Article

Quantitative Evaluation of the Batwing Effect on Depth Measurement in Coherence Scanning Interferometry for Rectangular Gratings

1
School of Electronic and Optical Engineering, Nanjing University of Science and Technology, Nanjing 210094, China
2
Jiangsu Institute of Metrology, Nanjing 210023, China
3
Shanghai Key Laboratory of Online Test and Control Technology, Shanghai 201203, China
*
Author to whom correspondence should be addressed.
Photonics 2024, 11(4), 341; https://doi.org/10.3390/photonics11040341
Submission received: 27 February 2024 / Revised: 26 March 2024 / Accepted: 3 April 2024 / Published: 8 April 2024
(This article belongs to the Special Issue Optical Interferometry)

Abstract

:
Coherence scanning interferometry (CSI) is a widely used non-contact method for measuring areal surface topography. The calculation of groove depth typically employs the ‘W/3 rule’ specified in ISO 5436-1. However, the batwing effect causes overshoots near the groove edges, which can introduce noise singularities in the selected W/3 region for depth calculation, thereby affecting the measurement of rectangular grating depth. This paper introduces the definition of batwing height and width and proposes a simulation model that considers various factors, such as the center wavelength, spectrum of the illuminating light, shadow effect, and numerical aperture, to analyze their influence on depth measurement. The simulation results demonstrate that the batwing width is the primary factor influencing depth measurement for small grating periods. Specifically, a decrease in numerical aperture increases the batwing width, leading to larger depth measurement errors, while a decrease in the center wavelength reduces the batwing width, resulting in smaller depth measurement errors. The influence of the illumination spectrum and shadow effect on depth measurement is found to be minor. Experimental validation using a step standard consisting of rectangular gratings with different periods and depths confirms the agreement between the experimental and simulated results. The proposed method provides a quantitative evaluation of depth measurement accuracy in CSI.

1. Introduction

Coherence scanning interferometry (CSI) is a widely used technique for areal surface topography measurement [1]. The International Organization for Standardization (ISO) recommends employing the ‘W/3 rule’ for calculating the groove depth of rectangular gratings [2]. However, the batwing effect, characterized by an overshoot at the edges of the grooves, can have a significant impact on the accuracy of depth measurements [3,4]. Importantly, as the period of the rectangular grating decreases, the batwing extends beyond the region of W/3 near the edge of the groove [5], leading to significant errors in depth measurement for small grating periods. Therefore, it is necessary to establish and analyze the relationship between batwing and depth measurement errors. The overshoot profiles of the batwing are influenced by various factors, such as the center wavelength, spectrum of the illuminating light, shadow effect, and numerical aperture (NA). The above can be quantitatively analyzed to identify the dominant factor that affects depth measurement errors. This paper specifically examines grating depths smaller than the coherence length of the illuminating light, as the batwings cease to appear beyond certain depths [4].
Modeling is an effective approach for investigating correlations between key factors and measurement outcomes. Many previous works have established models to investigate these relationships. Peter De Groot devised a computationally efficient theoretical model that analyzed the relationship between the illumination geometry, the spectrum of the illuminating light and the interference signals [6]. Additionally, He established an elementary Fourier optics model to analyze the relationship between surface and main features of CSI topography images [7]. Harasaki developed a diffraction model elucidating the relationship between the coherence length of the illuminating light and the batwing effect [4]. Xie proposed a Fourier model for analyzing the relationship between the lateral resolution and the transfer function of vertical scanning white-light interferometers [8,9]. Tobias presented a 3D simulation model investigating the transfer characteristics of confocal microscopes regarding 2D surface structures [10]. Su proposed a physics-based virtual CSI based on the foil model and obtained an inverse filtering is shown to reduce measurement errors [11,12]. Peter Lehmann suggested that the three-dimensional transfer function of the CSI is crucial for the correct analysis and reconstruction of the surface of the object [13]. Although the aforementioned works explain the correlation between main factors and measured topography, and some have addressed the impact of the batwing effect on step height measurements as well as the elimination of this effect in reconstructed morphology [14,15], none specifically investigate the relationship between the batwing effect and measurement error when utilizing the ‘W/3 rule’ for depth calculation.
The batwing is usually taken as a pseudo signal, and it is directly filtered or the topography where the batwing occurred is directly ignored. But whether the batwing effect will affect the depth calculation is still unknown, even if the overshot on the topography seems like to be filtered out. It is evident that batwings within the selected W/3 region intuitively influence depth measurement errors, but the exact relationships remain indirect. In this paper, we present a method that aims to determine the height and width of the batwing. Additionally, we seek to establish the correlation between the batwing effect and the errors encountered in depth measurements. Additionally, a simulation model is proposed and utilized to analyze these correlation relationships, ultimately identifying the primary factors that contribute to errors in measuring the grating depth.

