Article

The New and the Old: Platform Cross-Validation of Immunoaffinity MASS Spectrometry versus ELISA for PromarkerD, a Predictive Test for Diabetic Kidney Disease

Proteomics International, Perth, WA 6009, Australia
*
Author to whom correspondence should be addressed.
Proteomes 2020, 8(4), 31; https://doi.org/10.3390/proteomes8040031
Submission received: 30 September 2020 / Revised: 22 October 2020 / Accepted: 27 October 2020 / Published: 28 October 2020
(This article belongs to the Special Issue Proteomics: Technologies and Their Applications)

Abstract

:
PromarkerD is a proteomics derived test for predicting diabetic kidney disease that measures the concentrations of three plasma protein biomarkers, APOA4, CD5L and IBP3. Antibodies against these proteins were developed and applied to a multiplexed immunoaffinity capture mass spectrometry assay. In parallel, and facilitating current clinical laboratory workflows, a standard ELISA was also developed to measure each protein. The performance characteristics of the two technology platforms were compared using a cohort of 100 samples, with PromarkerD test scores demonstrating a high correlation (R = 0.97). These technologies illustrate the potential for large scale, high throughput clinical applications of proteomics now and into the future.

1. Introduction

The application of proteomics technologies for the measurement of protein concentrations has a long history including the use of optical density measurements, SDS-PAGE, amino acid analysis and a range of mass spectrometry techniques. This simple application of technology has increasing importance as research programmes continue producing novel and often complex panels of biomarkers for disease detection and monitoring. As proteomic technology has advanced so too have the methods by which proteomics can be applied to protein concentration measurement. The requirements to go deeper into the proteome to measure that elusive biomarker have produced an arms race between the expanding technology capabilities and the increased sensitivity and specificity demanded of novel biomarker detection. This manuscript describes methods both new and old that make up a repertoire of techniques that can be used to address the fundamental principles of accurate, reproducible protein measurement for specific diseases.
Chronic kidney disease (CKD) affects one in three adults with diabetes, accounting for 40,000 deaths and $100 billion (USD) in healthcare spending [1]. The current usual-care tests are the urinary albumin:creatinine ratio (ACR) and the blood estimated glomerular filtration rate (eGFR). While these tests can independently determine a person’s kidney function, they can provide conflicting diagnoses and their predictive power to determine a patient’s disease progression is minimal [2].
A proteomics biomarker discovery workflow was undertaken [3] to develop the PromarkerD test that can successfully predict four years in advance that a patient will develop chronic kidney disease (CKD) [4,5]. A panel of three protein biomarker concentrations (Apolipoprotein A4 (APOA4), CD5 antigen-like (CD5L) and Insulin-like growth factor-binding protein 3 (IBP3)) and three clinical variables (age, hdl-cholesterol and eGFR) are combined in a predictive algorithm [5]. The original assay workflow involved firstly immunodepletion of the top 14 plasma proteins, and then diafiltration, reduction, alkylation, digestion and targeted mass spectrometry [3]. This process was suited to a research test that could validate the utility of the biomarker and clinical measurements but ultimately was not sufficient for current clinical application where a more rapid and cost-effective assay was required.
To provide a higher-throughput and more robust test, two complementary but distinct new methods were developed. An immunoaffinity mass spectrometry assay (IAMS) and an enzyme-linked immunosorbent assay (ELISA). Both were derived from the development of antibodies against the three biomarker proteins.
The IAMS method uses a bead-bound antibody mixture to capture all three biomarkers from a plasma sample before washing, reduction, alkylation and digestion on the bead. The resulting peptide solution is then analysed by LCMSMS and with the use of calibration protein standards, a concentration of each biomarker is determined. The advantages of this method are the selectivity and hence cleanliness and sensitivity achievable with an antibody capture step and the subsequent multiplexing capability of multiple reaction monitoring (MRM). Independent of this study, the IAMS method was shown to be reproducible and provide equivalent results to the original immunodepletion MS method [6].
The ELISA technology, in use for close to 50 years [7,8], utilises pairs of antibodies developed against recombinant forms of the biomarker proteins. Each plasma biomarker concentration is derived from a standard curve of the recombinant protein with each biomarker measured in a separate well from a distinct dilution of the original plasma sample. The advantages of an ELISA system for clinical biomarker measurements are well known [9,10] and at present is the default method for measuring protein concentrations in many clinical laboratories around the world.
Collectively the IAMS and ELISA methods were tested for aspects of assay reproducibility, robustness and stability before comparison with a cohort of 100 plasma samples on each platform. These newly developed PromarkerD assays represent different strategies and workflows that are applicable in achieving high-throughput, robust, reproducible measurements of protein biomarkers that can be applied to any potential biomarker of any disease.

2. Materials and Methods

2.1. Reagents

Chemicals were from Sigma unless otherwise stated. NP-40, Tris (2-carboxyethyl) phosphine hydrochloride (TCEP) from Thermofisher; Iodoacetamide (IAM) from Astral Scientific. Synthetic isotopically labelled peptides (AQUA, Sigma, St. Louis, MO, USA) were LEPYADQL-R(13C6,15N4) from APOA4 protein, LVGGDNL-C(CAM)-SG-R(13C615N4) from CD5L protein, and FLNVLSP-R(13C6,15N4) from IBP3 protein. A standard reference plasma was created by combining EDTA plasma from three healthy volunteers before aliquoting and storage at −80 °C.

2.2. Clinical Samples

All clinical plasma samples were provided by the Fremantle Diabetes Study (FDS), a longitudinal observational cohort [11]. EDTA (Ethylenediaminetetraacetic acid) plasma was collected from all patients after an overnight fast and stored at −80 °C until required. The FDS protocol was approved by the South Metropolitan Area Health Service Human Research Ethics Committee. All subjects gave informed consent before participation. The clinical characteristics of the 100-person cohort from FDS are shown in Supplementary Table S1. These participants were similar in age, gender, BMI, diabetes duration, fasting plasma glucose and glycated hemoglobin (HbA1c) to those in the larger cohort used for the development and validation of PromarkerD (all p > 0.05) [5].

2.3. Antibody Production

A monoclonal antibody targeting the biomarker APOA4 was developed by the Monash Antibody Technologies Facility (MATF, Monash University, Melbourne, Australia) and the second antibody targeting APOA4 was developed by CDI laboratories (Mayaguez, PR, USA). Both pairs of monoclonal antibodies targeting CD5L and IBP3 were developed by CDI laboratories (Mayaguez, PR, USA). The hybridoma cell lines for all antibodies were used for the production of purified antibodies by the Monoclonal Antibody Facility of the Harry Perkins Institute of Medical Research (Perth, Australia), tested for specificity and provided for use.

2.4. Bead-Antibody Production

Batches of magnetic bead-antibody conjugates were made from Dynabeads® (M-270 Epoxy beads, Thermofisher, Waltham, MA, USA) according to the manufacturer’s instructions at a ratio of 50:1 beads:antibody. The bead-antibody conjugate batches (“Ab-beads”) were quality control tested and stored in PBS at a concentration of beads at 10 mg/mL at 2–8 °C until use. Long term stability testing indicates a shelf life of at least 5 months at 2–8 °C (data not shown).

2.5. Standards and Controls for IAMS Method

A calibrator standard was prepared by combining the recombinant protein biomarkers APOA4, CD5L and IBP3 (Sino Biologicals, Beijing, China) in phosphate-buffered saline (PBS) with human serum albumin (HSA, 5%, Sigma) as a carrier protein, at concentrations of 18.3, 0.686 and 0.178 µg/mL respectively. Once processed alongside plasma samples they represented effective plasma concentrations of 91.5, 3.45 and 0.892 µg/mL. The synthetic isotopically labelled peptides were used for signal normalisation. The standard reference plasma (stored at −80 °C) was processed (4 separate aliquots) with every batch of samples to allow the monitoring of assay performance and quality control (QC) measurements. The acceptance criteria for a batch of samples (96 well plate) was that the standard reference plasma biomarker concentrations interpolated from the calibrator standard were within 2 standard deviations of the rolling averages of the biomarker concentrations from all the batches for which data had been acquired.

2.6. Sample Processing for IAMS Method

Equal volumes of “Ab-beads” were pooled from the three individual stock solutions, then the liquid removed using a magnet to hold the beads. The beads were then resuspended in a volume of PBS to provide 150 µL volume of beads in each well corresponding to 120 µg of each antibody-bead conjugate per well. The following steps were carried out with a robotic handling system (Janus, Perkin Elmer, Waltham, MA, USA) unless otherwise stated. The Ab-beads were transferred from a trough to a 96 well plate (2 mL round bottom, Thermofisher). The calibrator standard (N = 4 replicates, 50 µL), the reference plasma (N = 4 replicates, 10 µL) and the samples (10 µL) were added to the plate along with blanks (PBS, 50 µL) and a double blank (200 µL PBS, no Ab-beads). All plasma samples had an additional 40 µL of PBS added to make all final volumes to 200 µL. The plate was incubated at 37 °C for 90 min with intermittent shaking (Thermomixer, Eppendorf, Hamburg, Germany) to keep beads suspended in solution. The liquid was removed with the plate on the magnet and the beads washed with 800 µL of 50 mM triethylammonium bicarbonate (TEAB), 150 mM NaCl, 0.1% NP-40 and then a further wash with 800 µL of 50 mM TEAB. After removing the liquid, the beads were resuspended in 56 µL of 50 mM TEAB, 5 mM TCEP containing 400 fmoles of each of three synthetic 13C15N labelled peptides (one for each target protein) and incubated at 55 °C for 20 min (Thermomixer, Eppendorf). To each well was added 6 µL of 100 mM IAM and incubated at room temperature (RT) in the dark for 20 min. Sequencing grade trypsin (Sigma, catalog no. 1418475001) was then added to each well (10 µL of 0.05 µg/µL in Milli-Q water) and incubated at 37 °C for 16 hrs. The solution was transferred into a clean 96 well plate (300 µL V bottom, Greiner) for LCMSMS analysis.

2.7. Targeted (MRM) LCMSMS Analysis for IAMS Method

The system used was a Shimadzu Prominence HPLC system with Loading pump (40 µL/min) to load the sample onto a trap column and two Nano pumps operating at a combined 5 µL/min to provide the analytical gradient. The column was kept at 40 °C and the 5 µL/min flow was directed to a QTRAP 5500 mass spectrometer (Sciex, Framingham, MA, USA) equipped with a 50-micron electrode and grounding unit (Sciex) for the Turbo-V ion source. From the V bottom plate, 25 µL of each sample for analysis was injected onto a MicroLC Guard Column C18 (Sciex) flowing isocratically at 2% (v/v) Acetonitrile, 0.1% (v/v) formic acid at 40 µL/min for 2 min. The flow was then directed from the microflow pumps (5 µL/min) through the guard column into a ChromXP C18, 3 µm 120 Å 300 micron ID, 5 cm column (Sciex) with a gradient of 10–40% acetonitrile, 0.1% formic acid over 2 min. The flow was ramped to 98% acetonitrile over 0.2 min, held at that concentration for 0.2 min before returning to the starting conditions of 10% acetonitrile, 0.1% formic acid over 0.1 min. The total runtime was 8 min. The mass spectrometry settings were as follows: Source Temp 250 °C; IonSpray Voltage 5500; Curtain Gas 25; Collision Gas Medium; GS1 and GS2 25; Entrance Potential 10; Collision Cell Exit Potential 14; Q1 Resolution Low; Q3 Resolution Low. The transition settings are shown in Supplementary Table S2.

2.8. Data Analysis for IAMS Method

Raw data files were imported into Skyline (v 3.6, MacCoss Lab Software, University of Washington, Seattle, WA, USA) and peaks integrated to provide peak areas for each of the peptides and their corresponding labelled version with a signal to noise (S/N) >5 for all peaks based on the Total Area and Total Background values provided by Skyline. The calibrator replicate ratios of unlabelled:labelled peptide peak area for each biomarker protein were averaged and the known concentration of the calibrator protein was used to determine the concentrations of each protein in unknown samples. In this way, the labelled peptide is used to normalise the signal across all samples analysed while the calibrator protein standards provide the concentrations.

2.9. Linearity, Detection Range, Stability, Reproducibility and Precision Testing

Linearity and detection range of the assay was demonstrated with a dilution series of the recombinant biomarker proteins in PBS and 5% HSA that was processed and analysed as for plasma samples.
To assess sample stability to freeze/thaw and other temperature variations the following experiments were carried out.
  • Two duplicates of 3 independent plasma samples (stored at −80 °C) were defrosted to 4 °C and then N = 3 samples were processed at 1 hour versus the duplicate N = 3 samples left at 4 °C for 24 h before processing.
  • Two duplicates of 3 independent plasma samples were defrosted with N = 3 kept at RT for 1 h before processing versus the duplicate N = 3 samples left for 24 h at RT before processing.
  • Three independent plasma samples were defrosted versus a set of N = 3 replicates that had two more additional freeze-thaw cycles (at least 1 hr frozen at −80 °C before thaw) performed on them before processing.
  • Previously injected replicates of processed plasma (N = 86) were left in the HPLC autosampler for a further 24 h after their initial analyses before being re-injected.
  • The precision testing was achieved with N = 4 replicates of a reference plasma processed in a single batch for an intra-day comparison and then this was repeated over twenty separate days for an inter-day comparison.

2.10. ELISA

The ELISA assay for the detection of APOA4, CD5L and IBP3 was developed by TGR BioSciences (Adelaide, SA, Australia). Antibodies were supplied as described above under “Antibody production”. Briefly, the ELISA uses the CaptSure™ technology (TGR BioSciences, Adelaide, SA, Australia) based on the principle of traditional sandwich format, but with a more efficient, faster, and simpler assay protocol than standard sandwich immunoassay. The analyte and both chemically tagged antibodies are added at the same time to the microplate that is pre-coated with the CaptsureTM reagent (TGR BioSciences) to immobilise the tagged antibody/biomarker complex. After a 1 hr incubation period, unbound assay reagents and analytes are washed away, and immuno-complexes are detected at an optical density of 450 nm. A 7-point standard curve for each biomarker covers the range of detection in plasma with different dilutions of the plasma required for each biomarker detection. Low and High QCs of recombinant protein are present on the plate to ensure assay integrity.

2.11. Cross-Platform Comparisons

A cohort of 100 patient samples from the FDS cohort were analysed by both the IAMS method and the developed ELISA. The biomarker concentrations from each method were then compared using Bland Altman plot analysis to determine the agreement between the two assays [12]. The mean bias between the two methods for each biomarker was determined from the Bland Altman difference plot. Any bias observed was then adjusted by applying the mean difference to the biomarker concentrations measured by ELISA. A hypothesis test for equality (Student’s t-test) was used to assess whether there was a statistical difference between concentrations measured by the two methods. The null hypothesis was that the bias is equal to zero (i.e., there is no difference between the two methods), against the alternative hypothesis that it is not equal to zero. Significant test p-values resulted in rejecting the null hypothesis and concluding that the bias is different from zero. The PromarkerD risk scores were calculated for each patient sample from each analysis method, using the adjusted ELISA concentrations. The two PromarkerD scores were then compared by the scatter plot and the correlation between the two methods assessed. An allowable difference of 5% was used to assess the number of subjects with larger differences between the methods. The acceptable percentage of subjects within the 5% difference was set at >90% of the cohort.

3. Results

Antibodies against the three biomarkers used in the PromarkerD test were produced to use them to develop an IAMS method and a more traditional ELISA test. The IAMS assay required a single antibody coupled to a magnetic bead while the ELISA used a pair of antibodies for each biomarker.

3.1. Immunoaffinity LCMSMS Assay (IAMS)

Antibodies against each biomarker were separately coupled to activated beads and firstly tested for their ability to bind purified recombinant protein and then the native proteins in plasma. After optimisation to produce a reproducible signal response the assay was developed in a 96 well format that allowed for the implementation of automated processing on a robotic handling platform for liquid handling steps. The use of magnetic beads enabled the easy removal of solutions with no meaningful loss of bead material. The original immunodepletion method used a nano LCMS (400 nL/min) with a 90 min run length. In contrast, the IAMS LCMS was optimised to an 8 min run using a flow rate of 5 µL/min into the ion source with a source needle of 50-micron internal diameter to provide spray stability at this flow rate. This method is shown in Figure 1 with a comparison to the ELISA method.
The linearity, limits of detection (LOD), limits of quantification (LOQ) and range of the assay were provided from a six-point dilution series of the three biomarkers processed by the assay [6]. The plasma working range and the LOD are shown in Table 1 (alongside equivalent values from the ELISA) with the R2 value for all three biomarkers >0.98. These quantification ranges covered the range of concentrations found in plasma samples in general using each method.
The robustness of the PromarkerD IAMS assay was demonstrated by temperature stability testing and intra/inter-assay comparison from biomarker measurements of a reference plasma. The temperature stability of the biomarker proteins was assessed using replicates of 3 independent plasma samples that were left for 1 h or 24 h at 4 °C or RT before processing and analysis. Similarly, multiple freeze-thaw cycles were tested on replicates of the 3 plasma samples as well as the stability of processed samples for 24 h at 4 °C to simulate delays in the injection of processed samples. For ligand binding assays (such as this IAMS method) the acceptance criteria, as defined by the FDA bioanalytical guidelines [13] is an accuracy within 20% of the nominal value and a precision CV% of <20% for the measurement of each biomarker. The results indicate that all biomarker comparisons between RT or 4 °C, between 1 h or 24 h, and between up to three freeze/thaw cycles, met these criteria [6].
Intra-assay (N = 4) and inter-assay (N = 20) comparisons of the biomarker concentrations were also demonstrated on the same reference plasma and shown in Table 2 as well as equivalent ELISA values.
The variability of both intra-assay or inter-assay measurements across 20 separate assays spanning a timeframe of two months had CVs < 11%. Collectively the stability and inter/intraday assay results demonstrate the robustness and reproducibility of measurement using the IAMS methodology.

3.2. ELISA Development and Testing

The ELISA assay was developed by TGR laboratories (Adelaide, Australia) from antibodies supplied to them. The ELISA utilised the CaptsureTM technology, a simple, one-step procedure, making it faster and easier to carry out than standard sandwich immunoassays. The basic method is compared to the IAMS method in Figure 1.
The concentration measurements were based on interpolation from a 7-point dilution series of each of the three recombinant protein standards with the R2 value of the fitted equation for each standard curve ≥0.99. Each biomarker had a separate plasma dilution factor and hence were run in separate wells. The working ranges and dilutions were shown in Table 1 with the IAMS data.
The stability of the plasma protein biomarkers was tested over multiple freeze-thaw cycles with five different plasma samples. All three biomarkers demonstrated robustness for up to 3 freeze-thaw cycles with average CV’s across the 5 samples and 4 measurements (0, 1×, 2×, 3× freeze thaws) to be 9.0%, 3.5% and 4.3% for APOA4, CD5L and IBP3 respectively.
Intra-assay precision was determined by analysing up to 3 biological replicates of plasma in the same assay on the same day. Inter-assay precision was determined by analysing the average of up to 3 biological replicates of plasma in 3 different assays. The precision of the ELISA in measuring the three biomarkers was shown in Table 2 with equivalent IAMS data.

3.3. IAMS vs ELISA Platform Comparison

The PromarkerD risk score is a prognostic prediction of a rapid decline in renal function. To prove the utility of the PromarkerD test as either an immunoaffinity (IAMS) or immunocapture (ELISA) method the two platforms were both used to measure biomarker concentrations from 100 plasma samples from the FDS collection. After a Bland Altman analysis of each protein, ELISA biomarker concentrations were adjusted (Figure 2, Figure 3 and Figure 4). The PromarkerD risk score was calculated and the results compared between the two methods, with a coefficient of correlation, R = 0.97 achieved (Figure 5).

4. Discussion

When creating an assay to test the efficacy of a biomarker to diagnose a disease state the initial considerations are naturally the practical considerations of accurately measuring the values required. Assay length, reproducibility in other laboratories, stability and clinical aspirations are often not addressed and take a backseat to the research-grade requirements. In the case of the PromarkerD test for chronic kidney disease, these factors have been addressed in this study and the two variations of the initial research method using immunodepletion, namely an immunoaffinity capture method and an ELISA method, are reflective of these considerations.

4.1. The “New” IAMS Immunoaffinity LCMSMS Assay

This manuscript has initially described the IAMS method with the conversion from a week-long immunodepletion method from processing to data analysis, into a two-day assay output for a batch of samples. This IAMS method was transformed from a labour intensive and expensive immunodepletion workflow that targeted three biomarkers from within a still complex sub-proteome of plasma, to an elegant multiplexing immunoaffinity capture-based methodology providing a clean and selective sample comprising only the biomarkers of interest.
An essential component of the viability of a clinical assay is the stability of the analytes under expected or potential conditions that might influence either the processing or detection characteristics. The temperature and time variables tested for this assay demonstrated the stable behaviour of the biomarkers under these conditions. Any assay for a clinical application to be employed over an extended period of time needs to be reproducible, both intra-assay and inter-assay, as demonstrated in this study.

4.2. The ‘Old’ ELISA

The ELISA technology is a tried and trusted method for clinical biomarker measurement and is typically the method of choice for clinical laboratories. To complement the IAMS method and provide options for commercial applications, the PromarkerD test was also converted into an ELISA format. Due to the large protein concentration range across the three biomarkers, it was not possible to multiplex the plasma processing as each biomarker used a different dilution of the plasma. However, a single plasma sample was diluted in series to provide detection of each of the three biomarkers in three separate plates in parallel. The stability of the plasma biomarkers was tested with multiple freeze-thaw cycles showing no significant deviation from the nominal values of a no-freeze thaw experiment. The inter-assay and intra-assay experiments also showed minimal deviation for expected values demonstrating the robustness of the ELISA platform in this application.

4.3. The ‘New’ versus the “Old”: IAMS and ELISA

The IAMS and ELISA platforms were tested against each other on a cohort of 100 plasma samples. Comparability of the individual PromarkerD biomarker concentrations were examined and a Bland Altman statistical analysis used to show the biomarker concentration data displayed no statistically significant difference between the two methods. However, differences in the relative performance of the two technologies between proteins were observed, specifically assays for CD5L and IBP3 gave higher correlations in sample concentration measurements than APOA4. These differences in performance merit investigation in future studies. Of note, and contrary to expectations, the ELISA’s exhibited greater dynamic range than the IAMS assays, which also merits further investigation.
The three proteins themselves showed comparable stability to variations in sample storage conditions, including freeze-thaw cycling, across both methods. This is a prerequisite for reproducible sample testing in a clinical environment where shipping conditions need to be flexible.
Individual protein concentration measurements were followed by a comparison of the PromarkerD risk score for progression of diabetic kidney disease for each of the 100 samples. There was high correlation between the “new” IAMS and “old” ELISA platforms with >90% of the cohort having risk scores within 5% of each platform’s score. This demonstrates the ability of either of these technologies to be used as clinical level assays for the measurement of PromarkerD biomarkers.

4.4. Protein Biomarker Assays in the Clinic

The power of proteomics is leading to the identification of a vast number of biomarkers for numerous diseases, with a Google Scholar search of the terms “proteomics” and “biomarker” yielding 300,000 hits. Despite this number, few proteomic biomarker panels have successfully transitioned from discovery to clinical use [14,15], with limited examples available: OVA2 for ovarian cancer [16], VeriStrat for lung cancer [17], Vectra for rheumatoid arthritis [18].
The major obstacles to this transition remain assay reproducibility and current clinical workflows which preference easy-to-use, low-cost technologies and fast turnaround times. Contrasting this, the omics revolution is highlighting the benefits of measuring multiple biomarkers simultaneously to achieve more accurate diagnoses of the disease state [19,20]. Although ELISA based platforms continue to drive current clinical practice, the assays themselves remain slow to develop, vulnerable to cross-reactivity, and difficult to multiplex.
The technology comparison described here emphasises that new developments in clinical proteomics, such as IAMS, can be applied effectively and in relatively high throughput to the measurement of multiplex biomarker panels. Importantly the approach uses the same concepts in sample enrichment which have proved so effective and reliable by ELISA, but with more adaptable and precise instrumentation.

5. Conclusions

Determining the concentration of a protein biomarker accurately, efficiently and precisely, is a requirement for any assay that relies on that measurement in making contributions to clinical decision making. We suggest the successful application of the three-protein PromarkerD test by immunoaffinity mass spectrometry offers a glimpse of the potential of this technique in clinical practice in the decades to come. Although each of the PromarkerD test methods discussed in this manuscript have different approaches, they all can provide a prognostic prediction of diabetic nephropathy outcome and identify at-risk patients allowing earlier intervention and monitoring of disease progression. This test and the technologies to implement it are a useful clinical tool to improve patient outcomes.

Supplementary Materials

The following are available online at https://www.mdpi.com/2227-7382/8/4/31/s1, Table S1: The baseline demographic and clinical characteristics of the 100 Fremantle Diabetes Study participants included in the current study. Table S2: Transition settings for mass spectrometry analysis.

Author Contributions

S.B. wrote the draft of the manuscript and was involved with the study design of all experiments and performing the immunoaffinity assays. J.I. was involved with running the immunoaffinity assays. T.C. assisted with the development and optimisation of the immunoaffinity assay. K.P. was involved with statistical analysis and Bland Altman plots. R.L. was involved with the design of the experiments, data analysis and editing of the manuscript. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Acknowledgments

The 100-person cohort was sourced from the Fremantle Diabetes Study Phase II (NHMRC grants #513781 and #1042231). The MS analyses were performed using NCRIS (National Collaborative Research Infrastructure Strategy) enabled Bioplatforms Australia infrastructure at the Western Australian Proteomics Facility, located at the Harry Perkins Institute for Medical Research. TGR Biosciences (Adelaide, Australia) developed the ELISA and performed an ELISA analysis of samples.

Conflicts of Interest

Scott Bringans, Jason Ito, Tammy Casey, Kirsten Peters, Richard Lipscombe are employees of Proteomics International, with Scott Bringans and Richard Lipscombe holding shares in the company. Proteomics International is a beneficiary of patent PCT/AU2011/001212 that relates to biomarkers described in this manuscript.

Abbreviations

Ab-beadsAb-beads magnetic beads conjugated with monoclonal antibodies to the biomarker proteins
ACR Albumin:creatinine ratio
APOA4 Apolipoprotein A4
CD5LCD5 antigen-like
CKD Chromic kidney disease
CV Coefficient of variation
eGFR Estimated glomerular filtration rate
EDTA Ethylenediaminetetraacetic acid
ELISA Enzyme-linked immunosorbent assay
FDA Food and Drug Administration
FDS Fremantle Diabetes Study
HSAHuman serum albumin
IAMIodoacetamide
IAMS Immunoaffinity mass spectrometry
IBP3Insulin-like growth factor-binding protein 3
LCMS Liquid chromatography mass spectrometry
LCMSMSLiquid chromatography mass spectrometry/mass spectrometry
LODLimits of detection
LOQ Limit of quantification
MRM Multiple reaction monitoring
N Number
NaCl Sodium chloride
PBS Phosphate buffered saline
QC Quality control
R Coefficient of correlation
R2 Coefficient of determination
RT Room temperature
SDS-PAGESodium dodecyl sulphate-polyacrylamide gel electrophoresis
S/N Signal to noise
TCEP Tris(2-carboxyethyl) phosphine hydrochloride
TEABTriethylammonium bicarbonate

References

  1. 2019–National-Chronic-Kidney-Disease-Fact-Sheet. Available online: https://www.cdc.gov/kidneydisease/pdf/2019_National-Chronic-Kidney-Disease-Fact-Sheet.pdf. (accessed on 7 May 2020).
  2. Dunkler, D.; Gao, P.; Lee, S.F.; Heinze, G.; Clase, C.M.; Tobe, S.; Teo, K.K.; Gerstein, H.; Mann, J.F.E.; Oberbauer, R. Risk prediction for early CKD in type 2 diabetes. Clin. J. Am. Soc. Nephrol. 2015, 10, 1371–1379. [Google Scholar] [CrossRef] [PubMed]
  3. Bringans, S.D.; Ito, J.; Stoll, T.; Winfield, K.; Phillips, M.; Peters, K.; Davis, W.A.; Davis, T.M.E.; Lipscombe, R.J. Comprehensive mass spectrometry based biomarker discovery and validation platform as applied to diabetic kidney disease. EuPA Open Proteom. 2017, 14, 1–10. [Google Scholar] [CrossRef] [PubMed]
  4. Peters, K.E.; Davis, W.A.; Ito, J.; Winfield, K.; Stoll, T.; Bringans, S.D.; Lipscombe, R.J.; Davis, T.M.E. Identification of novel circulating biomarkers predicting rapid decline in renal function in type 2 diabetes: The fremantle diabetes study phase II. Diabetes Care 2017, 40, 1548–1555. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Peters, K.E.; Davis, W.A.; Ito, J.; Bringans, S.D.; Lipscombe, R.J.; Davis, T.M.E. Validation of a protein biomarker test for predicting renal decline in type 2 diabetes: The fremantle diabetes study phase II. J. Diabetes Complicat. 2019, 107–406. [Google Scholar] [CrossRef] [PubMed]
  6. Bringans, S.; Ito, J.; Casey, T.; Thomas, S.; Peters, K.; Crossett, B.; Coleman, O.; Ebhardt, H.; Pennington, S.; Lipscombe, R. A robust multiplex immunoaffinity mass spectrometry assay (PromarkerD) for clinical prediction of diabetic kidney disease. Clin. Proteom. 2020, 17, 37. [Google Scholar] [CrossRef] [PubMed]
  7. Engvall, E.; Perlmann, P. Enzyme-linked immunosorbent assay (ELISA) quantitative assay of immunoglobulin G. Immunochemistry 1971, 8, 871–874. [Google Scholar] [CrossRef]
  8. Weemen, B.K.V.; Schuurs, A.H.W.M. Immunoassay using antigen-enzyme conjugates. FEBS Lett. 1971, 15, 232–236. [Google Scholar] [CrossRef] [Green Version]
  9. Frantzi, M.; Bhat, A.; Latosinska, A. Clinical proteomic biomarkers: Relevant issues on study design & technical considerations in biomarker development. Clin. Transl. Med. 2014, 3, 7. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  10. Füzéry, A.K.; Levin, J.; Chan, M.M.; Chan, D.W. Translation of proteomic biomarkers into FDA approved cancer diagnostics: Issues and challenges. Clin. Proteom. 2013, 10, 13. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  11. Davis, T.M.; Bruce, D.G.; Davis, W.A. Cohort profile: The fremantle diabetes study. Int. J. Epidemiol. 2013, 42, 412–421. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  12. Bland, J.M.; Altman, D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 327, 307–310. [Google Scholar] [CrossRef]
  13. Bioanalytical Method Validation Guidance for Industry. 2018. Available online: https://www.fda.gov/files/drugs/published/Bioanalytical-Method-Validation-Guidance-for-Industry.pdfv (accessed on 1 September 2020).
  14. Borrebaeck, C.A.K. Precision diagnostics: Moving towards protein biomarker signatures of clinical utility in cancer. Nat. Rev. Cancer 2017, 17, 199–204. [Google Scholar] [CrossRef] [PubMed]
  15. Srivastava, A.; Creek, D.J. Discovery and validation of clinical biomarkers of cancer: A review combining metabolomics and proteomics. Proteomics 2019, 19, 1700448. [Google Scholar] [CrossRef] [PubMed]
  16. Coleman, R.L.; Herzog, T.J.; Chan, D.W.; Munroe, D.G.; Pappas, T.C.; Smith, A.; Zhang, Z.; Wolf, J. Validation of a second-generation multivariate index assay for malignancy risk of adnexal masses. Am. J. Obs. Gynecol. 2016, 215, 82.e1–82.e11. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  17. Buttigliero, C.; Shepherd, F.A.; Barlesi, F.; Schwartz, B.; Orlov, S.; Favaretto, A.G.; Santoro, A.; Hirsh, V.; Ramlau, R.; Blackler, A.R.; et al. Retrospective assessment of a serum proteomic test in a phase III study comparing Erlotinib plus placebo with Erlotinib plus Tivantinib (MARQUEE) in previously treated patients with advanced non-small cell lung cancer. Oncol. 2019, 24, e251–e259. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  18. Savvateeva, E.; Smoldovskaya, O.; Feyzkhanova, G.; Rubina, A. Multiple biomarker approach for the diagnosis and therapy of rheumatoid arthritis. Crit. Rev. Clin. Lab. Sci. 2020, 1–12. [Google Scholar] [CrossRef] [PubMed]
  19. Gool, A.V.; Corrales, F.; Čolović, M.; Krstić, D.; Oliver-Martos, B.; Martínez-Cáceres, E.; Jakasa, I.; Gajski, G.; Brun, V.; Kyriacou, K.; et al. Analytical techniques for multiplex analysis of protein biomarkers. Expert Rev. Proteom. 2020, 17, 257–273. [Google Scholar] [CrossRef] [PubMed]
  20. Adamcova, M.; Šimko, F. Multiplex biomarker approach to cardiovascular diseases. Acta Pharm. Sin. 2018, 39, 1068–1072. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Comparison of the immunoaffinity mass spectrometry assay (IAMS) and ELISA platforms for analysis of three PromarkerD protein biomarkers from plasma.
Figure 1. Comparison of the immunoaffinity mass spectrometry assay (IAMS) and ELISA platforms for analysis of three PromarkerD protein biomarkers from plasma.
Proteomes 08 00031 g001
Figure 2. Difference plot (A) before Bland Altman correction comparing APOA4 concentrations derived from IAMS and ELISA platforms. Scatter plot (B) after adjusting ELISA concentration values with a Bland Altman analysis (adj = adjustment of +47.93 µg/mL to the ELISA value).
Figure 2. Difference plot (A) before Bland Altman correction comparing APOA4 concentrations derived from IAMS and ELISA platforms. Scatter plot (B) after adjusting ELISA concentration values with a Bland Altman analysis (adj = adjustment of +47.93 µg/mL to the ELISA value).
Proteomes 08 00031 g002
Figure 3. Difference plot (A) before Bland Altman correction comparing CD5L concentrations derived from IAMS and ELISA platforms. Scatter plot (B) after adjusting ELISA concentration values with a Bland Altman analysis. (adj = adjustment of +0.53 µg/mL to the ELISA value).
Figure 3. Difference plot (A) before Bland Altman correction comparing CD5L concentrations derived from IAMS and ELISA platforms. Scatter plot (B) after adjusting ELISA concentration values with a Bland Altman analysis. (adj = adjustment of +0.53 µg/mL to the ELISA value).
Proteomes 08 00031 g003
Figure 4. Difference plot (A) before Bland Altman correction comparing IBP3 concentrations derived from IAMS and ELISA platforms. Scatter plot (B) after adjusting ELISA concentration values with a Bland Altman analysis. (adj = adjustment of division by 9.52 µg/mL to the ELISA value).
Figure 4. Difference plot (A) before Bland Altman correction comparing IBP3 concentrations derived from IAMS and ELISA platforms. Scatter plot (B) after adjusting ELISA concentration values with a Bland Altman analysis. (adj = adjustment of division by 9.52 µg/mL to the ELISA value).
Proteomes 08 00031 g004
Figure 5. Scatter plot comparing PromarkerD risk predictions based on IAMS and ELISA platforms.
Figure 5. Scatter plot comparing PromarkerD risk predictions based on IAMS and ELISA platforms.
Proteomes 08 00031 g005
Table 1. IAMS and ELISA working ranges and dilutions.
Table 1. IAMS and ELISA working ranges and dilutions.
ProteinIAMSELISA
Plasma Working Ranges (µg/mL)LOD (µg/mL)Assay Working Ranges (ng/mL)Plasma Dilution FactorPlasma Working Ranges (µg/mL)
APOA437.5–2009.406.64–4251 in 4002.60–170
CD5L0.104–10.00.1000.16–101 in 16000.26–16.0
IBP30.0104–1.000.0100.63–401 in 2000.13–8.0
Table 2. IAMS and ELISA intra-assay and inter-assay variability. Precision based on the average % CV.
Table 2. IAMS and ELISA intra-assay and inter-assay variability. Precision based on the average % CV.
ProteinIAMSELISA
Intra-Assay (N = 4)
Precision (% CV)
Inter-Assay (N = 20)
Precision (% CV)
Intra-Assay (N = 3)
Precision (% CV)
Inter-Assay (N = 3)
Precision (% CV)
APOA49.49.44.22.8
CD5L7.69.83.43.9
IGFBP35.610.52.711.8
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Bringans, S.; Peters, K.; Casey, T.; Ito, J.; Lipscombe, R. The New and the Old: Platform Cross-Validation of Immunoaffinity MASS Spectrometry versus ELISA for PromarkerD, a Predictive Test for Diabetic Kidney Disease. Proteomes 2020, 8, 31. https://doi.org/10.3390/proteomes8040031

AMA Style

Bringans S, Peters K, Casey T, Ito J, Lipscombe R. The New and the Old: Platform Cross-Validation of Immunoaffinity MASS Spectrometry versus ELISA for PromarkerD, a Predictive Test for Diabetic Kidney Disease. Proteomes. 2020; 8(4):31. https://doi.org/10.3390/proteomes8040031

Chicago/Turabian Style

Bringans, Scott, Kirsten Peters, Tammy Casey, Jason Ito, and Richard Lipscombe. 2020. "The New and the Old: Platform Cross-Validation of Immunoaffinity MASS Spectrometry versus ELISA for PromarkerD, a Predictive Test for Diabetic Kidney Disease" Proteomes 8, no. 4: 31. https://doi.org/10.3390/proteomes8040031

APA Style

Bringans, S., Peters, K., Casey, T., Ito, J., & Lipscombe, R. (2020). The New and the Old: Platform Cross-Validation of Immunoaffinity MASS Spectrometry versus ELISA for PromarkerD, a Predictive Test for Diabetic Kidney Disease. Proteomes, 8(4), 31. https://doi.org/10.3390/proteomes8040031

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop