Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards
Abstract
:1. Introduction
2. Materials and Methods
2.1. System Characterization
2.2. Spatial Resolution
2.3. Sensitivity
2.4. Count Rate Measurements and Scatter Fraction
2.5. Imaging Studies
2.5.1. NEMA Micro-PET IQ and Derenzo Phantoms Used for PET Imaging
2.5.2. Micro-PET Phantoms Used for MRI Imaging
2.5.3. Animal Studies
3. Results
3.1. Spatial Resolution
3.2. Sensitivity
3.3. Count Rate Measurements and Scatter Fraction
3.4. Imaging Studies
3.4.1. NEMA Micro-PET Image Quality Phantom Study
3.4.2. NEMA Micro-PET Image Quality Phantom Study
3.4.3. Animal Studies
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Vaquero, J.J.; Kinahan, P. Positron emission tomography: Current challenges and opportunities for technological advances in clinical and preclinical imaging systems. Annu. Rev. Biomed. Eng. 2015, 17, 385–414. [Google Scholar] [CrossRef] [PubMed]
- Sun, X.; Cai, W.; Chen, X. Positron emission tomography imaging using radiolabeled inorganic nanomaterials. Acc. Chem. Res. 2015, 48, 286–294. [Google Scholar] [CrossRef] [PubMed]
- Lu, Y.; Yang, K.; Zhou, K.; Zhang, Q.; Pang, B.; Ren, Q. Development of a SiPM-based PET imaging system for small animals. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 2014, 743, 30–38. [Google Scholar] [CrossRef]
- Xu, J.; Xie, S.; Zhang, X.; Tao, W.; Yang, J.; Zhao, Z.; Weng, F.; Huang, Q.; Yi, F.; Peng, Q. A preclinical PET detector constructed with a monolithic scintillator ring. Phys. Med. Biol. 2019, 64, 155009. [Google Scholar] [CrossRef]
- Ma, T.; Xu, T.; Liu, H.; Wei, Q.; Peng, F.; Deng, Z.; Gong, G.; Gong, H.; Wang, S.; Liu, Y. Development of a SiPM based preclinical PET SPECT imaging system imaging system. J. Nuclear Med. 2017, 58 (Suppl. 1), 397. [Google Scholar]
- Schug, D.; Lerche, C.; Weissler, B.; Gebhardt, P.; Goldschmidt, B.; Wehner, J.; Dueppenbecker, P.M.; Salomon, A.; Hallen, P.; Kiessling, F. Initial PET performance evaluation of a preclinical insert for PET/MRI with digital SiPM technology. Phys. Med. Biol. 2016, 61, 2851. [Google Scholar] [CrossRef]
- Amirrashedi, M.; Sarkar, S.; Ghafarian, P.; Hashemi Shahraki, R.; Geramifar, P.; Zaidi, H.; Ay, M.R. NEMA NU-4 2008 performance evaluation of Xtrim-PET: A prototype SiPM-based preclinical scanner. Med. Phys. 2019, 46, 4816–4825. [Google Scholar] [CrossRef]
- Yao, R.; Lecomte, R.; Crawford, E.S. Small-animal PET: What is it, and why do we need it? J. Nucl. Med. Technol. 2012, 40, 157–165. [Google Scholar] [CrossRef]
- Lancelot, S.; Zimmer, L. Small-animal positron emission tomography as a tool for neuropharmacology. Trends Pharmacol. Sci. 2010, 31, 411–417. [Google Scholar] [CrossRef]
- Yamamoto, S.; Imaizumi, M.; Kanai, Y.; Tatsumi, M.; Aoki, M.; Sugiyama, E.; Kawakami, M.; Shimosegawa, E.; Hatazawa, J. Design and performance from an integrated PET/MRI system for small animals. Ann. Nucl. Med. 2010, 24, 89–98. [Google Scholar] [CrossRef]
- Vandenberghe, S.; Marsden, P.K. PET-MRI: A review of challenges and solutions in the development of integrated multimodality imaging. Phys. Med. Biol. 2015, 60, R115. [Google Scholar] [CrossRef]
- NEMA NU 4-2008; Performance Measurements of Small Animal Positron Emission Tomographs; National Electrical Manufacturers Association: Rosslyn, VA, USA, 2008; pp. 1–23.
- Goertzen, A.L.; Bao, Q.; Bergeron, M.; Blankemeyer, E.; Blinder, S.; Cañadas, M.; Chatziioannou, A.F.; Dinelle, K.; Elhami, E.; Jans, H.-S. NEMA NU 4-2008 comparison of preclinical PET imaging systems. J. Nucl. Med. 2012, 53, 1300–1309. [Google Scholar] [CrossRef]
- Buchert, R.; Bohuslavizki, K.H.; Fricke, H.; Mester, J.; Clausen, M. Performance evaluation of PET scanners: Testing of geometric arc correction by off-centre uniformity measurement. Eur. J. Nucl. Med. 2000, 27, 83–90. [Google Scholar] [CrossRef]
- Schmitz, R.E.; Harrison, R.L.; Stearns, C.W.; Lewellen, T.K.; Kinahan, P.E. Optimization of noise equivalent count rate performance for a partially collimated PET scanner by varying the number of septa. IEEE Trans. Med. Imaging 2007, 26, 935–944. [Google Scholar] [CrossRef]
- Sanchez, F.; Moliner, L.; Correcher, C.; Gonzalez, A.; Orero, A.; Carles, M.; Soriano, A.; Rodriguez-Alvarez, M.J.; Medina, L.; Mora, F. Small animal PET scanner based on monolithic LYSO crystals: Performance evaluation. Med. Phys. 2012, 39, 643–653. [Google Scholar] [CrossRef]
- Pajak, M.Z.; Volgyes, D.; Pimlott, S.L.; Salvador, C.C.; Asensi, A.S.; McKeown, C.; Waldeck, J.; Anderson, K.I. NEMA NU4-2008 performance evaluation of Albira: A two-ring small-animal PET system using continuous LYSO crystals. Open Med. J. 2016, 3, 12–26. [Google Scholar] [CrossRef]
- Grant, A.M.; Deller, T.W.; Khalighi, M.M.; Maramraju, S.H.; Delso, G.; Levin, C.S. NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system. Med. Phys. 2016, 43, 2334–2343. [Google Scholar] [CrossRef]
- Hallen, P.; Schug, D.; Weissler, B.; Gebhardt, P.; Salomon, A.; Kiessling, F.; Schulz, V. PET performance evaluation of the small-animal Hyperion IID PET/MRI insert based on the NEMA NU-4 standard. Biomed. Phys. Eng. Express 2018, 4, 065027. [Google Scholar] [CrossRef]
- Hsu, D.F.; Ilan, E.; Peterson, W.T.; Uribe, J.; Lubberink, M.; Levin, C.S. Studies of a next-generation silicon-photomultiplier–based time-of-flight PET/CT system. J. Nucl. Med. 2017, 58, 1511–1518. [Google Scholar] [CrossRef]
- Stortz, G.; Thiessen, J.D.; Bishop, D.; Khan, M.S.; Kozlowski, P.; Retière, F.; Schellenberg, G.; Shams, E.; Zhang, X.; Thompson, C.J. Performance of a PET insert for high-resolution small-animal PET/MRI at 7 tesla. J. Nucl. Med. 2018, 59, 536–542. [Google Scholar] [CrossRef]
- Visser, E.P.; Disselhorst, J.A.; Brom, M.; Laverman, P.; Gotthardt, M.; Oyen, W.J.; Boerman, O.C. Spatial resolution and sensitivity of the Inveon small-animal PET scanner. J. Nucl. Med. 2009, 50, 139–147. [Google Scholar] [CrossRef] [PubMed]
- Baghaei, H.; Uribe, J.; Li, H.; Wang, Y.; Wong, W.-H. Comparison of the SSRB, MSRB, and FORE methods with the 3DRP algorithm using data from a high resolution PET scanner. In Proceedings of the 2000 IEEE Nuclear Science Symposium, Conference Record (Cat. No. 00CH37149), Lyon, France, 15–20 October 2000; Volume 192, pp. 15/199–115/202. [Google Scholar]
- Watson, C.C. New, faster, image-based scatter correction for 3D PET. IEEE Trans. Nucl. Sci. 2000, 47, 1587–1594. [Google Scholar] [CrossRef]
- Kemp, B.J.; Hruska, C.B.; McFarland, A.R.; Lenox, M.W.; Lowe, V.J. NEMA NU 2-2007 performance measurements of the Siemens Inveon™ preclinical small animal PET system. Phys. Med. Biol. 2009, 54, 2359. [Google Scholar] [CrossRef] [PubMed]
- Gu, Z.; Taschereau, R.; Vu, N.; Wang, H.; Prout, D.; Silverman, R.; Bai, B.; Stout, D.; Phelps, M.; Chatziioannou, A. NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph. Phys. Med. Biol. 2013, 58, 3791. [Google Scholar] [CrossRef]
- García Hernández, T.; Vicedo González, A.; Ferrer Rebolleda, J.; Sánchez Jurado, R.; Roselló Ferrando, J.; Brualla González, L.; Granero Cabañero, D.; del Puig Cozar Santiago, M. Performance evaluation of a high resolution dedicated breast PET scanner. Med. Phys. 2016, 43, 2261–2272. [Google Scholar] [CrossRef]
- Zhang, J.; Maniawski, P.; Knopp, M.V. Performance evaluation of the next generation solid-state digital photon counting PET/CT system. EJNMMI Res. 2018, 8, 97. [Google Scholar] [CrossRef]
- Queiroz, M.A.; Wollenweber, S.D.; von Schulthess, G.; Delso, G.; Veit-Haibach, P. Clinical image quality perception and its relation to NECR measurements in PET. EJNMMI Phys. 2014, 1, 103. [Google Scholar] [CrossRef]
- Bao, Q.; Newport, D.; Chen, M.; Stout, D.B.; Chatziioannou, A.F. Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards. J. Nucl. Med. 2009, 50, 401–408. [Google Scholar] [CrossRef]
- Constantinescu, C.C.; Mukherjee, J. Performance evaluation of an Inveon PET preclinical scanner. Phys. Med. Biol. 2009, 54, 2885. [Google Scholar] [CrossRef]
- Belcari, N.; Camarlinghi, N.; Ferretti, S.; Iozzo, P.; Panetta, D.; Salvadori, P.A.; Sportelli, G.; Del Guerra, A. NEMA NU-4 performance evaluation of the IRIS PET/CT preclinical scanner. IEEE Trans. Radiat. Plasma Med. Sci. 2017, 1, 301–309. [Google Scholar] [CrossRef]
- Prasad, R.; Ratib, O.; Zaidi, H. Performance evaluation of the FLEX triumph X-PET scanner using the national electrical manufacturers association NU-4 standards. J. Nucl. Med. 2010, 51, 1608–1615. [Google Scholar] [CrossRef]
- Cañadas, M.; Sanz, E.R.; Vives, M.O.; Vaquero, J.J.; Desco, M.; Vicente, E.; Udías, J.M.; Romero, L. Performance evaluation for 68 Ga and 18 F of the ARGUS small-animal PET scanner based on the NEMA NU-4 standard. In Proceedings of the IEEE Nuclear Science Symposuim & Medical Imaging Conference, Knoxville, TN, USA, 30 October–6 November 2010; pp. 3454–3457. [Google Scholar]
- Lage, E.; Vaquero, J.J.; Sisniega, A.; España, S.; Tapias, G.; Abella, M.; Rodriguez-Ruano, A.; Ortuño, J.E.; Udias, A.; Desco, M. Design and performance evaluation of a coplanar multimodality scanner for rodent imaging. Phys. Med. Biol. 2009, 54, 5427. [Google Scholar] [CrossRef]
- Prasad, R.; Ratib, O.; Zaidi, H. NEMA NU-04-based performance characteristics of the LabPET-8™ small animal PET scanner. Phys. Med. Biol. 2011, 56, 6649. [Google Scholar] [CrossRef]
- Liu, Q.; Li, C.; Liu, J.; Krish, K.; Fu, X.; Zhao, J.; Chen, J.C. Performance evaluation of a small-animal PET/CT system based on NEMA NU 4–2008 standards. Med. Phys. 2021, 48, 5272–5282. [Google Scholar] [CrossRef]
- Sarnyai, Z.; Nagy, K.; Patay, G.; Molnár, M.; Rosenqvist, G.; Tóth, M.; Takano, A.; Gulyás, B.; Major, P.; Halldin, C. Performance evaluation of a high-resolution nonhuman primate PET/CT system. J. Nucl. Med. 2019, 60, 1818–1824. [Google Scholar] [CrossRef]
- Spinks, T.; Karia, D.; Leach, M.; Flux, G. Quantitative PET and SPECT performance characteristics of the Albira Trimodal pre-clinical tomograph. Phys. Med. Biol. 2014, 59, 715. [Google Scholar] [CrossRef]
PET Specifications | |
---|---|
Parameter | Specifications |
Photon Multiplier (PM) | SiPM |
Detector crystal | Continuous LYSO |
PM element/crystal | 12 × 12 = 144 |
Number of detector rings | 3 |
Transaxial FOV | 80 mm |
Axial FOV | 150 mm |
Detector ring inner diameter | 105 mm |
Number of individual crystals | 24 |
Number of crystals in a ring | 8 |
Number of modules in PET scanner | 24 |
Crystal size | Truncated pyramid |
48 × 48 mm at entrance window | |
50 × 50 mm at base 10 mm thick | |
Data collection method | List mode (or) raw data |
Reconstruction method | 3D-MLEM |
MRI Specifications | |
Magnetic field strength | 7 Tesla |
Inner diameter | 105 mm |
Outer diameter | 169 mm |
Maximum gradient strength | 900 mT/m |
Maximum linear slew rate | 4200 T/m/s |
Analysis Method | Minimum (kBq/mL) | Maximum (kBq/mL) | Average Concentration (kBq/mL) | %STD |
---|---|---|---|---|
Uniformity | 536.8 | 951.6 | 739.9 | 8.8 |
Rod Diameter (mm) | Recovery Coefficient | %STD |
---|---|---|
1 | 0.17 | 28.4 |
2 | 0.41 | 17.7 |
3 | 0.69 | 14.2 |
4 | 0.85 | 9.3 |
5 | 0.94 | 7.2 |
Analysis Method | Chamber | SOR | %STD |
---|---|---|---|
Spillover ratio (SOR) | Air | 0.14 | 24.7 |
Water | 0.17 | 28.2 |
Analysis Methods | Bruker’s Small Animal 7T PET/MRI | IRIS PET/CT | Inveon | FLEX Triumph X-PET (2D-FBP) | FLEX Triumph X-PET (2D-OSEM) | Argus PET | VrPET/CT | Lab PET-8TM | MetisTM PET/CT | XTRIM-PET | LFER 150 PET/CT | |
---|---|---|---|---|---|---|---|---|---|---|---|---|
Uniformity | %STD | 8.8 | 7.0 | 5.3 | 6.3 | 6.0 | 6.0 | 15.3 | 7.0 | 10.9 | 3.82 | 3.3 |
Recovery Coefficient (RC) | 1 mm | 0.17 | 0.14 | 0.17 | 0.13 | 0.15 | 0.28 | 0.22 | 0.13 | 0.22 | 0.12 | 0.08 |
2 mm | 0.41 | 0.58 | 0.48 | 0.38 | 0.43 | 0.77 | 0.62 | 0.32 | 0.50 | 0.40 | 0.75 | |
3 mm | 0.69 | 0.73 | 0.72 | 0.54 | 0.56 | 0.88 | 0.72 | 0.80 | 0.81 | 0.68 | 1.05 | |
4 mm | 0.85 | 0.82 | 0.54 | 0.67 | 0.68 | 0.98 | 0.75 | 0.83 | 0.88 | 0.86 | 1.07 | |
5 mm | 0.94 | 0.89 | 0.93 | 0.85 | 0.88 | 0.92 | 0.75 | 0.96 | 0.89 | 0.90 | 1.06 | |
Spillover ratio (SOR) | Air | 24.7% | 11% | −0.6% | 9.3% | 6.1% | 13% | 8.5% | 13.7% | 17.7% | 35% | 12.6% |
Water | 28.2% | 16% | 1.7% | 11% | 8.2% | 15% | 9.3% | 24.4% | 19.4% | 25% | 15.9% |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Doss, K.K.M.; Mion, P.E.; Kao, Y.-C.J.; Kuo, T.-T.; Chen, J.-C. Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards. Electronics 2022, 11, 2194. https://doi.org/10.3390/electronics11142194
Doss KKM, Mion PE, Kao Y-CJ, Kuo T-T, Chen J-C. Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards. Electronics. 2022; 11(14):2194. https://doi.org/10.3390/electronics11142194
Chicago/Turabian StyleDoss, Kishore Krishnagiri Manoj, Pei En Mion, Yu-Chieh Jill Kao, Tsung-Ter Kuo, and Jyh-Cheng Chen. 2022. "Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards" Electronics 11, no. 14: 2194. https://doi.org/10.3390/electronics11142194
APA StyleDoss, K. K. M., Mion, P. E., Kao, Y. -C. J., Kuo, T. -T., & Chen, J. -C. (2022). Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards. Electronics, 11(14), 2194. https://doi.org/10.3390/electronics11142194