1. Introduction
Arrays of ionization detectors are an important tool for the quality assurance of radiotherapy treatments and are widely used in radiotherapy quality assurance [
1,
2]. Several arrays have been developed, relying on different detection technologies like air-filled ionization chambers, solid-state detectors, or liquid-filled ionization chambers [
3,
4,
5,
6,
7,
8,
9,
10,
11,
12,
13].
Recombination of charge carriers can affect the response of ionization detectors used for the dosimetry of radiotherapy fields. Recombination in air and liquid ionization chambers is well understood, and there exist validated models and correction methods (like the two-voltage-method in air-filled ionization chambers for moderate dose rates) that can be employed to correct the reading of an ionization detector [
14,
15]. These methods can be applied in static situations when the dose rate in the detector does not change with time.
However, in many clinical situations, the above condition does not hold. When an array of detectors is used to verify a radiation treatment, each detector is irradiated with a varying dose rate depending on the configuration of the fields conforming the treatment. In this situation, recombination cannot be properly corrected. While this may not be an issue when measuring dose distributions in conventional radiotherapy (because of low dose rates and low recombination effects in most detectors), it may have important consequences when measuring treatments delivered by flattening-filter-free (FFF) linacs [
16], or ultra-high dose rate (UHDR) radiotherapy [
17,
18].
In this work, we present a method for correcting online the recombination effects in arrays of ionization chambers irradiated with time-varying dose rates. The method is based on the characterization of the dose rate/recombination response of the detector, and the measurement of the instant ionization current in the detector, rather than the integrated charge. Recombination in ionization chambers irradiated with dynamic fields has been studied by other authors. In particular, Boag theoretically studied recombination in an ionization chamber irradiated by a swept beam [
19]. Palmans et al. studied recombination in ionization chambers irradiated by Tomotherapy dynamic beams and developed a dynamic method of correction based on the calculation (with Monte Carlo) and integration of dynamic dose rates in the detector [
20].
2. Methods and Materials
2.1. Online Correction Method: Pulsed Radiation
2.1.1. Recombination in Pulsed Beams
The collection efficiency of a radiation detector irradiated with a pulsed beam is well described by the equation of Boag in the near-saturation regime (
) [
21]:
The variable
u depends on the polarization voltage,
V, dose per pulse,
, and the geometry and properties of the sensitive material. We can just summarize this dependence as,
where
c is proportional to
.
Equations (
1) and (
2) are behind the development of the two-voltage-method for air-filled detectors irradiated with pulsed beams [
20,
22]. In the case of liquid-filled detectors, the ionized charge per pulse depends on the applied voltage due to initial recombination, and more complex methodologies are necessary to obtain the collection efficiency [
14,
15,
23].
In the case of a significant contribution of free electrons to the collected charge, the following could be used alternatively [
24]
with
, and
p the free-electron fraction in the chamber.
2.1.2. Analytic Online Correction Method—Pulsed Beams
In a pulsed beam with period
T and dose per pulse
, we can define the average dose rate as:
Equally, we can define the average ionization current associated with the full charge collected without recombination effects:
where
e stands for the elementary charge,
m is the mass of the active volume of the chamber and
W is the average energy per electron–ion pair released in the medium.
The chamber actual average electrical current can be obtained through the integration in a time interval of duration
in the form:
where
is the actual instantaneous readout current. This moving average value,
, can suffer fluctuations depending on the size of the time integration interval
, the actual shape of the instantaneous current of the detector after each pulse, and due to variations in time of the irradiation delivery (i.e., dose per pulse and pulse repetition rate). However, if
is sufficiently large compared with pulse period
T, and considering that irradiation conditions are steady during that interval, the relative standard deviation of
could be reasonably small for the application of the method proposed in this work. For example, taking
provides a standard deviation for
below 1%.
Under these conditions, we may assume that for a given average ionization current
there is an associated charge collection efficiency
, and thus we can write that
where
is the collection efficiency associated to the ionization current
, which is proportional to the dose rate in the detector,
.
We can use the two-voltage method, or any other method, to determine the collection efficiency at a reference dose per pulse,
, or, equivalently, at a given ionization reference current
. We will denote the value of
u in this reference condition as
. Then, the variable
u at a different dose per pulse
verifies that:
consequently,
From Equation (
7), we can express
as
and substituting in Equation (
9), we can obtain
We can solve the above equation for
to obtain:
This expression can be used to correct online the readout signal. The correction factor only depends on the own value of
and a single parameter,
, which can be experimentally determined. Thus, we can write that:
2.1.3. Considerations regarding the Application to Liquid-Filled Ionization Chambers
In order to apply Boag’s Equation (
1) (and correction methods based on it), there must be no overlapping between charges ionized by consecutive pulses, i.e., charge collection must take less than the time between consecutive pulses. This constraint is easily satisfied in air-filled chambers, where the mobilities of charge carriers are large. However, this may not be the case in liquid-filled detectors, where charge collection takes ∼10 ms operating under standard conditions, while the time between pulses may be below that value in clinical LINACs.
However, even if Boag theory cannot be formally applied, it was shown in [
25] that the functional dependence of the collection efficiency on the dose per pulse does not change with overlapping. In this situation, in the near-saturation region we can write:
where
.
Because the functional dependence of the collection efficiency on the dose per pulse does not change, correction methods based on this dependence, like the two-dose-rate method [
14], can still be employed. This is not the case with correction methods based on the functional dependence of the collection efficiency on the voltage (like the three-voltage method [
23]). The online correction method developed in the previous section can also be applied, with the particularity that the two-dose-rate method is used to obtain the value of
.
A further consideration needs to be taken into account in LICs, regarding the extended collection time of the charge released by a pulse and the integration time. As mentioned in the above discussion, the charge collection time in LICs is not negligible in comparison with the time between pulses (in fact, it may be larger). In addition, due to the high density of the ionization medium, the readout signal is large, and integration times may be limited to values of the order of ms (depending on the readout electronics). As the integration time, the collection time, and the time between pulses are of the same order of magnitude, fluctuation on the readout signal may appear which are not caused by variations of the dose per pulse (this is easily noticeable in
Section 3.1 for measurements taken with the detector presented in [
13]). Such fluctuations, which are not caused by variations of the dose per pulse, should be minimized to improve the accuracy of the online correction method. As discussed in
Section 2.1.2, one should set
to avoid this issue. If this is not technically possible, replacing the readout current (
) with a moving average,
(emulating a longer integration time), will reduce the undesired fluctuations:
This moving average will be used in the analysis of experimental data of a LIC in
Section 3.1.
2.2. Online Correction Method: Continuous Radiation
2.2.1. Recombination in Continuous Beams
Nowadays, continuous radiation is far less used than pulsed radiation in radiotherapy. Nonetheless, we shall present the method for continuous beams (e.g., cobalt-60 units) for the sake of completeness. In this situation, the method becomes simpler, because there is no need for averaging to counteract the pulsed structure of the beam. The collection efficiency of a radiation detector irradiated with continuous radiation is well described by the Greening formula [
26,
27] when
:
where
V is the polarization voltage, and
a is a parameter that depends on the physicochemical properties of the ionization medium and is proportional to the dose rate,
. The first expression shows the explicit dependence on the polarization voltage, but for the purpose of this work, the second expression, with
, is more adequate.
This equation has been used to develop recombination correction methods in static situations, like the very well known two-voltage method for gas ionization chambers in continuous beam [
20,
28,
29], and the three-voltage and two-dose rate methods in LICs [
23,
30].
2.2.2. Analytic Online Correction Method—Continuous Beams
We define the ionization current
in this case as:
We will assume that the variations in time of the dose rate are sufficiently slow as to keep the equilibrium charge carrier distributions across the detector at each time t.
The current in the detector is given by:
where
is the readout current,
is the ionization current, and
f the collection efficiency. As
, we can write the collection efficiency as a function of
.
