1. Introduction
In the process of designing, manufacturing, and applying critical structural components, it is necessary to test and evaluate the comprehensive performance of their materials and structures, such as the distribution of internal and external stresses and deformations of structural components under certain conditions of loading, temperature, etc. [
1,
2,
3]. Typically, the stress can be obtained by monitoring the strain generated in the elastic range of the member material [
4]. Through strain testing, the degree and cause of structural deformation can be analyzed, which is of great significance to the quality monitoring of structural components.
I In recent years, a variety of high-precision strain sensors have emerged in the field of sensing and testing, specifically including two categories of contact and non-contact. After years of development, resistive strain gauges have become a more mature measurement method, and as the earliest measurement technology applied in the field of strain monitoring, with a wide range of characteristics such as simple testing and reliability, it has been widely used in a variety of engineering strain testing, commonly used in the calibration of other new sensors [
5,
6]. However, at the same time, the use of strain gauges as point sensors in the large-scale range sensing and monitoring of the fabrication structure is complex, and under high temperature and high humidity, electromagnetic interference and other complex environments are not applicable [
7,
8]. With the increase in the complexity of testing needs, fiber optic sensing technology has accelerated in development. The Fiber Bragg Grating (FBG) is one of the most representative sensors in fiber optic sensors and is resistant to electromagnetic interference, corrosion, and other complex environmental factors. Multiple gratings in series can be used to carry out quasi-distributed sensing measurements relatively easily. In recent years, with the further in-depth research on FBG sensors, the application of FBG sensors has expanded from strain and temperature to various fields. By characterizing FBGs into different types, we can extend them to other parameter measurements, such as acceleration, displacement, and pressure. These are the hot application fields of grating sensors at present [
9,
10].
Based on the distributed sensor technology of optical frequency domain reflection (OFDR), the backscattered Rayleigh signal on the optical fiber link is retrieved through the linear swept light source combined with the coherent detection technology, and the strain and temperature sensing test with high precision and high spatial resolution is realized. Using conventional single-mode optical fiber as sensor, it is easy to realize large-scale deployment and realize point-to-point high-precision and high-resolution distributed sensing, which is often used in the fields of structural health detection, nondestructive testing of composite materials, battery temperature testing and so on in civil engineering. With the advantages of OFDR technology, it provides a new idea for the design of various sensing test schemes [
11,
12]. At present, with the development of FBG and OFDR, FBG and OFDR have been widely used in the measurement of temperature, strain, stress and other physical parameters [
13,
14,
15]. As the basic application of optical fiber sensing technology, strain measurement, as a key index, needs further calibration and analysis.
As an elastic sensitive element, the beam with equal strength has the advantages of structural stability and convenient processing, and it is an isosceles triangle in shape. When the vertex of the triangle is subjected to concentrated load, the free end changes in deflection, and the stress and strain on the section at any distance from the vertex are constant. It is commonly used in various stress and strain detection tests.
In this paper, the standard Beams of equal strength calibration system with mature the Electric Strain Gauge (ESG) and two fiber optic sensing technologies, FBG and OFDR, was used to carry out accuracy calibration experiments and related analysis. After several Beams of equal strength loading and unloading tests, the fiber-optic single-point strain data with the Beams of equal strength loading and unloading changes showed good continuity and linearity, and the cyclic stability was better. The error of strain test data was less than 2% after repeated loading.
To accurately describe the strain distribution and numerical accuracy of the Beams of equal strength, a finite element simulation model was used to compare and analyze with the measured data, which showed an error of less than 5% and good linearity of the respective strain test curves. As the number of loading and unloading cycles increased, the local strain jumps in the ESG measurement data became more prominent, and the error of the strain test data was less than 4% after repeated loading. After repeated loading and unloading tests, it was verified that the fiber grating and distributed fiber exhibited good stability in terms of the repeatability accuracy of the strain test, while the strain gauges exhibited local strain shifts after multiple repeated loading and unloading cycles. This experiment further verified that OFDR distributed fiber optic sensing technology, FBG, and ESG were essentially comparable in terms of strain test accuracy, and also validated the high repeatability stability of fiber optic sensing technology. The calibration method provided technical support for the basic application of fiber optic sensing technology.
3. Experimental Setup
In the experiment, three types of sensors were sequentially deployed onto an isostress beam for testing, as illustrated in
Figure 5, which shows the schematic diagram of the sample deployment. Equal strength beam means that the maximum normal stress on each cross section of the beam is equal and reaches the allowable stress of the material. The material of equal strength beam used in this experiment is aluminum alloy, and the working size is
L × B × H = 295 × 9.25 × 3 mm. The strain calculation formula for beams with equal strength is:
where
L,
H and
E are the length, thickness and elastic modulus of beams with equal strength, respectively (
E = 72 GPa). where
B is the width of the load-bearing part and
σ is the stress on the beam section. Tensile stress, p is the acting force on the beam with equal strength.
Among them, the Fiber Bragg Grating (FBG) system utilized a fiber optic grating with a central wavelength of 1565 nm as the sensor, the Optical Frequency Domain Reflectometry (OFDR) system employed a bend-insensitive polyimide optical fiber, and the Electric Strain Gauge (ESG) was equipped with a sensor featuring a sensitive grid length of 3 mm, as depicted in the actual sensor image shown in
Figure 6. In this test, a fiber grating with a central wavelength of 1565 nm, a strain gauge with a sensitive grating length of 3 mm and a single-mode polyimide fiber with a length of 20 cm are used as strain test sensors. The experimental process involved loading weights incrementally, with each increment consisting of a 100 g weight, totaling 15 levels of loading.
In this experiment, the strain gauge demodulator is DH3818Y static stress and strain test and analysis system produced by Jiangsu Donghua Test Technology Co., Ltd., (Jingjiang, China) with a strain range of ±10,000 με, a resolution of ±0.1 με and a static sampling rate of 5 Hz. The fiber grating demodulator is OCI series equipment produced by Wuhan Megasense Technologies Co., Ltd. (Wuhan, China). The strain and temperature measurement is realized by calculating the center wavelength offset of grating. The strain range is ±4000 με, the spatial resolution is in the nanometer level, and the demodulation rate is 2 kHz. OFDR system uses OSI series equipment produced by Wuhan Megasense Technologies Co., Ltd. (Wuhan, China). The OFDR system uses a narrow-band linear swept-frequency light source, with a working wavelength of 1530–1570 nm, a strain range of ±12,000 με, a spatial resolution of 1mm, a maximum sampling rate of 100 Hz and a strain accuracy of ±1 με. The above three sensing demodulation devices are shown in
Figure 7.
During the experiment, the isostress beam’s state without any weights placed was recorded as the zero state. The weights were incrementally added by 100 g, and the strain test data for each stable state after loading were recorded. For data recording, a grating demodulator, a strain demodulator, and an OFDR device were used sequentially to record and save the strain test data for the grating, strain gauge, and distributed optical fiber strain tests.
4. Results and Discussion
When loading with a 200 g weight, as depicted in
Figure 8, strain test data from the three types of sensors were obtained. It can be observed that both the strain gauge and the grating, after being demodulated by the demodulator, output a single strain value, whereas the strain test data from the distributed optical fiber is continuously distributed. The strain value measured by the strain gauge represents the average strain data within its sensitive grid section, and the strain value measured by the grating is determined by the strain size caused by the central wavelength shift within its grating area. Therefore, during data processing, it is necessary to average all the strain data from the sensing points within the optical fiber section corresponding to the sensitive grid of the strain gauge. This average is then used as the measurement result for the optical fiber, which is compared with the measurement data from the fiber grating and the strain gauge.
Finite element analysis was conducted on the equi-strength beam used in the experiment. At the same location on the equi-strength beam, a strain increase of 47 micro-strain (με) was observed for every 100 g weight added, as shown in
Figure 9 and
Figure 10.
Table 1 shows the test loading data, with 100 g weights per stage and 15 stages in total. Three groups of test data are recorded, and the average value is plotted as shown in
Figure 11, in which the horizontal axis represents the increased weight and the vertical axis represents the measured strain value. The strain test data from three sensors are fitted linearly in turn, and the fitting coefficients (r) are 0.99813, 0.99889 and 0.99873 respectively. The good linear relationship shows that the test data in this experiment is accurate and effective. When comparing the strain test data from 1565 nm fiber grating, distributed fiber and 3 mm strain gauge, the error is less than 2%. Compared with the data of finite element theory analysis, the error between the measured data and the data of finite element theory analysis is less than 5%.
Multiple repeated loading and unloading cycles were conducted on the three sensors: the fiber grating, the distributed optical fiber, and the strain gauge. A 100 g weight was used for each loading/unloading stage, totaling eight stages. Based on the strain test results from the three sensors as depicted in
Figure 12 and
Figure 13, the following conclusions can be drawn. The single-point strain data from the fiber grating and the distributed optical fiber exhibited good continuity and linearity with changes in loading and unloading, demonstrating better cyclic stability. After repeated loading, the strain test data error was less than 2%. In contrast, the strain gauge measurements showed increasingly noticeable local strain jumps with an increasing number of loading and unloading cycles. After multiple repeated loading cycles, the strain test data error was less than 4%. This, apart from being attributed to the intrinsic performance of the strain gauge, may also be related to the adhesive properties of the strain gauge. The adhesive layer between the sensitive grating of the strain gauge and the substrate, and the gluing between the substrate and the test piece will cause the strain gauge data to drift when they are displaced and degummed.
It can be clearly seen from
Figure 13 above that the strain data measured by strain gauge began to drift obviously after repeated loading of the test piece for many times. Under normal circumstances, there are two kinds of situations that lead to the deviation of test data: 1: Repeated loading causes irreversible damage to the test piece itself, which leads to data drift; 2: The stability of repeated loading test of the sensor itself is weak, and some irreversible strain remains in the sensor itself after repeated loading. Through the loading test of the equal strength beam specimen alone, after repeated loading tests, the aluminum alloy specimen of the equal strength beam is in good structural condition without damage, so the deviation of the strain gauge data caused by the specimen itself is ruled out, which further ensures the reliability and accuracy of the test.
In order to avoid the contingency of the selected loading data, the test data are statistically processed again and plotted as shown in
Table 2. The FBG, PI and ESG sensors are repeatedly loaded with 800 g weight, and an average value is taken every 40 groups of data, and the error rate of strain test data after repeated loading is analyzed and calculated.
5. Conclusions
This paper presents precision calibration experiments and related analyses using a standardBeams of equal strength system equipped with a mature ESG, along with two types of fiber optic sensing technologies: FBG and OFDR. After multipleBeams of equal strength loading and unloading tests, it was found that the fiber optic single-point strain data exhibited good continuity and linearity with changes in loading and unloading, and better cyclic stability, with the error in strain test data after multiple repeated loading being less than 2%. Additionally, the finite element theoretical analysis data were within a 5% error margin of the actual measured data, indicating a good linearity of the strain test curves with R2 values greater than 0.998. Furthermore, repeated loading and unloading experiments demonstrated the good stability of FBG and OFDR in terms of strain testing repeatability and precision, while the strain gauge showed local strain shifts after multiple repeated loading and unloading.
The calibration experiments and data analysis in this paper validate the stability of OFDR distributed fiber optic sensing technology and FBG in strain testing accuracy, providing foundational technical support for the development of applications of fiber optic sensing technologies.