Article

Design and Construction of a Device to Evaluate the Performance of Variable Orifice Flow Meters (VOFM)

by
William Prado Martínez
1,*,
Juan Felipe Arroyave Londoño
2,* and
Jefferson Vásquez Gómez
2
1
Department of Mecathronics Engineering, Universidad Tecnológica de Pereira, Pereira 66001, Colombia
2
Department of Mechanical Tecnology, Universidad Tecnológica de Pereira, Pereira 66001, Colombia
*
Authors to whom correspondence should be addressed.
Inventions 2023, 8(5), 110; https://doi.org/10.3390/inventions8050110
Submission received: 10 July 2023 / Revised: 13 August 2023 / Accepted: 25 August 2023 / Published: 30 August 2023
(This article belongs to the Special Issue Recent Advances in Fluid Mechanics and Transport Phenomena)

Abstract

:
This work presents a low-cost device for evaluating Variable Orifice Flow Meters (VOFM) used in medical mechanical ventilation applications. Specifically, the equipment was used to assess the impact of length and thickness on pressure drop for different flows in a rectangular geometry VOFM. A total of six VOFMs, with three different lengths and two different thicknesses, were evaluated. All VOFMs were stimulated with an airflow ranging from 0 L . m i n 1 to 90 L . m i n 1 , with increments of approximately 2 L . m i n 1 . The experiments conducted with the device showed a strong relationship between pressure drop P and flow rate Q in the evaluated VOFMs, with two different zones: one exhibiting non-linear behavior and another showing linear behavior. The results suggest that increased length and decreased thickness lead to higher sensitivity. However, it is essential to reduce the cross-sectional area to mitigate nonlinear effects of the sensor.

1. Introduction

Differential pressure flow meters are widely used devices for measuring fluid flow in pipelines due to their ease of installation and maintenance as well as their capability to measure different fluids. These devices operate based on the pressure drop when the fluid passes through a restriction [1], as shown in Figure 1.
According to the experiments, the fluid increases its velocity. It decreases its pressure as it passes through a reduced area, while it decreases its velocity and increases its pressure if the area is increased. Equation (1), which relates the pressure drop P = P 1 P 2 and the flow Q through the restriction in Figure 1, can be derived by applying the Bernoulli equation and the principle of mass conservation. This non-linear equation predicts that the flow is proportional to the square root of the pressure drop and has been widely used for flow measurement using devices with fixed area restrictions, such as the Venturi tube, fixed orifice plate, and nozzle:
Q = v 1 A 1 = A 2 1 A 2 A 1 2 2 ρ P    
where A 1 is the inlet area of the flowmeter, A 2 is the passage area of flow obstruction, and ρ is the fluid density.
However, in industrial applications, deviations in measurement often arise due to factors such as flow profile and fluid properties, among others. Therefore, an experimentally determined correction coefficient is required, in addition to making adjustments related to the design, installation, and use of differential pressure meters to compensate for these deviations [2,3,4]. The abovementioned requirements are detailed in the ISO 5167-1 standard [5], establishing flow measurement standards in piping systems. The objective of this standard is to ensure the reliability of measurements.
In the case of medical applications, such as monitoring the respiratory function of patients connected to mechanical ventilation systems and anesthesia machines, flow meters with fixed area restrictions can generate a significantly high-pressure drop compared to the operating pressure [6,7]. This pressure drop results in an effort that hinders the efficient evacuation of gases from the patient during the expiratory phase, disrupting the respiratory cycle’s naturalness [8]. This additional effort can make it challenging to obtain reliable and accurate measurements, which in turn hampers the ability of medical professionals to characterize the mechanical properties of the pulmonary system, understand the interaction between the patient and the mechanical ventilator, and accurately monitor changes in respiratory muscle activity [9,10,11]. On the other hand, an inaccurate measurement of flow can lead to inadequate patient ventilation therapy, which in the case of air and oxygen mixtures, can result in conditions such as irreversible pulmonary fibrosis and retrolental fibroplasia, among others [12]. Additionally, medical flow meters are susceptible to condensing moisture or secretions [13] and contamination by viruses and bacteria, which is why these meters must undergo strict sterilization processes for use on different patients or be limited to single-patient use, thereby increasing their cost [14,15].
The Variable Orifice Meter (VOFM) has been developed to reduce expiratory effort and cost. It has gained popularity due to its simplicity, robustness, low manufacturing cost, linear response, and ease of integration with electronic sensors and digital filters. The VOFM offers several advantages, including fast and accurate response, silent operation, and the ability to operate in a wide range of conditions, such as condensation or secretions in the gas flow exhaled by a patient [16,17]. The VOFM is a flexible membrane placed transversely in the respiratory circuit, with one or more cantilever beams cut on it. When the gas Q v o f m passes through the VOFM membrane, it creates a pressure drop P that flexes the beams and increases the effective flow area. As a result, the expiratory effort is reduced.
A review conducted on Google Patents and Espacenet found approximately 40 patents related to VOFMs, of which nine, shown in Table 1, present claims that improve the linear relationship between Q v o f m and P . The simplest reported shape is the rectangular geometry, while the geometry with the most operating claims is the rhomboidal geometry.
The patent US4083245 [18] details that the rhomboidal shape of the beam generates a more linear relationship between pressure drop and flow than other shapes, such as rectangular or circular [19]. This means that a differential pressure gauge can indicate the flow, which is advantageous in measurement directly. On the other hand, patents US4006634A [20] and US4989456A [21] argue that using a single beam instead of multiple beams significantly reduces the noise present in the designs due to vibration and resonance effects. Regarding the location of the pressure taps, patent US4006634A states that they should be placed on the side of the beam’s fixed end to avoid affecting the measurement due to gas vortices or other unwanted effects that could distort the pressure reading and consequently affect the accuracy. Table 2 presents the type of material and thickness stated in each invention. The use of Mylar® (biaxially oriented polyethylene terephthalate) is common in several of the mentioned patents due to its beneficial characteristics such as dimensional stability, flatness, chemical resistance, fatigue resistance, and heat resistance up to 230 °C, good electrical resistance, and adequate mechanical strength [22,23,24]. Mylar® is commercially available in thicknesses ranging from 0.001 inches to 0.014 inches.
Despite the extensive development of VOFM, the technical and scientific literature does not report experiments evaluating the impact of the length and thickness of the VOFM beam on its response. Therefore, this work proposes the construction of a device to assess the influence of beam length and thickness on the behavior of a variable orifice meter for medical applications. The design of the device and the experimental results obtained for six rectangular VOFM designs will be presented.

Theoretical Flow Rate Q

The theoretical flow rate Q considered in this work corresponds to the calculation based on Equation (1) for an incompressible fluid flowing through a restriction without considering energy losses due to fluid conditions or restriction geometry [27]. The pressure difference exerts a force per unit length on the cantilever beam, resulting in its deflection. The greater the deflection of the beam, the larger the opening area for gas flow. Additionally, the deflection model [28] of a beam with a distributed load due to steady fluid flow [29], as employed in this work with the rectangular configuration shown in Figure 2, is calculated from Equation (2), which predicts deflection as a function of pressure drop:
y = 3 L 4 2 E e 3     P
where B is the width at the support, L the length, e the thickness, I the moment of inertia, E the elastic modulus, and t cutting thickness, as shown in Figure 2a,b.
According to Figure 2b, the opening area A 2 can then be calculated by adding to the area of the cut A e = 2 L + B t and the increment of the area due to deflection, as shown in Equation (3):
A 2 = 2 L + B   t 2 + B + L 3 L 4 2 E e 3     P   2  
Therefore, Equation (4) for the predicted theoretical flow rate Q for the VOFM can be derived as follows:
Q = B + L 3 L 4 2 E e 3     P   2 + 2 L + B   t 2       1 B + L 3 L 4 2 E e 3     P   2 + 2 L + B   t 2       A 1 2 2 ρ P

2. Measurement System

Taking into account the flow and pressure variables described in the models for evaluating the behavior of beams in variable orifice flow meters, we have designed and developed a comprehensive measurement system. This system comprises a gas circuit with a container specially designed to secure the beam of the plates. Additionally, an electronic system dedicated to the accurate reading of the sensors and actuator control has been implemented. To ensure effective data tracking and timely storage, we have incorporated a software platform that enables continuous monitoring and capturing of relevant information.
Furthermore, given that the VOFM is intended for medical use, design considerations for the measurement system were based on the guidelines provided by the National Institute of Food and Drug Surveillance of Colombia (INVIMA), which include the following: the mechanical ventilation conditions [30,31], recommendations from Colombian scientific societies [32,33], international health agencies [34,35,36], and the guidelines provided by the World Health Organization (WHO) [37].
Figure 3 shows the device to Evaluate the Performance of Variable Orifice Flow Meters (VOFM) used in this work, which consists of a gas circuit with a container housing the VOFM, an electronic system for sensor reading and actuator control, and software for data monitoring and storage.

2.1. Gas Circuit

The gas from a compressed air network is regulated into the VOFM container using a proportional solenoid valve. Then, the gas flows into container (b), where the pressure taps are located, P 1 and P 2 .
The VOFM container, depicted in Figure 4, was designed using two symmetrical cone-shaped parts. According to patent EP-0331773A1 [38], the cone angle is set at 23.61°. A flanged connection and an O-ring gasket ensure the container’s airtightness. The positioning of the pressure taps on the upper section of the container enables the VOFM to measure bidirectional flow effectively, unaffected by condensation and the downward orientation of the VOFM beam which reduces the effects of turbulence on the pressure taps. The container was fabricated using 3D printing technology and medical-grade resin.

2.2. Electronic System

The electronic system was developed using an Arduino Due® and three PCBs (Printed Circuit Boards) designed and manufactured in the Universidad Tecnológica de Pereira. The Arduino Due® and the three boards connected in a four-level arrangement, as shown in Figure 5a. The Arduino Due processing board (b) is responsible for processing sensor information, establishing communication with the data acquisition software, and controlling the operation sequence.
The power board (c) regulates the voltage levels of 3.3 V and 5 V that power the sensors from a 24 V 5A switch-type power supply.
On board (d), the proportional valve control circuit is situated, comprising an LM358 operational amplifier and a TIP120 transistor. This circuit governs the proportional valve’s opening Yongchuang YCLT21-1GBVV-SC11. For this purpose, the control circuit regulates the current flowing through the coil based on the analog voltage signal DAC0, ranging from 0 V to 3.3 V, supplied by the Arduino Due, as illustrated in Figure 6.
The proportional valve Yongchuang YCLT21-1GBVV-SC11 operates with a maximum coil current of 270 mA at Full Scale (FS) based on a power calculation of 6.5 W and a voltage of 24 VDC. When subjected to a supply pressure of 400 kPa (58 psi) and a current of 85% FS (230 mA), the resulting flow rate from the proportional valve is approximately 110 ± 10 L . m i n 1 . Additionally, under the same supply pressure, the minimum current required to initiate the valve’s proportional opening is approximately 30% FS (80 mA).
The instrumentation board (e) is designed to connect a differential pressure sensor SPD1 to measure the pressure drop P across the VOFM and a hot wire flow meter TSI to measure the Q v o f m with port X4-1 and gas temperature T with port X4-2.
The circuits on the instrumentation board, as shown in Figure 7, demonstrate the use of 0.1 uF capacitances to filter out high-frequency signals that could affect the measurements. The differential pressure sensor SPD2 in Figure 7c, to which the P 1 and P 2 pressure taps of the VOFM are connected, is the SSCDRRN100MDAA5 model, which has an operating range of ±10 kPa (±1.45 PSI) with an accuracy of 0.25% [19]. The PCB in Figure 7d displays three circuits similar to those in Figure 7c. In SPD1 and SPD4, a 3.3 V-supplied differential pressure sensor can be installed as a backup in case of SPD2 failure. In SPD3, an HSCMANN100PGAA3 sensor was installed to measure the supply pressure to the proportional valve, which has an operating range of 0 kPa to 689.5 kPa (0 PSIG–100 PSIG). On the other hand, the hot wire flowmeter used as a reference instrument is a TSI-840120 Rev:E model with an accuracy of ±2.5% of the reading.
A program for Arduino Due corrects the flow error caused by temperature, oxygen (O2) percentage in the air, and electrical noise. This program is based on the correction equation provided by the manufacturer of the hot wire sensor and incorporates an averaging filter routine that takes 20 flow measurements. The flow error is calculated as E Q = Q s p Q v o f m , where Q s p is SetPoint flow value or desired flow value; it is used by a PID algorithm implemented on the Arduino Due to correct the errors by comparing the actual flow value measured by the hot wire sensor.

2.3. Monitoring Software

Lastly, a SCADA interface was developed using Indusoft Web Studio Educational® 8.1 software to operate the device and store the data. The software was installed on a laptop with the following specifications: Intel Core i5, 8 GB RAM, and 512 GB SSD storage. An RS232 serial communication between the Arduino Due and the application hosted on a computer was implemented. The communication driver allows receiving sensor measurement information at a baud rate of 115,200 bps. The driver also enables parameterizing the SetPoint flow value to be delivered by the proportional valve. Figure 8 shows the adjustment parameters, the selection buttons for MANUAL or AUTOMATIC operation, and the readings from the sensors on the developed interface.
If the behavior of a VOFM is to be evaluated, the MANUAL operation mode is selected in the SCADA. In this mode, it is necessary to provide the parameters (inspiration time T i , respiratory rate (FR), maximum inspiratory pressure (PIP), positive end-expiratory pressure (PEEP), and tidal volume (Vc) to calculate the desired flow value for a proportional valve opening. Simultaneously, the information collected (pressure drop across the MOV, supply network pressure, gas temperature, humidity, O2 percentage, and the hot-wire meter flow value) by the monitoring software is stored in a TXT file as a historical record of the system’s operation so that it can be subsequently used by other software or databases.
If the operation of a mechanical ventilation system is to be simulated, the AUTOMATIC operation mode is selected. This mode corresponds to constant-flow ventilation, limited by pressure and cycled by time.

3. Experiment

With the aim of assessing the performance of the VOFM, air coming from the proportional valve was passed through the VOFM to record the pressure drop and, subsequently, through the wire flow meter to determine the flow. In total, six variable orifice flow meters were evaluated, which were cut on Mylar sheets according to a factorial design. This design considered a width B = 15   m m with three different lengths ( L 1 = 15   m m , L 2 = 19.5   m m , and L 3 = 22.5   m m ) and two different thicknesses ( e 1 = 0.1016   m m and e 2 = 0.1905   m m ). All VOFMs were stimulated with an airflow ranging from 0 L . m i n 1 to 90 L . m i n 1 , with increments of approximately 2 L . m i n 1 . For each airflow value Q v o f m , a total of n = 20 repetitions of the pressure drop P across the VOFM were recorded. The fluid used was air with an oxygen percentage of approximately 21% and with a temperature during the experiments of 25 °C ± 2 °C.
The measure of dispersion S P of the pressure drop P was calculated using Equation (5):
S P = P P ¯ 2 n 1                            
The deflection of the beam in the VOFM was validated by measuring the deflection using images and comparing the results with the theoretical deflection.

4. Results

4.1. Pressure Drop and Deflection

Since Equation (2) predicts that the deflection of the beam is directly proportional to the pressure drop, an experiment was conducted to experimentally evaluate this relationship in a beam with dimensions A1 = 15 mm and L2 = 19.5 mm. For this purpose, the deflection of the beam y r was measured using images as shown in Figure 9, and it was compared with the theoretical deflection y from Equation (2).
The results in Table 3 show that, for Q v o f m 31.0   L . m i n 1 , the error between the measured and calculated deflection is less than 5.4%, while, for Q v o f m 21.3   L . m i n 1 ), the error is higher than 5.4%. Typically, mechanical ventilation equipment for medical applications passes flow calibration tests if the error is below 10%.

4.2. Relationship between Pressure Drop P and Flow Rate Q v o f m

Figure 10a displays the results obtained for three lengths of a beam with thickness e 1 = 0.1016   m m , while Figure 10b shows the results for a thickness of e 2 = 0.1905   m m . In the mentioned figures, it is evident that at low flows, the pressure drop with respect to flow is not linear, while if the flow is increased, the behavior becomes linear.
The transition between linear and non-linear behaviors was obtained through linear regression analysis and residual error analysis. For linear regression, a mathematical model of the form P ¯ = γ 0 + γ 1 Q v o f m , was used, and a set of data far from the non-linear region was considered. In this case, we considered the data for which Q v o f m 30   L . m i n 1 . Table 4 and Table 5 show an R 2 close to one for all dimensions, indicating that the mathematical model predicts a strong relationship between the pressure drop and the flow.
Figure 11 shows the residual error E r = P P ¯ , where P is the experimental pressure drop and y P ¯ is the expected value from the regression. The transition to linear behavior is depicted in Figure 11 with dashed vertical lines. On the left side of the lines, the behavior of P with respect to Q v o f m is non-linear and similar to gas flow behavior through a fixed orifice plate.
On the right side, the linear regression model exhibits increasing heteroscedasticity. This behavior is associated with turbulent flow regimes that increase variance. Table 6 presents the transition flow values obtained from Figure 11, where the transition between the non-linear and linear zones occurs. Figure 12 shows the transition flow rate as a function of the area A e .

5. Discussion

Considering the previous results, we can now identify more clearly how the length of the VOFM beam influences the pressure drop behavior at different flows. The experiments show two zones: a non-linear zone and a linear behavior zone. In the non-linear zone, which is present at low flows, the pressure difference exerted on the plate is insufficient to overcome the material stiffness, meaning that gas flows through the area of the cut slots A e = ( 2 L + B ) t without causing appreciable deflection. As a result, the pressure drop at low flows in the variable orifice meter is proportional to the square of the flow rate, similar to the behavior of a fixed orifice plate.
The area A e can be significantly reduced by reducing the thickness of the cut slot t more than by variations in L o B because of the following:
A e t A e L
The transition flow is also affected by the thickness. According to the deflection equation for a cantilever beam, the stiffness of the beam increases as its thickness increases. An increase in the thickness of the VOFM results in less deflection and a smaller opening for fluid passage, increasing the flow resistance. Therefore, a higher flow is required for the transition if the thickness is increased. This finding highlights the importance of thickness in the design and operation of variable orifice meters.
Even though the linear regression model showed values close to one for the coefficient of determination R2, the measure of dispersion S P of the measurements increased proportionally with the flow rate, indicating that the errors are not constant. This suggests that the obstruction generates vortices and turbulence through the orifice plate. These vortices and turbulence affect the velocity distribution and flow structure, which can modify the Reynolds number in the region near the orifice plate.
We calculated the Reynolds number for the circular pipe upstream of the VOFM us-ing the equations for a circular section to verify any potential turbulence influence that could affect the VOFM beam. According to the Reynolds number, we found that, at a flow rate of around 35 L . m i n 1 , the transition from laminar to turbulent flow begins, and at a flow rate of 67 L . m i n 1 , the flow regime will be turbulent. For this reason, we can assert that the increased variance in the pressure drop is due to the flow regime. However, we did not calculate the Reynolds number directly at the VOFM section. Therefore, we consider it important to design mathematical models and conduct experiments to assess the flow regime precisely at the VOFM, considering the variable geometry. This would enable the optimization of the VOFM beam geometries for potential plate vibrations and, thus, enhance its accuracy as a flow measurement sensor.
For this reason, we can assert that the increased variance in pressure drop is due to the flow regime. However, we did not calculate the Reynolds number directly at the VOFM section. Therefore, we consider it important to design mathematical models and conduct experiments to assess the flow regime precisely at the VOFM, considering the variable geometry. This would enable the optimization of the VOFM beam geometries for potential plate vibrations and, thus, enhance its accuracy as a flow measurement sensor.
With the device, we have discovered that increasing the length of the beam and reducing its thickness can enhance sensitivity. However, geometry changes need to be adjusted according to the requirements of mechanical ventilation. This is crucial because the treatment for mechanical ventilation varies significantly between patients such as human neonates and adults. This is even more critical if the VOFM is applied to pets or animals, as the variability in ventilation requirements is extensive.

