Evaluation of the Milling Accuracy of Zirconia-Reinforced Lithium Silicate Crowns Fabricated Using the Dental Medical Device System: A Three-Dimensional Analysis
Abstract
:1. Introduction
2. Materials and Methods
2.1. Study Design and Experimental Condition
2.2. Selection of the Master Model and Acquisition of Impression Scanning Data
2.3. Fabrication of the Crown Using the Dental Milling Machine
2.4. Scanning of the Outer and Inner Surfaces of Crowns
2.5. Three-Dimensional Accuracy Analysis
2.6. Statistical Analysis
3. Results
4. Discussion
5. Conclusions
- After processing of ceramic restorations using a CAD/CAM system, lithium disilicate was shown to have superior milling accuracy compared to zirconia-reinforced lithium silicate.
- According to the results from the qualitative analysis, a positive error appearing in the internal and external parts of the crown may not sufficiently match the fitting of the prosthesis, resulting in micro-discrepancies. In addition, a negative error may loosen the prosthesis due to excessive deletion.
- According to the results from the quantitative analysis, milling accuracy was within 120 μm for all types of ceramics, thus confirming their clinical applicability.
Author Contributions
Funding
Acknowledgments
Conflicts of Interest
References
- Kang, W.; Park, J.-K.; Kim, W.-C.; Kim, H.-Y.; Kim, J.-H. Effects of Different Thickness Combinations of Core and Veneer Ceramics on Optical Properties of CAD-CAM Glass-Ceramics. BioMed Res. Int. 2019, 2019, 331529. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Yuan, J.C.-C.; Barão, V.A.R.; Wee, A.G.; Alfaro, M.F.; Afshari, F.S.; Sukotjo, C. Effect of brushing and thermocycling on the shade and surface roughness of CAD-CAM ceramic restorations. J. Prosthet. Dent. 2018, 119, 1000–1006. [Google Scholar] [CrossRef] [PubMed]
- Drago, C.J. Two new clinical/laboratory protocols for CAD/CAM implant restorations. J. Am. Dent. Assoc. 2006, 137, 794–800. [Google Scholar] [CrossRef] [PubMed]
- Kang, S.-Y.; Lee, H.-N.; Kim, J.-H.; Kim, W.-C. Evaluation of marginal discrepancy of pressable ceramic veneer fabricated using CAD/CAM system: Additive and subtractive manufacturing. J. Adv. Prosthodont. 2018, 10, 347–353. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Kim, D.-Y.; Kim, J.-H.; Kim, H.-Y.; Kim, W.-C. Comparison and evaluation of marginal and internal gaps in cobalt–chromium alloy copings fabricated using subtractive and additive manufacturing. J. Prosthodont. Res. 2018, 62, 56–64. [Google Scholar] [CrossRef]
- Kang, S.-Y.; Park, J.-H.; Kim, J.-H.; Kim, W.-C. Three-dimensional trueness analysis of ceramic crowns fabricated using a chairside computer-aided design/manufacturing system: An in vitro study. J. Prosthodont. Res. 2020, 64, 152–158. [Google Scholar] [CrossRef]
- Ng, J.; Ruse, D.; Wyatt, C. A comparison of the marginal fit of crowns fabricated with digital and conventional methods. J. Prosthet. Dent. 2014, 112, 555–560. [Google Scholar] [CrossRef]
- Shamseddine, L.; Mortada, R.; Rifai, K.; Chidiac, J.J. Fit of pressed crowns fabricated from two CAD-CAM wax pattern process plans: A comparative in vitro study. J. Prosthet. Dent. 2016, 118, 49–54. [Google Scholar] [CrossRef]
- De Kok, P.; Pereira, G.; Fraga, S.; De Jager, N.; Venturini, A.B.; Kleverlaan, C.J. The effect of internal roughness and bonding on the fracture resistance and structural reliability of lithium disilicate ceramic. Dent. Mater. 2017, 33, 1416–1425. [Google Scholar] [CrossRef]
- Ritter, R.G. Multifunctional Uses of a Novel Ceramic-Lithium Disilicate. J. Esthet. Restor. Dent. 2010, 22, 332–341. [Google Scholar] [CrossRef] [PubMed]
- Alkadi, L.; Ruse, N.D. Fracture toughness of two lithium disilicate dental glass ceramics. J. Prosthet. Dent. 2016, 116, 591–596. [Google Scholar] [CrossRef] [PubMed]
- Tinschert, J.; Natt, G.; Mautsch, W.; Augthun, M.; Spiekermann, H. Fracture resistance of lithium disilicate-, alumina-, and zirconia-based three-unit fixed partial dentures: A laboratory study. Int. J. Prosthodont. 2001, 14, 231–238. [Google Scholar] [PubMed]
- Anadioti, E.; Aquilino, S.A.; Gratton, D.G.; Holloway, J.A.; Denry, I.; Thomas, G.W.; Qian, F. 3D and 2D Marginal Fit of Pressed and CAD/CAM Lithium Disilicate Crowns Made from Digital and Conventional Impressions. J. Prosthodont. 2014, 23, 610–617. [Google Scholar] [CrossRef] [PubMed]
- Chai, H.; Lawn, B.R. A universal relation for edge chipping from sharp contacts in brittle materials: A simple means of toughness evaluation. Acta Mater. 2007, 55, 2555–2561. [Google Scholar] [CrossRef]
- Mounajjed, R.; Salinas, T.J.; Ingr, T.; Azar, B. Effect of different resin luting cements on the marginal fit of lithium disilicate pressed crowns. J. Prosthet. Dent. 2018, 119, 975–980. [Google Scholar] [CrossRef]
- Colombo, M.; Gallo, S.; Poggio, C.; Ricaldone, V.; Arciola, C.R.; Scribante, A. New Resin-Based Bulk-Fill Composites: In vitro Evaluation of Micro-Hardness and Depth of Cure as Infection Risk Indexes. Materials 2020, 13, 1308. [Google Scholar] [CrossRef] [Green Version]
- Sieper, K.; Wille, S.; Kern, M. Fracture strength of lithium disilicate crowns compared to polymer-infiltrated ceramic-network and zirconia reinforced lithium silicate crowns. J. Mech. Beheav. Biomed. Mater. 2017, 74, 342–348. [Google Scholar] [CrossRef]
- Elsaka, S.E.; Elnaghy, A.M. Mechanical properties of zirconia reinforced lithium silicate glass-ceramic. Dent. Mater. 2017, 32, 908–914. [Google Scholar] [CrossRef]
- Janyavula, S.; Lawson, N.; Cakir, D.; Beck, P.; Ramp, L.C.; Burgess, J.O. The wear of polished and glazed zirconia against enamel. J. Prosthet. Dent. 2013, 109, 22–29. [Google Scholar] [CrossRef]
- Oh, S.H.; Kim, S.G. Effect of abutment shade, ceramic thickness, and coping type on the final shade of zirconia all-ceramic restorations:in vitrostudy of color masking ability. J. Adv. Prosthodont. 2015, 7, 368–374. [Google Scholar] [CrossRef] [Green Version]
- Stefani, A.; Brito, R.B.; Kina, S.; Andrade, O.S.; Carvalho, A.A.; Ambrosano, G.M.B.; Giannini, M. Bond Strength of Resin Cements to Zirconia Ceramic Using Adhesive Primers. J. Prosthodont. 2016, 25, 380–385. [Google Scholar] [CrossRef] [PubMed]
- Chavali, R.; Nejat, A.H.; Lawson, N.C. Machinability of CAD-CAM materials. J. Prosthet. Dent. 2017, 118, 194–199. [Google Scholar] [CrossRef] [PubMed]
- Lawson, N.C.; Bansal, R.; Burgess, J.O. Wear, strength, modulus and hardness of CAD/CAM restorative materials. Dent. Mater. 2016, 32, e275–e283. [Google Scholar] [CrossRef]
- Bosch, G.; Ender, A.; Mehl, A. A 3-dimensional accuracy analysis of chairside CAD/CAM milling processes. J. Prosthet. Dent. 2014, 112, 1425–1431. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Kirsch, C.; Ender, A.; Attin, T.; Mehl, A. Trueness of four different milling procedures used in dental CAD/CAM systems. Clin. Oral Investig. 2017, 21, 551–558. [Google Scholar] [CrossRef] [PubMed]
- Ender, A.; Mehl, A. Accuracy of complete-arch dental impressions: A new method of measuring trueness and precision. J. Prosthet. Dent. 2013, 109, 121–128. [Google Scholar] [CrossRef] [Green Version]
- Patzelt, S.B.; Emmanouilidi, A.; Stampf, S.; Strub, J.R.; Att, W. Accuracy of full-arch scans using intraoral scanners. Clin. Oral Investig. 2014, 18, 1687–1694. [Google Scholar] [CrossRef]
- Uhm, S.H.; Kim, J.-H.; Jiang, H.B.; Woo, C.-W.; Chang, M.; Kim, K.-N.; Bae, J.-M.; Oh, S.H. Evaluation of the accuracy and precision of four intraoral scanners with 70% reduced inlay and four-unit bridge models of international standard. Dent. Mater. J. 2017, 36, 27–34. [Google Scholar] [CrossRef] [Green Version]
- Kim, C.M.; Kim, S.R.; Kim, J.H.; Kim, H.Y.; Kim, W.C. Trueness of milled prostheses according to number of ball-end mill burs. J. Prosthet. Dent. 2016, 115, 624–629. [Google Scholar] [CrossRef]
- Srinivasan, M.; Cantin, Y.; Mehl, A.; Gjengedal, H.; Müller, F.; Schimmel, M. CAD/CAM milled removable complete dentures: An in vitro evaluation of trueness. Clin. Oral Investig. 2017, 21, 2007–2019. [Google Scholar] [CrossRef] [Green Version]
- International Organization for Standardization (ISO). Standard Atmospheres for Conditioning and/or Testing -Specifications; ISO 554: 1976; ISO: Geneva, Switzerland, 1976. [Google Scholar]
- International Organization for Standardization (ISO). Dentistry–Digitizing Devices for CAD/CAM Systems for Indirect Dental Restorations: Test Method for Assessing Accuracy; ISO 12836: 2015; ISO: Geneva, Switzerland, 2015. [Google Scholar]
- Rezende, C.E.E.; Borges, A.F.S.; Gonzaga, C.-C.; Duan, Y.; Rubo, J.H.; Griggs, J.A. Effect of cement space on stress distribution in Y-TZP based crowns. Dent. Mater. 2017, 33, 144–151. [Google Scholar] [CrossRef] [PubMed]
- Chen, X.-P.; Xiang, Z.-X.; Song, X.-F.; Yin, L. Machinability: Zirconia-reinforced lithium silicate glass ceramic versus lithium disilicate glass ceramic. J. Mech. Behav. Biomed. Mater. 2020, 101, 103435. [Google Scholar] [CrossRef] [PubMed]
- Kokubo, Y.; Tsumita, M.; Kano, T.; Sakurai, S.; Fukushima, S. Clinical marginal and internal gaps of zirconia all-ceramic crowns. J. Prosthodont. Res. 2011, 55, 40–43. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- McLean, J.W.; Von Fraunhofer, J.A. The estimation of cement film thickness by an in vivo technique. Br. Dent. J. 1971, 131, 107–111. [Google Scholar] [CrossRef] [PubMed]
Group | Materials | Shade | Basic Composition | Manufacturer |
---|---|---|---|---|
AM | Lithium disilicate | HT A2 | SiO2, Li2O, P2O5, Al2O3 other oxides and colorants | HASS Corp |
IPS | Lithium disilicate | HT A2 | SiO2, Li2O, K2O, MgO, Al2O3, P2O5 other oxides | Ivoclar Vivadent AG |
CEL | Zirconia-reinforced lithium silicate | HT A2 | SiO2, LiO2, ZrO2, P2O5, Al2O3, K2O, CeO2, other oxides | Sirona Dentsply |
Group | B (°C) | S (min) | t1 (°C/min) | T (min) | H (min) | V1 (°C) | V2 (°C) | L (°C) |
---|---|---|---|---|---|---|---|---|
AM | 400 | 6.00 | 60 | 810 | 15.00 | 550 | 810 | 680 |
IPS | 403 | 6.00 | 90 | 830 | 10.00 | 550 | 830 | 710 |
CEL | 500 | 3:30 | 60 | 820 | 1:00 | off | off | 750 |
Trueness Results (RMS) for the Outer and Inner Surface of the Ceramic Crowns | ||||||
---|---|---|---|---|---|---|
Group | Outer Surface | Inner Surface | ||||
Mean ± SD | 95% CI | p-Value | Mean ± SD | 95% CI | p-Value | |
AM | 38.30 ± 4.20 a,b | 35.97–40.63 | <0.001 | 58.76 ± 6.55 a | 55.13–62.38 | <0.016 |
IPS | 34.89 ± 4.74 a | 32.26–37.51 | 59.42 ± 8.89 a,b | 54.49–64.34 | ||
CEL | 40.38 ± 3.32 b | 38.54–42.21 | 67.12 ± 3.76 b | 65.03–69.20 |
Trueness Results (RMS) for the Internal and External Part of the Ceramic Crowns | ||||||
---|---|---|---|---|---|---|
Group | Internal Part | External Part | ||||
Mean ± SD | 95% CI | p-Value | Mean ± SD | 95% CI | p-Value | |
AM | 50.20 ± 7.37 a | 46.12–54.28 | <0.001 | 85.73 ± 19.30 a | 75.05–96.41 | <0.006 |
IPS | 41.32 ± 4.42 b | 38.87–43.76 | 108.11 ± 12.94 b | 100.9–115.3 | ||
CEL | 54.38 ± 3.72 a | 52.32–56.45 | 103.34 ± 12.40 ab | 96.48–110.2 |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Kang, S.-Y.; Yu, J.-M.; Lee, J.-S.; Park, K.-S.; Lee, S.-Y. Evaluation of the Milling Accuracy of Zirconia-Reinforced Lithium Silicate Crowns Fabricated Using the Dental Medical Device System: A Three-Dimensional Analysis. Materials 2020, 13, 4680. https://doi.org/10.3390/ma13204680
Kang S-Y, Yu J-M, Lee J-S, Park K-S, Lee S-Y. Evaluation of the Milling Accuracy of Zirconia-Reinforced Lithium Silicate Crowns Fabricated Using the Dental Medical Device System: A Three-Dimensional Analysis. Materials. 2020; 13(20):4680. https://doi.org/10.3390/ma13204680
Chicago/Turabian StyleKang, Seen-Young, Ji-Min Yu, Jun-Seok Lee, Ki-Sook Park, and Seung-Youl Lee. 2020. "Evaluation of the Milling Accuracy of Zirconia-Reinforced Lithium Silicate Crowns Fabricated Using the Dental Medical Device System: A Three-Dimensional Analysis" Materials 13, no. 20: 4680. https://doi.org/10.3390/ma13204680
APA StyleKang, S. -Y., Yu, J. -M., Lee, J. -S., Park, K. -S., & Lee, S. -Y. (2020). Evaluation of the Milling Accuracy of Zirconia-Reinforced Lithium Silicate Crowns Fabricated Using the Dental Medical Device System: A Three-Dimensional Analysis. Materials, 13(20), 4680. https://doi.org/10.3390/ma13204680