Article

Electrochemical Characterization and Detection of Lead in Water Using SPCE Modified with BiONPs/PANI

1
Department of Chemistry, School of Physical and Chemical Sciences, Faculty of Natural and Agricultural Sciences, North-West University (Mafikeng Campus), Private Bag X2046, Mmabatho 2735, South Africa
2
Material Science Innovation and Modelling (MaSIM) Research Focus Area, Faculty of Natural and Agricultural Sciences, North-West University (Mafikeng Campus), Private Bag X2046, Mmabatho 2735, South Africa
3
Department of Geography and Environmental Sciences, North-West University, Mafikeng Campus, Mmabatho 2735, South Africa
4
Institute of Nanotechnology and Water Sustainability, College of Science, Engineering and Technology, University of South Africa, Johannesburg 1710, South Africa
*
Author to whom correspondence should be addressed.
Nanomaterials 2021, 11(5), 1294; https://doi.org/10.3390/nano11051294
Submission received: 9 April 2021 / Revised: 2 May 2021 / Accepted: 4 May 2021 / Published: 14 May 2021

Abstract

:
The need for constant assessment of river water qualities for both aquatic and other biological survival has emerged a top priority, due to increasing exposure to industrial pollutants. A disposable screen print carbon electrode was modified with a conductive polymer (PANI) and Zn and/or Cu oxides NPs, obtained through bioreduction in citrus peel extracts (lemon and orange), for ultra-sensitive detection of PB2+, in the Crocodile River water sample. The synthesized materials were characterized with Fourier-transform infra-red spectroscopy (FTIR), ultra-violet visible spectroscopy (UV-Vis), and scanning electron microscopy (SEM). The SPC-modified electrodes designated as SPCE/LPE/BiONPs/PANI and SPCE/OPE/BiONPs/PANI were characterized using cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) and eventually deployed in the electrochemical detection of PB2+ in water using square wave voltammetry (SWV) technique. The electrochemical responses of the modified electrodes for both CV and EIS in 0.1 M HCl demonstrated enhanced performance relative to the bare SPCE. A detection and quantification limit of 0.494 ppb and 1.647 were obtained at SPCE/LPE/BiONPs/PANI, respectively, while a detection and quantification limit of 2.79 ppb and 8.91 ppb, respectively, were derived from SPCE/OPE/BiONPs/PANI. The relative standard deviations (RSD) for SPC electrode at a 6.04 µM PB2+ analyte concentration was 4.76% and 0.98% at SPCE/LPE/BiONPs/PANI and SPCE/LPE/BiONPs/PANI, respectively. The effect of copper, zinc, iron, cobalt, nickel, and magnesium on the stripping peaks of PB2+ at SPCE/OPE/BiONPs/PANI, showed no significant change except for cobalt, with about 17.67% peak current drop. The sensors were assessed for possible determination of PB2+ in spiked river water samples. The average percentage recovery and RSD calculated were 94.25% and 3.74% (n = 3) at SPCE/LPE/BiONPs/PANI and, 96.70% and 3.71% (n = 3) at SPCE/OPE/BiONPs/PANI, respectively. Therefore, the fabricated sensor material could be used for environmental assessment of this highly toxic heavy metal in the aquatic system

1. Introduction

Water is essential to living and plays vital roles in countless human activities. Globally, there has been an increasing demand for potable water both for industrial and commercial purpose. However, rapid growth in population and attendant industrial growth in developing economies have emerged as key drivers into a global water scarcity. Based on the existing population growth rate, by the year 2050, the population of the world is projected to get close to 9 billion and the magnitude of the water crisis will keep rising [1].
Naturally sourced water is, as a rule, largely unsuitable to be used without any form of treatment, particularly for consumption, a sequel to the exposure to a range of contaminants, which could be organic, inorganic, pathogenic, etc. [2,3,4,5]. It is projected that 3.2% of global deaths could be attributed to unsafe water resulting from poor hygiene and sanitation [6] and four-fifths of the diseases in humans globally are as a result of consumption of contaminated water, acute hygiene deficiency and inadequate sanitary knowledge [7].
Currently, of the water contaminants, heavy metal pollution is fast becoming a serious environmental concern due to their increasing outlets, namely electroplating, mining, chemical plants, metallurgy, household, and agriculture wastewater in the modern society [8,9]. Additionally, heavy metals exposure, even in trace amounts, could pose great risk to humans, exerting harmful effects on the environment and other ecological receptors [10,11,12,13,14]. The environmental monitoring and control of exposure is consequently of great importance and has drawn much attention in many research frontiers. However, the detection of heavy metals is a complicated procedure, often involving widespread processes of preparation of sample leading to the measurement, when analytical techniques are employed [15]. Such standard techniques for the detection of heavy metal ions such as inductively coupled plasma optical emission spectrometry (ICP-OES) [16,17,18], UV–Vis spectrometry, [19,20,21] atomic absorption/emission spectroscopy [22,23,24] and laser-induced breakdown spectroscopy (LIBS) [25,26], would not be suitable for in situ, fast, easy and low-cost operation.
The growing and mandatory need for real-time on-site tracking of quality of water and the environment necessitates an alternative reliable and sensitive technique which is affordable and exerts less pressure on the environment [27]. The electrochemical monitoring technique has emerged that promising platform for handy, affordable substitute with high selectivity and a low limit of detection [28,29]. However, the choice of applicable electrochemical sensors in environmental assessment of heavy metals, using various electrochemical techniques could undermine their applicability [30]. The most effective and reliable electrode materials are reported hybrid materials, incorporating various precursors with diverse complementary and supplementary properties [31,32,33]. Following the development of screen printing technology, mass production of screen-printed electrodes (SPEs) has enjoyed extensive achievement [34]. SPEs share similar electrochemical properties with traditional bulk electrodes and possess the superiority of on-site sensing ability, real-time quantification, and ease of operation over the bulk electrodes.
Generally, electrodes modified with nanoparticles (NPs) have attracted huge interest owing to their exceptional chemical, electronic, and physical properties, which no other materials for sensing can attain or the equivalent counterpart of the bulk materials. Using DPASV, Honeychurch and co-workers employed unmodified SPCEs in detection of lead [35,36] in pond waters and copper [36] in waters and serum. The LOD calculated were 2.5 ppb and 8.2 ppb, respectively. Using the SWASV, Guell et al. [37] detected simultaneously cadmium and lead in seawater with the SPCEs. They realized an LOD of 2.9 ppb and 1.8 ppb for deposition time of 120 s. Aragay et al. [38] examined the temperature dependence of electrochemical stripping activity of a HM sensor using nanoparticles of carbon. They realized a multiple detection of copper, lead, mercury, and cadmium ions with higher efficiency, as a result of the availability of larger surface area for the metal deposition and electrons transfer enhanced as a sequel to the increase in edge-like planes. Sommers et al., previously employed polyaniline (PANI), polyanilinepoly (2,2′–dithiodianiline) (PANI-PDTDA) or PANI-methylene blue [39,40,41] in detecting metals (mercury, nickel, lead, and cadmium), realizing a LOD of for 56 ppb for mercury after 120 s deposition time using DPASV as EC techniques.
In this work, the unique properties of environmentally friendly and low-cost bicomponent mixture of CuO and ZnO, named hereinafter bioxide NPs (BIONPs) and PANI blend, were employed in the modification of SPCE for quantification of lead ion in Crocodile River, South Africa, close to a point surrounded by agricultural activities, using SWV technique.

2. Materials and Methods

2.1. Materials

Citrus limon (Lemon) and Citrus sinensis (sweet orange) peels were precursor plant part used. Potassium hexacyanoferrate (III), (K3[Fe (CN)6]; 99% purity), zinc nitrate (Zn (NO3)2, and zinc acetate [(CH3.COO)2 Zn.2H2O all of which have analytical purity, being acquired from Sigma-Aldrich, (Chemie GmbH, Steinheim, Germany). Dimethyl sulfoxide (DMSO, (CH3)2SO; 99%), N, N-Dimethylformamide (DMF, HCON(CH3)2, Aniline (C6H7N; ≥99% purity), and ortho-phosphoric acid (85%), and sodium hydroxide (NaOH; 99% purity), were bought from Emsure Iso, Merck KGaA, (Darmstadt, Germany). HCl (30%) of analytical purity was gotten from Promark Chemicals (Johannesburg, South Africa). Cupric nitrate tri-hydrate (Cu (N03)2.3H2O and Zinc nitrate (Zn (NO3)2 of analytical purity was purchased from SAARCHEM pty ltd (Gauteng, South Africa) and, Sigma-Aldrich (St. Louis, MO, USA), respectively. Potassium chloride (KCl; 3 mol/L) electrolyte was acquired from Metrohm Ltd., (cH-9100 Herisau, Ionenstrasse, Switzerland) and used in the electrochemical studies. Distilled and deionized water (Table 1) were obtained from the laboratory water from purite and vent filter MPKO1, Millipore S.A.S 67,120 (Molsheim, France), respectively. Nitric acid, (HNO3, 55%) was purchased from SAAR, Friestand Drive, Modderfontein, Gauteng, 1645 South Africa. Cadmium (II) chloride (CdCl2; ≥99%), mercury (II) chloride (HgCl2;), and lead nitrate, (Pb(NO3)2; 99.5%), were gotten from Fluker, Sigma-Aldrich, 3050Spruce Street, St Louis, MO, 63103, USA, edutrade, (Gaunteng, South Africa), and, Glassworld, Robertville, Johannesburg, South Africa, respectively.. Ammonium persulphate (APS; (NH4)2S2O8) was acquired from Sigma-Aldrich (St. Louis, MO, USA). In total, 0.1 M buffer solution of the probe were prepared using K3[Fe(CN)6] and KCl. Then, 0.1 M solution each of HCl and NaOH were employed to adjust the value of the pH of the buffer solution. However, 1 M NaOH was used in the optimization of the pH for the electrochemical detection of the analytes to reduce dilution of the solution.

2.2. Material Characterization

The materials were characterized using, Carry 300, UV-Vis Spectrophotometer, Agilent Technologies, Germany, spectroquant Prove300, Merck KGaA, (Darmstadt, Germany), and UV–Vis Uviline 9400 (Sl Analytics, Hattenbergstr.10, D-55122 Mainz, Germany) in the investigation of the optical properties of the nanomaterial fabricated. FTIR (Opus Alpha-P, Brucker Corporation, Billerica, MA, USA) being employed in the evaluation of the functional groups that intermingled with the metal salts and hence performed the role of a reducing or capping agent or both. Quanta FEG 250 ESEM, (ThermoFisher Scientific, Hillsboro, OR, USA) operating on an acceleration voltage of 15.0 kV was employed to describe the surface structure of the nanomaterials prepared. DropSense (analysed with Metrohm Dropview 200 and 8400 series), incorporated with screen plate electrode (SPE) having a 4 mm carbon working electrode (WE) diameter at its centre, carbon counter electrode (CE) and an Ag pseudo-reference (Ag/AgCl) electrode (RE) was used in electrochemical studies. A potential window of −1000 mV to 1200 mV, was employed in the cyclic voltammetry study, using a common scan step of 10 mV/s. The same range of windows was also used in the electrochemical detection of the target Pb2+ metal.

2.3. Synthesis of Bioxide NPs

The bioxide NPs were prepared by simple bioreduction method using citrus peels extracts in an alkaline medium. The preparation of the orange and lemon peels extract used (OPE and LPE, respectively), have been detailed in our previous works [30,42]. In total, 0.1 M (100 mL) each of the precursors salts, Cu(NO3)2 and Zn(NO3)2 were added at the same time into two different conical flask of appropriate volume and heated above 90 °C under a magnetic stirring. Then, 20 mL peel extracts from lemon and orange were added each into the different conical flasks containing the equal ratio of the precursors. Colour changes were noted for any possible bireduction within the first 10 min and followed up with UV-Vis run. After 10 min, about 5–10 mL 1 M solution of NaOH was introduced into each the mixture under vigorous stirring. Immediately, the colour of the mixtures changed from bright light blue through series of colour changes to whitish yellow, indicating the bioreduction of both Zn and Cu. Another sample was collected for UV-Vis confirmation and subsequently the routine was repeated at hourly interval, while trying to keep the reactions conditions constant. After 3 h, the mixtures were cooled down, with thick pale whitish black precipitate settling immediately at the base of the conical flask. The separate mixtures were later centrifuged each for 15 min at 6000× g rpm, and washed thoroughly at least twice with distilled water for another 5 min at the same speed, leaving the precipitate settling at the base. The samples were collected, and oven dried above 200 °C for more than 5 h in a thoroughly washed crucible. The dry samples were collected and kept in airtight vessels for further use.

2.4. Synthesis of PANI

A total of 0.5 g aniline was added to 15 mL of 1 M HCl solution at room temperature and stirred over a magnetic stirrer, and the mixture was cooled down to <5 °C in an ice bath. Then, 1.28 g of APS in another 15 mL of 1 M HCl was cooled to <5 °C was added stepwise to the aniline solution, then stirred at such temperature range, in an ice bath for 4 h, resulting in the formation of green precipitate. The mixture was allowed to settle overnight and thoroughly washed with deionized water. The precipitate was dried in the oven at 50 °C for 24 h. To get de-doped polymer, the dried polymer powder was introduced into 50 mL of IM aqueous ammonium solution and mixed over a stirrer for 24 h at room temperature. The resultant wet precipitate was filtered, thoroughly washed using distilled water, and dried with the oven at 50 °C for about 48 h to get obtain emeraldine base, denoted as de-doped PANI hereafter.

2.5. Modification of the SPCE

Drop dry approach was used to treat each of the SPCE (disposable) modified. About 2 mg of the as prepared metal oxide NPs (MO NPs), and metal oxide NPs and PANI blend (MO NPs/PANI), were each dissolved in about 3–6 drops of DMSO and sonicated at room temperature. Typically, 20–30 µL of the resultant MO NPs, and MO NPs/PANI mixtures were each cast separately on the carbon working electrode and air-dried overnight to obtain SPCE modified with MO NPs, named hereinafter SPCE/CPE/MO NPs and a second SPCE modified with MO NPs/PANI, named hereinafter SPCE/CPE/MO NPs/PANI. Working electrodes hence characterized were, (i) unmodified SPCE, otherwise called bare SPCE, (ii) LPE/ZnONPs modified SPCE, hence termed SPCE/LPE/ZnONPs, (iii) OPE/ZnO NPs modified SPCE, designated hence as SPCE/OPE/ZnONPs, (iv) LPE/CuO NPs modified SPCE named hereinafter SPCE/LPE/CuONPs, (iv) OPE/CuONPs modified SPCE named hereinafter SPCE/OPE/CuONPs, (v) LPE/BIONPs modified SPCE named hereinafter SPCE/LPE/BIONPs, (vi) OPE/BIONPs modified SPCE named hereinafter SPCE/OPE/BIONPs, (vii) PANI modified SPCE named hereinafter SPCE/PANI, (viii) LPE/BIO NPs/PANI modified SPCE named hereinafter SPCE/LPE/BIONPs/PANI and (ix) OPE/BIONPs/PANI modified SPCE termed henceforth SPCE/OPE/BIONPs-PANI.

2.6. Cyclic Voltammetric and Impedance Measurements (EIS)

The electrochemical responses of the engineered nanomaterials modified SPCE were subsequently studied using CV in 0.01 M KeFe(CN)6 probe in 0.1 M KCl at a pH value adjusted to approximately 7.0. Further, CV and EIS behaviour of fabricated electrodes were investigated in the electrolytes (pH ≈ 1.0), prepared using distilled water. During each study, the solution of either the probe or the electrolyte (40–50 µL) was placed on the surface of the electrodes to cover just the CE, RE and WE and scanned using the Dropview 200 and 8400 model software. All experimentations were done at atmospheric room conditions. The methodical procedures were easy and straightforward.

2.7. Electrochemical Tracing of Heavy Metal Ions

Pre-concentration of the electrode was performed on the SPC electrodes in the electrolyte (0.1 M HCl) containing the target Pb2+ analyte at a defined concentration, optimally selected deposition potential, deposition time, frequency, amplitude and incremental step of 4 mV. In practical terms, the modified electrodes were cycled using CV for about 20 times and conditioned for not less than 600 s in the SWV mode with at least 10 scans, to reduce the background current. The analysis of Pb2+ was performed with square wave voltammetry (SWV), which was dependent on the signal stability. All experimental procedures and measurements were performed in most cases in triplets at room temperature, and no oxygen removal was involved. Each measurement comprised of three distinct steps: electrode conditioning, analytical and background scan. For each addition of the analyte in the electrolyte, the solution was bubbled over time to ensure proper mixing before each measurement. The different steps were automated using the Dropview software. Both peak heights and peak derivatives were used for quantification. The real sample analysis was done by spiking the river water in 0.1 M HCl with the measured solution of the analyte.

3. Results and Discussion

3.1. UV-Vis (Ultraviolent Visible Spectroscopy)

The appearance of two distinguishable peaks at around 360 nm and 470 nm for the colloidal ZnO NPs and CuO NPs for both CPEs, respectively, suggests the formation of a core-shell NPs [43,44,45]. This result agrees with the finding reported by Bayahia [46] and Bae [47]. The computed band gaps using Tauc plots for ZnO were 3.34 eV, and 3.30 eV for the OPE, and LPE BiONPs, respectively. These values are comparable with that of their monoxide counterparts [30,42]. Using the same Tauc plot, the computed band gap for CuO were 2.44 eV, and 2.45 eV and are consistent the band gap values of Cu2O reported in another work [48]. However, while the absorption peak (λmax) remained largely the same for both the monoxides NPs (MoONPs) and the BIONPs, the value of Eg was higher for the BiONPs than that of the MoONPs in each of their equivalent oxides. This suggests that the BIONPs will have reduced particle size in relation to their counterpart MoONPs as confirmed in the XRD analysis the subsequent sessions. This improvement in particle morphology is expected to bring some interesting contribution in the size dependent electrochemical behaviours of the engineered NPs.
The as-prepared PANI exhibited two distinct absorption bands at around 332 nm and 633 nm, corresponding to that of pure PANI (Figure 1). The λmax between 330 nm–360 nm, and 600–650 nm are typical of two chromophores, denoting oxidation states of emeraldine with y = 1, y = 0.5 [49]. The peaks of absorption, near 330 nm–360 nm is ascribed to the p–p* transitions in the aromatic rings. The absorption bands near 600–650 nm resemble that of the intramolecular electronic transitions between quinoid and benzenoid units. When PANI is doped with acids, the quinoid bands demonstrate hypochromism as presented in Figure 1, is suggestive of the existence of configuration of random coil typical of PANI salts. The representative low wavelength polaron bands near 400–440 nm as a result of the conductive nature of PANIs are typical of PANI salts only [50].

3.2. FTIR

The FTIR of the citrus peels mediated nanoparticles (CPE/NPs) was carried out and presented (Figure 2a,b) to explore the phytochemicals responsible for the bioreduction, capping and effective stabilization of the composites of ZnO/CuO NPs. The results of FTIR indicate that the citrus peels contain bioactive compounds with functional groups corresponding to that of flavonoids, protein, phenols, dietary fibre and non-phenolic carotenoïds, monoterpenes, alcohols, esters and carbonyl compounds, limonene. [51]. The FTIR of the NPs reported, presents peaks corresponding to that of CuO and ZnO and the potential phytochemicals that could have been responsible for the effective reduction, capping and stabilization of the prepared CPE/NPs. The IR peaks in the region of 920 cm−1, 1013 cm−1, 1389 cm−1, 1600 cm−1 and 3284 cm−1, correspond to –CH out of plane bending vibration of trans or E-alkene, vibrations of carboxylic acids and bending frequencies of C–O, C–OH bending (in-plane), carbonyl stretches, and –OH stretching vibration on the surface of the bioxide NPs, respectively [52,53,54]. From the IR compared to that of the citrus peel, it is obvious that the functional group responsible for the three desired roles of bioreduction, capping and stabilization by electron donation, corresponds to that of flavonoids, proteins, esters, soluble sugars, carboxylic acid from phenolic acids, alcohol and carbonyl groups. A typical IR absorption bands for the OPE and LPE mediated NPs showed common strong peaks at 493 cm−1 for all CuONPs/CPE, typical of Cu-O stretching vibration [55,56] and diversified band of, 620 cm−1 for ZnO NPs of both BiONPs, 622 and 627 cm−1 for the, ZnO/LPE and ZnO/OPE, respectively, which is typical to that of ZnO band [57,58,59].
The spectral evaluation of the FT-IR for the prepared PANI as seen in Figure 3 has shown that the vibrational peaks in the region between 3400–3200 cm−1 is the stretching vibration of N-H band, 3100–2800 cm−1 region corresponds to the C-H band, and broad band around 2800–2300 cm−1 specifies the existence of iminium planes with dopant ions on PANI matrix [60,61]. The typical vibrations in the regions between 1590–1560 cm−1 (1589 cm−1) and 1500–1490 cm−1 designated the presence of the of backbone of the PANI, springing from the stretching manners of the quinoid and the benzenoid rings [62,63]. The peaks close to 1300 cm−1 are representative of the stretching vibrations of strong C-N in PANI, while the peaks close to 1100–1160 cm−1 correspond to aromatic C-H in-plane deformation and prominent band close to 790–830 cm−1 is commonly attributed to the out-of-plane bending pattern of aromatic CAH groups in 1, 4-disubstituted aniline units [64,65,66,67]. This is indicative of the head-to-tail juxtaposition common with PANI. The peak around 500 cm−1 typifies another bending vibration of the backbone of PANI.

3.3. SEM

The morphology of the synthesized NPs is shown in Figure 4. The LPE/BiONPs showed rough surfaced agglomerated balls of clusters that could result in smaller colloidal dimensioned NPs in appropriate solvents, while that of the OPE/BiONPs showed agglomerated spherical clusters having even surfaces, evidently devoid of cracks.
Figure 4 represents the surface morphology of the chemically prepared PANI, which indicated that the PANI has a coarse surface, which is similar to earlier literature reports [68,69]. The microstructure morphology of the PANI is considered as being of a positive effect on its particle specific surface area and the counter anions diffusion [68]; hence, it is profitable for the amplification of specific capacitance of PANI electrode. On the other hand, the rough surface may influence negatively the PANI’s conductivity [69].

3.4. Cyclic Voltammetry of the NPs Modified SPCE in 10 mM [Fe (CN)6]4− Probe

The electrochemical responses of the SPCE and the nanoparticles modified SPCE, were explored. The individual and comparative voltammograms presented in Figure 5 reveals redox peaks at the bare SPE and all the modified electrodes that are well-defined. The additional redox peaks at 0.66 V, −0.59 V; 0.69 V, −0.53 V; and 0.66 V, −0.54 V for SPCE/LPE/CuO NPs, SPCE/OPE/CuO NPs and SPCE/OPE/BIO NPs could be attributed to the reduction of Cu2+ to Cu0 and oxidation of Cu0 to Cu2+, which underscores the effective amplification of the performance of the unmodified electrodes.
The peak anodic currents for the SPCE/LPE/CuONPs, SPCE/OPE/CuONPs, SPCE/CPE/ZnO NPs, SPCE/PANI, SPCE/LPE/BIONPs and SPCE/OPE/BIONPs are approximately 1.31, 1.0, and 1.16, 1.79, 1.56 and 2.78 times greater than the bare SPCE, respectively.
It is expected that the blend of NPs with high surface area and high sensitivity, and less porous conductive PANI polymer will bring some enhancement in the electro-activity of the bare SPCE. These BIO NPs were mixed with conductive PANI polymer in approximately 2:1 ratio (BIO NPs: PANI) and evaluated for possible superior performance, relative to just BIO NPs as illustrated in Figure 6. In Table 2, the redox peak current was higher in SPCE/LPE/BIONPs than that of the SPCE/LPE/BIONPs/PANI though. This is likely due to the electrostatic repulsion between the negatively charged nitrogen ions in both probe and PANI. IN the absence of such repulsive force in the 0.1 M HCl electrolyte, the SPCE/LPE/BIONPs/PANI, gave a much higher current response with defined redox peaks than both the bare and SPCE/LPE/BIONPs.

3.5. Electrochemical Detection of Pb2+ Using SWV Technique

3.5.1. Optimization

To obtain the best SWV results of the target Pb2+ analyte it was necessary to optimize the following sensing parameters, namely the pH, deposition time and potential, frequency and pulse amplitude.

3.5.1.1. Choice of Supporting Electrolyte

The importance of the best choice of supporting electrolyte has been commonly reported in the literature [70,71,72,73]. Hence, seven potential electrolytes were considered. These include (0.1 M) acetic acid (CH3COOH), nitric acid (HNO3), potassium chloride (KCl), ortho-phosphoric acid (H3PO4), sodium chloride (NaCl), phosphate buffer (PBS), and hydrogen chloride (HCl). H2SO4 was omitted for concerns of reacting with lead nitrate to form insoluble lead salt. The cyclic voltammetry was run using a common potential window of −1000 mV to 1200 mV, potential step fixed at 0.01 V at a scan rate of 0.1 V/s, as shown in Figure 7. Generally, the PANI based electrode exhibits dissimilar cyclic voltammetric response in different acid electrolytes which may be as a result of the diverse size and charge of acid [69].
Table 3 shows that all the electrolytes demonstrated redox ability with defined peaks except in acetic acid. The oxidation peak around 0.12–0.15 V observed in HNO3 and PBS typifies the conversion of leucoemeraldine base to emeraldine salt [40]. HNO3, ortho-phosphoric acid and NaOH demonstrated redox peaks that indicate the oxidation or reduction of either of the PANI peaks, but not both, which may undermine selectivity of the electrode, and more so at a lower peak current. HCl and KCl demonstrated similar current response, but at a higher scan rate, one of the characteristic peaks of PANI disappeared in KCl but remained in HCl. The implication is that, in HCl electrolyte, the electrode may be more selective towards target ions than in KCl electrolyte, even though the later appeared to have higher peak current response. Hence, HCl was chosen as a better electrolyte for this study.

Electrochemical Characterization of the SPCE Modified Electrodes in 0.1 M HCl Electrolyte

Figure 8 shows the cyclic voltammetry of the unmodified and modified BIONPs-PANI composites of the SPCE in 0.1 M HCl electrolyte solution. The electrochemical redox behaviour of the modified electrodes was investigated over the potential window of −1000 mV to 1200 mV. The forward and backward scan on the bare SPCE electrodes in the electrolyte showed only solvent decomposition current without any pronounced redox peaks. The SPCE/LPE/BIONPs, followed a similar trend but with a less pronounced reduction potential peak at around 0.23 V, which could be attributed to reduction of Cu2+ to Cu0. In SPCE/OPE/BIONPs, solvent decomposition currents were observed at more negative potential values than positive potential values.
The current response far outstripped the upper limit of the drop-sense current window resulting in voltammetry that looks similar to solvent decomposition at the central potential range. However, an oxidation peak was noticeable at around −0.2 V, which could be attributed to oxidation of Cu0 to Cu2+ salt [74]. At SPCE/LPE/BIONPs/PANI and SPCE/OPE/BIONPs/PANI also, solvent decomposition was observed at more positive or negative potentials. In the forward scan, two anodic peak currents designated as I and II were seen at about 0.14 V and 0.53 V for SPCE/LPE/BIONPs/PANI and 0.16 V and 0.48 V for SPCE/OPE/BIONPs/PANI, respectively (Figure 8d–e). In the backward scan, two pronounced reduction peaks designated as I’ and II’ were around −0.08 V and 0.24 V for SPCE/LPE/BIONPs/PANI and 0.00 V and 0.33 for SPCE/OPE/BIONPs/PANI, respectively. The redox peaks at I/I’ and II/II’ are similar to that of PANI reported in the previous works for SPCE and other bulk electrodes. The oxidation peaks at I and II in the forward scan are typical of to the conversion of leucoemeraldine base to emeraldine salt and the emeraldine salt to pernigraniline salt forms. While the backward scan peaks represented by I’ and II’’ are characteristic of the transformation of pernigraniline salt to emeraldine salt and emeraldine salt to leucoemeraldine base, respectively [40,75].
Typically, electrodes modified by conducting polymers, follow three processes which are: (1) a heterogeneous transfer of electrons between the electrode surface and the layer of the polymer; (2) electrons diffusions along the chain of the polymer, and (3) the solution species diffusions to the electrode [40,76,77,78,79]. In addition, changes in the kinetic properties for electrodes of various polyanilines (PANIs) is probable and it is basically dependent on synthesis conditions such as the nature of supporting electrolyte employed, electrosynthesis solution pH, the concentration of the monomer, and potential applied [40,77], which were all justified in the optimization.
Electrochemical impedance is a successful technique for the study of the interfacial properties of electrode modified surface. The characteristic impedance of any electrode system is essentially dependent on the accumulative contribution of numerous parameters, namely (1) electrolyte resistance (Rs), (2) charge transfer resistance (Rct) between the solution and the electrode surface, (3) Warburg element (Zw), and (4) double layer capacitance (Cdl) (caused by the interface between the surface of the electrode and the solution). The engagement of a constant phase element in place of the capacitance is necessary in the optimization of the fit in the experiment, which is attributable to the deviation from idealist nature of the electrode [80,81]. The impedance complexity can be explained as the total of the real (Zrel) and the imaginary (Zw) element, originating from the cell resistance and capacitance. In addition, for the interest of providing further comprehensive report on the impedance of the SPCE modified electrode, an adapted Randles equivalent circuit fitting was preferred in fitting the measured results. Hence, the Randle equivalent circuit was employed in fitting the impedance data in Figure 9 with the given circuit parameters: solution resistance (Rs), charge transfer resistance (Rct), the double layer capacitance (Cdl), and Warburg impedance (Zw). The interfacial properties of the electrode were represented by the double layer capacitance and the charge transfer resistance, which are essentially regulated by the modification of the surface of the electrodes.
As noted in the CV in the 0.1 M HCl electrolyte solution, in the previous section, the presence of the nanocomposite on the SPCE surface drastically reduced the Rct values: 175 kΩ (SPCE) to 100 Ω (SPCE/LPE/BiONPs/PANI), equivalent to 99.94% reduction and 65 kΩ (SPCE/OPE/BiONPs/PANI) corresponding to 62.86% reduction (Figure 9). This development springs from the enhanced electroactive surface area and interfacial kinetics of the nanocomposites on the surface of the SPCE. An Rct of 100 Ω and 65 kΩ indicates a synergic blending of the CPE/BiONPs and PANI. The Nyquist plots data demonstrated that the modifying layers were effectively deployed onto the electrode surface.

3.5.1.2. Effect of pH

Figure 10a–c shows the current response in SWV mode of the different concentrations of Pb2+ solution in the 0.1 M HCl analyte at SPCE/OPE/BIONPs/PANI, deposition time of 30 s, and deposition potential of −1100 mV, and prior to using optimized conditions. The electrodes showed two stripping peaks at about −0.568 V and −0.476 V, indicative of two binding sites with different energy levels, as a result of the heterogeneity of SPCEs [82]. The site at −0.568 V, showed higher current response with increasing concentration of Pb2+ throughout the study, suggesting that it is the site with the higher energy site with strongest adsorption [83]. As seen and expected in the current response, the anodic stripping peak current increased with increasing Pb2+ concentration. However, there are other measuring parameters that could affect the current response apart from the concentration of the target analyte and one of them is the pH of the solution. The anodic current is favourably dependent on the target analyte solution. Hence, the effect of the pH of on the stripping current of the Pb 2+ was investigated over the acidic region as most literature reports indicate that the acid medium favours the electrochemical stripping of Pb2+, Cd2+, and Hg2+ [82]. Figure 10d shows the magnitude of the current response on SWV mode at various pH of the analyte solution. At pH of 1, the change in magnitude of the peak current (∆ip) was highest and hence was selected as the optimal pH for the electrochemical quantification of the Pb2+ analyte, which could be attributed to reduction in the H+ ions in solution that grow weaker the complexes established between the Pb2+ ions and amino groups [83]. This also agrees with other literature report [40]. Besides the steadily decreasing ∆ip is also the concern of rise in background current with increasing pH.

3.5.1.3. Effect of Deposition Time

Another important parameter optimized is the deposition time. To obtain the optimum deposition time, the SWVs were taken at different times, 10, 20, 30, 40, 60, 90, 120, 180, and 210 s, as shown in Figure 11. It is apparent that the anodic peak current continued to increase with increasing deposition time somewhat sharply until 180 s. Hence, 180 s was chosen as the optimum deposition time. This however is not to say that the highest peak current was obtained at that time, but indicates a saturating of the electrode surface, due to the electro-deposition of the Pb0.

3.5.1.4. Effect of Deposition Potential

A potential range guided by other literature reports were chosen from −0.8 to −1.2 V and examined (Figure 12). Peak currents were observed at two stripping potentials, higher one at about −0.57 V and a second lower one at −0.49 V, due to the heterogeneity of the SPCE. The peak current increased with increasing negative potential till −1.3 V at both stripping potentials. It was necessary to keep the deposition potential at a moderate potential, to reduce the possible interference of the second lower peak and hence, 1.2 V was selected as the optimal deposition potential.

3.5.1.5. Effect of Frequency and Square Wave Amplitude

Figure 13 represents the dependence of the stripping peaks on the frequency of the voltammetry. At higher frequency, the anodic stripping peak currents increases with a skew to the potential. Unfortunately, the background also increases with this increase in the peak current. The difference between the SWV ipa for the frequency of 8 Hz and 15 Hz is small compared to the background current generated. In actual sense, the ∆ip was higher at the frequency of 8 Hz, hence it was used as the optimum frequency. Figure 14 shows the dependence of the ipa on the SW amplitude. Similar to the case of the frequency, the SWV ipa continued to rise with attendant increase in the background, as the amplitude increased. However, an appropriate SW amplitude of 25 mV was used as the optimal amplitude in the sequence of the experiment.

3.6. Calibration of Sensors and Detection of Limit Determination

Calibration plots were generated from the square wave voltammograms of the SPCEs and examined under the conditions chosen and summarized in Table 4. With the aim of evaluating the ability of the SPCEs to detect the smallest introduction of the metal inorganic pollutant, the curves were generated by spiking a fixed volume (10 mL) of the 0.1 M HCl electrolyte solution with small amount of the standard laboratory prepared concentrations of the target metal ion solution.
Figure 15 and Figure 16 show the SWV of the different Pb2+ concentrations (a) and the resultant calibration curves (b). The ipa increased linearly with increasing Pb2+ concentration on the two SPCEs represented by the following equation:
i pa = 8.16 Conc Pb 2 + 8.61 × 10 6   ( R 2   =   0.979 )   for   SPCE / LPE / BIONPs / PANI
i pa = 8.17 Conc Pb 2 + 1.84 × 10 6   ( R 2   =   0.997 )   for   SPCE / OPE / BIONPs / PANI
The limit of detection and quantification were evaluated based on the relation in Equations (3) and (4)
LOD = 3σ/S
LOQ = 10σ/S
where σ is the standard deviation of 6 consecutive SWV blank and S is the calibration curve slope.
The LOD and LOQ were computed using Equations (1) and (2) for Pb2+ to be 1.49 nM (0.494 ppb) and 4.90 nM at SPCE/LPE/BIONPs/PANI (n = 9), respectively. While the LOD and LOQ computed for Pb2+ at SPCE/OPE/BIONPs/PANI (n = 8) were 8.40 nM (2.79 ppb) and 28.00 nM (8.91 ppb), respectively. Apparently the SPCE/LPE/BIONPs/PANI demonstrated superior electrochemical sensitivity with a broader linear range than the SPCE/OPE/BIONPs/PANI towards Pb2+, however with lower R2 value. The LODs for the SPCEs are well above the limit for Pb2+ concentration in drinking water by WHO (10 ppb) and US-EPA (15 ppb).
Table 5 shows the comparison of the electrochemical performance of carbon-based electrodes in the detection of Pb2+, and demonstrates the applicability of the SPCE/LPE/BIONPs/PANI and SPCE/OPE/BIONPs/PANI in environmental monitoring of Pb2+ in water.

3.7. Evaluation of the Precision of the SPCEs

Precision of the SPCEs were evaluated upon measuring the stripping responses of the samples (n = 8) having Pb2+ concentration as shown in Figure 17. The relative standard deviations (RSD) are presented computed at SPCE/LPE/BIONPs/PANI and SPCE/OPE/BIONPs/PANI are 4.76% and 0.98%, respectively. These very low RSD show that electrodes were highly stable and could be used in environmental monitoring of these HMs. It is reported that the stability is a result of the presence of PANI, which covers the surface of the NPs, thereby blocking surface active compound present in the bulk solution.

3.8. Real Sample Analysis

The electrochemical sensor fabricated was subjected to detection of Pb2+ in real water sample as shown in Figure 18 and Figure 19. The water was collected from a crocodile river in South Africa with the coordinate shown in Table 6. The experiment was done in triplicates (n = 3), with the precision typified in Figure 20 and the average of the peak current responses was used for the calibration plots. The slopes of the calibration plots are represented by the following equation on the two electrodes:
ipa = 5.37 [Pb2+] − 5.03 × 10−6 SPCE/LPE/BIONPs/PANI
ipa = 8.15 [Pb2+] − 4.89 × 10−6 SPCE/OPE/BIONPs/PANI
The results as shown in Table 7 show a good recovery rate of 104.32% and 103.32% for SPCE/LPE/BIONPs/PANI, and SPCE/OPE/BIONPs/PANI, respectively, when the river water sample was spiked with 3.6 µM of Pb2+ concentration. The average recovery rates for the two electrodes are, however, 94.25% and 96.70%, in that same order. The t-test also shows a p-value at 95% confidence level that is less than 0.05 for both calibrations from the two electrodes (Figure 18b and Figure 19b), showing high precision of the calibration coefficients despite the number of different concentrations (spikes) considered. Using SWV, the fabricated sensor detected Pb2+ concentration of 0.727 µM (240.78 ± 0.18 ppb) and 0.683 µM (226.21 ± 0.29 ppb) at SPCE/LPE/BIONPs/PANI, and SPCE/OPE/BIONPs/PANI, respectively. The results are relatively comparable. However, the ICP-MS analysis result showed the concentration of Pb2+ in the river was below the instrument’s limit of detection. The ability of the engineered electrodes to detect Pb2+ in this river water samples, hence underscores the superior sensitivity and selectivity of the electrochemical techniques employed in environmental monitoring strategy. Worthy to mention is that the Pb2+ concentration detected in the river samples is well above the recommended unit for safe drinking water by WHO but well below the standard set by other environmental agencies for Pb2+ in sea water. It could hence be argued that run off waters from the agricultural and mining activities close to the site was responsible for the high concentration of Pb2+ in the fresh river water sampled.

3.9. Interference Studies

The selectivity of the fabricated SPCE/OPE/BIONPs/PANI towards the target analyte was examined in the presence of six other metals commonly associated with, Pb2+, at equal molar concentrations and presented in Table 8 below. The results showed that the electrode was highly selective towards the target analyte in the presence of these metals. However, Co2+ reduced the stripping peak current of Pb2+ by 17.67%. The SPCE/LPE/BIONPs/PANI was so susceptible to interference of these metals that we consider further details into that a future work to be carried out.

4. Conclusions

A common environmental citrus waste peel of lemon and orange were successfully employed in the green synthesis of ZnO and CuO blends to form bioxide of BiONPs in an alkaline medium. The synthesized NPs were characterized using UV-Vis, SEM, and FTIR, which showed basic spectroscopic properties that were consistent with the monoxide components of the BiONPs. This posits that nanotechnology could be used to recycle organic environmental wastes, hitherto poorly managed. Citrus peels extracts mediated BiONPs/PANI polymer were exploited in the modification of the SPCE and an improved electrochemical response relative to the unmodified SPCE and SPCE/PANI electrodes using CV and EIS characterization was realized. The presence of the nanocomposite on the SPCE surface drastically reduced the Rct value of the bare SPCE from 175 kΩ to 100 Ω at SPCE/LPE/BiONPs/PANI, which is equivalent to 99.94% reduction, and 175 kΩ (SPCE) to 65 kΩ at SPCE/OPE/BiONPs/PANI, corresponding to 62.86% reduction. A detection limit of 0.494 ppb and 1.647 were obtained at SPCE/LPE/BiONPs/PANI, respectively, while a detection and quantification limit of 2.79 ppb and 8.91 ppb, respectively, were derived from SPCE/OPE/BiONPs/PANI. The RSD computed for both electrodes, were less than 5%. The SPCE/OPE/BiONPs/PANI showed mostly negligible reduction in peak current in the presence of interfering metals. These further underscore the sensitivity, stability, selectivity and the applicability of the tools and technique used. An optimal potential voltage of −1.2 V for 180 s in 0.1 M HCl supporting electrolyte was employed. The two working electrodes developed showed good applicability in real river water sample analysis. Both electrodes showed limits of detection for Pb2+ that are far below the standard recommended by WHO and US-EPA in drinking water and comparable to other reported works. Therefore, the employed technique and electrodes modified with eco-friendly, simple and low-cost synthesized nanocomposites, could be employed as a good on the site, low cost and easy to operate strategy in the environmental sensing application.

Author Contributions

O.E.F. conceptualized and designed the work and was part of the manuscript write-up. E.C.O. carried out the experiments, interpreted some of the results. E.C.O., S.C.N., O.E.F. and E.E.E. were involved in the manuscript preparation. All authors have read and agreed to the published version of the manuscript.

Funding

NRF-Thutuka grant (UID: 117009).

Data Availability Statement

Data for the study is available upon request from the authors.

Acknowledgments

Authors acknowledged the assistance of MASIIM of North-West University, NRF-Thutuka grant (UID: 117009) and the Higher Degree of North-West University, Mafikeng Campus are also acknowledged.

Conflicts of Interest

Authors confirm that they have no conflict of interest.

References

  1. Boretti, A.; Rosa, L. Reassessing the projections of the world water development report. NPJ Clean Water; Nature Publishing Group: London, UK, 2019; Volume 2, pp. 1–6. [Google Scholar]
  2. Dojlido, J.; Best, G.A. Chemistry of Water and Water Pollution; Ellis Horwood Limited: Hemstead, UK, 1993. [Google Scholar]
  3. Moore, J.W. Inorganic Contaminants of Surface Water: Research and Monitoring Priorities; Springer Science & Business Media: Berlin, Germany, 2012. [Google Scholar]
  4. Bao, L.-J.; Maruya, K.A.; Snyder, S.A.; Zeng, E.Y. China’s water pollution by persistent organic pollutants. Environ. Pollut. 2012, 163, 100–108. [Google Scholar]
  5. Olaolu, D.; Akpor, O.B.; Akor, C.O. Pollution indicators and pathogenic microorganisms in wastewater treatment: Implication on receiving water bodies. Int. J. Environ. Prot. Policy 2014, 2, 205–212. [Google Scholar] [CrossRef]
  6. World Health Organization; UniCeF. Progress on sanitation and drinking-water: 2010 update. In Progress on Sanitation and Drinking-water: 2010 Update; World Health Organization: Geneva, Switzerland, 2010; p. 60. [Google Scholar]
  7. Punia, P.; Bharti, M.K.; Chalia, S.; Dhar, R.; Ravelo, B.; Thakur, P.; Thakur, A. Recent advances in synthesis, characterization, and applications of nanoparticles for contaminated water treatment—A review. Ceram. Int. 2020, 47, 1526–1550. [Google Scholar] [CrossRef]
  8. Baysal, A.; Ozbek, N.; Akman, S. Determination of trace metals in waste water and their removal processes. Waste Water Treat. Technol. Recent Anal. Dev. 2013, 145–171. [Google Scholar] [CrossRef] [Green Version]
  9. Noor, S.M.; Ahmad, N.; Khattak, M.; Mukhtar, A.; Badshah, S.; Khan, R. Removal of Heavy Metal from Wastewater: A Review of Current Treatment Processes. J. Adv. Res. Mater. Sci. 2019, 58, 1–9. [Google Scholar]
  10. Kampa, M.; Castanas, E. Human health effects of air pollution. Environ. Pollut. 2008, 151, 362–367. [Google Scholar] [CrossRef]
  11. Khan, A.; Khan, S.; Khan, M.A.; Qamar, Z.; Waqas, M. The uptake and bioaccumulation of heavy metals by food plants, their effects on plants nutrients, and associated health risk: A review. Environ. Sci. Pollut. Res. 2015, 22, 13772–13799. [Google Scholar]
  12. Liu, J.; Liu, Y.J.; Liu, Y.; Liu, Z.; Zhang, A.N. Quantitative contributions of the major sources of heavy metals in soils to ecosystem and human health risks: A case study of Yulin, China. Ecotoxicol. Environ. Saf. 2018, 164, 261–269. [Google Scholar] [CrossRef] [PubMed]
  13. Rehman, K.; Fatima, F.; Waheed, I.; Akash, M.S.H. Prevalence of exposure of heavy metals and their impact on health consequences. J. Cell. Biochem. 2018, 119, 157–184. [Google Scholar] [CrossRef]
  14. Tchounwou, P.B.; Yedjou, C.G.; Patlolla, A.K.; Sutton, D.J. Heavy metal toxicity and the environment. Mol. Clin. Environ. Toxicol. 2012, 101, 133–164. [Google Scholar]
  15. Matong, J.M.; Nyaba, L.; Nomngongo, P.N. Determination of As, Cr, Mo, Sb, Se and V in agricultural soil samples by inductively coupled plasma optical emission spectrometry after simple and rapid solvent extraction using choline chloride-oxalic acid deep eutectic solvent. Ecotoxicol. Environ. Saf. 2017, 135, 152–157. [Google Scholar] [CrossRef]
  16. Schunk, P.F.T.; Kalil, I.C.; Pimentel-Schmitt, E.F.; Lenz, D.; de Andrade, T.U.; Ribeiro, J.S.; Endringer, D.C. ICP-OES and micronucleus test to evaluate heavy metal contamination in commercially available Brazilian herbal teas. Biol. Trace Elem. Res. 2016, 172, 258–265. [Google Scholar] [CrossRef] [PubMed]
  17. Massadeh, A.M.; Allah, A.; Al-Massaedh, T.; Kharibeh, S. Determination of selected elements in canned food sold in Jordan markets. Environ. Sci. Pollut. Res. 2018, 25, 3501–3509. [Google Scholar] [CrossRef]
  18. Hou, H.; Zeinu, K.M.; Gao, S.; Liu, B.; Yang, J.; Hu, J. Recent advances and perspective on design and synthesis of electrode materials for electrochemical sensing of heavy metals. Energy Environ. Mater. 2018, 1, 113–131. [Google Scholar] [CrossRef] [Green Version]
  19. Mehder, A.; Habibullah, Y.; Gondal, M.; Baig, U. Qualitative and quantitative spectro-chemical analysis of dates using UV-pulsed laser induced breakdown spectroscopy and inductively coupled plasma mass spectrometry. Talanta 2016, 155, 124–132. [Google Scholar] [CrossRef] [PubMed]
  20. Luce, M.S.; Ziadi, N.; Gagnon, B.; Karam, A. Visible near infrared reflectance spectroscopy prediction of soil heavy metal concentrations in paper mill biosolid-and liming by-product-amended agricultural soils. Geoderma 2017, 288, 23–36. [Google Scholar] [CrossRef]
  21. Tang, B.-C.; Fu, H.-Y.; Yin, Q.-B.; Zhou, Z.-Y.; Shi, W.; Xu, L.; She, Y.-B. Combining near-infrared spectroscopy and chemometrics for rapid recognition of an Hg-contaminated plant. J. Spectrosc. 2016, 2016, 1–7. [Google Scholar] [CrossRef] [Green Version]
  22. Yu, J.; Yang, S.; Sun, D.; Lu, Q.; Zheng, J.; Zhang, X.; Wang, X. Simultaneously determination of multi metal elements in water samples by liquid cathode glow discharge-atomic emission spectrometry. Microchem. J. 2016, 128, 325–330. [Google Scholar] [CrossRef]
  23. Thongsaw, A.; Chaiyasith, W.C.; Sananmuang, R.; Ross, G.M.; Ampiah-Bonney, R.J. Determination of cadmium in herbs by SFODME with ETAAS detection. Food Chem. 2017, 219, 453–458. [Google Scholar] [CrossRef]
  24. Gitet, H.; Hilawie, M.; Muuz, M.; Weldegebriel, Y.; Gebremichael, D.; Gebremedhin, D. Bioaccumulation of heavy metals in crop plants grown near Almeda Textile Factory, Adwa, Ethiopia. Environ. Monit. Assess. 2016, 188, 1–8. [Google Scholar] [CrossRef]
  25. dos Santos Augusto, A.; Batista, É.F.; Pereira-Filho, E.R. Direct chemical inspection of eye shadow and lipstick solid samples using laser-induced breakdown spectroscopy (LIBS) and chemometrics: Proposition of classification models. Anal. Methods 2016, 8, 5851–5860. [Google Scholar] [CrossRef]
  26. Gondal, M.A.; Shemis, M.A.; Khalil, A.A.; Nasr, M.M.; Gondal, B. Retracted Article: Laser produced plasma diagnosis of carcinogenic heavy metals in gallstones. J. Anal. At. Spectrom. 2016, 31, 506–514. [Google Scholar] [CrossRef]
  27. Bigham, T.; Dooley, J.S.; Ternan, N.G.; Snelling, W.J.; Castelán, M.H.; Davis, J. Assessing microbial water quality: Electroanalytical approaches to the detection of coliforms. TracTrends Anal. Chem. 2019, 121, 115670. [Google Scholar] [CrossRef]
  28. Xu, K.; Chen, Q.; Zhao, Y.; Ge, C.; Lin, S.; Liao, J. Cost-effective, wireless, and portable smartphone-based electrochemical system for on-site monitoring and spatial mapping of the nitrite contamination in water. Sens. Actuators B Chem. 2020, 319, 128221. [Google Scholar] [CrossRef]
  29. Lambrou, T.P.; Anastasiou, C.C.; Panayiotou, C.G.; Polycarpou, M.M. A low-cost sensor network for real-time monitoring and contamination detection in drinking water distribution systems. IEEE Sens. J. 2014, 14, 2765–2772. [Google Scholar] [CrossRef]
  30. Okpara, E.C.; Fayemi, O.E.; Sherif, E.-S.M.; Junaedi, H.; Ebenso, E.E. Green Wastes Mediated Zinc Oxide Nanoparticles: Synthesis, Characterization and Electrochemical Studies. Materials 2020, 13, 4241. [Google Scholar] [CrossRef] [PubMed]
  31. Abbas, W.; Akhtar, N.; Liu, Q.; Li, T.; Zada, I.; Yao, L.; Naz, R.; Zhang, W.; Mazhar, M.E.; Zhang, D. Facilely green synthesis of 3D nano-pyramids Cu/carbon hybrid sensor electrode materials for simultaneous monitoring of phenolic compounds. Sens. Actuators B Chem. 2019, 282, 617–625. [Google Scholar] [CrossRef]
  32. Go, M.; Hwang, B.; Lim, S. Highly reliable mulberry paper (Hanji)-based electrode with printed silver nanowire/zinc oxide hybrid for soft electronics. Mater. Manuf. Process. 2019, 34, 1605–1611. [Google Scholar] [CrossRef]
  33. Zhou, Q.; Kim, J.-N.; Han, K.-W.; Oh, S.-W.; Umrao, S.; Chae, E.J.; Oh, I.-K. Integrated dielectric-electrode layer for triboelectric nanogenerator based on Cu nanowire-Mesh hybrid electrode. Nano Energy 2019, 59, 120–128. [Google Scholar] [CrossRef]
  34. Waheed, A.; Mansha, M.; Ullah, N. Nanomaterials-based electrochemical detection of heavy metals in water: Current status, challenges and future direction. TrAC Trends Anal. Chem. 2018, 105, 37–51. [Google Scholar] [CrossRef]
  35. Kalpana, V.; Devi Rajeswari, V. A review on green synthesis, biomedical applications, and toxicity studies of ZnO NPs. Bioinorg. Chem. Appl. 2018, 2018, 1–12. [Google Scholar] [CrossRef]
  36. Honeychurch, K.C.; Hawkins, D.M.; Hart, J.P.; Cowell, D.C. Voltammetric behaviour and trace determination of copper at a mercury-free screen-printed carbon electrode. Talanta 2002, 57, 565–574. [Google Scholar] [CrossRef]
  37. Güell, R.; Aragay, G.; Fontàs, C.; Anticó, E.; Merkoçi, A. Sensitive and stable monitoring of lead and cadmium in seawater using screen-printed electrode and electrochemical stripping analysis. Anal. Chim. Acta 2008, 627, 219–224. [Google Scholar] [CrossRef]
  38. Aragay, G.; Pons, J.; Merkoçi, A. Enhanced electrochemical detection of heavy metals at heated graphite nanoparticle-based screen-printed electrodes. J. Mater. Chem. 2011, 21, 4326–4331. [Google Scholar] [CrossRef]
  39. Somerset, V.; Iwuoha, E.; Hernandez, L. Stripping voltammetric measurement of trace metal ions at screen-printed carbon and carbon paste electrodes. Procedia Chem. 2009, 1, 1279–1282. [Google Scholar] [CrossRef] [Green Version]
  40. Somerset, V.; Leaner, J.; Mason, R.; Iwuoha, E.; Morrin, A. Development and application of a poly (2, 2′-dithiodianiline)(PDTDA)-coated screen-printed carbon electrode in inorganic mercury determination. Electrochim. Acta 2010, 55, 4240–4246. [Google Scholar] [CrossRef]
  41. Somerset, V.; Leaner, J.; Mason, R.; Iwuoha, E.; Morrin, A. Determination of inorganic mercury using a polyaniline and polyaniline-methylene blue coated screen-printed carbon electrode. Int. J. Environ. Anal. Chem. 2010, 90, 671–685. [Google Scholar] [CrossRef]
  42. Okpara, E.C.; Fayemi, O.E. Comparative study of spectroscopic and cyclic voltammetry properties of CuONPs from citrus peel extracts. Mater. Res. Express 2019, 6, 105056. [Google Scholar] [CrossRef]
  43. Raveendran, P.; Fu, J.; Wallen, S.L. Completely “green” synthesis and stabilization of metal nanoparticles. JACS 2003, 125, 13940–13941. [Google Scholar] [CrossRef] [PubMed]
  44. Zhang, G.; Du, M.; Li, Q.; Li, X.; Huang, J.; Jiang, X.; Sun, D. Green synthesis of Au–Ag alloy nanoparticles using Cacumen platycladi extract. RSC Adv. 2013, 3, 1878–1884. [Google Scholar] [CrossRef]
  45. Duan, S.; Wang, R. Bimetallic nanostructures with magnetic and noble metals and their physicochemical applications. Prog. Nat. Sci. Mater. Int. 2013, 23, 113–126. [Google Scholar] [CrossRef] [Green Version]
  46. Bayahia, H.; Al-Ghamdi, M.; Hassan, M.; Amna, T. Facile Synthesis of ZnO-Cu2O Composite Nanoparticles and Effect of Cu2O Doping in ZnO on Antimicrobial Activity. Mod. Chem. Appl. 2017, 5, 2. [Google Scholar] [CrossRef] [Green Version]
  47. Bae, K.-L.; Kim, J.; Lim, C.K.; Nam, K.M.; Song, H. Colloidal zinc oxide-copper (I) oxide nanocatalysts for selective aqueous photocatalytic carbon dioxide conversion into methane. Nat. Commun. 2017, 8, 1156. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  48. Wang, Y.; Lany, S.; Ghanbaja, J.; Fagot-Revurat, Y.; Chen, Y.; Soldera, F.; Horwat, D.; Mücklich, F.; Pierson, J. Electronic structures of Cu2O, Cu4O3, and CuO: A joint experimental and theoretical study. Phys. Rev. B 2016, 94, 245418. [Google Scholar] [CrossRef] [Green Version]
  49. Liu, P. Synthesis and characterization of organo-soluble conducting polyaniline doped with oleic acid. Synth. Met. 2009, 159, 148–152. [Google Scholar] [CrossRef]
  50. Dhivya, C.; Vandarkuzhali, S.A.A.; Radha, N. Antimicrobial activities of nanostructured polyanilines doped with aromatic nitro compounds. Arab. J. Chem. 2019, 12, 3785–3798. [Google Scholar] [CrossRef] [Green Version]
  51. M’hiri, N.; Ioannou, I.; Ghoul, M.; Mihoubi Boudhrioua, N. Phytochemical characteristics of citrus peel and effect of conventional and nonconventional processing on phenolic compounds: A review. Food Rev. Int. 2017, 33, 587–619. [Google Scholar] [CrossRef]
  52. Nagaraj, E.; Karuppannan, K.; Shanmugam, P.; Venugopal, S. Exploration of Bio-synthesized Copper Oxide Nanoparticles Using Pterolobium hexapetalum Leaf Extract by Photocatalytic Activity and Biological Evaluations. J. Clust. Sci. 2019, 30, 1157–1168. [Google Scholar] [CrossRef]
  53. Coates, J. Interpretation of infrared spectra, a practical approach. Encycl. Anal. Chem. Appl. Theory Instrum. 2006. [Google Scholar] [CrossRef]
  54. Yang, A.-l.; Li, S.-p.; Wang, Y.-j.; Wang, L.-l.; Bao, X.-c.; Yang, R.-q. Fabrication of Cu2O@ Cu2O core–shell nanoparticles and conversion to Cu2O@ Cu core–shell nanoparticles in solution. Trans. Nonferrous Met. Soc. China 2015, 25, 3643–3650. [Google Scholar] [CrossRef]
  55. Xu, Y.; Chen, D.; Jiao, X.; Xue, K. Nanosized Cu2O/PEG400 composite hollow spheres with mesoporous shells. J. Phys. Chem. C 2007, 111, 16284–16289. [Google Scholar] [CrossRef]
  56. Gliemann, G.K. Nakamoto: Infrared and Raman Spectra of Inorganic and Coordination Compounds; John Wiley and Sons: New York, NY, USA; Chichester, UK; Brisbane, Australia; Toronto, ON, Canada, 1978. [Google Scholar]
  57. Wahab, R.; Mishra, A.; Yun, S.-I.; Kim, Y.-S.; Shin, H.-S. Antibacterial activity of ZnO nanoparticles prepared via non-hydrolytic solution route. Appl. Microbiol. Biotechnol. 2010, 87, 1917–1925. [Google Scholar] [CrossRef] [PubMed]
  58. Muthukumaran, S.; Gopalakrishnan, R. Structural, FTIR and photoluminescence studies of Cu doped ZnO nanopowders by co-precipitation method. Opt. Mater. 2012, 34, 1946–1953. [Google Scholar] [CrossRef]
  59. Akintelu, S.A.; Folorunso, A.S. A review on green synthesis of zinc oxide nanoparticles using plant extracts and its biomedical applications. BioNanoScience 2020, 10, 848–863. [Google Scholar] [CrossRef]
  60. Chan, H.; Ng, S.; Sim, W.; Tan, K.; Tan, B. Preparation and characterization of electrically conducting copolymers of aniline and anthranilic acid: Evidence for self-doping by X-ray photoelectron spectroscopy. Macromolecules 1992, 25, 6029–6034. [Google Scholar] [CrossRef]
  61. Chan, H.; Ng, S.; Sim, W.; Seow, S.; Tan, K.; Tan, B. Synthesis and characterization of conducting poly (o-aminobenzyl alcohol) and its copolymers with aniline. Macromolecules 1993, 26, 144–150. [Google Scholar] [CrossRef]
  62. Guo, Y.; Zhou, Y. Polyaniline nanofibers fabricated by electrochemical polymerization: A mechanistic study. Eur. Polym. J. 2007, 43, 2292–2297. [Google Scholar] [CrossRef]
  63. Kuramoto, N.; Geniès, E.M. Micellar chemical polymerization of aniline. Synth. Met. 1995, 68, 191–194. [Google Scholar] [CrossRef]
  64. Šeděnková, I.; Trchová, M.; Blinova, N.V.; Stejskal, J. In-situ polymerized polyaniline films. Preparation in solutions of hydrochloric, sulfuric, or phosphoric acid. Thin Solid Film. 2006, 515, 1640–1646. [Google Scholar] [CrossRef]
  65. Stejskal, J.; Sapurina, I.; Trchová, M.; Prokeš, J.; Křivka, I.; Tobolková, E. Solid-state protonation and electrical conductivity of polyaniline. Macromolecules 1998, 31, 2218–2222. [Google Scholar] [CrossRef]
  66. Li, G.-R.; Feng, Z.-P.; Zhong, J.-H.; Wang, Z.-L.; Tong, Y.-X. Electrochemical synthesis of polyaniline nanobelts with predominant electrochemical performances. Macromolecules 2010, 43, 2178–2183. [Google Scholar] [CrossRef]
  67. Zhang, L.; Dong, S. The electrocatalytic oxidation of ascorbic acid on polyaniline film synthesized in the presence of camphorsulfonic acid. J. Electroanal. Chem. 2004, 568, 189–194. [Google Scholar] [CrossRef]
  68. Li, H.; Wang, J.; Chu, Q.; Wang, Z.; Zhang, F.; Wang, S. Theoretical and experimental specific capacitance of polyaniline in sulfuric acid. J. Power Sources 2009, 190, 578–586. [Google Scholar] [CrossRef]
  69. Zhu, H.; Peng, S.; Jiang, W. Electrochemical properties of PANI as single electrode of electrochemical capacitors in acid electrolytes. Sci. World J. 2013, 2013, 940153. [Google Scholar] [CrossRef]
  70. Rodriguez-Sanchez, L.; Blanco, M.; Lopez-Quintela, M. Electrochemical synthesis of silver nanoparticles. J. Phys. Chem. B 2000, 104, 9683–9688. [Google Scholar] [CrossRef]
  71. Forzani, E.S.; Li, X.; Zhang, P.; Tao, N.; Zhang, R.; Amlani, I.; Tsui, R.; Nagahara, L.A. Tuning the Chemical Selectivity of SWNT-FETs for Detection of Heavy-Metal Ions. Small 2006, 2, 1283–1291. [Google Scholar] [CrossRef]
  72. Xu, R.-X.; Yu, X.-Y.; Gao, C.; Jiang, Y.-J.; Han, D.-D.; Liu, J.-H.; Huang, X.-J. Non-conductive nanomaterial enhanced electrochemical response in stripping voltammetry: The use of nanostructured magnesium silicate hollow spheres for heavy metal ions detection. Anal. Chim. Acta 2013, 790, 31–38. [Google Scholar] [CrossRef] [PubMed]
  73. Liu, M.; Guan, Q.; Liu, S. Nitrogen-doped hollow carbon spheres for electrochemical detection of heavy metal ions. Ionics 2018, 24, 2783–2793. [Google Scholar] [CrossRef]
  74. Romero-Cano, L.A.; Zárate-Guzmán, A.I.; Carrasco-Marín, F.; González-Gutiérrez, L.V. Electrochemical detection of copper in water using carbon paste electrodes prepared from bio-template (grapefruit peels) functionalized with carboxyl groups. J. Electroanal. Chem. 2019, 837, 22–29. [Google Scholar] [CrossRef]
  75. Song, E.; Choi, J.-W. Conducting polyaniline nanowire and its applications in chemiresistive sensing. Nanomaterials 2013, 3, 498–523. [Google Scholar] [CrossRef]
  76. Ozoemena, K. Recent advances in analytical electrochemistry; Transworld Research Network: Trivandrum, India, 2007. [Google Scholar]
  77. Zanello, P.; Nervi, C.; De Biani, F.F. Inorganic Electrochemistry: Theory, Practice and Application; RCS: London, UK, 2019. [Google Scholar]
  78. Girault, H.H. Analytical and Physical Electrochemistry; EPFL Press: New York, NY, USA, 2004. [Google Scholar]
  79. Bard, A.J.; Faulkner, L.R.; Leddy, J.; Zoski, C.G. Electrochemical Methods: Fundamentals and Applications; John Wiley and Sons Inc.: New York, NY, USA, 2001; Volume 1. [Google Scholar]
  80. Olowu, R.A.; Ndangili, P.M.; Baleg, A.A.; Ikpo, C.O.; Njomo, N.; Baker, P.; Iwuoha, E. Spectroelectrochemical dynamics of dendritic poly (propylene imine)-polythiophene star copolymer aptameric 17 β-estradiol biosensor. Int. J. Electrochem. Sci. 2011, 6, 1686–1708. [Google Scholar]
  81. Honeychurch, K.C.; Hart, J.P.; Cowell, D.C. Voltammetric Behavior and Trace Determination of Lead at a Mercury-Free Screen-Printed Carbon Electrode. Electroanal. Int. J. Devoted Fundam. Pract. Asp. Electroanal. 2000, 12, 171–177. [Google Scholar] [CrossRef]
  82. Byrne, J.T.; Rogers, L. Critical interpretation of electrodeposition studies involving traces of elements. J. Electrochem. Soc. 1951, 98, 457. [Google Scholar] [CrossRef]
  83. Jiokeng, S.L.; Dongmo, L.M.; Ymélé, E.; Ngameni, E.; Tonlé, I.K. Sensitive stripping voltammetry detection of Pb (II) at a glassy carbon electrode modified with an amino-functionalized attapulgite. Sens. Actuators B Chem. 2017, 242, 1027–1034. [Google Scholar] [CrossRef]
  84. Silwana, B.; van der Horst, C.; Iwuoha, E.; Somerset, V. Screen-printed carbon electrodes modified with a bismuth film for stripping voltammetric analysis of platinum group metals in environmental samples. Electrochim. Acta 2014, 128, 119–127. [Google Scholar] [CrossRef]
  85. Honeychurch, K.C.; Hart, J.P.; Cowell, D.C.; Arrigan, D.W. Voltammetric studies of lead at calixarene modified screen-printed carbon electrodes and its trace determination in water by stripping voltammetry. Sens. Actuators B Chem. 2001, 77, 642–652. [Google Scholar] [CrossRef]
  86. Molinero-Abad, B.; Izquierdo, D.; Pérez, L.; Escudero, I.; Arcos-Martínez, M. Comparison of backing materials of screen printed electrochemical sensors for direct determination of the sub-nanomolar concentration of lead in seawater. Talanta 2018, 182, 549–557. [Google Scholar] [CrossRef]
  87. Koudelkova, Z.; Syrovy, T.; Ambrozova, P.; Moravec, Z.; Kubac, L.; Hynek, D.; Richtera, L.; Adam, V. Determination of zinc, cadmium, lead, copper and silver using a carbon paste electrode and a screen printed electrode modified with chromium (III) oxide. Sensors 2017, 17, 1832. [Google Scholar] [CrossRef] [PubMed]
  88. Shah, A.; Zahid, A.; Khan, A.; Iftikhar, F.J.; Nisar, J.; Fernandez, C.; Akhter, M.S.; Almutawah, A.A.; Kraatz, H.-B. Development of a highly sensitive electrochemical sensing platform for the trace level detection of lead ions. J. Electrochem. Soc. 2019, 166, B3136. [Google Scholar] [CrossRef]
  89. Dali, M.; Zinoubi, K.; Chrouda, A.; Abderrahmane, S.; Cherrad, S.; Jaffrezic-Renault, N. A biosensor based on fungal soil biomass for electrochemical detection of lead (II) and cadmium (II) by differential pulse anodic stripping voltammetry. J. Electroanal. Chem. 2018, 813, 9–19. [Google Scholar] [CrossRef]
  90. Tayeb, I.A.; Razak, K.A. Development of gold nanoparticles modified electrodes for the detection of heavy metal ions. J. Phys. Conf. Ser. 2018, 1083, 012044. [Google Scholar] [CrossRef]
  91. Torres-Rivero, K.; Torralba-Cadena, L.; Espriu-Gascon, A.; Casas, I.; Bastos-Arrieta, J.; Florido, A. Strategies for surface modification with Ag-shaped nanoparticles: Electrocatalytic enhancement of screen-printed electrodes for the detection of heavy metals. Sensors 2019, 19, 4249. [Google Scholar] [CrossRef] [Green Version]
  92. Sun, Y.-F.; Sun, J.-H.; Wang, J.; Pi, Z.-X.; Wang, L.-C.; Yang, M.; Huang, X.-J. Sensitive and anti-interference stripping voltammetry analysis of Pb (II) in water using flower-like MoS2/rGO composite with ultra-thin nanosheets. Anal. Chim. Acta 2019, 1063, 64–74. [Google Scholar] [CrossRef] [PubMed]
Figure 1. The UV-Vis of (a) Comparison of the synthesis for the two CPE/NPs (b) PANI.
Figure 1. The UV-Vis of (a) Comparison of the synthesis for the two CPE/NPs (b) PANI.
Nanomaterials 11 01294 g001
Figure 2. FTIR spectral of (a) citrus peels extract (CPE) and (b) CPE/NPs.
Figure 2. FTIR spectral of (a) citrus peels extract (CPE) and (b) CPE/NPs.
Nanomaterials 11 01294 g002
Figure 3. FTIR spectral of PANI.
Figure 3. FTIR spectral of PANI.
Nanomaterials 11 01294 g003
Figure 4. The SEM images of (a) LPE/BiONPs (b) OPE/BiONPs, (c), PANI.
Figure 4. The SEM images of (a) LPE/BiONPs (b) OPE/BiONPs, (c), PANI.
Nanomaterials 11 01294 g004aNanomaterials 11 01294 g004b
Figure 5. Individual and CV comparison of 0.01 M [Fe(CN)6]4− in 0.1 M KCl at pH ≈ 7 on the (a) bare SPE, modified (b) SPCE/LPE/CuONPs (c) SPCE/OPE/CuONPs (d) SPCE/CPE/ZnONPs, (e) SPCE/PANI, (f) SPCE/LPE/BiONPs, (g) SPCE/OPE/BiONPs, and (h) modified SPCE (scan rate = 0.01 V/s, Potentials vs. Ag/AgCl).
Figure 5. Individual and CV comparison of 0.01 M [Fe(CN)6]4− in 0.1 M KCl at pH ≈ 7 on the (a) bare SPE, modified (b) SPCE/LPE/CuONPs (c) SPCE/OPE/CuONPs (d) SPCE/CPE/ZnONPs, (e) SPCE/PANI, (f) SPCE/LPE/BiONPs, (g) SPCE/OPE/BiONPs, and (h) modified SPCE (scan rate = 0.01 V/s, Potentials vs. Ag/AgCl).
Nanomaterials 11 01294 g005aNanomaterials 11 01294 g005b
Figure 6. The comparative voltammogram of the SPCE, SPCE/LPE/BiONPs and SPCE/LPE/BiONPs/PANI in (a) 0.1 M HCl (scan rate = 0.1 V/s) and (b) probe (scan rate = 0.01 V/s). Potentials vs. Ag/AgCl.
Figure 6. The comparative voltammogram of the SPCE, SPCE/LPE/BiONPs and SPCE/LPE/BiONPs/PANI in (a) 0.1 M HCl (scan rate = 0.1 V/s) and (b) probe (scan rate = 0.01 V/s). Potentials vs. Ag/AgCl.
Nanomaterials 11 01294 g006
Figure 7. CV of the 0.1 M supporting electrolyte at the SPCE/OPE/LBIONPs, (scan rate = 0.1 V/s, Potentials vs. Ag/AgCl).
Figure 7. CV of the 0.1 M supporting electrolyte at the SPCE/OPE/LBIONPs, (scan rate = 0.1 V/s, Potentials vs. Ag/AgCl).
Nanomaterials 11 01294 g007
Figure 8. The CV of the (a) SPCE (b) SPCE/LPE/BiONPS (c) SPCE/OPE/BiONPs (d) SPCE/LPE/BiONPs/PANI (e) SPCE/OPE/BiONPs/PANI and (f) Comparative voltamogram of the electrodes in 0.1 M HCl, (scan rate = 0.1 V/s, pH = 1.0, Potentials vs. Ag/AgCl).
Figure 8. The CV of the (a) SPCE (b) SPCE/LPE/BiONPS (c) SPCE/OPE/BiONPs (d) SPCE/LPE/BiONPs/PANI (e) SPCE/OPE/BiONPs/PANI and (f) Comparative voltamogram of the electrodes in 0.1 M HCl, (scan rate = 0.1 V/s, pH = 1.0, Potentials vs. Ag/AgCl).
Nanomaterials 11 01294 g008aNanomaterials 11 01294 g008b
Figure 9. Electrochemical Impedance Spectroscopy of the Bare and the Modified Electrodes.
Figure 9. Electrochemical Impedance Spectroscopy of the Bare and the Modified Electrodes.
Nanomaterials 11 01294 g009
Figure 10. Typical SWV of different Pb2+ concentrations in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, (a) 6.04 µM, (b) 60.4 µM, (c) 6.04 µM and 60.4 µM (d). The influence of pH on the stripping peak current of 6.04 µM Pb2+, 1100 mV deposition potential, 30 s deposition time, (Potential vs. Ag/AgCl).
Figure 10. Typical SWV of different Pb2+ concentrations in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, (a) 6.04 µM, (b) 60.4 µM, (c) 6.04 µM and 60.4 µM (d). The influence of pH on the stripping peak current of 6.04 µM Pb2+, 1100 mV deposition potential, 30 s deposition time, (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g010aNanomaterials 11 01294 g010b
Figure 11. SWVs showing impact of deposition times on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, 1100 mV deposition potential, 30 s deposition time, (Potential vs. Ag/AgCl).
Figure 11. SWVs showing impact of deposition times on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, 1100 mV deposition potential, 30 s deposition time, (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g011
Figure 12. SWVs showing impact of deposition potential on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, −800 mV to −1300 mV deposition potential, 180 s deposition time, (Potential vs. Ag/AgCl).
Figure 12. SWVs showing impact of deposition potential on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, −800 mV to −1300 mV deposition potential, 180 s deposition time, (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g012
Figure 13. SWVs showing influence of frequency on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, 1200 mV deposition potential, 180 s deposition time, (Potential vs. Ag/AgCl).
Figure 13. SWVs showing influence of frequency on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, 1200 mV deposition potential, 180 s deposition time, (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g013
Figure 14. SWVs showing influence of pulse amplitude on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, 1200 mV deposition potential, 180 s deposition time, (Potential vs. Ag/AgCl).
Figure 14. SWVs showing influence of pulse amplitude on the stripping peak current of 6.0 × 10−5 M Pb2+ in 0.1 M HCl electrolyte on SPCE/OPE/BIONPs/PANI, 1200 mV deposition potential, 180 s deposition time, (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g014
Figure 15. (a) SWVs of different concentrations of Pb2+ on SPCE/LPE/BIONPs/PANI. (b) Calibration curves of the peak current vs. different concentrations of Pb2+. (Potential vs. Ag/AgCl).
Figure 15. (a) SWVs of different concentrations of Pb2+ on SPCE/LPE/BIONPs/PANI. (b) Calibration curves of the peak current vs. different concentrations of Pb2+. (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g015
Figure 16. (a) SWVs of different concentrations of Pb2+ on SPCE/OPE/BIONPs/PANI. (b) Calibration curves of the peak current vs. different concentrations of Pb2+. Potential vs. Ag/AgCl).
Figure 16. (a) SWVs of different concentrations of Pb2+ on SPCE/OPE/BIONPs/PANI. (b) Calibration curves of the peak current vs. different concentrations of Pb2+. Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g016
Figure 17. SWV anodic peak current magnitude for eight consecutive runs of 6.04 µM of Pb2+ at (a) SPCE/LPE/BIONPs/PANI. (b) SPCE/OPE/BIONPs/PANI.
Figure 17. SWV anodic peak current magnitude for eight consecutive runs of 6.04 µM of Pb2+ at (a) SPCE/LPE/BIONPs/PANI. (b) SPCE/OPE/BIONPs/PANI.
Nanomaterials 11 01294 g017
Figure 18. Represent (a) the SWV of the different spikes with Pb2+ concentration and (b) the resultant calibration curve at SPCE/LPE/BiONPs/PANI at optimum conditions. (Potential vs. Ag/AgCl).
Figure 18. Represent (a) the SWV of the different spikes with Pb2+ concentration and (b) the resultant calibration curve at SPCE/LPE/BiONPs/PANI at optimum conditions. (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g018
Figure 19. Represent (a) the SWV of the different spikes with Pb2+ concentration and (b) the resultant calibration curve at SPCE/OPE/BiONPs/PANI at optimum conditions. (Potential vs. Ag/AgCl).
Figure 19. Represent (a) the SWV of the different spikes with Pb2+ concentration and (b) the resultant calibration curve at SPCE/OPE/BiONPs/PANI at optimum conditions. (Potential vs. Ag/AgCl).
Nanomaterials 11 01294 g019
Figure 20. The precision of the electrodes towards the Pb2+ analyte concentration at (a) SPCE/LPE/BiONPs/PANI and (b) SPCE/OPE/BiONPs/PANI.
Figure 20. The precision of the electrodes towards the Pb2+ analyte concentration at (a) SPCE/LPE/BiONPs/PANI and (b) SPCE/OPE/BiONPs/PANI.
Nanomaterials 11 01294 g020
Table 1. Ultra-pure water quality (type 1) used.
Table 1. Ultra-pure water quality (type 1) used.
Water Resistivity MΩ cm @ 25 °CIonic Concentration (µg/L)
18.2<1
17.0<2
16.0<3
15.0<5
10.0<10
Table 2. The peak summary for the comparative voltammogram of the SPCE, SPCE/LPE/BiONPs and SPCE/LPE/BiONPs/PANI in the probe.
Table 2. The peak summary for the comparative voltammogram of the SPCE, SPCE/LPE/BiONPs and SPCE/LPE/BiONPs/PANI in the probe.
ELECTRODESEpa (V)Ipa (µA)Epc (V)Ipc (µA)∆Ep (V) i p a i p c
SPCE0.19053.3790.020−66.2290.170.806
SPCE/PANI0.22095.6200.020−113.9100.120.840
SPCE/LPE/BIONPs0.19283.5870.0925−83.3240.101.003
SPCE/LPE/BIONPs/PANI0.18259.4210.0765−60.1810.110.987
Table 3. Peak properties of the investigated electrolyte.
Table 3. Peak properties of the investigated electrolyte.
ElectrolyteIp (µA)Potential WindowCurve Coverage
Area
Acetic acid-−1.0 to 1.2 74.444
Nitric acid80.164−1.0 to 1.2173.414
Potassium chloride134.245−1.0 to 1.2169.226
Hydrogen chloride 110.445−1.0 to 1.2140.949
Phosphate Buffer Solution -−1.0 to 1.2192.919
Ortho-phosphoric acid65.245−1.0 to 1.2111.044
Sodium hydroxide73.164−1.0 to 1.2203.585
Table 4. Optimal conditions for the electrochemical deposition of Pb2+ analyte.
Table 4. Optimal conditions for the electrochemical deposition of Pb2+ analyte.
Optimal Conditions
Deposition time180 s
Electrolyte0.1 M HCL
Deposition potential−1.2 V
pH1.01
Frequency8 Hz
Potential Step4 mV
Pulse Amplitude25 mV
Table 5. The comparison of LODs and LOQs of the Pb2+ analytes at the SPCE/LPE/BiONPs/PANI and SPCE/OPE/BiONPs/PANI, with other reported modified electrodes, given optimum conditions. Potential vs. Ag/AgCl.
Table 5. The comparison of LODs and LOQs of the Pb2+ analytes at the SPCE/LPE/BiONPs/PANI and SPCE/OPE/BiONPs/PANI, with other reported modified electrodes, given optimum conditions. Potential vs. Ag/AgCl.
ElectrodesSensitivity
(µA/µM)
STD. DEV.LOD
(ppb)
LOQ
(ppb)
Linear Range
(µM)
Method
Analysis
Deposition Time (s)Electrolyte/
pH
Reference
BiF-SPCE--0.008 0.0270–0.1 AdDPSV1800.01 M
Ammonium Buffer
pH = 9.2
[84]
Bare-SPCE--2.58.336.3- 24DPASV1500.1 M HCl[81]
CAL-SPCEs--5.016.670–0.01DPASV51.0 M ammonia buffer at pH 11.5[85]
AuNP/ERGO-SPCE0.157-0.110.360.002–0.014DPASV3000.02 M HCl[86]
Cr (III) oxide -SPCE--3100.03–2.42SWASV100 [87]
Mercury film SPE * 0.33-1.86.000.029–0.30SWASV120 0.6 M NaCl
pH 8
[37]
DPTGCE--0.6952.3211–45 SWASV120 0.1 M HCl pH 3[88]
GCE/SWCNTs/BE23.983-33.1110.33-DPASV1200.1 M acetate buffer pH 4.5[89]
AuNPs-SPCE0.376-4.6215.545–25SWASV-0.2 M acetate buffer[90]
SPCNFEs0.1-2.89.332–100DPASV180acetic acid/acetate buffer pH 4.5[91]
MoS2/rGO-GCE50.80-1.595.3016.55–264.8SWASV1200.1 M NH4Cl-HCl
pH 4.0
[92]
Cr-CPE18.75-3100.03–2.42SWASV1002 M
acetate buffer, pH 5
[87]
SPCE/LPE/BIONPs/PANI8.160.0040.491.470.45–5.43SWV1800.1 M HCl
pH = 1
This work
SPCE/OPE/BIONPs/PANI3.460.0292.798.910–4.03SWV180This work
DPASV: differential pulse anodic stripping voltammetry. SWASV: square wave anodic stripping voltammetry; AdDPSV: Adsorptive differential pulse anodic stripping voltammetry, AuNP gold nanoparticles, ERGO electroreduced graphene oxide. CAL: calixarenes BiF: Bismuth Film, MB: methylene blue, PANI-PDTDA: Polyaniline-poly (2,2′-dithiodianiline); *: Simultaneous detection on modified SPE electrodes, DPTGCE: 1-dodecanoyl-3-phenylthiourea glassy carbon electrode; GCE/SWCNTs/BE: Glassy carbon electrode –Single walled carbon nanotubes biomass electrode; SPCNFEs: Screen-printed carbon nanofiber electrodes; MoS2/rGO-GCE: Molybdenum dichalcogenides reduced graphene oxide glassy carbon electrode; Cr-CPE: Chromium (III) oxide modified carbon paste electrode.
Table 6. The coordinates of the Crocodile River Water Sampling Point.
Table 6. The coordinates of the Crocodile River Water Sampling Point.
Sampling PointCoordinatesSite ID
125° 40′ 52.6 S
27° 48′ 12.6 E
Agriculture/Mining
Table 7. The calibration coordinates, working range, LOD, recovery rate and RSD data for the electrodes.
Table 7. The calibration coordinates, working range, LOD, recovery rate and RSD data for the electrodes.
SPCEsR2a ± Sab ± Sb
(10−6)
Working Range (µM)LOD
(µM)
Amount
Added
(µM)
Amount Found
(µM)
RSD
(%)
ta-StatPa-Value
LPE/BIONPs/PANI0.985.37 ± 0.185.03 ± 0.680.9–3.620.733.623.74 (104.32%)3.69.7820.01
LPE/BIONPs/PANI0.998.15 ± 0.29−4.9 ± 0.720.9–3.620.683.623.71 (103.32%)3.114.0510.005
Table 8. The interference of other metals to the simultaneous detection of the three-target analyte at SPCE/OPE/BiONPs/PANI, given optimum conditions. (Potential vs. Ag/AgCl).
Table 8. The interference of other metals to the simultaneous detection of the three-target analyte at SPCE/OPE/BiONPs/PANI, given optimum conditions. (Potential vs. Ag/AgCl).
InterferencePb2+ RECOVERED
Mg2+ 99.9999
Cu2+100
Co2+82.33
Fe2+100
Zn2+98.07
Ni2+ 100
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Okpara, E.C.; Nde, S.C.; Fayemi, O.E.; Ebenso, E.E. Electrochemical Characterization and Detection of Lead in Water Using SPCE Modified with BiONPs/PANI. Nanomaterials 2021, 11, 1294. https://doi.org/10.3390/nano11051294

AMA Style

Okpara EC, Nde SC, Fayemi OE, Ebenso EE. Electrochemical Characterization and Detection of Lead in Water Using SPCE Modified with BiONPs/PANI. Nanomaterials. 2021; 11(5):1294. https://doi.org/10.3390/nano11051294

Chicago/Turabian Style

Okpara, Enyioma C., Samuel Che Nde, Omolola E. Fayemi, and Eno E. Ebenso. 2021. "Electrochemical Characterization and Detection of Lead in Water Using SPCE Modified with BiONPs/PANI" Nanomaterials 11, no. 5: 1294. https://doi.org/10.3390/nano11051294

APA Style

Okpara, E. C., Nde, S. C., Fayemi, O. E., & Ebenso, E. E. (2021). Electrochemical Characterization and Detection of Lead in Water Using SPCE Modified with BiONPs/PANI. Nanomaterials, 11(5), 1294. https://doi.org/10.3390/nano11051294

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop