Programming Optimization in Implantable Cardiac Monitors to Reduce False-Positive Arrhythmia Alerts: A Call for Research
Abstract
:1. Introduction
2. Materials and Methods
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
Abbreviations
References
- Brignole, M.; Moya, A.; de Lange, F.J.; Deharo, J.C.; Elliott, P.M.; Fanciulli, A.; Fedorowski, A.; Furlan, R.; Kenny, R.A.; Martín, A.; et al. ESC Scientific Document Group. 2018 ESC Guidelines for the diagnosis and management of syncope. Eur. Heart J. 2018, 39, 1883–1948. [Google Scholar] [CrossRef] [PubMed]
- Ko, D.; Dai, Q.; Flynn, D.B.; Bosch, N.A.; Helm, R.H.; Monahan, K.M.; Andersson, C.; Anderson, C.D.; Walkey, A.J. Meta-Analysis of Randomized Clinical Trials Comparing the Impact of Implantable Loop Recorder Versus Usual Care After Ischemic Stroke for Detection of Atrial Fibrillation and Stroke Risk. Am. J. Cardiol. 2022, 162, 100–104. [Google Scholar] [CrossRef] [PubMed]
- Kheiri, B.; Sanghai, S.; Alhamoud, H.; Osman, M.; Stecker, E.; Henrikson, C.A.; Nazer, B. Meta-Analysis of Implantable Loop Recorder for Atrial Fibrillation Detection in Patients With Ischemic Stroke. Am. J. Cardiol. 2021, 161, 115–116. [Google Scholar] [CrossRef] [PubMed]
- Ciconte, G.; Giacopelli, D.; Pappone, C. The Role of Implantable Cardiac Monitors in Atrial Fibrillation Management. J. Atr. Fibrillation 2017, 10, 1590. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Bisignani, A.; De Bonis, S.; Mancuso, L.; Ceravolo, G.; Bisignani, G. Implantable loop recorder in clinical practice. J. Arrhythmia 2019, 35, 25–32. [Google Scholar] [CrossRef]
- Aggarwal, G.; Aggarwal, S.; Alla, V.; Narasimhan, B.; Ryu, K.; Jeffery, C.; Lakkireddy, D. Subcutaneouscardiac Rhythm Monitors: A Comprehensive Review. J. Atr. Fibrillation 2021, 13, 2387. [Google Scholar]
- Afzal, M.R.; Mease, J.; Koppert, T.; Okabe, T.; Tyler, J.; Houmsse, M.; Augostini, R.S.; Weiss, R.; Hummel, J.D.; Kalbfleisch, S.J.; et al. Incidence of false-positive transmissions during remote rhythm monitoring with implantable loop recorders. Heart Rhythm 2020, 17, 75–80. [Google Scholar] [CrossRef]
- Zanotto, G.; D’Onofrio, A.; Della Bella, P.; Solimene, F.; Pisanò, E.C.; Iacopino, S.; Dondina, C.; Giacopelli, D.; Gargaro, A.; Ricci, R.P. Organizational model and reactions to alerts in remote monitoring of cardiac implantable electronic devices: A survey from the Home Monitoring Expert Alliance project. Clin. Cardiol. 2019, 42, 76–83. [Google Scholar] [CrossRef] [Green Version]
- Maggi, R.; Rafanelli, M.; Ceccofiglio, A.; Solari, D.; Brignole, M.; Ungar, A. Additional diagnostic value of implantable loop recorder in patients with initial diagnosis of real or apparent transient loss of conscious- ness of uncertain origin. Europace 2014, 16, 1226–1230. [Google Scholar] [CrossRef] [Green Version]
- Maines, M.; Zorzi, A.; Tomasi, G.; Angheben Catanzariti, D.; Piffer, L.; Del Greco, M. Clinical impact, safety, and accuracy of the remotely monitored implantable loop recorder Medtronic Reveal LINQTM. Europace 2018, 20, 1050–1057. [Google Scholar] [CrossRef]
- Sanna, T.; Diener, H.C.; Passman, R.S.; di Lazzaro, V.; Bernstein, R.A.; Morillo, C.A.; Rymer, M.M.; Thijs, V.; Rogers, T.; Beckers, F.; et al. CRYSTAL AF Investigators. Cryptogenic stroke and underlying atrial fibrillation. N. Engl. J. Med. 2014, 370, 2478–2486. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Ciconte, G.; Saviano, M.; Giannelli, L.; Calovic, Z.; Baldi, M.; Ciaccio, C.; Cuko, A.; Vitale, R.; Giacopelli, D.; Conti, M.; et al. Atrial fibrillation detection using a novel three-vector cardiac implantable monitor: The atrial fibrillation detect study. Europace 2017, 19, 1101–1108. [Google Scholar] [CrossRef] [PubMed]
- Hindricks, G.; Pokushalov, E.; Urban, L.; Taborsky, M.; Kuck, K.H.; Lebedev, D.; Rieger, G.; Pürerfellner, H. XPECT Trial Investigators. Performance of a new leadless implantable cardiac monitor in detecting and quantifying atrial fibrillation: Results of the XPECT trial. Circ. Arrhythmia Electrophysiol. 2010, 3, 141–147. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Nölker, G.; Mayer, J.; Boldt, L.H.; Seidl, K.; VAN Driel, V.; Massa, T.; Kollum, M.; Brachmann, J.; Deneke, T.; Hindricks, G.; et al. Performance of an Implantable Cardiac Monitor to Detect Atrial Fibrillation: Results of the DETECT AF Study. J. Cardiovasc. Electrophysiol. 2016, 27, 1403–1410. [Google Scholar] [CrossRef]
- Lortz, J.; Varnavas, V.; Weißenberger, W.; Erbel, R.; Reinsch, N. Maintaining accurate long-term sensing ability despite significant size reduction of implantable cardiac monitors. Pacing Clin. Electrophysiol. 2016, 39, 1344–1350. [Google Scholar] [CrossRef]
- O’Shea, C.J.; Middeldorp, M.E.; Hendriks, J.M.; Brooks, A.G.; Harper, C.; Thomas, G.; Emami, M.; Thiyagarajah, A.; Feigofsky, S.; Gopinathannair, R.; et al. Remote Monitoring of Implantable Loop Recorders: False-Positive Alert Episode Burden. Circ. Arrhythmia Electrophysiol. 2021, 14, e009635. [Google Scholar] [CrossRef]
- Arrocha, A.; Klein, G.J.; Benditt, D.G.; Sutton, R.; Krahn, A.D. Remote electrocardiographic monitoring with a wireless implantable loop recorder: Minimizing the data review burden. Pacing Clin. Electrophysiol. 2010, 33, 1347–1352. [Google Scholar] [CrossRef]
- Søgaard, P.; Behrens, S.; Konyi, A.; Taborsky, M.; Christiansen, P.D.; Jacobsen, P.K.; Faluközy, J.; Malczynski, J.; Brachmann, J.; Samol, A.; et al. Transmission and loss of ECG snapshots: Remote monitoring in implantable cardiac monitors. J. Electrocardiol. 2019, 56, 24–28. [Google Scholar] [CrossRef]
- De Coster, M.; Demolder, A.; De Meyer, V.; Vandenbulcke, F.; Van Heuverswyn, F.; De Pooter, J. Diagnostic accuracy of R-wave detection by insertable cardiac monitors. Pacing Clin. Electrophysiol. 2020, 43, 511–517. [Google Scholar] [CrossRef]
- Bisignani, A.; De Bonis, S.; Mancuso, L.; Ceravolo, G.; Giacopelli, D.; Pelargonio, G.; Lanza, G.A.; Crea, F.; Bisignani, G. Are implantable cardiac monitors reliable tools for cardiac arrhythmias detection? An intrapatient comparison with permanent pacemakers. J. Electrocardiol. 2020, 59, 147–150. [Google Scholar] [CrossRef]
- Ciconte, G.; Vicedomini, G.; Giacopelli, D.; Calovic, Z.; Conti, M.; Spinelli, B.; Saviano, M.; Santinelli, V.; Pappone, C. Feasibility of remote monitoring using a novel long-dipole insertable cardiac monitor. JACC Clin. Electrophysiol. 2018, 4, 559–561. [Google Scholar] [CrossRef] [PubMed]
- Mariani, J.A.; Weerasooriya, R.; van den Brink, O.; Mohamed, U.; Gould, P.A.; Pathak, R.K.; Lin, T.; Conradie, A.; Illes, P.; Pavia, S.; et al. Miniaturized implantable cardiac monitor with a long sensing vector (BIOMONITOR III): Insertion procedure assessment, sensing performance, and home monitoring transmission success. J. Electrocardiol. 2020, 60, 118–125. [Google Scholar] [CrossRef] [PubMed]
- Bisignani, G.; De Bonis, S.; Bisignani, A.; Mancuso, L.; Giacopelli, D. Sensing performance, safety, and patient acceptability of long dipole cardiac monitor: An innovative axillary insertion. Pacing Clin. Electrophysiol. 2018, 41, 277–283. [Google Scholar] [CrossRef] [PubMed]
- Reinsch, N.; Ruprecht, U.; Buchholz, J.; Diehl, R.R.; Kälsch, H.; Neven, K. The BioMonitor 2 insertable cardiac monitor: Clinical experience with a novel implantable cardiac monitor. J. Electrocardiol. 2018, 51, 751–755. [Google Scholar] [CrossRef]
- Lacour, P.; Dang, P.L.; Huemer, M.; Parwani, A.S.; Attanasio, P.; Pieske, B.; Boldt, L.H.; Haverkamp, W.; Blaschke, F. Performance of the new BioMonitor 2-AF insertable cardiac monitoring system: Can better be worse? Pacing Clin. Electrophysiol. 2017, 40, 516–526. [Google Scholar] [CrossRef]
- Ooi, S.-Y.; Ng, B.; Singarayar, S.; Hellestrand, K.; Illes, P.; Mohamed, U.; Razak, S.; Weerasooriya, R.; BioMonitor 2 Study Investigators. BioMonitor 2 Pilot Study: Early experience with implantation of the Biotronik BioMonitor 2 implantable cardiac monitor. Heart Lung Circ. 2018, 27, 1462–1466. [Google Scholar] [CrossRef]
- Piorkowski, C.; Busch, M.; Nölker, G.; Schmitt, J.; Roithinger, F.X.; Young, G.; Taborsky, M.; Herrmann, G.; Schmitz, D. Clinical evaluation of a small implantable cardiac monitor with a long sensing vector. Pacing Clin. Electrophysiol. 2019, 42, 1038–1046. [Google Scholar] [CrossRef] [Green Version]
- Forleo, G.B.; Amellone, C.; Sacchi, R.; Lombardi, L.; Lucciola, M.T.; Scotti, V.; Viecca, M.; Schiavone, M.; Giacopelli, D.; Giammaria, M. Factors affecting signal quality in implantable cardiac monitors with long sensing vector. J. Arrhythmia 2021, 37, 1061–1068. [Google Scholar] [CrossRef]
- Nedios, S.; Romero, I.; Gerds-Li, J.H.; Fleck, E.; Kriatselis, C. Precordial electrode placement for optimal ECG monitoring: Implications for ambulatory monitor devices and event recorders. J. Electrocardiol. 2014, 47, 669–676. [Google Scholar] [CrossRef]
- Carretta, D.M.; Troccoli, R.; Spadaro Guerra, A.; De Vita, M.; Giacopelli, D.; D’Agostino, C. How to solve T-wave oversensing in an implantable cardiac monitor with a long sensing vector? Pacing Clin. Electrophysiol. 2020, 43, 1579–1581. [Google Scholar] [CrossRef]
- Lee, I.O.; Ukena, C.; Böhm, M.; Buob, A.; Neuberger, H.R. Underdetection of ventricular tachycardia by an implantable cardioverter-defibrillator due to a locally prolonged arrhythmia cycle length. Clin. Res. Cardiol. 2013, 102, 923–925. [Google Scholar] [CrossRef] [PubMed]
- Kokubun, H.; Miyauchi, Y.; Inada, S.; Shimizu, W. Underdetection of Ventricular Fibrillation by Wearable Cardioverter-Defibrillator Leading to Sudden Death. Circ. J. 2020, 85, 79. [Google Scholar] [CrossRef] [PubMed]
Parameter | Range of Values | Standard |
---|---|---|
Sensing Consult | Standard | Standard |
Sense after large PVCs | ||
Sense small PVCs | ||
Sense short intervals | ||
T-wave suppression | ||
Atrial fibrillation (AF) | ON; OFF | ON |
AF sensitivity | Low; Medium; High | Medium |
RR variability limit | 6; 9; 12; 15; 18% | 12% |
Confirmation time | 1… (1)… 6; 10; 20; 30 min | 6 min |
Bigeminy rejection | OFF; Standard; Aggressive | Standard |
High ventricular rate (HVR) | ON; OFF | ON |
HVR limit | 100… (10)… 200 bpm | 180 bpm |
HVR counter | 8… (4)… 24; 32; 48 cycles | 16 cycles |
Bradycardia | ON; OFF | ON |
Brady zone limit | 30… (5)… 80 bpm | 40 bpm |
Brady duration | 5… (5)… 30 s | 10 s |
Asystole duration | ON; OFF | ON |
Asystole duration | 2… (1)… 10 s | 3 s |
Sudden rate drop (SRD) | ON; OFF | OFF |
SRD rate decrease | 20… (10)… 70% | 50% |
SRD sensitivity | Low; Medium; High | Medium |
Study Population (n = 8) | |
---|---|
Age, years | 57.5 ± 23.2 |
Female, n (%) | 3 (37%) |
ICM indication, n (%) | |
Unexplained syncope | 4 (50%) |
Unexplained palpitations | 3 (37%) |
Cryptogenic stroke | 1 (13%) |
Comorbidities, n (%) | |
Hypertension | 5 (63%) |
Coronary artery disease | 1 (13%) |
R wave amplitude, mV | 0.41 ± 0.22 |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Guarracini, F.; Testolina, M.; Giacopelli, D.; Martin, M.; Triglione, F.; Coser, A.; Quintarelli, S.; Bonmassari, R.; Marini, M. Programming Optimization in Implantable Cardiac Monitors to Reduce False-Positive Arrhythmia Alerts: A Call for Research. Diagnostics 2022, 12, 994. https://doi.org/10.3390/diagnostics12040994
Guarracini F, Testolina M, Giacopelli D, Martin M, Triglione F, Coser A, Quintarelli S, Bonmassari R, Marini M. Programming Optimization in Implantable Cardiac Monitors to Reduce False-Positive Arrhythmia Alerts: A Call for Research. Diagnostics. 2022; 12(4):994. https://doi.org/10.3390/diagnostics12040994
Chicago/Turabian StyleGuarracini, Fabrizio, Martina Testolina, Daniele Giacopelli, Marta Martin, Francesco Triglione, Alessio Coser, Silvia Quintarelli, Roberto Bonmassari, and Massimiliano Marini. 2022. "Programming Optimization in Implantable Cardiac Monitors to Reduce False-Positive Arrhythmia Alerts: A Call for Research" Diagnostics 12, no. 4: 994. https://doi.org/10.3390/diagnostics12040994
APA StyleGuarracini, F., Testolina, M., Giacopelli, D., Martin, M., Triglione, F., Coser, A., Quintarelli, S., Bonmassari, R., & Marini, M. (2022). Programming Optimization in Implantable Cardiac Monitors to Reduce False-Positive Arrhythmia Alerts: A Call for Research. Diagnostics, 12(4), 994. https://doi.org/10.3390/diagnostics12040994