Next Article in Journal
Instantaneous Best Integer Equivariant Position Estimation Using Google Pixel 4 Smartphones for Single- and Dual-Frequency, Multi-GNSS Short-Baseline RTK
Next Article in Special Issue
Design of Dust-Filtering Algorithms for LiDAR Sensors Using Intensity and Range Information in Off-Road Vehicles
Previous Article in Journal
Measuring Indoor Occupancy through Environmental Sensors: A Systematic Review on Sensor Deployment
Previous Article in Special Issue
Commercial Off-the-Shelf Components (COTS) in Realizing Miniature Implantable Wireless Medical Devices: A Review
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Comparison of Three Prototypes of PPG Sensors for Continual Real-Time Measurement in Weak Magnetic Field †

Institute of Measurement Science, Slovak Academy of Sciences, 841 04 Bratislava, Slovakia
*
Author to whom correspondence should be addressed.
This study is an extended version of the paper entitled “Two Realizations of the Wearable PPG Sensor Working in Reflectance Mode for Measurement in Weak Magnetic Field” presented at 8th International Electronic Conference on Sensors and Applications (ECSA-8), 1–15 November 2021.
Sensors 2022, 22(10), 3769; https://doi.org/10.3390/s22103769
Submission received: 15 April 2022 / Revised: 11 May 2022 / Accepted: 14 May 2022 / Published: 16 May 2022

Abstract

:
This paper is focused on investigation of three developed prototypes of sensors based on the photoplethysmography (PPG) principle for continual measurement of the PPG signal in the magnetic field environment with the inherent radiofrequency and electromagnetic disturbance. The tested prototypes differ in the used optical part of the PPG sensor and their working mode, control unit, power supply, and applied Bluetooth (BT) communication methods. The main aim of the current work was motivated by finding suitable and universal parameter settings for PPG signal real-time recording in different working mode conditions. Comparative measurements in laboratory conditions by certified commercial pulse oximeter and blood pressure monitor (BPM) devices show good stability and proper accuracy of finally determined heart rate values. The supplementary investigation certifies the necessity of the placement of the pressure cuff of the BPM device on the opposite arm than the tested PPG sensor. Measurement experiments inside the scanning area of the running weak field magnetic resonance scanner verify proper function and practical usability of sensed PPG signals for further processing and analysis in all three prototype cases. Additional testing shows that the BT transmission in the scanning area has no visible influence on the quality of the finally obtained scanner images.

1. Introduction

During a non-invasive medical examination in a magnetic resonance imaging (MRI) scanner, patients are exposed to the mechanical vibration of the gradient-coil system and the resulting acoustic noise. These factors have a negative influence on the subject’s cardiovascular system and evocate a mental stress [1] that can be detected and evaluated by continual monitoring of blood pressure (BP) and heart rate (HR) parameters. This monitoring is often carried out using wearable sensors based on the photoplethysmography (PPG) [2,3]. The PPG signals can be sensed and recorded from various body positions: finger, ear, wrist, arm, neck, nose, chest, and forehead [4], so different practical realizations of PPG sensors are necessary. Various parameters identifying the physiological and mental stress can be derived from the PPG signal: pulse transit time [2], pulse wave velocity [5], blood oxygen saturation, cardiac output [6], etc. MRI tomography is also used to obtain vocal tract shapes during the articulation of speech sounds for the articulatory synthesis [7]. Here, the stress-evoked vocal cord tension has an influence on the recorded speech signal [8] by modifying its supra-segmental and spectral features; therefore, it can bring about errors and inaccuracy in calculation of 3D models of the human vocal tract [9]. Analysis, mapping, detection, and evaluation of this negative effect were the basis of our previous [10,11] as well as present [12] research investigation. For this reason, three prototypes of PPG sensors with real-time Bluetooth (BT) transmission were subsequently developed. Due to the static magnetic field inside the MRI scanner accompanied by an inherent radiofrequency (RF) and electromagnetic (EM) disturbance [13], a special realization of sensors must be used for PPG signal pickup.
Our present work was practically motivated by seeking suitable and universal parameter settings for PPG signal real-time recording in different working mode conditions including proper function in a weak magnetic field environment. Before starting this development, we had already tested commercial fitness bracelets and smart watches. These types of devices work in the reflectance mode and product HR parameter values that are not very precise and are not designed for function in these conditions (in addition with inherent RF and EM disturbance). Next, majority of these commercial devices are not able to send the PPG signal samples to an external device (the HR values only). For the purpose of this study, we need to evaluate fast and slow changes in the heart rate as well as changes in the width of systolic/diastolic pulses to determine special PPG signal properties for stress or pain detection. A possible solution seems to be using an oximeter device working in the transmission mode—more expensive types enable real-time visualization of the sensed PPG wave. These devices have better resistance to interference, but their practical problem originates from the experimental arrangement which must be fulfilled: a PPG/oximeter device placed inside the scanning area of the MRI tomograph must communicate wirelessly with an external device which is located outside the shielding cage of the whole MRI equipment. All oximeters, which we had at our disposal for preliminary testing, had a poor, unstable BT data connection thru the metal cage, often the connection cannot be established whatsoever. Therefore, we finally decided to develop new sensor prototypes which can solve all problems and limitation conditions. Prior to full experimental usage of the developed prototypes of PPG sensors, the evaluation of precision and functional stability by comparative measurements must be performed in the laboratory as well as in real conditions inside the running MRI device. In our last work presented in [12], two prototypes of wearable PPG sensors working in a reflectance mode were compared. The comparative measurements performed with the certified commercial pulse oximeter (POXI) device in the normal laboratory conditions confirm sufficient stability and precision of the determined HR values. Additional analysis of the influence of the used smoothing method on the finally obtained PPG signal properties was also realized in the previous work. First-step measurements inside the scanning area of the working MRI device with low magnetic field verify practical functionality of both PPG sensors and their potential usability in further measurements.
The current extension (improvements) consists in the basic fact that this article compares three developed PPG sensor prototypes where the third one works in a transmittance mode and is practically unwearable. It means that all three investigated prototypes differ in the used optical part of the PPG sensor working in reflectance/transmittance modes, control units based on Arduino boards [14], power supplies, used BT modules, and communication standards—but all three prototypes enable functioning in the low magnetic field environment with RF and EM disturbance. The first part of the current work is focused on description and comparison consisting of following 4 points: (1) characterization of types of PPG waveforms and their basic features; (2) detailed description of a method practically used for real-time calculation of HR values and determination of PPG signal properties enabling their statistical analysis in the frame of post-processing operations; (3) summarization of common characteristics and differences of three tested PPG sensors; (4) basic description of created application for Windows platform which controls real-time sensing of the PPG wave, signal preprocessing (filtering), determination of HR values, and final storing in a Wave format. In the second part, this study comprises following 6 measurement experiments: (1) measurement of the sensor’s mean direct current using different power supply of 5/3.7 V in the 3 operation modes; (2) the calibration measurement of determined HR values from PPG signals sensed by all 3 tested PPG sensors together with parallel measurements by a certified blood pressure monitor (BPM) and OXI devices; (3) an analysis of necessity of proving a pressure effect of the inflated BPM cuff by its quantification and mapping during the simultaneous measurement of the PPG signal consecutively picked up from fingers of both hands; (4) testing of the stability and quality of BT connection between the tested PPG sensor located inside the scanning area of the MRI device and the control device located outside the shielding cage; (5) an analysis of the influence of the BT transmission in the scanning area on the quality of the finally obtained MR images; (6) a verification of practical functionality of all 3 tested PPG sensor prototypes in the scanning area of the running open-air MRI scanner based on a static field with magnetic induction up to 0.2 T.

