1. Introduction
Worldwide, more than 1.7 million cardiovascular devices were implanted each year, and the incidence is expected to rise in the future [
1,
2]. The foreign materials can trigger a local immune response in the host involving the coagulation cascade, complement system, platelets, as well as immune cells, and provide fertile conditions for bacterial biofilm formation and consequently result in medical device-associated infection [
2]. After the initial colonization, adhered bacteria can form a biofilm leading to a contamination of the implant material. This contamination can compromise the functionality and performance of the medical device itself but also affects the integration into the surrounding biosystem. Furthermore, this situation can cause an activation of the inflammatory response and poses a serious risk for the patient in terms of systemic infection, sepsis, and ultimately death [
3]. Up to 50–70% of healthcare-associated infections are caused by medical device-associated infection, posing a serious issue for clinical therapies [
2]. Bacterial contamination of cardiovascular implants or infections originating from medical device implantation often lead to infective endocarditis (IE), which represents a clinically challenging entity requiring multidisciplinary approaches for the diagnosis and therapy. IE with mortality rates of 20 to 25% remain a clinical problem despite advances in antimicrobial as well as surgical therapies. Implant materials are especially prone to bacterial adhesion and need up to 10,000 times less bacterial exposure to get contaminated in comparison to native materials. A possible reason could be that the artificial implant materials, rather than the native tissue, have no adaptive response towards the bacterial adhesion [
4,
5,
6]. In particular, heart valve leaflets were affected by bacterial colonialization and manifestation of IE. Once a mature bacterial biofilm has established, the bacteria in the biofilm are more resistant to antibiotics and antibiotic treatment is often not efficacious. The last option to eradicate the infection is implant removal because there is the risk of leaving a small fraction of non-growing persister bacterial cells that survive antibiotic therapy and may potentially reconstitute the biofilm after the antibiotic treatment is discontinued [
3]. About 10 to 30% of IE cases are caused by Streptococcus and Enterococcus species, whereas
Staphylococcus aureus (
S. aureus) is the most frequent pathogen, occurring in 70% of the IE cases, and accounts for around 30% of healthcare-associated infections [
2]. Therefore, new implant material strategies or functionalizations are needed to prevent medical device-associated infections and supplement the currently available therapies, especially for cardiovascular implant as well as intravascular catheter-related infections, which is the most common cause of nosocomial bacteremia [
7].
To date, only 4 antimicrobial technologies are reported with clinical outcomes for orthopedic implants: silver and iodine coatings, gentamicin poly-D, L-lactide (PDLLA) coating and a fast-resorbable hydrogel coating with Defensive Antibacterial Coating (DAC) based on hyaluronan and poly-L-lactide (PLLA) [
2]. In vitro models are not able to simulate the complex biological response to a foreign material in the bloodstream and in particular the elimination of the bacteria by the immune cells. In addition, animal models are commonly used to analyze antimicrobial agents in bacterial endocarditis on native heart valves. These in vivo models are very laborious and did not examine the antimicrobial response of implant materials or test materials in the bloodstream [
8,
9].
In this study, we describe an in vivo mouse model to test new implant materials in the bloodstream regarding their antimicrobial properties in the bloodstream. This model is easy to handle and involves no unnecessary suffering for the animals because the tested material is implanted minimally invasive into the tail vein as a micro-stent. The setup was created for a basic simulation of an IE situation originating from an infected implant or an intravascular catheter-related infection. Therefore, two clinically critical situations can be modeled in a simplified form: (i) the subsequent situation in IE after a bacterial biofilm has been established and the therapy options get complicated as well as (ii) catheter-related bloodstream infections associated with high mortality and also causing severe comorbidities such as IE.
2. Materials and Methods
2.1. Implant Design and Production
The used implants were designed as micro-stents in the form of tubes composed of the thermoplastic silicone polycarbonate elastomer (TSPCU; ChronoSil80A, AdvanSource Biomaterials, Wilmington, NC, USA) with a length of 10 mm, an inner diameter of 0.2 and outer diameter of 0.45 mm. The TSPCU micro-stents were generated via extrusion (HAAKE MiniLab II, Thermo Fisher Scientific, Karlsruhe, Germany) and cut with medical scissors (Braun, Melsungen, Germany) to the final length of 10 mm. Afterwards, the micro-stent was spray-coated with poly-L-lactide (PLLA; Resomer L210, Evonik Industries AG, Darmstadt, Germany) dissolved in chloroform (J.T. Baker, Radnor, PA, USA) and dried for 1 week at 37 °C in a vacuum drying chamber. The final PLLA coat thickness was about 20 µm. The diameter of the used micro-stent before and after the coating process was measured with a biaxial laser scanner (ODAC 32 XY, Zumbach Electronic AG, Orpund, Switzerland). The micro-stents were threaded on a 6-0 polypropylene suture material (Prolene, Ethicon, Norderstedt, Germany) for better handling. Micro-stents were sterilized for 10 s in 70% ethanol and rinsed 3 times in 1 mL sterile phosphate buffered saline (PBS, pH 7.4; Thermo Fisher, Schwerte, Germany). Handling of the micro-stents was performed in an aseptic environment.
2.2. Animals and Bacterial Culture
All animal experiments were performed according to the German animal protection guidelines that have been approved by the local animal care and use committee (7221.3-1-069/19). Female C57BL/6 mice with an age of 8–12 weeks were purchased from Janvier Labs (Saint Berthevin, France) and used in the experiments. All mice were kept on a standard diet and a 12 h day-night rhythm. A clinical endocarditis isolate of Staphylococcus aureus FR20 (gift from Institute for Microbiology University Greifswald) was used as bacterial strain. S. aureus was cultured in Luria broth (LB; Sigma-Aldrich, Taufkirchen, Germany) at 37 °C with 150 rpm shaking.
2.3. Infection of the Implants
To ensure an exponentially growing bacterial culture, an overnight culture of S. aureus was diluted 1:50 in LB. The pre-culture was incubated at 37 °C with 150 rpm shaking until an optical density OD600nm of 0.5–0.7 was reached. Optical density was measured in a 96-well plate (Greiner Bio-One, Frickenhausen, Germany) using a micro-plate reader (FLUOstar Omega, BMG LABTECH GmbH, Ortenberg, Germany). For the assessment of the implant contamination concentration different protocols with varying incubation time and bacteria concentration of the incubation solution were tested. After the infection with the bacteria, the implants were washed 3 times with 1 mL PBS, transferred to a new tube containing PBS with 0.1% Tween-20 (Sigma-Aldrich, Taufkirchen, Germany), placed in a sonification bath (USC300T, VWR International, Darmstadt, Germany) for 5 min and the detached bacteria were determined by plating a serial dilution of the bacterial culture on LB agar plates. For the implant infection, the pre-culture was diluted to an OD600nm of 0.1 or 0.5. The sterilized implants were incubated in 1 mL diluted pre-culture for 30 min at 37 °C and 150 rpm shaking, followed by a wash step of 1 mL PBS repeated 3 times. The incubation of the implants with the diluted pre-culture having an OD600nm of 0.1 resulted in a total bacteria load on the implant of 1 × 106 and with an OD600nm of 0.5 resulting in 5 × 106 CFU (colony forming units). CFUs were determined by plating a serial dilution of the bacterial culture on LB agar plates. Up to the implantation, the infected micro-stents were stored in PBS at 4 °C for a maximum of 3 h.