6. Conclusions

This study has introduced an innovative, cost-effective device specifically designed for the assessment of Variable Orifice Flow Meters (VOFM) of the beam type, commonly used in medical mechanical ventilation applications. The device consists of three mod-ules: an experimental setup, an electronic control board for data acquisition and control of the equipment, and software for running the system and processing data using SCADA.
The measurement device can record the following: the supply pressure from 0 to 100 psi; the gas temperature from 0 °C to 50 °C; the oxygen concentration from 0 to 100%; the relative hu-midity from 0 to 100%; the pressure drops across the plate from −0.1 bar to 0.1 bar; gas flow rate from 0 to 100 L . m i n 1 ; and the proximal pressure to the patient from −0.1 bar to 0.1 bar. On the other hand, software parameterization allows the simulation of operating conditions of a mechanical ventilator for living beings by defining parameters such as inspiration time (Ti), tidal volume (Vi), and respiratory frequency (FR).
This devised system and its methodological approach have demonstrated their efficacy as invaluable tools for the precise and systematic evaluation of Variable Orifice Flow Meters in medical applications. The findings not only enhance our understanding of the factors influencing pressure drop in these configurations but also hold vital implications for the optimization and design of mechanical ventilation systems.
Finally, to utilize VOFM in mechanical ventilation and functional respiratory monitoring, it is necessary to perform the following: investigate the impact of various geometries, minimize turbulence formation in the orifice plate, and utilize digital signal filtering techniques for pressure signals.

Author Contributions

Conceptualization, W.P.M. and J.F.A.L.; methodology, W.P.M. and J.F.A.L.; software, W.P.M.; validation, J.F.A.L. and J.V.G.; formal analysis, W.P.M. and J.F.A.L.; investigation, W.P.M. and J.F.A.L.; resources, W.P.M. and J.F.A.L.; data curation, W.P.M. and J.F.A.L.; writing—original draft preparation, W.P.M.; writing—review and editing, W.P.M. and J.F.A.L.; visualization, W.P.M. and J.F.A.L.; supervision, W.P.M. and J.F.A.L.; project administration, W.P.M. and J.F.A.L.; funding acquisition, W.P.M. and J.F.A.L. All authors have read and agreed to the published version of the manuscript.

Funding

This research received funding from Universidad Tecnológica de Pereira.

Data Availability Statement

No data other than those presented in the paper are available.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Hardy, J.E.; Hylton, J.O.; McKnight, T.E.; Remenyik, C.J.; Ruppel, F.R. Flow Measurement Methods and Applications; John Wiley & Sons: Hoboken, NJ, USA, 1999; ISBN 978-0-471-24509-4. [Google Scholar]
  2. Baker, R.C. Flow Measurement Handbook: Industrial Designs, Operating Principles, Performance, and Applications; Cambridge University Press: Cambridge, UK, 2000. [Google Scholar]
  3. Shustrova, M.; Fafurin, A.; Baytimirov, A. The Question of Dynamic Gas Flow Measuring. In Proceedings of the 2019 International Conference on Industrial Engineering, Applications and Manufacturing (ICIEAM), Sochi, Russia, 25–29 March 2019; pp. 1–6. [Google Scholar]
  4. Tomaszewska-Wach, B.; Rzasa, M. A Correction Method for Wet Gas Flow Metering Using a Standard Orifice and Slotted Orifices. Sensors 2021, 21, 2291. [Google Scholar] [CrossRef] [PubMed]
  5. BS EN ISO 5167-1; Measurement of Fluid Flow by Means of Pressure Differential Devices Inserted in Circular Cross-Section Conduits Running Full: Part 1. General Principles and Requirements. British Standards Institution: London, UK, 2021.
  6. Chavarri, J.A.C.; Ruiz, C.G.R.; Gómez Amador, A.M.; Cardenas, B.J.M.A.; Anaya, S.C.; Lozano Jauregui, J.H.; Hinostroza, A.T.; Jiménez de Cisneros y Fonfría, J.J. Mathematical Analysis of a Low Cost Mechanical Ventilator Respiratory Dynamics Enhanced by a Sensor Transducer (ST) Based in Nanostructures of Anodic Aluminium Oxide (AAO). Mathematics 2022, 10, 2403. [Google Scholar] [CrossRef]
  7. Kudrna, P. Effects of Orifice Plate as a Flow Sensor in an Endotracheal Tube on the Ability of Elimination CO2 during HFJV—Animal Study. Appl. Sci. 2021, 11, 4210. [Google Scholar] [CrossRef]
  8. Kwon, J.; Kang, E.; Shin, S.; Lee, B.; Ko, M.; Kim, S.; Lee, S. Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit. Medicina 2023, 59, 965. [Google Scholar] [CrossRef] [PubMed]
  9. Vales, S.B.; Gómez, L.R. Fundamentos de la Ventilación Mecánica; Marge Books: Barcelona, Spain, 2023; ISBN 978-84-15340-50-8. [Google Scholar]
  10. Eckhout, G.V.; Bhatia, S. Another Cause of Difficulty in Ventilating a Patient. J. Clin. Anesth. 2003, 15, 137–139. [Google Scholar] [CrossRef] [PubMed]
  11. Nicolò, A.; Massaroni, C.; Schena, E.; Sacchetti, M. The Importance of Respiratory Rate Monitoring: From Healthcare to Sport and Exercise. Sensors 2020, 20, 6396. [Google Scholar] [CrossRef] [PubMed]
  12. Duprez, F.; Barile, M.; Bonus, T.; Cuvelier, G.; Ollieuz, S.; Mashayekhi, S.; Legrand, A. Accuracy of Medical Oxygen Flowmeters: A Multicentric Field Study. Health 2014. [Google Scholar] [CrossRef]
  13. Kazmina, A.S.; Makukha, V.K. Testing of the Airflow Sensor Honeywell AWM42300V for Using It in Spirometer. In Proceedings of the 2017 IEEE Conference of Russian Young Researchers in Electrical and Electronic Engineering (EIConRus), St. Petersburg and Moscow, Russia, 1–3 February 2017; pp. 36–39. [Google Scholar]
  14. Pintavirooj, C.; Maneerat, A.; Visitsattapongse, S. Emergency Blower-Based Ventilator with Novel-Designed Ventilation Sensor and Actuator. Electronics 2022, 11, 753. [Google Scholar] [CrossRef]
  15. Chen, X.; Liu, C.; Yang, D.; Liu, X.; Hu, L.; Xie, J. Highly Accurate Airflow Volumetric Flowmeters via PMUTs Arrays Based on Transit Time. J. Microelectromech. Syst. 2019, 28, 707–716. [Google Scholar] [CrossRef]
  16. Tardi, G.; Massaroni, C.; Saccomandi, P.; Schena, E. Experimental Assessment of a Variable Orifice Flowmeter for Respiratory Monitoring. J. Sensors 2015, 2015, e752540. [Google Scholar] [CrossRef]
  17. Montagna, F.; Schena, E.; Massaroni, C.; Caciotti, C.; Presti, D.L.; Silvestri, S. Influence of the Length of Lead Lines on the Response of a Variable Orifice Meter: Analysis of Sensitivity and Settling Time. In Proceedings of the 2018 IEEE International Symposium on Medical Measurements and Applications (MeMeA), Rome, Italy, 11–13 June 2018; pp. 1–6. [Google Scholar]
  18. Osborn, J.J. Variable Orifice Gas Flow Sensing Head. U.S. Patent US4083245A, 11 April 1978. [Google Scholar]
  19. Ciobanu, C.I.; Schaeffer, D. Variable Orifice Flow Sensor. U.S. Patent US5970801A, 26 October 1999. [Google Scholar]
  20. Billette, R.J.; Zias, A.R. Flow Meter. U.S. Patent US4006634, 8 February 1977. [Google Scholar]
  21. Stupecky, J. Variable Area Obstruction Gas Flow Meter. U.S. Patent US5038621A, 13 August 1991. [Google Scholar]
  22. Beer, R.H.M.D.; Kuypers, M.H.; Holland, J.; Schotte, H. Flow Resistance Tube for Gas Flow Meter. E.P. Patent EP0331772A1, 13 September 1989. [Google Scholar]
  23. Bonassa, J. Flow Sensor with Double Obstruction. U.S. Patent US20090064794A1, 21 September 2010. [Google Scholar]
  24. Friberg, H.; Daescher, J. Respiratory Flow Sensor. U.S. Patent US10905357B2, 2 February 2021. [Google Scholar]
  25. Ganseman, J.; Ayoubi, A.; Ayoubi, N.; Bellaire, T.; Brodkin, I.; Dosil, V.; Findlay, J.; Halwani, F.; Willms, A. Spirometer Breathing Tube with Compound Membrane. U.S. Patent CA2803645A1, 5 January 2012. [Google Scholar]
  26. Franetzki, M. Flow Measuring Device. U.S. Patent US3989037, 2 November 1976. [Google Scholar]
  27. Funaki, T.; Kawashima, K.; Fujita, T.; Kagawa, T. Characteristic Analysis of Variable Orifice Flow Meter. Proc. JFPS Int. Symp. Fluid Power 2002, 2002, 83–88. [Google Scholar] [CrossRef]
  28. Wang, Q.; Cui, J.; Tang, Y.; Pang, L.; Chen, K.; Zhang, B. Research on a Precision Calibration Model of a Flexible Strain Sensor Based on a Variable Section Cantilever Beam. Sensors 2023, 23, 4778. [Google Scholar] [CrossRef] [PubMed]
  29. Ruan, C.; Chen, M.; Yu, Y.; Zhang, Y.; Wang, J.; Zhang, Z.; Yang, J.; Zhu, S.; Qu, B. Research on Very-Low-Frequency Hydroacoustic Acoustic Velocity Sensor Based on DFB Fiber Laser. Photonics 2023, 10, 463. [Google Scholar] [CrossRef]
  30. Gutiérrez Muñoz, F. Ventilación Mecánica. Acta Médica Peru. 2011, 28, 87–104. [Google Scholar]
  31. Williams, D.; Flory, S.; King, R.; Thornton, M.; Dingley, J. A Low Oxygen Consumption Pneumatic Ventilator for Emergency Construction during a Respiratory Failure Pandemic. Anaesthesia 2010, 65, 235–242. [Google Scholar] [CrossRef] [PubMed]
  32. Capítulo 7. Manejo Del Paciente COVID-19 en UCI. Available online: https://amci.org.co/images/consenso/Capitulo_7_Manejo_del_paciente_COVID-19_en_UCI.pdf (accessed on 10 May 2023).
  33. Chica-Meza, C.; Peña-López, L.A.; Villamarín-Guerrero, H.F.; Moreno-Collazos, J.E.; Rodríguez-Corredor, L.C.; Lozano, W.M.; Vargas-Ordoñez, M.P. Cuidado respiratorio en COVID-19. Acta Colomb. Cuid. Intensivo 2020, 20, 108–117. [Google Scholar] [CrossRef]
  34. Información Sobre Prototipos de Respiradores. Pruebas de Seguridad Y Requisitos de Investigacion Clinica. Available online: https://www.aemps.gob.es/informa/notasInformativas/productosSanitarios/2020/NI-PS_11-2020-respiradores-pruebas-investigaciones.pdf?x33378 (accessed on 15 May 2023).
  35. Specification for Ventilators to Be Used in UK Hospitals during the Coronavirus (COVID-19) Outbreak. Available online: https://www.gov.uk/government/publications/specification-for-ventilators-to-be-used-in-uk-hospitals-during-the-coronavirus-covid-19-outbreak (accessed on 6 July 2023).
  36. Center for Devices and Radiological Health. Enforcement Policy for Ventilators and Accessories and Other Respiratory Devices during the Coronavirus Disease 2019 (COVID-19) Public Health Emergency. Available online: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/enforcement-policy-ventilators-and-accessories-and-other-respiratory-devices-during-coronavirus (accessed on 6 July 2023).
  37. Especificaciones Técnicas de Dispositivos Médicos para la Gestión de Casos de COVID-19 en los Servicios de Salud—OPS/OMS|Organización Panamericana de la Salud. Available online: https://www.paho.org/es/documentos/especificaciones-tecnicas-dispositivos-medicos-para-gestion-casos-covid-19-servicios (accessed on 6 July 2023).
  38. Beer, R.H.M.D.; Kuypers, M.H.; Holland, J.; Schotte, H. Strömungswiderstandsrohr für Gasströmungsmesser. Patent No. EP0331773A1, 13 September 1989. [Google Scholar]
Figure 1. Restriction due to section change. Source: Authors.
Figure 1. Restriction due to section change. Source: Authors.
Inventions 08 00110 g001
Figure 2. Variable orifice flow meter: (a) geometry of variable orifice flow meter; (b) deflected beam. Source: Authors.
Figure 2. Variable orifice flow meter: (a) geometry of variable orifice flow meter; (b) deflected beam. Source: Authors.
Inventions 08 00110 g002
Figure 3. Measurement device: (a) schematic of a measurement device; (b) experimental setup. Source: Authors.
Figure 3. Measurement device: (a) schematic of a measurement device; (b) experimental setup. Source: Authors.
Inventions 08 00110 g003
Figure 4. VOFM Container. Source: Authors.
Figure 4. VOFM Container. Source: Authors.
Inventions 08 00110 g004
Figure 5. Electronic system: (a) four-level arrangement: (b) Arduino Due processing board for processing information, communication, and controlling the operation sequence; (c) power board; (d) proportional valve control board; (e) instrumentation board. Source: Authors.
Figure 5. Electronic system: (a) four-level arrangement: (b) Arduino Due processing board for processing information, communication, and controlling the operation sequence; (c) power board; (d) proportional valve control board; (e) instrumentation board. Source: Authors.
Inventions 08 00110 g005
Figure 6. Control circuit for the proportional solenoid valve: (a) electronic diagram of proportional solenoid valve; (b) PCB of the control circuit for the proportional solenoid valve. Source: Authors.
Figure 6. Control circuit for the proportional solenoid valve: (a) electronic diagram of proportional solenoid valve; (b) PCB of the control circuit for the proportional solenoid valve. Source: Authors.
Inventions 08 00110 g006
Figure 7. Circuit for the pressure sensors: (a) electronic diagram to hot wire flowmeter; (b) electronic diagram to pressure sensors SPD1 and SPD4; (c) electronic diagram to pressure sensor SPD2; (d) PCB for the pressure sensors and hot wire flow meter. Source: Authors.
Figure 7. Circuit for the pressure sensors: (a) electronic diagram to hot wire flowmeter; (b) electronic diagram to pressure sensors SPD1 and SPD4; (c) electronic diagram to pressure sensor SPD2; (d) PCB for the pressure sensors and hot wire flow meter. Source: Authors.
Inventions 08 00110 g007
Figure 8. Monitoring software interface. Source: Authors.
Figure 8. Monitoring software interface. Source: Authors.
Inventions 08 00110 g008
Figure 9. Photograph of the beam deflection. Source: Authors.
Figure 9. Photograph of the beam deflection. Source: Authors.
Inventions 08 00110 g009
Figure 10. Results for three lengths of a beam pressure drop P and flow rate Q v o f m : (a) thickness e1 = 0.1016 mm; (b) thickness e2 = 0.1905 mm. Source: Authors.
Figure 10. Results for three lengths of a beam pressure drop P and flow rate Q v o f m : (a) thickness e1 = 0.1016 mm; (b) thickness e2 = 0.1905 mm. Source: Authors.
Inventions 08 00110 g010
Figure 11. Residual errors for VOFMs with different lengths and thickness: (ac) with a thickness e1 = 0.1016 mm; (df) with a thickness e2 = 0.1905 mm. Source: Authors.
Figure 11. Residual errors for VOFMs with different lengths and thickness: (ac) with a thickness e1 = 0.1016 mm; (df) with a thickness e2 = 0.1905 mm. Source: Authors.
Inventions 08 00110 g011
Figure 12. Transition flow as a function of A e . Source: Authors.
Figure 12. Transition flow as a function of A e . Source: Authors.
Inventions 08 00110 g012
Table 1. Displays the beam shapes used in the patents.
Table 1. Displays the beam shapes used in the patents.
Multiple beamsInventions 08 00110 i001
US9931056B2
Inventions 08 00110 i002
US4006634A
Inventions 08 00110 i003
US3989037A
Inventions 08 00110 i004
US4989456A
Single beamInventions 08 00110 i005
US5970801A
Inventions 08 00110 i006
US4083245
Inventions 08 00110 i007
EP-0331773A1
Inventions 08 00110 i008
US10905357B2
Table 2. Proposed materials in patents for VOFMs.
Table 2. Proposed materials in patents for VOFMs.
PatentMaterialThickness
US4083245ASynthetic rubber0.9 mm
US4006634AMaterial with Young’s modulus 3 × 107 psi-
US9931056B2 [25]Flexible material-
US3989037A [26]Polymer0.05 mm
EP0331773A1Mylar®50 μm y 125 μm
US5970801AMetallic-
US4989456 AStainless steel series 3000.0254 mm (0.001 in)
US7798016B2Mylar®0.1 mm–0.15 mm
US10905357B2 Mylar®0.127 mm–0.1778 mm
(0.005 in–0.007 in)
Table 3. Results of the beam deflection.
Table 3. Results of the beam deflection.
Q v o f m   ( L . m i n 1 ) P   ( P a ) y r   ( m m ) y   ( m m ) Error
4.20.40.10.00495.6%
11.913.80.20.16716.6%
21.340.30.40.48922.2%
31.069.50.80.8435.4%
40.6102.31.21.2413.4%
51.2129.71.51.5734.8%
61.2158.81.91.9261.4%
71.6189.42.22.2984.4%
81.2230.22.72.7922.6%
89.3252.73.13.0651.1%
Table 4. Linear regression for thickness e1 = 0.1016 mm.
Table 4. Linear regression for thickness e1 = 0.1016 mm.
A (mm)L (mm) γ γ 0 R2
15155.5045−43.744640.9990
1519.53.0528−24.752420.9989
1522.51.1228−4.9817510.9892
Table 5. Linear regression for thickness e2 = 0.1905 mm.
Table 5. Linear regression for thickness e2 = 0.1905 mm.
A (mm)L (mm) γ γ 0 R2
15158.0832−95.3990.9962
1519.54.3410−67.6390.9988
1522.52.8709−38.4350.9969
Table 6. Transition flows.
Table 6. Transition flows.
e 1   =   0.1016   m m e 2   =   0.1905   m m
L1 = 15 mm15 L . m i n 1 30 L . m i n 1
L2 = 19.5 mm17 L . m i n 1 36 L . m i n 1
L3 = 22.5 mm20 L . m i n 1 40 L . m i n 1
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Prado Martínez, W.; Arroyave Londoño, J.F.; Vásquez Gómez, J. Design and Construction of a Device to Evaluate the Performance of Variable Orifice Flow Meters (VOFM). Inventions 2023, 8, 110. https://doi.org/10.3390/inventions8050110

AMA Style

Prado Martínez W, Arroyave Londoño JF, Vásquez Gómez J. Design and Construction of a Device to Evaluate the Performance of Variable Orifice Flow Meters (VOFM). Inventions. 2023; 8(5):110. https://doi.org/10.3390/inventions8050110