Next Article in Journal
Interferometric Wavefront Sensing System Based on Deep Learning
Next Article in Special Issue
Rapid Analysis of Chemical Composition and Physical Properties of Gemstones Using LIBS and Chemometric Technique
Previous Article in Journal
Opportunities and Threats of Mediterranean Evergreen Sclerophyllous Woody Species Subjected to Extreme Drought Events
Previous Article in Special Issue
Laser-Induced Breakdown Spectroscopy for Determination of Spectral Fundamental Parameters
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Quantitative Analysis of Pig Iron from Steel Industry by Handheld Laser-Induced Breakdown Spectroscopy and Partial Least Square (PLS) Algorithm

1
Istituto per la Scienza e Tecnologia dei Plasmi (ISTP-CNR)-Sede di Bari, 70126 Bari, Italy
2
Primary and Environmental Laboratories, ArcelorMittal Italia S.p.A., Taranto Steelworks, 74123 Taranto, Italy
3
Physics Institute, Federal University of Mato Grosso do Sul, Campo Grande, MS 79070-900, Brazil
*
Author to whom correspondence should be addressed.
Appl. Sci. 2020, 10(23), 8461; https://doi.org/10.3390/app10238461
Submission received: 12 October 2020 / Revised: 13 November 2020 / Accepted: 25 November 2020 / Published: 27 November 2020
(This article belongs to the Special Issue Laser Induced Plasma/Breakdown Spectroscopy)

Abstract

:
One of the main objectives in the steel production process is to obtain a blast furnace pig iron of good quality and at the lowest possible cost. In general, the quality of pig iron is evaluated on the basis of its chemical composition determined by X-ray fluorescence laboratory equipment. In the present study, the performance of a handheld (h) laser-induced breakdown spectroscopy (LIBS) instrument in the identification and the quantification of the relevant elements C, Mn, P, Si, and Ti in forty-six blast furnace pig iron samples was tested successfully. The application of two different models, i.e., univariate and multivariate partial least square (PLS) calibration and validation, to the whole LIBS data set showed that the latter approach was much more efficient than the former one in quantifying all elements considered, especially Si and Ti.

1. Introduction

The main metallic materials and intermediate products involved in iron and steel industry include pig iron, cast iron, carbon steel, stainless steel, and tool steel, all of which require continuous, accurate, and precise analyses to keep the entire production process under control. In particular, pig iron consists of a carbon-rich intermediate product resulting from a mixture of iron ore, coke, and limestone burnt together in a blast furnace, which is then refined into steel in an oxygen furnace [1,2]. The main goal of the production process is to obtain pig iron of good quality at the lowest possible cost. Many factors affect the quality of pig iron, among which are the quality of ferrous burden materials and coke as well as the fuels and the methods used in the blast furnace process [3]. In particular, the content of the elements Si, Mn, P, and S, which depend on their abundance in ferrous burden materials [3], and Ti, which critically affects its precipitation with saturated carbon [4], have great importance for the quality of pig iron. According to relevant customer requirements and standards, the chemical composition of pig iron should include, besides the basic element, Fe, and a high C content, a number of other alloying elements that are Si at 0.50–0.80% wt., Mn at >0.20% wt., P at <0.11% wt., and S at <0.030% wt. for the production of rail steel and 0.14% wt. for other kinds of steel [3,5]. However, every steel plant around the world requires a different elemental composition. For example, the best performance of the blast furnace operating at the ArcelorMittal Italia S.P.A. steel plant located in Taranto, Italy, where the samples examined in this work were collected, is achieved at an optimal Si content in pig iron of 0.60% wt.
Wavelength-dispersive X-ray fluorescence (WDXRF) spectrometry is a well-established analytical technique in iron and steel industry and is routinely used to analyse raw materials, sinters, slags, and the elemental composition of cast iron and steels when coupled with optical emission spectrometry (OES) [2]. Although this technique can achieve great accuracy and precision, the instrumentation is not portable and is quite expensive. Furthermore, in iron and steel industry, the samples to be analysed must be representative of an enormous body of material, such as a shipload of iron ore, a stockpile of raw material, and intermediate and final products from the plant. This aspect involves a very marked reduction in size in deriving a laboratory-sized sample of a few kg from many thousands of tons of original material, thus it is important to implement an analytical procedure that can minimize the errors associated with sampling [2]. In recent years the demand for lower limits of detection (LOD), improved analytical reliability, strict control of alloy composition, and capability to place the analytical instrumentation as close as possible to the process so to facilitate and speed up any corrective actions during production have prompted manufacturers to design more sensitive and flexible instruments that could be easily incorporated into transportable equipment.
Laser-induced breakdown spectroscopy (LIBS) has been used widely as a versatile alternative analytical technique in the steel industry, e.g., for iron ore selection, process control, and iron slag analysis [6,7,8,9,10]. Furthermore, in the last years, the technique has progressed so to allow the development of handheld, self-contained commercial instruments able to analyse samples of various nature directly on site [11,12,13,14]. Although a set of recent manufacturer application notes have illustrated how the handheld (h) LIBS analysers could be used for material identification in steel industry [15,16], thus far, few scientific papers have been published on this topic. Afgan et al. [17] used a first generation specifically designed NanoLIBS-Q hLIBS instrument produced by B&W Tek (Newark, DE, USA) for steel analysis. Calibration curves were developed for Cr, C, Si, Cu, Mn, Ni, Mo, and V using the dominant factor-based partial least square regression (PLSR) method with and without spectral standardization. Average absolute measurement errors of 0.019, 0.039, 0.013, and 0.001% were achieved respectively for Si, Cr, Mn, and Ni. Senesi et al. [18] used a second generation B&W Tek (Newark, DE, USA) NanoLIBS instrument to qualitatively analyse pig iron in open air. The emission lines of Si, Mn, and Fe were measured within the 280–315 nm spectral window, with results comparing favourably to those achieved using a typical LIBS benchtop apparatus. In particular, the Si emission line of interest could be rapidly identified with a high degree of confidence. Very recently, Poggialini et al. [19] improved substantially the analytical performance of a Bruker EOS 500 hLIBS instrument (Billerica, MA, USA) for steel analysis using a nonlinear calibration approach based on a set of artificial neural networks (ANNs), one optimized for the determination of major elements in the alloy and the others specialized for the analysis of minor components. Although tests performed on steel samples used for instrument internal calibration achieved an accuracy comparable with results of ANN treatments, the latter ones yielded much more accurate results when unknown samples were tested.
The present work aimed to evaluate the performance of hLIBS supported by two different calibration models, i.e., univariate calibration and multivariate PLS calibration, in the identification and the quantification of the elements C, Mn, P, Si, and Ti in a number of blast furnace pig iron samples, in comparison to that of X-ray fluorescence (XRF) analysis. The predictive capability of each calibration model was evaluated by calculating the Pearson’s correlation coefficient (R), root mean square error (RMSE), and mean absolute prediction error (MAPE) and LODs, whereas the leave one out cross validation (LOO-CV) method was used to perform external validation.

2. Materials and Methods

2.1. Pig Iron Samples

Forty-six pig iron samples were collected from the blast furnaces operating at the industrial and manufacturing steel plant of ArcelorMittal Italia S.P.A. located in Taranto, Italy. Each sample was analysed in the form of a circular disc of 35 mm diameter and 7 mm thickness with one face flat and the other rough and presenting small shrinkages. The matrix element in all samples was Fe at a content of over 90%.

2.2. Handheld LIBS Instrumentation and Analytical Procedure

The B&W Tek (Newark, DE, USA) NanoLIBS hLIBS instrument has been described in detail by Senesi et al. [20]. Briefly, the instrument consisted of a miniature-diode-pumped, solid-state, short-pulsed laser emitting at the wavelength of 1064 nm and operating at a maximum pulse energy of 150 µJ with a pulse duration of 500 ps and a high repetition rate comprised between 1 and 5 KHz. The system was equipped with a low-resolution compact spectrometer operating in the non-gated mode and covering the range from 180 to 800 nm with an overall resolution of 0.4 nm. The whole setup was enveloped in a lightweight handheld body with mass and dimensions of about 1.8 Kg and 26 × 10 × 30 cm, respectively. The instrument was provided with a rechargeable Li-ion battery that allowed up to 8 h autonomy for field operations. The system was equipped with a rastering scan that provided a spatial averaging. The measurements were performed by placing the instrument against the sample surface, and the analysis was started via a trigger. A sensor installed in the instrument head controlled automatically the laser output so that the laser could operate only when the head was in contact with the sample. Ten different positions were analysed on the sample, and a total of 150 spectra corresponding to 1800 laser shots were acquired and averaged to obtain a single spectrum for each position (Figure 1). Then, the spectra were analysed qualitatively to identify the emission lines using the National Institute of Standards and Technology (NIST) database [21].

2.3. XRF Analysis

The reference concentrations of C, Si, Mn, P, and Ti in the pig iron samples were determined by a MagiX FAST simultaneous XRF spectrometer (PANalytical B.V., Almelo, The Netherlands) equipped with a 300 µm brass filter, a 1 mm Pb beam-stop, and a 4 kW Rh Super Sharp end-window Tube set at 60 kV and 66 mA to obtain the optimal measurement of heavier elements.

