Two-Dimensional Laser-Align Device for Ultrasound-Guided Injection
Abstract
:1. Introduction
2. Materials and Methods
2.1. Device Fabrication
2.2. Device Performance
2.3. Statistics
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Acknowledgments
Conflicts of Interest
Appendix A
Subject | Task | Order | Target Depth (cm) | Total Duration (s) | Withdraw Times (#) | Duration of Last Insertion (s) | |
---|---|---|---|---|---|---|---|
Inexperienced Group | 1 | FH | I | 1 | 64 | 4 | 7 |
1DLA | II | 1.5 | 15 | 1 | 15 | ||
2DLA | III | 2 | 10 | 1 | 10 | ||
2 | FH | I | 2 | 61 | 7 | 11 | |
1DLA | II | 1.5 | 58 | 5 | 13 | ||
2DLA | III | 1 | 42 | 2 | 14 | ||
3 | FH | III | 2 | 31 | 3 | 9 | |
1DLA | II | 1 | 8 | 1 | 8 | ||
2DLA | I | 1.5 | 26 | 2 | 11 | ||
4 | FH | III | 1 | 44 | 7 | 2 | |
1DLA | I | 2 | 36 | 3 | 5 | ||
2DLA | II | 1.5 | 23 | 2 | 6 | ||
5 | FH | II | 2 | 17 | 5 | 3 | |
1DLA | I | 1 | 19 | 3 | 4 | ||
2DLA | III | 1.5 | 11 | 1 | 11 | ||
6 | FH | I | 1.5 | 35 | 6 | 3 | |
1DLA | III | 1 | 18 | 3 | 5 | ||
2DLA | II | 2 | 7 | 1 | 7 | ||
7 | FH | II | 1 | 46 | 4 | 6 | |
1DLA | III | 1.5 | 5 | 1 | 5 | ||
2DLA | I | 2 | 10 | 1 | 10 | ||
8 | FH | II | 1.5 | 3 | 1 | 3 | |
1DLA | I | 2 | 11 | 1 | 11 | ||
2DLA | III | 1 | 19 | 1 | 19 | ||
9 | FH | III | 1.5 | 4 | 1 | 4 | |
1DLA | II | 2 | 37 | 3 | 7 | ||
2DLA | I | 1 | 15 | 1 | 15 | ||
10 | FH | II | 1 | 11 | 1 | 11 | |
1DLA | III | 2 | 8 | 1 | 8 | ||
2DLA | I | 1.5 | 40 | 2 | 20 | ||
11 | FH | III | 1.5 | 133 | 13 | 5 | |
1DLA | I | 1 | 64 | 5 | 21 | ||
2DLA | II | 2 | 24 | 1 | 24 | ||
12 | FH | I | 2 | 141 | 17 | 2 | |
1DLA | III | 1.5 | 30 | 4 | 4 | ||
2DLA | II | 1 | 36 | 3 | 7 | ||
13 | FH | I | 2 | 153 | 7 | 13 | |
1DLA | II | 1.5 | 61 | 3 | 12 | ||
2DLA | III | 1 | 18 | 1 | 18 | ||
14 | FH | III | 1 | 44 | 5 | 3 | |
1DLA | I | 1.5 | 41 | 3 | 10 | ||
2DLA | II | 2 | 26 | 2 | 14 | ||
Experienced Group | 1 | FH | III | 1.5 | 11 | 2 | 3 |
1DLA | II | 2 | 80 | 3 | 20 | ||
2DLA | I | 1 | 90 | 2 | 30 | ||
2 | FH | I | 1.5 | 15 | 2 | 5 | |
1DLA | II | 1 | 24 | 2 | 10 | ||
2DLA | III | 2 | 12 | 1 | 12 | ||
3 | FH | II | 2 | 20 | 3 | 5 | |
1DLA | III | 1.5 | 16 | 2 | 7 | ||
2DLA | I | 1 | 45 | 1 | 45 | ||
4 | FH | I | 1 | 3 | 1 | 3 | |
1DLA | II | 1.5 | 20 | 2 | 7 | ||
2DLA | III | 2 | 38 | 2 | 12 | ||
5 | FH | II | 2 | 2 | 1 | 2 | |
1DLA | I | 1 | 10 | 1 | 10 | ||
2DLA | III | 1.5 | 18 | 1 | 18 | ||
6 | FH | III | 1.5 | 5 | 2 | 2 | |
1DLA | II | 1 | 27 | 1 | 27 | ||
2DLA | I | 2 | 24 | 1 | 24 |
References
- Scholten, H.; Pourtaherian, A.; Mihajlovic, N.; Korsten, H.; Bouwman, R.A. Improving needle tip identification during ultrasound-guided procedures in anaesthetic practice. Anaesthesia 2017, 72, 889–904. [Google Scholar] [CrossRef] [PubMed]
- Infiniti Plus™. In-Plane Ultrasound Needle Guides. Available online: https://www.civco.com/catalog/ultrasound-needle-guides/infiniti-plus-ultrasound-needle-guides/ (accessed on 16 July 2019).
- Ungi, T.; Sargent, D.; Moult, E.; Lasso, A.; Pinter, C.; McGraw, R.C.; Fichtinger, G. Perk tutor: An open-source training platform for ultrasound-guided needle insertions. IEEE Trans. Biomed. Eng. 2012, 59, 3475–3481. [Google Scholar] [CrossRef] [PubMed]
- Najafi, M.; Abolmaesumi, P.; Rohling, R. Single-camera closed-form real-time needle tracking for ultrasound-guided needle insertion. Ultrasoun. Med. Biol. 2015, 41, 2663–2676. [Google Scholar] [CrossRef] [PubMed]
- Hoßbach, M.; Noll, M.; Wesarg, S. Simplified stereo-optical ultrasound plane calibration. Int. Soc. Opt. Photonics 2013. [Google Scholar] [CrossRef]
- Hemmerling, T.M.; Taddei, R.; Wehbe, M.; Cyr, S.; Zaouter, C.; Morse, J. Technical communication: First robotic ultrasound-guided nerve blocks in humans using the Magellan system. Anesth. Analg. 2013, 116, 491–494. [Google Scholar] [CrossRef] [PubMed]
- Collins, G.B.; Fanou, E.-M.; Young, J.; Bhogal, P. A comparison of free-hand vs laser-guided long-axis ultrasound techniques in novice users. Br. J. Radiol. 2013, 86, 20130026. [Google Scholar] [CrossRef] [PubMed]
- Tsui, B.C.H. Facilitating needle alignment in-plane to an ultrasound beam using a portable laser unit. Reg. Anesth. Pain Med. 2007, 32, 84–88. [Google Scholar] [CrossRef] [PubMed]
- Liu, S.-Y.; Liao, M.-Y.; Wei, T.-S.; Yang, C.-P.; Chen, L.-W.; Lin, S.-J.; Ding, Y.-S.; Chang, Y.-J. Laser assisted ultrasound guided aspiration improves procedure time and reduces number of withdrawals. Ultrasonics 2008, 48, 647–651. [Google Scholar] [CrossRef] [PubMed]
- Daehee, H.; Lan Seo, Y.; Choi, C.S.; Chul Kim, H.; Yoon, D.Y.; Bae, S.; Moon, J.; Kim, S.H.; Soo Kim, S.; Han, H. A steerable guiding device: The new method in ultrasound guidance. Investig. Radiol. 2002, 37, 626–631. [Google Scholar] [CrossRef]
- Infiniti Plus™. Disposable Needle Guide Replacement Kits. Available online: https://www.civco.com/catalog/ultrasound-needle-guides/infiniti-plus-disposable-needle-guide-replacement-kits/ (accessed on 25 June 2019).
- eTRAX. Needle Tip Tracking Guidance System—12GA. Available online: https://www.civco.com/catalog/fusion-tracking-technology/etrax-needle-tip-tracking-guidance-12ga/ (accessed on 25 June 2019).
- Clear Guide Scenergy. Available online: https://clearguidemedical.com/ (accessed on 25 June 2019).
Group | Subjects | Orders | Target depths | |
---|---|---|---|---|
Total Duration | Inexperienced | 35.83 ± 22.18, p = 0.133 | 35.83 ± 10.90, p = 0.134 | 35.83 ± 3.71, p = 0.779 |
Experienced | 25.56 ± 16.34, p = 0.471 | 25.56 ± 5.99, p = 0.535 | 25.56 ± 8.20, p = 0.385 | |
Withdrawals | Inexperienced | 3.31 ± 1.90, p = 0.281 | 3.31 ± 0.85, p = 0.234 | 3.31 ± 0.36, p = 0.975 |
Experienced | 1.67 ± 0.43, p = 0.206 | 1.67 ± 0.36, p = 0.076 | 1.67 ± 0.24, p = 0.331 | |
Last Duration | Inexperienced | 9.43 ± 3.72, p = 0.120 | 9.43 ± 0.78, p = 0.582 | 9.43 ± 0.53, p = 0.883 |
Experienced | 13.44 ± 4.75, p = 0.925 | 13.44 ± 2.12, p = 0.804 | 13.44 ± 5.69, p = 0.162 |
Group | FH | 1DLA | 2DLA | ||||
---|---|---|---|---|---|---|---|
Total Duration | Inexperienced | 56.21 ± 48.67 | p = 0.012 * | 29.36 ± 19.90 | p = 0.968 | 21.93 ± 10.91 | p = 0.207 |
Experienced | 9.33 ± 6.60 | 29.50 ± 23.24 | 37.83 ± 25.90 | ||||
Withdrawals | Inexperienced | 5.79 ± 4.36 | p = 0.026 * | 2.64 ± 1.39 | p = 0.274 | 1.50 ± 0.63 | p = 0.718 |
Experienced | 1.83 ± 0.69 | 1.83 ± 0.69 | 1.33 ± 0.47 | ||||
Last Duration | Inexperienced | 5.86 ± 3.60 | p = 0.179 | 9.14 ± 4.73 | p = 0.312 | 13.29 ± 5.23 | p = 0.050 * |
Experienced | 3.33 ± 1.25 | 13.50 ± 7.46 | 23.50 ± 11.54 |
Group | FH–1DLA | FH–2DLA | 1DLA–2DLA | |
---|---|---|---|---|
Total Duration | Inexperienced | p = 0.210 | p = 0.039 * | p = 0.541 |
Experienced | p = 0.026 * | p = 0.015 * | p = 0.485 | |
Withdrawals | Inexperienced | p = 0.019 * | p = 0.001 ** | p = 0.039 * |
Experienced | p = 1.000 | p = 0.310 | p = 0.310 | |
Last Duration | Inexperienced | p = 0.039 * | p < 0.001 ** | p = 0.044 * |
Experienced | p = 0.002 ** | p = 0.002 ** | p = 0.093 |
© 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Shiao, E.C.; Kuo, P.-L. Two-Dimensional Laser-Align Device for Ultrasound-Guided Injection. J. Clin. Med. 2019, 8, 1048. https://doi.org/10.3390/jcm8071048
Shiao EC, Kuo P-L. Two-Dimensional Laser-Align Device for Ultrasound-Guided Injection. Journal of Clinical Medicine. 2019; 8(7):1048. https://doi.org/10.3390/jcm8071048
Chicago/Turabian StyleShiao, Eric Chiwei, and Po-Ling Kuo. 2019. "Two-Dimensional Laser-Align Device for Ultrasound-Guided Injection" Journal of Clinical Medicine 8, no. 7: 1048. https://doi.org/10.3390/jcm8071048
APA StyleShiao, E. C., & Kuo, P. -L. (2019). Two-Dimensional Laser-Align Device for Ultrasound-Guided Injection. Journal of Clinical Medicine, 8(7), 1048. https://doi.org/10.3390/jcm8071048