Abstract

Conception and Development of Microfabricated Elements for Microfluidic Analytical Devices †

by
Irene Lara-Ibeas
1,*,
Christina Andrikopoulou
1,
Vincent Person
2,
Alireza Sharifi
2,3,
Dimitris Valougeorgis
3,
Stéphane Colin
4 and
Stéphane Le Calvé
1,2
1
Institute of Chemistry and Processes for Energy, Environment and Health (ICPEES), University of Strasbourg, 67087 Strasbourg, France
2
In’Air Solutions, 25 Rue Becquerel—Bâtiment 71A, 67087 Strasbourg, France
3
School of Engineering, Department of Mechanical Engineering, University of Thessaly, 38334 Volos, Greece
4
INSA Toulouse, Institute Clément Ader, 31400 Toulouse, France
*
Author to whom correspondence should be addressed.
Presented at the 5th International Symposium on Sensor Science (I3S 2017), Barcelona, Spain, 27–29 September 2017.
Proceedings 2017, 1(8), 841; https://doi.org/10.3390/proceedings1080841
Published: 4 December 2017
Portability and low energy consumption are of a great importance for on-site real-time monitoring of indoor air quality. Therefore, this work is focused on the miniaturization of the elements which are the bulkiest and require the highest energy supply in analytical devices. More specifically, micro-fabrication techniques were deployed so that a pre-concentrator and a gas chromatography micro-column with integrated heaters and temperature sensors were developed on silicon wafers of small dimensions of 20 mm × 12 mm and 50 mm × 50 mm, respectively.
Heating elements and sensors were fabricated by thermal evaporation of titaniumand gold. In the GC micro-column, four resistances of 82 Ω allow a temperature increase up to 200 °C in 150 s whereas two 430 Ω resistances function as sensors. In the case of the pre-concentrator, three different designs containing heaters of 50–446 Ω and sensors of 117–555 Ω were proposed in order to investigate the most appropriate configuration to achieve a temperature increase up to 250 °C in 10 s.
Furthermore, a study of wet etching kinetics of Si <100> with KOH (40%) was performed to evaluate its suitability for the fabrication of micro-channels. Silicon etching rates were determined to be between 14.7 and 63.1 µm/h at different temperatures from 50–80 °C. Afterwards, three GC micro-columns of 1, 2 and 3 m in length containing 106 and 146 μm deep microchannels were etched using KOH (40%) at 70 °C, thus demonstrating that wet etching is an alternative and cheaper technique to fabricate micro-columns.

Acknowledgments

This work was supported through the European project ELCOD. The ELCOD project was implemented as a part of the INTERREG V Oberrhein/Rhin Supérieur program and was supported by the European Regional Development Fund (ERDF) and the co-financed project partners Region Grand Est in France and the countries of Baden-Württemberg and Rhineland-Palatinate. This project has also received funding from the Clean Sky 2 Joint Undertaking under the European Union’s Horizon 2020 research and innovation programme under grant agreement No 687014. Microfabrication techniques during this work were deployed in the nanotechnology platform STnano (IPCMS – CNRS, University of Strasbourg).

Share and Cite

MDPI and ACS Style

Lara-Ibeas, I.; Andrikopoulou, C.; Person, V.; Sharifi, A.; Valougeorgis, D.; Colin, S.; Calvé, S.L. Conception and Development of Microfabricated Elements for Microfluidic Analytical Devices. Proceedings 2017, 1, 841. https://doi.org/10.3390/proceedings1080841

AMA Style

Lara-Ibeas I, Andrikopoulou C, Person V, Sharifi A, Valougeorgis D, Colin S, Calvé SL. Conception and Development of Microfabricated Elements for Microfluidic Analytical Devices. Proceedings. 2017; 1(8):841. https://doi.org/10.3390/proceedings1080841

Chicago/Turabian Style

Lara-Ibeas, Irene, Christina Andrikopoulou, Vincent Person, Alireza Sharifi, Dimitris Valougeorgis, Stéphane Colin, and Stéphane Le Calvé. 2017. "Conception and Development of Microfabricated Elements for Microfluidic Analytical Devices" Proceedings 1, no. 8: 841. https://doi.org/10.3390/proceedings1080841

APA Style

Lara-Ibeas, I., Andrikopoulou, C., Person, V., Sharifi, A., Valougeorgis, D., Colin, S., & Calvé, S. L. (2017). Conception and Development of Microfabricated Elements for Microfluidic Analytical Devices. Proceedings, 1(8), 841. https://doi.org/10.3390/proceedings1080841

Article Metrics

Back to TopTop