3.1. Best Sensor Configuration
In this part, the objective is to determine the optimal SPR sensor configuration by comparing different multilayer designs without the presence of the analyte (SARS-CoV-2). The configurations are evaluated based on key metrics: percentage attenuation, full FWHM, and percentage sensitivity enhancement. These metrics, displayed in
Figure 2 and
Table S3, allow us to calculate the intrinsic performance of each configuration and identify the most effective design before introducing the target virus.
Figure 2a shows the SPR reflectance curves for each configuration (Sys
0 to Sys
9) as a function of the angle of incidence, ranging from 60° to 80°. The position and shape of each SPR dip provide insight into the resonance conditions and sensitivity of each configuration. In the basic configuration, Sys
0, a sharp and narrow SPR dip is observed at approximately 67°, reflecting the interaction between the prism–silver interface and the surrounding water medium. This sharp dip in Sys
0 suggests moderate sensitivity, as the simple silver layer structure provides a minimal enhancement to the evanescent field interaction with the medium. In contrast, the addition of silicon nitride and graphene layers shifts the SPR dip to slightly higher angles and broadens the dip width, as observed in Sys
6, Sys
8, and Sys
9, where SPR dips occur around 70° to 72°. The broader SPR dips in these configurations indicate a greater interaction between the multilayer structure and the evanescent field, enhancing the sensor’s sensitivity [
27].
Attenuation, defined as the percent decrease in reflectance, indicates the strength of the SPR signal. A higher attenuation percentage corresponds to a stronger interaction between the evanescent field and the medium. As shown in
Figure 2b and detailed in
Table S3, the basic system Sys
0 (prism/silver/water) exhibits minimal attenuation (0.023%). For consistency, PBS was used as the medium for configurations Sys
1 through Sys
9, as PBS mimics physiological conditions and provides a more realistic refractive index for future experiments involving SARS-CoV-2 [
19]. Despite the absence of an analyte, the introduction of silicon nitride and graphene layers significantly increases attenuation. For instance, Sys
6 (prism/silver/silicon nitride/graphene/PBS) achieves an attenuation of 7.81%, while Sys
8 and Sys
9 reach 7.93% and 6.59%, respectively. These results confirm that the multilayer design enhances signal strength through better interaction with the evanescent field, even in the absence of the target analyte.
FWHM, shown in
Figure 2c and
Table S3, is a crucial metric in SPR biosensors, as it indicates the sharpness of the SPR dip; lower FWHM values suggest a more precise resonance, which is typically desirable [
28]. The basic configuration Sys
0 exhibits a relatively narrow FWHM (0.88°), consistent with a simple prism–silver design in water. For the configurations using PBS, the addition of Si
3N
4 and graphene layers slightly broadens the FWHM due to enhanced interaction with the evanescent field. Sys
6 shows an FWHM of 1.82°, while Sys
8 and Sys
9 exhibit the broadest values, 1.91° and 1.85°, respectively. Although increased FWHM indicates slightly reduced dip sharpness, it also suggests stronger interactions within the layered structure, which may enhance sensitivity upon analyte introduction.
Sensitivity enhancement (
Figure 2d), measured as the percentage increase in SPR response relative to Sys
0 in water, provides a baseline measure for evaluating the impact of each layer. Sys
0 was selected as the baseline because it represents the simplest configuration in water, devoid of additional layers and reflective of the least complex SPR setup. The configurations using PBS, specifically Sys
2 through Sys
9, exhibit varying sensitivity enhancements depending on the number and type of layers included. Among these, Sys
3, Sys
8, and Sys
9 show the highest sensitivity enhancements, with values of 5.19%, 5.66%, and 5.69%, respectively (
Table S3). While Sys
9 demonstrates the highest sensitivity enhancement, practical considerations, such as the complexity of manufacturing this configuration, led to the selection of Sys
8 as the best configuration. To remark, the observed drop in sensitivity for Sys
4 and Sys
5 likely results from the absence of a silicon nitride layer, which plays a critical role in stabilizing the graphene and enhancing the refractive index contrast needed for optimal SPR performance. When silicon nitride is introduced, as in Sys
6 through Sys
9, the system achieves more effective resonance conditions, resulting in higher sensitivity.
Then, based on percentage attenuation, FWHM, and percentage sensitivity enhancement metrics, Sys8 was selected as the optimal configuration for future analyte testing. This configuration, comprising prism/silver/silicon nitride/graphene/ssDNA layers with PBS as the medium, offers high sensitivity and stability while avoiding the manufacturing complexities associated with Sys9:
Graphene directly on silver and topped with Si3N4 is unconventional, posing adhesion issues. This unusual stacking order requires precise control to avoid layer instability.
Binding ssDNA to graphene is less straightforward than to silicon nitride, often needing intermediary molecules or additional treatments, which increases fabrication steps and variability.
Differences in thermal expansion among silver, graphene, and Si3N4 can lead to strain or delamination, especially under handling or environmental changes.
The precise deposition and functionalization required for each layer make it challenging to scale Sys9 for commercial production, as it requires stringent quality control to maintain sensor performance.
These factors suggest Sys8 is more practical for scalable, reproducible manufacturing while still offering high sensitivity.
3.2. Optimization of Silver Thickness
Figure 3a shows the SPR reflectance curves for varying silver thicknesses from 40 to 65 nm with PBS as the medium. The baseline configuration, Ag
base (Sys
8 with 55 nm of silver in water), is included for comparison. As the silver thickness increases, the SPR dip shifts slightly and narrows, reflecting improved coupling with the evanescent field. However, there is a trade-off between narrowing the SPR dip and managing the attenuation and sensitivity, as excessive thickness can hinder light penetration into the metal layer.
Figure 3b and
Table S4 present the attenuation percentages for each silver thickness. A noticeable increase in attenuation occurs at higher silver thicknesses, particularly at 65 nm, where attenuation reaches 38.87%. The lowest attenuation is observed at 50 nm (0.43%), significantly reducing reflectance loss compared to thicker silver layers. This reduced attenuation at 50 nm suggests a more efficient SPR response, where the sensor maintains signal strength without excessive energy loss.
The FWHM values, shown in
Figure 3c and
Table S4, indicate the sharpness of the SPR dip, with lower values reflecting sharper resonance and potentially higher signal-to-noise ratio. FWHM decreases as the silver thickness increases, with the lowest FWHM observed at 65 nm (1.64°). However, the trade-off at 65 nm is higher attenuation, which can impact sensitivity. At 50 nm, the FWHM is 2.24°, balancing dip sharpness with manageable attenuation. This suggests that 50 nm provides an acceptable compromise, maintaining a defined resonance peak without the excessive broadening seen at lower thicknesses.
Figure 3d and
Table S4 illustrate the sensitivity enhancement for each silver thickness relative to Ag
base.
Figure S1 shows that sensitivity enhancement increases quasi-linearly with silver thickness, yet the whole improvement remains minimal, reaching a maximum of only 0.82% at 65 nm. At 50 nm, the sensitivity enhancement is 0.74%, offering a slight improvement over the baseline without the high attenuation and reduced FWHM associated with thicker layers. This minor increase in sensitivity suggests that thicker silver layers provide diminishing returns and that maximizing thickness beyond 50 nm is not cost-effective for practical applications.
Based on the analysis of these metrics, a silver thickness of 50 nm is selected as the optimal configuration. This thickness provides the lowest attenuation (0.43%), a balanced FWHM (2.24°), and a manageable sensitivity enhancement (0.74%). This combination ensures that the sensor maintains a robust SPR response with minimized energy loss, sufficient dip sharpness, and acceptable sensitivity.
3.3. Optimization of Silicon Nitride Thickness
To further optimize the sensor’s performance, the thickness of the silicon nitride (Si
3N
4) layer was varied from 5 to 20 nm, and its effects on SPR metrics were analyzed.
Figure 4 and
Table S5 display the results obtained.
Figure 4a shows the SPR reflectance curves as a function of the angle of incidence, with S
3N
4_base as the reference configuration in water with 5 nm thickness. As the Si
3N
4 thickness increases, the SPR dip shifts toward higher angles and broadens. This shift and broadening indicate enhanced interaction with the evanescent field, which can increase sensitivity. However, excessive thickness (such as 20 nm) leads to a pronounced broadening, which may negatively affect the sensor’s precision and lead to high attenuation.
Figure 4b and
Table S5 present the attenuation percentages for each Si
3N
4 thickness. A noticeable increase in attenuation is observed as the thickness increases, especially at 20 nm, where attenuation reaches 90.64%. This high attenuation indicates substantial energy loss, which can reduce sensor effectiveness dramatically. Conversely, at 15 nm, attenuation remains relatively low (8.30%), offering a reasonable balance between signal strength and energy retention. This manageable attenuation value, at 15 nm, suggests an optimal balance, enabling effective SPR response without significant signal loss.
Figure 4c and
Table S5 display the FWHM values for each thickness. The FWHM increases with the Si
3N
4 thickness, reaching a maximum of 14.16° at 20 nm, indicating a broad and less distinct SPR dip. For practical biosensing, a sharp resonance (lower FWHM) is preferable, as it correlates with higher signal-to-noise ratios. At 15 nm, the FWHM is 4.99°, which is broader than lower thicknesses but still within an acceptable range (<5°) for effective sensing.
Figure 4d and
Table S5 illustrate the sensitivity enhancement percentage for each Si
3N
4 thickness, while
Figure S2 shows a linear trend in sensitivity as Si
3N
4 thickness increases. Sensitivity enhancement improves with greater thickness, peaking at 19.82% for 20 nm. However, this increase in sensitivity comes at the cost of significant attenuation and an excessively broad FWHM at 20 nm, making it less practical. At 15 nm, sensitivity enhancement reaches 15.09%, providing a substantial improvement over lower thicknesses while avoiding excessive attenuation and broad FWHM seen at 20 nm.
Based on these metrics, a Si3N4 thickness of 15 nm is selected as the optimal configuration. This thickness provides a balanced performance with manageable attenuation (8.30%), an FWHM of 4.99° (ensuring adequate resonance sharpness), and a high sensitivity enhancement of 15.09%. The choice of 15 nm achieves a practical compromise between maximizing sensitivity and maintaining a sharp, low-loss SPR response.
3.4. Optimization of the Number of Graphene Layers
To determine the optimal configuration, we analyze the sensor’s performance with one to six graphene layers.
Figure 5 and
Table S6 summarize these results, while
Figure S3 shows the sensitivity enhancement trend.
Figure 5a presents the SPR reflectance curves as the number of graphene layers increases. As more graphene layers are added, the SPR dip shifts to higher angles and broadens, indicating greater interaction with the evanescent field. However, excessive broadening with multiple layers reduces the dip sharpness, which critically impacts sensor precision. The single-layer graphene configuration (L1) maintains a distinct SPR dip at approximately 82°, balancing sensitivity and dip sharpness.
Figure 5b and
Table S6 display the attenuation percentages for each configuration. Attenuation rises sharply as the graphene layers increase, with a notable jump to 66.15% at six layers. Single-layer graphene (L1) has the lowest attenuation at 8.30%, indicating minimal reflectance loss and efficient SPR response. This low attenuation with single-layer graphene suggests that it supports a strong SPR signal while retaining energy, crucial for maximizing signal-to-noise ratios.
The FWHM values, shown in
Figure 5c and
Table S6, increase as the graphene layer count rises. At one layer, the FWHM is 4.99°, which is within an acceptable range for achieving a sharp resonance. With additional layers, FWHM increases significantly, reaching 9.58° at five layers and 13.8° at six layers, indicating a substantial loss in resonance sharpness. Maintaining a sharp resonance is critical for high-precision detection, and single-layer graphene offers an optimal balance between sensitivity and FWHM.
Sensitivity enhancement, displayed in
Figure 5d and
Table S6, increases linearly up to four layers, with a maximum of 3.27% at five layers. However, beyond four layers, sensitivity enhancement plateaus and even begins to decrease slightly. Single-layer graphene (L1) provides a sensitivity enhancement of 1.28%, which is lower than configurations with multiple layers but is balanced by the advantages of lower attenuation and narrower FWHM.
Figure S3 highlights this quasi-linear trend, indicating that while additional graphene layers initially improve sensitivity, they also introduce diminishing returns and adverse effects on attenuation and FWHM at higher counts.
To stress again, single-layer graphene (L1) is selected as the optimal configuration. This choice is justified by the following considerations:
The low attenuation of single-layer graphene minimizes signal loss, ensuring a strong SPR response.
With a manageable FWHM, single-layer graphene maintains a sharp resonance dip, balancing precision and sensitivity.
Although sensitivity enhancement is higher with more layers, as reported in [
25], the improvement with single-layer graphene is acceptable, especially given the diminishing returns and high attenuation observed with additional layers.
Then, single-layer graphene provides the best balance across these metrics, supporting efficient SPR interaction without the excessive losses and broadening seen in multilayer graphene configurations.
3.5. Optimization of Thiol-Tethered ssDNA Thickness
Now, we examine the impact of ssDNA layer thicknesses from 3.2 to 50 nm.
Figure 6 and
Table S7 summarize the results, while
Figure S4 highlights the sensitivity enhancement trend. Particularly,
Figure 6a shows the SPR reflectance curves as the ssDNA thickness increases. With thicker ssDNA layers, the SPR dip shifts toward higher angles and broadens. The baseline configuration, ssDNA
3.2nm_base (ssDNA in water), maintains a sharp dip, but with increasing thickness, excessive broadening is observed, especially at 30 nm and 50 nm. This broadening negatively affects the sensor’s precision, as it reduces the sharpness of the resonance dip.
Figure 6b and
Table S7 evidence the effect of increasing ssDNA thickness on attenuation. Attenuation rises sharply with layer thickness, reaching 96.93% at 50 nm. At 5 nm, attenuation is relatively low (10.34%), providing a strong SPR response with manageable energy loss. This low attenuation at 5 nm suggests that it balances sensitivity with signal clarity, making it preferable over thicker layers that result in substantial reflectance loss.
The FWHM values in
Figure 6c and
Table S7 show an increasing trend with ssDNA thickness. Starting from 4.99° at 3.2 nm, FWHM rises to 5.20° at 5 nm, remaining within an acceptable range for maintaining a sharp resonance. Beyond 5 nm, FWHM increases considerably, reaching 9.29° at 20 nm and an excessive 53.86° at 50 nm. The 5 nm thickness offers a balance, allowing for a well-defined SPR dip essential for accurate and reliable biosensing.
Figure 6d and
Table S7 demonstrate the sensitivity enhancement for each ssDNA thickness, while
Figure S4 shows a linear trend in sensitivity up to 20 nm. Sensitivity peaks at 6.84% for a 20 nm thickness. However, this comes at the cost of high attenuation (80.48%) and broad FWHM (9.29°). For practical purposes, the 5 nm thickness provides a reasonable sensitivity enhancement (2.27%) without the detrimental increase in attenuation and FWHM associated with thicker ssDNA layers. After 20 nm, sensitivity enhancement declines sharply, making thicker layers less effective. Therefore, the 5 nm ssDNA layer offers an optimal trade-off, preserving sensitivity while maintaining low attenuation and a sharp resonance.
3.6. Application of the Optimized SPR Biosensor for SARS-CoV-2 Sensing
Table S8 presents the optimized parameters for Sys
8. The structure comprises a BK7 prism with a refractive index of 1.5151, forming the base substrate for efficient SPR excitation [
29]. Over this, a 50 nm silver layer (refractive index 0.056253 + 4.2760i) is used to generate the SPR effect, achieving a balance between signal strength and resonance sharpness. A 15 nm layer of silicon nitride, with a refractive index of 2.0394, serves as a dielectric spacer, enhancing the sensor’s sensitivity by optimizing the evanescent field interaction. Single-layer graphene (0.34 nm, refractive index 2.7611 + 1.6987i) is then added to amplify the SPR signal without significant attenuation, while a 5 nm thiol-tethered ssDNA layer (refractive index 1.462) provides selectivity for SARS-CoV-2 detection through specific binding interactions. We point out that the sensor is calibrated with PBS (refractive index 1.334) as the sensing medium, which closely mimics physiological conditions. To detect SARS-CoV-2, the refractive index of the virus in PBS is presented for varying concentrations, ranging from 1.340 at 150 mM to 1.355 at 525 mM. These values, based on a study by Kumar et al. [
19], demonstrate a linear relationship between refractive index and viral concentration, enabling quantitative analysis of SARS-CoV-2 levels.
Figure 7a shows the SPR reflectance curves for various SARS-CoV-2 concentrations (150 to 525 mM) in PBS, with the black curve representing the baseline optimized system in PBS without virus adsorption (refractive index of 1.334). As the virus concentration increases, corresponding to refractive indices of 1.340, 1.345, 1.350, and 1.355, the SPR dip shifts toward higher angles, demonstrating the sensor’s response to changes in the refractive index due to virus presence [
30]. This shift indicates successful detection, as increased virus concentration enhances the binding events on the sensor surface.
Figure 7b and
Table S9 demonstrate the increasing trend in attenuation as the virus concentration rises. Starting from 18.11% at 150 mM, attenuation escalates to 78.47% at 525 mM. This high attenuation suggests that higher concentrations of SARS-CoV-2 significantly dampen the reflected light, indicating a strong interaction between the evanescent field and the viral particles. However, this also highlights a drawback: excessive attenuation at higher concentrations can lead to signal saturation, reducing the sensor’s efficiency in detecting further increases in virus concentration. This saturation effect is visible as a plateau in sensitivity at higher concentrations.
The FWHM values in
Figure 7c and
Table S9 indicate broadening of the SPR dip as virus concentration increases. Starting from 5.64° at 150 mM, the FWHM rises to 8.62° at 525 mM. This broadening suggests a decrease in dip sharpness, which may affect detection precision. While some increase in FWHM is expected with higher analyte concentrations due to enhanced binding, excessive broadening could reduce the sensor’s ability to resolve small changes in virus concentration, as the signal becomes less distinct.
Figure 7d and
Table S9 show sensitivity enhancement percentages for each concentration. Sensitivity initially rises from 2.25% at 150 mM to 5.11% at 400 mM, following a linear trend up to this concentration, as observed in
Figure S5. Beyond 400 mM, sensitivity enhancement plateaus, reaching 4.74% at 525 mM. This plateau suggests that the sensor’s responsiveness diminishes at higher concentrations, potentially due to near-complete saturation of the ssDNA functional layer. This behavior implies that while the sensor remains sensitive at low to moderate concentrations, its effectiveness decreases as it approaches higher viral loads.
Furthermore, the linear sensitivity trend up to 400 mM indicates that Sys8 operates effectively within this range, providing reliable measurements and clear differentiation between concentrations. However, the observed plateau and high attenuation beyond 400 mM suggest that the sensor’s operational range may be limited, as increased virus concentrations lead to diminishing returns in sensitivity, broad FWHM, and excessive attenuation. Therefore, the optimized Sys8 configuration is best suited for low to moderate SARS-CoV-2 concentrations, where it achieves a balance between sensitivity, sharpness, and manageable attenuation.
To further emphasize the potential of the optimized Sys
8 configuration, we report different performance metrics in
Figure 8 and
Table S10, including variation angle (Δθ), sensitivity (S), detection accuracy (DA), and figure of merit (FoM). Actually, the SPR angle shift is a critical indicator of responsiveness to refractive index changes upon SARS-CoV-2 adsorption [
25]. As shown in
Figure 8a and
Table S10, Δθ increases progressively from 1.85° at 150 mM to 4.2° at 400 mM, reflecting the biosensor’s strong sensitivity to changes in viral concentration. This significant angle shift, particularly at 400 mM, implies a high level of viral particle interaction with the functionalized ssDNA layer. However, the reduction in Δθ to 3.9° at 525 mM suggests the onset of surface saturation, where the sensor’s available binding sites reach capacity, limiting additional angle shifts. This plateau effect is typical in biosensors and highlights Sys
8’s optimal concentration range for quantitative analysis, ideally up to 400 mM.
Sensitivity, displayed in
Figure 8b and
Table S10 as the angular shift per refractive index unit (°/RIU), is highest at 315.91°/RIU at 275 mM, demonstrating Sys
8’s high responsiveness to moderate concentrations. This sensitivity level is advantageous in diagnostic settings where precise measurement of viral load is necessary, especially in early-stage infections or low-viral-load scenarios [
31,
32]. Sensitivity decreases to 185.71°/RIU at 525 mM, reflecting the sensor’s diminishing capacity to detect finer concentration changes at high viral loads. This decrease suggests that Sys
8 is best suited for detecting SARS-CoV-2 in the 150–275 mM range, where it achieves sharp and clear angular shifts. This characteristic makes Sys
8 valuable for early diagnostics, allowing healthcare providers to detect lower virus concentrations with high accuracy.
Detection accuracy, which represents the sensor’s ability to differentiate between varying viral concentrations, peaks at 0.60 at 400 mM, as shown in
Figure 8c and
Table S10. The increased DA from 0.33 at 150 mM to 0.60 at 400 mM suggests that Sys
8 is highly effective in quantifying SARS-CoV-2 at these concentrations. However, DA declines to 0.45 at 525 mM, representing the trends in Δθ and S, and indicating potential saturation. Despite this, Sys
8 maintains a high degree of precision up to 400 mM, which is crucial for applications requiring accurate viral load quantification, such as patient monitoring or assessing infection progression. This level of accuracy reinforces Sys
8’s capability as a reliable diagnostic tool, especially within the concentration range where viral load is most clinically relevant.
The figure of merit, which balances sensitivity with signal resolution, is highest at 54.66 RIU
−1 at 150 mM and decreases to 21.55 RIU
−1 at 525 mM, as detailed in
Figure 8d and
Table S10. The high QF at lower concentrations reflects Sys
8’s ability to maintain a sharp, well-defined SPR dip, essential for distinguishing slight variations in virus concentration. As concentration increases, however, the broadening of the SPR dip reduces QF, compromising resolution. The gradual decay in QF from 54.66 RIU
−1 to 37.45 RIU
−1 up to 400 mM still supports Sys
8’s effectiveness for low and moderate concentrations. This makes Sys
8 especially valuable for scenarios where clarity and precision are crucial, such as detecting small shifts indicative of early infection or changes in treatment efficacy.
In this context, the analyzed performance metrics of Sys
8 position it as a versatile biosensor for SARS-CoV-2 detection across a range of clinically relevant concentrations. Its high sensitivity, especially at low to moderate viral loads, makes it a prime candidate for early diagnostic applications. Furthermore, the observed linearity in sensitivity up to 400 mM (as shown in
Figure S5) suggests that Sys
8 can reliably quantify viral load without the need for complex calibration models, which simplifies real-time monitoring in clinical environments.
3.7. Comparison with Previous Reports
To contextualize the performance of Sys
8 configuration, a comparison with other reported SPR-based systems for virus detection [
33,
34,
35,
36] was conducted.
Table 1 presents key metrics for sensitivity, stability, linearity, responsivity, and detection limit, highlighting Sys
8’s competitive standing within the field of viral diagnostics.
Sys
8 demonstrated a high sensitivity of 315.91°/RIU, comparable to the Ag-based multilayer system developed by Uddin et al. [
36] for SARS-CoV-2 detection, which exhibited a sensitivity of 320.50°/RIU. This level of sensitivity is critical for detecting low concentrations of viral particles, emphasizing Sys
8’s utility in applications requiring precise viral load monitoring, especially at early infection stages. Notably, Sys
8 outperformed other configurations, such as the gold-antibody-based SPR biosensor for hepatitis A virus (HAV) by Santos et al. [
33], which achieved a sensitivity of 265.00°/RIU.
In terms of stability, Sys
8 displayed high operational consistency, attributed to the combined use of silicon nitride and graphene layers, which provide enhanced structural and optical robustness. This stability aligns with the high stability reported by Santos et al. [
33] for their gold-antibody-based system but exceeds that of the nano-layered SARS-CoV-2 sensor by Moznuzzaman et al. [
34], which displayed moderate stability. Such stability is crucial for sustained diagnostic applications where sensor durability under repeated measurements is essential.
Regarding linearity, Sys
8 exhibited a linear response across a concentration range of 150–400 mM, which is on par with the Ag-based multilayer sensor by Uddin et al. [
36] and slightly narrower than the nano-layered SPR sensor by Moznuzzaman et al. [
34], which reported a range of 200–500 mM. However, Sys
8’s linearity range covers a clinically relevant concentration spectrum, enabling accurate quantification of SARS-CoV-2 across various viral loads encountered in diagnostic settings.
The responsivity of Sys
8 was measured at 4.2° at 400 mM, slightly higher than the 4.1° achieved by Uddin et al. [
36] for a similar concentration. This level of responsivity highlights Sys
8’s capability to detect incremental changes in viral concentration, an essential feature for real-time monitoring applications. This responsivity aligns well with clinical needs, facilitating rapid response to fluctuations in viral load. Lastly, Sys
8 demonstrated a detection limit of 150 mM, comparable to the Ag-based sensor by Uddin et al. [
36] and lower than the detection limit of 200 mM reported by Moznuzzaman et al. [
34] for their nano-layered configuration. A lower detection limit enables Sys
8 to detect minimal viral concentrations, making it particularly suitable for early diagnostic applications where low viral loads are expected.