We can use the two-voltage-method, or any other method, to determine the collection efficiency at a reference dose rate,
, and the value of
b, which we will refer to as
. The collection efficiency at any other dose rate/ionization current can be written as:
is not known, but
is measured, and by using Equation (
19) we can write:
This equation can be solved to obtain an expression for the collection efficiency as a function of the measured ionization current
:
where
in continuous irradiation. With this expression, we can correct the readout current as:
2.3. Correction Methods beyond Boag and Greening Limits
In the previous two sections, we developed the online correction method from the analytic expressions of Boag and Greening for pulsed and continuous beams, respectively. We obtained simple closed-form expressions for the proposed correction method. However, in many experimental situations, the expressions presented above do not hold. For example, in gas-filled ionization chambers, it is customary to include the effects of initial recombination and recombination by diffusion in the overall collection efficiencies [
20,
29], even though these contributions are smaller than general/volume recombination. Additionally, in gas chambers, charge multiplication may play a role too, and it should be considered when analyzing the response of the chamber [
20]. The addition of these terms will lead to modifications of the methods presented above, but the methodology would remain the same.
More interestingly, in situations where dose rates/doses per pulse are very large, volume recombination can be quite important, and the Boag/Greening analytic expressions will fail to describe the dependence of the collection efficiency on the dose rate/dose per pulse. Those expressions are only valid in the near-saturation region, with . This may be the case in FFF radiotherapy, with doses per pulse reaching ≃ 2 mGy, and especially UHDR radiotherapy, with doses per pulse that can be several orders of magnitude higher. In these situations, the densities of charge carriers in the ionization medium are much larger, field screening effects become important, and the analytic description of the collection efficiency is no longer valid.
However, in this situation, the dependence of the collection efficiency on the dose per pulse (ionization current) can be characterized experimentally,
where
are a set of parameters that describe the dependence (they can arise from an interpolation of experimental data, or from a phenomenological function fitting the dose per pulse dependence, as in [
31]). After the collection efficiency is accurately characterized (i.e., the function
g is known, either analytically or numerically), the collection efficiency can be obtained following the procedure described in the previous section, by writing the equation as
and solving for
.
2.4. Experimental Data
The presented methods were applied to measurements taken with the liquid-filled ionization chamber presented in [
13]. This array has 2041 detectors with 2.5 × 2.5 mm
effective area and 0.5 mm thickness. A total of 43 × 43 detectors cover a central area of 107.5 × 107.5 mm
. The operation voltage is +1500 V, yet, despite this operation voltage, recombination is still important for large dose rates [
13,
25]. The readout is done with VLSI electrometers (Sens-Tech Ltd, Egham, UK), which communicate with a computer through an Ethernet interface. The readout can be performed at rates >1000 Hz, even though in practice we usually employed values of 20–250 Hz (4 to 50 ms integration times). The readout software includes an option (
sequence mode) that allows saving every single acquisition, in addition to the total integrated signal.
Measurements were performed at Hospital General de Valencia-ERESA in a Truebeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA). We analyzed the verification of a complex treatment, a lymph node tumor treatment consisting of 15 fields. The treatment was delivered at 600 MU/min. The acquisition was saved in sequence mode and analyzed offline (the online correction method was not implemented in the readout software) to correct for recombination with the proposed method.
2.5. Numerical Simulations
We further investigated the application of the proposed methods by performing numerical simulations of the response of both air-filled ionization chambers and liquid-filled ionization chambers to a time-varying dose rate irradiation. We have used an in-house developed software that numerically solves the transport equations of charge carriers in an ionization detector. The software has been validated in previous publications [
23,
25]. The main parameters included in the simulation are presented in
Table 1. The detectors that we have simulated correspond to the LIC presented in [
13], and the PTW “729” and “1500” arrays [
9]. We would like to note here that there are different versions of the 729 detector, which operate at different voltages. We have simulated the detector operating at 400 V, as shown in
Table 1.
The dose-rate profile of the irradiation was taken from the clinical lymph node treatment presented in
Section 2.4. This clinical treatment delivers a maximum dose per pulse of ∼0.5 mGy. For the simulation study, the dose per pulse in each segment was scaled up (and the irradiation time scaled down accordingly) to reach maximum values of 1 mGy, 2 mGy and 50 mGy while keeping constant the total dose delivered. This was done to explore dose per pulse regimens with high recombination. The first two values correspond to typical dose per pulse values encountered in FFF radiotherapy, while the latter corresponds to values encountered in UHDR radiotherapy.
4. Discussion
In this work, we have presented a method for real-time recombination correction in the dosimetry of dose rate varying radiotherapy treatments with arrays of ionization chambers. The method is based on the measurement of the instantaneous readout current (rather than only the total integrated charge) and the automatic correction according to such measurements. We have investigated the application of the method both with simulations of the response of ionization detectors (liquid- and air-filled chambers) and with preliminary experimental measurements taken with a liquid-filled ionization chamber, showing that the method seems feasible and accurate for the correction for recombination in dose-rate-varying radiotherapy verifications. Nonetheless, further experimental validation of this method is necessary to properly implement this methodology, in particular in different dose-per-pulse/dose-rate regimes and with different detection technologies.
The method presented in this work can be used to correct the readout current in real time, without the need to save large amounts of data to perform the correction offline. This would require to perform the correction operation at a maximum frequency of ≃kHz. While in this work we did not perform the correction of the experimental data online, as the online recombination method was not implemented in the readout software at the time when the measurements were performed, we verified at a later stage that performing such operation online, at a frequency of 100–250 Hz is feasible with the software running on a standard personal computer.
The analytic method that we developed for pulsed beams, based on Boag formula, seems adequate for correcting in the near saturation region (collection efficiencies ≃0.9 or higher). This range includes most applications in conventional radiotherapy and FFF radiotherapy, both for liquid- and air-filled ionization chambers). For UDHR deliveries, where collection efficiencies can be well below the near-saturation region, Boag theory may not be reliable, and it may be necessary to use different parametrizations, as in Equation (
3) or Equation (
24). In addition, the pulse structure of UHDR deliveries may be quite different to the pulse structure used in conventional/FFF radiotherapy (which we have used in this work). In such situations, it is important to pay special attention to the selection of the integration time of the detector to avoid fluctuations and pulse overlapping. This issue has not been fully investigated in this work due to the lack of realistic delivery sequences for UHDR treatments.
The methods that we propose in this work are similar to those described in [
20], where the authors studied recombination in ionization chambers irradiated by dynamic Tomotherapy beams. They considered not only the temporal variation of the dose rate on the detector, but also the spatial variation of the dose rate (caused by dose gradients especially in the penumbra of the beams). However, in that paper, the authors relied on a Monte Carlo calculation of the dose rate on the detector, rather than on the measurement of the ionization current, which allows us to develop an online correction method like the one that we present in this paper. The effect of dose gradients on the volume of the detectors is ignored in this work. In particular, in the analyzed experimental data, the small pixel size (2.5 × 2.5 mm
) should minimize this effect. Nonetheless, the dose-gradient in each detector of an array can be inferred from the measurement of the detectors in its neighborhood, opening the door for the real-time correction of this effect if necessary. However, this was beyond the scope of this work, where we have focused on the correction for recombination. The latter effect is bound to be more important when performing measurements at large dose rates/doses per pulse.
The method proposed in this work may improve the dosimetry of radiotherapy treatments with ionization chamber arrays. The potential is limited for air ionization chambers in conventional and FFF radiotherapy, because such detectors can be designed to present negligible recombination effects. This is not the case with LICs, where recombination losses in FFF radiotherapy can easily reach ∼3–5%. Therefore, the correction of that effect can allow to obtain better agreements in relative dosimetry measurements (the normalization of the dose distribution to the maximum/reference dose would perturb the low-dose regions and the penumbra of the dose distributions). In addition, this would allow to perform accurate absolute dosimetry comparisons, rather than relying on relative dosimetry.
Where this method may offer the most potential is in the verification of UHDR treatments, as it would facilitate the use of ionization chamber arrays in a situation where extremely high doses per pulse would limit the application on these devices. There is a need to develop novel methodologies for the dosimetry of UHDR beams (for example, the AAPM has recently established a task force to deal with this issue). The application in such situations will still require more experimental and simulation studies, paying special attention to the pulse structure of those beams. Such detailed studies were beyond the scope of this work, which has been focused on presenting the method. Future studies of the application to UHDR deliveries will be carried out with specific experimental data.