2.4. Implantation Procedure
Female C57BL/6 mice were anesthetized with a mixture of ketamine (75 mg ketamine hydrochloride per kg bodyweight; bela-pharm GmbH, Vechta, Germany) and xylazine (10 mg xylazine hydrochloride per kg bodyweight; Rompun® 2%, Bayer Healthcare GmbH, Leverkusen, Germany) via intra peritoneal application. After reaching a deep unconscious state, the mice were placed in a tail vein restrainer (Tailveiner, AgnThos, Lidingö, Sweden), the tail was disinfected via Octenisept spray (Schülke & Mayr GmbH, Norderstedt, Germany) and the vein was manually compressed before puncture. The implants were injected in the lateral tail vein via a syringe applicator with a 21 G cannula (Braun, Melsungen, Germany). The implants were pushed out by a stainless-steel wire (0.4 mm; Rayher, Laupheim, Germany) fixed at the syringe plunger. The animals were divided into 3 groups: (a) receiving a sterile implant; (b) infected implant with 1 × 106 CFU S. aureus and (c) 5 × 106 CFU S. aureus. For each group, the implants were kept in the mice for 6 h, 24 h and 48 h with n = 5 animals per group and time. Animals given a sterile implant served as controls.
2.5. Bacterial Load Determination
For the analysis of the bacterial load, spleens were homogenized by disruption of the organ with a 5 mL syringe plunger (Braun, Melsungen, Germany) in a 70 µm cell strainer (Falcon, Schwerte, Germany) and flushed with PBS. Serial dilutions of the homogenized organs as well as harvested blood were plated on LB agar plates. The CFU were counted after 24 h incubation at 37 °C.
2.6. Cytokine Quantification
The concentration of Interleukin-6 (IL-6), granulocyte colony-stimulating factor (G-CSF) and C-reactive protein (CRP) in the blood plasma was performed via a magnetic bead luminex assay (Bio-techne GmbH, Wiesbaden, Germany). Blood plasma was retained after 10 min centrifugation at 4000× g and snap frozen in liquid nitrogen. For the luminex assay, the blood plasma was diluted 1:2 and the assay was performed according to the manufacturer’s instructions. Quantification was calculated using the standards provided and measuring a standard curve according to the manufacturer’s instructions. The readout was done with the Luminex® 100/200™ System and the xPONENT® 3.1.871.0 software (Luminex Corperation, Austin, TX, USA).
2.7. Gene Expression Analysis
RNA of the tissue adjacent to the implants was isolated with the NucleoSpin RNA kit (Macherey-Nagel, Düren, Germany). For this propose, the mouse tail was cut into a section retaining only the implant and the tail vein with the implant in these sections was flushed with the lysis buffer at least 3 times using a syringe with a 21 G needle (Braun, Melsungen, Germany). RNA concentration was determined with the Colibri micro-volume spectrometer (Berthold Technologies GmbH & Co.KG, Bad Wildbad, Germany). cDNA synthesis was performed using the RevertAid First Strand cDNA Synthesis Kit (Thermo Fisher Scientific, Schwerte, Germany) following the manufacturer’s instructions. Gene expression analysis was accomplished via quantitative real-time PCR (qPCR) using TaqMan gene expression master mix (Applied Biosystems, Thermo Fisher Scientific, Schwerte, Germany) and TaqMan assays (see
Table 1), all purchased from Thermo Fisher Scientific. The qPCR was performed using the QuantStudio 3 (Thermo Fisher Scientific, Schwerte, Germany) under the following reaction conditions: 50 °C for 2 min followed by 95 °C for 10 min, 40 cycles at 95 °C for 15 s, and 60 °C for 1 min. The ΔΔCt method was employed for relative quantification.
Gapdh served as endogenous control.
2.8. Histological Staining
Mouse tail sections were fixed in 4% formaldehyde (Süsse Labortechnik, Gudensberg, Germany) and embedded in paraffin after removal of the bone tissue (USEDECALC solution, Medite, Burgdorf, Germany). Histological sections with a thickness of 4–6 µm were cut with the microtome (Hyrax M55 rotary microtome, Carl Zeiss, Oberkochen, Germany). The slides were deparaffinized by two-time incubation for 10 min in xylene (JT Baker, Phillipsburg, NJ, USA) and rehydrated by a serial dilution of ethanol (100%, 95%, 90%, 80%, 70%) and deionized water for 3 min each. For the Masson Trichrome staining, the Trichrome Stain (Masson) and, for the gram staining, Gram Staining Kit (both Sigma-Aldrich, Taufkirchen, Germany) were used according to the manufacturer’s instructions. The DeadEnd™ Fluorometric TUNEL System (Promega, Walldorf, Germany) enabled labelling of the apoptotic tissue of the samples after following the manufacturer’s instructions. The TUNEL sections were counterstained with DAPI to visualize cell nuclei and embedded in Fluoroshield™ mounting medium (Sigma-Aldrich, Taufkirchen, Germany). Image acquisition was performed with the Primovert inverted microscope equipped with AxioCamMRc (Carl Zeiss, Oberkochen, Germany) for the Masson Trichrome as well as gram staining and with the confocal laser scanning microscope LSM 780 (Carl Zeiss, Oberkochen, Germany) for the TUNEL staining.
2.9. Statistical Analysis
For all experiments, the statistical analyses were carried out with GraphPad Prism5 software (GraphPad Software Inc., La Jolla, USA). Results are presented as means ± SD. Data analyses were performed using the two-tailed Mann–Whitney U test. Asterisks are used in the figures to specify statistical significance (*: p < 0.05; **: p < 0.01; ***: p < 0.001).
4. Discussion
The use of antimicrobial materials could help to reduce implant-related infections and prevent subsequent challenging treatments [
10].
The used micro-stent coating material PLLA is a well-known and established coating material for cardiovascular devices and controlled drug release systems [
11] including direct mixing or adsorption of antibiotics or drug encapsulation in microspheres or core-shell porous nanofibers to avoid burst-release [
12,
13,
14]. Therefore, this material is an ideal candidate for the validation of our in vivo model for testing antimicrobial materials in the bloodstream for cardiovascular devices. The micro-stent can be coated with other polymeric material or chemically functionalized, e.g., with antibiotics, antimicrobial peptides, or antifouling coatings. Hence, the model is highly adaptable to test different materials in an easy-to-handle and non-laborious manner. This in vivo method enables the imitation of intravascular catheter-related infection and the pathology of an established IE in a simplified form and therefore facilitates investigation of new treatment options as well as material innovations for this clinically difficult condition. Our study covers two crucial and dangerous consequences of the biofilm formation: (i) the local inflammation caused by the adhered bacteria as biofilm and (ii) the systemic inflammation caused by the circulating bacteria reflecting the cell spreading and detachment from the mature biofilm.
The use of a lethal dose of bacteria in the reported range of ~1–3 × 10
7 CFU [
15] was avoided or not desired. Especially considering the higher susceptibility to contamination of implant materials, a formed bacterial biofilm is hard to treat and therefore testing of new implant materials with an already established bacterial biofilm can demonstrate their antimicrobial potential. A lethal dose may obscure the local immune response to the biofilm with the overactive systemic immune response. With the investigated non-lethal ICC, a bacterial clearance of the circulating bloodstream bacteria by the mice was shown after 24 h. IL-6 and CRP as acute phase reactants are elevated in response to infection, with IL-6 as the main inducer of CRP expression [
16]. Both cytokines were elevated in response to the infected implants after 6 h. Subsequently, due to the fact that the IL-6 concentration decreased after 24 h, the CRP level also returned to baseline values. G-CSF stimulates the proliferation and maturation of granulocyte precursors and activates granulocyte function, which is essential for the elimination of bacteria during the early state of bacterial infection [
17]. This likely reflects the clearance of the bacteria from the bloodstream after 24 h and signals an immune response that is still present but only slightly activated after 48 h. For IE, IL-6 as well as G-CSF levels were reported to increase in the human blood plasma [
18]. G-CSF and IL-6 has been shown to be involved in neutrophil production as well. Neutrophils are the immune cells which play an important role in the host defense against microorganisms [
19]. Thus, the measured ICC dependent increases of IL-6, CRP as well as G-CSF after 6 h highlight the short systemic activation of the immune response of the mice after receiving the infected implants, which also declined again after clearance of the blood circulating bacteria after 24 h.
The common testing methods for antimicrobial materials are subcutaneous implantations in mice [
20,
21] or rabbits [
22]. However, these models take only into account the local infection with indirect contact to the cardiovascular system. IE is usually simulated in time-consuming rabbit [
8], rat [
23] or mouse [
9] models based on surgical valve trauma by placing a polyurethane catheter at the aortic root and subsequent infection with bacteria. These models do not include the testing of materials.
In addition, we saw in our experiments thrombus aggregates, probably consisting of clotted blood as well as immune cells and bacteria, formed near the infected or injured endothelial site. These clots were also found as vegetations, which is a common pathogenesis of IE. These vegetations can substantially influence the function of the heart valves in IE [
10,
24] and their constitution as well as formation can also be analyzed with our in vivo model especially regarding the initiation of the local inflammatory response.
S. aureus is known to induce the cell surface expression of ICAM-1 as well as VEGF secretion in endothelial cells as a pro-inflammatory response to mediate monocyte adhesion and recruitment [
25,
26]. VEGF is also essential in angiogenesis for promoting endothelial survival [
27]. This pro-inflammatory and pro-coagulant phenotype of the endothelial tissue is observed in our in vivo model and is typical for endovascular infections including IE.
Antibiotic-releasing materials are under intensive research but pose the risk of having only a short window of activity and afterwards only deliver subinhibitory doses that can lead to the development of antibiotic-resistant bacteria [
28]. The in vivo model presented in this study can be extended to any implantation duration to additionally investigate drug-resistant behavior of the bacteria against the tested material. Antimicrobial material grafting or functionalization implement their antimicrobial function via immobilization of antimicrobial peptides [
29] or physicochemical changes such as polycationic [
30] or zwitterionic coatings [
31] that may interrupt the net negative charge of the bacteria membrane. But under in vivo conditions in the bloodstream, various other physical, chemical and biological conditions may act on these materials that might affect the physicochemical properties and therefore impact the life-span of the antimicrobial properties. These functional analyses can also be performed with the presented in vivo model. Nonetheless, materials preventing the adhesion of bacteria via hydrophobic or nanostructured surfaces [
28] cannot be directly tested with the presented in vivo model because it will probably not be possible to sufficiently contaminate these implants with bacteria. But the in vivo conditions affecting the surface modification and its antimicrobial life-time can be simulated by this model and the antimicrobial properties can be examined in vitro after the explantation. Metal-based antimicrobial coatings, e.g., silver or copper coatings can also be tested with this in vivo model. Silver is one of the most investigated antimicrobial agents and is incorporated in materials in elemental, salt, complexed or nanoparticle forms. The silver cations disrupt the function of the bacterial cell membrane and metabolic proteins. But regarding the low threshold concentration for cytotoxicity and the low bio- as well as hemocompatibility index of the metal-ions, this poses some greater difficulties for these coatings in the application of cardiovascular devices [
28,
29]. Finally, various bacterial strains can be investigated with the presented in vivo model and also compared between each other. Therefore strain-specific antimicrobial differences can be highlighted and analyzed for their mechanisms in causing and developing IE.
After all, our simplified model has limitations regarding the direct translation of the results to humans. The human immune system is more complex than the immune system of mice [
32]. In the case of catheter-related infection, scar tissue formed around the injection site may also have an impact on the infection event and the body’s response to the injected material. The relation between the bacteria number that burdens the host also cannot be transferred in its entirety from mice to humans. However, results from a mouse model can provide indications of possible responses in the human organism.