2. Definitions and Models

Figure 1 illustrates the ‘W/3 rule’ and provides definitions for the batwing height and width. In this figure, W represents the measured width of the groove. Least square lines are fitted to regions A and B, and subsequently to region C [2].
The groove depth d is determined by the distance between the fitted lines.
Z = α h X + β h + h δ   .
where α h , β h , and h are fitted via the least squares method to a profile. The variable δ takes the value +1 in regions A and B and the value −1 in region C. The depth of the groove d is twice the estimated value of h.
The batwing height h b is defined as the difference in height between the highest point on the batwing profile and the fitted line.
h b = α h x h z h + β h + h δ 1 + α h 2   .
where ( x h , z h ) represents the coordinates of the highest point on the batwing profile. The batwing width w b is defined as the distance between the two intersection points of the batwing profile and the fitted line, with these points being the closest intersections on either side of the batwing.
w b = x 1 x 2   .
where x 1 and x 2 are the two intersection points of the batwing profile and the fitted line.
It is important to note that as the period of rectangular gratings decreases, the batwing extends beyond the W/3 region near the groove edge, affecting the accuracy of depth measurements in regions A, B, and C where least square lines are fitted. Additionally, the batwing width and height vary depending on the parameters of the microstructures being tested and the specifications of the CSI instrument. Therefore, it is necessary to analyze the influence of the batwing effect on groove depth measurements of rectangular gratings using CSI and the ‘W/3 rule’. Establishing a model provides an effective and quantitative approach to understand the relationship between influencing factors and depth measurement errors.
The simulation process is depicted in Figure 2. Initially, the nominal period and depth of the rectangular grating are provided as inputs. Subsequently, the depth scan index, modulation factor, illumination spectrum, and shadow factor are combined using the interference intensity equation to generate the corresponding interferograms. The envelope evaluation method is then employed to reconstruct the grating profile [16]. Finally, the ‘W/3 rule’ is applied to calculate the depth of the rectangular grating and the height and width of the batwing from the measured profile.
The object taken in the simulation is a rectangular grating H ( x , y ) with period Λ and groove depth d [8]. The illumination spectrum S ( λ ) is assumed to follow a Gaussian distribution [17].
We assume that the depth of the grating is significantly smaller than the width of the coherent envelope, and that the power on the pupil plane is uniformly distributed. In contrast to Xie’s definition [9], we define shadow factor Γ as the ratio of reflected power to incident power at the edge of the grating. Figure 3 illustrates the shadow effect at the edge of the rectangular grating, which is expressed as:
Γ = 1 r m a x < r θ 2 r θ r m a x r 2 π 2 arccos r θ / r d r + 2 0 r θ 2 π r d r / ( π r m a x 2 ) r m a x > r θ ,
with
α m a x = arcsin A N ,
r m a x = d tan ( α m a x ) ,
r 0 , r m a x , α 0 , α m a x ,
r θ = x n Λ Λ , x ( n Λ Λ + 0.5 Λ , n Λ Λ + 0.75 Λ ) n Λ Λ x . x ( n Λ Λ + 0.75 Λ , n Λ Λ + Λ ) n Λ = 0 , 1 , 2
where r θ represents the local coordinate of the grating bottom region, A N represents the numerical aperture of the objective, and n Λ represents the number of periods. The refractive index of air is approximately 1. This paper specifically focuses on grating depths that are significantly smaller than the depth of field. As a result, the influence of scanning depth on the shadow factor can be disregarded.
The modulation of the coherence signal is influenced by the temporal and spatial coherence, which are consequences of the illumination spectrum and the NA. The modulation factor is expressed as
A δ z = 1 , δ z δ m i n exp δ z δ m i n ,   δ z > δ m i n ,
δ z = m z s c a n .
where δ z is the scan step, z s c a n represents the depth of scan, and m is the depth scan index. A ( δ z ) is used to characterize the modulation effect. With
δ m i n = min δ N A , δ t ,
δ N A = 0.703 λ / σ 2 A N 2 , δ t = 0.703 λ 0 2 / δ λ .
where σ represents the spatial coherence coefficient, δ N A indicates spatial coherence, and δ t indicates coherence length of the illumination spectrum, respectively [17]. According to Abdulhalim’s theory [18], the simplified interference formula within this model from a planar object, denoted as I ( δ z , λ , α ) , represents the integration of the raw interferogram over the depth scan index, incident angles, and illumination spectrum. The raw interferogram does not take diffraction effects into account and is expressed as follows:
I r a w , O R ( x , y ) = Γ I r a w ( x , y ) , x n Λ Λ + 0.5 Λ , n Λ Λ + Λ I r a w ( x , y ) , x n Λ Λ , n Λ Λ + 0.5 Λ ,
with
I r a w ( x , y , δ z ) = λ m i n λ m a x 0 α m a x I δ z + H x , y , λ , α A δ z d α d λ ,
where λ m a x and λ m i n correspond to the longest and shortest wavelengths of the illumination spectrum, respectively.
To calculate the point spread function (PSF), we can utilize the analytical calculation method described by Goodman [19], which involves performing a Fourier transform of the pupil function. In our approach, we use the focal plane PSF to approximate the out-of-focus PSF. For circular pupils, the P S F ( x ,   y ) , after eliminating the constant factor, can be expressed as follows:
P S F ( x , y ) = 2 J 1 u x , y u x , y 2 ,
u x , y = 2 π A N x 2 + y 2 1 / 2 / λ 0 .
In this equation, λ 0 is the central wavelength, x and y represent coordinates on the planar object, and J1 represents the Bessel function of the first kind and order one. As described by Abdulhalim [17], when the grating depth is smaller than the coherence length, the maximum lateral coherence distance between two points in the object plane is δ r c , as clarified in Equation (18). To reduce computational complexity, we focus our calculations on the intensity within the lateral coherence distance. This allows us to concentrate on the partial point spread function P S F p (x, y), which can be formulated as follows:
P S F p x , y = P S F x , y ,   x 2 + y 2 δ r c     0 , x 2 + y 2 > δ r c   ,
δ r c = 1.22 λ 0 / σ A N .
According to Fourier optics, the diffraction effect can be expressed as follows:
I i m a g e x , y = I r a w , O R ( x , y ) P S F p ( x , y ) .
where * represents the convolution operator. I i m a g e (x, y) represents a series of stacked interferograms that consider the diffraction effect. The envelope method is applied to reconstruct the grating profile [16]. Finally, the ‘W/3 rule’ is utilized to calculate the depth measurement of the rectangular grating.
It is worth noting that the boundaries of the simulation model are applicable to gratings with relatively low depths, which should be smaller than the center wavelength.

3. Simulations and Experiments

Figure 4a,b illustrate the batwing width and height values for different periods and depths of the grating. These figures also demonstrate the independence between batwing height and width. Figure 4c illustrates the dominant influence of the width-to-period ratio (WPR) on depth measurement errors, which represents the ratio between the batwing width and the grating period. When the WPR is smaller than 1/6, the impact of the batwing on depth measurement errors can be disregarded. Figure 4d shows the positive correlation between the height-to-depth ratio (HDR), which indicates the ratio between the batwing height and the grating depth, and depth measurement errors when the WPR values are similar and exceed 1/6. Figure 4e,f display the depth measurement errors of rectangular gratings with different depths and periods, assessed using the envelope evaluation method [16]. The depth measurement errors typically decrease as the grating period increases, except for certain instances near the depth values of d = 0.25 λ0 and d = 0.75 λ0. Grating depths smaller than or approximately equal to 0.25 λ0 (and also 0.75 λ0) result in negative depth measurement errors, whereas depths larger than or approximately equal to 0.25 λ0 (and also 0.75 λ0) yield positive errors [20,21,22].
Figure 5a depicts the depth measurement errors with AN = 0.3 λ0 = 0.6 μm. As the NA decreases, the batwing width increases, leading to large measurement errors for small grating periods, especially when the depth is close to λ0. On the other hand, Figure 5b shows the depth measurement errors with AN = 0.55 λ 0 = 0.5 μm. Decreasing the center wavelength leads to a reduction in the batwing width, resulting in smaller depth measurement errors for small grating periods. Moreover, we conducted simulations to examine the influence of altering the illumination spectrum and shadow factor on the depth measurement error, and we observed that these variations closely resemble the patterns depicted in Figure 4e. Therefore, we can conclude that these factors have a minimal influence on the depth measurement error, and therefore, we have not included the corresponding figures in this paper.
The Bruker step standard with a calibrated height of 0.4598 μm was measured using a Mirau objective with AN = 0.55 and LED light sources with λ0 = 0.6 μm and δ λ = 0.08 μm. Figure 6 presents a comparison between simulated and experimental results. In Figure 6a, the black line represents the experimental profile, and the green line represents the simulated profile. In Figure 6b–e, the black lines represent the experimental signals, and the green lines represent the simulated signals. These signals correspond to points A–D in Figure 6a, respectively. We assumed σ to be 0.5. R1 and R2 were constant and assumed to be 1.
Table 1 presents the values for Figure 6a. The simulated and experimental results exhibit good agreement, as shown in Figure 6a. The simulated values for batwing height and width are 0.3068 μm and 0.2530 μm, respectively. In comparison, the experimental values for batwing height and width are 0.3130 μm and 0.4987 μm, respectively.
The difference between simulated and experimental values for batwing width is probably due to the fact that the experimental standard step is quasi-perpendicular. Whether batwings appear or not depends not on the total groove depth but on the change of depth involved in the diffraction region [21]. The simulated and experimental results for profile depth are 0.460 μm and 0.45934 μm, respectively. The relative errors with respect to the nominal value are 0.04% and −0.10%.
The comparison results between simulated and experimental envelope signals are presented in Figure 6b–e, revealing a good match between the two sets of data. In Figure 6c,d, the envelope of the signal exhibits asymmetry, which can be attributed to the presence of diffraction and shadow effect. A shadow effect occurs at the edge of the bottom region because the acceptance angle of the microscope objective is smaller than the maximal incident angle and that of the reflected rays passing through the optical system on the bottom [21]. As shown in Figure 3, the structure of the step standard prevents rays with high incident and reflected angles. And the signal at the edge of the top part is also influenced by the diffraction effect.
The resolution standard RS-N manufactured by Simetrics GmbH (Chemnitz, Germany) provides a set of rectangular gratings with different periods and depths. The periods range from 0.3 to 6 μm and the depths range from 0.1472 to 0.1896 μm. The standard was calibrated by the Metrological Large Range Scanning Probe Microscope (Met LR-AFM) based on the Nano Measuring Machine (NMM, GmbH, Wien, Germany) in the cleanroom centre of PTB (Physikalisch-Technische Bundesanstalt, Braunschweig, Germany) [23].
As stated in ISO 25178-603, when the lateral resolution is limited by the lateral sampling interval of the image sensor system, the smallest period that can be measured should cover five pixels [24]. However, for high magnification systems, the lateral resolution is limited by the optical resolution. For CSI, the smallest measured period is two times of the Rayleigh criterion [25]. Thus, for the CSI Mirau objective with AN = 0.55 and λ0 = 0.6 μm, the smallest period that can be measured is about 1.2 μm. Moreover, the Nyquist sampling theory is satisfied, ensuring accurate data acquisition.
The simulated and experimental profiles in Figure 7 demonstrate strong agreement. To evaluate the similarity between these profiles, the cross-correlation function is utilized. Table 2 presents the computed cross-correlation values for Figure 7a–d, which indicate a decreasing trend as the period decreases. This relationship is further supported by Table 2 and Figure 8e, where the simulated and experimental depths, calculated using the ‘W/3 rule’ for gratings with varying periods, exhibit larger errors compared to the calibrated depth.
Figure 8a,b show the simulated and experimental batwing width and height for different grating periods, revealing consistent agreement. Table 3 presents the computed cross-correlation values for Figure 8a–f. Notably, Figure 8c,d,f illustrate that the WPR values remain relatively unchanged for gratings with similar depths but different periods. As mentioned previously, when the WPR values are similar and exceed 1/6, there is a positive correlation between the depth measurement errors and HDR values. This relationship arises from the fact that the batwing effect occupies a significant portion of the measured profile for small-period gratings, resulting in considerable depth measurement errors [5]. As the grating period decreases, the influence of grating diffraction effects on experimental measurements becomes more significant, leading to substantial depth measurement errors for small-period gratings.