2. Methods

2.1. Photoplethysmography—Types, Basic Properties, and Principles of Sensing

The photoplethysmography is a noninvasive optical method used to monitor pulsatile blood flow in arteries. A typical PPG waveform consists of two components: a non-pulsatile or direct current (DC) component originates from light absorption in the non-pulsating tissue. The superimposed pulsatile or alternating current (AC) component is generated due to light absorption in the pulsatile arterial blood and synchronously varies with the beating of the heart, so it is a simple tool for the heart rate (HR) monitoring. The magnitude of the AC component is much smaller—typically about 2% of the DC component. The PPG sensors can be placed at various body locations, for example, finger, ear, wrist, arm, chest, face. Apart from the HR and HR variability, other physiological parameters can be estimated from the PPG signal, for example, arterial blood oxygen saturation denoted as SpO2 when measured by the POXI device. Here two PPG signals are picked up and SpO2 is calculated using the ratio of normalized AC components of both signals [3].
A crucial step in the PPG pulse wave analysis is calculation of signal derivatives that may be used for determination of pulse wave features. The original pulse wave (raw photoplethysmogram—PTG) as well as its first derivative (called as velocity photoplethysmogram—VPG) and its second derivative (known as acceleration photoplethysmogram—APG) can be used to identify points of interest [15]. In all PPG signals, each photoplethysmogram cycle contains two local maxima representing systolic and diastolic peaks. These peaks provide valuable information about the pumping action of the heart—see an example in Figure 1 together with a detailed region of interest (ROI) with denoted heart pulse periods (THP).
As documented in Figure 1, the raw PPG signal does not seem suitable for local maxima determination. On the other hand, the first- and the second-order derivatives of this PPG signal are more informative due to more pronounced local extremes. The PPG waveform for the purpose of classification or prediction processes could also be used in the form of an image in a similar way as this approach is applied in the case of real-time electrocardiogram processing [16].
A conventional PPG sensor consists of a light emitter—light source (LS) and a photo detector (PD). Most commonly, the LS element is formed by one or more light-emitting diodes (LEDs). The PDs usually contain photodiodes, less commonly phototransistors or photoresistors. In the transmission mode, the LS and PD are placed on two opposite sides of the measured human tissue, for example, on a finger or an earlobe. In the reflection mode, the LS and PD measuring the intensity of the reflected and backscattered light are placed on the same side of the body surface and practically any skin area can be used here. The transmission PPG probes usually have a form of a finger ring (FR), a finger clip (FC), or an ear clip (EC)—see documentary photos in Figure 2a–c. The finger types are mostly used in medical applications—the FC is useful for pulse oximetry, as its clip exerts a pressure sufficient to eliminate the venous component in the tissue under the probe [17]. The EC realization is inconvenient for long-term measurement and recording but it is preferred in heart rate monitoring in situations when hands of the examined person cannot remain at rest [18].
The reflectance sensors, offering higher flexibility for PPG signal measurement from different locations on the body, are more suitable for the non-invasive wearable long-time monitoring devices. This type of PPG sensor is the most universal with practical placement mainly on fingers (typically fixed on a finger by an elastic/textile ribbon), as documented in Figure 2d, and on a wrist as various multi-functional smart watches or fitness bracelets. High-priced devices have three pieces of LEDs arranged in a triangle and multiple (three or more) sensing PHDs, as documented by a photo in Figure 2e. In addition, the PPG signal, together with the determined HR values, can be transmitted via the wireless connection to an external device (smartphone, tablet, etc.) for storage or next processing. On the other hand, there exist also specialized reflectance in-ear PPG sensors (placed in the auditory canal) which must be individually customized depending on the ear channel proportions [19]. These types of PPG sensors, mainly used in the systems for SpO2 continual monitoring, are usually connected to the POXI device by an electric cord [20].
The picked-up PPG signal has a typical amplitude modulation (ripple) with a partially linear trend (LT). A de-trending operation [15] is important in the case of a raw PPG signal, but it is usually not applied on VPG and APG signals. In addition, the PPG signal generally contains a noise component and can be often partially disturbed or degraded [19]. Therefore, the sensed PPG signal must be filtered prior to its processing and determination of HR values. A smoothing approach suitable for real-time processing of this relatively slow signal is a moving average (MA) filter with rectangular weighting. This basic smoothing method working in the time domain is used to filter out high frequency fluctuations in the PPG signal [24]. The proper choice of the MA window length is important—too long a window causes unacceptable decrease in the signal amplitude and loss of details, whilst a very short window does not deliver the required filtering effect. Analysis of MA filtering impact on the PPG signal properties was already analyzed in our previous work [12]. From our practical experiments, it follows that the Nx parameter representing the half length of the MA window should be selected in the interval of 4–12 samples (if the sampling frequency of 125 Hz is applied).
For an analysis of morphological features of individual PPG pulse waves, a relatively high sampling frequency fS (higher than 500 Hz) is required [25]. On the other hand, in continuous long-term screening and monitoring, sampling at lower frequencies may be used, for example, 128 Hz in a wrist-based wearable device for atrial fibrillation detection [26]. Commercial, wearable PPG sensors use typical sampling frequencies between 50 and 100 Hz [3]. For the purpose of this study, we determined and analyzed the energetic, time, and statistical parameters of the sensed PPG waves representing signal properties together with heart rate values. Using different fS from the investigated range cannot change the subsequently detected pulse period and the finally determined HR; only the precision of the systolic and diastolic peaks decreases for lower fS.

2.2. Determination of PPG Signal Properties and Heart Rate Calculation

In the frame of off-line post-processing and analysis of the PPG wave, the signal envelope can be determined by low pass filtering of the squared input signal. In our case, the upper and lower envelopes (EHI, ELOW) are determined by the root-mean-square (RMS) method using a sliding window. The basic energetic parameters: amplitude, signal range, and signal modulation, are determined next. These parameters are subsequently used in the heart rate calculation process. For calculation of the mean heart peak amplitude (HPAMPL), the maximum (LpMAX) and minimum (LpMIN) levels of these peaks must be determined together with the offset level of the PPG signal (LOFS). The absolute HPAMPL is calculated as:
H P AMPL = ( L p MAX + L p MIN ) / 2 L OFS ¯ ,
where the L OFS ¯ is the mean signal offset value.
Then, the relative heart peak ripple HPRIPP of a PPG wave is calculated using LpMAX and LpMIN level values previously determined:
HPRIPP = (LpMAXLpMIN)/LpMAX × 100 [%].
Finally, the relative PPG signal range (PPGRANGE) can be determined from the mean heart peak amplitude in percentage as:
PPGRANGE = HPAMPL/ADRES × 100 [%],
where the ADRES parameter depends on the resolution of the currently used A/D converter for digitalization of the analogue signal from the optical sensor and practically represents an actual numerical range of the processed PPG signal.
The HR determination algorithm must be simple and fast to be universally used in real-time processing of a continually sensed PPG signal. The HR may be estimated from the PPG signal by various methods [5,19,27]. The specialty of our approach to HR determination lies in the processing of a positive-valued PPG signal with the range of 0 to ADRES instead of a signal of both polarities within the range from −1 to 1. The currently proposed algorithm determines HR values from the picked-up PPG signal in the following way: (1) the PPG signal threshold LTHRESH is set; (2) the pulse wave is clipped binary; (3) the heart pulse periods THP are determined; (4) the HR values are calculated. The signal threshold LTHRESH is given by a fixed value LBASIC or by an adaptive threshold LADAPT. The LBASIC threshold value should cross the lower third of the PPG signal range predicted according to the measured PPG signal parameters. The LADAPT threshold value is specified as smaller than the smallest peak of all of the heart systolic pulses, see visualization in a documentary example in Figure 3.
The clipping operation produces a sequence cPPG (n) of values 1/0 corresponding to the input signal samples above/below LTHRESH as:
c PPG ( n ) = { 1 y ( n ) L TRESH 0 y ( n ) L TRESH   1     n     M ,
where M is a total number of processed samples of a PPG signal. The heart pulse periods THP in samples are determined from this clipped sequence as the length of two adjacent segments of ones T1P and zeros T0P (THP = T1P + T0P). Using the sampling frequency fs, HR values are calculated as:
HR = 60 × fs/THP [min−1].
The whole process of HR calculation is graphically demonstrated by an example in Figure 4.
To describe stability of the calculated HR values, a relative variability HRVAR is calculated based on the mean HR and the standard deviation HRSTD as:
HRVAR = (HRSTD/HRMEAN) × 100 [%].
For specifying the accuracy of the determined HR values from the PPG signal (HRPPG) a comparison with those measured by blood pressure monitor (BPM) and/or oximeter devices (HRCTRL) is used. When the control HR measurement by a BPM device is applied, the HRCTRL parameter represents one value HRBPM obtained for the whole measuring interval with duration TDUR. In the case of using an oximeter which produces continual values of HR during the whole measurement, the HRCTRL is defined as the mean value. In this way we obtain the differential parameter HRDIFF relative to the mean HRPPG ( H R PPG ¯ ) in [%] defined as:
H R DIFF = ( H R CTRL H R PPG ) / H R PPG ¯   ×   100   [ % ] .
The acquired HRDIFF values are statistically processed for final graphical comparison in the form of bar-graphs, boxplots, and histograms.
The HRDIFF parameters can only be determined by the steady PPG signal. Usually, the measurement by the BPM device and the PPG signal sensing is realized on the opposite arm/hand to minimize the pressure effect on the blood vessels in the arm. In this case, we observed a higher ripple of the PPG wave during the measurement time interval (between BPM start and BPM end) and a small decrease in the signal amplitude (range). This phenomenon has minimal influence on the finally determined HR values as shown in an example in Figure 5. If the arm for BP measurement and the hand for parallel PPG signal recording are the same, the recorded PPG cycles may be missing or having very low amplitude due to compressed blood vessels in the arm. For a detection of these events, we created the so called no-pulse signal function HPNOS with the value of “1” for no HR pulses present in the PPG signal and the value of “0” otherwise. For practical calculation of this function, the envelope threshold ETHRESH based on the possible minimum upper and lower envelopes EHI, ELOW must be determined from the whole processed PPG wave consisting of P determined PPG cycles. Then, each systolic pulse amplitude Apj is compared with the ETHRESH value:
H P NOS = { 1 A p j E THRESH , E TRESH 0 A p j > | E THRESH |   1     j     P .
The final no-signal time duration TNOS is calculated from the whole PPG signal with a total time duration TDUR as a duration of the time interval in which the HPNOS value is “1”–see the red dash-dotted line in Figure 6a. HR values detected from the PPG signal are shown in Figure 6b. For the time interval TNOS when HPNOS is equal to 1, the HR values are not defined, so the maximum possible value is substituted there (HRmax = 130 min−1). The simultaneously picked-up pressure characteristics of the tested BPM device is then used for determination of the following three temporal parameters: TPUP—time interval of increasing the air pressure in the measuring cuff, TPDWN—time interval of squeezing the air out of the cuff, and TPMEAS—time duration of the whole measuring process, as documented in Figure 6c.

3. Objects, Experiments, and Results

3.1. Principal Description of Developed Prototypes of PPG Sensors

The PPG sensor for real-time continuous PPG signal measurement principally consists of 4 functional blocks: (1) optical part with LS transmitter and PD receiver together with the analog interface for basic pre-processing of an electrical signal picked up by the PD, (2) microcontroller part containing an analog-to-digital (A/D) converter and a serial communication interface, (3) wireless communication part for receipt/transmission of commands/data from/to an external device, (4) power supply part. The sensor’s power supply can be realized by rechargeable battery cells or power banks to prevent any galvanic connection with other devices powered by a standard AC line of 230 V and 50 Hz (in Europe). To enable its operation in a low magnetic field, the sensor must be made of a non-ferromagnetic material including the power supply. In addition, due to a strong RF disturbance in the scanning area of the MRI device, all of the parts must be shielded.
The first prototype of the PPG signal sensing tool (further called “PPG-EP”) operating in the transmission mode is based on the optical sensor HRM-2511E by Kyoto Electronic Co., China, practically realized in the form of a rubber finger ring (see the photo in Figure 2a). The output from the PHD is pre-amplified and double filtered in a cascade connection using the Easy Pulse sensor v 1.1 module (ER-CDE10301E) by Embedded Lab, Williamsburg, VA, USA. Each of the 2 cascade blocks in this analogue interface consists of a passive RC high-pass filter (HPF) with the cut-off frequency of about 0.5 Hz and an active low-pass filter (LPF) with the cut-off frequency set to 3.4 Hz [20]. The output from the second LPF block is passed to the input of a 10-bit A/D converter integrated in the 8-bit processor ATmega328P by Atmel Company. This processor is a core element of the microprocessor unit—the Arduino Uno v. 3.0 board [28] together with an USB interface which can also be supplemented by the GSM module for connection to a mobile data gateway [29]. For our purpose, the BT communication module HC-06 [30] is used. It supports the bi-directional data transfer in BT 2.0 standard at 2.4 GHz with the maximum baud rate of 115,200 bps. All 3 parts are powered via the USB port by 5 V power bank AlzaPower Source 20,000 Quick Charge 3.0 by Alza.cz—see the overview assembled photo in Figure 7a. Due to the described type of the analogue interface, this PPG sensor produces the second derivative of the PPG wave in the final effect (compare PPG wave types in Figure 1).
The next two prototypes of the PPG sensor are based on the optical Pulse Sensor Amped (Adafruit 1093) [31] working in a reflectance mode and having an analogue interface integrated on board with a light transmitter and receiver. The integrated analogue interface consists of one step pre-amplifier, passive high-pass/active low-pass signal filters with the same cut-off frequencies as Easy Pulse module, so the finally produced PPG wave represents the first derivative of the originally sensed PPG signal. These two prototype realizations differ in the used microcontroller board, the BT communication module, and the power supply method—see Table 1 and Table 2.
The second compared sensor in this study (further called “PPG-PS1”) was developed using the Arduino Nano v. 3.0 board [32], which is also based on the processor ATmega328P working with the clock frequency fCLK of 16 MHz, and an USB interface. For communication with the control device, the sensor is connected to the mentioned BT module HC-06 (see photo in Figure 7b). All 3 components are powered via the USB port by a cable from an external 5 V power bank THAZER.
The last developed PPG sensor (further called “PPG-BLE”) is based on Arduino Pro Mini v. 2.0 board [33] also with the processor ATmega328 but without the USB interface and running at fCLK = 8 MHz. For serial communication, the BT module MLT-BT05 by Techonics Ltd. (Lahore, Pakistan) [34] working in the BT4.1 BLE standard was applied. The 3.7 V rechargeable polymer-lithium-ion (Li-Po) cell was used for sensor powering. The battery is practically mounted directly on the top of the shielding aluminum box. The applied assemblage and realized shielding by aluminum boxes for both prototypes of PPG sensors working in a reflectance mode are documented in photos in Figure 7c.
In summary, it can be specified that all of the 3 investigated PPG sensor realizations:
  • have a common basic structure consisting of four functional blocks;
  • use Arduino boards based on the processor ATmega328P with integrated 10-bit A/D converters, so the theoretical maximum ADRES value is 210 = 1024;
  • enable real-time PPG signal sensing in the low magnetic field with RF disturbance environment;
  • the sensor’s body and optical part are covered by aluminum boxes;
  • work in the slave mode: after initialization they wait for commands from the master device via the BT connection;
  • PPG signal sensing in 2 operating modes: (1) PPG wave pick up in data blocks with the length of NMEAS = {1k, 4k, 10k, and 25k} using 16-bit binary data samples; (2) continual PPG signal measurement that begins and ends by <Start> and <Stop> commands;
  • work with the control application “PPGsens7BT.exe” developed for Windows platform (successfully tested under XP, 7, and 10 versions) created for the master device to control the PPG sensor;
  • a service program implemented in the microcontroller board supports measurement and data transmission;
  • adjustable time delay to read the analog signal from the optical sensor, to perform 10-bit A/D conversion and data transmission to the control device;
  • transmission of data blocks with the length of NMEAS = 1k-samples are used for monitoring and display of the currently sensed PPG signal. For practical PPG signal measurement, the setting of NMEAS = 10k or 24k-samples is usually applied enabling a 1-shot storage of the PPG wave with duration of 80 or 240 s (using fS = 125 Hz). If longer time durations of sensed PPG signals are required, the signal data can be stored directly to an output wave file. Other requested time durations of sensed PPG signals can be solved using the direct storage to an output wave file.
On the other hand, Table 1 summarizes differences in an architecture, basic components, and mechanical realizations, and Table 2 shows detailed differences in electrical and functional parameters of all three PPG sensors.

3.2. Description of the Control Application Based on Windows Platform

The control application PPGsens7BT.exe was created for the master device to control the whole process of the PPG signal acquisition and post-processing consisting of: (1) real-time monitoring and displaying of the PPG signal, (2) continuous PPG signal measurement for the selected sampling frequency fS and the post-processing operations. This second operation mode enables: (a) direct saving of the unprocessed PPG data to a file on the hard disk of the control device, (b) saving of the acquired PPG data to an internal memory buffer for further processing such as filtering and HR determination; in this way the modified data can be stored off-line to the file(s) by an application user. In this case, a fixed number of samples NMEAS is automatically transmitted from the PPG sensor to the control device. The received PPG signal is stored in a Wave format (with 16-bit quantization, mono, PCM coding), so the signal must be first converted to the bipolar relative representation (in the range from −1 to 1). The stored PPG signal records can be further processed and analyzed off-line in the Matlab program environment.
The developed application enables communication with a PPG sensor in three modes: (1) “off-line”: without any BT communication—only the stored PPG waves can be shown, analyzed, and processed; (2) “automatic”: with automatic connection to a pre-defined type of an external PPG sensor (by initial setting of program parameters stored in *.INI file)—real-time PPG signal monitoring as well as continual measurement including subsequent storing of received data is enabled; (3) “manual” mode beginning with the application start without BT connection, proceeding with a manually established connection to a chosen type of a PPG sensor and communication parameters (serial channel type, baud rate, BT 2.0/BT 4.1 BLE protocol, etc.) working similar to the automatic mode. In addition, within this mode it is also possible to perform manual disconnection and/or again to create a connection with another PPG sensor using different parameters. An example of a screen copy of the main control window of the application PPGsens7BT (version 5.19) is shown in Figure 8a. The activated setting windows with showed listing during establishing of the BT connection with the PPG sensor working at the standard 4.1 BLE—transmitted and received AT commands for MLT-BT05 module can be seen in Figure 8b.

3.3. Performed Measurements and Analyses

The main aim of all of the performed measurements and analyses was to find suitable and the most universal setting of parameters for PPG signal real-time recording as well as off-line post-processing in different working mode conditions including proper function in a low magnetic field environment. The comparison of three prototypes of the developed PPG sensors was realized in four phases:
  • Preliminary testing and verification of the functionality of the PPG sensors in cooperation with the control application for three operation modes: (1) without BT connection (NC), (2) after established connection to the control device (CE), (3) during real-time transmission of PPG signal samples to the control master device (MC). This phase was accompanied by a measurement of the sensor’s mean DC using a different power supply of 5/3.7 V in the 3 mentioned operation modes;
  • Testing of the precision and stability of the HR values determined from the PPG signal by comparative measurement with another commercial HR measurement device (portable BPM) and by a POXI device, all in the normal laboratory conditions. In addition, a supplementary analysis of the influence of the BP/HR parallel measurement by a BPM device on the same hand as the PPG signal recording was performed;
  • Analyzing the functionality, quality, and stability of the BT connection between the tested PPG sensor located inside the scanning area of the MRI device and the control device located outside the shielding cage. Additional testing of the influence of the BT transmission in the scanning area on the quality of the finally obtained MR images;
  • First-step measurement of the PPG signals in a low magnetic field environment of the MRI device, analysis of the PPG signal properties.
In the third and fourth phases, the experiments were performed under three different conditions: (1) the door of the shielding cage open (OD) and the MRI device executing no scan sequence (NS), (2) the door of the cage closed (CD) and NS, (3) CD and a MR scan sequence is running (SR). Testing of the quality of obtained MR images was only realized in the SR condition.

3.4. Experimental Conditions

In the frame of our previous research [10,11], we have also compared results of HR values measured by three tested BPMs with those determined from the PPG signal. The best results with minimal dispersion and approximately zero mean value of calculated relative differences were achieved by the automatic blood pressure monitor BP A150-30 AFIB by Microlife AG, Swiss Corporation, Widnau/Switzerland. Hence, this type of BPM was used in the present experiments [35]. To prevent the already discussed possible negative influence of an inflated pressure cuff of BPM on a tested person’s blood system, the PPG signal was picked up from a forefinger of the opposite hand—sees an arrangement photo in Figure 9a. In the second type of comparative measurement experiments, the PPG signal was sensed and measured in parallel for calibration of the determined HR values with the help of the oximeter Berry BM1000C [36] by Shanghai Berry Electronic Tech Co., Ltd., Shanghai, China. This POXI device works in a transmission mode and enables also recording of the blood oxygen saturation and transfer of the HR values to the control device (tablet) via BT connection. PPG sensors were successively attached to little fingers of both hands by an elastic ribbon, the POXI device on the forefingers as documented in Figure 9b. The recording of PPG signals as well as oximeter values lasted for 80 s. The HR values obtained in this way were subsequently processed off-line and analyzed statistically. In the currently realized comparative measurements two female and six male healthy volunteer persons (authors themselves and their colleagues with average age of 50 years) were joined.
In the supplementary experiments with a parallel measurement by a BPM device and a PPG sensor placed on the same hand, the PPG sensor was worn on a forefinger—a practically used arrangement is shown in a photo in Figure 9c.
The third and fourth phase of the experiments were realized in a low magnetic field environment of the open-air MRI device E-scan Opera by Esaote S.p.A. [13] that is located at the Institute of Measurement Science, Slovak Academy of Sciences in Bratislava. The static field with magnetic induction of 0.178 T is formed between two parallel permanent magnets of this MRI scanner. It is placed in a metal cage to suppress high-frequency interference. This cage is made of a 2-mm thick steel plate with evenly spaced holes of 2.5-mm diameter in a 5-mm grid to eliminate the propagation of the electromagnetic field to the surrounding space of the control room [13].
In the preliminary experiments, PPG sensors were located inside the scanning area of the MRI device and no testing person was present; only a testing water phantom with a special testing grid area (TGA) was placed in the RF coil for generating MR images. The PPG sensor’s body, together with its optical part, were laid on a patient’s bed (see photos in Figure 10). The quality of BT connection was evaluated by the Received Signal Strength Indicator (RSSI) parameter representing an estimated measure of power level that an RF client device receives from an access point [37]. At larger distances, the signal becomes weaker and the wireless data rates become slower, leading to a lower overall data throughput. Measurement of the RSSI parameter was performed in three mentioned operating conditions {OD&NS, CD&NS, CD&SR}; during measurement in the CD&SR condition, the Hi-res SE-HF scan sequence with TE = 26 ms, TR = 500 ms, and sagittal orientation was executed. The distance between the PPG sensor and the control device was Dx = 225 cm, both connected devices were in the approximate height of 75 cm from the floor. For each of the sensor prototypes in every condition, the RSSI value was measured three times in [dBm] during the BT connection establishment (see the documentary screen copy of the communication setting window of the PPGsens7BT application in Figure 8b). For usage in final comparison, the mean value was determined from these three obtained values. The same experimental arrangement was used in the investigation of the influence of BT transmission on the quality of MR images. Only the PPG-EP prototype in the SR condition was tested in this investigation because it consists of the BT module HC-06 working with higher levels of the communication signal energy than the BT module MLT-BT05 operating in the BT 4.1 BLE standard. Therefore, we suppose that the assumed RF disturbance will be principally higher than in the case of the prototype PPG-BLE.
In the first PPG signal measurement, the testing person was lying on a patient’s bed with a head placed near the RF sensing coil in the middle of the scanning area (see the arrangement photo in Figure 11). For testing of the sensor PPG-EP operating in the transmission mode, the rubber finger ring with optical part of the sensor was put on the left/right hand forefingers. The sensor’s body and the supply power bank were loosely laid on a patient’s bed near the left leg. In the case of testing of the sensor prototypes working in the reflectance mode (PPG-PS1 and PPG-BLE), the PPG sensor body with a power supply source was mounted on the left/right hand’s wrist and the optical part of the sensor was fixed on a forefinger by an elastic ribbon. Within the CD&SR measuring condition, the scan sequence 3D-CE (with TE = 30 ms, TR = 40 ms; 3D phases = 8) was running. In this way, six PPG signal records per a tested person were formed. In the frame of this measurement, small databases of PPG signals from eight healthy volunteers were collected and further processed. The examined persons were the authors themselves and their colleagues in the age between 20 and 59 years—4 females and 4 males.

4. Discussion of Obtained Results

The preliminary experiments in the standard laboratory conditions confirm the practical functionality of all three tested realizations of PPG sensors with real-time BT data transfer mastered from the control application running on the laptop with Windows 10 OS. Due to the required real-time function of all 3 investigated PPG sensor prototypes, the recommended sampling frequencies were fS = {100, 125, 200, 250, and 500 Hz}. On the other hand, sensing of the PPG signal with fS higher than 250 Hz is only reasonable in the case of the systolic pulse width determination with higher accuracy. In our case the precise shape of peaks is not so relevant; rather, only the detected heart pulse period is necessary for correct HR value calculation. Therefore, the setting of fS = 125 Hz was finally chosen for use in further measurement experiments. A measurement of the sensor’s mean DC shows relatively great independency on the used operating voltage, type of a power supply, and a microcontroller board—see the values in Table 3. Chosen types and practically used power banks and Li-Po battery cells with their capacity guarantee safe power supplying for long-term PPG signal sensing experiments.
In the frame of the second phase experiments, the accuracy and stability (fluctuation) of the heart rate determined from the PPG signal was compared with discrete values obtained by the BPM device as well as with continual values transmitted from the oximeter. Statistical results of comparative measurements by the OXI device shown in Figure 12 demonstrate the lowest HRVAR values for the PPG-EP sensor and the highest ones in the case of the PPG-PS1 prototype. HR fluctuation of the used OXI device was comparable with the PPG-EP sensor but it was lower than the values obtained by PPG sensors working on a reflectance principle. Essential differences were detected according to the used hand (PPG signal sensed on the left hand had higher variation than from the right hand) but not according to the gender of a tested person (male vs. female). These results are in principal correspondence with others obtained by comparative measurements with the help of the BPM device as documented by numerical results in Table 4. However, HR stability parameters obtained from parallel measurements by the OXI device have shown greater HRVAR and HRDIFF values. Both parallel measurements have shown that the PPG-PS1 realization enables HR determination with positive HRDIFF variation of the measured values while the PPG-EP and PPG-BLE prototypes give negative HRDIFF variation as can be seen on the histogram in Figure 12c and in the right part of Table 4.
The additionally obtained experimental results confirm our assumption that the parallel measurement by the BPM device on the same arm has significant negative influence on the PPG signal sensed from the fingers which is manifested by generation of a no-signal time interval with the duration TNOS. In the final effect, differences were observed in blood circulation in the measured arm. Analysis of the determined TNOS values shows that time intervals are the longest for the PPG-EP prototype with an optical sensor working in the transmittance mode. Shorter TNOS intervals were observed in the case of prototypes using the reflectance optical sensor but there are no significant differences between PPG-PS1 and PPG-BLE prototypes. In contrast, all three prototypes differ in TNOS values depending on the hand type—see statistical results for all tested persons in Figure 13.
Preliminary mapping of conditions has confirmed the possibility of functional wireless BT connection and data transfer through the shielding cage of the tested MRI device with lower signal amplitude. It is well documented by the obtained RSSI values for three tested experimental conditions provided in Table 5. The minimum RSSI of −94 dBm was reached for the PPG-BLE sensor realization when the MR scan sequence was running, but also in this case the BT connection was stable, and the received PPG signal was usable (without any disturbance or artifacts). An investigation about the influence of the BT transmission inside the scanning area of the running MRI device on the scanning process confirms our working presumption.
For an analysis of the influence of BT transmission during the scanning process of the running MRI device on the quality of the obtained MR images, only the PPG-EP prototype with HC-06 BT module was tested. The water phantom presented in Figure 10b was used for this testing. The obtained results in the form of MR images of this water phantom rotated by 180 degrees can be seen in Figure 14. The achieved quality factors of the images are the same (Q.F. = 132), but careful observation of these images reveals soft horizontal lines inside the phantom’s special grid area that are more visible in Figure 14b,c than in Figure 14a. Finally, we can generalize that the application of PPG sensing has a minimal influence on the quality of the obtained MR images.
First-step measuring experiments confirm practical usability of all three proposed sensors for long-term sensing of the PPG signal in the magnetic field environment with additional RF and EM disturbance. To guarantee secure serial BT communication between the PPG sensor and the control device through the shielding cage the baud rate must be decreased, especially in the case of the PPG-BLE realization (max. at 57,600 bps—see the fourth column in Table 2). The numerical comparison of PPG signal properties and HR parameters in Table 6 shows that the running scan process (CD&RS) has practically minimum influence on the sensed PPG signal. Higher HRVAR values can be observed in the case of a group of female persons whose stress factor due to scanning inside the MRI device was probably greater. All 3 tested PPG sensors produce relatively stable PPG signals with the ripple up to 20%; the highest HRVAR values were detected for the PPG-PS1 prototype, the minimum values in the case of the PPG-EP sensor. The results of a detailed analysis of properties of the PPG signal sensed from a male person with the help of the PPG-EP prototype in three sensing conditions {OD&NS, CD&NS, CD&RS} are presented graphically in Figure 15a,b. The obtained results are principally in correspondence with the summary ones for {CD&NS/CD&RS} states. A decrease in a PPG signal range and an increase in HPRIPP, and HRVAR parameters can be caused by a progressive stress effect on a tested person. The calculated histograms of z-scores of HR values for both hands show no significant differences in three sensing conditions. Nevertheless, histograms for the PPG signal sensed in OD&NS condition contain curves that are smoothed more than for CD&RS condition as documented in Figure 15c.

5. Conclusions

Investigated PPG sensor prototypes were developed especially for measurement in the magnetic field environment. The comparative measurements performed in laboratory conditions using the BPM and OXI devices show that all three tested realizations of PPG sensors have acceptable and similar accuracy of HR determination. The detailed statistical analysis provides the highest variance of HRDIFF values in PPG signals sensed by the PPG-PS1 type. The PPG-EP prototype with the optical sensor part working in a transmittance mode produces the smallest variance. There are essential differences between HRVAR values determined from PPG signals of the left and right hands as presented by the graphs in Figure 12 and by numerical results in Table 4. The supplementary analysis has shown that the cuff must be placed on the opposite arm for proper BP measurement while the pressure effect on the PPG signal was practically none.
Additional analysis based on application of a special water phantom confirms our presumption that BT transmission in the scanning area of the running MRI device has practically no influence on the quality of the obtained scanner images. Finally, the performed measurements using testing persons lying inside the MRI tomograph during the scanning process have verified proper function of all three PPG sensors in the magnetic field with RF and EM disturbance and usability of the sensed PPG signals for further processing and analysis.
The PPG-BLE prototype with lower supply voltage produces a PPG signal with lower PPGRANGE and subsequently worse signal-to-noise ratio. On the other hand, the acquisition should be battery saving to enable long-term PPG signal recording. Therefore, this requirement is fulfilled by using the low energy PPG-BLE sensor—as shown by the PPG sensors’ mean DC values in Table 3.
The main limitation of this study lies in the fact that only a small group of tested persons participated in the measurement of PPG signals due to a bad pandemic situation in our country and limited possibilities for experiment realizations, so these results cannot be generalized. Only healthy vaccinated people could participate (authors themselves and their colleagues from IMS SAS) for collecting the PPG signal databases. The second limitation lies in the fact that our open-air MRI device was devised for standard medical practice, but our institute does not hold a certificate for real patient examination, so it can only be used for non-clinical and non-medical research.
We plan to realize more measurement experiments inside the MRI device in the near future. Properties of the PPG signals picked up from fingers and wrists will be compared with HR values measured by examined persons’ smart watches and fitness bracelets. The whole-body MRI device TMR-96 also located at our institute will be used for PPG sensor prototype testing. In this MRI scanner with magnetic inductance of 0.1 T, the gradient system is formed by solenoid coils flown by stronger currents resulting in stronger electromagnetic disturbance. In the case of possibly using of PPG sensors in MRI devices working with strong magnetic induction up 3 T, we will require further structural and material modifications of both the optical part of the PPG sensor and its body. The aforementioned limitation of using only non-ferromagnetic parts and components must be strictly retained, because even small metal pieces can cause damage of the whole device or bring about any injury of a tested person. Finally, practically applicable power supply methods for PPG sensors have not yet been analyzed; here, the possibility of external power supply using a power-bank is excluded.

Author Contributions

Conceptualization and methodology, J.P. and A.P.; data collection and processing, J.P.; writing—original draft preparation, J.P. and A.P.; writing—review and editing, A.P.; project administration, I.F.; funding acquisition, I.F. All authors have read and agreed to the published version of the manuscript.

Funding

This work was funded by the Slovak Scientific Grant Agency project VEGA2/0003/20 and the Slovak Research and Development Agency project APVV-19-0531.

Institutional Review Board Statement

Ethical review and approval were waived for this study, due to testing authors themselves and colleagues from IMS SAS. No personal data were saved, only PPG signals used in this research.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data supporting reported results are not readily available because they can be used only for research purposes and our paper must be cited where our data are used. Requests to access the data should be directed to the corresponding author Jiří Přibil ([email protected]).

Acknowledgments

We would like to thank all our colleagues who participated in the PPG signal recording as well as the BP and HR measurement experiments.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Steckner, M.C. A review of MRI acoustic noise and its potential impact on patient and worker health. eMagRes 2020, 9, 21–38. [Google Scholar] [CrossRef]
  2. Celka, P.; Charlton, P.H.; Farukh, B.; Chowienczyk, P.; Alastruey, J. Influence of mental stress on the pulse wave features of photoplethysmograms. Healthc. Technol. Lett. 2020, 7, 7–12. [Google Scholar] [CrossRef] [PubMed]
  3. Charlton, P.H.; Marozas, V. Wearable photoplethysmography devices. In Photoplethysmography: Technology, Signal Analysis, and Applications; Kyriacou, P.A., Allen, J., Eds.; Elsevier: London, UK, 2022; pp. 401–438. [Google Scholar] [CrossRef]
  4. Jarchi, D.; Salvi, D.; Tarassenko, L.; Clifton, D.A. Validation of instantaneous respiratory rate using reflectance PPG from different body positions. Sensors 2018, 18, 3705. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Rundo, F.; Conoci, S.; Ortis, A.; Battiato, S. An advanced bio-inspired photoplethysmography (PPG) and ECG pattern recognition system for medical assessment. Sensors 2018, 18, 405. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  6. Allen, J. Photoplethysmography and its application in clinical physiological measurement. Physiol. Meas. 2007, 28, R1–R39. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Mainka, A.; Platzek, I.; Mattheus, W.; Fleischer, M.; Müller, A.S. Three-dimensional vocal tract morphology based on multiple magnetic resonance images is highly reproducible during sustained phonation. J. Voice 2017, 31, e11–e504. [Google Scholar] [CrossRef] [PubMed]
  8. Hansen, H.L.; Patil, S. Speech under stress: Analysis, modeling and recognition. In Speaker Classification I; Muller, C., Ed.; Springer: Berlin/Heidelberg, Germany, 2007; Volume LNAI 4343, pp. 108–137. [Google Scholar] [CrossRef]
  9. Schickhofer, L.; Malinen, J.; Mihaescu, M. Compressible flow simulations of voiced speech using rigid vocal tract geometries acquired by MRI. J. Acoust. Soc. Am. 2019, 145, 2049–2061. [Google Scholar] [CrossRef] [PubMed]
  10. Přibil, J.; Přibilová, A.; Frollo, I. Physiological impact of vibration and noise in an open-air magnetic resonance imager: Analysis of a PPG signal of an examined person. Proceedings 2020, 42, 14. [Google Scholar] [CrossRef] [Green Version]
  11. Přibil, J.; Přibilová, A.; Frollo, I. Comparative measurement of the PPG signal on different human body positions by sensors working in reflection and transmission modes. Eng. Proc. 2020, 2, 69. [Google Scholar] [CrossRef]
  12. Přibil, J.; Přibilová, A.; Frollo, I. Two realizations of the wearable PPG sensor working in reflectance mode for measurement in weak magnetic field. Eng. Proc. 2021, 10, 55. [Google Scholar] [CrossRef]
  13. E-Scan Opera. Image Quality and Sequences Manual; Revision 830023522; Esaote S.p.A.: Genoa, Italy, 2008. [Google Scholar]
  14. Smith, A.G. Introduction to Arduino: A Piece of Cake; CreateSpace Independent Publishing Platform: Scotts Valley, CA, USA, 2011; 170p. [Google Scholar]
  15. Elgendi, M. PPG Signal Analysis: An Introduction Using MATLAB, 1st ed.; CRC Press: Abingdon, OX, USA, 2021; pp. 27–36. [Google Scholar]
  16. Caesarendra, W.; Hishamuddin, T.A.; Lai, D.T.C.; Husaini, A.; Nurhasanah, L.; Glowacz, A.; Alfarisy, G.A.F. An embedded system using convolutional neural network model for online and real-time ECG signal classification and prediction. Diagnostics 2022, 12, 795. [Google Scholar] [CrossRef] [PubMed]
  17. Nitzan, M.; Ovadia-Blechman, Z. Physical and physiological interpretations of the PPG signal. In Photoplethysmography: Technology, Signal Analysis, and Applications; Kyriacou, P.A., Allen, J., Eds.; Elsevier: London, UK, 2022; pp. 319–339. [Google Scholar] [CrossRef]
  18. Ahmed, H.; Rony, R.J. Understanding self-reported stress among drivers and designing stress monitor using heart rate variability. Qual. User Exp. 2021, 6, 4. [Google Scholar] [CrossRef]
  19. Blazek, V.; Venema, B.; Leonhardt, S.; Blazek, P. Customized optoelectronic in-ear sensor approaches for unobtrusive continuous monitoring of cardiorespiratory vital signs. Int. J. Ind. Eng. Manag. 2018, 9, 197–203. [Google Scholar] [CrossRef]
  20. Pulse Oximeter Model 2500 A. Operator’s Manual. © 2014 Nonin Medical, Inc., USA. Available online: https://www.nonin.com/products/palmsat-2500/ (accessed on 7 June 2020).
  21. Easy Pulse Sensor (Version 1.1). Overview-Embedded Lab. Available online: http://embedded-lab.com/blog/easy-pulse-version-1-1-sensor-overview (accessed on 6 June 2019).
  22. Kettler Consul Super–Art. Nr. 07954-900. User Manual. Available online: https://www.libble.eu/kettler/c457749/ (accessed on 14 July 2020).
  23. Garmin Vívoactive® 3. Available online: https://www.garmin.com/en-US/p/571520#specs (accessed on 25 May 2020).
  24. Hui, M.L.; Nisar, H.; Ho, Y.K.; Chiong, T.P. A video based heart rate monitoring system for drivers using photoplethysmography signal. In Proceedings of the 2018 IEEE Region 10 Humanitarian Technology Conference, Colombo, Sri Lanka, 6–8 December 2018; p. 8629825. [Google Scholar] [CrossRef]
  25. Mejía-Mejía, E.; Allen, J.; Budidha, K.; El-Hajj, C.; Kyriacou, P.A.; Charlton, P.H. Photoplethysmography signal processing and synthesis. In Photoplethysmography: Technology, Signal Analysis, and Applications; Kyriacou, P.A., Allen, J., Eds.; Elsevier: London, UK, 2022; pp. 69–145. [Google Scholar] [CrossRef]
  26. Bonomi, A.G.; Schipper, F.; Eerikäinen, L.M.; Margarito, J.; Van Dinther, R.; Muesch, G.; De Morree, H.M.; Aarts, R.M.; Babaeizadeh, S.; McManus, D.D.; et al. Atrial fibrillation detection using a novel cardiac ambulatory monitor based on photo-plethysmography at the wrist. J. Am. Heart Assoc. 2018, 7, e009351. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  27. Neshitov, A.; Tyapochkin, K.; Smorodnikova, E.; Pravdin, P. Wavelet analysis and self-similarity of photoplethysmography signals for HRV estimation and quality assessment. Sensors 2021, 21, 6798. [Google Scholar] [CrossRef] [PubMed]
  28. Arduino UNO R3. Available online: https://docs.arduino.cc/hardware/uno-rev3 (accessed on 20 July 2020).
  29. Suwarjono, S.; Wayangkau, I.H.; Istanto, T.; Rachmat, R.; Marsujitullah, M.; Hariyanto, H.; Caesarendra, W.; Legutko, S.; Glowacz, A. Design of a home fire detection system using Arduino and SMS gateway. Knowledge 2021, 1, 61–74. [Google Scholar] [CrossRef]
  30. Arduino and Bluetooth Module HC-06. Available online: https://www.aranacorp.com/en/arduino-and-bluetooth-module-hc-06/ (accessed on 21 October 2020).
  31. Pulse Sensor Amped Product (Adafruit 1093): World Famous Electronics LLC. Ecommerce Getting Starter Guide. Available online: https://pulsesensor.com/pages/code-and-guide (accessed on 16 July 2020).
  32. Arduino Nano Official Store. Available online: http://arduino.cc/en/Main/ArduinoBoardNano (accessed on 21 October 2020).
  33. Using the Arduino Pro Mini 3.3V. Available online: https://learn.sparkfun.com/tutorials/using-the-arduino-pro-mini-33v (accessed on 7 April 2021).
  34. MLT-BT05 4.0 Bluetooth Serial Communication Module. Available online: https://www.techonicsltd.com/product/mlt-bt05-ble4-0/ (accessed on 12 May 2021).
  35. Microlife BP A150 AFIB. Available online: https://www.microlife.cz/produkty/krevni-tlak/automaticke-tlakomery/bp-a150-afib (accessed on 12 May 2019).
  36. China Berry Hospital Oxycare Pulse Oximeter for Kids. Available online: https://berrymedical.en.made-in-china.com/product/zsuQEvPLsqcJ/Ch (accessed on 12 May 2021).
  37. RSSI-Received Signal Strength Indication. Available online: https://www.rfwireless-world.com/Terminology/RSSI-Received-Signal-Strength-Indication.html (accessed on 3 February 2022).
Figure 1. Example of three basic types of photoplethysmograms: (a) 1-k sample of the raw photoplethysmogram together with its first/second derivatives, (b) detailed 100 sample ROI around the 10th PPG cycle together with determined maxima corresponding to systolic and diastolic peaks and a denoted heart pulse period THP.
Figure 1. Example of three basic types of photoplethysmograms: (a) 1-k sample of the raw photoplethysmogram together with its first/second derivatives, (b) detailed 100 sample ROI around the 10th PPG cycle together with determined maxima corresponding to systolic and diastolic peaks and a denoted heart pulse period THP.
Sensors 22 03769 g001
Figure 2. Examples of practical realizations of optical parts of a PPG sensor: (a) rubber finger ring [21], (b) plastic finger clip [20], (c) plastic ear clip [22], (d) reflectance PPG sensor with analogue interface fixed on an index finger by an elastic ribbon, (e) back side of the Garmin watch VivoActive 3 with three green LEDs and one PD inside—loosely put on a wrist [23].
Figure 2. Examples of practical realizations of optical parts of a PPG sensor: (a) rubber finger ring [21], (b) plastic finger clip [20], (c) plastic ear clip [22], (d) reflectance PPG sensor with analogue interface fixed on an index finger by an elastic ribbon, (e) back side of the Garmin watch VivoActive 3 with three green LEDs and one PD inside—loosely put on a wrist [23].
Sensors 22 03769 g002
Figure 3. Visualization of adaptive signal level threshold determination: (a) whole 10-k sample PPG signal with determined LpMAX, LpMIN levels, HPAMPL and heart ripple parameters, and finally calculated LADAPT level, (b) ROI within 80th–89th PPG cycles with localized systolic peaks, determined LpMAX/LpMIN values, and global LOFS value; fs = 100 Hz.
Figure 3. Visualization of adaptive signal level threshold determination: (a) whole 10-k sample PPG signal with determined LpMAX, LpMIN levels, HPAMPL and heart ripple parameters, and finally calculated LADAPT level, (b) ROI within 80th–89th PPG cycles with localized systolic peaks, determined LpMAX/LpMIN values, and global LOFS value; fs = 100 Hz.
Sensors 22 03769 g003
Figure 4. An example of HR determination: (a) 1k-samples of a PPG wave together with signed heart pulse periods THP, lengths T1P, T0P, and a clipped sequence cPPG (b) HR values corresponding to periods THP together with a final mean HR; LTHRESH = 750, fs = 125 Hz.
Figure 4. An example of HR determination: (a) 1k-samples of a PPG wave together with signed heart pulse periods THP, lengths T1P, T0P, and a clipped sequence cPPG (b) HR values corresponding to periods THP together with a final mean HR; LTHRESH = 750, fs = 125 Hz.
Sensors 22 03769 g004
Figure 5. Visualization of a sensed PPG wave with parallel measurement by the BPM device on different arms in three phases (before, during, and after the BPM measurement): (a) sensed PPG wave with calculated signal range, envelopes, and ripple, (b) HR values from the PPG signal compared with HRBPM together with HRVAR and HRMEAN values; a PPG sensor is worn on a left forefinger, a BPM device on a right arm.
Figure 5. Visualization of a sensed PPG wave with parallel measurement by the BPM device on different arms in three phases (before, during, and after the BPM measurement): (a) sensed PPG wave with calculated signal range, envelopes, and ripple, (b) HR values from the PPG signal compared with HRBPM together with HRVAR and HRMEAN values; a PPG sensor is worn on a left forefinger, a BPM device on a right arm.
Sensors 22 03769 g005
Figure 6. Visualization of a PPG wave sensed from the left little finger with parallel BP measurement on the same arm: (a) the PPG signal and its envelopes together with determined TNOS ≅ 10 s, (b) HR values determined from the PPG signal, (c) the pressure characteristic measured by the BPM device with determined TPUP, TPDWN, and TPMEAS time intervals, BP and HR values.
Figure 6. Visualization of a PPG wave sensed from the left little finger with parallel BP measurement on the same arm: (a) the PPG signal and its envelopes together with determined TNOS ≅ 10 s, (b) HR values determined from the PPG signal, (c) the pressure characteristic measured by the BPM device with determined TPUP, TPDWN, and TPMEAS time intervals, BP and HR values.
Sensors 22 03769 g006
Figure 7. Assemblage with aluminum covering of: (a) prototype of the PPG-EP sensor, (b) the body and optical part of the PPG-PS1 sensor, (c) the body, optical part of the PPG-BLE sensor, and the Li-Po battery used for power supply.
Figure 7. Assemblage with aluminum covering of: (a) prototype of the PPG-EP sensor, (b) the body and optical part of the PPG-PS1 sensor, (c) the body, optical part of the PPG-BLE sensor, and the Li-Po battery used for power supply.
Sensors 22 03769 g007
Figure 8. Screen copy of the control application PPGsens7BT v.5.19: (a) main operating window with a displayed PPG wave (4th frame of 4-k sample data block, fS = 250 Hz) and a determined mean HR value in beats per minute, (b) the parameter setting window documenting the process of BT connection creation with the PPG-BLE sensor based on the BT module MLT-BT05.
Figure 8. Screen copy of the control application PPGsens7BT v.5.19: (a) main operating window with a displayed PPG wave (4th frame of 4-k sample data block, fS = 250 Hz) and a determined mean HR value in beats per minute, (b) the parameter setting window documenting the process of BT connection creation with the PPG-BLE sensor based on the BT module MLT-BT05.
Sensors 22 03769 g008
Figure 9. Arrangements of comparative measurements of HR values from the PPG wave and by the BPM and POXI devices: (a) with parallel measurement by the BPM device on the right hand and the PPG-EP sensor on a forefinger of the left hand; (b) parallel sensing of the PPG signal via the POXI with BT data transfer to a tablet and by the PPG-BLE sensor using a forefinger and a pinkie of the left hand; (c) parallel measurement by the BPM device and by the PPG-PS1 sensor worn on a forefinger-both placed on the left hand.
Figure 9. Arrangements of comparative measurements of HR values from the PPG wave and by the BPM and POXI devices: (a) with parallel measurement by the BPM device on the right hand and the PPG-EP sensor on a forefinger of the left hand; (b) parallel sensing of the PPG signal via the POXI with BT data transfer to a tablet and by the PPG-BLE sensor using a forefinger and a pinkie of the left hand; (c) parallel measurement by the BPM device and by the PPG-PS1 sensor worn on a forefinger-both placed on the left hand.
Sensors 22 03769 g009
Figure 10. Arrangement of RSSI parameter measurement inside the MRI device E-Scan Opera: (a) overall photo with currently tested PPG-BLE sensor, (b) detail of used water phantom with a special testing grid area inside the RF receiving/transmitting coil.
Figure 10. Arrangement of RSSI parameter measurement inside the MRI device E-Scan Opera: (a) overall photo with currently tested PPG-BLE sensor, (b) detail of used water phantom with a special testing grid area inside the RF receiving/transmitting coil.
Sensors 22 03769 g010
Figure 11. Measurement of a person lying in the MRI device Opera using the PPG-EP sensor with BT data transmission to a control device located outside the shielding metal cage.
Figure 11. Measurement of a person lying in the MRI device Opera using the PPG-EP sensor with BT data transmission to a control device located outside the shielding metal cage.
Sensors 22 03769 g011
Figure 12. Statistical results of comparative measurements by the oximeter device for all three tested PPG sensor prototypes: (a) bar-graph of mean variances of HR values separately for both hands, (b) boxplot of basic statistical parameters of HRDIFF values joint for both hands, (c) histograms of HRDIFF values for left and right hands together; all tested persons.
Figure 12. Statistical results of comparative measurements by the oximeter device for all three tested PPG sensor prototypes: (a) bar-graph of mean variances of HR values separately for both hands, (b) boxplot of basic statistical parameters of HRDIFF values joint for both hands, (c) histograms of HRDIFF values for left and right hands together; all tested persons.
Sensors 22 03769 g012
Figure 13. Statistical comparison of TNOS time intervals for tested PPG sensors: (a) boxplot of TNOS values for left hand forefingers, (b) boxplot for right hand forefingers, (c) mean TNOS values for both hands together; all tested persons.
Figure 13. Statistical comparison of TNOS time intervals for tested PPG sensors: (a) boxplot of TNOS values for left hand forefingers, (b) boxplot for right hand forefingers, (c) mean TNOS values for both hands together; all tested persons.
Sensors 22 03769 g013
Figure 14. MR images of the water phantom with the testing grid area obtained in three BT communication conditions of the PPG-EP sensor with the HC-06 BT module placed in the scanning area of the MRI device: (a) without any BT connection, (b) after established connection to the control device, (c) during data transmission to the control master device; executed Hi-res SE-HF scan sequences with TE = 26 ms, TR = 500 ms, and sagittal orientation.
Figure 14. MR images of the water phantom with the testing grid area obtained in three BT communication conditions of the PPG-EP sensor with the HC-06 BT module placed in the scanning area of the MRI device: (a) without any BT connection, (b) after established connection to the control device, (c) during data transmission to the control master device; executed Hi-res SE-HF scan sequences with TE = 26 ms, TR = 500 ms, and sagittal orientation.
Sensors 22 03769 g014
Figure 15. Comparison of PPG parameters using the PPG-EP prototype for three conditions inside the MRI device—both hands separately: (a) mean values of PPGRANGE, HPRIPP, and HRVAR parameters for the left hand (b) the right hand, (c) histograms of z-scores calculated from HR values; PPG sensor was worn on the index fingers of L/R hands of a male.
Figure 15. Comparison of PPG parameters using the PPG-EP prototype for three conditions inside the MRI device—both hands separately: (a) mean values of PPGRANGE, HPRIPP, and HRVAR parameters for the left hand (b) the right hand, (c) histograms of z-scores calculated from HR values; PPG sensor was worn on the index fingers of L/R hands of a male.
Sensors 22 03769 g015
Table 1. Differences in an architecture, basic components, and mechanical realizations of all three PPG sensor prototypes.
Table 1. Differences in an architecture, basic components, and mechanical realizations of all three PPG sensor prototypes.
Sensor PrototypeOptical Sensor PartMounting MethodWearableArduino Board (1)BT Module/StandardSensor’s Body Dimensions (L × W × H)Sensor’s Body Weight
PPG-EPHRM-2511Erubber ringNoUno v. 3.0HC-06/BT 2.0105 × 70 × 80 mm445 g
PPG-PS1Adafruit 10931” aluminum target fixed by elastic ribbonYesNano v. 3.0HC-06/BT 2.080 × 20 × 10 mm38 g
PPG-BLEAdafruit 1093YesPro Mini v. 2.0MLT-BT05/BT4.1 BLE40 × 25 × 15 mm40 g (2)
(1) All Arduino boards are based on the processor ATmega328 with 10-bit A/D converters; (2) Including mounted Li-Po battery cell.
Table 2. Differences in electrical and functional parameters for three tested PPG sensors.
Table 2. Differences in electrical and functional parameters for three tested PPG sensors.
Sensor PrototypeSensor Working ModeProcessor fCLKMax. BT Baud RateSupported fS for PPG SignalsPower Supply
(Voltage/Capacity)
PPG-EPtransmittance16 MHz115,200 bps{100, 125, 200, 250, 500, 1000} Hz5 V/22,000 mAh
PPG-PS1reflectance16 MHz115,200 bps{100, 125, 200, 250, 500, 1000} Hz5 V/2200 mAh
PPG-BLEreflectance8 MHz57,600 bps{100, 125, 200, 250, 500} Hz3.7 V/125 mAh
Table 3. Comparison of PPG sensors’ mean DC values for three functional states and used power supplies.
Table 3. Comparison of PPG sensors’ mean DC values for three functional states and used power supplies.
Sensor TypeSupplying MethodFunctional State
NCCEMC
PPG-EP5 V Power bank via USB53 mA60 mA68 mA
PPG-PS15 V Power bank via USB18 mA26 mA30 mA
PPG-BLE3.7 V Li-Po battery cell13.5 mA14 mA17 mA
Table 4. Mean HR stability parameters calculated from the PPG signal measured in parallel by the BPM device for all tested persons.
Table 4. Mean HR stability parameters calculated from the PPG signal measured in parallel by the BPM device for all tested persons.
Sensor (A)/ParametersHRVAR [%]HRDIFF [%]
LeftRightBothLeftRightBoth
PPG-EP3.162.662.91−1.48−1.24−1.36
PPG-PS14.353.784.062.521.892.21
PPG-BLE3.763.613.68−1.82−1.16−1.49
(A) Tested PPG sensors worn on forefingers of left/right hand, BPM device on an opposite arm.
Table 5. Mean RSSI values joined for two BT communication modules measured in three experimental conditions.
Table 5. Mean RSSI values joined for two BT communication modules measured in three experimental conditions.
BT Module (A)/ConditionOD&NSCD&NSCD&RS
HC-06 (5 V, BT 2.0)−60 dBm−90 dBm−91 dBm
MLT-BT05 (3.3 V, BT 4.1)−86 dBm−92 dBm−94 dBm
(A) PPG-EP and PPG-PS1 prototypes consist of the same BT module HC-06; PPG-BLE includes the MLT-BT05 module.
Table 6. PPG signal properties and mean HR parameters for two measurement conditions inside the MRI device for groups of male and female tested persons separately.
Table 6. PPG signal properties and mean HR parameters for two measurement conditions inside the MRI device for groups of male and female tested persons separately.
Parameter/Condition (A)PPG-EP (CD&NS/CD&RS)PPG-PS1 (CD&NS/CD&RS)PPG-BLE (CD&NS/CD&RS)
MaleFemaleMaleFemaleMaleFemale
PPGRANGE [%]55/5249/4750/4845/4245/4341/40
HPRIPP [%]13/1415/1615/1718/2015/1617/18
HRVAR [%]1.8/1.92.4/3.22.9/4.73.4/6.22.7/3.83.1/5.5
(A) Mean values both hands together.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Přibil, J.; Přibilová, A.; Frollo, I. Comparison of Three Prototypes of PPG Sensors for Continual Real-Time Measurement in Weak Magnetic Field. Sensors 2022, 22, 3769. https://doi.org/10.3390/s22103769

AMA Style

Přibil J, Přibilová A, Frollo I. Comparison of Three Prototypes of PPG Sensors for Continual Real-Time Measurement in Weak Magnetic Field. Sensors. 2022; 22(10):3769. https://doi.org/10.3390/s22103769

Chicago/Turabian Style

Přibil, Jiří, Anna Přibilová, and Ivan Frollo. 2022. "Comparison of Three Prototypes of PPG Sensors for Continual Real-Time Measurement in Weak Magnetic Field" Sensors 22, no. 10: 3769. https://doi.org/10.3390/s22103769

APA Style

Přibil, J., Přibilová, A., & Frollo, I. (2022). Comparison of Three Prototypes of PPG Sensors for Continual Real-Time Measurement in Weak Magnetic Field. Sensors, 22(10), 3769. https://doi.org/10.3390/s22103769

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop