1. Introduction
Light flicker, being one of the fundamental power quality parameters, is caused by low-frequency voltage amplitude fluctuations. This distortion of the enlightenment caused by voltage fluctuations has an irritating effect on people, which is measured by the method recommended in the IEC (International Electrotechnical Commission) Standard 61000-4-15 [
1]. However, it is known that, in its present form, the IEC flickermeter is suffering from some deficiencies [
2,
3].
One of these deficiencies is that the source and direction of the flicker can’t be determined. Some solutions are provided in the literature [
2,
3]. Another problem with the IEC flickermeter is that there can be divergences in the measurements from the human perception [
4,
5]. Low pass filters (LPF) and Root-Mean-Square (RMS) computations are introduced to be used instead of blocks 4 and 5 of the IEC flickermeter to reduce this problem and improvement has been reported with this method in [
5].
Another shortcoming of the IEC flickermeter is that it is adjusted only for the incandescent lamp [
2]. However, studies show that modern lamps, such as compact fluorescents and LEDs, have different enlightenment response to fluctuating voltages [
6,
7,
8,
9,
10], and the IEC flickermeter operates only for the incandescent lamp response. It has been emphasized that the standardized model for the incandescent lamp cannot measure the light flicker caused by other types of lamps [
6]. Tests on the amount of light flicker measured from different types of lamps show that even though the flicker severity measured from the input voltage exceeds the limits, the actual light flicker caused by the LEDs are much below the limits while compact fluorescent lamps (CFLs) exceed the limits a little, but not significantly [
8]. It is clearly shown that 24 different LED lamps exhibit different light intensity variations caused by the same voltage fluctuation and it is also stated that a new flicker curve is required because LED lamps cause less light intensity variation for the voltage fluctuations below 10 Hz. [
9]. It has been stated that the lamps do not exhibit linear behaviors. Some have a higher voltage fluctuation amplitude than the incandescent lamps, while others have smaller. It was observed that synthetic and real voltage signals with similar fluctuation values did not give the same flicker values. For this reason, there is no guarantee that an energy saver lamp is less sensitive than incandescent lamps at every voltage fluctuation level [
10]. Correcting calculation errors caused by instantaneous changes has been tried with a hybrid intelligent flicker intensity estimation method and it has been shown that the proposed method is better than other tested intelligent methods [
11]. In another study, the analytical model of the IEC flickermeter was proposed, and this model was tested in the field with two different loads causing flicker in a laboratory environment and the method is reported to detect the short-term flicker severity values (
) accurately [
12]. Another algorithm, which uses the fundamental frequency and harmonics to detect the flicker frequency, is presented and stated to give good
estimates on the synthetic data [
13].
So far, all of the work in the literature listed above depends on the Amplitude Modulation (AM) modulated signal model for the voltage signal with flicker. However, this does not actually fit the voltage signal in the field. A large number of interharmonic components added to the fundamental and the harmonic components also cause flicker and the flicker model should be formulated as to this reality. In a study using a new model considering this phenomenon, successful
estimations are obtained [
14].
In another work, it is stated that there is a relation between the 2nd harmonics and the
but the IEC flickermeter cannot measure it since it is filtering out all high-frequency components, which is stated to be a deficiency of the IEC flickermeter [
15]. In a recent study based on RMS measurement for flicker estimation, it is stated that, as the frequency increases, the ripple will decrease and therefore the VPD based flickermeter should be used [
16]. It is said that, while incandescent filament lamps are sensitive to the interharmonics around the fundamental frequency, LEDs and CFLs are also sensitive to high frequency interharmonics around the odd harmonic frequencies to which incandescent lamps are insensitive [
17]. LEDs and CFL lamps, which are more sensitive to VPD, in a VPD-based flickermeter proposed by [
18], better flicker estimations are obtained for these lamps, which are also robust to phase-angle jumps [
18]. In [
19], a modified flickermeter is designed to sense the high frequency interharmonics causing the flicker, which provides more realistic results compared to the classical RMS voltage computation based methods designed for incandescent lamps [
19].
Certain variable frequency drives (VFDs) are known to add interharmonics to the system, which cause voltage flicker. An interharmonic flicker curve is proposed to obtain the flicker-induced interharmonic components depending on the VPD in [
20]. It has been shown that CFLs and LEDs themselves produce visible flicker by forming components around the fundamental harmonic through the rectifiers of the interharmonic components around the 3rd and 5th harmonics [
21]. In [
21,
22,
23], it is shown that high-frequency interharmonics for different lamps cause significant flicker and it is reported that the lamps produced should be tested in the range of 0.1 Hz and 2.5 KHz (if possible up to 9 KHz). Average, RMS, and VPD-based flickermeters are used and results are quite similar for frequency components less than 100 Hz, whereas, for larger frequencies, good results are obtained with only the VPD method [
24]. A flickermeter that is recommended in [
25] detects only the high-frequency interharmonic components. In another study, the VPD-based flickermeter is designed to combine individual effects of all the interharmonic components [
26]. In [
27], an FFT-based flickermeter robust to the fundamental frequency deviation is proposed and it is shown that the total instantaneous flicker sensation can be found as the sum of the individual instantaneous flicker sensations generated by each low-frequency interharmonic component contributing to flicker. In that work, however, high-frequency interharmonic components are not taken into consideration [
27].
It is understood from all of these studies that there is a need for a flickermeter, which is designed for all types of lamps and which senses the flicker caused by the interharmonics in the entire frequency spectrum, even if there are instantaneous changes in the signal. In a previous work of the authors, a new flickermeter and a flicker curve have been proposed to meet the above shortcomings by uncovering the disadvantages of recommended flickermeters in the literature. It is also shown analytically that the light flicker should be determined not by using the AM signaling model as in the IEC flickermeter but the interharmonic signal model. It should also be noted that not only the low-frequency interharmonic components but also the high frequency interharmonic components have almost the same effect as the low frequencies. In the case of the flicker caused by the high-frequency interharmonics, VPD flickermeter detects the flicker while IEC flickermeter doesn’t. Hence, a new VPD flickermeter with a new flicker curve considering also the high-frequency interharmonics is suggested in the previous work [
28]. This new flickermeter is shown to detect the flicker caused by high-frequency interharmonics and to be much more robust to the fundamental frequency deviations, such as in cases of electric arc furnace loads [
26,
29].
In this paper, a simplified VPD flickermeter based on the spectral decomposition method in [
27] is proposed using the relative amplitude in the flicker curve obtained in [
28]. The maximum value of the instantaneous flicker sensitivity,
, values are obtained using 0.2 s (10 cycles of the 50 Hz fundamental frequency) windows overlapping nine cycles. Here, different
calculation methods consider the fact that interharmonic components are not evenly distributed around the fundamental and harmonics while spectral decomposition is used to detect
. After selecting the best one among these
calculation methods, flicker estimations are computed with field data obtained from a substation supplying an electric arc furnace plant.
This paper consists of seven parts.
Section 2 briefly addresses the deficiencies of IEC and VPD flickermeters. In
Section 3, simplification by mathematical inference for the VPD flickermeter is shown. In
Section 4, the proposed simplified VPD flickermeter method is described. The results obtained using both synthetic and field data are shown in
Section 5 and
Section 6. The results are highlighted in the last chapter.
2. Problem Definition
Deficiencies of the IEC and VPD flickermeters can be summarized as follows:
IEC flickermeter cannot sense flicker resulting from high frequency interharmonic components.
IEC flickermeter models the signal with flicker as an AM modulated signal, which results in evenly distributed interharmonics around the fundamental and the harmonics and this does not fit the case of actual transmission system voltage waveforms.
IEC flickermeter usually suffers from fundamental frequency deviations, which is a fact of the electric system due to time-varying load-generation balance.
VPD flickermeter proposed in [
26], which detects high-frequency components and deems flicker as a signal with interharmonic components instead of an AM modulated signal, is incompatible with the IEC flickermeter.
VPD flickermeter in [
26] has a highly faulty response in case of fundamental frequency deviation. Hence, a new flicker curve is needed to cope with this problem.
The aim of this work is to introduce a new spectral decomposition based flicker computation method to take care of all these deficiencies.
In [
27], it is shown that the instantaneous flicker sensation can be computed as the summation of the flicker effects of each individual frequency component in the voltage spectrum. In Equation (1), the effect of the
ith frequency component on the instantaneous flicker sensation, where
represents the ratio of the
ith flicker frequency amplitude to the fundamental frequency component amplitude [
27]:
In Equation (1), (
)
IEC corresponds to the
values resulting in unity instantaneous flicker sensation at the output of Block-4 of the IEC flickermeter.
is the frequency response of the filters in the IEC 61000-4-15 flickermeter at each flicker component
. The resultant
value according to [
27] is obtained as the sum of
values for all interharmonic components existing in the voltage spectrum as given in (2):
However, this computation method assumes that interharmonic frequency components are evenly distributed around the fundamental frequency. In cases of fundamental frequency deviations and highly time-varying loads such as electric arc furnaces,
computation can be varied to obtain closer values to the actual instantaneous flicker sensation values as suggested in (3)–(5):
where
values are the spectral corrected amplitudes of the interharmonics formed around the fundamental frequency causing the flicker, “
N” is the number of flicker components around the fundamental and
are the relative amplitude values for each frequency to obtain unity instantaneous flicker sensation value for the VPD flickermeter [
28].
Table 1 provides a list of
values causing unity
for all interharmonic frequencies starting from 1 Hz to 25 Hz (which correspond to 50 Hz ± interharmonic frequency in the voltage frequency spectrum).
values in
Table 1 are half the values given in the IEC Standard 61000-4-15 since the total effect of the beat frequencies around the fundamental are listed in the standard. Here, in
Table 1, effects of the individual interharmonic frequencies are listed, which is actually one of the advantages of the proposed method in this paper, i.e., different frequencies causing the same beat effect are allowed to have their own effects on the measured flicker.
If the fundamental frequency does not shift and the sampling rate is chosen to be an integer multiple of this frequency, all
values suggested in Equations (3)–(5) are equal to each other if their frequencies are in the same distance to the fundamental frequency. However, in cases of highly time-varying loads such as electric arc furnaces resulting in time-varying voltages, discrete Fourier Transform based methods cannot detect frequency components accurately. Moreover, interharmonic components of actual voltage signals are not evenly distributed around the fundamental frequency, which leads to erroneous calculations for the methods based on the modulated signal model given of the IEC flickermeter [
1]. To overcome this situation, both spectral correction and different
calculation methods have been focused on and different results are obtained.
3. Simplification of VPD Flickermeter
When the flicker is expressed with the interharmonic signal model, more precise flicker detection can be performed for both the new generation lamps using rectifiers such as CFLs, LEDs, and the incandescent lamps. To achieve this, VPD flickermeter, which detects flicker using a down-up sampling method, is used [
26]. Block diagram of the VPD flickermeter is shown in
Figure 1a. Also the IEC 61000-4-15 flickermeter is represented in
Figure 1b for comparison.
In the VPD flickermeter illustrated in
Figure 1a, block 2 and block 3 of the IEC flickermeter recommended in [
1], as in
Figure 1b, are changed [
26]. Instead of these blocks, sampling is achieved with the sampling frequency
and
M is an integer such that
fs is an integer multiple of the fundamental frequency
. Then, a “rectification” is applied that is mathematically taking the “absolute value”. Peak voltage is then obtained by “down sampling” by
or
. When the peak point is determined, “up sampling” is performed by
, keeping the two peaks inside one cycle and replacing all other samples by zeros. As the last step, a low-pass filter (LPF), with cut-off frequencies of 50 and 60 Hz for 50 and 60 Hz systems, is used in this flickermeter, while the IEC 61000-4-15 flickermeter uses these of 35 and 42 Hz, respectively. The rest of the system is the same as the IEC flickermeter [
26]. In addition, a high pass filter (HPF) with cut-off frequencies of 0.05 Hz is used in this flickermeter. LPF and HPF flickermeter constitutes band pass filter (BPF). The rest of the flickermeter is the same as IEC 61000-4-15 flickermeter.
New generation lamps such as LEDs and CFLs are more susceptible to peak amplitude fluctuation and therefore the VPD flickermeter method is preferred when they are used. This is because these lamps, excluding the incandescent lamps, carry the high-frequency interharmonic components to the interharmonic frequencies around the fundamental and the odd harmonic due to their operation principles.
The signal with flicker is modeled as in (6) using the interharmonic signal model:
in (6) symbolises the system amplitude, is the fundamental frequency, is relative the amplitude of the interharmonic signal, is the frequency of this interharmonic, “h” is an odd number, is the odd harmonic, is the flicker frequency of the interharmonic and represents the component of the flicker.
Equation in (6) can be rewritten as in (7):
If
v(
t) is defined as the input signal in
Figure 1a, then the rectified signal in
Figure 1a (output of Rectification block) can be expressed as a square wave function multiplied by the sinusoidal expression [
26]. Square wave
s(
t) is given as in (8) using the Taylor series expansion:
Hence, a rectified form of
or
can be expressed as given in (9):
The final equations in (10) and (11), the first part of (9) with the singular harmonics
, and the second part of it with the interharmonics as
, are, respectively:
Equations (10) and (11) are expressed as discrete time signals
and
in (12) and (13), respectively:
In (12) and (13), “k” is the discrete time index, is the first sampling time; (L = 0, 1, 2, ...), where is independent of the semantic value k and is because of an odd integer notation “n”, and thus value is obtained as the “DC” term.
(
h = 2
p + 1,
p = 0, 1, 2, 3, …) and
(due to peak detection at each quarter period of the fundamental) are expressed in (13) and the variables are given as follows:
expression will then be passed through the Band Pass Filter (BPF) and Low Pass Filter (LPF), so that the high frequency components and the DC term will be filtered and, as a result, final expression will be as given in (17):
If the first
cosine expression in (17) is explicitly rewritten, (18) is obtained.
The
phase value in this equation is neglected and then the cosine term in (18) can be written as in (20) using the trigonometric identity given in (19):
If it is used according to the Equation (18), it will be like (20):
Furthermore, if
is substituted in the expression,
cosine expression in (21) is obtained because
:
When
h =
1,
3,
5,
… and
n =
1,
3,
5,
…. (
h − n)
is even and cosine and sine components with (
h − n)
terms cancel out and denoting them by
and
respectively, (22) is obtained:
It is obvious that
= 0 for all values of even (
h − n), so that the left side of the expression
is equal to zero. The expression
will be as given in (23) depending on the value of (
h − n)
:Due to (23), (22) can be expressed as given in (24):
Similarly, the second cosine expression in (17) can be explicitly written as in (25):
If
is omitted and the expression (25) is derived according to (19), (26) is obtained:
Furthermore, if
is substituted in the expression, this
cosine expression is obtained as given in (27), since
:
When
h =
1,
3,
5,
… and
n =
1,
3,
5,
…., (
h − n) is even and cosine and sine components with (
h − n)
terms cancel out and denoting them by
and
, respectively, as (28):
It is clear that
= 0 for all values of (
h + n), so that the left side of the expression
becomes zero.
is expressed as in (29) according to the values of (
h + n):
Then, the expression in (28) can be written as given in (30). In addition, using the expressions (23) and (24),
can be written as in (31):
Considering the expressions (23) and (29), the changes of the expressions
and
will be examined as in (32) assuming that
, respectively. Note that only the components around the fundamental frequency exist in Equation (31). In this equation,
is the frequency response of the flicker frequency component of the filters in the IEC 61000-4-15 flickermeter:
Therefore, Equation (31) can be expressed as in (33):
It is obvious that this expression is independent of the “n” value and has positive and negative values, respectively, as given in (32).
If the statement in (33) is written again as in (31), then (34) follows:
The end part of the total expression in (34) can be approximated as in (35):
Therefore, when the expression in (35) is substituted in (34), (36) is obtained:
This expression shows the output of the weighting filter following BPF (HPF with 0.05 Hz cut-off frequency + LPF with 50 Hz cut-off frequency) in
Figure 1a. When the square of the expression in (36) is taken, it can be shown as in (37):
The
frequency component constituting the fluctuation is passed through the last LPF with a time constant of
as recommended in the IEC Standard [
1] to obtain the instantaneous flicker sensation
value in (38) [
27]:
in (38) is the frequency response of all the filters in the VPD flickermeter at the frequency
. The
,
,
expressions in (36)–(38) are obtained for a signal with a single interharmonic component causing flicker. As stated in [
27], there is more than one flicker component in the actual case of a power network. (38) can be expressed as the summation of the different frequency components causing the flicker as expressed in (39):
The expression in (39) shows the output of BPF in
Figure 1a. When the square of (39) is taken,
can be written as in (40):
In Equation (40), components with the frequency
, “
”, and “
” constitute the double-frequency components, the frequency-difference components and the frequency-summation components, respectively. As stated in [
27], if these components are negligible, the expression in (40) can be written as in (41):
The
value, which is the value of the instantaneous flicker sensation, is obtained by passing through the last LPF with a time constant of
as in (42) [
1]:
In (42), N is the number of components to be calculated at frequencies to around the fundamental frequency. Note that the beat frequencies around the odd harmonics are reflected to the interharmonics around the fundamental by using the VPD flickermeter, hence they are already taken care of.
Based on the values of
given in
Table 1 and the Equation (1), individual effect of each interharmonic component (
Smax(i)) is computed as in (43), using the VPD flickermeter values in
Table 1:
The overall
value is then given as the sum of each
value as in (44):
Hence, it can be concluded that the proposed flickermeter has lower computational complexity compared to the VPD flickermeter based on (43) and (44).
4. Proposed Simplified VPD Flickermeter Method
The block diagram of the proposed VPD-flickermeter, including the spectral verification scheme used, is given in
Figure 2. In the proposed method, a comb filter is first applied on the input signal
v[
k], to extract the fundamental frequency with the interharmonics in the neighborhood of
Hz and only the interharmonics around the odd harmonics, but not the harmonics themselves. Comb filter is obtained by using LPF, HPF and notch filter groups in parallel, specifically HPF and LPF filters with
ve
cut-off frequencies, and notch filters with stop frequency
at harmonics, respectively. Zero-crossing detection is applied on the filtered signal, and both the fundamental frequency “
” and the starting sample “
”, which corresponds to a zero-crossing, for the DFT and the exact start time
for generating synthetic signal are determined [
27,
29,
30]. (45) is obtained with the help of (34) to find the rectified signal with the determined starting point and the starting time:
where
: starting from
points and
k runs from
0 to
N − 1, and
samples. Window length is selected as 0.2 s. The other variables in (45) are listed below:
(sample points of the signal starting at the exact zero-crossing),
n: harmonic grade,
M: number of harmonics up to fs/2,
: fundamental frequency.
indicates the rectified signal. If the expression is a signal at (6) starting at the point “” and in the case of frequency deviation, this approach will yield erroneous results and the synthetic waveform in is generated to compensate for the error.
In (46), the “
A” value in the expression
is the actual amplitude of the fundamental component for each window and is obtained by using the DFT and it is used to obtain the amplitude of the synthetically generated signal [
30].
is used as the corrected starting point to prevent any spectral leakage due to fundamental frequency deviations:
In order to obtain the interharmonic components around both the fundamental frequency and the other harmonics,
is obtained in the frequency domain by subtracting the DFTs of signals
v1 and
v2 from each other [
30]:
Due to the rectification process in (9), if there is a component around an odd harmonic, it is collected around the fundamental frequency. It should be noted that, unlike the DFT-based flickermeter proposed in [
27], the flicker component here is shifted to the spectrum at 0–25 and 75–100 Hz. For example, if there is an interharmonic component of 60 Hz, it will form a component at |50 − 60| = 10 Hz and |150 − 60| = 90 Hz originating from the multiplication with the 3rd harmonic. The flicker frequency of this component will be at |
n*50 − 60|=10 Hz when
n = 1 is selected, as mentioned earlier.
Next, by taking the components from 5 to 25 Hz and 75 to 95 Hz and replacing the IEC
expression in (1) with the VPD
value in
Table 1, the instantaneous flicker sensation values of each flicker component are obtained for the three methods (
S1,
S2, and
S3) and then
and
values are computed.
It is expected that these values will be equal to each other if the fundamental frequency does not deviate and a good sampling is performed with integer multiples of this frequency. However, since the flicker components occur with highly time-varying characteristics, the DFT components are not identically distributed at two sides of the fundamental frequency component, which leads to erroneous calculations. To overcome this problem, both spectral correction and different calculation methods are considered and the results are compared. It has been shown that the best results are obtained with the method.
6. Proposed VPD Flickermeter Response to Field Data
The field data used in this work has been obtained from the electricity transmission system of Turkey by the power quality monitoring devices developed through the National Power Quality Project of Turkey [
32]. The data has been collected from a transformer substation supplying an electric arc furnace (EAF) plant; therefore, it is rich in highly time-varying interharmonics and hence suffers from significant flicker. It is 10-min data collected from three different phases simultaneously at a sampling frequency of 3.2 KHz.
For all phases of the collected voltage waveform, frequency variation is calculated. In addition, instantaneous flicker sensation values (
) for 10 min are obtained using the proposed method, the IEC flickermeter and the flickermeter proposed in [
27], which does not consider the high-frequency flicker components. All results are plotted together in
Figure 3,
Figure 4 and
Figure 5 for both the whole 10-min measurement period and also for a sample 1-min period to observe the details for comparison purposes.
values are calculated for 0.2-s (10 cycle of the fundamental) windows overlapping 9-cycles, hence generating one
value every cycle and average of each 10-cycle is taken to obtain one
S value at each 0.2-s window. The averages have been calculated by using the
method considering the simplified IEC flickermeter [
27] and the proposed VPD flickermeter.
It is observed in
Figure 3,
Figure 4 and
Figure 5 that the results of the IEC flickermeter and the simplified IEC flickermeter are very close to each other. This shows that the results obtained by the
method using the method of [
27] are consistent with the results obtained from the IEC flickermeter. However, as can be seen from the tests, using synthetic data that the flicker produced by the high frequency components cannot be detected. In addition, to the knowledge of the authors, there is no work in the literature to determine the effect of the high-frequency interharmonics when the fundamental frequency is deviating. It is seen that the
results obtained with the proposed VPD flickermeter have similar tendencies of increase and decrease with the results of the IEC flickermeter and the simplified IEC flickermeter in [
27]. The differences can be interpreted as the effect of the high-frequency components detected by the proposed VPD flickermeter, which can’t be detected with the IEC flickermeter and the simplified IEC flickermeter in [
27]. For this reason, the values of
and
obtained with the VPD flickermeter are usually somewhat higher.
Table 9 shows the
values obtained for the three different flickermeters used. In a similar manner, higher
values are obtained because the proposed VPD flickermeter senses the flicker caused by the high frequency interharmonic components.