4. Conclusions

Accurately quantifying the impact of the batwing effect on grating depth measurements using CSI is crucial for evaluating measurement results. To accomplish this, we determined the height and width of the batwing based on the characteristics that can affect depth calculations from the topography or profile, while utilizing the ‘W/3 rule’. Moreover, we have developed a comprehensive simulation model that integrates the shadow effect, diffraction effect, temporal coherence, and spatial coherence.
Extensive simulations and experimental investigations have revealed significant influences of the NA and the center wavelength on depth measurements. Conversely, the illumination spectrum and shadow effect exert minimal influence on depth measurements. Significantly, we have established that the batwing effect results in substantial depth measurement errors when the WPR values exceed 1/6. Moreover, we have observed a positive correlation between the depth measurement errors and HDR values when the WPR values are similar and exceed 1/6. However, based on our experience, if the grating period exceeds five times the Rayleigh criterion, the ‘W/3 rule’ can accurately determine the grating depths. The proposed model provides a valuable means of evaluating the accuracy of CSI in measuring the depths of rectangular gratings.
In this study, the Fourier model is utilized. When the grating period exceeds twice the Rayleigh criterion and the height is less than the central wavelength, the disparity between the model simulation and experimental results is minimal. However, as the grating period decreases, the inconsistency between the model simulation and experimental results grows. To address the impact of batwing effects on CSI measurement depths for grating periods less than twice the Rayleigh criterion, it is essential to incorporate the effects of grating diffraction and scattering into the model.

Author Contributions

Conceptualization, J.Z. and Q.Y.; methodology, X.F.; software, L.C.; validation, L.L., Y.F. and Z.X.; data curation, J.M.; writing—original draft preparation, J.Z.; writing—review and editing, Q.Y. and Z.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Natural Science Foundation of China (62175107, 62205148, 62305160); National Key Research and Development Program (2022YFF0706302); Shanghai Natural Science Foundation (21ZR1483100); Program of Shanghai Academic/Technology Research Leader (21XD1425000); China Postdoctoral Science Foundation (No. 2023M731399); Jiangsu Funding Program for Excellent Postdoctoral Talent (No. JB23056); and State Administration for Market Regulation Science and Technology Program (No. 2023MK038).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available upon request from the corresponding author.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Leach, R. Advances in Optical Surface Texture Metrology; IOP Publishing: Philadelphia, PA, USA, 2020. [Google Scholar]
  2. ISO 5436-1: 2000; Geometrical Product Specifications (GPS)—Surface Texture: Profile Method; Measurement Standards—Part 1: Material Measures. International Organization for Standardization: Geneva, Switzerland, 2000.
  3. Sheppard, C.; Carlini, A.; Matthews, H. Three-dimensional imaging of phase steps. Optik 1988, 80, 91–94. [Google Scholar]
  4. Harasaki, A.; Wyant, J.C. Fringe modulation skewing effect in white-light vertical scanning interferometry. Appl. Opt. 2000, 39, 2101–2106. [Google Scholar] [CrossRef] [PubMed]
  5. Hagemeier, S.; Pahl, T.; Breidenbach, J.; Lehmann, P. A novel cubic-exp evaluation algorithm considering non-symmetrical axial response signals of confocal microscopes. Microsc. Res. Tech. 2023, 86, 1012–1022. [Google Scholar] [CrossRef] [PubMed]
  6. de Groot, P.; de Lega, X.C. Signal modeling for low-coherence height-scanning interference microscopy. Appl. Opt. 2004, 43, 4821–4830. [Google Scholar] [CrossRef] [PubMed]
  7. de Groot, P.J.; de Lega, X.C. Fourier optics modeling of interference microscopes. JOSA A 2020, 37, B1–B10. [Google Scholar] [CrossRef] [PubMed]
  8. Xie, W.; Lehmann, P.; Niehues, J. Lateral resolution and transfer characteristics of vertical scanning white-light interferometers. Appl. Opt. 2012, 51, 1795–1803. [Google Scholar] [CrossRef] [PubMed]
  9. Xie, W.; Lehmann, P.; Niehues, J.; Tereschenko, S. Signal modeling in low coherence interference microscopy on example of rectangular grating. Opt. Express 2016, 24, 14283–14300. [Google Scholar] [CrossRef] [PubMed]
  10. Pahl, T.; Hagemeier, S.; Bischoff, J.; Manske, E.; Lehmann, P. Rigorous 3D modeling of confocal microscopy on 2D surface topographies. Meas. Sci. Technol. 2021, 32, 094010. [Google Scholar] [CrossRef]
  11. Su, R.; Thomas, M.; Liu, M.Y.; Drs, J.; Bellouard, Y.; Pruss, C.; Coupland, J.; Leach, R. Lens aberration compensation in interference microscopy. Opt. Lasers Eng. 2020, 128, 9. [Google Scholar] [CrossRef]
  12. Su, R.; Leach, R. Physics-based virtual coherence scanning interferometer for surface measurement. Light Adv. Manuf. 2021, 2, 120–135. [Google Scholar] [CrossRef]
  13. Lehmann, P.; Pahl, T. Three-dimensional transfer function of optical microscopes in reflection mode. J. Microsc. 2021, 284, 45–55. [Google Scholar] [CrossRef] [PubMed]
  14. Li, Y.; Yang, Y.; Liang, Y. Batwing elimination algorithm based on the Haar wavelet in white-light interferometry. Appl. Opt. 2024, 63, 611–616. [Google Scholar] [CrossRef] [PubMed]
  15. Sang, M.; Du, X.; Wang, S.; Xu, T.; Dong, J.; Liu, T. Gap-matching algorithm with the impCEEMDAN in scanning white-light interference microscopy. Opt. Express 2020, 28, 15101–15111. [Google Scholar] [CrossRef] [PubMed]
  16. Sun, Y.; Yu, H.; Ma, J.; Hu, Q.; Huang, X.; Zhou, J.; Xie, P.; Yang, Z.; Lei, L.; Fu, Y. Spurious fringe processing for dielectric metasurface profile measurement using white-light scanning interferometry. Appl. Opt. 2021, 60, 215–223. [Google Scholar] [CrossRef] [PubMed]
  17. Abdulhalim, I. Spatial and temporal coherence effects in interference microscopy and full-field optical coherence tomography. Ann. Phys. 2012, 524, 787–804. [Google Scholar] [CrossRef]
  18. Abdulhalim, I. Theory for double beam interference microscopes with coherence effects and verification using the Linnik microscope. J. Mod. Opt. 2001, 48, 279–302. [Google Scholar] [CrossRef]
  19. Goodman, J.W. Introduction to Fourier Optics; Roberts and Company Publishers: Greenwood Village, CO, USA, 2005. [Google Scholar]
  20. Xie, W. Transfer Characteristics of White Light Interferometers and Confocal Microscopes. Ph.D. Thesis, Universität Kassel, Kassel, Germany, 2017. [Google Scholar]
  21. Lehmann, P.; Tereschenko, S.; Xie, W. Fundamental aspects of resolution and precision in vertical scanning white-light interferometry. Surf. Topogr. Metrol. Prop. 2016, 4, 024004. [Google Scholar] [CrossRef]
  22. Xie, W.; Hagemeier, S.; Woidt, C.; Hillmer, H.; Lehmann, P. Influences of edges and steep slopes in 3D interference and confocal microscopy. In Proceedings of the Optical Micro-and Nanometrology VI, Brussels, Belgium, 26 April 2016; pp. 220–232. [Google Scholar]
  23. Dai, G.; Pohlenz, F.; Danzebrink, H.-U.; Xu, M.; Hasche, K.; Wilkening, G. Metrological large range scanning probe microscope. Rev. Sci. Instrum. 2004, 75, 962–969. [Google Scholar] [CrossRef]
  24. ISO 25178-603: 2013; Geometrical Product Specifications (GPS)—Surface Texture: Areal—Part 603: Nominal Characteristics of Non-Contact (Phase-Shifting Interferometric Microscopy) Instruments. International Organization for Standardization: Geneva, Switzerland, 2013.
  25. Niehues, J.; Lehmann, P. Improvement of lateral resolution and reduction of batwings in vertical scanning white-light interferometry. In Proceedings of the Optical Measurement Systems for Industrial Inspection VII, Munich, Germany, 26 May 2011; pp. 278–285. [Google Scholar]
Figure 1. The definitions of the batwing height and width.
Figure 1. The definitions of the batwing height and width.
Photonics 11 00341 g001
Figure 2. The illustrations of simulation model procedure.
Figure 2. The illustrations of simulation model procedure.
Photonics 11 00341 g002
Figure 3. The illustrations of object cone angle obstructed at the groove edge.
Figure 3. The illustrations of object cone angle obstructed at the groove edge.
Photonics 11 00341 g003
Figure 4. Simulation results with AN = 0.55 λ0 = 0.6 μm: (a) batwing width values; (b) batwing height values; (c) WPR values; (d) HDR values; (e) depth measurement errors (3D view) calculated via envelope evaluation method; (f) depth measurement errors for different height values.
Figure 4. Simulation results with AN = 0.55 λ0 = 0.6 μm: (a) batwing width values; (b) batwing height values; (c) WPR values; (d) HDR values; (e) depth measurement errors (3D view) calculated via envelope evaluation method; (f) depth measurement errors for different height values.
Photonics 11 00341 g004
Figure 5. Depth measurement errors (3D view) calculated via envelope evaluation method: (a) AN = 0.3 λ0 = 0.6 μm; (b) AN = 0.55 λ 0 = 0.5 μm.
Figure 5. Depth measurement errors (3D view) calculated via envelope evaluation method: (a) AN = 0.3 λ0 = 0.6 μm; (b) AN = 0.55 λ 0 = 0.5 μm.
Photonics 11 00341 g005
Figure 6. Comparison of simulated and experimental result with d = 0.4598 μm and λ0 = 0.6 μm. (a) The simulated and experimental profiles. Simulated and experimental envelope signals: (b) far from the edge on the top position for point A; (c) close to the edge on the top position for point B; (d) close to the edge on the bottom position for point C; (e) far from the edge on the bottom position for point D.
Figure 6. Comparison of simulated and experimental result with d = 0.4598 μm and λ0 = 0.6 μm. (a) The simulated and experimental profiles. Simulated and experimental envelope signals: (b) far from the edge on the top position for point A; (c) close to the edge on the top position for point B; (d) close to the edge on the bottom position for point C; (e) far from the edge on the bottom position for point D.
Photonics 11 00341 g006
Figure 7. Comparison of simulated and experimental profiles: (a) Λ = 6 μm, d = 0.1896 μm; (b) Λ = 3 μm, d = 0.1869 μm; (c) Λ = 2 μm, d = 0.1823 μm; (d) Λ = 1.2 μm, d = 0.1738 μm.
Figure 7. Comparison of simulated and experimental profiles: (a) Λ = 6 μm, d = 0.1896 μm; (b) Λ = 3 μm, d = 0.1869 μm; (c) Λ = 2 μm, d = 0.1823 μm; (d) Λ = 1.2 μm, d = 0.1738 μm.
Photonics 11 00341 g007
Figure 8. Comparison of simulated and experimental results: (a) batwing width; (b) batwing height; (c) WPR values; (d) HDR values; (e) depth values; (f) depth errors.
Figure 8. Comparison of simulated and experimental results: (a) batwing width; (b) batwing height; (c) WPR values; (d) HDR values; (e) depth values; (f) depth errors.
Photonics 11 00341 g008
Table 1. Comparison of simulated and experimental batwing width, height, and depth errors for step standard.
Table 1. Comparison of simulated and experimental batwing width, height, and depth errors for step standard.
Simulated Batwing Width/μmExperimental Batwing Width/μmSimulated Batwing Height/μmExperimental Batwing Height/μmSimulated Depth Errors/%Experimental Depth Errors/%
0.25300.31300.30680.49870.04−0.10%
Table 2. Comparison of calibrated depth, simulated depth, experimental depth, and cross-correlation values for different grating periods.
Table 2. Comparison of calibrated depth, simulated depth, experimental depth, and cross-correlation values for different grating periods.
Granting Period/μmCalibrated Depth/μmSimulated Depth/μmExperiment Depth/μmCross-Correlation Values
60.18960.189670.189720.93
30.18690.187960.187780.89
20.18230.184330.184820.79
1.20.17380.179180.180670.72
Table 3. Comparison of simulated and experimental batwing width, height, and depth errors for different grating periods.
Table 3. Comparison of simulated and experimental batwing width, height, and depth errors for different grating periods.
Granting Period/μmSimulated Batwing Width/μmExperimental Batwing Width/μmSimulated Batwing Height/μmExperimental Batwing Height/μmSimulated Depth Errors/%Experimental Depth Errors/%
60.29120.28060.02580.02430.03810.0644
30.28560.28430.02520.02400.5670.470
20.28220.27940.02470.02671.111.38
1.20.26170.26060.01880.02663.103.96
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Zhang, J.; Yuan, Q.; Fan, X.; Chen, L.; Lei, L.; Fu, Y.; Xu, Z.; Ma, J.; Gao, Z. Quantitative Evaluation of the Batwing Effect on Depth Measurement in Coherence Scanning Interferometry for Rectangular Gratings. Photonics 2024, 11, 341. https://doi.org/10.3390/photonics11040341

AMA Style

Zhang J, Yuan Q, Fan X, Chen L, Lei L, Fu Y, Xu Z, Ma J, Gao Z. Quantitative Evaluation of the Batwing Effect on Depth Measurement in Coherence Scanning Interferometry for Rectangular Gratings. Photonics. 2024; 11(4):341. https://doi.org/10.3390/photonics11040341

Chicago/Turabian Style

Zhang, Jiale, Qun Yuan, Xiaoxin Fan, Lu Chen, Lihua Lei, Yunxia Fu, Zhiyi Xu, Jianqiu Ma, and Zhishan Gao. 2024. "Quantitative Evaluation of the Batwing Effect on Depth Measurement in Coherence Scanning Interferometry for Rectangular Gratings" Photonics 11, no. 4: 341. https://doi.org/10.3390/photonics11040341

APA Style

Zhang, J., Yuan, Q., Fan, X., Chen, L., Lei, L., Fu, Y., Xu, Z., Ma, J., & Gao, Z. (2024). Quantitative Evaluation of the Batwing Effect on Depth Measurement in Coherence Scanning Interferometry for Rectangular Gratings. Photonics, 11(4), 341. https://doi.org/10.3390/photonics11040341

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop