1. Introduction
The “respect” approach to mitral valve repair, which involves the implantation of artificial neochordae to constrain and correct excessive leaflet motion, is increasingly favoured over resecting abnormal leaflet tissue [
1]. Emerging evidence indicates that this approach may result in improved haemodynamics and long-term survival [
2]. However, the success of this method hinges on the precise fixation and sizing of the neochordae to ensure optimal tension and leaflet coaptation. Traditionally, surgeons have manually constructed these chords, a process that is potentially prone to error and can be time-consuming during the critical cross-clamp period.
Novel systems such as Chord-X Pre-Measured Loops (On-X Life Technologies, Inc., Austin, TX, USA) offer a simplified and reproducible method for using artificial chordae, eliminating the need for manual construction by the surgeon and facilitating straightforward implantation. To date, only one published series of 17 patients has reported on the use of this technique [
3]. Our study aims to expand the knowledge base of this innovative mitral valve repair technology by documenting our experience and early repair outcomes in our first 40 patients.
2. Materials and Methods
2.1. Surgical Technique
The mitral valve and subvalvular apparatus are exposed using the surgeon’s preferred operative approach. The visual inspection identifies the diseased valve leaflets requiring neochordae support and the most suitable papillary muscle(s) for chordal loop attachment. The ideal chordal loop size is measured from the fibrous portion of the papillary muscle to an adjacent, un-diseased leaflet to achieve optimal leaflet coaptation. Measurements are made using the Chord-X Chordal Sizer or a disposable ruler. Once the appropriate size is selected, the corresponding number of Chord-X Pre-Measured Loops sets are introduced to the sterile field. These can be used immediately, requiring no further surgeon or scrub team preparation.
Each Chord-X Pre-Measured Loops set includes a central pledget with two free needles exiting from one side and three pre-configured chordal loops (with attached needles) emerging from the opposite side (
Figure 1). The threads are composed of expanded polytetrafluoroethylene (ePTFE). The two free needles anchor to the fibrous portion of the papillary muscle and tie onto a second free pledget (
Figure 2A). The chordal loops are subsequently sutured via their associated needles to the diseased mitral valve leaflets and tied (
Figure 2B,C). The number of knots and tension in the knot does not change the neochordae size. Although each set has three loops, attaching all loops to the valve is not mandatory. Unused loops can be discarded without compromising the system’s integrity or strength. Post-cardiopulmonary bypass (CPB) trans-oesophageal echocardiography (TOE) is used to confirm the adequacy of repair (
Figure 2D).
2.2. Study Design
We designed a retrospective observational cohort study to investigate the short-to-medium-term outcomes of using Chord-X Pre-Measured Loops. From May 2020 to February 2024, three surgeons in Geelong, Victoria, Australia, performed 40 consecutive mitral valve repair operations using Chord-X Pre-Measured Loops. Patients were selected for this operative technique based on pre-operative echocardiography, indicating that their mitral valve pathology was amenable to repair. Patients who underwent mitral valve repair during the study period without neochordae (e.g., annuloplasty only) were not included in this study.
All 40 patients with neochordae were included in this retrospective study and analysed. Data were collected retrospectively from the patient’s medical records, including operative notes, discharge paperwork, and post-operative echocardiography reports. The duration and frequency of follow-up echocardiography were determined by the patient’s treating cardiologist, and only data from the latest follow-up were evaluated. All data extraction, cleaning, and statistical analysis were performed using R version 4.3.0 (R Core Team, Vienna, Austria).
Barwon Health Human Research Ethics Committee (QA/104853/VICBH-2023-405717(v1), 13 December 2023) approved this study and granted a waiver of written consent due to the negligible risk of this research.
3. Results
All 40 patients were referred for surgery due to severe mitral regurgitation identified via transthoracic (TTE) echocardiography, with most patients undergoing a follow-up TOE to further clarify the mechanism of regurgitation. The underlying pathology was myxomatous valve degeneration in all cases, with no instances of rheumatic or ischemic mitral disease in this dataset (
Table 1). All but three procedures were performed electively. Approximately one-third of patients were asymptomatic despite severe mitral valve disease, but these individuals exhibited evidence of early left ventricular dilatation and thus met the indications for surgical restoration (
Table 2). Most patients had preserved left ventricular function, and the mean pulmonary artery pressure was within the normal range. Other than the single emergency case, these patients represented a low-risk surgical cohort (EuroSCORE II < 3%). Patient demographics are further detailed in
Table 1.
The choice of operative approach was primarily influenced by favourable thoracic anatomy and the absence of combined procedures, with six patients undergoing a minimally invasive approach (via right lateral mini-thoracotomy). Nine patients underwent concomitant procedures (six CABG, one tricuspid valve repair, and two bi-atrial MAZE). Intraoperative inspection of the mitral valve confirmed P2 leaflet pathology in 85% of patients, with P3 involvement (concomitant or standalone) in 55% (
Table 2). Although pre-operative echocardiography (either transthoracic or trans-oeosphageal) determined leaflet prolapse as the underlying mechanism in the majority of cases, many patients were found to have abnormal or ruptured chordae on intra-operative inspection. The anterior leaflet was implicated in only one patient in this series who had prolapse of the entire anterior complex; aside from requiring longer neochordae, no major modifications to the implantation technique were necessary.
Most Chord-X loops were anchored to the posterior papillary muscle, with only 11 patients requiring chords attached to the anterior papillary muscle. In general, the papillary muscle attachment was chosen so that neochordae do not cross the mitral valve ‘midline’ and thus respect native chordal arrangements [
4,
5]. For disease of the posterior mitral leaflets, 16 mm pre-measured loops were the most appropriate size for 85% of patients, with the remainder requiring 12 mm loops (
Table 3). In the single patient with bi-leaflet prolapse, 20 mm loops were necessary in order to support the anterior leaflet, mimicking the increased length of anterior leaflet chordae observed in vivo [
6]. Successful repair necessitated only one prosthesis (i.e., ≤3 loops) in 80% of patients (
Figure 3). Whilst all patients received an annuloplasty in addition to their neochordae, extra procedures to restore leaflet integrity (e.g., repair of leaflet clefts) were only required in five cases.
Intraoperative echocardiography was utilised for every procedure, and no greater than mild MR was confirmed in all patients at the conclusion of the procedure (
Table 4). One of the earlier patients (4th in this series) required a return onto CPB after weaning off, as the Chord-X prosthesis had dislodged from the papillary muscle. The leaflet attachments were all intact; thus, only the papillary muscle attachment needed to be re-anchored. It was reattached to the same papillary muscle head (but slightly deeper) a second time with no further complications, and a durable long-term repair was achieved without the need for re-operation.
Surgical recovery was uneventful for most patients, with a median length of stay of seven days in the hospital. One patient required a return to the theatre for control of bleeding, which resulted in a prolonged stay. Five patients were readmitted within 30 days of operation, either due to arrhythmia or non-cardiac concerns. No readmission was related to heart failure or valve deterioration, and there were no early mortalities.
Postoperative echocardiography data was obtained in all 40 patients at varying time points, and data from the latest follow-up examination is summarized in
Table 5. The median time until the latest follow-up echocardiogram was 33 days, with 15 patients having data available greater than three months post-operatively. The longest post-procedure echocardiography follow-up was 1288 days. At the latest follow-up, 75% of patients had nil or trace MR, and only 25% had mild MR. No patient to date has had greater than mild MR or required reintervention to their mitral valve.
4. Discussion
Since David first popularised the use of ePTFE neochordae for mitral valve repair [
7], various techniques have been suggested for how surgeons should create and implant these neochordae. The most prevalent methods include the use of simple interrupted sutures [
8], a continuous running suture [
1], or the Leipzig loop technique [
9]. No method has been proven to be strictly superior to another, and each method has its own advantages and disadvantages, considering factors such as time taken to construct neochordae, procedural complexity of construction, technical difficulty of achieving correct chordal length, and ease of implantation into the patient. While no in vivo studies compare the efficacy of these techniques, a biomechanical analysis in 2021 demonstrated that each of these configurations can withstand forces far greater than those seen under physiological conditions [
10]. Thus, it is up to the surgeon to choose a technique that they can reliably master in order to produce a long-lasting repair.
The Chord-X Pre-Measured Loops system is an iteration of the Leipzig loop technique, whereby multiple chordal loops are pre-created and arise from a single pledget, which will be anchored to a papillary muscle. The primary advantage of this technique is that the loops can be constructed on a back table prior to the initiation of CPB, therefore limiting cross-clamp time [
5]. Additionally, all the loops are assembled to be of identical size, ensuring that a consistent repair can be achieved over a wide leaflet area. This is in contrast to the interrupted or continuous suture techniques, where extra technical difficulty is involved in ensuring that each implanted neochord is of the correct length.
Nonetheless, there are several drawbacks of the traditional Leipzig technique, which a premeasured loop system aims to resolve. Firstly, the manually constructed Leipzig loops rely heavily on accurate TOE measurements, as the loops are already constructed prior to intra-operative valve inspection. Secondly, only a certain number of loops are created prior to cross-clamping, and if further loops are needed (due to inaccurate sizing, a misplaced stitch, or a failed repair), then these can be time-consuming to create and contribute to extra ischaemic time. Our experience has shown us that the Chord-X system can circumnavigate these potential weaknesses. Once the mitral valve has been inspected and the appropriate loop size is determined, the Chord-X Pre-Measured Loops can be immediately opened and sutured in place without further configurations. This efficiency proved especially useful, for example, in our patient who required another period of myocardial arrest in order to re-repair the valve. Utilising a system that can be easily implanted or adapted in the case of a failed first repair might influence the decision to re-repair compared to yielding to a valve replacement and thus have long-lasting implications for the patient.
To our knowledge, this study represents the largest case series involving a novel pre-measured chordal loop system and the only study to include post-discharge outcomes. Our most notable finding was that all surgeons at our institution were able to achieve satisfactory mitral valve repair in all cases, as demonstrated by excellent intraoperative echocardiography results and no patient requiring conversion to mitral valve replacement. These repairs proved to be durable after the immediate post-operative period, with no patients having greater than mild MR in their follow-up scans. Additionally, these results were achieved by surgeons with a relatively low annual volume of mitral valve repairs. Large database studies have suggested that low-volume surgeons are at higher risk of conversion to valve replacement [
11,
12], and thus, a system that mitigates this risk is worth adopting. Significantly, no additional or specialised training was required for surgeons to utilise the Chord-X Pre-Measured Loops system effectively. Each of the surgeon’s first experiences with Chord-X is included in this study, and these patients exhibited comparable results to the rest of the cohort.
We hope that our early success with this system will encourage surgeons to adopt Chord-X Pre-measured Loops or similar novel systems into their mitral repair practice. In the case of single leaflet disease, adequate repair required only one prosthesis to be utilised in almost all patients (
Figure 2), highlighting the ease of use and cost-effective nature of such a system. Conversely, in cases of complex bi-leaflet prolapse, a system capable of reliably producing loops of various lengths is invaluable and can be adapted to any patient’s anatomy. Given that our study and that of Gillinov et al. [
3] have established short-term safety data of Chord-X Pre-measured Loops, future research should focus on the long-term durability of these systems. Additionally, it might be beneficial to directly compare surgeon experience with pre-measured versus manually constructed loops in terms of ease of repair or bypass times.
5. Conclusions
Overall, our results underscore the safety and reproducibility of pre-measured loop systems, particularly the Chord-X Pre-Measured Loops system. Surgeons with relatively lower mitral case volumes were able to implement this system effectively in a low-risk cohort with primarily posterior leaflet prolapse. We acknowledge the study’s limitations, including the small sample size, retrospective design, and relatively short echocardiographic follow-up period. Nevertheless, we hope these findings will encourage surgeons to incorporate these novel systems into their clinical practice. Further adoption of Chord-X Pre-Measured Loops will facilitate future studies comparing this technology with traditional surgeon-constructed chords, providing long-term data.