Review

Metallurgical Tests in Endodontics: A Narrative Review

Department of Oral and Maxillofacial Sciences, La Sapienza University of Rome, 00161 Rome, Italy
*
Author to whom correspondence should be addressed.
Bioengineering 2022, 9(1), 30; https://doi.org/10.3390/bioengineering9010030
Submission received: 19 December 2021 / Revised: 7 January 2022 / Accepted: 11 January 2022 / Published: 12 January 2022

Abstract

:
Since there are no reviews of the literature on this theme, the aim of this narrative review is to summarize the metallurgical tests used in endodontics, pointing out their functional use and their pros and cons and giving readers a user-friendly guide to serve as an orientation aid in the plethora of metallurgical tests. With this purpose, a literature search for articles published between January 2001 and December 2021 was conducted, using the electronic database PubMed to collect all published articles regarding the metallurgical tests used in endodontics for the evaluation of NiTi rotary instruments. The search was conducted using the following keywords: “metallurgy”, “differential scanning calorimetry” (DSC), “X-ray diffraction” (XRD), “atomic force microscopy” (AFM), “energy-dispersive X-ray spectroscopy” (EDS), “focused ion beam analysis” (FIB) and “Auger electron spectroscopy” (AES) combined with the term “endodontics” or “NiTi rotary instruments”. Considering the inclusion and exclusion criteria, of the 248 articles found, only 81 were included in the narrative review. According to the results, more than 50% of the selected articles were published in one of the two most relevant journals in endodontics: International Endodontic Journal (22.2%) and Journal of Endodontics (29.6%). The most popular metallurgical test was DSC, with 43 related articles, followed by EDS (33 articles), AFM (22 articles) and XRD (21 articles). Few studies were conducted using other tests such as FIB (2 articles), micro-Raman spectroscopy (4 articles), metallographic analysis (7 articles) and Auger electron spectroscopy (2 articles).

1. Introduction

It could be stated that the modern era of endodontics began after the introduction of the nickel–titanium (NiTi) alloy as the material of choice for the manufacturing of endodontic instruments [1]. From that moment, the way of conceiving endodontics changed considerably, although its basic principles have remained the same and could be mostly summarized in three key processes: chemo-mechanically disinfecting the root canal system to achieve a reduction in bacteria as far as possible; obtaining a stable root canal filling to entomb residual bacteria and to isolate the endodontic system from the periapical tissues; obtaining a stable coronal restoration to avoid secondary infections of the root canal system [2,3].
The widespread use of NiTi rotary instruments, which have almost completely replaced stainless-steel (SS) manual instruments, fundamentally arose from the two most characteristic features of NiTi alloy: superelasticity and the shape memory effect. The former is defined as the ability of the alloy to store stress up to 8% without being plastically deformed, remaining in the elastic region of deformation by the creation of a stress-induced phase, called stress-induced martensite (SIM) [4,5]. The latter allows NiTi alloy to “memorize” a pre-imposed form and to return to it on heating, as a consequence of the transition from the martensitic crystallographic phase to the austenitic phase [6,7].
In the light of the above, NiTi rotary instruments have allowed the simplification of shaping procedures, increasing the predictability and the effectiveness of endodontic treatments due to the possibility of manufacturing endodontic instruments with a greater taper and using them at higher speed without compromising their flexibility and their mechanical behavior, in a way that is not possible for SS instruments [8,9,10,11].
To date, the mechanical properties of NiTi instruments have been investigated in detail and summarized in several reviews of the literature [5,7,12], most recently in the review by Zanza et al. [13]. This review emphasized the mechanical behavior of NiTi instruments during instrumentation procedures, highlighting the current knowledge regarding cyclic fatigue and torsional resistance, flexibility, centering and shaping ability and canal transportation of endodontic instruments and criticizing the most common testing devices and their limitations. As pointed out by the authors, the main drawback of the published articles was the limited analysis performed to compare NiTi systems. In most cases, in fact, the authors performed only static tests, failing to consider the dynamic behavior of NiTi instruments and drawing misleading conclusions [13]. Moreover, in the majority of the published articles, the mechanical evaluation of NiTi instruments was not accompanied by metallurgical tests, giving only a partial understanding of the topic. For this reason, Silva et al. proposed a multimethod approach for the evaluation of different instruments, in which not only the static behavior but also the dynamicity and the metallurgical features of endodontic instruments were investigated, guaranteeing more relevant results with clinical significance [14]. Moreover, the metallurgical investigation of NiTi rotary instruments is fundamental for an in-depth comprehension of their mechanical properties, and it allows clinicians, researchers and engineers to understand the chemo-physical rationale behind certain mechanical behaviors of endodontic instruments, improving knowledge on this theme. The endodontic literature is full of mechanical comparisons between different instruments [13], with authors trying to determine which is the best-performing instrument in terms of cyclic fatigue resistance, torsional resistance, shaping ability, centering ability and cutting efficiency; however, as demonstrated in the previously mentioned reviews, these results are in most cases isolated from an in-depth investigation of the metallurgical properties of the instruments, giving only a partial view of the results of mechanical testing of instruments and their performance.
Despite the importance of metallurgical tests in endodontics, to date there are no reviews of the literature on this topic, despite the fact that in recent years it has increasingly captured the attention of researchers.
Therefore, the aim of this narrative review was to summarize the metallurgical tests used in endodontics, pointing out their functional use, their pros and cons and the future perspectives, in order to offer readers a user-friendly guide to navigating the topic of endodontic metallurgy with ease, since this field may be tricky to understand for endodontists. Moreover, this review, in addition to the review published by Zanza et al. in a previous issue of this journal, aims to complete the general overview of the current tests used in endodontics, in terms of both the mechanical and the metallurgical behavior of NiTi rotary instruments [13].

2. Materials and Methods

An online search was conducted in the peer-review journals listed in PubMed to retrieve in vitro and laboratory studies regarding investigations into the metallurgical characteristics of NiTi endodontic rotary instruments, using the following search query: (“metallurgic” [All Fields] OR “metallurgical” [All Fields] OR “metallurgically” [All Fields]) AND (“endodontal” [All Fields] OR “endodontic” [All Fields] OR “endodontical” [All Fields] OR “endodontically” [All Fields] OR “endodontics” [MeSH Terms] OR “endodontics” [All Fields]) AND 2001/01/01:3000/12/31 [Date—Publication].
Then, a specific search was performed for each metallurgical test evidenced in the above-mentioned search (i.e., differential scanning calorimetry (DSC), X-ray diffraction (XRD), atomic force microscopy (AFM), energy-dispersive X-ray spectroscopy (EDS or EDX), focused ion beam analysis (FIB) and Auger electron spectroscopy (AES)), respectively, using the following search queries: (“calorimetry, differential scanning” [MeSH Terms] OR (“calorimetry” [All Fields] AND “differential” [All Fields] AND “scanning” [All Fields]) OR “differential scanning calorimetry” [All Fields] OR (“differential” [All Fields] AND “scanning” [All Fields] AND “calorimetry” [All Fields])) AND ((“nitinol” [Supplementary Concept] OR “nitinol” [All Fields] OR “nickel titanium” [All Fields]) AND (“rotaries” [All Fields] OR “rotary” [All Fields]) AND (“instrument” [All Fields] OR “instrument s” [All Fields] OR “instrumentation” [MeSH Subheading] OR “instrumentation” [All Fields] OR “instruments” [All Fields] OR “instrumented” [All Fields] OR “instrumenting” [All Fields])) AND 2001/01/01:3000/12/31 [Date—Publication]; (“x ray diffraction” [MeSH Terms] OR (“x ray” [All Fields] AND “diffraction” [All Fields]) OR “x ray diffraction” [All Fields] OR “x ray diffraction” [All Fields]) AND ((“nitinol” [Supplementary Concept] OR “nitinol” [All Fields] OR “nickel titanium” [All Fields]) AND (“rotaries” [All Fields] OR “rotary” [All Fields]) AND (“instrument” [All Fields] OR “instrument s” [All Fields] OR “instrumentation” [MeSH Subheading] OR “instrumentation” [All Fields] OR “instruments” [All Fields] OR “instrumented” [All Fields] OR “instrumenting” [All Fields])) AND 2001/01/01:3000/12/31 [Date—Publication]; (“spectrometry, x ray emission” [MeSH Terms] OR (“spectrometry” [All Fields] AND “x ray” [All Fields] AND “emission” [All Fields]) OR “x-ray emission spectrometry” [All Fields] OR “energy dispersive x ray spectroscopy” [All Fields]) AND ((“nitinol” [Supplementary Concept] OR “nitinol” [All Fields] OR “nickel titanium” [All Fields]) AND (“rotaries” [All Fields] OR “rotary” [All Fields]) AND (“instrument” [All Fields] OR “instrument s” [All Fields] OR “instrumentation” [MeSH Subheading] OR “instrumentation” [All Fields] OR “instruments” [All Fields] OR “instrumented” [All Fields] OR “instrumenting” [All Fields])) AND 2001/01/01:3000/12/31 [Date—Publication]; (“accommodation, ocular” [MeSH Terms] OR (“accommodation” [All Fields] AND “ocular” [All Fields]) OR “ocular accommodation” [All Fields] OR “focusing” [All Fields] OR “focused” [All Fields] OR “focuses” [All Fields]) AND (“ions” [MeSH Terms] OR “ions” [All Fields] OR “ion” [All Fields]) AND “beam” [All Fields] AND (“endodontal” [All Fields] OR “endodontic” [All Fields] OR “endodontical” [All Fields] OR “endodontically” [All Fields] OR “endodontics” [MeSH Terms] OR “endodontics” [All Fields]) AND 2001/01/01:3000/12/31 [Date—Publication]; (“auger” [All Fields] OR “augers” [All Fields]) AND (“electron s” [All Fields] OR “electrone” [All Fields] OR “electrons” [MeSH Terms] OR “electrons” [All Fields] OR “electron” [All Fields]) AND (“spectroscopies” [All Fields] OR “spectroscopy s” [All Fields] OR “spectrum analysis” [MeSH Terms] OR (“spectrum” [All Fields] AND “analysis” [All Fields]) OR “spectrum analysis” [All Fields] OR “spectroscopy” [All Fields]) AND (“endodontal” [All Fields] OR “endodontic” [All Fields] OR “endodontical” [All Fields] OR “endodontically” [All Fields] OR “endodontics” [MeSH Terms] OR “endodontics” [All Fields]) AND 2001/01/01:3000/12/31 [Date—Publication]; (“photoelectron spectroscopy” [MeSH Terms] OR (“photoelectron” [All Fields] AND “spectroscopy” [All Fields]) OR “photoelectron spectroscopy” [All Fields] OR “x ray photoelectron spectroscopy” [All Fields]) AND (“endodontal” [All Fields] OR “endodontic” [All Fields] OR “endodontical” [All Fields] OR “endodontically” [All Fields] OR “endodontics” [MeSH Terms] OR “endodontics” [All Fields]) AND 2001/01/01:3000/12/31 [Date—Publication]; (“microscopy, atomic force” [MeSH Terms] OR (“microscopy” [All Fields] AND “atomic” [All Fields] AND “force” [All Fields]) OR “atomic force microscopy” [All Fields] OR (“atomic” [All Fields] AND “force” [All Fields] AND “microscopy” [All Fields])) AND (“endodontal” [All Fields] OR “endodontic” [All Fields] OR “endodontical” [All Fields] OR “endodontically” [All Fields] OR “endodontics” [MeSH Terms] OR “endodontics” [All Fields]) AND 2001/01/01:3000/12/31 [Date—Publication].

2.1. Inclusion and Exclusion Criteria

The inclusion criteria for this review were articles from peer-reviewed journals indexed in PubMed and written in English from January 2001 to December 2021 that reported at least one in vitro or laboratory test regarding the investigation of metallurgical characteristics of endodontic instruments. Thus, studies that did not meet the above inclusion criteria, such as articles not written in English, published before January 2001 or not reporting a metallurgical test of NiTi rotary endodontic instruments were excluded. Moreover, all articles in which the metallurgical tests were performed on a wire blank and not on manufactured endodontic instruments were also excluded, as well as articles on SS manual instruments.

2.2. Search Methodology

The titles and abstracts of all articles identified from the electronic searches in PubMed were examined by two authors (A.Z. and R.R.) in order to eliminate, in the first instance, articles that clearly failed to meet the inclusion criteria. Full-text copies of all remaining articles were further examined independently by both authors to establish whether the inclusion criteria were met. The investigators met and reviewed the remaining list of articles.

3. Results

The final list of articles generated after electronic searching included 248 studies. After the first screening and the discarding of duplications, 137 of these articles were obtained for full-text review. After full-text review, a total of 81 articles were included (Table 1).
Of those 81 articles, 45 articles were published in the two most relevant journals regarding endodontics: International Endodontic Journal with 19 articles (22.2%) and Journal of Endodontics with 26 articles (29.6%) (Figure 1).
According to the collected articles, the most popular metallurgical investigation was DSC, followed by EDS, AFM, XRD, metallurgical analysis, micro-Raman spectroscopy, FIB, AES and electrochemical potential measurements (Figure 2).

4. Discussion

In the last decades, NiTi rotary instruments have become the most widespread endodontic instruments, increasing their use over SS manual instruments. This is mainly due to their undisputable advantages in comparison to SS instruments, i.e., the superelastic behavior and the shape memory effect of NiTi alloys that allow the manufacturing of endodontic instruments with a greater taper, cutting efficiency and performance. All these characteristics guarantee the simplification of instrumentation procedures, an increase in the predictability and effectiveness of endodontic treatments and a reduction in the time of endodontic treatments [5,7,12,13,95,96,97,98,99]. Regarding the increased performance of NiTi rotary instruments and their mechanical behavior, several studies have been published and summarized in many literature reviews [5,7,12,13]. Nevertheless, to date, there are no literature reviews regarding the metallurgy of NiTi rotary instruments and the metallurgical tests used in endodontics. For this reason, the aim of this narrative review was to summarize all the tests used in endodontics and to investigate the properties of NiTi alloy, giving readers and researchers an overview of this theme.
The metallurgy of endodontic instruments is fundamental for an in-depth understanding of their mechanical behavior, for the evaluation of recently introduced instruments and for comparisons between them. Unsurprisingly, of the 81 selected articles, more than 50% were published in one of the two most relevant journals in endodontics (JOE and IEJ), demonstrating the importance of metallurgical tests in addition to mechanical tests when a comparison between instruments is performed. Therefore, to the best of our knowledge, there is no equal correspondence between articles related to the comparison between different instruments and articles in which metallurgical tests were performed. In fact, performing a literature search in PubMed regarding articles on mechanical evaluation of NiTi rotary instruments using the keywords “(cyclic fatigue OR shaping ability OR torsional resistance OR flexibility OR centering ability OR cutting efficiency) AND endodontic rotary instruments” led to a total of about 670 articles being found. Comparing these results with the number of articles regarding the metallurgical investigation of NiTi rotary instruments, it is clear that in most cases the comparison between endodontic instruments was performed using only mechanical evaluation and not considering an investigation of their metallurgy. This could lead to a partial and limited view of the topic and in some cases to misleading conclusions. In our opinion, in order to improve the quality of the research, one or more metallurgical tests should always be performed when researchers intend to compare different instruments, thus performing multimethod research that investigates the whole topic. Probably, the lack of articles on the metallurgy of endodontic rotary instruments is due to two main factors: the high cost of the devices used for these types of tests and the need for the presence of specialized technicians in these investigations. Moreover, endodontic researchers may have difficulties in choosing the most appropriate test for their aims, since metallurgy is not their field of expertise. Hence, the aim of this narrative review was to summarize the metallurgical tests used in endodontics and to offer the readers a user-friendly guide to navigating the topic of endodontic metallurgy with ease, helping researchers to choose the most appropriate test for their scope.
The following is a list of all the metallurgical tests performed for the evaluation of NiTi rotary instruments from January 2001 to December 2021.

4.1. Differential Scanning Calorimetry

The most commonly used metallurgical test in endodontics was undoubtedly DSC. The pioneering research that introduced it for the first time in the endodontic field was conducted by Brantley et al. in 2002 [91]. DSC, as stated by the authors, was used to evaluate the structure of the NiTi alloys via accurate measurement of the difference in thermal power supplied to a test specimen and an inert control specimen heated at the same rate. More precisely, both specimens were cooled to temperatures ranging from −70 °C to −130 °C and subsequently underwent a heating/cooling cycle in the range of about −130 °C to 110 °C, with an isotherm step at 100–110 °C for one minute. Structural transformations and transition temperatures in the NiTi alloys were revealed as endothermic peaks on the heating DSC curves and as exothermic peaks on the cooling DSC curves. Moreover, DSC provides the enthalpy change during transformation and the temperature ranges of the transformation between the three crystallographic phases: austenite, R-phase and martensite [91]. One of the key phases of DSC analysis is the specimen preparation. The pans (aluminum or SS) are not able to contain the whole length of instruments; thus, they must be cut into parts and then placed in the pans. The cutting must be performed under a constant water-cooling jet and with a slow-speed diamond bur to minimize mechanical stresses that might change the crystallographic phase of the NiTi instruments [84].
DSC was mainly used to determine the enthalpy changes and transition temperature (martensitic transformation starting point (Ms), martensitic transformation finishing point (Mf), reverse transformation starting point (As) and reverse transformation finishing point (Af)); however, it was also used to investigate the modification of the NiTi alloy after clinical use. The results clearly showed that the enthalpy changes of as-received instruments were higher than those of instruments with simulated clinical use, showing that instruments used during instrumentation procedures are subjected to considerable stresses and are significantly changed in microstructure [18,90]. Moreover, the stored stress after instrumentation significantly modifies the transition temperatures, reducing the Af and Mf temperatures and increasing the Ms temperature [65,90].

4.2. Energy-Dispersive X-ray Spectroscopy

X-ray energy-dispersive spectroscopy (EDS) is an analytical technique used for the elemental analysis or chemical characterization of a sample and was used to identify and characterize semi-quantitatively the chemical elements of the NiTi alloy and the presence of deposits on the instrument surfaces. Therefore, EDS was used to determine the near-equiatomic composition of NiTi instruments and to detect any other materials present as inclusions in the alloy or as adherent deposits on the surface [76,93].
In the last decades, EDS was mainly used to determine the importance of differences in terms of percentages of Ni and Ti in the alloy used for the manufacture of NiTi rotary instruments; however, its relevance is still dubious. Alapati et al. reported that differences in the wire blanks in terms of percentages of Ni and Ti, in addition to the machining procedures, can alter the metallurgical characteristics of NiTi instruments and can influence their mechanical performance [75]. However, the potential role of different weight percentages of nickel still remains uncertain, since there are no studies that evaluate the influence of this factor on the mechanical properties of the NiTi instrument while maintaining unchanged all other factors such as instrument design and heat treatment. Thus, it is difficult to determine whether there is a crucial effect of a single factor (Ni and Ti percentage) or as is more likely, a combination of different correlated factors, when determining the final mechanical properties. Furthermore, Zinelis et al. reported that the thermomechanical history, as well as the instrument design, has a much more crucial effect on the final mechanical behavior in comparison to the compositional deviations of NiTi instruments [73].
EDS analysis has also been used to investigate the composition of the as-received NiTi instrument surface and its variation after clinical use, cycles of sterilization or immersion in endodontic solutions. Several studies demonstrated that the manufacturing procedures of NiTi instruments, such as machining operations and heating of the alloy, lead to the formation of irregularities and cause the presence of adherent material on the instrument’s surface, to which dentine adheres after root canal shaping [4,76,93]. For this reason, particular attention during cleaning procedures before sterilization or the single use of endodontic instruments is necessary to reduce the possibility of cross contamination, since the adherence of dentine to the deposits may prevent appropriate sterilization of NiTi instruments and enhance the risk of cross infection between patients [59,93].
Moreover, Spagnuolo et al. evaluated the effect of autoclaving cycles on the chemical composition of NiTi alloy, stating that the percentage of nickel and titanium was affected by these procedures, with a decrease in nickel and titanium for ProTaper Universal and a decrease in nitrogen and titanium with an increase in nickel for AlphaKyte [67]. As mentioned before, the immersion of NiTi instruments in endodontic solutions could also alter their surface composition, leaving sodium and chlorine elements on the surface, with a consequent decrease in the weight percentage of nickel and titanium and a corresponding increase in the weight percentage of sodium and chlorine elements [58]. However, to date, there are no data on the influence of these surface changes on the mechanical behavior of endodontic instruments. Thus, further studies are required.
Hence, DSC, EDS and XRD should be considered as important types of analysis in completely understanding the different mechanical behaviors of NiTi rotary instruments.

4.3. Atomic Force Microscopy

Atomic force microscopy was introduced for the first time as a tool for the evaluation of NiTi instrument surfaces by Valois et al. [86]. Its advantages are fundamentally its ease of use, low cost and molecular-level resolution of structural detail, providing three-dimensional images with high spatial resolution, as demonstrated by several studies using AFM [100,101]. Qualitative and quantitative information on the NiTi instrument’s topography are provided via the detection of several forces between the probing tip used in this analysis and the sample [86]. For this reason, AFM could be considered as part of the scanning probe microscopy family, with the ability to reconstruct the surfaces of NiTi instruments three-dimensionally (as sets of x, y and z values). These sets are then analyzed with dedicated digital software to give all the data pertaining to the examined surface in quantitative form, using vertical topographic parameters [79,100,102]. The most widely used evaluation parameters for describing the topographic characteristics of a surface are the arithmetic mean roughness (AMR), maximum height (MH) and root mean square (RMS) [79,82,86].
AFM was used to evaluate the surfaces irregularities of as-received endodontic instruments and their changes after clinical use. Valois et al. stated that all tested instruments showed surface irregularities arising from the manufacturing procedure [86], and this has been confirmed by other articles published in subsequent years [45,63]. Moreover, it was stated that root canal shaping and the consequent multiple autoclave cycles can cause the deformation of NiTi instrument surfaces, increasing their irregularities and thus their roughness [42,79,82].
Another potential area of application of AFM was the determination of surface irregularities of NiTi instruments caused by their immersion in irrigating solutions such as sodium hypochlorite (NaOCl) and ethylenediaminetetraacetic acid (EDTA). Ametrano et al., Saglam et al., Fayyad et al. and Uslu et al. reported that the immersion of NiTi instruments in NaOCl caused a slight but significant degradation of the surface, increasing their roughness. Although with different intensities, even EDTA and chlorhexidine affect the instrument’s surface regularity [16,41,64,68,71].

4.4. X-ray Diffraction

X-ray diffraction (XRD) is one of the most valuable tools for the determination of the crystallographic structure of materials, in particular for NiTi alloy. XRD and DSC are able to provide a complete description of the NiTi alloy, both in terms of crystallographic organization (austenite, martensite and R-phase) and transition temperatures between each phase. Khun et al. explained in detail the operation of XRD analysis as follows:
“Planes (hkl) of atoms constructively interfere with X-rays and diffraction occurs: Bragg’s law, λ = 2d sinθ, allows the calculation of interplanar spacings or d-spacings from the angular location of XRD peaks, θ (degree). Comparison of XRD data to known standards is used to identify phases. Miller indices, hkl integers, are assigned to XRD peaks. Miller indices describe the orientation of planes of atoms to the unit cell of a material’s crystal structure”
[94]
Thus, XRD is mainly used to evaluate the crystallographic phase organization of the NiTi alloy of endodontic instruments and to assess the influence of heat treatments on the constitution of the phases in the instruments. However, the main disadvantage of this test is the impossibility of establishing the amount of each phase in the samples with sufficient precision. In other words, is not possible to know the exact percentage of austenite, martensite and R-phase in NiTi rotary instruments; only the most represented crystallographic organization of NiTi [33].

4.5. Metallographic Analysis

A metallographic analysis is usually performed using scanning electron microscopy to identify the grains of austenite and martensite in the NiTi alloy. Thus, this kind of test is fundamentally used to confirm and complete the results of XRD and EDS and to detect any inclusions in the alloy. In order to disclose the microstructure of the NiTi matrix, samples must be embedded in epoxy resin, ground, polished and then etched with 60% nitric acid, 10% fluorhydric acid and 30% acetic acid at room temperature for 5 s [103]. However, Generali et al. proposed a different method for the preparation of the specimen for metallographic analysis in order to reduce the artefacts that might have arisen from metallographic etching with hydrofluoric-acid-based compounds, that consisted of vibro-polishing and plasma-cleaning of the resin-mounted specimen instead of the etching procedure [28].

4.6. Micro-Raman Spectroscopy, Focused Ion Beam Analysis and Auger Electron Spectroscopy

Micro-Raman spectroscopy and focused ion beam analysis were methods used to investigate the surface of NiTi instruments to assess the presence of surface layers or coatings. In the endodontic literature, there were only four studies that used micro-Raman spectroscopy and two studies that used FIB analysis in their methodology, and all of these were published by the same research group [22,26,28,44].
Auger electron spectroscopy is a through-thickness method that is based on ion bombardment. It is able to evaluate the chemical composition of the alloy in the surface layers, up to a depth of a few micrometers, in contrast to FIB/SEM and micro-Raman analyses [26].

5. Conclusions

Metallurgical tests are fundamental for thoroughly understanding the mechanical behavior of NiTi rotary instruments, particularly when the aim of the study is to compare the performance of different instruments. Moreover, they allow clinicians, researchers and engineers to understand the chemo-physical rationale behind certain mechanical behaviors of endodontic instruments, giving further information for developing knowledge in this field.

Author Contributions

Conceptualization, A.Z. and R.R.; methodology, D.D.N.; software, M.S.; validation, G.M.; formal analysis, L.T.; investigation, A.Z.; resources, L.T.; data curation, D.D.N.; writing—original draft preparation, A.Z. and R.R.; writing—review and editing, M.S. and D.D.N.; visualization, G.M.; supervision, L.T. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Walia, H.M.; Brantley, W.A.; Gerstein, H. An initial investigation of the bending and torsional properties of Nitinol root canal files. J. Endod. 1988, 14, 346–351. [Google Scholar] [CrossRef]
  2. Ng, Y.L.; Mann, V.; Rahbaran, S.; Lewsey, J.; Gulabivala, K. Outcome of primary root canal treatment: Systematic review of the literature—Part 1. Effects of study characteristics on probability of success. Int. Endod. J. 2007, 40, 921–939. [Google Scholar] [CrossRef]
  3. Ng, Y.L.; Mann, V.; Rahbaran, S.; Lewsey, J.; Gulabivala, K. Outcome of primary root canal treatment: Systematic review of the literature—Part 2. Influence of clinical factors. Int. Endod. J. 2008, 41, 6–31. [Google Scholar] [CrossRef] [PubMed]
  4. Thompson, S.A. An overview of nickel-titanium alloys used in dentistry. Int. Endod. J. 2000, 33, 297–310. [Google Scholar] [CrossRef] [Green Version]
  5. Tabassum, S.; Zafar, K.; Umer, F. Nickel-Titanium Rotary File Systems: What’s New? Eur. Endod. J. 2019, 4, 111–117. [Google Scholar] [CrossRef] [PubMed]
  6. Ounsi, H.F.; Nassif, W.; Grandini, S.; Salameh, Z.; Neelakantan, P.; Anil, S. Evolution of Nickel-titanium Alloys in Endodontics. J. Contemp. Dent. Pract. 2017, 18, 1090–1096. [Google Scholar] [CrossRef]
  7. Zupanc, J.; Vahdat-Pajouh, N.; Schäfer, E. New thermomechanically treated NiTi alloys—A review. Int. Endod. J. 2018, 51, 1088–1103. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  8. Sadeghi, S. Shaping ability of NiTi rotary versus stainless steel hand instruments in simulated curved canals. Med. Oral Patol. Oral Cir. Bucal 2011, 16, e454–e458. [Google Scholar] [CrossRef]
  9. Htun, P.H.; Ebihara, A.; Maki, K.; Kimura, S.; Nishijo, M.; Tokita, D.; Okiji, T. Comparison of torque, force generation and canal shaping ability between manual and nickel-titanium glide path instruments in rotary and optimum glide path motion. Odontology 2020, 108, 188–193. [Google Scholar] [CrossRef]
  10. Cheung, G.S.; Liu, C.S. A retrospective study of endodontic treatment outcome between nickel-titanium rotary and stainless steel hand filing techniques. J. Endod. 2009, 35, 938–943. [Google Scholar] [CrossRef] [PubMed]
  11. Del Fabbro, M.; Afrashtehfar, K.I.; Corbella, S.; El-Kabbaney, A.; Perondi, I.; Taschieri, S. In Vivo and In Vitro Effectiveness of Rotary Nickel-Titanium vs. Manual Stainless Steel Instruments for Root Canal Therapy: Systematic Review and Meta-analysis. J. Evid. Based Dent. Pract. 2018, 18, 59–69. [Google Scholar] [CrossRef]
  12. Gavini, G.; Santos, M.D.; Caldeira, C.L.; Machado, M.E.L.; Freire, L.G.; Iglecias, E.F.; Peters, O.A.; Candeiro, G.T.M. Nickel-titanium instruments in endodontics: A concise review of the state of the art. Braz. Oral Res. 2018, 32, e67. [Google Scholar] [CrossRef]
  13. Zanza, A.; D’Angelo, M.; Reda, R.; Gambarini, G.; Testarelli, L.; Di Nardo, D. An Update on Nickel-Titanium Rotary Instruments in Endodontics: Mechanical Characteristics, Testing and Future Perspective—An Overview. Bioengineering 2021, 8, 218. [Google Scholar] [CrossRef]
  14. Silva, E.; Martins, J.N.R.; Lima, C.O.; Vieira, V.T.L.; Braz Fernandes, F.M.; De-Deus, G.; Versiani, M.A. Mechanical Tests, Metallurgical Characterization, and Shaping Ability of Nickel-Titanium Rotary Instruments: A Multimethod Research. J. Endod. 2020, 46, 1485–1494. [Google Scholar] [CrossRef] [PubMed]
  15. Martins, J.N.R.; Silva, E.; Marques, D.; Belladonna, F.; Simões-Carvalho, M.; Camacho, E.; Braz Fernandes, F.M.; Versiani, M.A. Comparison of design, metallurgy, mechanical performance and shaping ability of replica-like and counterfeit instruments of the ProTaper Next system. Int. Endod. J. 2021, 54, 780–792. [Google Scholar] [CrossRef] [PubMed]
  16. Jose, J.; Khandelwal, A.; Siddique, R. Qualitative Assessment of the Surface Topographic Changes of XP-endo Shaper and TruNatomy files after exposure to Sodium Hypochlorite and Ethylenediaminetetraacetic Acid. Eur. Endod. J. 2021, 6, 197–204. [Google Scholar] [CrossRef] [PubMed]
  17. Jo, H.J.; Kwak, S.W.; Kim, H.C.; Kim, S.K.; Ha, J.H. Torsional Resistance of Heat-Treated Nickel-Titanium Instruments under Different Temperature Conditions. Materials 2021, 14, 5295. [Google Scholar] [CrossRef] [PubMed]
  18. Van Pham, K. Differential scanning calorimetric investigations of three rotary nickel-titanium instrument systems before and after simulated clinical uses. BMC Oral Health 2021, 21, 488. [Google Scholar] [CrossRef]
  19. Martins, J.N.R.; Silva, E.; Marques, D.; Belladonna, F.G.; Simões-Carvalho, M.; da Costa, R.P.; Ginjeira, A.; Braz Fernandes, F.M.; Versiani, M.A. Comparison of five rotary systems regarding design, metallurgy, mechanical performance, and canal preparation—A multimethod research. Clin. Oral Investig. 2021, 1–12. [Google Scholar] [CrossRef]
  20. Kalyoncuoğlu, E.; Keskin, C.; Keleş, A.; Aydemir, H. Metallurgical characterisation and torsional resistance of blue thermomechanically treated nickel titanium instruments after simulated ex vivo retreatment procedure. Aust. Endod. J. 2021. [Google Scholar] [CrossRef]
  21. Martins, J.N.R.; Silva, E.; Marques, D.; Belladonna, F.; Simões-Carvalho, M.; Vieira, V.T.L.; Antunes, H.S.; Braz Fernandes, F.M.B.; Versiani, M.A. Design, metallurgical features, mechanical performance and canal preparation of six reciprocating instruments. Int. Endod. J. 2021, 54, 1623–1637. [Google Scholar] [CrossRef]
  22. Azizi, A.; Prati, C.; Schiavon, R.; Fitzgibbon, R.M.; Pirani, C.; Iacono, F.; Pelliccioni, G.A.; Spinelli, A.; Zamparini, F.; Puddu, P.; et al. In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments. Eur. Endod. J. 2021, 6, 90–97. [Google Scholar] [CrossRef] [PubMed]
  23. Keskin, C.; Sivas Yilmaz, Ö.; Keleş, A.; Inan, U. Comparison of cyclic fatigue resistance of Rotate instrument with reciprocating and continuous rotary nickel-titanium instruments at body temperature in relation to their transformation temperatures. Clin. Oral Investig. 2021, 25, 151–157. [Google Scholar] [CrossRef]
  24. Martins, J.N.R.; Silva, E.; Marques, D.; Pereira, M.R.; Ginjeira, A.; Silva, R.J.C.; Braz Fernandes, F.M.; Versiani, M.A. Mechanical Performance and Metallurgical Features of ProTaper Universal and 6 Replicalike Systems. J. Endod. 2020, 46, 1884–1893. [Google Scholar] [CrossRef] [PubMed]
  25. Martins, J.N.R.; Nogueira Leal Silva, E.J.; Marques, D.; Ginjeira, A.; Braz Fernandes, F.M.; De Deus, G.; Versiani, M.A. Influence of Kinematics on the Cyclic Fatigue Resistance of Replicalike and Original Brand Rotary Instruments. J. Endod. 2020, 46, 1136–1143. [Google Scholar] [CrossRef] [PubMed]
  26. Generali, L.; Malovo, A.; Bolelli, G.; Borghi, A.; La Rosa, G.R.M.; Puddu, P.; Lusvarghi, L.; Rota, A.; Consolo, U.; Pedullà, E. Mechanical Properties and Metallurgical Features of New Green NiTi Reciprocating Instruments. Materials 2020, 13, 3736. [Google Scholar] [CrossRef] [PubMed]
  27. Weyh, D.J.; Ray, J.J. Cyclic fatigue resistance and metallurgic comparison of rotary endodontic file systems. Gen. Dent. 2020, 68, 36–39. [Google Scholar]
  28. Generali, L.; Puddu, P.; Borghi, A.; Brancolini, S.; Lusvarghi, L.; Bolelli, G.; Consolo, U.; Pedullà, E. Mechanical properties and metallurgical features of new and ex vivo used Reciproc Blue and Reciproc. Int. Endod. J. 2020, 53, 250–264. [Google Scholar] [CrossRef]