2.4. Data Pre-Treatment

In order to reduce the noise from experimental fluctuations and matrix effects, a pre-treatment was carried out on each acquired spectrum. The first step consisted in applying the standard normal variate (SNV) procedure [22], i.e., an algebraic manipulation of the spectra to have a total sum fixed at zero and the standard deviation unitary (Equation (1)):
y i = y i y ¯ s t d
where y i is the transformed wavelength intensity, y i is the original one, and the constants y ¯ and s t d are, respectively, the mean and the standard deviation of the spectrum.
In the second step, a Savitzky–Golay (SG) filter was applied to smooth the data, reduce the noise, and enhance the performance of the PLS model [23]. In particular, the polynomial SG smoothing method was applied and consisted in selecting N points in the spectrum and adjusting a polynomial function of order P. This adjustment generated new data points that were used to generate a new spectrum. This process was repeated by sweeping the entire spectrum, and the values of N and P were optimized to obtain the smallest validation error.
In the third step, the spectrum was cut into the selected ranges covering the emission lines of the elements C, Mn, P, Si, and Ti of interest. The emission lines were chosen based on their intensity in the spectrum and on the absence or minimization of interferences due to other transitions. The whole process was performed manually using the averaged spectra and the NIST database reference [21].

2.5. Statistical Analysis

To assess the quality of calibration curves, which directly compare the intensity obtained from the spectrum with the reference elemental concentration of the sample measured by XRF, the Pearson’s R coefficient (Equation (2)) and the RMSE (Equation (3)) were used:
R = i = 1 N ( x i x ¯ σ x ) j = 1 N ( y j y ¯ σ y ) N 1
where x i and y i are the wavelength and the LIBS intensity values, respectively, x ¯ , y ¯ and σ x , σ y are the mean and standard deviation from the distributions, and N is the number of points; and
R M S E = i = 1 N ( y r e f i y p r e d i ) 2 N
where y r e f i and y p r e d i are, respectively, the reference and the predicted concentration values for the model, and N is the number of points. The R value measures the degree of correlation between two variables and ranges from −1 (perfect negative correlation) to +1 (perfect positive correlation). The RSME basically measures the difference between predicted and measured values. Thus, the values of R and RSME provide complementary information for model evaluation.
Thus, the analytical performance of each calibration method used for the quantification of a given element can be evaluated by the R value of the best fitting line between predicted and nominal values (R = 1 in the case of perfect agreement), the intercept value (ideally equal to zero), and the RMSE of the prediction (equal to zero in case of perfect agreement).
Further, to evaluate the validation curve, the MAPE values was calculated (Equation (4)), which measures the average percentage error of the value predicted by the model compared to the reference value:
M A P E = 100 × 1 N i = 1 N | y r e f i y p r e d i y r e f i |
where N is the total number of samples, and y r e f i and y p r e d i are, respectively, the reference and the predicted concentration values used in the model. The MAPE value provides the quantification error expected for tested samples.
The LOD values were calculated using the calibration curve according to the following Equation (5):
L O D = 3 × σ α
where σ is the error of the linear coefficient, and α represents the slope of the calibration curve. The LOD represents the minimum concentration value of the analyte that can be measured by the technique used, i.e., the technique sensitivity as the detection of concentrations below the LOD are not possible.
In order to avoid possible overfitting, which may occur if data are adjusted very well to the training data but exhibit a low predictive capacity, they were validated by the LOO-CV method by which one sample is taken out from the training set and used later for testing. In this way, a concentration can be predicted for the external sample and compared with the reference value. The process was repeated by leaving each sample out of the training so that all samples could be validated by the model trained with the remaining N-1 samples. Although this validation process is more complicated computationally, it can achieve a result more reliable than that obtained by the k-fold validation method, as the former one is independent from the random selection of the training set. As the test sample does not participate in the training set, it works as an external measure. The only relationship between the test sample and the training data set was that the measurements were carried out in a similar way (same day and equipment). Although the ideal procedure would be to validate the protocol using measurements in situ and perform it on different days, this is strictly required only in the final stages of implementation of the equipment for industrial applications. However, to present the potential of spectroscopic techniques applications, the cross-validation approach is widely used and is considered an excellent evaluation parameter.

2.6. Univariate and Multivariate Analyses

The first stage of analysis was performed by calculating the area of specific emission lines for each element of interest in the raw spectra. The area below each line was fitted to a Gaussian function by subtracting the background [24], whereas the area of interfered emission lines was calculated by fitting multiple Gaussian functions [25]. A curve was constructed for each emission line, and the R coefficient and the RMSE values were calculated for the calibration curve and MAPE values for the validation curve. Only the emission lines that showed the best R coefficients in the calibration are discussed in Section 3. For calibration curves with very low R (<0.60), the validation process was not performed, as the associated error was very high.
The multivariate analysis of data was performed on the spectra after SNV normalization and SG filter by the PLS method [26,27], which takes into account any possible interdependence of the emission lines of each element of interest and both the background, allowing one to reduce possible matrix effects [28] and the emission lines of other elements [29]. In this study, the system was optimized to operate only in the spectral range width of emission lines of the elements of interest, which included also variables related to the background and occasionally emissions of other elements close to the transition. SG parameters were optimized to reduce the validation error. To evaluate the quality of analysis, the coefficient R and RMSE values of the calibration curves and MAPE values of the validation curves were calculated.
As the emission line(s) of each element of interest selected after data pre-treatment procedure presented a slight deviation (in the order of 0.1 nm), probably due to the limited spectral resolution, from those in the NIST database, in multivariate analysis, the point of highest intensity of the line was selected as the centre of the transition. The selected spectral width of the emission line was optimized also to minimize the validation error. Thus, the model included part of the background and possible neighbouring transitions and interferences, which is a procedure desirable in multivariate analysis, as the inclusion of extra variables can help the model to circumvent laser fluctuation problems, matrix effects, and interferences [29]. A workflow summarizing the analytical steps performed is shown in Figure 2.

3. Results and Discussion

The surface morphology of the craters formed after two consecutive laser shots on a typical pig iron sample (Figure 3a) was investigated by optical microscopy. Images obtained (Figure 3b) indicated that the energy density used was sufficient to provide sample ablation and crater formation only after two laser shots. Further, the laser-impact areas in Figure 3b showed a central intensely ablated region and a rim around it where some droplets and cracks were present. Some black spots could be also observed with pinholes appearing on the surface. Further, the surface colour changed to black, and the surface reflectivity decreased significantly. The crater diameters were estimated to be about 300 μm (Figure 3b).
The full broadband LIBS emission spectra (one example is shown in Figure 4) acquired directly within few seconds by a point and shoot operation without any sample preparation were used to identify the elements present in the pig iron samples. Although all spectra featured an intense background, especially in the wavelength range from 200 to 300 nm where multiple lines of the major matrix element Fe appeared, the main elemental components of the samples, i.e., C, Mn, P, Si, and Ti, could be revealed. The emission lines chosen for further analysis as a result of data pre-treatment procedure were: Si I (251.432 nm, 263.128 nm, 288.158 nm); Ti II (307.522 nm, 334.903 nm, 336.121 nm); Mn II (257.610 nm, 296.595 nm) and Mn I (404.876 nm); P I (215.294 nm, 253.560 nm); C I (193.027 nm, 247.856 nm).

3.1. Univariate Analysis

The univariate calibration curves achieved using the emission areas of the elements C, Mn, and P featured a poor linearity with too low (R < 0.90) correlation coefficients, which affected the accuracy and the precision of the prediction and the robustness of the model. Differently, the univariate calibration and cross-validation curves constructed using the areas of the emission lines of Si I at 288.158 nm and Ti II at 334.903 nm and 334.940 nm, which were those that presented the best results in the calibration, versus the reference XRF concentrations were more satisfactory (Figure 5). In the case of Ti, the two emission lines considered were overlapping, and a unique area was calculated which included both lines. An R coefficient of 0.95 for Si and 0.86 for Ti and an RMSE of 237.05 for Si and 126.10 for Ti were achieved for calibration with a MAPE of 18.04% for Si and 19.40% for Ti for validation. Although these values could be considered satisfactory, they would have limited application in real cases, i.e., univariate calibration curves could be used only for a rapid estimation of some elements in the samples but not for an accurate estimation of unknown samples.
Furthermore, building univariate calibration curves by plotting LIBS emission line areas vs. percentage of weight concentration might introduce matrix effect errors because the same numerical concentration in different matrices might correspond to a different weight concentration on dependence on the stoichiometry of the sample. Thus, the univariate model yielded an unsatisfactory outcome, therefore it was discarded, and the PLS multivariate calibration approach was attempted to improve the analysis of LIBS spectra in both broad spectral range and selected emission lines to possibly allow the technique to be used in industrial application. In multivariate analysis, the choice of weight or number concentration for the reference samples does not make a difference, because all emission lines are analysed simultaneously.

3.2. Multivariate Analysis

Before applying the PLS model, the LIBS spectra were normalized using the SNV method and an SG filter to smooth and reduce noise. In multivariate analysis, all the elemental emission lines selected were used simultaneously together with a portion of the lateral background, and the number of components used in the model was optimized to achieve the smallest error.
Figure 6 shows raw LIBS spectra for all the samples compared with detailed SNV and SNV + SG spectra in a specific spectral region (310–365 nm) and illustrates how the filter directly affects the transitions and the background attenuating fluctuations. As a verification criterion to estimate the noise reduction achieved by the applied filters, three spectral configurations were considered, i.e., raw without any treatment, SNV normalized, and SNV-SG corrected of the transition of Fe II (266.663 nm) (Figure 7). This transition was chosen because it was apparently not inferring interferences, and the concentration of Fe in the sample was very high and practically constant. To ensure an identical analytical condition for the three configurations, a baseline correction and normalization was performed to guarantee that the set of spectra remained within a minimum of zero and a maximum of one in intensity. This process did not affect the fluctuation among spectra (Figure 7), as it was applied to the set of spectra as a whole and not individually to each spectrum. A sum of pixel intensities was performed, which resulted in a transition intensity/area value (Riemann sum) for each spectrum. Then, the mean of these intensities and the standard deviation were calculated. The percentage value of the standard deviation related to the mean value was 31.5% for the raw spectra set, 3.7% for the SNV set, and 1.9% for the SNV-SG set. These results show that the process as a whole was able to reduce global spectral fluctuations and probably was a collaborative factor in improving the multivariate model.
The calibration and the validation curves of the elements C, Mn, P, Si, and Ti constructed by applying PLS to the areas of the corresponding elemental lines versus the certified concentrations are shown in Figure 8. The R and the RMSE values achieved for the calibration curves and MAPE and LOD for the validation curves using univariate and multivariate models (Table 1) indicated a significant improvement achieved by PLS and PLS + SG data treatment. In particular, the R (above 0.93) and the RMSE (below 0.07, with the exception of C) values of the calibration curves constructed using the PLS approach were very satisfactory for all elements, and especially for Si and Ti. Additionally, the MAPE percentage values achieved in the validation by PLS (below 15% for all elements except Mn) were also satisfactory.
Although the hLIBS approach might present intrinsic laser fluctuations leading to an unpredictable interaction between radiation and matter, overfitting processes could be totally avoided by performing an efficient cross-validation procedure. Thus, the error due to equipment and laser matter interaction could be minimized, but the error associated with the reference technique could not.

4. Conclusions

The primary qualitative factor of pig iron, which specifies its class and features required by the steel plant, is its chemical composition. In particular, one of the most important indicators of the technical level and the quality of steel production is the Si content achieved in the intermediate pig iron produced in the blast furnace. This study demonstrates that broadband LIBS analysis performed by an hLIBS instrument associated with PLS treatment of spectral data is able to identify and quantify different elements, specifically C, Mn, P, Si, and Ti, in pig iron samples. Thus, the capacity of an hLIBS instrument to perform spatially resolved, multi-element, in-situ, online analysis of element concentrations in open air without any sample pre-treatment provides an obvious advantage with respect to the use of a traditional laboratory bench-top apparatus.
Although hLIBS cannot be considered yet totally satisfactory in terms of accuracy, the time reduction from sampling to result delivery and the limited sample preparation no doubt contribute to a more rapid and efficient control of the steel-making process. The use of the hLIBS instrument is thus expected to streamline operations in the blast furnace, gain real-time information of sample quality, and allow a rapid decision of which samples to consider for further laboratory analyses, thus saving time and costs and minimizing transportation to the laboratory. Further, testing of industrial materials by portable instruments is often performed in field harsh conditions, thus easy portability, high robustness, and fast data acquisition are required for these instruments to incorporate easily the analytical process into the production and the processing workflows. Finally, a variety of industrial applications of hLIBS systems can be expected by the progress of spectral data evaluation tools.

Author Contributions

Conceptualization, G.S.S.; Data curation, B.S.M.; Formal analysis, G.S.S.; Investigation, G.S.S. and B.S.M.; Methodology, G.S.S. and B.S.M.; Resources, A.B.; Supervision, G.S.S.; Validation, B.S.M.; Writing—original draft, G.S.S.; Writing—review & editing, G.S.S., O.D.P. and B.S.M. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Acknowledgments

The authors acknowledge Davide Manzini from MADAtec s.r.l. and B&W Tek for providing the NanoLIBS handheld instrument used in this work.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Mei, Y.; Cheng, S.; Niu, Q.; Xu, W.; Ge, J. A review on transfer mechanism and influence factors of silicon in blast furnace. Ironmak. Steelmak. 2020, 47, 246–262. [Google Scholar] [CrossRef]
  2. Bonvin, D. X-ray fluorescence spectrometry in the iron and steel industry. In Encyclopedia of Analytical Chemistry; Meyers, R.A., Ed.; John Wiley & Sons Ltd.: Chichester, UK, 2000; pp. 9009–9028. [Google Scholar]
  3. Kardas, E. The effect of quality of ferrous burden materials on the quality of pig iron. In Proceedings of the Metal, 21st International Conference on Metallurgy and Materials, Brno, Czech Republic, 23–25 May 2012. [Google Scholar]
  4. Gao, L.-Z.; Ma, T.-X.; Hu, M.-J.; Yan, Z.-M.; Lü, X.-W.; Hu, M.-W.L. Effect of titanium content on the precipitation behavior of carbon-saturated molten pig iron. Int. J. Min. Met. Mater. 2019, 26, 483–491. [Google Scholar] [CrossRef]
  5. PN-EN 10001:1996. Pig Iron—Definition and Classification; Polish Committee of Standardization: Warszawa, Poland, 1996. [Google Scholar]
  6. Bette, H.; Brysch, A.; Kraushaar, M.; Monch, I.; Peter, L.; Sturm, V. Laser-induced breakdown spectrometry—Applications for production control and quality assurance in the steel industry. Spectrochim. Acta Part B 2001, 56, 637–649. [Google Scholar]
  7. Noll, R.; Sturm, V.; Aydin, U.; Eilers, D.; Gehlen, C.; Hohne, M.; Lamott, A.; Makowe, J.; Vrenegor, J. Laser-induced breakdown spectroscopy—From research to industry, new frontiers for process control. Spectrochim. Acta Part B 2008, 63, 1159–1166. [Google Scholar] [CrossRef]
  8. Noll, R. Laser-Induced Breakdown Spectroscopy: Fundamentals and Applications; Springer: Berlin/Heidelberg, Germany, 2012. [Google Scholar]
  9. Khater, M.A. Laser-induced breakdown spectroscopy for light elements detection in steel: State of the art. Spectrochim. Acta Part B 2013, 81, 1–10. [Google Scholar] [CrossRef]
  10. Wang, Z.; Deguchi, Y.; Shiou, F.; Yan, J.; Liu, J. Application of laser-induced breakdown spectroscopy to real-time elemental monitoring of iron and steel making processes. ISIJ Int. 2016, 56, 723–735. [Google Scholar] [CrossRef] [Green Version]
  11. Noll, R.; Fricke-Begemann, C.; Brunk, M.; Connemann, S.; Meinhardt, C.; Scharun, M.; Sturm, V.; Makowe, J.; Gehlen, C. Laser-induced breakdown spectroscopy expands into industrial applications. Spectrochim. Acta Part B 2014, 93, 41–51. [Google Scholar] [CrossRef]
  12. Noll, R.; Fricke-Begemann, C.; Connemann, S.; Meinhardt, C.; Sturm, V. LIBS analyses for industrial applications—An overview of developments from 2014 to 2018. J. Anal. Spectrom. 2018, 33, 945–956. [Google Scholar] [CrossRef] [Green Version]
  13. Senesi, G.S.; Harmon, R.S.; Hark, R.R. Field-portable and handheld LIBS. In Laser-Induced Breakdown Spectroscopy, 2nd ed.; Singh, J.P., Thakur, S.N., Eds.; Elsevier: Amsterdam, The Netherlands, 2020; pp. 537–560. [Google Scholar]
  14. Senesi, G.S.; Harmon, R.S.; Hark, R.R. Field-portable and handheld LIBS: Historical review, current status and future prospects. Spectrochim. Acta Part B 2021, 175, 106013. [Google Scholar] [CrossRef]
  15. SciAps, Inc. Application Note, In-Field Measurement of Mn:C Ratios for Steel Embrittlement with Z LIBS Handheld Analyzer. 2017. Available online: https://www.sciaps.com/wp-content/uploads/2016/07/SciAps-ApNote_Mn-C-ratios-steel.pdf?x61443 (accessed on 5 October 2020).
  16. SciAps, Inc. Application Note, Carbon Analysis in Stainless and Carbon Steels with Handheld LIBS. 2018. Available online: https://www.sciaps.com/wp-content/uploads/2018/04/SciAps_ApNote_LIBSCarbonMARCH2018.pdf?x24702 (accessed on 5 October 2020).
  17. Afgan, M.S.; Hou, Z.; Wang, Z. Quantitative analysis of common elements in steel using a handheld μ-LIBS instrument. J. Anal. At. Spectrom. 2017, 32, 1905–1915. [Google Scholar] [CrossRef]
  18. Senesi, G.S.; Dilecce, G.; Bove, A.; De Pascale, O. Laser-induced breakdown spectroscopy (LIBS) in open air applied to the elemental analysis, specifically Si, in pig iron samples. Comparison from a handheld instrument and a bench-top apparatus. In Proceedings of the Thirteenth Frontiers in Low-Temperature Plasma Diagnostics & First Frontiers in Low-Temperature Plasma Simulations, Bad Honnef, Germany, 13–16 May 2019. [Google Scholar]
  19. Poggialini, F.; Campanella, B.; Legnaioli, S.; Pagnotta, S.; Raneri, S.; Palleschi, V. Improvement of the performances of a commercial hand-held laser-induced breakdown spectroscopy instrument for steel analysis using multiple artificial neural networks. Rev. Sci. Instrum. 2020, 91, 073111. [Google Scholar] [CrossRef] [PubMed]
  20. Senesi, G.S.; Manzari, P.; Consiglio, A.; De Pascale, O. Identification and classification of meteorites using a handheld LIBS instrument coupled with a fuzzy logic-based method. J. Anal. Spectrom. 2018, 33, 1664–1675. [Google Scholar] [CrossRef]
  21. NIST Database. 2020. Available online: https://physics.nist.gov/PhysRefData/ASD/LIBS/libsform.html (accessed on 10 September 2020).
  22. Syvilay, D.; Wilkie-Chancellier, N.; Trichereau, B.; Texier, A.; Martinez, L.; Serfaty, S.; Detalle, V. Evaluation of the standard normal variate method for Laser-Induced Breakdown Spectroscopy data treatment applied to the discrimination of painting layers. Spectrochim. Acta Part B 2015, 114, 38–45. [Google Scholar] [CrossRef]
  23. Yang, Y.; Li, C.; Liu, S.; Min, H.; Yan, C.; Yang, M.; Yu, J. Classification and identification of brands of iron ores using laser-induced breakdown spectroscopy combined with principal component analysis and artificial neural networks. Anal. Methods 2020, 12, 1316–1323. [Google Scholar] [CrossRef]
  24. Marangoni, B.S.; Silva, K.S.G.; Nicolodelli, G.; Senesi, G.S.; Cabral, J.S.; Villas-Boas, P.R.; Silva, C.S.; Teixeira, P.C.; Nogueira, A.R.A.; Benites, V.M.; et al. Phosphorus quantification in fertilizers using laser induced breakdown spectroscopy (LIBS): A methodology of analysis to correct physical matrix effects. Anal. Methods 2016, 8, 78–82. [Google Scholar] [CrossRef] [Green Version]
  25. Nicolodelli, G.; Marangoni, B.S.; Cabral, J.S.; Villas-Boas, P.R.; Senesi, G.S.; Dos Santos, C.H.; Romano, R.A.; Segnini, A.; Lucas, Y.; Montes, C.R.; et al. Quantification of total carbon in soil using laser-induced breakdown spectroscopy: A method to correct interference lines. Appl. Opt. 2014, 53, 2170–2176. [Google Scholar] [CrossRef] [PubMed]
  26. Andrade, J.M.; Cristoforetti, G.; Legnaioli, S.; Lorenzetti, G.; Palleschi, V.; Shaltout, A.A. Classical univariate calibration and partial least squares for quantitative analysis of brass samples by laser-induced breakdown spectroscopy. Spectrochim. Acta Part B 2010, 65, 658–663. [Google Scholar] [CrossRef]
  27. Tucker, J.M.; Dyar, M.D.; Schaefer, M.W.; Clegg, S.M.; Wiens, R.C. Optimization of laser induced breakdown spectroscopy for rapid geochemical analysis. Chem. Geol. 2010, 277, 137–148. [Google Scholar] [CrossRef]
  28. Takahashi, T.; Thornton, B. Quantitative methods for compensation of matrix effects and self-absorption in Laser Induced Breakdown Spectroscopy signals of solids. Spectrochim. Acta Part B 2017, 138, 31–42. [Google Scholar] [CrossRef]
  29. Pochon, A.; Desaulty, A.-M.; Bailly, L. Handheld laser-induced breakdown spectroscopy (LIBS) as a fast and easy method to trace gold. J. Anal. Spectrom. 2020, 35, 254–264. [Google Scholar] [CrossRef]
Figure 1. Sketch representing the laser-induced breakdown spectroscopy (hLIBS) analysis.
Figure 1. Sketch representing the laser-induced breakdown spectroscopy (hLIBS) analysis.
Applsci 10 08461 g001
Figure 2. Workflow of the analytical steps performed.
Figure 2. Workflow of the analytical steps performed.
Applsci 10 08461 g002
Figure 3. The pig iron circular disc showing the rough face with small shrinkages (a); laser beam craters (red circles) produced on the surface (b).
Figure 3. The pig iron circular disc showing the rough face with small shrinkages (a); laser beam craters (red circles) produced on the surface (b).
Applsci 10 08461 g003
Figure 4. Typical LIBS spectrum acquired with the NanoLIBS hLIBS instrument for a pig iron sample.
Figure 4. Typical LIBS spectrum acquired with the NanoLIBS hLIBS instrument for a pig iron sample.
Applsci 10 08461 g004
Figure 5. Univariate calibration (a,c) and validation (b,d) curves for Si and Ti versus certified X-ray fluorescence (XRF) concentrations.
Figure 5. Univariate calibration (a,c) and validation (b,d) curves for Si and Ti versus certified X-ray fluorescence (XRF) concentrations.
Applsci 10 08461 g005
Figure 6. Raw, standard normal variate (SNV), and SNV + Savitzky–Golay (SG) LIBS spectra in a specific spectral region (310–365 nm).
Figure 6. Raw, standard normal variate (SNV), and SNV + Savitzky–Golay (SG) LIBS spectra in a specific spectral region (310–365 nm).
Applsci 10 08461 g006aApplsci 10 08461 g006b
Figure 7. Raw, SNV, and SNV + SG LIBS spectra of the transition of Fe II at 266.663 nm.
Figure 7. Raw, SNV, and SNV + SG LIBS spectra of the transition of Fe II at 266.663 nm.
Applsci 10 08461 g007aApplsci 10 08461 g007b
Figure 8. Calibration (a,c,e,g,i) and validation (b,d,f,h,j) curves for the elements C, Mn, P, Si, and Ti constructed by partial least square (PLS) treatment of data versus XRF certified concentrations.
Figure 8. Calibration (a,c,e,g,i) and validation (b,d,f,h,j) curves for the elements C, Mn, P, Si, and Ti constructed by partial least square (PLS) treatment of data versus XRF certified concentrations.
Applsci 10 08461 g008aApplsci 10 08461 g008b
Table 1. Results achieved by applying univariate, SNV + PLS and SNV + SG + PLS models.
Table 1. Results achieved by applying univariate, SNV + PLS and SNV + SG + PLS models.
CalibrationCalibrationCalibrationValidation
ElementRRMSELOD (%)MAPE (%)
Si (Univariate)0.95237.050.19518.04
Ti (Univariate)0.86126.100.01719.40
Mn (Univariate)0.12nullnullNull
P (Univariate)0.26nullnullNull
C (Univariate)0.19nullnullNull
Si (SNV + PLS)0.990.1510.0620.49
Ti (SNV + PLS)0.980.0120.00622.08
Mn (SNV + PLS)0.54nullnullNull
P (SNV + PLS)0.850.080.037.87
C (SNV + PLS)0.960.10.631.89
Si (SG + SNV + PLS)0.990.0670.0614.07
Ti (SG + SNV + PLS)0.990.00260.002611.61
Mn (SG + SNV + PLS)0.930.030.03115.42
P (SG + SNV + PLS)0.960.0160.0167.63
C (SG + SNV + PLS)0.970.580.581.3
The best SG were with seven points and order three polynomium. R: Pearson’s correlation coefficient; RMSE: root mean square error; LOD: limit of detection; MAPE: mean absolute prediction error.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Senesi, G.S.; De Pascale, O.; Bove, A.; Marangoni, B.S. Quantitative Analysis of Pig Iron from Steel Industry by Handheld Laser-Induced Breakdown Spectroscopy and Partial Least Square (PLS) Algorithm. Appl. Sci. 2020, 10, 8461. https://doi.org/10.3390/app10238461

AMA Style

Senesi GS, De Pascale O, Bove A, Marangoni BS. Quantitative Analysis of Pig Iron from Steel Industry by Handheld Laser-Induced Breakdown Spectroscopy and Partial Least Square (PLS) Algorithm. Applied Sciences. 2020; 10(23):8461. https://doi.org/10.3390/app10238461

Chicago/Turabian Style

Senesi, Giorgio S., Olga De Pascale, Aldo Bove, and Bruno S. Marangoni. 2020. "Quantitative Analysis of Pig Iron from Steel Industry by Handheld Laser-Induced Breakdown Spectroscopy and Partial Least Square (PLS) Algorithm" Applied Sciences 10, no. 23: 8461. https://doi.org/10.3390/app10238461

APA Style

Senesi, G. S., De Pascale, O., Bove, A., & Marangoni, B. S. (2020). Quantitative Analysis of Pig Iron from Steel Industry by Handheld Laser-Induced Breakdown Spectroscopy and Partial Least Square (PLS) Algorithm. Applied Sciences, 10(23), 8461. https://doi.org/10.3390/app10238461

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop