Next Article in Journal
Genome-Scale Metabolic Modelling Approach to Understand the Metabolism of the Opportunistic Human Pathogen Staphylococcus epidermidis RP62A
Previous Article in Journal
Glycosphingolipid Levels in Urine Extracellular Vesicles Enhance Prediction of Therapeutic Response in Lupus Nephritis
 
 
Article
Peer-Review Record

Comparative Evaluation of Plasma Metabolomic Data from Multiple Laboratories

Metabolites 2022, 12(2), 135; https://doi.org/10.3390/metabo12020135
by Shin Nishiumi 1,*,†, Yoshihiro Izumi 2,*,†, Akiyoshi Hirayama 3,*,†, Masatomo Takahashi 2, Motonao Nakao 2, Kosuke Hata 2, Daisuke Saigusa 4,5, Eiji Hishinuma 5,6, Naomi Matsukawa 5,6, Suzumi M. Tokuoka 7, Yoshihiro Kita 7,8, Fumie Hamano 7,8, Nobuyuki Okahashi 9, Kazutaka Ikeda 10, Hiroki Nakanishi 11, Kosuke Saito 12, Masami Yokota Hirai 13, Masaru Yoshida 14, Yoshiya Oda 7, Fumio Matsuda 9 and Takeshi Bamba 2add Show full author list remove Hide full author list
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Metabolites 2022, 12(2), 135; https://doi.org/10.3390/metabo12020135
Submission received: 7 January 2022 / Revised: 24 January 2022 / Accepted: 28 January 2022 / Published: 1 February 2022

Round 1

Reviewer 1 Report

This is an excellent paper in my opinion. It hi-lites the issues which could befall large metabolite studies needing multiple machines/ laboratories. As this paper states there can be significant differences in the metabolites, including quantitive measurements. It is a warning. As the authors observe, standardisation can lead to a reduction between differences in laboratories/machines, although its may not be sufficient.

Although the issues have been well described, perhaps there could be  a bit more written, added,  on solutions, to help workers in the field, rather than describing the issues/problems. hence I suggest minor revision.

Its an extremely well written paper, the English is excellent.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 2 Report

Manuscript title: Comparative evaluation of plasma metabolomic data from multiple laboratories.

 

General comment: In this paper the authors investigated the effects of several factors, among which measurement, data analysis and sample preparation on the inter-laboratory differences in metabolomic data with the aim of facilitating the understanding of the problems in metabolomics studies involving multiple laboratories. Before publication on Metabolimics, I suggest minor revisions.

L104-L109: I suggest the authors to better explain the main goal of their work at the end of the Introduction section, since it’s not enough clear.

Section 2.1. The authors title this section “Experimental design”. This can be misleading, as experimental design or DoE is typically referred to experiments aim at predicting the outcome by introducing a change of the preconditions, which is represented by one or more independent variables.

Moreover, in this section they should clearly report the number of laboratories involved in the study and the differences among the methods that they used.

L153: “[S-2]/([S-1]+[S-3])×100” Please, clarify what S stands for.

Figure 1: The figure is informative since it helps the reader to understand the operating procedure of the experiments. However, the authors should at least indicate the total number of laboratories that participated in the study.

Table 3 and Table 4. “The percentages (%) in the parentheses in columns 5 and 6 indicate the ratio of the numbers in column 5 or 6 to column 2.” Only 4 columns are reported in these tables. Please, carefully check.

L320: “Many of the relative ratios (human/mouse) were lower in SFC/MS (Methods E–F) than in LC/MS (Methods A–D)..” I suggest the authors to argue this point, as in general for some classes of compounds such as alkyl polyglyceryl ethers SFC offer a better separation compared to LC, which coupled with MS provided very useful structural information, ensuring also more sensitive results in quantification. To this regard, see and cite (2021). Analysis of surfactants by mass spectrometry: Coming to grips with their diversity. MASS SPECTROMETRY REVIEWS, (2021). p. 1-32, ISSN: 0277-7037, doi: 10.1002/mas.21735.

 

 

Author Response

Please see the attachment.

Author Response File: Author Response.pdf