.010 | p = 0.964 |
Acquisition time vs. personal smartphone confidence | r = 0.074 | p = 0.732 | r = 0.107 | p = 0.645 |
Brand | Product | Source (as of October 2022) |
---|---|---|
Arcturus Labs | Magnifi photoadapter | Discontinued (www.arcturuslabs.com (accessed on 20 October 2022)) |
Celestron | NexYZ universal smartphone adapter | https://www.celestron.de/ce_de/nexyz-universaler-3-achsen-smartphone-adapter.html (accessed on 20 October 2022) |
Digital Eye Center | Universal smartphone slit lamp adapter w/ sleeves | https://www.digitaleyecenter.com/product/universal-smartphone-slit lamp-adapter (accessed on 20 October 2022) |
Eye2Mobile | PHONEdock | https://www.eye2mobile.com/ (accessed on 20 October 2022) |
Eyecapp (formerly Cruise Ophthalmic) | Smartphone adapter | Discontinued (https://www.eyecapp.com) (accessed on 20 October 2022) |
Eye Photo Doc | Smartphone universal case | https://www.eyephotodoc.com/Price_of_Eyephotodoc_iPhone_systems.html (accessed on 20 October 2022) |
HI Resolution Enterprises | Snapzoom universal digiscoping adapter | https://www.snapzooms.com/shop/6v3efy0iew39g7e8vbx3tspwjykz95/6v3efy0iew39g7e8vbx3tspwjykz95 (accessed on 20 October 2022) |
Keeler | Portable slit lamp iPhone 4 imaging adapter (3010-P-7010) | Discontinued |
Kowa Optimed | Smartoscope Vario universal smartphone adapter | https://www.kowaoptic.com/de/smartoscope-vario-universal-smartphone-adapter?c=84 (accessed on 20 October 2022) |
New Vision | Smartphone photographic adapter | http://www.4vision.cn/P_view.asp?pid=191 (accessed on 20 October 2022) |
oDocs | Slit Lamp adapter | https://odocseyecare.shop/products/odocs-slit lamp-surgical-microscope-adapter (accessed on 20 October 2022) |
Optimetrics | Smartphone digital slit lamp metal adapter | https://optimetrics.com/home/1265-smart-phone-digital-slit lamp-adapter.html (accessed on 20 October 2022) |
Orion Telescopes & Binoculars | SteadyPix telescope photo adapter for iPhones | https://www.telescope.com/Astrophotography/Astrophotography-Accessories/Orion-SteadyPix-Telescope-Photo-Adapter-for-iPhone/c/4/sc/61/p/101445.uts (accessed on 20 October 2022) |
Seiler Medical | Smartphone adaptor (IPH-VA) | https://www.seilermicro.com/products/accessories/smart-phone-adaptor (accessed on 20 October 2022) |
Skylight Healthcare Systems | SkyLight smartphone adapter | Discontinued (https://opticsmag.com/skylight-scope-microscope-cell-phone-adapter (accessed on 20 October 2022)) |
TigerLens | Smartphone adapter | Discontinued (www.tigerlens.com (accessed on 20 October 2022)) |
TTI Medical | ACCU-BEAM universal smartphone adapter (8100SP) | https://ttimedical.com/products/digital-adapters/smartphone-adapter (accessed on 20 October 2022) |
Welch Allyn/Hillrom (now part of Baxter Inc.) | iExaminer adapter for iPhones (11840) | Discontinued (https://www.welchallyn.com/en/microsites/iexaminer.html (accessed on 20 October 2022)) |
Zarf Enterprises | Slit lamp digital camera adapters | http://www.zarfenterprises.com (accessed on 20 October 2022) |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Muth, D.R.; Blaser, F.; Foa, N.; Scherm, P.; Mayer, W.J.; Barthelmes, D.; Zweifel, S.A. Smartphone Slit Lamp Imaging—Usability and Quality Assessment. Diagnostics 2023, 13, 423. https://doi.org/10.3390/diagnostics13030423
Muth DR, Blaser F, Foa N, Scherm P, Mayer WJ, Barthelmes D, Zweifel SA. Smartphone Slit Lamp Imaging—Usability and Quality Assessment. Diagnostics. 2023; 13(3):423. https://doi.org/10.3390/diagnostics13030423
Chicago/Turabian StyleMuth, Daniel Rudolf, Frank Blaser, Nastasia Foa, Pauline Scherm, Wolfgang Johann Mayer, Daniel Barthelmes, and Sandrine Anne Zweifel. 2023. "Smartphone Slit Lamp Imaging—Usability and Quality Assessment" Diagnostics 13, no. 3: 423. https://doi.org/10.3390/diagnostics13030423
APA StyleMuth, D. R., Blaser, F., Foa, N., Scherm, P., Mayer, W. J., Barthelmes, D., & Zweifel, S. A. (2023). Smartphone Slit Lamp Imaging—Usability and Quality Assessment. Diagnostics, 13(3), 423. https://doi.org/10.3390/diagnostics13030423