  29. Arias, A.; Macorra, J.C.; Govindjee, S.; Peters, O.A. Effect of gamma-ray sterilization on phase transformation behavior and fatigue resistance of contemporary nickel-titanium instruments. Clin. Oral Investig. 2020, 24, 3113–3120. [Google Scholar] [CrossRef]
  30. Khabadze, Z.; Mordanov, O.; Balashova, M.; Stolov, L.; Pangratyan, A.; Bokova, R.; Nazhmudinov, S.; Solimanov, S.; Kuznetsova, A.; Adzhieva, A.; et al. Laboratory Rational of Changes in the Crystal Lattice of Nickel-Titanium Endodontic Rotary Files in Autoclaving. Int. J. Dent. 2020, 2020, 8386215. [Google Scholar] [CrossRef]
  31. Almeida, G.C.; Guimarães, L.C.; Resende, P.D.; Buono, V.T.L.; Peixoto, I.F.C.; Viana, A.C.D. Torsional behaviour of Reciproc and Reciproc blue instruments associated with their martensitic transformation temperatures. Int. Endod. J. 2019, 52, 1768–1772. [Google Scholar] [CrossRef]
  32. Staffoli, S.; Grande, N.M.; Plotino, G.; Özyürek, T.; Gündoğar, M.; Fortunato, L.; Polimeni, A. Influence of environmental temperature, heat-treatment and design on the cyclic fatigue resistance of three generations of a single-file nickel-titanium rotary instrument. Odontology 2019, 107, 301–307. [Google Scholar] [CrossRef] [PubMed]
  33. Garcia, P.R.; Resende, P.D.; Lopes, N.I.A.; Peixoto, I.; Buono, V.T.L.; Viana, A.C.D. Structural Characteristics and Torsional Resistance Evaluation of WaveOne and WaveOne Gold Instruments after Simulated Clinical Use. J. Endod. 2019, 45, 1041–1046. [Google Scholar] [CrossRef] [PubMed]
  34. Generali, L.; Borghi, A.; Lusvarghi, L.; Bolelli, G.; Veronesi, P.; Vecchi, A.; Consolo, U.; Becce, C.; Bertoldi, C.; Sassatelli, P. Evaluation of the usage-induced degradation of Genius and Reciproc nickel-titanium reciprocating instruments. Odontology 2019, 107, 473–481. [Google Scholar] [CrossRef] [PubMed]
  35. Arias, A.; Hejlawy, S.; Murphy, S.; de la Macorra, J.C.; Govindjee, S.; Peters, O.A. Variable impact by ambient temperature on fatigue resistance of heat-treated nickel titanium instruments. Clin. Oral Investig. 2019, 23, 1101–1108. [Google Scholar] [CrossRef]
  36. Khalil, W.A.; Natto, Z.S. Cyclic fatigue, bending resistance, and surface roughness of ProTaper Gold and EdgeEvolve files in canals with single- and double-curvature. Restor. Dent. Endod. 2019, 44, e19. [Google Scholar] [CrossRef]
  37. Üreyen Kaya, B.; Erik, C.E.; Kiraz, G. Atomic force microscopy and energy dispersive X-ray spectrophotometry analysis of reciprocating and continuous rotary nickel-titanium instruments following root canal retreatment. Microsc. Res. Tech. 2019, 82, 1157–1164. [Google Scholar] [CrossRef]
  38. Arias, A.; Macorra, J.C.; Govindjee, S.; Peters, O.A. Correlation between Temperature-dependent Fatigue Resistance and Differential Scanning Calorimetry Analysis for 2 Contemporary Rotary Instruments. J. Endod. 2018, 44, 630–634. [Google Scholar] [CrossRef]
  39. Pedullà, E.; Lo Savio, F.; La Rosa, G.R.M.; Miccoli, G.; Bruno, E.; Rapisarda, S.; Chang, S.W.; Rapisarda, E.; La Rosa, G.; Gambarini, G.; et al. Cyclic fatigue resistance, torsional resistance, and metallurgical characteristics of M3 Rotary and M3 Pro Gold NiTi files. Restor. Dent. Endod. 2018, 43, e25. [Google Scholar] [CrossRef]
  40. Pereira, E.S.J.; Amaral, C.C.F.; Gomes, J.; Peters, O.A.; Buono, V.T.L.; Bahia, M.G.A. Influence of clinical use on physical-structural surface properties and electrochemical potential of NiTi endodontic instruments. Int. Endod. J. 2018, 51, 515–521. [Google Scholar] [CrossRef]
  41. Uslu, G.; Özyürek, T.; Yılmaz, K. Effect of Sodium Hypochlorite and EDTA on Surface Roughness of HyFlex CM and HyFlex EDM Files. Microsc. Res. Tech. 2018, 81, 1406–1411. [Google Scholar] [CrossRef] [PubMed]
  42. Özyürek, T.; Yılmaz, K.; Uslu, G.; Plotino, G. The effect of root canal preparation on the surface roughness of WaveOne and WaveOne Gold files: Atomic force microscopy study. Restor. Dent. Endod. 2018, 43, e10. [Google Scholar] [CrossRef]
  43. Yılmaz, K.; Uslu, G.; Özyürek, T. Effect of multiple autoclave cycles on the surface roughness of HyFlex CM and HyFlex EDM files: An atomic force microscopy study. Clin. Oral Investig. 2018, 22, 2975–2980. [Google Scholar] [CrossRef] [PubMed]
  44. Iacono, F.; Pirani, C.; Generali, L.; Bolelli, G.; Sassatelli, P.; Lusvarghi, L.; Gandolfi, M.G.; Giorgini, L.; Prati, C. Structural analysis of HyFlex EDM instruments. Int. Endod. J. 2017, 50, 303–313. [Google Scholar] [CrossRef] [PubMed]
  45. Inan, U.; Gurel, M. Evaluation of surface characteristics of rotary nickel-titanium instruments produced by different manufacturing methods. Niger. J. Clin. Pract. 2017, 20, 143–146. [Google Scholar] [CrossRef] [Green Version]
  46. Cai, J.J.; Tang, X.N.; Ge, J.Y. Effect of irrigation on surface roughness and fatigue resistance of controlled memory wire nickel-titanium instruments. Int. Endod. J. 2017, 50, 718–724. [Google Scholar] [CrossRef]
  47. Shim, K.S.; Oh, S.; Kum, K.; Kim, Y.C.; Jee, K.K.; Chang, S.W. Mechanical and Metallurgical Properties of Various Nickel-Titanium Rotary Instruments. Biomed. Res. Int. 2017, 2017, 4528601. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  48. Kalyoncuoğlu, E.; Keskin, C.; Uzun, İ.; Bengü, A.S.; Guler, B. Scanning electron microscopy with energy dispersive X-ray spectrophotometry analysis of reciprocating and continuous rotary nickel-titanium instruments following root canal retreatment. J. Oral Sci. 2016, 58, 401–406. [Google Scholar] [CrossRef] [Green Version]
  49. Aminsobhani, M.; Khalatbari, M.S.; Meraji, N.; Ghorbanzadeh, A.; Sadri, E. Evaluation of the Fractured Surface of Five Endodontic Rotary Instruments: A Metallurgical Study. Iran. Endod. J. 2016, 11, 286–292. [Google Scholar] [CrossRef]
  50. de Vasconcelos, R.A.; Murphy, S.; Carvalho, C.A.; Govindjee, R.G.; Govindjee, S.; Peters, O.A. Evidence for Reduced Fatigue Resistance of Contemporary Rotary Instruments Exposed to Body Temperature. J. Endod. 2016, 42, 782–787. [Google Scholar] [CrossRef] [Green Version]
  51. Aun, D.P.; Peixoto, I.; Houmard, M.; Buono, V.T.L. Enhancement of NiTi superelastic endodontic instruments by TiO2 coating. Mater. Sci. Eng. C Mater. Biol. Appl. 2016, 68, 675–680. [Google Scholar] [CrossRef]
  52. Pirani, C.; Iacono, F.; Generali, L.; Sassatelli, P.; Nucci, C.; Lusvarghi, L.; Gandolfi, M.G.; Prati, C. HyFlex EDM: Superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined NiTi rotary instruments. Int. Endod. J. 2016, 49, 483–493. [Google Scholar] [CrossRef]
  53. Shen, Y.; Zhou, H.; Coil, J.M.; Aljazaeri, B.; Buttar, R.; Wang, Z.; Zheng, Y.F.; Haapasalo, M. ProFile Vortex and Vortex Blue Nickel-Titanium Rotary Instruments after Clinical Use. J. Endod. 2015, 41, 937–942. [Google Scholar] [CrossRef] [PubMed]
  54. Hieawy, A.; Haapasalo, M.; Zhou, H.; Wang, Z.J.; Shen, Y. Phase Transformation Behavior and Resistance to Bending and Cyclic Fatigue of ProTaper Gold and ProTaper Universal Instruments. J. Endod. 2015, 41, 1134–1138. [Google Scholar] [CrossRef]
  55. Nair, A.S.; Tilakchand, M.; Naik, B.D. The effect of multiple autoclave cycles on the surface of rotary nickel-titanium endodontic files: An in vitro atomic force microscopy investigation. J. Conserv. Dent. 2015, 18, 218–222. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  56. Can Sağlam, B.; Görgül, G. Evaluation of surface alterations in different retreatment nickel-titanium files: AFM and SEM study. Microsc. Res. Tech. 2015, 78, 356–362. [Google Scholar] [CrossRef] [PubMed]
  57. Braga, L.C.; Faria Silva, A.C.; Buono, V.T.; de Azevedo Bahia, M.G. Impact of heat treatments on the fatigue resistance of different rotary nickel-titanium instruments. J. Endod. 2014, 40, 1494–1497. [Google Scholar] [CrossRef]
  58. Prasad, P.S.; Sam, J.E.; Kumar, A.; Kannan. The effect of 5% sodium hypochlorite, 17% EDTA and triphala on two different rotary Ni-Ti instruments: An AFM and EDS analysis. J. Conserv. Dent. 2014, 17, 462–466. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  59. Pirani, C.; Paolucci, A.; Ruggeri, O.; Bossù, M.; Polimeni, A.; Gatto, M.R.; Gandolfi, M.G.; Prati, C. Wear and metallographic analysis of WaveOne and reciproc NiTi instruments before and after three uses in root canals. Scanning 2014, 36, 517–525. [Google Scholar] [CrossRef]
  60. Türker, S.A.; Sağlam, B.C.; Koçak, M.M.; Koçak, S. The effect of glide path on the surface quality of new and used rotary and reciprocating single files: OneShape versus WaveOne. Scanning 2014, 36, 608–613. [Google Scholar] [CrossRef]
  61. Pirani, C.; Ruggeri, O.; Cirulli, P.P.; Pelliccioni, G.A.; Gandolfi, M.G.; Prati, C. Metallurgical analysis and fatigue resistance of WaveOne and ProTaper nickel-titanium instruments. Odontology 2014, 102, 211–216. [Google Scholar] [CrossRef] [PubMed]
  62. Nakagawa, R.K.; Alves, J.L.; Buono, V.T.; Bahia, M.G. Flexibility and torsional behaviour of rotary nickel-titanium PathFile, RaCe ISO 10, Scout RaCe and stainless steel K-File hand instruments. Int. Endod. J. 2014, 47, 290–297. [Google Scholar] [CrossRef]
  63. Fatma, Y.; Ozgur, U. Evaluation of surface topography changes in three NiTi file systems using rotary and reciprocal motion: An atomic force microscopy study. Microsc. Res. Tech. 2014, 77, 177–182. [Google Scholar] [CrossRef] [PubMed]
  64. Fayyad, D.M.; Mahran, A.H. Atomic force microscopic evaluation of nanostructure alterations of rotary NiTi instruments after immersion in irrigating solutions. Int. Endod. J. 2014, 47, 567–573. [Google Scholar] [CrossRef]
  65. Shen, Y.; Coil, J.M.; Zhou, H.; Zheng, Y.; Haapasalo, M. HyFlex nickel-titanium rotary instruments after clinical use: Metallurgical properties. Int. Endod. J. 2013, 46, 720–729. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  66. Shen, Y.; Coil, J.M.; Zhou, H.M.; Tam, E.; Zheng, Y.F.; Haapasalo, M. ProFile Vortex instruments after clinical use: A metallurgical properties study. J. Endod. 2012, 38, 1613–1617. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  67. Spagnuolo, G.; Ametrano, G.; D’Antò, V.; Rengo, C.; Simeone, M.; Riccitiello, F.; Amato, M. Effect of autoclaving on the surfaces of TiN-coated and conventional nickel-titanium rotary instruments. Int. Endod. J. 2012, 45, 1148–1155. [Google Scholar] [CrossRef]
  68. Sağlam, B.C.; Koçak, S.; Koçak, M.M.; Topuz, O. Effects of irrigation solutions on the surface of ProTaper instruments: A microscopy study. Microsc. Res. Tech. 2012, 75, 1534–1538. [Google Scholar] [CrossRef]
  69. Yamazaki-Arasaki, A.; Cabrales, R.; Santos, M.; Kleine, B.; Prokopowitsch, I. Topography of four different endodontic rotary systems, before and after being used for the 12th time. Microsc. Res. Tech. 2012, 75, 97–102. [Google Scholar] [CrossRef]
  70. Shen, Y.; Zhou, H.M.; Zheng, Y.F.; Campbell, L.; Peng, B.; Haapasalo, M. Metallurgical characterization of controlled memory wire nickel-titanium rotary instruments. J. Endod. 2011, 37, 1566–1571. [Google Scholar] [CrossRef]
  71. Ametrano, G.; D’Antò, V.; Di Caprio, M.P.; Simeone, M.; Rengo, S.; Spagnuolo, G. Effects of sodium hypochlorite and ethylenediaminetetraacetic acid on rotary nickel-titanium instruments evaluated using atomic force microscopy. Int. Endod. J. 2011, 44, 203–209. [Google Scholar] [CrossRef]
  72. Viana, A.C.; Chaves Craveiro de Melo, M.; Guiomar de Azevedo Bahia, M.; Lopes Buono, V.T. Relationship between flexibility and physical, chemical, and geometric characteristics of rotary nickel-titanium instruments. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 2010, 110, 527–533. [Google Scholar] [CrossRef] [PubMed]
  73. Zinelis, S.; Eliades, T.; Eliades, G. A metallurgical characterization of ten endodontic Ni-Ti instruments: Assessing the clinical relevance of shape memory and superelastic properties of Ni-Ti endodontic instruments. Int. Endod. J. 2010, 43, 125–134. [Google Scholar] [CrossRef] [PubMed]
  74. Condorelli, G.G.; Bonaccorso, A.; Smecca, E.; Schäfer, E.; Cantatore, G.; Tripi, T.R. Improvement of the fatigue resistance of NiTi endodontic files by surface and bulk modifications. Int. Endod. J. 2010, 43, 866–873. [Google Scholar] [CrossRef] [PubMed]
  75. Alapati, S.B.; Brantley, W.A.; Iijima, M.; Schricker, S.R.; Nusstein, J.M.; Li, U.M.; Svec, T.A. Micro-XRD and temperature-modulated DSC investigation of nickel-titanium rotary endodontic instruments. Dent. Mater. 2009, 25, 1221–1229. [Google Scholar] [CrossRef] [PubMed]
  76. Shen, Y.; Coil, J.M.; McLean, A.G.; Hemerling, D.L.; Haapasalo, M. Defects in nickel-titanium instruments after clinical use. Part 5: Single use from endodontic specialty practices. J. Endod. 2009, 35, 1363–1367. [Google Scholar] [CrossRef] [PubMed]
  77. Bonaccorso, A.; Schäfer, E.; Condorelli, G.G.; Cantatore, G.; Tripi, T.R. Chemical analysis of nickel-titanium rotary instruments with and without electropolishing after cleaning procedures with sodium hypochlorite. J. Endod. 2008, 34, 1391–1395. [Google Scholar] [CrossRef]
  78. Alves-Claro, A.P.; Claro, F.A.; Uzumaki, E.T. Wear resistance of nickel-titanium endodontic files after surface treatment. J. Mater. Sci. Mater. Med. 2008, 19, 3273–3277. [Google Scholar] [CrossRef]
  79. Valois, C.R.; Silva, L.P.; Azevedo, R.B. Multiple autoclave cycles affect the surface of rotary nickel-titanium files: An atomic force microscopy study. J. Endod. 2008, 34, 859–862. [Google Scholar] [CrossRef]
  80. Topuz, O.; Aydin, C.; Uzun, O.; Inan, U.; Alacam, T.; Tunca, Y.M. Structural effects of sodium hypochlorite solution on RaCe rotary nickel-titanium instruments: An atomic force microscopy study. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 2008, 105, 661–665. [Google Scholar] [CrossRef]
  81. Li, U.M.; Iijima, M.; Endo, K.; Brantley, W.A.; Alapati, S.B.; Lin, C.P. Application of plasma immersion ion implantation for surface modification of nickel-titanium rotary instruments. Dent. Mater. J. 2007, 26, 467–473. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  82. Inan, U.; Aydin, C.; Uzun, O.; Topuz, O.; Alacam, T. Evaluation of the surface characteristics of used and new ProTaper Instruments: An atomic force microscopy study. J. Endod. 2007, 33, 1334–1337. [Google Scholar] [CrossRef]
  83. Hayashi, Y.; Yoneyama, T.; Yahata, Y.; Miyai, K.; Doi, H.; Hanawa, T.; Ebihara, A.; Suda, H. Phase transformation behaviour and bending properties of hybrid nickel-titanium rotary endodontic instruments. Int. Endod. J. 2007, 40, 247–253. [Google Scholar] [CrossRef]
  84. Alexandrou, G.B.; Chrissafis, K.; Vasiliadis, L.P.; Pavlidou, E.; Polychroniadis, E.K. SEM observations and differential scanning calorimetric studies of new and sterilized nickel-titanium rotary endodontic instruments. J. Endod. 2006, 32, 675–679. [Google Scholar] [CrossRef]
  85. Alexandrou, G.; Chrissafis, K.; Vasiliadis, L.; Pavlidou, E.; Polychroniadis, E.K. Effect of heat sterilization on surface characteristics and microstructure of Mani NRT rotary nickel-titanium instruments. Int. Endod. J. 2006, 39, 770–778. [Google Scholar] [CrossRef]
  86. Valois, C.R.; Silva, L.P.; Azevedo, R.B. Atomic force microscopy study of stainless-steel and nickel-titanium files. J. Endod. 2005, 31, 882–885. [Google Scholar] [CrossRef]
  87. Miyai, K.; Ebihara, A.; Hayashi, Y.; Doi, H.; Suda, H.; Yoneyama, T. Influence of phase transformation on the torsional and bending properties of nickel-titanium rotary endodontic instruments. Int. Endod. J. 2006, 39, 119–126. [Google Scholar] [CrossRef]
  88. Tripi, T.R.; Bonaccorso, A.; Condorelli, G.G. Fabrication of hard coatings on NiTi instruments. J. Endod. 2003, 29, 132–134. [Google Scholar] [CrossRef]
  89. Tripi, T.R.; Bonaccorso, A.; Rapisarda, E.; Tripi, V.; Condorelli, G.G.; Marino, R.; Fragalà, I. Depositions of nitrogen on NiTi instruments. J. Endod. 2002, 28, 497–500. [Google Scholar] [CrossRef] [PubMed]
  90. Brantley, W.A.; Svec, T.A.; Iijima, M.; Powers, J.M.; Grentzer, T.H. Differential scanning calorimetric studies of nickel-titanium rotary endodontic instruments after simulated clinical use. J. Endod. 2002, 28, 774–778. [Google Scholar] [CrossRef] [PubMed]
  91. Brantley, W.A.; Svec, T.A.; Iijima, M.; Powers, J.M.; Grentzer, T.H. Differential scanning calorimetric studies of nickel titanium rotary endodontic instruments. J. Endod. 2002, 28, 567–572. [Google Scholar] [CrossRef]
  92. Kuhn, G.; Jordan, L. Fatigue and mechanical properties of nickel-titanium endodontic instruments. J. Endod. 2002, 28, 716–720. [Google Scholar] [CrossRef]
  93. Martins, R.C.; Bahia, M.G.; Buono, V.T. Surface analysis of ProFile instruments by scanning electron microscopy and X-ray energy-dispersive spectroscopy: A preliminary study. Int. Endod. J. 2002, 35, 848–853. [Google Scholar] [CrossRef] [PubMed]
  94. Kuhn, G.; Tavernier, B.; Jordan, L. Influence of Structure on Nickel-Titanium Endodontic Instruments Failure. J. Endod. 2001, 27, 516–520. [Google Scholar] [CrossRef]
  95. Zanza, A.; Seracchiani, M.; Di Nardo, D.; Reda, R.; Gambarini, G.; Testarelli, L. A Paradigm Shift for Torsional Stiffness of Nickel-Titanium Rotary Instruments: A Finite Element Analysis. J. Endod. 2021, 47, 1149–1156. [Google Scholar] [CrossRef] [PubMed]
  96. Di Nardo, D.; Zanza, A.; Seracchiani, M.; Donfrancesco, O.; Gambarini, G.; Testarelli, L. Angle of Insertion and Torsional Resistance of Nickel–Titanium Rotary Instruments. Materials 2021, 14, 3744. [Google Scholar] [CrossRef] [PubMed]
  97. Zanza, A.; Seracchiani, M.; Reda, R.; Di Nardo, D.; Gambarini, G.; Testarelli, L. Role of the Crystallographic Phase of NiTi Rotary Instruments in Determining Their Torsional Resistance during Different Bending Conditions. Materials 2021, 14, 6324. [Google Scholar] [CrossRef] [PubMed]
  98. Chugh, V.K.; Patnana, A.K.; Chugh, A.; Kumar, P.; Wadhwa, P.; Singh, S. Clinical differences of hand and rotary instrumentations during biomechanical preparation in primary teeth-A systematic review and meta-analysis. Int. J. Paediatr. Dent. 2021, 31, 131–142. [Google Scholar] [CrossRef]
  99. Gambarini, G.; Seracchiani, M.; Zanza, A.; Miccoli, G.; Del Giudice, A.; Testarelli, L. Influence of shaft length on torsional behavior of endodontic nickel-titanium instruments. Odontology 2021, 109, 568–573. [Google Scholar] [CrossRef]
  100. Jandt, K. Atomic force Microscopy of Biomaterials Surfaces and Interfaces. Surf. Sci. 2001, 491, 303–332. [Google Scholar] [CrossRef]
  101. Paredes, J.I.; Martínez-Alonso, A.; Tascón, J.M.D. Application of scanning tunneling and atomic force microscopies to the characterization of microporous and mesoporous materials. Microporous Mesoporous Mater. 2003, 65, 93–126. [Google Scholar] [CrossRef]
  102. Assender, H.; Bliznyuk, V.; Porfyrakis, K. How Surface Topography Relates to Materials’ Properties. Science 2002, 297, 973–976. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  103. Pirani, C.; Cirulli, P.P.; Chersoni, S.; Micele, L.; Ruggeri, O.; Prati, C. Cyclic fatigue testing and metallographic analysis of nickel-titanium rotary instruments. J. Endod. 2011, 37, 1013–1016. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Schematic representation of the 81 articles according to publication journals. JOE (Journal of Endodontics); IEJ (International Endodontic Journal); EEJ (European Endodontic Journal), BMC (BMC Oral Health); CLOI (Clinical Oral Investigation); AEJ (Australian Endodontic Journal); GD (General Dentistry); IJD (International Journal of Dentistry); Odon. (Odontology); RDE (Restorative Dentistry and Endodontics); MRT (Microscopy Research and Technique); NJCP (Nigerian Journal of Clinical Practice); BRI (BioMed Research International); JOS (Journal of Oral Science); Iran. Endod. Jour. (Iranian Endodontic Journal); MSE (Materials Sciences and Engineering); JCD (Journal of Conservative Dentistry); Scann. (Scanning); OOOE (Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology); Den. Mat. (Dental Materials); JMS (Journal of Materials Science: Materials in Medicine); Den. Mat. Journ (Dental Materials Journal).
Figure 1. Schematic representation of the 81 articles according to publication journals. JOE (Journal of Endodontics); IEJ (International Endodontic Journal); EEJ (European Endodontic Journal), BMC (BMC Oral Health); CLOI (Clinical Oral Investigation); AEJ (Australian Endodontic Journal); GD (General Dentistry); IJD (International Journal of Dentistry); Odon. (Odontology); RDE (Restorative Dentistry and Endodontics); MRT (Microscopy Research and Technique); NJCP (Nigerian Journal of Clinical Practice); BRI (BioMed Research International); JOS (Journal of Oral Science); Iran. Endod. Jour. (Iranian Endodontic Journal); MSE (Materials Sciences and Engineering); JCD (Journal of Conservative Dentistry); Scann. (Scanning); OOOE (Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology); Den. Mat. (Dental Materials); JMS (Journal of Materials Science: Materials in Medicine); Den. Mat. Journ (Dental Materials Journal).
Bioengineering 09 00030 g001
Figure 2. Schematic representation of the metallurgical tests used in the 81 selected articles from 2001 to 2021. XRD (X-ray diffraction), DSC (differential scanning calorimetry), EDS (energy-dispersive X-ray spectroscopy), FIB (focused ion beam analysis), AES (Auger electron spectroscopy), XPS (X-ray photoelectron spectroscopy) and AFM (atomic force microscopy); electr. potential (electrochemical potential measurement); metal. analysis (metallographic analysis).
Figure 2. Schematic representation of the metallurgical tests used in the 81 selected articles from 2001 to 2021. XRD (X-ray diffraction), DSC (differential scanning calorimetry), EDS (energy-dispersive X-ray spectroscopy), FIB (focused ion beam analysis), AES (Auger electron spectroscopy), XPS (X-ray photoelectron spectroscopy) and AFM (atomic force microscopy); electr. potential (electrochemical potential measurement); metal. analysis (metallographic analysis).
Bioengineering 09 00030 g002
Table 1. Summary of all articles that met the inclusion criteria, in chronological order from more recent to older. XRD (X-ray diffraction), DSC (differential scanning calorimetry), EDS (energy-dispersive X-ray spectroscopy), FIB (focused ion beam analysis), AES (Auger electron spectroscopy), XPS (X-ray photoelectron spectroscopy) and AFM (atomic force microscopy).
Table 1. Summary of all articles that met the inclusion criteria, in chronological order from more recent to older. XRD (X-ray diffraction), DSC (differential scanning calorimetry), EDS (energy-dispersive X-ray spectroscopy), FIB (focused ion beam analysis), AES (Auger electron spectroscopy), XPS (X-ray photoelectron spectroscopy) and AFM (atomic force microscopy).
First Author and Publication YearJournalMetallurgical TestsSelected Instruments
Martins (2021) [15]International Endodontic JournalDSC and EDSProTaper Next and X-File
Jose (2021) [16]European Endodontic JournalAFM and EDSXP-endo Shaper and TruNatomy
Jo (2021) [17]MaterialsDSCOneShape, OneCurve, WaveOne Gold and HyFlex EDM
Van Pham (2021) [18]BMC Oral HealthDSCReciproc, HyFlex CM, and Neoniti A1
Martins (2021) [19]Clinical Oral InvestigationsEDS and DSCProTaper Universal, U-File, ProTaper Gold, Premium Taper Gold, Go-Taper Flex,
Kalyoncuoğlu (2021) [20]Australian Endodontic JournalXRD and DSCReciproc and Reciproc Blue
Martins (2021) [21]International Endodontic JournalDSC, EDSReciproc, Reciproc Blue, One Files, One Files Blue, Reverso Silver, and WaveOne Gold
Azizi (2021) [22]European Endodontic JournalEDS, micro-Raman spectroscopy, FIB, XRD, metallographic analysis and DSC (Buono)OneShape and OneCurve
Keskin (2021) [23]Clinical Oral InvestigationsDSCRotate, Reciproc Blue, Reciproc, and Mtwo
Martins (2020) [24]Journal of EndodonticsEDS and DSCProTaper Universal, EdgeTaper, U-File, Go-Taper Universal, Super Files, Multitaper and Pluri Taper.
Martins (2020) [25]Journal of EndodonticsEDS and DSCProTaper Universal, ProTaper Gold, U-File, Super Files and Super Files Blue
Generali (2020) [26]MaterialsEDS, FIB, micro-Raman spectroscopy, XRD, DSC and AES with depth profilingProcodile and Reziflow
Silva (2020) [14]Journal of EndodonticsDSC and EDSNeoNiti, HyFlex EDM, ProTaper Gold and ProTaper Universa
Weyh (2020) [27]General DentistryEDSProTaper Gold, EdgeTaper Platinum, ProTaper Universal, EdgeTaper, Vortex Blue and EdgeSequel Sapphire.
Generali (2020) [28]International Endodontic JournalEDS, FIB, micro-Raman spectroscopy, metallographic analysis, XRD, DSC and depth-sensing indentation testReciproc and Reciproc Blue
Arias (2020) [29]Clinical Oral InvestigationsDSCHyFlex EDM and TRUShape
Khabadze (2020) [30]International Journal of DentistryEDS and XRDProTaper Universal
Almeida (2019) [31]International Endodontic JournalDSCReciproc and Reciproc Blue
Staffoli (2019) [32]OdontologyDSCOne Curve and One Shape New Generation, One Shape
Garcia (2019) [33]Journal of EndodonticsXRD and DSCWaveOne and WaveOne Gold
Generali (2019) [34]OdontologyEDS, DSC, metallographic analysis, XRD and depth-sensing indentation testEndo-Eze™ Genius and Reciproc
Arias (2019) [35]Clinical Oral InvestigationsDSCEdgeSequel Sapphire and Vortex Blue
Khalil (2019) [36]Restorative Dentistry and EndodonticsAFM and EDSEdgeEvolve and ProTaper Gold
Üreyen Kaya (2019) [37]Microscopy Research and TechniqueEDS and AFMProTaper Retreatment and WaveOne Gold
Arias (2018) [38]Journal of EndodonticsDSCHyflex EDM and TRUShape
Pedullà (2018) [39]Restorative Dentistry and EndodonticsDSCM3 Rotary and M3 Pro Gold
Pereira (2018) [40]International Endodontic JournalEDS and electrochemical potential measurements ProTaper Universal, ProTaper Next, Typhoon, Hyflex EDM and Vortex Blue
Uslu (2018) [41]Microscopy Research and TechniqueAFMHyFlex CM and HyFlex EDM
Özyürek (2018) [42]Restorative Dentistry and EndodonticsAFMWaveOne and WaveOne Gold
Yılmaz (2018) [43]Clinical Oral InvestigationsAFMHyFlex EDM and HyFlex CM
Iacono (2017) [44]International Endodontic JournalXRD, DSC, EDS, micro-Raman spectroscopy and depth-sensing indentation testHyFlex EDM and HyFlex CM
Inan (2017) [45]Nigerian Journal of Clinical PracticeAFMTwisted Files and Mtwo
Cai (2017) [46]International Endodontic JournalAFMHyFlex CM and M3 Rotary
Shim (2017) [47] BioMed Research InternationalDSCProFile, K3, One Shape, ProTaper Next, Reciproc, WaveOne, HyFlex CM and Twisted File
Kalyoncuoğlu (2016) [48]Journal of Oral ScienceEDSReciproc and ProTaper Retreatment
Aminsobhani (2016) [49]Iranian Endodontic JournalEDS, DSC and XRDNeoniti, iRaCe, Mtwo, Twisted File and ProTaper Next
de Vasconcelos (2016) [50]Journal of EndodonticsDSCProTaper Universal, HyFlex CM, TRUShape and Vortex Blue
Aun (2016) [51]Materials Sciences and Engineering DSC and XRDiRaCe
Pirani (2016) [52]International Endodontic JournalEDS and metallographic analysisHyflex EDM
Shen (2015) [53]Journal of EndodonticsDSCProFile Vortex and Vortex Blue
Hieawy (2015) [54]Journal of EndodonticsDSCProTaper Universal and ProTaper Gold
Nair (2015) [55]Journal of Conservative DentistryAFMMtwo and ProTaper Universal
Can Sağlam (2015) [56]Microscopy Research and TechniqueAFMProTaper Retreatment, R-endo and Mtwo retreatment
Braga (2014) [57]Journal of EndodonticsEDS and DSCEndoWave, ProFile Vortex, HyFlex CM, Typhoon and ProTaper Universal
Prasad (2014) [58]Journal of Conservative DentistryAFM and EDSProTaper Universal and iRaCe
Pirani (2014) [59]ScanningEDS and metallographic analysisWaveOne and Reciproc
Türker (2014) [60]ScanningAFMOneShape and WaveOne
Pirani (2014) [61]OdontologyEDS and metallographic analysisWaveOne primary and ProTaper Universal
Nakagawa (2014) [62]International Endodontic JournalEDS, DSC and XRDPathFile, Race ISO 10 and Scout RaCe
Fatma (2014) [63]Microscopy Research and TechniqueAFMProTaper Universal, Reciproc and WaveOne
Fayyad (2014) [64]International Endodontic JournalAFMProFile GT, Twisted File, RaCe and HeroShaper
Shen (2013) [65]International Endodontic JournalDSC and XRDHyFlex CM
Shen (2012) [66]Journal of EndodonticsDSC and XRDProFile Vortex
Spagnuolo (2012) [67]International Endodontic JournalEDS and AFMProTaper Universal and AlphaKite
Sağlam (2012) [68]Microscopy Research and TechniqueAFMProTaper Universal
Yamazaki-Arasaki (2012) [69]Microscopy Research and TechniqueAFMK3, ProTaper Universal, Twisted Files and BioRace
Shen (2011) [70]Journal of EndodonticsDSC, XRD, EDS and metallographic analysisEndoSequence, ProFile, ProFile Vortex, Twisted Files, Typhoon and Typhoon™ CM.
Ametrano (2011) [71] International Endodontic JournalAFMProTaper Universal
Viana (2010) [72]Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and EndodontologyEDS, XRD and DSCProTaper Universal, K3, and EndoSequence.
Zinelis (2010) [73]International Endodontic JournalEDS and XRDEndoSequence, FlexMaster, Hero 642, K3, Liberator, NRT, ProFile and ProTaper Universal
Condorelli (2010) [74]International Endodontic JournalXPS, EDS and XRDRaCe
Alapati (2009) [75]Dental MaterialsXRD and DSCProFile GT, ProTaper Universal, K3 and Quantec
Shen (2009) [76]Journal of EndodonticsEDSProFile Series 29, ProFile and ProTaper Universal
Bonaccorso (2008) [77] Journal of EndodonticsEDS and AESRaCe
Alves-Claro (2008) [78] Journal of Materials Science: Materials in MedicineXPSNitiflex
Valois (2008) [79]Journal of EndodonticsAFMGreater Taper and ProFile
Topuz (2008) [80]Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and EndodontologyAFMRaCe
Li (2007) [81]Dental Materials JournalXPS and DSCProTaper Universal
Inan (2007) [82]Journal of EndodonticsAFMProTaper Universal
Hayashi (2007) [83]International Endodontic JournalDSCPrototype Ni–Ti rotary instruments
Alexandrou (2006) [84]Journal of EndodonticsDSCProFile and FlexMaster
Alexandrou (2006) [85]International Endodontic JournalDSCNRT
Valois (2005) [86]Journal of EndodonticsAFM Greater Taper and Quantec
Miyai (2005) [87]International Endodontic JournalDSCEndoWave, Hero 642, K3, ProFile and ProTaper Universal
Tripi (2003) [88]Journal of EndodonticsXPS and XRDProFile GT
Tripi (2002) [89]Journal of EndodonticsXPS and XRDProFile GT
Brantley (2002) [90]Journal of EndodonticsDSCProFile and Lightspeed
Brantley (2002) [91]Journal of EndodonticsDSCProFile and Lightspeed
Kuhn (2002) [92]Journal of EndodonticsDSCProFile and Hero
Martins (2002) [93]Journal of EndodonticsEDSProFile
Kuhn (2001) [94]Journal of EndodonticsXRDProFile and Hero
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Zanza, A.; Seracchiani, M.; Reda, R.; Miccoli, G.; Testarelli, L.; Di Nardo, D. Metallurgical Tests in Endodontics: A Narrative Review. Bioengineering 2022, 9, 30. https://doi.org/10.3390/bioengineering9010030

AMA Style

Zanza A, Seracchiani M, Reda R, Miccoli G, Testarelli L, Di Nardo D. Metallurgical Tests in Endodontics: A Narrative Review. Bioengineering. 2022; 9(1):30. https://doi.org/10.3390/bioengineering9010030

Chicago/Turabian Style

Zanza, Alessio, Marco Seracchiani, Rodolfo Reda, Gabriele Miccoli, Luca Testarelli, and Dario Di Nardo. 2022. "Metallurgical Tests in Endodontics: A Narrative Review" Bioengineering 9, no. 1: 30. https://doi.org/10.3390/bioengineering9010030

APA Style

Zanza, A., Seracchiani, M., Reda, R., Miccoli, G., Testarelli, L., & Di Nardo, D. (2022). Metallurgical Tests in Endodontics: A Narrative Review. Bioengineering, 9(1), 30. https://doi.org/10.3390/bioengineering9010030

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop