Review

Current and Emerging Methods of Antibiotic Susceptibility Testing

School of Mechanical Engineering, Korea University of Technology and Education, Cheonan, Chungnam 31253, Korea
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Diagnostics 2019, 9(2), 49; https://doi.org/10.3390/diagnostics9020049
Submission received: 18 March 2019 / Revised: 28 April 2019 / Accepted: 28 April 2019 / Published: 3 May 2019
(This article belongs to the Special Issue Diagnosis of Bacterial Pathogens)

Abstract

:
Antibiotic susceptibility testing (AST) specifies effective antibiotic dosage and formulates a profile of empirical therapy for the proper management of an individual patient’s health against deadly infections. Therefore, rapid diagnostic plays a pivotal role in the treatment of bacterial infection. In this article, the authors review the socio-economic burden and emergence of antibiotic resistance. An overview of the phenotypic, genotypic, and emerging techniques for AST has been provided and discussed, highlighting the advantages and limitations of each. The historical perspective on conventional methods that have paved the way for modern AST like disk diffusion, Epsilometer test (Etest), and microdilution, is presented. Several emerging methods, such as microfluidic-based optical and electrochemical AST have been critically evaluated. Finally, the challenges related with AST and its outlook in the future are presented.

1. Introduction

Antibiotic resistance is defined as the genetic ability of bacteria to encode the resistance genes that counterfeit the inhibitory effect of potential antibiotics for survival [1]. It can be developed either intrinsically by natural recombination and integration into the bacterial genome, or it can be acquired through horizontal gene mutation events such as conjugation, transformation, and transduction [2]. The prominent events in the generation of bacterial resistance include inactivation of the porin channel, modification of antibiotic targets, and neutralizing antibiotic efficacy through enzymatic action [3]. Thus, the understanding of the genetic makeover and the morpho-anatomical changes in bacteria are of prime importance to counteracting the resistance mechanism.
The discovery of antibiotics was paradigm-altering, as it was not only an effective tool against chronic infections but also opened new avenues for drug industries. Global antibiotic statistics suggested an increase of 35% in the antibiotic consumption between 2000 and 2010, and the current antibiotic industry stands at USD 39.8 billion (up to 2015). Russia, India, China, Brazil, and South Africa are major contributing countries, where 76% of the rise in antibiotic consumption has been estimated [4]. The changes experienced in the enhanced consumption of antibiotics over the past decade remain unprecedented, and this is chiefly a result of the emergence of new diseases. Alternately, the increase in antibiotic consumption and industrialization might be due to overuse or misuse of antibiotics recommended by physicians/self-medication at the time of infection [5,6]. A recent report on the casualties related to antibiotic resistance by the world health organization (WHO) depicted an alarming 700,000 lives per year currently, and predicts a disturbing 10 million/year by 2050, ensuring that antibiotic resistance will be the most prevalent cause of death [7]. Adding to this, WHO also forewarns the severity of antibiotic resistance, stating that “it threatens the achievements of modern medicine, a post-antibiotic era—in which common infections and minor injuries can kill—is a very real possibility for the 21st century” [8].
Minimum inhibitory concentrations (MICs) of various antimicrobial susceptibility testing (AST) are categorized by various international agencies. These MIC guidelines determine whether an antibiotic is susceptible or not. The Clinical and Laboratory Standards Institute (CLSI) provides the most popular guidelines, which are based on pharmacokinetic–pharmacodynamic (PK-PD) properties and mechanisms of resistance [9]. Most European countries follow the MIC cut-offs based on PK-PD properties, and the epidemiological MIC cut-offs (ECOFFS) as determined by the European Committee on Antimicrobial Susceptibility Testing (EUCAST). The MIC breakpoints recommended by EUCAST are generally higher than the CLSI. Because of the modifications in the guidelines, the results are substantially changed, such as higher ceftazidime resistance in Klebsiella pneumonia and ESBL-producing Escherichia coli (E. coli). Moreover, the CLSI guidelines re accessible for non-members as a package of three documents for USD 500 annually, while EUCAST guidelines are freely available on the EUCAST website [9].
A sharp surge in bacteria-encoding resistance is occurring worldwide, jeopardizing the efficacy of antibiotics that have saved millions of lives [10]. The antibiotics which have threatened bacteria for decades are under grave threat themselves. Owing to large societal repercussions of multidrug resistance and the significantly reduced development of drugs, it is mandatory to determine the microbes which need more attention than others for drug development. Consequently, WHO has developed a priority list of the pathogens, and stratified the list into critical, high, and medium priorities. Carbapenem-resistant Pseudomonas aeruginosa and Acinetobacter baumanii, carbapenem-resistant and third-generation cephalosporin-resistant Enterobacteriaceae were placed in critical-priority bacteria. The high priority bacteria included vancomycin-resistant Enterococcus faecium, methicillin-resistant Staphylococcus aureus (MRSA), clarithromycin-resistant Helicobacter pylori, fluoroquinolone-resistant Campylobacter spp., penicillin-resistant Streptococcus pneumoniae, ampicillin-resistant Haemophilus influenzae, and fluoroquinolone-resistant Shigella spp. [11] (Figure 1). Resistance against beta-lactam antibiotics like penicillin is widespread, while resistance against other drugs, such as vancomycin and fluoroquinolone, are less frequently observed (Supplementary Table S1). The infections arising from resistant bacteria, because of mutations, might present themselves with harsher symptoms than their predecessors (Supplementary Table S1). Although novel drugs have shown much promise against these resistant bacteria, their rapid diagnostic is still a huge concern (Supplementary Table S1).
In this review, we concisely discuss the advantages and limitations of various tools for AST (Supplementary Table S2). We first review the phenotypic methods for AST like diffusion, dilution, and automated AST tools. Central aspects of genotype-based AST methods, including susceptibility diagnostics based by polymerase chain reaction (PCR) and DNA microarray, are addressed. Then, an overview of several emerging approaches such as fluorescent, colorimetric, and electrochemical microfluidic sensors, with their related caveats, are discussed (Supplementary Table S2).

2. Phenotypic AST Methods

2.1. Diffusion

The disk diffusion method is the gold standard for confirming the susceptibility of bacteria. Standardized disk diffusion was introduced by Bauer and Kirby’s experiments in 1956, after finalizing all aspects of optimization by changing physical conditions [12]. In this method, the isolated bacterial colony is selected, suspended into growth media, and standardized through a turbidity test. The standardized suspension is then inoculated onto the solidified agar plate, and the antibiotic-treated paper is tapped on the inoculated plate. The disc containing the antibiotic is allowed to diffuse through the solidified agar, resulting in the formation of an inhibition zone after the overnight incubation at 35 °C. Thereafter, the size of the inhibition zone formed around the paper disc is measured; the size of the inhibition zone corresponds to the concentration of antibiotic (Figure 2) [12,13]. Assessing and determining the susceptibility of bacteria generally takes 16–24 h. Several diffusion-based experiments have been performed prior to the standardized disk diffusion method. In the 1920s, Fleming was the pioneering contributor to AST. Fleming’s gutter method was the first method of antibiotic analysis where antibiotic was dispensed into a gutter made on solid agar that allowed the antibiotics to diffuse through it [14]. A modification to this design, called the “Oxford cup method,” was subsequently developed by Abraham et al. in 1941, where the gutter was replaced with a glass cup for diffusion [15]. Simultaneously, in the 1940s, Pope (1940), Foster and Woodruff (1943), and Vincet and Vincet (1944) used an antibiotic-impregnated paper disc for the diffusion of antibiotics [16,17]. These methods were hindered by inaccurate analysis due to evaporation, difficulty in handling, sterilization, and cumbersome operation [18]. Moreover, a single antibiotic was focused on susceptibility testing (i.e., penicillin). In later years, the introduction of effective drugs and convenient means of susceptibility testing have prevailed with the increase of the deadly infections. Therefore, variations of the method have been adopted to expand its versatility and utility. In 1947, Hoyt, Levine, and Bondi introduced penicillin tablets and the standard 6.5-mm disk method separately to emphasize multiple targets [19,20]. In the 1950s, experiments by Gould and Bowie (1952) and Stokes (1955) enabled the differentiation between susceptible and resistant bacteria through the multiple disk diffusion technique [21,22]. All the proposed methods were inaccurate, unsuitable, and unreliable for routine testing because of discrepancies in results obtained from different labs [23]. Hence, in 1961, several organizations (especially WHO) made several efforts to address the need for a standardized method for antibiotic susceptibility testing. Later, in the year 1966, Bauer and Kirby’s method was confirmed as a standard method for susceptibility testing. This method has potential for the routine testing of susceptibility in clinical laboratories. Furthermore, the method is widely accepted because it offers a simple, cost-effective protocol for the detection of multiple targets [24]. However, along with these advantages, it also has some significant drawbacks: only semi-automation is available (Sirscan), insufficient data availability for many bacteria (strains of Pseudomonas, Bacillus, and Corynebacterium), and it has a poor performance when analyzing slow-growing and fastidious bacteria [25,26]. Influenced by many physiochemical factors like evaporation, solubility, pH, temperature, and nutrient media, additional limitations restrict its suitability for accurate diagnostics [27].
Recently, the emergence of various instruments for analyzing the zone of inhibition has added to the reliability of the disk diffusion results by reducing variability due to operator handling and interpretation. The camera or scanner takes the picture, and the inbuilt image analysis software displays the zone of inhibition and compares the obtained results with the various guidelines present in the database. Accuzone (AccuMed International, Hillsboro, Oregon, USA), Biomic (Giles Scientific, Santa Barbara, California, USA), Mastascan Elite (Mast, Bootle, UK), and Sirscan (Becton Dickinson, Oxford, UK) are a few of the instruments capable of analyzing the zone of inhibition, but all differ in data input, analysis, ease of use, and presentation of results [28].

2.2. Dilution

Dilution was one of the earliest tools in microbiological practice, starting in the early 1870s, and it allows the growth and identification of bacterial populations in suspension [31]. Pasteur, Lister, Koch, and Ehrlich were listed as the pioneers in the field of bacteriology, and they worked on the concept of macrodilution [32]. William Roberts and John Tyndall further contributed to the macrodilution method and observed bacterial growth in a diluted medium [33]. The two basic types of dilution are microdilution and macrodilution, wherein broth and agar are the most commonly used mediums. In broth dilution, consecutive two-fold dilutions (1, 2, 4, 8, and 12 µL) of antibiotics are made and dispensed into micro-centrifuge tubes containing bacterial growth medium, followed by making up the final volume by adding the medium and incubating overnight at 35 °C. Finally, the growth examination is carried out for setting the breakpoint through the turbidity of culture media (Figure 2) [34,35]. In agar dilution, antibiotics are diluted into the agar medium, followed by plate formation and application of bacterial cells to the surface of the agar plate.
In the early 20th century, various scientists made efforts to introduce serial dilution. They set the dilution factor in terms of geometric progression, and derived the generalized mathematical equation for interpreting the dilution results [36,37,38,39,40]. In 1929, Alexander Fleming performed the serial dilution technique to understand the activity of antibiotics. In this technique, two-fold dilution of antibiotics is mixed with a pre-inoculated liquid medium to determine antibiotic actions by checking the turbidity. In 1942, Fleming modified the previous protocol by using pH instead of turbidity to identify antibacterial activity. In the same year, Rammelkamp and Maxon introduced broth macro dilution, or the “tube dilution method”, which is regarded as the standardized dilution method for both minimum inhibitory concentration (MIC) and AST. CLSI recommends guidelines to set the breakpoints. The first attempt regarding AST was made by Schmith and Reymann using agar medium during the 1940s [41].
Microdilution is a miniaturized prototype of the macrodilution method where susceptibility testing is performed on disposable 96-well microtiter plates, where each well has a sample capacity of ~0.1 mL (Figure 2) [42]. To dispense the samples into microwells, mechanized dispensers are used to avoid the handling error. After overnight incubation, growth and MIC are assessed through specialized optical instruments. This method has been well standardized for most fastidious bacteria [13].
The central drawback of dilution methods is the requirement of a large volume of reagents. Apart from that, other potential limitations include: experimental space, tedious dilution steps (macrodilution), the possibility of false positive results due to long incubation times [43], chances of cross-contamination, bacterial incompatibility for growth, and the inability of discriminating viable and nonviable bacteria. Maintaining the recommended optimum testing parameters like pH, temperature, media, and length of incubation are additional hurdles, and a control viability plate is mandatory in tests to achieve practical clinical relevance [44].

2.3. Etest

Epsilometer testing (Etest) is another significant development for the routine analysis of widespread antibiotic resistance in bacteria. In the late 1980s, Bolmström and Eriksson developed this test [45]. AB BIODISK manufactured the first Etest plastic strip to inspect multiple antibiotics on a single platform in 1991 (Figure 2) [45]. Etest plastic strips are coated with pre-defined antibiotic concentrations, and the corresponding interpretive MIC ranges are marked on the surface and back of the strip, respectively. For detection, multiple strips are placed on a pre-inoculated streaked agar plate, followed by an overnight incubation; elliptical inhibition zones appear around the strips, indicating the MIC at the intersection point between the inhibition zone and the strip edge [46]. The simplicity, accuracy, and reliability of the Etest makes it appropriate and convenient for Food and Drug Administration (FDA) approved commercialization [47]. The ability of convenient interpretations of MIC under diverse physical conditions made the Etest a preferential method over standardized disk diffusion and dilution techniques in clinical laboratories for AST [30,48].
In the 1990s, series of comparative studies with the other standardized techniques, for instance, agar dilution and diffusion, and broth dilution, established the significance of the Etest. Many strains of H. pylori, Neisseria gonorrhoeae, Enterococcus spp., and many other clinical isolates were tested by Etest and compared with standard methods, resulting in a good correlation in the range of 91%–99% [49,50,51]. Recently, in 2016, methicillin-sensitive S. aureus (MSSA) and methicillin-resistant S. aureus (MRSA) isolates were examined with an Etest to determine the MIC of ceftaroline. The results were compared with broth microdilution (BMD) and showed an excellent agreement of more than 95% [52,53]. Multiple cultures of Campylobacter spp. against seven antibiotics were also evaluated by Etest to determine their resistance [54]. All these results demonstrated the reliability and importance of the Etest in evaluating MICs of a wide range of antibiotics over the present standardized methods, especially for slow-growing bacteria (Campylobacter jejuni, H. pylori) and rare fastidious bacteria (S. pneumoniae and Neisseria spp.). One of the significant advantages of the Etest is its sensitivity; it can detect extended-spectrum beta-lactamase (ESBL), even in a trace amount [55]. Additionally, accurate resistance strains can easily be easily quantified in laboratories/hospitals due to the stable concentration gradient of antibiotics marked on the Etest strip.
Besides several advantages, there are some limitations that cannot be ignored, primarily related to the inaccurate and inconsistent behavior of the Etest for certain antibacterial agents, such as Penicillin, ciprofloxacin, ofloxacin, and rifampicin [56]. Some additional demerits that make the Etest complicated for routine test analyses are: pH-sensitive coated antibiotics, expensive batch performance, strip storage, and laboratory set-up for proper plate inoculation and incubation [57].

2.4. Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry (MALDI-TOF MS)

MALDI-TOF MS, introduced in 2000, is another sensitive method for bacterial identification. High sensitivity and accuracy are the key characteristics that make it a useful method for clinical relevance. Assorted studies reveal its significance in discriminating MRSA, MSSA, and other bacterial strains where susceptible and resistant bacteria have been evaluated through spectral peak analysis. Even the subtle difference in expression profiles have been noticed in isogenic strains of S. aureus [58,59]. The efficiency of MALDI-TOF MS has been further investigated on vancomycin-resistant Enterococci, where sensitivity higher than 90% has been recorded. Furthermore, analysis of multiple targets with different resistant strains of Pseudomonas spp. against ciprofloxacin, tobramycin, and meropenem have been identified efficiently [60]. The newly developed MALDI Biotyper antibiotic susceptibility test rapid assay (MBT-ASTRA) is a more-straightforward and cost-effective modulation of MALDI-TOF MS used for both AST and MIC determination [61]. Despite all the advantage of MALDI-TOF MS, the expensive nature of the instrument and its maintenance are prime disadvantages for mass application.

2.5. Automated Systems

Since the dawn of automated technologies in the 1980s, antibiotic susceptibility tests have been perpetually improvised and, hence, have superseded conventional phenotypic methods [13]. Automation, simplicity, and compactness are the major reasons for their widespread acceptance in diagnostics. Computer integration has allowed online analysis and data sharing, which is a giant leap for results validation, especially in remote areas [62]. Among the developed automated systems, MicroScan WalkAway (Beckman Coulter, Inc. Atlanta, Georgia, USA) (1980), Micronaut (Merlin, berlin, Germany) (1990), the avantage test (Abbott Laboratories, Irving, Texas, USA) (1980), Vitek 2 (bioMe’rieux, Marcy-l’Étoile, France) (2000), Phoenix (BD Diagnostics, Franklin Lakes, New jersey, USA) (2001), and Sensititre ARIS 2X (Trek Diagnostic Systems, Oakwood Village, Ohio, USA) (2004) are the major FDA approved systems for AST. Vitek and Pheonix detect growing bacteria on the basis of turbidity, whereas comparable automated systems like MicroScan WalkAway (Beckman Coulter, Inc. Atlanta, Georgia, USA) and Sensititre ARIS 2X (are based on fluorescence emission of the growing bacteria. The resistance in gram-negative, gram-positive, and Streptococcus strains of bacteria can easily be estimated through Phoenix, and Vitek 2, but, MicroScan WalkAway and Sensititre ARIS 2XESBL (Trek Diagnostic Systems, Oakwood Village, Ohio, USA) are capable of detecting the extended-spectrum beta-lactamase (ESBL)-producing strains in the species mentioned above [63]. Micronaut and Avantage are capable of accurate direct susceptibility testing for gram-positive and gram-negative bacteria, respectively [64,65].
Detection incompatibility for many bacteria/antibiotics using Sensititre ARIS 2 and Vitek 1 have led to the development of the updated Sensititre ARIS 2X and Vitek 2, which have better performances and are applicable to a broader range of bacteria/antibiotics. Presently, all the automated systems are incorporated with advanced expert system software for enhanced performance and online data processing [28]. Every automated system has a specific panel capacity and an average time for performing detection, which varies from 40 to 100 wells, and can vary in time such as 20, 12 and 9 h, respectively. Certain models such as Phoenix AP, and Vitek 2Xl, are dedicated towards automated inoculation, enhanced card capacity and compactness.
The aforementioned systems lack reproducibility, sensitivity, and reliability compared with the existing traditional methods. Moreover, an inability to test a wide range of clinically relevant bacteria (e.g., S. pneumonia), antimicrobial agents (e.g., vancomycin), and heteroresistant isolates, as well as a limited panel capacity and the high cost of instruments and consumables, are all significant issues that restrict these systems from frequent analysis [66].

3. Genotypic AST Methods

Molecular or genotypic AST are the effective direct methods that eliminate tedious bacterial cultures, long incubation, chances of contamination, and the spreading of deadly infections [67]. PCR, DNA microarray and DNA chips, and loop-mediated isothermal amplification (LAMP) are some of the genotypic techniques for the detection of antibiotic resistance. Mutational assessment of methicillin resistance in Staphylococcus spp., vancomycin resistance in Enterococcus spp., and multi-antibiotic (isoniazid, rifampin, streptomycin, pyrazinamide, and the fluoroquinolones) resistance in Mycobacterium spp. have been successfully estimated through various genotypic techniques.
PCR is one of the most efficient and rapid molecular tools for quantification and profiling of bacterially infectious genes. The first report on PCR diagnostic application was published by Saiki et al. [68]. The general methodology of PCR includes cycles of denaturation, annealing of the primers, and elongation of the primers by a thermostable DNA polymerase in a compatible buffer containing nucleotides, ions, and so on. Each cycle of amplification doubles the target DNA molecule. The amplified target can be confirmed for the presence of resistance genes through electrophoresis, southern blotting, restriction fragment-length polymorphism, single-strand conformation polymorphism (SSCP), DNA fingerprinting, molecular beacons, and other DNA sequencing analysis methods (Figure 3) [68,69]. Another tool developed on the basis of PCR is LAMP, which has also been used for the evaluation of AST. In LAMP, the gene of interest is amplified at a constant temperature of 60–65 °C using a Bst DNA polymerase instead of Taq polymerase because of strong strand displacement activity (required in isothermal techniques) [70].
DNA microarrays and DNA chips are the other promising technologies utilized for screening susceptibility [72]. DNA arrays employ cDNA fragment probes on nylon membrane, where each DNA chip has a glass or silicon platform for probe binding. The specific hybridization of the labeled probe with the target and its recognition help to determine the resistance. Determination of isoniazid resistance in M. tuberculosis has been carried out successfully through DNA microarrays and chips [73,74]. Colorimetric detection and multiplexing are the attractive features of these techniques.
Genotypic methods are generally attributed to the rapid, direct, sensitive, and specific detection of resistance genes, but they also suffer from severe drawbacks that diminish their clinical utility. These drawbacks include: (i) the individual antimicrobial agents to be tested need a specific assay for detection; (ii) only potential/key resistance genes can be detected, which are often not relevant due to coincidental mutations; (iii) there is a lack of sensitivity towards the patients with latent infections, or when only a few organisms are present in a sample; (iv) the genetic mechanism/profile for the resistance of all bacteria is not yet defined; (v) the occurrence of false-positive results due to contamination of the test sample might be expected; (vi) they require expensive reagents and machinery with specific maintenance conditions; and most importantly (vii) all the tools have a prerequisite of skilled personnel [67].

4. Emerging Methods for AST

Microfluidics-based diagnostics are one of the most promising emerging tools for AST. Microfluidics is an evolving field characterized by the manipulation of fluids in micro-volume, thereby offering portability, cost-effectiveness, multiplexing, reproducibility, and a controllable environment in an in vitro system [12]. The concept was first introduced in the semiconductor and micro-electromechanical systems (MEMS) industries, then further extended to the field of biomedical research [75]. Integrated microfluidic devices, generally referred to as micro total analysis systems (µTAS), are proficient in performing molecular diagnostics [76].
The quantity of samples has always been the biggest challenge for biological studies in general, and pathological analysis in particular. Since microfluidics is capable of dealing with the minimal quantity of samples, it has therefore emerged as a promising tool for pathologists. Currently, microfluidics platforms are capable of single-cell analysis, and can even analyze the single-cell interrogation of signaling networks in cultured cell lines. In the past decades, numerous conventional and automated strategies have been successfully embraced to diagnose the ever-increasing antibiotic resistance in bacteria [77]. Nevertheless, the most-practiced techniques, such as disk diffusion and microdilution, have a high chance of cross-contamination, laborious protocols, lengthy processing times, improper power supplies, and cumbersome set-ups that render their application in resource-limited regions, and unfortunately, these regions have a higher rate of resistance too [2,6]. The rise of microfluidics as a diagnostic tool has shown promise in addressing the abovementioned shortcomings [78,79].
Progression in optical imaging has developed various image sensors with high sensitivity and resolution that are capable of biological analysis. These imaging systems are frequently used for morphological and growth studies of bacteria. Generally, owing to real-time analysis and minimum culture dependency, microfluidic devices coupled with an optical sensor can perform AST and detect the MIC in few hours [80]. Recent reports based on single bacterial cell analysis have claimed that optical sensor-based nanofluidic (30 nl) can finish AST within 30 min [81]. This direct imaging of single bacterium requires simple sample preparation steps, but eliminates the tedious steps of continuous sample injection, loading of cells, and counting-based cell identification [81].
Fluorescence proteins and dyes are commonly used for tagging resistant biomarkers. These proteins/dyes can be of biological or chemical origin. One of the pioneer proteins used for imaging is green fluorescent protein (GFP). GFP, obtained from jellyfish Aequoria victoria, is essential for the noninvasive real-time monitoring of antimicrobial susceptibility [82]. Adding to this, the same group was able to evaluate multiple antibiotic sensitivities in real-time in polymicrobial culture bacterial strains (namely, E. coli, P. aeruginosa, and K. pneumoniae) simultaneously. Green and red fluorescent protein tagging have been utilized for real-time growth quantification in the presence of multiple antibiotics on a multiplexed microfluidic platform (Figure 4) [83]. All these methods are sensitive and reliable, but creating these recombinant bacteria requires molecular handling, which is a challenge for routine clinical observations. Fluorescent dyes are another means for optical fluorescence tagging. These chemicals, with all the benefits of the fluorescence protein, can avoid stearic hindrance due to their small size, and elimination of cloning or transformation. SYTOX green and resazurin (alamarBlue; inactive precursor of resazurin) are two common fluorescence indicator dyes used for viability testing in AST [84,85]. Similarly, resazurin can be utilized in colorimetric detection. When the culture media is supplemented with antibiotic and resazurin, a uniform intense blue color is obtained in the beginning. In the presence of resistance, the resazurin is reduced to resafurin, and the intense blue color changes to pink and leuco; in the absence of resistance in bacteria, the blue color sustains. This method is also translated to a microfluidic chip for MIC estimation for four different antibiotics against 20 clinical strains of Escherichia and Shigella [85].
Similarly, bioluminescence or ATP bioluminescence assay (ATP-BLA) is an enzyme-based approach mediated by luciferase enzyme that converts luciferin substrate to oxyluciferin in the presence of ATP, leading to an emission of light [87]. By using this phenomenon, the susceptibility of 13 different types of clinical strains present in urinary infections were evaluated against eight antibiotics on the microfluidic plate. When bacteria grew in the presence of antibiotics, resistant bacteria resulted in bioluminescence, whereas susceptible bacteria remained neutral. Both identification and susceptibility were obtained within 3–6 h [88,89].
Several electrochemical devices have been developed for AST. One of the most prominent works utilized AC electrokinetic fluid motion and Joule heating-induced temperature elevation for the electrochemical sensing of bacterial 16S rRNA (Figure 4) [86]. Real-time and rapid detection are possible, as 16s rRNA is highly specific for the bacterial pathogens in blood culture and does not need prior purification. Although 16S rRNA has indeed provided critical details about sensitive bacterial analysis, genetic complexities across kingdoms make it inappropriate for reproducible, clinically relevant, or point-of-care susceptibility testing. The use of electroactive chemicals (redox reagents) as probe molecules can be an interesting approach to elucidating bacterial susceptibility. In 2015, pyocyanin, a potential marker of cell viability and virulence, was studied for the electrochemical monitoring of the susceptibility of P. aeruginosa biofilms on a microfluidic device [90]. A good correlation between the electrical signal drop and the viability of P. aeruginosa cells in the presence of antibiotics was successfully demonstrated. The most recent electrochemical biosensor can perform AST within 90 min along with the label-free isolation of bacteria from whole blood samples [91]. Plastic-based microchips with printed electrodes capture the target bacteria with the help of specific antibodies. The electrochemical response to the captured bacteria is monitored in both the presence and absence of antibiotics.
The simplified blood culture system (SBCS) is an emerging tool in near-patient testing and surveillance tools for blood stream infections (BSI) [92]. SBCSs uses unprocessed samples, thereby, it requires zero sample preparation time. Furthermore, SBCSs combine detection, gram status, and identification in blood culture (BC) instruments, thus ensuring the evaluation of susceptibility within 8–12 h, whereas the conventional blood culture system requires up to 48 h because of incubation for colony generation and susceptibility testing. Moreover, owing to its simplified and time efficient nature, the SBCS can be used in resource-limited settings, which was not possible for conventional BC methods due to the lack of electricity, limited culture bottles, profuse dust, improper ambient temperature control, and lack of skilled personnel [92].

5. Challenges and Future Perspective

Primarily, the continuous flow of culturing media to feed cells is the biggest challenge for maintaining nutrient conditions, as a slight change in growth media due to evaporation can affect the bacterial growth and hence the accuracy of AST [93,94]. The accuracy of AST might be improved by considering the factors altering its pharmacokinetics (diffusion, metabolism, and elimination), which may result in unpredictable changes, and therefore, the MIC might change. Furthermore, to avoid the indiscriminate use of antibiotics and the evolution of antibiotic resistance, the dosage scheme should also consider the method of antibiotic administration (e.g., oral versus intravenous (IV) administration) and the site of infection, along with the AST results [95].
The applicability of molecular biology tools such as cloning and recombinant expression has enhanced the sensitivity of detection, and low concentrations of bacteria in clinical samples can be evaluated, but there are some serious limitations. Creating recombinant strains with these molecular genes is a troublesome process. Firstly, genetic analysis is cumbersome and prone to mutations due to frequent change in the resistant behavior of bacteria [96]. Therefore, prior knowledge of specific resistance genes before susceptibility testing is essential. Secondly, the genetic markers for all clinically relevant bacteria is still unknown, moreover, the known targets are not universal. Thirdly, advanced molecular biology skills and laboratory sets are important in dealing with recombinant technology. Alternatively, fluorescence label dyes are used to avoid molecular challenges. Although the use of dyes is simple and easy over other molecular techniques, the requirement of a high-resolution A charge-coupled device camera (CCD) and sophisticated instruments for signal amplification and observation are restricted to resource-limited areas. False-positive results due to changes in physical parameters are also a major problem. Additionally, immense versatility among the culture conditions of different bacterial species is a challenge for the development of a single platform for different bacteria and multiplexing [97,98].
Looking beyond the imaging requirements, researchers must also focus on other relevant unanswered questions in the context of diagnostic devices. This primarily includes what performance metrics will be essential to lessen the exposure of contamination from hospitals or biomedical research units, and secondarily, what advances will be necessary to reduce the incorporation of sophisticated external circuits, pumps, and pneumatic systems in maintaining the flow continuity in microfluidic platform To address these issues, the paper-based detection system seems to be an attractive approach in developing a cost-effective, automated, and incinerable platform for the determination of susceptibility. In the past decades, the emergence of paper-based microfluidics has proven its performance for biological assays [99], and it could be a bedrock for reducing contamination through easy disposal. Additionally, the inbuilt property of capillary action of paper can eliminate the integration of complicated pneumatic chambers and pumps. Therefore, more research work on paper microfluidic AST would be beneficial in the future. An absence of the simultaneous detection of multiple analytes and the need for simpler fabrication tools are further shortcomings of microfluidics. The collaboration of academic research and commercial firms with transparent technology distribution might offer a compelling solution for multiplexing and for more-straightforward fabrication. Hence, efforts to expand our understanding, especially for the development of a user-friendly device with multiplexing, would be valuable.
Smartphones and the technology that powers them are continually growing more advanced. Coupling fluorescent/colorimetric tools with the ubiquitous and ever-evolving smartphones will enable us with on-site monitoring, real-time database updates, and the generation of an antibiotic susceptibility map to help us understand the geographical prevalence of resistance. The use of smartphones is limited by their camera performances, which result in lower detection limits, especially in a colorimetric assay [100]. It is exciting to speculate that ongoing advancement will bring much higher resolution cameras coupled with better time-lapse technologies and offer morphological and biochemical measurements.
Loop-mediated isothermal amplification polymerase chain reaction (LAMP-PCR) has shown us the way to develop lateral-flow devices for the genetic detection of antibiotic resistance. Carbapenem-resistance in Acinetobacter baumanii was successfully evaluated by LAMP-PCR by amplifying the OXA-type carbapenemases and metallo-β-lactamases genes [101]. However, more studies are warranted to establish the applicability of LAMP. In coming years, there might be a rise in non-infecting but resistance-bearing mutants; genetic detection will be essential to screen out these mutants, and LAMP-based lateral flow devices will serve that purpose.

6. Conclusions

While all AST methods offer qualitative assessments using susceptible, intermediate, or resistant categories, certain methods specify qualitative and effective antibiotic dosage (e.g., minimum inhibitory concentration) and formulate a profile of empirical therapy for the proper management of individual patients’ health against deadly infections. A rise in antibiotic resistance is a certainty, therefore we must develop technologies that will permit rapid AST (within an hour) and are non-invasive (saliva- or urine-based) or minimally invasive.

Supplementary Materials

The following are available online at https://www.mdpi.com/2075-4418/9/2/49/s1.

Author Contributions

Z.A.K., M.F.S., and S.P. wrote and edited the manuscript. All the authors read and approved the final manuscript.

Funding

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIT) (NRF-2015R1C1A1A01054762).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Blair, J.M.; Webber, M.A.; Baylay, A.J.; Ogbolu, D.O.; Piddock, L.J. Molecular mechanisms of antibiotic resistance. Nat. Rev. Microbiol. 2015, 13, 42–51. [Google Scholar] [CrossRef] [PubMed]
  2. Martínez, J.L. Antibiotics and Antibiotic Resistance Genes in Natural Environments. Science 2008, 321, 365–367. [Google Scholar] [CrossRef] [PubMed]
  3. Lalitha, M. Manual on antimicrobial susceptibility testing. In Performance Standards for Antimicrobial Testing: Twelfth Informational Supplement; CLSI: Wayne, PA, USA, 2004; Volume 56238, pp. 454–456. [Google Scholar]
  4. Van Boeckel, T.P.; Gandra, S.; Ashok, A.; Caudron, Q.; Grenfell, B.T.; Levin, S.A.; Laxminarayan, R. Global antibiotic consumption 2000 to 2010: An analysis of national pharmaceutical sales data. Lancet Infect. Dis. 2014, 14, 742–750. [Google Scholar] [CrossRef]
  5. Lee, C.-R.; Cho, I.H.; Jeong, B.C.; Lee, S.H. Strategies to Minimize Antibiotic Resistance. Int. J. Environ. Res. Public Health 2013, 10, 4274–4305. [Google Scholar] [CrossRef]
  6. Ayukekbong, J.A.; Ntemgwa, M.; Atabe, A.N. The threat of antimicrobial resistance in developing countries: Causes and control strategies. Antimicrob. Resist. Infect. Control 2017, 6, 47. [Google Scholar] [CrossRef]
  7. Brogan, D.M.; Mossialos, E. A critical analysis of the review on antimicrobial resistance report and the infectious disease financing facility. Glob. Health 2016, 12, 8. [Google Scholar] [CrossRef]
  8. Viens, A.M.; Littmann, J. Is Antimicrobial Resistance a Slowly Emerging Disaster? Public Health Ethics 2015, 8, 255–265. [Google Scholar] [CrossRef]
  9. Kassim, A.; Omuse, G.; Premji, Z.; Revathi, G. Comparison of Clinical Laboratory Standards Institute and European Committee on Antimicrobial Susceptibility Testing guidelines for the interpretation of antibiotic susceptibility at a University teaching hospital in Nairobi, Kenya: A cross-sectional study. Ann. Clin. Microbiol. Antimicrob. 2016, 15, 21. [Google Scholar] [CrossRef] [PubMed]
  10. Ventola, C.L. The antibiotic resistance crisis: Part 1: Causes and threats. Pharm. Ther. 2015, 40, 277–283. [Google Scholar]
  11. Tacconelli, E.; Carrara, E.; Savoldi, A.; Harbarth, S.; Mendelson, M.; Monnet, D.L.; Pulcini, C.; Kahlmeter, G.; Kluytmans, J.; Carmeli, Y.; et al. Discovery, research, and development of new antibiotics: The WHO priority list of antibiotic-resistant bacteria and tuberculosis. Lancet Infect. Dis. 2018, 18, 318–327. [Google Scholar] [CrossRef]
  12. Bauer, A.W.; Kirby, W.M.; Sherris, J.C.; Turck, M. Antibiotic susceptibility testing by a standardized single disk method. Am. J. Clin. Pathol. 1966, 45, 493–496. [Google Scholar] [CrossRef] [PubMed]
  13. Reller, L.B.; Weinstein, M.; Jorgensen, J.H.; Ferraro, M.J. Antimicrobial Susceptibility Testing: A Review of General Principles and Contemporary Practices. Clin. Infect. Dis. 2009, 49, 1749–1755. [Google Scholar] [CrossRef]
  14. Fleming, A. On the antibacterial action of cultures of a penicillium, with special reference to their use in the isolation of B. influenzae. Br. J. Exp. Pathol. 1929, 10, 226. [Google Scholar] [CrossRef]
  15. Abraham, E.P.; Chain, E.; Fletcher, C.M.; Gardner, A.D.; Heatley, N.G.; Jennings, M.A.; Florey, H.W. Further observations on penicillin. Lancet 1941, 238, 177–189. [Google Scholar] [CrossRef]
  16. Foster, J.W.; Woodruff, H.B. Microbiological aspects of penicillin: I. Methods of assay. J. Bacteriol. 1943, 46, 187. [Google Scholar]
  17. Vincent, J.G.; Vincent, H.W.; Morton, J. Filter Paper Disc Modification of the Oxford Cup Penicillin Determination. Proc. Soc. Exp. Biol. Med. 1944, 55, 162–164. [Google Scholar] [CrossRef]
  18. Gavin, J.J. Analytical Microbiology: II. The Diffusion Methods. Appl. Microbiol. 1957, 5, 25. [Google Scholar]
  19. Hoyt, R.E.; Levine, M.G. A Method for determining Sensitivity to Penicillin and Streptomycin. Science (Washington) 1947, 171. [Google Scholar] [CrossRef]
  20. Bondi, A., Jr.; Spaulding, E.H.; Smith, D.E.; Dietz, C.C. A routine method for the rapid determination of susceptibility to penicillin and other antibiotics. Am. J. Med. Sci. 1947, 213, 221–225. [Google Scholar] [CrossRef]
  21. Gould, J.C. The determination of bacterial sensitivity to antibiotics. Edinb. Med. J. 1952, 59, 178–199. [Google Scholar] [PubMed]
  22. Stokes, E.J. Antibiotic sensitivity testing. Br. Med. J. 1971, 2, 707. [Google Scholar] [CrossRef]
  23. Mayrhofer, S.; Domig, K.J.; Mair, C.; Zitz, U.; Huys, G.; Kneifel, W. Comparison of broth microdilution, Etest, and agar disk diffusion methods for antimicrobial susceptibility testing of Lactobacillus acidophilus group members. Appl. Environ. Microbiol. 2008, 74, 3745–3748. [Google Scholar] [CrossRef]
  24. World Health Organization. Standardization of Methods for Conducting Microbic Sensitivity Tests: Second Report of the Expert Committee on Antibiotics [Meeting Held in Geneva from 11 to 16 July 1960]; WHO: Geneva, Switzerland, 1961. [Google Scholar]
  25. Patel, J.B.; Tenover, F.C.; Turnidge, J.D.; Jorgensen, J.H. Susceptibility test methods: Dilution and disk diffusion methods. In Manual of Clinical Microbiology, 10th ed.; American Society of Microbiology: Sterling, VA, USA, 2011; pp. 1122–1143. [Google Scholar]
  26. Hombach, M.; Zbinden, R.; Bottger, E.C. Standardisation of disk diffusion results for antibiotic susceptibility testing using the sirscan automated zone reader. BMC Microbiol. 2013, 13, 225. [Google Scholar] [CrossRef]
  27. Murray, P.R. Antibiotic Susceptibility Testing. Part I. Lab. Med. 1983, 14, 345–350. [Google Scholar] [CrossRef]
  28. Felmingham, D.; Brown, D.F. Instrumentation in antimicrobial susceptibility testing. J. Antimicrob. Chemother. 2001, 48, 81–85. [Google Scholar] [CrossRef]
  29. Sageerabanoo, S.; Malini, A.; Mangaiyarkarasi, T.; Hemalatha, G. Phenotypic detection of extended spectrum β-lactamase and Amp-C β-lactamase producing clinical isolates in a Tertiary Care Hospital: A preliminary study. J. Nat. Sci. Biol. Med. 2015, 6, 383–387. [Google Scholar] [CrossRef] [PubMed]
  30. Sader, H.S.; Pignatari, A.C. E test: A novel technique for antimicrobial susceptibility testing. Sao Paulo Med. J. Rev. Paul. Med. 1994, 112, 635–638. [Google Scholar] [CrossRef] [PubMed]
  31. Guardino, R.F. Early History of Microbiology and Microbiological Methods; Parenteral Drug Association: Wilmington, DE, USA, 2005. [Google Scholar]
  32. Rittenberg, S.C. Three Centuries of Microbiology. Hubert, A. Lechevalier and Morris Solotorovsky. McGraw-Hill, New York, 1965. viii + 536 pp. Paper, $4.95. Science 1965, 149, 530–531. [Google Scholar] [CrossRef]
  33. Poupard, J.A.; Rittenhouse, S.F.; Walsh, L.R. The evolution of antimicrobial susceptibility testing methods. In Antimicrobial Susceptibility Testing; Springer, Plenum Publishing: New York, NY, USA, 1994; pp. 3–14. [Google Scholar]
  34. Ericsson, H.M.; Sherris, J.C. Antibiotic sensitivity testing. Report of an international collaborative study. Acta Pathol. Microbiol. Scand. 1971, 90. [Google Scholar]
  35. Jorgensen, J.H.; Turnidge, J.D.; Washington, J.A.; Murray, P.R.; Pfaller, M.A.; Tenover, F.C.; Baron, E.J.; Yolken, R.H. Antibacterial susceptibility tests: Dilution and disk diffusion Methods. In Manual of Clinical Microbiology; Geo. F. Brooks Publisher: Washington, DC, USA, 1999. [Google Scholar]
  36. Phelps, E.B. A method of calculating the numbers of B. coli from the results of dilution tests. Am. J. Public Hyg. 1908, 18, 141. [Google Scholar]
  37. McCrady, M. Tables for rapid interpretation of fermentation-tube results. Public Health J. 1918, 9, 201–220. [Google Scholar]
  38. Reed, J. Report of Advisory Committee on Official Water Standards; Public Health Reports; Sage Publications, Inc.: Teller Road, AK, USA, 1925; Volume 40. [Google Scholar]
  39. Greenwood, M.; Yule, G.U. On the statistical interpretation of some bacteriological methods employed in water analysis. J. Hyg. 1917, 16, 36. [Google Scholar] [CrossRef]
  40. Fisher, R.A. Statistical methods for research workers. In Breakthroughs in Statistics; Springer, Oliver & Boyd: New York, NY, USA, 1992; pp. 66–70. [Google Scholar]
  41. Wheat, P.F. History and development of antimicrobial susceptibility testing methodology. J. Antimicrob. Chemother. 2001, 48, 1–4. [Google Scholar] [CrossRef]
  42. Tang, Y.-W.; Stratton, C.W. Advanced Techniques in Diagnostic Microbiology; Springer: Nashville, TN, USA, 2012. [Google Scholar]
  43. Waites, K.B.; Duffy, L.B.; Bébéar, C.M.; Matlow, A.; Talkington, D.F.; Kenny, G.E.; Totten, P.A.; Bade, D.J.; Zheng, X.; Davidson, M.K. Standardized methods and quality control limits for agar and broth microdilution susceptibility testing of Mycoplasma pneumoniae, Mycoplasma hominis, and Ureaplasma urealyticum. J. Clin. Microbiol. 2012, 50, 3542–3547. [Google Scholar] [CrossRef]
  44. Collins, A.M.; Craig, G.; Zaiman, E.; Roy, T.E. A comparison between disk-plate and tube-dilution methods for antibiotic sensitivity testing of bacteria. Can. J. Public Health/Rev. Can. Sante’e Publique 1954, 45, 430–439. [Google Scholar]
  45. Picard, J. Applied Veterinary Bacteriology and Mycology: Bacteriological Techniques; University of Pretoria, Afrivip: Pretoria, South Africa, 1990. [Google Scholar]
  46. Sanchez, M.L.; Jones, R.N. E test, an antimicrobial susceptibility testing method with broad clinical and epidemiologic application. Antimicrob. Newsl. 1992, 8, 1–7. [Google Scholar] [CrossRef]
  47. Joyce, L.F.; Downes, J.; Stockman, K.; Andrew, J.H. Comparison of five methods, including the PDM Epsilometer test (E test), for antimicrobial susceptibility testing of Pseudomonas aeruginosa. J. Clin. Microbiol. 1992, 30, 2709–2713. [Google Scholar]
  48. Baker, C.N.; Stocker, S.A.; Culver, D.H.; Thornsberry, C. Comparison of the E Test to agar dilution, broth microdilution, and agar diffusion susceptibility testing techniques by using a special challenge set of bacteria. J. Clin. Microbiol. 1991, 29, 533–538. [Google Scholar]
  49. Glupczynski, Y.; Labbé, M.; Hansen, W.; Crokaert, F.; Yourassowsky, E. Evaluation of the E test for quantitative antimicrobial susceptibility testing of Helicobacter pylori. J. Clin. Microbiol. 1991, 29, 2072–2075. [Google Scholar]
  50. Van Dyck, E.; Smet, H.; Piot, P. Comparison of E test with agar dilution for antimicrobial susceptibility testing of Neisseria gonorrhoeae. J. Clin. Microbiol. 1994, 32, 1586–1588. [Google Scholar]
  51. Schulz, J.E.; Sahm, D.F. Reliability of the E test for detection of ampicillin, vancomycin, and high-level aminoglycoside resistance in Enterococcus spp. J. Clin. Microbiol. 1993, 31, 3336–3339. [Google Scholar]
  52. Cantón, R.; Livermore, D.M.; Morosini, M.I.; Díaz-Regañón, J.; Rossolini, G.M. Etest® versus broth microdilution for ceftaroline MIC determination with Staphylococcus aureus: Results from PREMIUM, a European multicentre study. J. Antimicrob. Chemother. 2017, 72, 431–436. [Google Scholar] [CrossRef]
  53. Skov, R.; Smyth, R.; Larsen, A.R.; Bolmstrom, A.; Karlsson, A.; Mills, K.; Frimodt-Moller, N.; Kahlmeter, G. Phenotypic detection of methicillin resistance in Staphylococcus aureus by disk diffusion testing and Etest on Mueller-Hinton agar. J. Clin. Microbiol. 2006, 44, 4395–4399. [Google Scholar] [CrossRef] [PubMed]
  54. Hariharan, H.; Sharma, S.; Chikweto, A.; Matthew, V.; DeAllie, C. Antimicrobial drug resistance as determined by the E-test in Campylobacter jejuni, C. coli, and C. lari isolates from the ceca of broiler and layer chickens in Grenada. Comp. Immunol. Microbiol. Infect. Dis. 2009, 32, 21–28. [Google Scholar] [CrossRef]
  55. Falagas, M.E.; Karageorgopoulos, D.E. Extended-spectrum β-lactamase-producing organisms. J. Hosp. Infect. 2009, 73, 345–354. [Google Scholar] [CrossRef]
  56. Frosch, M.; Maiden, M.C.J. Handbook of Meningococcal Disease: Infection Biology, Vaccination, Clinical Management; John Wiley & Sons: Hoboken, NJ, USA, 2006. [Google Scholar]
  57. Balouiri, M.; Sadiki, M.; Ibnsouda, S.K. Methods for in vitro evaluating antimicrobial activity: A review. J. Pharm. Anal. 2016, 6, 71–79. [Google Scholar] [CrossRef]
  58. Bernardo, K.; Pakulat, N.; Macht, M.; Krut, O.; Seifert, H.; Fleer, S.; Hünger, F.; Krönke, M. Identification and discrimination of Staphylococcus aureus strains using matrix-assisted laser desorption/ionization-time of flight mass spectrometry. Proteomics 2002, 2, 747–753. [Google Scholar] [CrossRef]
  59. Hrabák, J.; Chudáčková, E.; Walková, R. Matrix-assisted laser desorption ionization–time of flight (MALDI-TOF) mass spectrometry for detection of antibiotic resistance mechanisms: From research to routine diagnosis. Clin. Microbiol. Rev. 2013, 26, 103–114. [Google Scholar] [CrossRef] [PubMed]
  60. Jung, J.; Eberl, T.; Sparbier, K.; Lange, C.; Kostrzewa, M.; Schubert, S.; Wieser, A. Rapid detection of antibiotic resistance based on mass spectrometry and stable isotopes. Eur. J. Clin. Microbiol. Infect. Dis. 2014, 33, 949–955. [Google Scholar] [CrossRef]
  61. Zimmermann, S.; Burckhardt, I. Development and Application of MALDI-TOF for Detection of Resistance Mechanisms. In MALDI-TOF and Tandem MS for Clinical Microbiology; John Wiley & Sons: Hoboken, NJ, USA, 2017; pp. 231–248. [Google Scholar]
  62. Richter, S.S.; Ferraro, M.J. Susceptibility testing instrumentation and computerized expert systems for data analysis and interpretation. In Manual of Clinical Microbiology, 10th ed.; American Society of Microbiology: Sterling, VA, USA, 2011; pp. 1144–1154. [Google Scholar]
  63. Sellenriek, P.; Holmes, J.; Ferrett, R.; Drury, R.; Storch, G.A. Comparison of MicroScan Walk-Away®, Phoenix™ and VITEK-TWO® Microbiology systems used in the identification and susceptibility testing of bacteria. In Proceedings of the Abstr 105th General Meeting of the American Society for Microbiology, Atlanta, GA, USA, 5–9 June 2005. [Google Scholar]
  64. Wright, D.N.; Matsen, J.M.; DiPersio, J.R.; Kirk, M.; Saxon, B.; Ficorilli, S.M.; Spencer, H.J. Evaluation of four newer antimicrobial agents in the Avantage susceptibility test system. J. Clin. Microbiol. 1989, 27, 2381–2383. [Google Scholar]
  65. Wellinghausen, N.; Pietzcker, T.; Poppert, S.; Belak, S.; Fieser, N.; Bartel, M.; Essig, A. Evaluation of the Merlin MICRONAUT System for Rapid Direct Susceptibility Testing of Gram-Positive Cocci and Gram-Negative Bacilli from Positive Blood Cultures. J. Clin. Microbiol. 2007, 45, 789–795. [Google Scholar] [CrossRef] [PubMed]
  66. Karlowsky, J.A.; Richter, S.S. Antimicrobial susceptibility testing systems. In Manual of Clinical Microbiology, 11th ed.; American Society of Microbiology: Sterling, VA, USA, 2015; pp. 1274–1285. [Google Scholar]
  67. Cockerill, F.R. Genetic methods for assessing antimicrobial resistance. Antimicrob. Agents Chemother. 1999, 43, 199–212. [Google Scholar] [CrossRef]
  68. Fluit, A.C.; Visser, M.R.; Schmitz, F.-J. Molecular detection of antimicrobial resistance. Clin. Microbiol. Rev. 2001, 14, 836–871. [Google Scholar] [CrossRef] [PubMed]
  69. Miller, M.B.; Tang, Y.-W. Basic Concepts of Microarrays and Potential Applications in Clinical Microbiology. Clin. Microbiol. Rev. 2009, 22, 611–633. [Google Scholar] [CrossRef]
  70. Li, Y.; Fan, P.; Zhou, S.; Zhang, L. Loop-mediated isothermal amplification (LAMP): A novel rapid detection platform for pathogens. Microb. Pathog. 2017, 107, 54–61. [Google Scholar] [CrossRef]
  71. Athamanolap, P.; Hsieh, K.; Chen, L.; Yang, S.; Wang, T.H. Integrated Bacterial Identification and Antimicrobial Susceptibility Testing Using PCR and High-Resolution Melt. Anal. Chem. 2017, 89, 11529–11536. [Google Scholar] [CrossRef]
  72. Frye, J.G.; Lindsey, R.L.; Rondeau, G.; Porwollik, S.; Long, F.; McClelland, M.; Jackson, C.R.; Englen, M.D.; Meinersmann, R.J.; Berrang, M.E.; et al. Development of a DNA microarray to detect antimicrobial resistance genes identified in the National Center for Biotechnology Information database. Microb. Drug Resist. 2010, 16, 9–19. [Google Scholar] [CrossRef]
  73. Gryadunov, D.; Mikhailovich, V.; Lapa, S.; Roudinskii, N.; Donnikov, M.; Pan’kov, S.; Markova, O.; Kuz’min, A.; Chernousova, L.; Skotnikova, O.; et al. Evaluation of hybridisation on oligonucleotide microarrays for analysis of drug-resistant Mycobacterium tuberculosis. Clin. Microbiol. Infect. 2005, 11, 531–539. [Google Scholar] [CrossRef] [PubMed]
  74. Huang, W.L.; Hsu, Z.J.; Chang, T.C.; Jou, R. Rapid and accurate detection of rifampin and isoniazid-resistant Mycobacterium tuberculosis using an oligonucleotide array. Clin. Microbiol. Infect. 2014, 20, O542–O549. [Google Scholar] [CrossRef] [PubMed]
  75. Sackmann, E.K.; Fulton, A.L.; Beebe, D.J. The present and future role of microfluidics in biomedical research. Nature 2014, 507, 181–189. [Google Scholar] [CrossRef]
  76. Park, S.; Zhang, Y.; Lin, S.; Wang, T.H.; Yang, S. Advances in microfluidic PCR for point-of-care infectious disease diagnostics. Biotechnol. Adv. 2011, 29, 830–839. [Google Scholar] [CrossRef]
  77. Ibrahim, W.A.; Marouf, S.A.; Erfan, A.M.; Nasef, S.A.; Jakee, J.K.E. The occurrence of disinfectant and antibiotic-resistant genes in Escherichia coli isolated from chickens in Egypt. Vet. World 2019, 12, 141–145. [Google Scholar] [CrossRef] [PubMed]
  78. Chin, C.D.; Laksanasopin, T.; Cheung, Y.K.; Steinmiller, D.; Linder, V.; Parsa, H.; Wang, J.; Moore, H.; Rouse, R.; Umviligihozo, G.; et al. Microfluidics-based diagnostics of infectious diseases in the developing world. Nat. Med. 2011, 17, 1015–1019. [Google Scholar] [CrossRef]
  79. Martinez, A.W.; Phillips, S.T.; Whitesides, G.M.; Carrilho, E. Diagnostics for the Developing World: Microfluidic Paper-Based Analytical Devices; ACS Publications: Sterling, VA, USA, 2009. [Google Scholar]
  80. Chen, C.H.; Lu, Y.; Sin, M.L.Y.; Mach, K.E.; Zhang, D.D.; Gau, V.; Liao, J.C.; Wong, P.K. Antimicrobial susceptibility testing using high surface-to-volume ratio microchannels. Anal. Chem. 2010, 82, 1012–1019. [Google Scholar] [CrossRef]
  81. Baltekin, Ö.; Boucharin, A.; Andersson, D.I.; Elf, J. Fast Antibiotic Susceptibility Testing (FASTest) based on single cell growth rate measurements. bioRxiv 2016. [Google Scholar] [CrossRef]
  82. Webb, J.S.; Barratt, S.R.; Sabev, H.; Nixon, M.; Eastwood, I.M.; Greenhalgh, M.; Handley, P.S.; Robson, G.D. Green fluorescent protein as a novel indicator of antimicrobial susceptibility in Aureobasidium pullulans. Appl. Environ. Microbiol. 2001, 67, 5614–5620. [Google Scholar] [CrossRef]
  83. Mohan, R.; Sanpitakseree, C.; Desai, A.V.; Sevgen, S.E.; Schroeder, C.M.; Kenis, P.J.A. A microfluidic approach to study the effect of bacterial interactions on antimicrobial susceptibility in polymicrobial cultures. RSC Adv. 2015, 5, 35211–35223. [Google Scholar] [CrossRef]
  84. Kalashnikov, M.; Lee, J.C.; Campbell, J.; Sharon, A.; Sauer-Budge, A.F. A microfluidic platform for rapid, stress-induced antibiotic susceptibility testing of Staphylococcus aureus. Lab Chip 2012, 12, 4523–4532. [Google Scholar] [CrossRef]
  85. Elavarasan, T.; Chhina, S.K.; Ash, M.P.; Sankaran, K. Resazurin reduction based colorimetric antibiogram in microfluidic plastic chip. Sens. Actuators B Chem. 2013, 176, 174–180. [Google Scholar] [CrossRef]
  86. Liu, T.; Lu, Y.; Gau, V.; Liao, J.C.; Wong, P.K. Rapid antimicrobial susceptibility testing with electrokinetics enhanced biosensors for diagnosis of acute bacterial infections. Ann. Biomed. Eng. 2014, 42, 2314–2321. [Google Scholar] [CrossRef]
  87. Marques, S.M.; Esteves da Silva, J.C. Firefly bioluminescence: A mechanistic approach of luciferase catalyzed reactions. IUBMB Life 2009, 61, 6–17. [Google Scholar] [CrossRef]
  88. Dong, T.; Zhao, X. Rapid identification and susceptibility testing of uropathogenic microbes via immunosorbent ATP-bioluminescence assay on a microfluidic simulator for antibiotic therapy. Anal. Chem. 2015, 87, 2410–2418. [Google Scholar] [CrossRef]
  89. Ivančić, V.; Mastali, M.; Percy, N.; Gornbein, J.; Babbitt, J.T.; Li, Y.; Landaw, E.M.; Bruckner, D.A.; Churchill, B.M.; Haake, D.A. Rapid antimicrobial susceptibility determination of uropathogens in clinical urine specimens by use of ATP bioluminescence. J. Clin. Microbiol. 2008, 46, 1213–1219. [Google Scholar] [CrossRef]
  90. Webster, T.A.; Sismaet, H.J.; Goluch, E.D. Electrochemically monitoring the antibiotic susceptibility of Pseudomonas aeruginosa biofilms. Analyst 2015, 140, 7195–7201. [Google Scholar] [CrossRef]
  91. Safavieh, M.; Pandya, H.J.; Venkataraman, M.; Thirumalaraju, P.; Kanakasabapathy, M.K.; Singh, A.; Prabhakar, D.; Chug, M.K.; Shafiee, H. Rapid Real-Time Antimicrobial Susceptibility Testing with Electrical Sensing on Plastic Microchips with Printed Electrodes. ACS Appl. Mater. Interfaces 2017, 9, 12832–12840. [Google Scholar] [CrossRef]
  92. Dailey, P.J.; Osborn, J.; Ashley, E.A.; Baron, E.J.; Dance, D.A.B.; Fusco, D.; Fanello, C.; Manabe, Y.C.; Mokomane, M.; Newton, P.N.; et al. Defining System Requirements for Simplified Blood Culture to Enable Widespread Use in Resource-Limited Settings. Diagnostics 2019, 9, 10. [Google Scholar] [CrossRef]
  93. Rodriguez, F.D.; Simonsson, P.; Alling, C. A method for maintaining constant ethanol concentrations in cell culture media. Alcohol Alcohol. 1992, 27, 309–313. [Google Scholar]
  94. Dawson, A.I. Bacterial Variations Induced by Changes in the Composition of Culture Media. J. Bacteriol. 1919, 4, 133–148. [Google Scholar] [PubMed]
  95. Schreckenberger, P.C.; Binnicker, M.J. Optimizing Antimicrobial Susceptibility Test Reporting. J. Clin. Microbiol. 2011, 49, S15–S19. [Google Scholar] [CrossRef]
  96. Martinez, J.L.; Baquero, F. Mutation Frequencies and Antibiotic Resistance. Antimicrob. Agents Chemother. 2000, 44, 1771–1777. [Google Scholar] [CrossRef]
  97. Khan, Z.A.; Siddiqui, M.F.; Park, S. Progress in antibiotic susceptibility tests: A comparative review with special emphasis on microfluidic methods. Biotechnol. Lett. 2019, 41, 221–230. [Google Scholar] [CrossRef]
  98. van Belkum, A.; Dunne, W.M., Jr. Next-generation antimicrobial susceptibility testing. J. Clin. Microbiol. 2013, 51, 2018–2024. [Google Scholar] [CrossRef]
  99. Kim, S.; Masum, F.; Jeon, J.S. Recent Developments of Chip-based Phenotypic Antibiotic Susceptibility Testing. BioChip J. 2019, 13, 43–52. [Google Scholar] [CrossRef]
  100. Huang, X.; Xu, D.; Chen, J.; Liu, J.; Li, Y.; Song, J.; Ma, X.; Guo, J. Smartphone-based analytical biosensors. Analyst 2018, 143, 5339–5351. [Google Scholar] [CrossRef]
  101. Vergara, A.; Zboromyrska, Y.; Mosqueda, N.; Morosini, M.I.; García-Fernández, S.; Roca, I.; Cantón, R.; Marco, F.; Vila, J. Evaluation of a Loop-Mediated Isothermal Amplification-Based Methodology To Detect Carbapenemase Carriage in Acinetobacter Clinical Isolates. Antimicrob. Agents Chemother. 2014, 58, 7538–7540. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Representation of the final ranking of antibiotic-resistant bacteria, adapted with permission from Tacconelli et al. [11]. CR = carbapenem resistant. 3GCR = third-generation cephalosporin resistant. VR = vancomycin resistant. MR = meticillin resistant. ClaR = clarithromycin resistant. FQR = fluoroquinolone resistant. PNS = penicillin non-susceptible. AmpR = ampicillin resistant.
Figure 1. Representation of the final ranking of antibiotic-resistant bacteria, adapted with permission from Tacconelli et al. [11]. CR = carbapenem resistant. 3GCR = third-generation cephalosporin resistant. VR = vancomycin resistant. MR = meticillin resistant. ClaR = clarithromycin resistant. FQR = fluoroquinolone resistant. PNS = penicillin non-susceptible. AmpR = ampicillin resistant.
Diagnostics 09 00049 g001
Figure 2. Representation of various conventional antibiotic susceptibility testing methods. (a) Disk diffusion, demonstrating of inhibition zones, adapted from Sageerabanoo [29]. (b) Etest gradient disk diffusion, adapted from Sader [30], under terms of the Creative Commons attribution license. (c,d) Broth macro and micro dilution, showing bacterial susceptibility based on optical density and (e) Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry (MADI-TOF MS), adapted from the MALDI Biotyper system (Bruker, Billerica, Massachusetts, United States), Laboratory Information System (LIS) and Laboratory Information Management System (LIMS).
Figure 2. Representation of various conventional antibiotic susceptibility testing methods. (a) Disk diffusion, demonstrating of inhibition zones, adapted from Sageerabanoo [29]. (b) Etest gradient disk diffusion, adapted from Sader [30], under terms of the Creative Commons attribution license. (c,d) Broth macro and micro dilution, showing bacterial susceptibility based on optical density and (e) Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry (MADI-TOF MS), adapted from the MALDI Biotyper system (Bruker, Billerica, Massachusetts, United States), Laboratory Information System (LIS) and Laboratory Information Management System (LIMS).
Diagnostics 09 00049 g002
Figure 3. Digital PCR-High Resolution Melt analysis (HRM)-based bacterial identification from mixed bacterial samples, reproduced with permission from [71], published by American Chemical Society, 2017. (SVM: Support-vector machine)
Figure 3. Digital PCR-High Resolution Melt analysis (HRM)-based bacterial identification from mixed bacterial samples, reproduced with permission from [71], published by American Chemical Society, 2017. (SVM: Support-vector machine)
Diagnostics 09 00049 g003
Figure 4. Demonstration of sensors for antibacterial susceptibility testing involving (a) an optical microfluidics biosensor, showing optical detection of microbial cultures, reproduced with permission from [83], published by RCS Advances, 2015; and (b) an electrochemical biosensor, detection was based on hybridization of the target bacterial 16S rRNA with a detector probe, adapted with permission from Liu [86] (under terms of the Creative Commons attribution license).
Figure 4. Demonstration of sensors for antibacterial susceptibility testing involving (a) an optical microfluidics biosensor, showing optical detection of microbial cultures, reproduced with permission from [83], published by RCS Advances, 2015; and (b) an electrochemical biosensor, detection was based on hybridization of the target bacterial 16S rRNA with a detector probe, adapted with permission from Liu [86] (under terms of the Creative Commons attribution license).
Diagnostics 09 00049 g004

Share and Cite

MDPI and ACS Style

Khan, Z.A.; Siddiqui, M.F.; Park, S. Current and Emerging Methods of Antibiotic Susceptibility Testing. Diagnostics 2019, 9, 49. https://doi.org/10.3390/diagnostics9020049

AMA Style

Khan ZA, Siddiqui MF, Park S. Current and Emerging Methods of Antibiotic Susceptibility Testing. Diagnostics. 2019; 9(2):49. https://doi.org/10.3390/diagnostics9020049

Chicago/Turabian Style

Khan, Zeeshan A., Mohd F. Siddiqui, and Seungkyung Park. 2019. "Current and Emerging Methods of Antibiotic Susceptibility Testing" Diagnostics 9, no. 2: 49. https://doi.org/10.3390/diagnostics9020049

APA Style

Khan, Z. A., Siddiqui, M. F., & Park, S. (2019). Current and Emerging Methods of Antibiotic Susceptibility Testing. Diagnostics, 9(2), 49. https://doi.org/10.3390/diagnostics9020049

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.