1. Introduction
Multi-photon microscopy (MPM) is a powerful technique for the minimally invasive imaging of live brains with subcellular resolution and for performing three-dimensional imaging with depths ranging from a few hundred microns to a few millimeters using two- and three-photon excitation processes [
1,
2,
3,
4,
5,
6]. Because of the nonlinear interaction between ultrashort pulses and the brain at the near-infrared wavelengths, MPM can provide the high-resolution imaging, as well as deeper penetration depth. The distinctive advantages of MPM adoption in neuroscience include the ability to visualize the brain activity and structure with subcellular resolution (even imaging fine structures such as spine and axons) in vivo at millimeter depths with or without labeling the different cell types in the brain. Although confocal and light sheet microscopy also provides subcellular resolution, they are only capable of imaging shallower depths compared to multiphoton microscopy. Because it is critical to understand the connectivity between neurons at distinct cortical layers when they process sensory and other kinds of information, multiphoton microscopy is particularly useful in neuroscience to address these types of questions.
A femtosecond pulsed laser is used as a laser source for the MPM. Generally, the ultrashort laser pulses can be generated either by using bulk or fiber-type laser systems. Titanium-doped Sapphire (Ti:S) lasers, which provide stability, short pulse widths in the femtosecond region, and extended wavelength tunability, are the most frequent conventional lasers used in MPM. Ti-S lasers produce wavelengths ranging from 700 to 1300 nm at a repetition rate of 80 MHz. In general, femtosecond laser pulses with this spectral range and repetition rate are appropriate for exciting most fluorophores for structural and functional brain imaging via two-photon interactions [
7]. When performing three-photon imaging of the brain, the laser characteristics differ dramatically from those used for two-photon microscopy. For three-photon microscopy, the number of photons that are absorbed during this process can be described as follows [
5,
6]:
where the terms are explained as follows:
n: number of excited photons,
: excitation cross-section,
P: average power,
: pulse width,
R: repetition rate,
NA: numerical aperture,
h: Planck ‘s constant,
c: speed of light, and
: wavelength.
Therefore, it is important to increase the laser power and lower the repetition rate and the pulse width in order to increase the number of photons excited in a three-photon process while keeping the numerical aperture and the wavelength constant. Specifically, the wavelength of the laser has been chosen at longer wavelength ranges, such as 1300 and 1700 nm. The laser average power is the key player in three-photon microscopy for brain imaging, because, the high average power can result in (a) tissue heating through linear absorption of longer excitation wavelengths, (b) nonlinear damage at the focal plane due to high peak power/intensity, (c) saturation of fluorophores due to the high pulse energy. Due to these effects, it is crucial to consider laser parameters and perform control experiments to make sure that it is safe to perform 3p imaging in the living brain [
6,
8]. Also, the repetition rates must be decreased to 0.5 to 2 MHz, and the pulse widths must be less than 50 fs [
5,
8,
9]. Significant progress has recently been made in the development of optical parametric amplifiers (OPAs) that can produce longer excitation wavelengths with moderate repetition rates at short pulse widths. Although these commercial lasers offer several multiphoton imaging choices in the brain [
10], they have several drawbacks, including high costs and sophisticated and bulky equipment. As a result, democratizing these imaging techniques in neuroscience labs is extremely difficult.
As an alternative to conventional ultrafast lasers, great progress has been made in fiber lasers, which can overcome various disadvantages of conventional lasers. In particular, fiber lasers are very small and compact, can be operated with air cooling, have a high pump-to-signal conversion efficiency, have a lower overall cost than conventional solid-state lasers, and can provide unconventional laser characteristics such as longer wavelengths, high repetition rates, variable pulse energy, and narrow pulse widths. Thus, in the recent decade, the use of fiber lasers for neuroscience applications has emerged. The simplest way to generate multiple wavelengths in a fiber laser is to adapt various types of gain fibers to achieve the required wavelength of light [
11,
12]. The ytterbium (Yb) fiber is employed to generate 1000 nm wavelengths [
13]. Similarly, the laser wavelength spectrum of 1550 nm is achieved by using Erbium (Er)-doped fiber as the gain medium [
14,
15] and the thulium (Tm)-doped fiber emits a laser wavelength spectrum of 1700 nm [
16]. Recent advancements in femtosecond fiber lasers have been reported in fiber laser oscillators as well as fiber amplifiers. In laser oscillators, important advancements have focused on the development of novel saturable absorbers (SAs), various cavity designs, similaritons, all-normal dispersion (ANDi) configurations, dual-wavelength generation, variable repetition rates, and multimode fiber laser oscillators [
17]. Similarly, innovative fiber amplifiers concentrate on chirped pulse amplification and direct amplification approaches for increasing the energy of ultra-short laser pulses. Eventually, the average power of the laser pulses increased to over 1 Watts [
18] at MHz repetition rates, which is essential for achieving multi-photon effects such as two- and three-photon fluorescence, as well as label-free imaging in the brain [
19].
In this review, we focus on recent advances in the development of ultrafast fiber lasers for neuroscience applications, with a specific emphasis on two- and three-photon fluorescence and label-free microscopy for structural and functional brain imaging. As a future perspective, we propose that new breakthroughs in fiber laser technology will cover a wide range of applications in the field of neuroscience. We believe that this review is critical for neuroscientists and laser scientists to understand the current state-of-the-art fiber lasers for performing multiphoton imaging in living brains, as well as to see potential new developments in fiber lasers to democratize their use in neuroscience labs.
The rest of this paper is organized as follows: In
Section 2, multiphoton microscopy for neuronal imaging is discussed. In
Section 3, the advances in ultrafast lasers for neural imaging are presented. In
Section 4, fiber laser-based multiphoton microscopy (FL- MPM) applications in neuroscience are discussed. Finally, the future directions and outlook, as well as the conclusion are presented in
Section 5 and
Section 6, respectively.
2. Multiphoton Microscopy (MPM) for Neuronal Imaging
Conventionally, several imaging techniques, including positron emission tomography (PET), ultrasonography, and functional magnetic resonance imaging (fMRI), have been proposed for structural and functional brain imaging. However, they are limited in terms of spatial and temporal resolution [
20]. These disadvantages can be solved by optical imaging techniques, such as confocal, light-sheet, two-photon (2P), and three-photon (3P) microscopes. The main advantage of these technologies is to have the capability of imaging the structure and function of the brain at sub-cellular and greater temporal resolutions. When we compare the spatial and temporal resolution capabilities of multiphoton, light sheet, and confocal microscopy techniques, they all exhibit almost the same performance. On the other hand, light sheet and confocal microscopy enable brain imaging at shorter depths, but 2P and 3P microscopy techniques can image all cortical layers, white matter, and even subcortical regions in intact brains [
5,
6,
21]. There are several advantages and challenges of 2P and 3P microscopy for performing structural and functional brain imaging. In minimally invasive 2P microscopy for functional brain imaging, it is very feasible to perform high-speed (kHz rate) and larger field of view (several millimeters) functional imaging mostly at superficial cortical layers [
22,
23,
24,
25]. However, it is challenging to perform 2P microscopy in deeper regions of the brain due to high out-of-focus signal generation. When compared to 2P microscopy, 3P microscopy may readily minimize the challenge of deep brain functional imaging by utilizing longer excitation wavelengths (1300, and 1700 nm) at moderate imaging speeds (5–10 Hz) and a smaller field of view [
4,
5,
6]. Therefore, the challenges of functional brain imaging with 3P microscopy are (a) to increase the imaging speed to kHz range at deeper brain regions, (b) to increase the imaging depth to subcortical regions, (c) to increase the field of view to image multiple brain regions simultaneously. For structural brain imaging, two-photon fluorescence microscopy (2PEF), three-photon fluorescence microscopy (3PEF), second-harmonic generation (SHG), and third-harmonic generation (THG) microscopy are utilized. The challenges of 2PEF, 3PEF, SHG, and THG microscopy for brain imaging include developing fluorophores with greater absorption peaks at longer excitation wavelengths and developing PMT modules that can have higher sensitivity outside the visible range of light to perform deeper brain imaging. A typical MPM experimental setup and different kinds of MPM modalities are explained in
Figure 1.
A typical MPM setup is shown in
Figure 1a. In a typical multiphoton microscope, a femtosecond pulsed fiber or a conventional laser is used as a laser source. The collimated light is directed to the pair of Galvo mirrors for scanning the beam in the two-dimensional space. Then, the light beam is passed through a scan and tube lens, which magnifies the beam to fill the back aperture of the microscope objective. Finally, the laser beam is transmitted by the dichroic and focused on the tissue through the objective with a beam size in the range of micrometers. The reflected light is directed to the same dichroic mirror which sends the beam to the highly sensitive photon multiplier tubes (PMTs). In front of each PMT, there are band-pass filters to select an appropriate emission wavelength for different kinds of MPM techniques (
Figure 1a). There are numerous varieties of MPM based on the degree of the nonlinear excitation process and the type of contrast mechanism. Two-photon excited fluorescence (2PEF), second-harmonic generation (SHG), three-photon excited fluorescence (3PEF), and third-harmonic generation (THG) microscopy techniques are described in
Figure 1b. When two lower energy photons are absorbed at the same time, electrons shift from the ground to the excitation energy state, resulting in two-photon excited fluorescence (2PEF). The electrons then lose some energy as they travel to a lower energy ground state level and produce another photon with a wavelength more than half that of the incident photons (
Figure 1b). Second-harmonic generation (SHG) occurs when two photons of identical energy excite an electron to a virtual energy state. When the electrons return to the ground state, they emit a photon with half the wavelength of the incident photons (
Figure 1b). In terms of three-photon interactions, three-photon excited fluorescence (3PEF) occurs when electrons absorb three photons simultaneously and move to the excitation state. The electrons then lose energy as they migrate from the higher energy excitation states to the ground state, releasing another photon with a wavelength greater than one-third of the excitation of the incident photons. On the other hand, third-harmonic generation (THG) is a parametric harmonic generation process similar to SHG in which three photons of the same energy are involved in the excitation process of electrons to a virtual energy state, and when the excited electrons return to the ground state, they emit photons at one-third of the excitation wavelength of incident photons. The important challenges of 2PEF, 3PEF, SHG, and THG microscopy for brain imaging include developing fluorophores with greater absorption peaks at longer excitation wavelengths and developing PMT modules that can have higher sensitivity outside the visible range of light to perform deeper brain imaging. Besides, femtosecond laser sources should possess versatile characteristics in order to enhance the MPM performance for neuroimaging applications.
3. The Advances in Ultrafast Fiber Lasers for MPM
The technology developments in ultrafast fiber lasers for MPM mainly focus on extending the excitation wavelengths (>1000 nm) to perform structural and functional brain imaging [
26,
27]. These applications require moderate repetition rates (<80 MHz) and higher pulse energies (>10 nJ) compared to conventional ultrafast lasers [
26,
27]. Finally, it is demonstrated that signal generation in two-photon (2p) and three-photon (3p) microscopy is inversely proportional to the pulse width [
6,
8]; thus, it is very advantageous to reduce the pulse width (<100 fs) in order to generate a stronger signal with these microscopy techniques.
The widespread adoption of two-photon microscopy was aided by the development of reliable near-infrared femtosecond-pulse sources, such as the titanium–sapphire laser. The benefits of using longer excitation wavelengths were confirmed using a Ti:S-pumped optical parametric oscillator (OPO) [
17]. But, the complete system is expensive, occupies more space, is complex, and requires expertise to handle the laser. This has led researchers to focus on the development of fiber lasers that possess all the merits of conventional laser systems, along with easy handling and low operational costs.
Table 1 summarizes the advantages and disadvantages of fiber lasers and traditional lasers. To be specific, the advantages of usage of fiber laser for neuroscience applications are as follows:
In fiber laser-based sources, the signal propagates through the fiber’s core, eliminating the need for alignment while maintaining the same beam quality;
By combining the fiber laser oscillator with a fiber amplifier, the average power can be scaled up;
The laser’s output travels through the fiber and it is easily connected to a miniscope or an endoscope;
The components used to construct the fiber lasers are commercially available, and they can be easily replaced, reducing the overall cost to an order of magnitude that of a commercial laser.
In this section, we focus on the principles of ultrafast lasers, various configurations of fiber lasers, fiber amplifiers, and fiber laser characteristics.
3.1. The Recent Developments in Ultrafast Fiber Lasers
3.1.1. Mechanism of Mode-Locking Pulse Generation
Generally, a femtosecond fiber laser consists of a fiber laser oscillator connected to one or more fiber amplifiers to reach the required pulse energy and pulse duration for MPM applications. The laser oscillator works on the principle of mode-locking to produce ultra-short laser pulses. In
Figure 2a, a typical fiber laser configuration is exemplified. The mode-locking oscillator consists of a continuous wave (CW) pump source which operates at 975 nm and emits a pump power of 500 mW. Based on the active fiber selection and its fiber emission spectrum as a gain medium, (for example ‘Yb’, ‘Er’, and ‘Tm’), the laser operating regime is determined. A wavelength division multiplexer (WDM) is utilized to combine the pump signal (975 nm) and the gain medium light (1064 nm). An isolator is also used to prevent cavity damage from backward reflection and to assure unidirectional operation. The light is then sent through a saturable absorber, which dampens low-intensity pulses while amplifying higher-intensity pulses across several round trips. To evaluate the laser properties inside the oscillator, a coupler is employed to split a part of the laser signal. In the last two decades, the advancements were more focused on the configuration of the laser cavity and enhancing the nonlinear behavior of the saturable absorber [
28].
The inset of
Figure 2a depicts the concept of mode-locking that produces ultra-short pulses by locking longitudinal modes in-phase with each other, which are equally spaced by c/2L inside the laser cavity. The modes whose corresponding frequencies fall within the gain bandwidth of the laser are amplified and lased out.
3.1.2. Various Operating Regimes
Figure 2b shows the various operating regimes of fiber laser oscillator which includes soliton, stretched pulse, similariton, and all-normal dispersion (ANDi) [
29,
30,
31]. Among them, soliton regimes are widely adopted in both conventional and fiber lasers for mode-locking the laser pulses. Soliton pulses are generated by balancing anomalous dispersion with Kerr non-linearity that arises in the fiber [
32]. Although the soliton pulses are very stable, they are limited by the smaller pulse energies that can be calculated by the soliton area theorem
(here,
is the pulse energy,
is the pulse duration,
is the nonlinear coefficient, and the group velocity dispersion is denoted as
), which are normally in the range of pJ. In order to meet higher pulse energy requirements, the similariton and ANDi regimes were discovered. “Similariton” laser pulses are parabolic in nature and can be generated when an arbitrary pulse propagates through a longer length of gain medium with normal dispersion inside the laser cavity. It possesses unique properties, including being free from wave breaking under strong nonlinearity and offering transform-limited compression. Consequently, high-energy pulses can be generated from the similariton laser cavity with good stability [
33].
3.1.3. Types of Saturable Absorbers
Another advancement in laser oscillator technology is the discovery of innovative saturable absorber technologies based on optical phenomena and materials, as shown in
Figure 2c. The saturable absorber is required in ultrafast lasers to initiate the mode-locking operation. Over the previous two decades, two types of SAs have been deployed. Real saturable absorbers comprise two main types: bulk semiconductor saturable absorber mirrors (SESAMs) [
34] and nanomaterial-based saturable absorbers (SAs). The latter category includes quantum dots [
35], single-walled carbon nanotubes [
36], graphene [
37], transition metal dichalcogenides (TMDs) like MoS2 and WS2 [
38], as well as topological insulators (TIs), black phosphorus (BP), and MXenes [
32], which are all subtypes of TMDs. In mode-locked lasers, SESAMs have constraints such as high costs and narrow bandwidth. On the other hand, nanomaterial-based SAs require a complicated fabrication technique and have a low damage threshold. Nonlinear polarization rotation (NPR), nonlinear optical loop mirror (NOLM), and nonlinear magnifying annular mirror (NALM) SAs [
32] from the second group can overcome these constraints. These SAs are highly resistant to high power and are commonly utilized in lasers of various wavelengths. These SAs have the disadvantage of being affected by environmental perturbations, and eventually, mode-locking is disturbed. The most recent development in SAs, based on nonlinear multimode interference (NL-MMI), presents an all-fiber, compact solution that is less affected by environmental effects [
39]. Among these, graphene and CNT SA-based ultrafast fiber laser research is particularly appealing for neuroscience applications due to its capacity to tune the laser across a wider range of wavelengths, achieve short pulse durations, and generate stable laser pulses. In MPM, a CNT-based femtosecond laser has recently been used to image biosamples using SHG, THG, 2PFE, and 3PFE [
40]. The advantages of a CNT-based femtosecond laser for MPM applications include lower costs and a longer wavelength of operation. In the artificial SA category, nonlinear polarization rotation SA-based lasers are widely used for MPM applications because of a stable generation of laser pulses with requisite energy and narrow pulse width [
26].
3.1.4. Tunable Repetition Rate and Wavelength
The predominant share of optogenetics and imaging research in neuroscience is heavily dependent on femtosecond lasers with pulse energies ranging from 500 nJ to 100
J and repetition rates spanning from a few hundred kHz to a few MHz for two-photon optogenetics [
41], a higher repetition rate (80 MHz) and lower pulse energies of (10–50 nJ) for two-photon imaging [
21], and a lower repetition rate (0.5–4 MHz) and higher pulse energies (1–2
J) for three-photon imaging [
6].
The mode-locked oscillator’s repetition rate is determined by its entire cavity length. Most fiber oscillator designs use long optical fibers with typical repetition rates in the 10 MHz range. The repetition rate can be lowered by inserting more passive fibers into the cavity. However, lowering the repetition rate of a ring-cavity oscillator below 1 MHz necessitates a large cavity length of approximately 200 m, which is impractical. Therefore, the most feasible technique to generate ultrashort pulses with a repetition rate of less than 10 MHz is to use an external acousto-optic pulse picker, as shown in
Figure 2d. By implementing this method, we can achieve a repetition rate of down to the kHz range. Another significant advancement in ultrafast fiber laser technology is the generation of longer wavelengths (>1050 nm), which can be accomplished by utilizing the soliton self-frequency shift (SSFS) [
42], supercontinuum generation, and multi-wavelength generation lasers. SSFS is effectively achieved by utilizing the intra-pulse Raman scattering effect in an optical fiber, which results in shifted longer wavelengths of above 1300 nm from the shorter pump wavelength. This is accomplished by moving energy from a higher frequency to a lower frequency. In order to achieve SSFS for the excitation wavelengths below 1300 nm, photonic crystal fibers (PCFs) are usually employed because their dispersion can be flexibly engineered. Furthermore, when suitable PCFs are pumped by a Yb-fiber laser, SSFS can lead to a soliton pulse with the center wavelength shifted beyond 1300 nm, even reaching 1700 nm [
43,
44]. The researchers’ focus has recently switched to constructing tunable ultrafast fiber laser sources for imaging multiple fluorophores in the brain. Therefore, they have been developing nonlinear fiber-optic technologies to develop tunable-wavelength ultrashort pulses from a fiber laser source. Supercontinuum (or) white light generation is normally caused by the interaction of the nonlinear effect of self-phase modulation and linear dispersion. As a result, the pulse leading edge has red-shifted components, and the trailing edge has blue-shifted components. Further, normal dispersion leads to increased separation between these red and blue components in the temporal domain, causing the pulse to broaden, whereas anomalous dispersion cancels it entirely and induces pulse stabilization in the form of an optical soliton, depending on the properties of the fiber used and the pump source wavelength [
45,
46]. Recently, this type of supercontinuum has been used in neuroscience imaging and optogenetic stimulation. A recent study revealed the development of a supercontinuum laser for imaging mouse brain tissue with GCaMP6s as well as perturbing it with optogenetic stimulation with C1V1-mCherry opsin [
47]. Multi-wavelength generation sources are necessary in neuroscience to perform two- and three-photon brain imaging with the same laser source. Furthermore, these laser sources allow us to perform optogenetic stimulation too [
48,
49].
3.2. Various Configurations of Ultrafast Lasers
All-normal dispersion (ANDi) laser and similariton laser configurations are well suited for MPM applications since the characteristics of the laser pulses meet the requirement of MPM for neuroscience applications.
Figure 3a depicts a schematic of an ANDi fiber laser implemented in a multiphoton microscope for neuronal imaging. An ANDi laser oscillator was recently built using multimode high-power pump diodes, along with passive and gain fibers. Without using any of the fiber amplifiers, this setup produced a pulse energy of 30 nJ with 81 fs pulse width and a repetition rate of 40 MHz. As a result, additional noises that are generated by fiber amplifiers can be avoided [
26].
Figure 3b shows a successful demonstration of employing an ultrafast source at 1550 nm, a repetition rate of 1 MHz, and a pulse duration of 500 fs that was used to generate an SSFS at 1617 nm using a photonic crystal rod integrated into a multiphoton microscope. The produced soliton pulse has a pulse width of 74.5 fs and a pulse energy of 141 nJ [
50]. Another way of generating a longer wavelength for three-photon microscopy was to utilize a material like Cr-forsterite (Cr-F) in the laser medium.
Figure 3c depicts a schematic experimental view of a Cr:F laser oscillator. The cavity configuration was a z-fold design achieved by inserting concave mirrors. They used a Yb-doped fiber laser as a pump source, with a maximum pump power of 14 W. An ultrashort pulse with a pulse width of 32 fs at 1260 nm, a repetition rate of 24 MHz, and a pulse energy of 22 nJ was produced by this cavity configuration, which operates in the soliton regime. The total cavity length was 6.22 m. Even though they can generate a pulse energy of 22 nJ without an amplifier stage, this pulse energy was insufficient for deep tissue imaging of the mouse brain [
51]. Very recently, a Raman shifter was used to shift the excitation wavelength to longer wavelengths for performing three- and four-photon microscopy in the brain. The schematic representation of the 1800 nm femtosecond fiber laser system is shown in
Figure 3d. Here, an Er-doped fiber laser oscillator was used, which produced laser pulses with a pulse duration of 492 fs, a repetition rate of 50 MHz, and an output power of 1 mW. The demonstrated system at 1800 nm was able to produce a 150 fs pulse duration. The pulses were amplified using a two-stage Tm: ZBLAN fiber amplifier. After the final stage, they increased pulse energy up to 1
J at 0.5 MHz [
16].
3.3. Fiber Amplifiers
3.3.1. Chirped Pulse Fiber Amplifier (CPA)
The schematic representation of CPA is shown in
Figure 4a, which consists of (1) an ultrafast mode-locked laser oscillator that generates seeding pulses; (2) a pulse stretcher that extends the seeding pulse to the picosecond level; (3) pre-amplifiers and power amplifiers; (4) a pulse picker used to adjust the repetition rate; and (5) a pulse compressor that dechirps the amplified pulse to nearly transform-limited pulse width [
15,
52].
3.3.2. Master-Oscillator Power-Amplifier (MOPA)
The schematic representation of MOPA system is depicted in
Figure 4b. Generally, the picosecond and nanosecond laser pulses are amplified with high average output power and peak power using MOPA. In this scheme, an ultrafast fiber oscillator generates stable pulses, which are then amplified by subsequent fiber amplifiers to increase average power and pulse energy. Chirped output pulses of the oscillator are directly fed to an amplifier without any stretcher unit. The master oscillator in a MOPA system only needs to provide low-power/energy pulses and can thus be carefully engineered to achieve low noise, high compactness, and extreme robustness. In subsequent fiber amplifiers, the weak seeding pulses are amplified by orders of magnitude in pulse energy or average power. In order to reduce the nonlinear effects, large-mode area fiber or double-clad gain medium is used in the power-amplifier stage [
53,
54].
3.3.3. Gain-Managed Nonlinear (GMN) Fiber Amplifier
Figure 4c shows a schematic illustration of GMN. Amplification is carried out in a new gain-managed regime that can provide pulses with high energy and a spectrum wider than the gain bandwidth. In this approach, a single-stage amplifier is sufficient to provide maximum pulse energy in the
J range while laser pulses can be as short as 50 fs. Recently, using a gain-managed nonlinear fiber amplifier, researchers were able to achieve a maximum pulse energy of 58 nJ, a laser repetition rate of 31 MHz, and a compressed pulse duration of 33 fs [
18].
3.4. Ultrafast Fiber Laser Characteristics
The repetition rate, pulse energy, pulse width, peak power, average power, spectral bandwidth, and central wavelength are all measurable properties of an ultrafast fiber laser system. In practice, tuning of these parameters is required for optimized laser and microscopy performance while performing wider, deeper, and faster brain imaging [
55] and optogenetics. Because of technical limits, maximizing one parameter always compromises another.
Figure 5 shows example measurements of some of these fiber laser parameters.
The wavelength spectrum is centered at 1033.2 nm as shown in
Figure 5a. The mode-locked laser oscillator pulse train is depicted in
Figure 5b, and the time interval between the single pulse is measured to be 46.2 ns.
Figure 5c shows the radio frequency (RF) spectrum of the laser output peak, which is centered at 21.35 MHz and has a signal-to-noise ratio (SNR) of 58 dB.
Figure 5d presents the measured autocorrelation trace of the mode-locked laser with pulse durations of 2 ps (inset) and 300 fs before and after the compression, respectively.
Figure 6a shows the laser pre-amplifier spectrum, which is centered at 1033.2 nm.
Figure 6b presents the power-amplifier output spectrum at 11 W average power. These two figures (
Figure 6a,b) show that there is a slight change in the spectral profile of the amplifier output, which could be due to the occurrence of a nonlinear effect. Amplified spontaneous emission (ASE) is normally present in any fiber amplifier system and this emission occurs due to the spontaneous emission during the amplification of laser pulses. In a high-power amplifier, this is considered as the amplifier noise, and it can limit the gain of the amplification. Therefore, the ASE should be within a particular limit to achieve high power amplification. In order to verify the presence of an ASE signal while scaling the power in the power amplifier, the output spectrum of the laser signal is measured and the pump spectrum exactly peaks at 975 nm and the strength of the ASE spectrum is very low, which is within the acceptable limit (
Figure 6c).
Figure 6d represents the output power versus input pump power of the power amplifier’s simulated and experimental characteristics [
52].
4. Fiber Laser-Based Multiphoton Microscopy (FL-MPM) Applications in Neuroscience
Despite significant advancements in fiber laser technologies, their use in neuroscience research has been limited. In this section, we will present research studies that used fiber lasers to accomplish fluorescence and label-free two- and three-photon imaging in diverse species’ brains. Ultrafast fiber lasers are also used in temporal focusing (TF) microscopy for two- and three-photon excitations, where the axial modulation is carried out to increase the axial confinement in a wide field without point-scanning which eventually enhances the optical sectioning in multiphoton microscopy. The diffraction gratings are used to separate the spatial frequencies and temporally focus the light. As a result, there is no nonlinear effect until the light reaches the focal plane where there is a spatiotemporal focusing of the beam. Recently, temporal focusing has been shown to perform functional and structural brain imaging via two- and three-photon excitation [
10]. In vivo multiphoton microscopy is an emerging imaging tool for performing structural and functional brain imaging within intact animals at cellular resolution. Moreover, 2PEF and SHG microscopy techniques have been used to perform structural and functional brain imaging in shallower depths. TPEF has been used to perform functional brain imaging via calcium indicators [
21] and structural brain imaging of elements such as neurons and microvasculature with fluorescent dyes [
26]. SHG microscopy is a label-free technique and is sensitive to structural features with broken inversion symmetry, such as collagen fibers in the skull and meninges [
56] and microtubules [
57]. On the other hand, 3PEF and THG microscopy techniques have been used to perform structural and functional brain imaging in deeper regions of the brain. More specifically, 3PEF microscopy has been used to perform functional brain imaging with calcium indicators [
5,
6], as well as to perform structural imaging of microvasculature, neurons, astrocytes with fluorescent dyes [
6,
9,
16]. THG is sensitive to the index of refraction and third-order susceptibility changes in the brain. Therefore, THG has been used to image the microvasculature as well as myelinated axonal tracts [
5,
6,
58].
4.1. Two-Photon Fluorescence Microscopy
4.1.1. Functional Imaging
Two-photon microscopy has been the workhorse of functional brain imaging with animal models at sub-cellular resolution. The majority of commercial Ti:sapphire oscillators with high repetition rates are used in two-photon brain imaging applications. The main advantage of commercial two-photon lasers involves their high wavelength tunability, which allows for the use of various calcium indicators for functional brain imaging. There has recently been development in producing calcium indicators that can be excited using standard fiber laser excitation wavelengths. In a recent study, researchers developed a jYCaMP, a calcium indicator suited for excitation wavelengths around 1030 nm [
27].
At layer 2/3 of the primary visual cortex, they utilized a commercial Ytterbium fiber laser (Spark Lasers) at 1040 nm with 31.25 MHz to accomplish dual color neuronal (Thy1:(jRGECO1a transgenic mice)) and axonal (jYCaMP) imaging as illustrated in
Figure 7a. Although commercial lasers have a high repetition rate, they can cause damage to the living brain since significant power is required to accomplish high-resolution deep brain imaging. Recently, an adaptive fiber laser source was developed to excite a region of interest, resulting in a 30-fold reduction in the power demand of two-photon functional brain imaging in awake mice with traditional calcium indicators such as GCaMP6s [
59]. Researchers developed a custom−made fiber laser based on SSFS at 1840 nm and used a BIBO crystal to generate a 920 nm wavelength at a repetition rate of 32 MHz. They compared the spontaneous activity of neurons at a 680
m depth with point scanning standard 2-photon imaging and an adaptive excitation source (AES), as shown in
Figure 7b. The field of view was 700 × 700
m, and the frame rate at 512 × 512 pixels was 30 Hz. They demonstrated that the photon number per neuron is 7 times greater than without AES, although using 4.5 times less average power with AES.
4.1.2. Structural Imaging
Since the calcium indicator development at fiber laser excitation wavelengths is still in its early stages, most of the early fiber laser research focused on structural fluorescence brain imaging with two-photon microscopy. At 40 MHz repetition rate, a seminal work used a multimode pump and double-clad fiber design to generate 30 nJ energy pulses with uncompressed average outputs of 1.2 W [
26].
The authors injected the Texas Red fluorophore retro-orbitally into transgenic YFP mice, which expressed yellow fluorescent protein in layer V pyramidal neurons. Consequently, they could observe the structure of layer V pyramidal neurons, including their apical dendrites and the microvasculature, up to a depth of 700–800
m, as shown in
Figure 8a. At the surface and the 750
m depth, the imaging power ranged from 20 to 200 mW. They also compared the structural imaging performance of their custom-made fiber laser to that of a commercial Ti:S laser at 920 nm. They used 380 mW of average power, as opposed to 200 mW, to image the neuronal soma at a depth of 900
m with the commercial laser and the fiber laser, respectively, under identical imaging settings. They were able to resolve neuronal soma with the fiber laser but not with the commercial laser at a 900
m depth. They also demonstrated that the fiber laser could resolve dendritic spines (about 1
m in size) up to a 200
m depth. Another work used a fiber laser with a Yb fiber amplifier and a diamond Raman laser to generate two wavelengths of 1060 and 1250 nm simultaneously [
60]. They used these two lasers to do neuronal and microvasculature imaging with a galvo–galvo scanner. The Raman laser was used to image the brain vasculature tagged with Alexafluor 647, and the Yb fiber laser was utilized to image the neuronal somata labeled with tdTomato up to a depth of 750
m as illustrated in
Figure 8b. Another seminal paper focused on the development of a fiber laser using circularly polarized soliton-frequency shift (SSFS) at 1617 nm to perform two-photon imaging of brain vasculature labeled with indocyanine green (IGC), as shown in
Figure 8c [
50]. At the laser exit, they generated pulses of 147 nJ pulse energy and 74.5 fs pulse width at a repetition rate of 1 MHz. They then demonstrated the ability to image the brain vasculature to a depth of 2 mm with a maximum pulse energy of 56 nJ as demonstrated in
Figure 8c.
4.2. Two-Photon Label-Free Imaging
Second-Harmonic Generation Microscopy
In addition to performing structural two-photon fluorescence microscopy in the brain, researchers also made significant advances in performing label-free second-harmonic generation (SHG) microscopy for performing structural two-photon microscopy in the brain. In particular, SHG contrasts only structural features with broken inversion symmetry, making it a suitable probe for buried interfaces, inducible molecular dipoles, polarized structures, and field-induced anisotropy. In a recent study, researchers developed a custom-made Cr:forsterite laser to perform both second- and third-harmonic generation imaging of fixed mouse brain acquired by a 3xTg mouse utilized as a mouse model for Alzheimer’s disease (AD). The laser generated a 1262 nm wavelength at a 105 MHz repetition rate. The pulse width was 38 fs and the power on the sample varied between 20 to 30 mW [
61]. Their goal was to achieve label-free imaging of amyloid beta plaques (A
) and neurofibrillary tangles (NFT), which are hallmarks of Alzheimer’s disease. They discovered that A
plaques generated both SHG and THG signals, whereas NFT produced predominantly SHG signals, and axons/dendrites produced mostly THG signals. They also showed that they could track the evolution of AD disease by comparing fixed mouse brain slices from different ages of 3xTg mice to age-matched samples from control mice. In another study, researchers developed an extended-cavity Cr:forsterite laser integrated with a photonic-crystal fiber solution frequency shifter and a periodically poled lithium niobate spectrum compressor for simultaneous harmonic generation and coherent Raman brain imaging in fixed mouse brain tissues [
62]. They developed a home-built Cr:forsterite laser oscillator with a central wavelength of 1250 nm and a pulse width of 40 fs. This combination produced laser pulses with an energy of 18 nJ at a repetition rate of 20 MHz, equal to 360 mW average power. They then used second- and third-harmonic generation microscopy with an average laser power of 20–30 mW to image brain slices from C57BL/6 mice. SHG microscopy was used in this study to visualize cell bodies as black patches representing the lowest density of interfaces and the highest optical homogeneity of the medium within the cell bodies. A smaller scale texture between the cell bodies, related to nerve fibers and dendritic structure, effectively generating a formed surface, is particularly efficient in SHG. In this study, they were able to visualize a distinctive layered structure of the hippocampus, matching the stratum moleculare, stratum granulosum, and polymorphic layer by using SHG microscopy.
Table 2 shows the fiber laser parameters that have been used for performing functional and structural imaging and their corresponding imaging depth achieved in 2PEF microscopy. Further, ultrafast fiber laser parameters used for SHG are also listed for the comparison with 2PEF microscopy. From the comparison, it can be concluded that the SHG requires less pulse energy with shallow depth imaging when compared to 2PEF, and the imaging is label free.
4.3. Three-Photon Fluorescence Microscopy
4.3.1. Functional Imaging
Although two-photon microscopy has shown significant potential for structural and functional brain imaging, it has major limits for deep-brain functional imaging due to high optical intensities and a poor signal-to-background ratio. Three-photon microscopy mitigates these challenges and has been used to image structural and functional features of deep cortical layers in the mouse brain [
6,
58] at subcellular resolution in our own studies. Nonetheless, these experiments were carried out using commercial solid-state lasers. Recent improvements have been achieved in the development of custom-built fiber lasers to perform three-photon microscopy for functional brain imaging. In a recent study, a Cr-forsterite oscillator with a center wavelength of 1260 nm was developed to perform 30-Hz functional three-photon brain imaging of Drosophila [
51].
The laser repetition rate in that study was 24 MHz, the hyperbolic secant pulse width was 32 fs, and the laser output power was 530 mW, equivalent to a pulse energy of 22 nJ. They monitored the neuronal activity of Kenyon cells tagged with GCaMP7f while electric shocking the Drosophila as shown in
Figure 9a. In another study [
59], researchers used transgenic jRGECO1a mice to perform 30 Hz functional three-photon imaging at a depth of 750
m with adaptive excitation source (AES) based on a self-soliton frequency shift at 1700 nm. They compared the performance of the fiber laser system with and without AES and they found that the photon number per neuron with AES is 30 times higher than that without AES, although the same power of 35 mW was used. The spontaneous activity of neurons was recorded and a signal-to-background ratio with AES was significantly higher compared to that without AES, as illustrated in
Figure 9b.
4.3.2. Structural Imaging
Similar to three-photon fluorescence functional imaging studies, in very recent studies, structural three-photon imaging in mouse brains has been performed. In a very recent study, researchers developed a custom-made fiber laser consisting of an Er-doped silica fiber laser and a Raman shift fiber in order to generate a wavelength of 1800 nm with 150 fs pulse width, and 1
J pulse energy at 0.5 MHz on the mouse brain [
16].
To demonstrate the performance of their system in vivo, they labeled the excitatory neurons of the mouse brain with TurboFP636 virally in the somatosensory cortex (S1). They varied the laser power between 50–85 mW while they were imaging from the surface to a 700
m depth, as shown in
Figure 10a. Their acquisition speed was 1 Hz at 512 × 512 pixels. In another study, researchers developed a two-color fiber laser at 1300 and 1700 nm wavelengths with optical parametric chirped pulse amplification (OPCPA). They used a Yb-doped fiber amplifier (YDFA) as a pump source at 1030 nm to generate a signal beam at 1700 nm as well as an idler beam at 2600 nm. They frequency-doubled 2600 nm to achieve 1300 nm wavelength output. This system generated ultrashort pulses at 69 fs with 0.71
J at 1300 nm and 65 fs with 3.1
J at 1700 nm output wavelengths at 1.25 MHz repetition rate, respectively. They imaged a fixed mouse brain up to a depth of 600
m first to show that their 3-photon dual-color system worked. Then, they performed dual-color three-photon imaging with HEK cells, chicken embryo spinal cord, and live adult zebrafish brain, as shown in
Figure 10b [
49].
4.4. Three-Photon Label-Free Imaging
Third-Harmonic Generation Microscopy
Although most previous studies in neuroscience focused on fluorescence three-photon microscopy with exogenous fluorophores, three-photon microscopy also provides label-free brain imaging via third-harmonic generation (THG), which provides contrast from myelinated axons and brain microvasculature.
In a recent study, a fiber laser with a master-oscillator power-amplifier (MOPA) at 1650–1700 nm wavelength range, 50 MHz repetition rate, 70 fs pulse width, and 400 mW average power was developed [
63]. Researchers used this fiber laser to image the fixed Rainbow-3 mouse cortex and brain stem. At depths of 20–150
m, they performed three-photon fluorescence (3PEF) imaging of tdTomato-labeled brain microvasculature and THG imaging of myelinated axons as presented in
Figure 11. They also demonstrated simultaneous 3PEF and THG imaging of a living brain with developing Drosophila embryos [
63]. THG has also been shown to provide pathological information regarding tumor-affected human brain tissues.
Table 3 shows the fiber laser parameters that have been used for performing functional and structural imaging and their corresponding imaging depth achieved in 3PEF microscopy. Further, ultrafast fiber laser parameters used for THG are also listed for the comparison with 3PEF microscopy. From the comparison, it can be concluded that the THG requires fewer energy pulses compared to 3PEF. On the other hand, THG offers label-free imaging. THG provides real-time images of unstained human brain tissue, highlighting features such as increased cellularity, nuclear pleomorphism, and rarefaction of neuropils [
64].