Review

Recent Developments and Future Perspective on Electrochemical Glucose Sensors Based on 2D Materials

by
Sithara Radhakrishnan
1,
Seetha Lakshmy
2,
Shilpa Santhosh
2,
Nandakumar Kalarikkal
2,3,4,
Brahmananda Chakraborty
5,6,* and
Chandra Sekhar Rout
1,*
1
Centre for Nano and Material Science, Jain University, Jain Global Campus, Jakkasandra, Ramanagara, Bangalore 562 112, Karnataka, India
2
International and Inter University Centre for Nanoscience and Nanotechnology, Mahatma Gandhi University, Kottayam 686 560, Kerala, India
3
School of Pure and Applied Physics, Mahatma Gandhi University, Kottayam 686 560, Kerala, India
4
School of Nanoscience and Nanotechnology, Mahatma Gandhi University, Kottayam 686 560, Kerala, India
5
High Pressure and Synchroton Radiation Physics Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400 085, Maharashtra, India
6
Homi Bhabha National Institute, Mumbai 400 094, Maharashtra, India
*
Authors to whom correspondence should be addressed.
Biosensors 2022, 12(7), 467; https://doi.org/10.3390/bios12070467
Submission received: 12 May 2022 / Revised: 17 June 2022 / Accepted: 21 June 2022 / Published: 28 June 2022

Abstract

:
Diabetes is a health disorder that necessitates constant blood glucose monitoring. The industry is always interested in creating novel glucose sensor devices because of the great demand for low-cost, quick, and precise means of monitoring blood glucose levels. Electrochemical glucose sensors, among others, have been developed and are now frequently used in clinical research. Nonetheless, despite the substantial obstacles, these electrochemical glucose sensors face numerous challenges. Because of their excellent stability, vast surface area, and low cost, various types of 2D materials have been employed to produce enzymatic and nonenzymatic glucose sensing applications. This review article looks at both enzymatic and nonenzymatic glucose sensors made from 2D materials. On the other hand, we concentrated on discussing the complexities of many significant papers addressing the construction of sensors and the usage of prepared sensors so that readers might grasp the concepts underlying such devices and related detection strategies. We also discuss several tuning approaches for improving electrochemical glucose sensor performance, as well as current breakthroughs and future plans in wearable and flexible electrochemical glucose sensors based on 2D materials as well as photoelectrochemical sensors.

1. Introduction

Diabetes was the seventh leading cause of mortality worldwide in 2016, according to the World Health Organization (WHO). Diabetes-related early death has been continuously rising. There were around 1.6 million deaths globally in 2019. According to the WHO, diabetes is a major cause of various health effects, including blindness, lethargy, kidney failure, strokes, numbness, and heart attacks [1]. Our mental pictures of diabetes are of overweight people and the elderly. According to current ICMR studies, over 96,000 children suffer from diabetes, with nearly 16,000 new cases diagnosed each year in children under the age of 14. As a result, there is a risk that diagnoses may be overlooked and treatment will be delayed, resulting in a potentially fatal illness [2]. Thus, glucose monitoring is an important tool for diabetes diagnosis [3].
Several types of biosensors have been developed for this purpose. They include invasive, non-invasive glucose monitoring systems, and continuous glucose monitoring systems (CGM). In this sector, reliable and accurate glucose monitoring is always a major difficulty. Recently, a report by Nava et al. provided a systematic review of the comparative diagnostic accuracy of CGM devices in preterm infants. They used a bivariate model to summarize the receiver operating characteristic curve (ROC) curve and extract the area under the curve (AUC). According to their findings, CGM systems have low sensitivity for detecting hypoglycemia in preterm newborns but good accuracy for identifying hyperglycemia [4]. As a result, new approaches are continually being developed, and minimally invasive glucose monitoring technologies have sparked a lot of interest in this field. A biosensor’s main components include:
(1)
A biological element that distinguish the analyte from other substances;
(2)
A transducer, which converts the biorecognition event into a quantifiable signal;
(3)
A processing system that converts a quantifiable signal into a readable signal [5].
Various types of transducers employed for the detection are calorimetric, thermometric, magnetic, optical, piezoelectric, electrochemical transducers, etc. Calorimetric transducers rely on changes in heat energy caused by chemical reactions between the bio-analyte and enzymes [6]. Thermistors are commonly used to detect temperature changes. Magnetic transducers detect changes in the magnetic properties of magnetic fields such as direction, intensity, and flux [7]. Optical transducers make use of the interaction of the optical field with bio components [8]. Piezoelectric transducers operate on the basis of a change in the frequency of oscillations that is proportional to the mass bound to the surface of the electrode [9]. Finally, electrochemical transducers rely on changes in electrical signals caused by electrochemical processes at the electrode surface, which are directly proportional to analyte concentration [5]. Among all these, electrochemical glucose sensors have a high potential due to several appealing features such as low cost, fast response time, high sensitivity, ease of operation, and excellent miniaturization and construction potential for portable glucose sensing devices for point-of-care (POC) applications [10]. The advantage and disadvantages are given below in Table 1.
The transduction techniques used in electrochemical glucose sensors are amperometry, conductometry; capacitive, impedance spectroscopy; voltammetry and potentiometry (ion-selective) field, and effect. Its features are discussed below in Table 2.
With the great advancements in nanotechnology over the last decade, there has been a surge in research interest in nanomaterial-based electrochemical glucose sensors. Combining glucose biosensors with the distinct properties of developing nanostructures based on two-dimensional (2D) advanced materials has resulted in significantly better sensor performance in terms of sensitivity and selectivity [15]. These recent advances have also improved the key sensor properties related to biofouling, long-term use, limit of detection, and wearability.
Reviews on electrochemical glucose sensors have already been published, encompassing their operational principles [16], history [17], important breakthroughs, and the issues these electrochemical glucose sensors confront. There are also papers discussing the utilization of nanostructures for electrochemical glucose sensors [10]. For example, graphene-based electrochemical sensors for the detection of glucose were well-reviewed by Zhang et al. [18]. After graphene, a large number of 2D materials were discovered and are also widely used in electrochemical glucose sensing. However, comprehensive reviews discussing the 2D materials for enzymatic and non-enzymatic electrochemical sensors are rare.
The first half of this review presents the mechanisms of electrochemical glucose sensors, followed by a discussion of recent advancements based on the use of 2D materials to improve glucose sensor performance. The second section focuses on wearable glucose sensors and provides a comprehensive analysis of different innovative non-invasive detection concepts, as well as associated limitations, challenges, and prospects. The fast advancement of wearable and mobile technologies has sparked considerable interest among researchers in non-invasive glucose biosensing. The last decade has also witnessed the development of photoelectrochemical glucose sensors. Despite its late beginnings, PEC glucose sensing has grown rapidly as a significant branch of electrochemical glucose detection. There are some noteworthy reviews available that state advancements in PEC sensing and performance. However, no review concentrating on PEC glucose sensing based on 2D materials has been described thus far.
To further understand the achievements of the last decade in depth, the readers must go through recent topics such as nanomaterials-based electrochemical glucose sensors, graphene-based electrochemical glucose sensors, PEC sensors, advances in wearable sensors, and so on [1,11,19,20].
This review covers the state of the art of electrochemical glucose sensors based on 2D materials developed primarily from 2010 onwards. Figure 1 shows the summary of the review.

Commercialy Available Electrochemical Glucose Sensors

The Graphene Market and 2D Materials Assessment Report (2021–2031) predicted 18 primary application areas and the transition of graphene biosensors from the laboratory to the commercial market in the coming decades. According to this assessment analysis, the graphene market is expected to expand from $100 million in 2020 to $700 million by 2031 [21]. This success will take time and will be the result of ongoing global research and commercialization initiatives. There is a race to commercialize graphene-based glucose sensors, and their introduction will transform the way diseases are diagnosed and treated. Patients will obtain test results quickly, diagnoses will arrive on-site in minutes, and treatment can begin immediately, saving important and precious time. For example, in partnership with integrated graphene, researchers from the University of Bath recently developed an electrochemical sensor whose function is unaffected by changes in pH or temperature. Boronic acid is used in the new sensor, which is connected to a graphene foam surface. On top, an electroactive polymer layer bind to the boronic acid. When glucose is present, it binds to the boronic acid competitively, displacing the polymer. The sensor generates an electric current proportional to the amount of polymer displaced, allowing the concentration of glucose in the sample to be precisely detected [22]. Even though this 2D-material-based glucose sensor shows quite satisfactory properties for glucose sensing, it should be noted that the electrochemical glucose sensors in the current market are still based on noble metal catalysts. However, there was a trend for graphene-based electrochemical glucose sensors. The global market trends and opportunities show that glucose sensors based on 2D materials are on an upward trend.

2. Electrochemical Glucose Sensors

2.1. Glucose Sensing Mechanism

Electrochemical biosensors are the most often used sensors today due to their improved sensitivity, repeatability, ease of maintenance, and low cost. Electrochemical glucose sensors are further classified as enzyme-based or non-enzyme-based. The non-enzymatic/enzymatic techniques are distinguished by the direct/indirect electrooxidation reaction of glucose molecules in the whole blood [23].
Glucose readings are frequently based on interactions with one of three enzymes: hexokinase, glucose oxidase (GOx), or glucose-1-dehydrogenase (GDH), with GOx and GDH interactions assisting in glucose self-monitoring biosensing.
For the electrochemical sensing of glucose, three detection approaches are available: detecting the amount of consumed oxygen, the amount of hydrogen peroxide produced by enzyme activity, or using a diffusible or immobilized mediator to transport electrons from the GOx to the electrode. GDH-pyrroloquinolinequinone (PQQ) and GDH-nicotinamide-adenine dinucleotide (NAD) are two frequently utilised GDH-based amperometric biosensor families [24]. The following sections will go over the various mechanisms involved in enzymatic and non-enzymatic glucose sensing.

2.1.1. The Sensing Mechanism of the Enzymatic Glucose Sensor

The history of the original glucose-sensing technology begins in 1962, with Clarks and Lyons’ description of the first enzyme-based electrode [25]. The first electrode’s basic premise was the oxidation of D-glucose by molecular oxygen by glucose oxidase entrapped on a semipermeable dialysis membrane supported by an oxygen electrode. Immobilized GOx promotes the oxidation of glucose molecules, forming gluconic acid and hydrogen peroxide. The redox cofactor flavin adenine dinucleotide (FAD), which reduces to FADH2 upon electron acceptance, contributes to GOX’s catalytic activity. H2O2 is formed as a result of cofactor regeneration. Hydrogen peroxide is oxidized at a catalytic, traditionally platinum (Pt) anode that counts the number of electron transfers as a function of the number of glucose molecules in the blood.
Clark patented his subsequent investigation with two electrode systems to monitor glucose sensing in 1970. Electro-inactive materials were transformed into electroactive species in the proposed system using enzyme-based sensing. The enzymes between the membrane and the first electrode transformed the substrate into an electroactive state despite the presence of interfering species in the sample. Still, the second electrode was extremely sensitive to these interfering species in the sample—the difference in currents allowed for successful glucose detection. The initial glucose sensor, however, was not a biosensor. The reflectometer served as the foundation for the early generations of all glucose-sensing devices. The Ames Reflectance Meter (ARM), launched in 1961, was never user-friendly or cost-effective. Clark’s work was moved to Yellow Springs to develop the first-generation commercial glucose sensor. The first commercial glucose sensor, introduced in 1975, was enzyme-based with an amperometric measurement of H2O2 from a human blood sample by the Yellow Springs Instrument Company (as Model 23A YSI analyzer) (Figure 2) [26,27].
With advancements in surface chemistry and device fabrication, the field of commercial electrochemical glucose biosensors has been advanced [28]. Many inorganic redox couples, mediators such as ferricyanide, ferrocene, tetrathiafulvalene (TTF), quinines, thionine tetracyanoquinodimethane (TCNQ), methyl viologen, and methylene blue, and organic colors, cater electrons from the enzyme layer to the working electrode surface [24]. Because the reaction with the enzyme is fast, ferrocene is an often-employed mediator. The mediator can function in a wide range of pH settings and has reversible electron kinetics. Second-generation glucose sensors outperformed first-generation glucose sensors from 1984 because the redox couple or mediator-based approach was generally quite sensitive to dissolved oxygen even at lower potentials and extreme interference ambience. ExacTech by Medisense Inc. was the first home-based self-monitoring glucose sensor to hit the market, following the discovery of a screen-printing process that enabled the creation of wearable, disposable, compact, and pocket-friendly electrodes. Each strip consisted of miniaturized screen-printed reference and working electrodes, with the working electrode coated with the important sensing components, including an electron-shuttle redox mediator, glucose oxidase linking agent and stabilizer [29]. The relevance of second-generation glucose detectors is still being debated; however, with the introduction of finger pricking home-based second-generation glucose sensors, the development of glucose meters accelerated tremendously.
Even the revolution of home-based diabetic-patient-friendly glucose metres using mediated systems did not contribute significantly to the nature of the biochemical structure of the GOx enzyme, the relative toxicity and solubility of the intermediates, and the general poor stability of these under varying pH, temperature, and humidity conditions. As a result, scientists all over the world have been working hard to improve performance by employing techniques that facilitate electron transfer between the GOx redox center and the electrode surface, such as the chemical modification of GOx with electron-relay groups, enzyme wiring of GOx by electron-conducting redox hydrogels, and the use of nanomaterials as electrical connectors, among others. Another dependable option, however, was the direct transmission of electrons from the enzyme sites to the electrode without the presence of intermediaries. Such needle-type implantable devices can be used for continuous glucose monitoring at low potentials near the enzyme’s redox potential. The absence of mediators allows the biosensors to pick glucose molecules better [24].
On the other hand, third-generation sensors are still in the early stages of development. However, those built on nano-mesoporous electrode surfaces hold promise but have substantial limitations. Only a few enzymes have been shown to transport electrons directly from the enzyme to the working electrode. However, these published papers on glucose detection do not provide sufficient evidence to explicate mediator-free glucose detection. Electrode-enzyme detection should be tuned for proper electron transfer. Carbon materials, including single-walled and multi-walled carbon tubes incorporated with nitrogen and phosphorous species, are better platforms for achieving high-performance biosensors for detecting glucose. The nature and amount of the nitrogen and phosphorus species combined in the carbon material surface can be selectively controlled. These species act as anchoring groups for the immobilization of the PQQ-GDH [30,31,32].

2.1.2. The Sensing Mechanism of Non-Enzymatic Glucose Sensors

As previously stated, enzymatic glucose sensors have certain inherent limitations, such as they require appropriate system storage, enzyme denaturation resulting from the change in environment, expensive preparation, protease digestion, time-consuming purification process, thermo-chemical deformation, high cost, poor reproducibility caused by extraneous oxygen sources, lack of stability in harsh conditions, which makes system sterilization difficult, and monotonous enzyme immobilization. The decreased performance produced by the immobilization process remains a barrier that must be overcome [33,34]. Nanomaterial-assisted electrochemical processes via non-enzymatic glucose sensing can fully describe the shortcomings of enzyme-based glucose sensing. As a result, non-enzymatic glucose sensors can be thought of as next-generation glucose detection meters. Most enzyme-free sensors are more sensitive than enzyme-based glucose detectors [35].
Non-enzymatic glucose sensors have a wide range of applications, including catalysis, biomedical devices, and environmental monitoring [36,37,38,39]. Since noble metals and related metal alloys have strong catalytic activity for non-enzymatic glucose electrooxidation, they have focused primarily on their applications [40]. There are currently two major approaches to improving the performance of enzyme-free glucose sensors: one is to change the topological structure of the material to increase the specific surface and electrochemically active sites, and the other is to form hybrids by combining or doping with other materials. Metals and their compounds, metal-organic frameworks (MOFs), conductive polymers, inorganic carbon materials and other materials used to construct glucose sensors form the sensor’s working electrode. This composite of two or more materials fully exploits their respective advantages in building glucose non-enzyme sensors, thereby boosting non-enzyme sensor performance, such as higher sensitivity, wider linear range, and lower detection limit.
Even though platinum nanoparticles are widely used in a non-enzymatic glucose sensor, there are still significant obstacles to overcome, such as poor selectivity, slow kinetics, and limited sensitivity. When platinum is changed with Co, Au, Ni, and Fe alloys and adorned with MWCNTs/carbon, the surface catalytic activity changes to improve stability and sensitivity. A specific synthesis method has been used to address the limitations of Pd BNP as well as for sensing glucose by considering effective combinations such as MOS2, CNTs, and so on [35]. However, noble metals are generally expensive, and the use of non-noble metals is always appreciated. As a result, non-enzymatic electro-oxidation of glucose has been widely investigated utilizing a variety of nanomaterials ranging from noble metals to non-noble metals such as Cu, Zn, Co, Mn, Fe, and Ni, as well as metal sulfides and oxides.
Increasing the exposed active surface of a non-enzymatic glucose sensor is an effective routine for improving analytical performance. As a result, many materials with favorable porosity (microporous, mesoporous, and even macroporous) are expected to find widespread use in the fabrication of satisfactory glucose sensing devices. Because of their better conductivity, high surface area, and improved electron transfer, nanomaterials supported on various substrates, including graphene and its derivatives-based substrates, can improve glucose electrooxidation and detection. A review published in 2020 summarizes the performance of 2D noble and non-noble materials pretty well [40].
Two-dimensional (2D) materials’ promising physicochemical features show them to be an emerging class of possibilities for application in sensing devices. They can also be applied to the IoT platform due to their high surface-area-to-volume ratio, leading to great sensitivity to target glucose molecules (Figure 3) [41]. Furthermore, due to their high mechanical strength, flexibility, and optical transparency, sensors based on 2D materials can also be miniaturized.
Physisorption and chemisorption are the two unique methods by which 2D materials interact with target molecules/ions. Physisorption is caused by molecules/ions interaction with 2D materials’ surface in the absence of covalent bonding, whereas chemisorption is caused by covalent bonding between molecules/ions and the surface of 2D materials. Non-covalent bonding, on the other hand, is favored for quick response. Because of its honeycomb structure, graphene material prefers to establish pi-bonds with molecules [42] rather than the van Der Waals force between MoS2 and molecules, which is indicative of TMDs [43]. Graphene’s thin single layer promotes high carrier layer mobility, making it an excellent 2D material for glucose sensing. The zero-band gap of the graphene substrate is responsible for the hazard-free charge flow [44]. In the case of glucose detection, however, decorating graphene, building a composite with other materials, or utilizing functionalized GO or rGO improve the sensitivity of graphene. N-doped graphene was produced by treating graphene with nitrogen plasma, while Cu-doped graphene was produced via electrodeposition that was adjusted by manipulating the deposition duration and voltage [45,46]. Both procedures produced repeated stable, sensitive, and selective results. Glucose sensing is catalyzed by GOx. However, GO is incompatible with organic polymers. The polymer N-succinimidyl acrylate is used to chemically modify GO with GOx. Figure 4 depicts the effect of several graphene electrodes employed under varying pH settings in an interfering environment [47].
Electrochemical etching, nano-band electrode fabrication, and the electrodeposition of metal nanoparticles are the three methods to decorate nanoparticles on surfaces. The sensing mechanism of Au-doped MoS2 is elucidated by Parlak et al. [48]. Au nanoparticle-doped materials proved to have better sensitivity to glucose. Au nanoparticles for glucose sensing enhancement were also studied by Su et al. [49]. The difference between Parlak et al. and Su et al. is that Parlak et al. dispersed Au nanoparticles with ultrasonication while Su et al. synthesized Au nanoparticles directly on MoS2 by the hydrothermal method. MoS2 nanosheets were dispersed in a 10 mM phosphate buffer solution and were treated with ultrasonication. Both samples were exclusively sensitive to glucose molecules. A vast study is carried out on MoS2 substrates by decorating them with different metals, including Ni, Cu, Au, Ag and Pt etc., adopting different synthesis techniques, including liquid exfoliation, solution method, hydrothermal method, and the electrodeposition method has been studied [50,51,52,53]. Bimetallic nanoparticles are difficult to synthesize, and the field is yet to be explored. Similarly, SnS2 is an n-type semiconductor having a bandgap of 2.18–2.44 eV and can also be employed for glucose and H2O2 detection [54]. They are resistant to acids and have good stability in the air, which has the capability of photocatalysts. After fabrication, the reproducibility and stability of the MWCNT–SnS2 sensor are demonstrated in Figure 5. Amperometric responses were shown and responded only to glucose while having no response to ascorbic acid and uric acid. All solutions were 0.1 mM in pH 6.0 phosphate buffer solution applying 0.43 V.

2.2. Factors Influencing the Electrochemical Glucose Sensing Performance and Monitoring

The electrical conductivity and geometry of the electrode material are the most important factors influencing an enzyme-free electrode’s electrochemical sensing potential. The higher the conductivity, the more electronic transfer and molecular adsorption there will be. The active sites accessible for glucose oxidation are determined by the shape and size of the electrode. Other factors that influence non-enzymatic glucose sensing potential include environmental conditions such as temperature, pressure, and humidity, physiological states that result in pH shifts and poisoning from species other than glucose molecules (interference effect), the electrode’s stability, reversibility, and biocompatibility, the cost of the material, which should be low in order to reduce the overall fabrication cost of the device, and the accuracy of the fabricated device. To evaluate the accuracy of glucose sensor performance, a set of parameters, including sensitivity, the response time, linear range limit of detection (LOD), operation potential, medium pH, selectivity, and so on, is utilized. Another factor influencing the electrocatalytic activity of enzyme-free glucose sensors is their nanomorphologies, which give a huge surface area for electrocatalytic activities.
When considering electrochemical sensing, the electrochemically active surface area is an important parameter to consider. It is also suggested that the increased surface area of two-dimensional (2D) or three-dimensional (3D) nanostructured materials allows for simple surface functionalization, allowing for an effective charge-transfer process. The main limitation of the non-enzymatic glucose sensing activity is its instability with the prolonged use of under-applied working potential. The 2D working electrode can prevent the sensing substrate from agglomerating, deforming, and collapsing [55,56]. Similarly, electrodes should be used with care to achieve precise glucose sensing. Because the pH of human blood ranges between 7.3 and 7.5, glucose-sensing operations should be carried out in physiological settings to optimize various electrochemical characteristics. Glucose levels in diabetic people vary. Even the same patient can experience significant glucose level fluctuations. They can range between 4 and 6 mM. As a result, a reliable glucose sensor should properly detect glucose concentrations across a wide range.
Ginsberg describes four types of errors that occur during blood glucose detection that can significantly impact detector effectiveness. Variations in strip production, strip storage, and ageing can all impair the accuracy of glucose detectors [57]. The efficiency of glucose detection varies from strip to strip when the strip is used as a source of error. Insufficient enzyme coverage can impair glucose meter performance. Excess enzyme on a thick level is utilized in strips to avoid this possibility of inaccuracy, and so slight decreases in the enzyme level or amount do not affect glucose measurements. Sometimes we receive strips with very small individual reaction wells (the sample chambers) (2–3 mm), and a well-size difference of 50 µm can lead to a 3% error. Similarly, the amount of mediator might have a negative impact on the performance of electrochemical glucose sensors. The electrode oxidizes the mediator, which results in the glucose signal. As a result, mediator decrease due to prolonged usage or inactivity can result in incorrect glucose detection. Glucose strips are constantly subjected to biological reactions. As a result, glucose testing strips always have a shelf life. Aside from these conditions, various additional physical factors, such as altitude and temperature (which are less predictable), have an impact on glucose-sensing capabilities. The only remedy for temperature factors is to provide an error warning under extreme conditions. It was discovered that altitude impaired GOx-based meters by 6–15%, whereas non-oxygen-dependent GDH devices had accuracy within 5%. The accuracy of the five strips tested at 8°C varied significantly, regardless of the enzyme base, with 5–7% errors either positively or negatively. Other sources of errors that cannot be adequately addressed are those caused by the patient while erroneously employing glucose detection strips and variations in hematocrit, the ratio of red blood cells to plasma in whole blood. The glucometer’s superior design can avoid the former inaccuracies, while the latter vary greatly in the blood. To avoid contamination and hence ensure accurate glucose testing, the patient should maintain proper cleanliness. The fourth source of mistakes, pharmacological sources of errors, is usually minimal when different drugs can influence the measured glucose level. Other sugars, such as maltose and xylose, can interfere with glucose dehydrogenase. The most hazardous chemical, however, is icodextrin.

2.3. Fabrication Procedure

Attaching nanostructures to the surface of the electrode while keeping their designed morphology, large surface area, and physicochemical properties for promising sensing applications is always a huge issue. Various synthesis procedures can be used to insert the catalysts onto a suitable surface to construct 1D, 2D, and 3D electrodes for enzyme-free glucose bio-detection. In 2008, Lee et al. fabricated a non-enzymatic silicon CMOS (complementary metal-oxide-semiconductor) integrable micro-biosensor [58]. Many advantages exist for silicon CMOS integrable microsensors, including reduced cost, smaller size/volume, low power consumption, mass production and implantable devices. First, an insulating layer was deposited over the silicon substrate. Using the dry-etching process, a Ti/Pt film was sputtered over the SiO2 layer and patterned to generate three distinct electrodes (counter CE, working WE and reference electrodes RE, Figure 6) of a three-electrode system. After that, the Pt WE were electroplated with nanoporous Pt. After introducing the sample into the C16EO8 mixture, liquid crystal templates of C16EO8 were generated over the Pt layer by lowering the temperature from 85 °C to various degrees. On top of the flat Pt electrode, a liquid crystalline hexagonal structure was produced at this stage. After removing the surfactant mould in deionised water, the nanoporous Pt electrode was created. Finally, the micro-sensor was created by screen printing Ag/AgCl paste over the RE. By comparing RFs and reproducibility of nanoporous Pt electrodes, the ideal electroplating conditions were established to be 35 mC/mm2, −0.12 V, and 25 °C, respectively. Increasing the applied voltage, charges, and temperature would reduce the roughness factor (RF) of Pt-based electrodes, leading to an increase in Pt grain size and overplating of the nanomolds generated. The optimized results reveal that the produced CMOS integrable non-enzymatic micro-sensors are promising for tiny handheld health care systems and continuous monitoring system applications in vitro and in vivo.
The cyclic voltammetry approach can be used to create disposable screen-printed carbon/nickel composites over indium tin oxide (ITO) electrodes for glucose detection without the usage of enzymes (DSPNCE) [59] (Figure 7). The prepared electrodes were notable for their lack of interference from common physiologic interferents such as ascorbic acid (AA) or uric acid (UA). As a result, this approach enabled the fabrication of a simple, disposable glucose biosensor.
The electrochemical deposition was used to successfully prepare cuprous oxide nanoscale thin-film electrodes with various shapes [60]. According to the results, the synthesized cuprous oxide had a greater purity, a tidy morphological structure, and a consistent grain size. This thin-film sensor having a sword-shaped dendrite, responds well to glucose and shows potential for use in sensors. Furthermore, it has a good linear relationship in the glucose concentration range of 1–20 mgL−1 with a detection limit of 0.337 mgL−1, a sensitivity of around 23.24 mA cm−2 mM−1, and high stability. For the first time, the template-assisted electrodeposition technique was used to create high-performance Co–Ni–Cu alloy nanotubes for non-enzymatic glucose detection [61]. The electrodeposition voltage and duration were adjusted for the production of high-quality ternary alloy nanotubes. The great performance is due to the high surface area of nanotube arrays as well as the synergistic effects of Ni, Co and Cu metal components. Furthermore, they demonstrated strong dependability and anti-jamming capability. Tafel plots revealed that alloying increased the exchange current density and reaction speed. These synergistic effects can also be inserted into 2D electrodes by alloying for improved glucose detection. Electrochemical techniques are also employed to make Ag/CNTs nanocomposite electrodes for non-enzymatic glucose sensors [62]. The SDS and COOH functionalized CNTs solution on the FTO electrodes generated an extremely dispersed solution for uniform and homogeneous spin coating of CNTs. Silver nanoparticles of different concentrations were electrodeposited on the manufactured CNTs electrode using cyclic voltammetry. The CNTs-coated electrode serves as a support system for the Ag nanoparticles’ highly electroactive surface area. During electrochemical analyses, the identification of glucose molecules improved as the number of Ag particles increased. These nanoparticles could be used to replace pricey electrodes in electrochemical glucose detection. The SDS and COOH functionalized CNTs solution on the FTO electrodes generated an extremely dispersed solution for homogeneous and uniform spin coating of CNTs. Silver nanoparticles of different concentrations were electrodeposited on the manufactured CNTs electrode using cyclic voltammetry. The CNTs-coated electrode serves as a support system for the Ag nanoparticles to enhance the highly electroactive surface area. During electrochemical analyses, the identification of glucose molecules improved as the number of Ag particles increased. These nanoparticles could be used to replace pricey electrodes in electrochemical glucose detection.
Electrodeposition onto a composite electrode consisting of reduced graphene oxide (rGO) mounted on a Ni foam substrate can be used to successfully manufacture uniform and porous CoNi2Se4 (prepared hydrothermally) [63]. This CoNi2Se4–rGO@NF hybrid electrode was used as an electrocatalyst for direct glucose oxidation, resulting in a high-performance non-enzymatic glucose sensor. The research also highlights the impact of decreasing anion electronegativity on improving electrocatalytic efficiency by lowering the voltage required for glucose oxidation and increasing the electrode’s selectivity for glucose molecules. A CuxO–NiO nanocomposite film was made using a unique modifying approach for non-enzymatic glucose determination [64]. For 15 min, an anodized Cu electrode was immersed in a solution of H2SO4, NiSO4, and CuSO4. Then, in a mixed solution of NiSO4 and CuSO4, a cathodization process with a step potential of -6 V was commenced, resulting in the creation of a porous Ni–Cu layer over the bare Cu electrode by electrodeposition facilitated by the release of H2 bubbles serving as soft templates. The electrodeposition procedure was optimized using experimental design software. The sensor showed high selectivity against some usual interfering species and high stability.
The unique nanostructure and high specific surface area of slack porous nano-TiO2 for efficient bio-molecule immobilization attract glucose sensors for electrode manufacturing. As a result, a sensor with an extremely low potential of 0.2 V was constructed using a nano-TiO2 coated aluminum tip electrode [65]. Surprisingly, Cu2O nanocubes [66] are also developed using the wet chemistry technique. A sensitivity value of 1040 µA/mMcm2 was evaluated for the electrode in the linear range from 0.007 to 4.5 mM, and a LOD of 3. 1 µM (S/N = 3). The electrode also showed satisfactory sensitivity towards other molecules, including sugar and ethanol.

3. 2D Materials for Electrochemical Glucose Sensing

3.1. Graphene

Since Novoselov and Geim discovered graphene in 2004, it has been a prominent carbon material. Graphene is a type of carbon material that features a honeycomb-like crystal lattice and an ultrathin one-atom-thick sheet structure. Graphene-based materials show potential application in electrochemical glucose sensing. Table 3 shows the properties of graphene that facilitate glucose sensing. The quantity of research publications and reviews on graphene and graphene-hybrids for electrochemical sensing has been steadily increasing in recent years. Different research groups have thoroughly examined graphene-based electrochemical glucose sensors. Zhang et al. [18] reviewed the current achievements in the preparation of graphene-based material for electrochemical glucose sensing. G. Gnana Kumar reviewed non-enzymatic electrochemical glucose sensors based on graphene [18]. According to these reviews, graphene is a promising material for electrochemical glucose detection. Comparing GO and rGO, pristine graphene has less interfacial contact; hence GO and rGO were widely utilized in high-performance glucose sensors. Since the application of graphene-based electrochemical glucose sensors was already reviewed in 2018 and 2017, we will focus here on the studies done after 2018.

3.1.1. Enzymatic Glucose Sensors Based on Graphene

Enzyme immobilization on various nanostructured materials (organic or inorganic composites, nanoparticles and 2D materials) has recently received considerable attention in electrochemical sensors because it improves enzyme activity and selectivity towards specific target analytes. Considering the combination of an organic or inorganic composite with a 2D material, Baek et al. modified the gold chip with Cu nanoflower- gold NPs-decorated GO nanofibers. Here GO nanofibers were electrospun over a gold chip that was further decorated with Au nanoparticles following Cu nanoflowers; then, the gold chip was covered with 1% Nafion as a binding agent. With the addition of glucose, it interacts with O2 in the presence of Cu-nanoflower@AuNPs-GO NFs to form H2O2 using an enzymatic catalytic process, resulting in a significant increase in electric current. The doping of AuNPs over GO NFs helps to enhance the sensitivity of glucose. The growth mechanism for Cu-nanoflower follows the nucleation and growth phases. While the Cu2+ ion interacts with the phosphate anion in PBS solution to form a copper phosphate crystal, the amide backbone in proteins (HRP and GOx) coordinates with the crystal to form the Cu–protein complex. This complex acts as a seed for the nanoflower, and the nuclei expand over time to form the nanoflower’s petals. A multi-layered nanoflower structure might be formed by a sequence of these successive processes. The Cu-nanoflower reacts with the glucose itself, forming a synergetic action with the GO NFs and AuNPs. Its stability was bolstered by the GO NFs as a backer. Because the 3D-structured AuNPs increase the surface area of the AuNPs, an intrinsic peroxidase-like activity could help HRP improve the electrochemical activities [67]. The immobilization of the enzyme over the nanoflowers such as Cu helps to enhance the sensitivity, durability as well as stability. In addition to this, GO NFs as a supporter also enhance the stability. This sensor shows excellent current even after 20 days suggesting the potential for field use.
Another electrochemical sensor was fabricated by Mao et al. He used rGO to improve the sensitivity and selectivity of the ZnO nanorod biosensor. In this work, the ZnO nanorods were hydrothermally synthesized over a polyethylene terephthalate (PET) substrate. The ZnO/PET working electrode was then coated with electrodeposited rGO, and AuNPs were scattered on the surface, resulting in ZnO/rGO/Au/PET. Lastly, the GOx was physically adsorbed on the electrode’s surface, yielding a GOx/rGO/ZnO/Au/PET glucose sensor with a sensitivity of 56.32 µA mM −1 cm−2 and a linear range of 0.1 to 12 mM [68]. However, under 10 cycles of bending, this working electrode exhibits good current responsiveness. However, after 15 cycles, the device’s performance is poor because GOx and rGO peel away from the ZNO nanorod, resulting in microcrack development [68]. Hossain and Slaughter have suggested a hybrid glucose biosensor employing MWCNTs and graphene with great sensitivity and selectivity. A one-step solvothermal approach was used to make a solution containing both MWCNTs functionalized with carboxylic groups and chemically-derived graphene. PtNPs were electrochemically deposited onto a thin layer formed by drop-casting this suspension onto an Au electrode. Finally, GOx was immobilized and coated with Nf on the nanostructured electrode. The hybrid biosensor was built with a sensitivity of 26.5 µA mM−1cm−2 and a linear detection range of 0.5 to 13.5 mM [69].

3.1.2. Non-Enzymatic Glucose Sensors Based on Graphene

A large number of transition metal nanoparticles such as platinum (Pt), gold (Au) and palladium (Pd) were synthesized and employed in non-enzymatic sensors. These metal nanoparticles have an electronic structure having unpaired-d electrons, and unfilled d-orbitals help in the electrocatalytic activity of glucose. In order to enhance the electrocatalytic activity of glucose, several supports have been investigated, among which GO serves as a suitable supporting material. Because of its high catalytic activity and stability, platinum is commonly employed as an electrocatalytic electrode material. Sakar and colleagues describe a unique design of glucose sensor with graphene Schottky diodes consisting of a graphene (G)/platinum oxide (PtO)/n-silicon (Si) heterostructure, taking into mind the benefits of combining graphene with the platinum electrode. They discovered that the platinum oxide film thickness affects the sensor’s sensitivity and that raising the PtO film thickness can increase the sensor’s sensitivity by up to 150%. This was ascribed to an increase in the number of active sites for glucose oxidation and the thickness of the graphene layer, resulting in increased charge carrier mobility and concentration. The device works by glucose molecule oxidation over the surface of the suggested heterostructure electrode due to the catalytic activity of the PtO thin film, as illustrated in Figure 8. Aside from that, the glucose molecules are broken down into gluconolactone, which creates gluconic acid, H2, and electrons. The current was measured from the graphene surface after applying a forward bias to the Si terminal. The physisorption interaction between PtO (metal-semiconductor Schottky junction) and graphene caused a shift in graphene’s fermi-level location and p-doping [70]. Here graphene is employed as both a protective and sieving layer to protect the PtO film and further enhance the sensor stability and selectivity.
Compared to Pt-based electrodes, the main advantage of using Au-based electrodes for glucose sensing is the higher current response, which allows for higher sensitivity and the ability to detect glucose at a neutral pH. However, the main drawback of Au-based electrodes is the low glucose oxidation efficiency on the Au electrode surface, which can be mitigated by using arrays of nanoelectrodes separated by non-electroactive materials. Furthermore, because these electrodes are better activated in alkaline solutions, they cannot be used in in vivo studies, they have surface contamination from anions such as phosphates and chlorides, and their selectivity is significantly smaller than Pt-based electrodes. Scandurra et al. prepared a graphene paper-based electrode using the dewetting technique. In this case, an 8-nm-thick Au layer was sputtered onto graphene paper before being dewetted with a laser. Dewetting with a laser-produced smaller AuNPs on the electrode surface. The sensor’s sensitivity was 1240 µA mM−1 cm−2 [71].
Combining the advantages of another transition-metal oxide palladium (Pd) and CNT with graphene nanoplates, Kiattisak et al. modified glassy carbon (GCE) using a nanocomposite of multi-walled carbon nanotubes wrapped with palladium nanoparticle-graphene nanoplatelets (PdNPs-GNPs/MWCNTs). This modified GCE was coupled with a flow injection amperometric detector where this sensor exhibits a detection limit of around 0.008 mM. The following reactions can be used to describe the mechanism of glucose oxidation by Pd nanoparticles.
Pd + Glucose → Pd-Glucose ads + H+ + e
Pd (OH)2 + glucose → Pd + gluconolactone + H2O
Pd + 2OH → Pd (OH)2 + 2e
Electrode materials such as Au and Pt are suitable for glucose detection, but they are costly. As a result, non-precious transition metals such as nickel (Ni) and copper (Cu) and their oxides have been studied. Despite the wide range of catalytic materials available for glucose sensors, nickel (Ni)-based nanomaterials have attracted researchers’ interest due to their high catalytic activity for glucose oxidation in alkaline medium, resulting in NiOOH and Ni(OH)2 species. Despite their high electrocatalytic activity, surface fouling from glucose oxidation compromises their stability. As a result, a conductive nanostructured substrate is required to improve electron transport and electrode stability against Ni glucose oxidation. Considering graphene’s excellent properties as a catalyst support material, Jothi et al. fabricated a non-enzymatic glucose sensor based on graphene nanoribbon/graphene sheet/nickel nanoparticles to prevent re-stacking and increase surface area. This Ni-based hybrid has a larger specific surface area, more active sites, and better electrical conductivity, allowing for easy ion transport and unrestricted OH- diffusion throughout the electrochemical process. The mechanism of glucose oxidation by Ni nanoparticles can be explained using the reactions listed below [72].
NiO(OH) + glucose → Ni(OH)2 + gluconolactone
Gluconolactone → gluconic acid
Among these Ni-based materials, Ni-based porous materials have a large surface area, and more active sites have been used as sensitive materials for glucose sensors. To exhibit the full potential of porous Ni materials, different kinds of substrates such as Ni foil, SPEC, and Cu foil have been used. Still, the electron transferability of porous Ni is still weak. Exploiting the properties of graphene, where graphene provides anchoring, conducting and separating actions, Ren et al. prepared a porous Ni over exfoliated graphene by using the hydrogen bubble method over the Cu foil substrate. This porous sensor shows high sensitivity due to the high electron transfer ability between the active reaction centre and electrode induced by the graphene [73]. Another Ni/graphene hybrid material was reported by Lavanya et al. where this hybrid material was synthesized using a simple in situ chemical reduction method. To begin, Ni2+ was adsorbed onto the GOR and GOS surfaces via the electrostatic interaction between functional groups (epoxy, carboxyl, and hydroxyl groups) and the Ni2+ on the GOR and GOS surfaces. The Ni(OH)2 developed in GOS after adding NaOH precipitant was then co-reduced with the help of hydrazine hydrate. The reducing agent contributes to the reduction of GOR and GOS, the conversion of Ni2+ ions to Ni0 nanoparticles, and the development of Ni nanoparticles in the GS/GNR network throughout this process. This sensor shows a detection limit of around 2.5 nM, and this superior catalytic performance of this sensor was due to the synergistic action of Ni nanoparticle and GS/GNR hybrid network. The Ni nanoparticles were uniformly dispersed throughout the GS/GNR hybrid to increase electrical conductivity, a greater surface area with more active sites, and unrestricted flow of OH- ions during the electrochemical process. The incorporated Ni nanoparticles are an effective catalytic active material for direct glucose oxidation, which improves electron transfer and results in a high lower detection limit, selectivity, and good sensitivity for glucose detection. Here during the co-reduction of GOR, GOS and Ni(OH)2, the abundant functional groups in the basal and edges of GOR and GOS offer additional anchoring sites for the incorporation of more Ni nanoparticles, i.e., catalytic site, and increase the rate of electron transfer between the electrode and glucose and thereby enhancing the performance of sensor [72].
According to recent research, bimetallic materials, particularly bimetal alloys, outperform monometallic counterparts in catalytic performance. Several studies on the performance of bimetallic materials for glucose sensing have been published. However, its performance can be improved by using a support material such as graphene. Deng et al. created a NiFe alloy nanoparticle/graphene oxide hybrid (NiFe/GO) based on this [74]. Rukiye et al. described another Ni-based bimetallic material in which the working electrode was decorated with monodisperse platinum–nickel nanocomposites-decorated on reduced graphene oxide (Pt/Ni@ rGO) manufactured utilizing a novel ultrasonic hydroxide aided reduction technique. This sensor shows excellent electrochemical activity with a sensitivity of around 171.92 μA mM−1 cm2 and LOD of 6.3 μM [75].
Among the transition metallic nanoparticles, metal sulfides, which are found as minerals in nature, have higher cycle stability than conducting polymers. Furthermore, because of the numerous sorts of structures, they have a lot of research potential in glucose sensors. Among these, CuS has attracted a lot of attention due to its exceptional fundamental property diversity, such as attractive photovoltaic capabilities, strong thermal stability, great transport properties, and so on. Yan et al. prepared CuS nanoflakes reduced graphene oxide (rGO/CuSNFs) nanocomposite to avoid its coagulation using a surfactant-free method. This sensor shows excellent sensitivity and a low-detection limit [76]. On the other hand, aggregation and low dispersion remain major issues in the synthesis of MNPs on graphene support, resulting in poor electrocatalytic performance and low stability.
Intrinsically conducting polymers (CPs) are one of the most relevant and extensively used materials for sensor modification due to their unique chemical and physical properties, such as adjustable architecture, adaptability, versatility, room stability, and sensitivity to surface changes in electrochemical activity with minor changes in its surface. Polypyrroles, polyanilines, and polythiophenes have received a lot of attention because of their good film-forming properties, electrical semiconductivity, great transparency in the visible range, and exceptional thermal and environmental stability [77]. CPs have recently been hybridized or mixed with graphene-based materials. When the materials are linked, each component’s combined optical, electrical, thermal, mechanical, chemical, or electrochemical capabilities can be exploited for chemical and biological sensing [78]. For glucose measurements, a composite comprising (PANI/PDPA) along with graphene nanosheets synthesized in a liquid–liquid (CHCl3/HCl) interface individual by Muthushankar et al. During the polymerization procedure, the PANI and PDPA chains grew nicely and were evenly distributed over the graphene nanosheets. The improved electro-catalytic activity of Gra-PANI-co-PDPA-ME towards glucose was achieved with greater sensitivity of 0.51 µA/µM at 5 s. The presence of aniline, graphene and diphenoquinone diamine (DPDI2+) together with their synergistic relationship improves electron transport for glucose oxidation. Interference tests further confirm that the constructed sensor is best suited for glucose sensing performance [79]. Silver (Ag) nanoparticles are considered a promising candidate because of their excellent stability and biocompatibility. Utilizing the advantages of this, Ag nanoparticle and carbon-based nanocomposite having excellent electrical conductivity and stability, Deshmukh et al. demonstrated an Ag-PANI/rGO nanocomposite using a simple hydrothermal method. This composite demonstrated a detection limit of 0.79 µM with a quick response time. They proved the viability of using Ag-PANI/rGO nanocomposites to detect glucose in real-world samples of organic fluids (milk, apple juice, mango juice, orange juice, and Coke). Here the introduction of rGO into PANI polymer improves the operational stability and electron transfer rate. This sensor shows stability up to 30 days [80]. Poly(3,4-ethylenedioxythiophene) (PEDOT) is another candidate that has piqued the interest of this group of CPs. PEDOT has been extensively used as an electrode material among CPs due to its low oxidation potential, good stability and mid-band gap energy. Considering the favorable properties of this polymer, a non-enzymatic glucose sensor based on an Au electrode was modified utilizing electroreduced graphene oxide (ERGO) and layered PEDOT by Mesut et al. The thin films of PEDOT–ERGO were made using a simple electrochemical approach based on progressively layering electrodeposition on an Au electrode, which was employed as a non-enzymatic glucose sensor in this study. The PEDOT–ERGO nanocomposite modified Au electrodes (Au–PEDOT–ERGO) were employed as electrocatalysts for voltammetry and amperometry glucose detection. This sensor exhibits a LOD of around 0.12 μM with a sensitivity of 696.9 μA mM−1 cm−2 [81].

3.2. MXene

MXene is suitable for building high-performance electrochemical glucose sensors because it has high hydrophilicity due to surface termination groups (O, OH, and F), great electrical conductivities, excellent ion intercalation behavior, facile functionalization, and dependable large-scale manufacture. Several high-quality reviews on MXenes have been published to date, suggesting the applicability of MXene in sensor fabrication [82,83,84,85]. However, no review focused on MXene-based electrochemical glucose sensors is available so far. Table 3 shows the properties of MXene facilitate in glucose sensing

3.2.1. Enzymatic Glucose Sensors Based on MXene

MXenes, similar to other 2D materials, can include additional materials such as metal nanoparticles, enzymes, CPs, and metal oxides, resulting in improved structural and electrical characteristics [86]. For example, without the need for external reducing agents, one-step hybridization of Au, Ag, and Pd nanoparticles from their respective aqueous solutions onto the surface of MXene was performed and employed as a substrate for SERS. Ti3C2 was the first MXene to be employed in constructing electrochemical glucose sensors, and Rakhi and her colleagues produced the first electrochemical sensor based on Ti3C2 in 2016. A nanocomposite of gold nanoparticles (Au) and MXene (Ti3C2Tx) for glucose detection has been reported. After in situ reductions of chloroauric acid with sodium borohydride, they deposited Au nanoparticles on the surface of MXene, and the nanocomposite was subsequently dispersed in Nafion. This composite was then drop casted on a glassy carbon electrode before being immobilized with glucose oxidase (GOx) to produce Au/GOx//MXene/Nafion/GCE. Au nanoparticles were critical in increasing the electron exchange between the electrode and the active center of GOx. After a first morphological examination, the biosensor was assessed using CV and amperometry. The glucose-sensing mechanism is based on the following Equations (6) and (7). GOx is composed of two extremely similar protein subunits and one coenzyme molecule, flavin adenine dinucleotide (FAD). FAD is reduced to FADH2 during the transfer process with 2e and 2H+ because it is present in the active site of the GOx. The biochemical reaction between GOx and glucose culminates in the conversion of glucose to glucono-D-lactone and the reduction of FAD to FADH2. FADH2 was then oxidized by dissolved oxygen to create H2O2 and form FAD.
GOx (FAD) + Glucose → GOx (FADH2) + glucono-D-lactone
GOx (FADH2) + O2 → GOx (FAD)+ H2O2
MXene, as well as Au nanoparticles, can both contribute to improved electron transfer kinetics between the electrode and active redox centers of the enzyme, while the large surface area with the distinctive layered architecture of Au/MXene nanocomposite effectively accommodates the enzyme. The amperometric response was shown to be linear over a concentration range of 0.1 to 18 μM, with a LOD as low as 5.9 μM. This sensor also shows long-term stability. Furthermore, in the presence of AA, UA and DA, this suggested biosensor is highly specific for glucose. The use of enzymes for glucose detection ensures a strong electrochemical response, particularly in complicated matrices. However, there are significant drawbacks to using enzymes, including the risk of enzyme inactivation, the reproducibility of enzyme immobilization, and the increased cost of analysis. Hence, the development of non-enzymatic sensors is becoming increasingly popular [87]. Using a mixing–drying process, researchers improved biosensing capabilities by combining MXene nanosheets with hydrophilic groups with the properties of graphene sheets to offer new functions. This three-dimensional (3D) porous hybrid film had a more open structure, which allowed glucose oxidase to enter through the inner pores, improving the hybrid film’s stable immobilization and retention of GOx. The constructed biosensor outperformed conventional 3D porous materials with sensitivities of 20.16 mM (LOD 0.13 mM) and 12.10 mM (LOD) 0.10 mM) in O2 saturated phosphate-buffered solution (PBS) and air-saturated PBS, respectively. This 3D porous film results in higher stability for GOX immobilization. This is due to (i) the enhanced hydrophilic property of this hybrid MG film, (ii) the open surface facilitates more access to GOx, and (iii) the excess 3D pores and hydrophilic wall of film this film provides a favorable microenvironment for GOX to stay in the pores and avoid re-dissolution in PBS [88].
In practical applications of glucose biosensing, the harmful intermediate product, H2O2, formed during enzymatic glucose oxidation usually hinders the action of GOx. Wu, M. et al. created a hybrid Ti3C2/poly-L-lysine (PLL)/glucose oxidase (GOx) nanohybrid that could catalyze the cascade processes of glucose oxidation and the intermediate H2O2 breakdown to address this issue. The PLL-modified MXene possesses a positive charge and exhibits excellent GOx loading capacity where the amine-groups in PLL are able to form a crosslink with the physically adsorbed enzyme to form MXene nanosheets covered with GOx/PLL. The Ti3C2 MXene was shown to be capable of catalyzing the breakdown reaction of H2O2, and when combined with the GOx, a cascade reaction in glucose decomposition was created. Ti3C2/PLL/GOx nanoreactors with higher catalytic activity were placed on glassy carbon electrodes to create a 2.6 μM LOD glucose biosensor [89].

3.2.2. Non-Enzymatic Glucose Sensors Based on MXene

Li et al., for the first time, reported a non-enzymatic glucose sensor based on MXene/Nickel–Cobalt layered double hydroxide (NiCo-LDH). They synthesized MXene over NiCo-LDH using a simple hydrothermal method. Because MXene is negatively charged, it can absorb cations and provide nucleation sites to form nanoparticles. The redox reaction in the following equation produces hydroxyl ions, which react with Co2+ and Ni2+ to make hydroxide monomers on the MXene sheet, and the monomers then react with others to produce primary particles in the hydrothermal reaction. In the meantime, some divalent cobalt is oxidized to trivalent cobalt. As the number of initial particles increased, they progressively collected and transformed into nanosheets coated on MXene.
4CH3OH + NO3 → 4HCHO + NH3 + OH + 2H2O
This glucose sensor has a broad linearity range (0.002 mM–4.096 mM), a LOD of around 0.53 µM, with a quick response time (3 s) at a working potential of 0.45 V (vs. SCE). Additionally, good selectivity, stability, and repeatability were obtained [90]. A non-enzymatic glucose sensor based on the MXene–Cu2O hybrid was recently studied by Gopal et al. This sensor has a linear range of 0.01–30 mM and an LOD of 2.83 µM [91].

3.3. TMD Based Electrochemical Glucose Sensors

Two-dimensional TMD nanosheets are appealing as electrochemical glucose sensors because of their high conductivity, large surface area, high signal/noise ratio and quick electron transfer kinetics and, most importantly, their practicality for forming composites.

3.3.1. Enzymatic Glucose Sensors Based on TMD

Other metal NPs, such as Au NPs, are utilized to alter MoS2 in addition to Ni and Cu NPs in enzymatic glucose sensors [92]. The use of AuNPs accelerates electron transport from the electrode to the immobilized enzyme. This enables GOx electrochemistry without the usage of an electron mediator. Thus, the GC modified with Au NPs and MoS2 was studied by Su et al. and showed enhanced electrocatalytic activity [93]. Another electrochemical glucose sensor was fabricated by Parlak et al. using a MoS2/Au NPs hybrid. Here, enzymes aided in directly transferring ions to the modified MoS2 surface, increasing its bioelectroactivity for redox processes. Figure 5 depicts a typical biosensor fabrication procedure in which AuNPs were self-assembled over the MoS2 nanosheets and subsequently connected to GOx for glucose detection in the presence of a mediating agent, ferrocene carboxylic acid. Because of their open 3D architectures with a significant interlayer space and surface area for enzyme immobilization, these electrodes demonstrated remarkable sensitivity to glucose. As a result, fast mass transport was accelerated, and glucose diffused quickly across the electrode surface. Despite the fact that the study revealed bioelectrocatalytic reactions of glucose, suggesting substantial improvements in biocatalysis at a MoS2-based enzymatic nanointerface, real-sample analysis and interferences were not performed, which could have demonstrated selectivity and further justified the analytical functionality of the biosensor. The constructed MoS2 enzymatic electrode, on the other hand, revealed a new potential for creating a unique 3D sensing substrate [48].
Because H2O2 is a consequence of many oxidative biological activities, measuring it with MoS2 electrodes allows for the identification of many more tiny molecules by integrating the sensor into lab-on-chip devices for intracellular detection. For instance, a glucose sensor based on MoS2 demonstrated strong activity towards H2O2. GOx was immobilized over the surface and used to detect glucose. The current response of this biosensor grew as the concentration of glucose climbed from 2.0 to 16.0 mM. Despite the fact that glucose oxidation consumes O2, the biosensor’s overall current response increased because both O2 and H2O2 contributed to the increase in reduction current. To explore the large surface area and enhanced electrical conductivity of 3D porous graphene aerogel, Jeong et al. fabricated a flow-injection biosensor device. They incorporated 2D MoS2 with 3D graphene aerogel using a facile hydrothermal approach to enhance the electrochemical sensing performance. Here the different porous structures of 3D MGA provide rapid, efficient pathways for electrons and ions, demonstrating excellent electrochemical performances. The following qualities contribute to the superior performance of these 3D MGA-based biosensors. At the interface, the interconnected network of MoS2 exposes extensive basal planes against the electrolyte solution, which enhances the H2O2 reduction activity of MoS2 nanosheets using oxygen substitution at exposed Mo edges. Thus, showing excellent selectivity and sensitivity [94,95].

3.3.2. Non-Enzymatic Glucose Sensors Based on TMD

Many researchers have worked ceaselessly on sensor fabrication for glucose detection employing MoS2 nanosurfaces among TMDs [96]. Massive efforts have been undertaken to tackle MoS2 difficulties such as low electrical conductivity and limited electrical transportation. The extended technique for other heteroatom doping such as B and N for increased semiconducting characteristics and catalytic reaction revealed an intriguing advancement with doping of Ni in the MoS2 framework. For the first time, Huang et al. fabricated an electrochemical glucose sensor based on Ni–MoS2 hybrid [50]. As an extension of this work, a functional hybrid of Ni-doped MoS2 and rGO was later fabricated. The electrochemical analysis demonstrated that the Ni-doped MoS2/rGO composite has a high electroactivity due to the efficient electron transport rates, highly exposed catalytic sites, electrical conduction efficiency, and large effective surface area of MoS2 and rGO. After successfully detecting glucose in human serum samples, the sensor demonstrated its great analytical value as a prospective catalyst for non-enzymatic investigation of additional metabolites such as cholesterol, H2O2 and lactose, among others [97]. In another case, Ji et al. modified glassy carbon with nickel (II) hydroxide nanoparticles and MoS2 film using an electrodeposition method. Here MoS2 act as catalytic support for the nickel (II) hydroxide nanoparticles and thereby enhancing the electrochemical activity of bare MoS2 and nickel (II) hydroxide NPs [98]. Huang et al. decorated MoS2 nanosheets using Cu nanoparticles. The prepared hybrid using the chemical reduction method and cyclic voltammetric and amperometric studies in alkaline media showed good sensitivity, 1055 µA mM−1 cm−2 [95]. Cu2O electrocatalytic activity has sparked interest in non-enzymatic glucose sensors, similar to Cu electrocatalytic activity. As a result, a new combination of 3D MoS2 nanoflowers and Cu2O metal oxide was developed for non-enzymatic glucose detection using amperometry. The electrode’s exceptional electrochemical activity could be attributed to synergistic effects offered by the peculiar structural characteristics of Cu2O–MoS2 nanohybrid enveloped in 3D flower-like MoS2 with thin interconnecting nanosheets and largely scattered Cu2O. The huge surface area of MoS2 nanosheets allowed for substantial Cu2O dispersion, boosting the amount of electrochemically active sites and, as a result, electrical conductivity and fast heterogeneous electron transfer rates. This sensor shows excellent sensitivity with an LOD around 1 µM [99]. Wu and their group have demonstrated the first MoS2-nanosheet-based electrochemical sensor. In the redox systems, exfoliated MoS2 nanosheets were electrochemically reduced in NaCl solution to generate reduced MoS2 (rMoS2) with a fast electron transfer rate and strong conductivity. The generated rMoS2 was used to detect glucose by immobilizing GOx and to specifically detect dopamine (DA) in the presence of UA and AA, as shown in Figure 9A. The reduction current at 0.33 V, arising from oxygen reduction, decreased as the glucose concentration increased from 0 to 20 mM [100].
Several reports on electrochemical glucose sensors based on MoS2 are available. Other TMD-based glucose sensors, on the other hand, are still being investigated. For the first time, Gayathri et al. reported a MoSe2 based non-enzymatic glucose sensor. Using a simple hydrothermal method, they decorated these MoSe2 nanosheets over a NiO nanorod. This composite modified GCE shows excellent electrochemical activity. The electro-catalytic mechanism for glucose oxidation is as given in following Figure 9B, and it follows according to the equation given below:
NiO + OH → NiO(OH) + e
NiO(OH) + glucose → Ni(OH)2 + gluconolactone
This composite had a linear response for glucose detection from 50 mM to 15.5 mM, with a LOD of 0.6 µM [101]. Another transition metal selenide-based non-enzymatic glucose sensor was reported by Mani et al. for the first time. The synthesized NiSe2 nanosheets using a facile-hydrothermal approach and modified the GC with these nanosheets show excellent activity towards glucose [102]. Comparing transition metal sulfides and selenides, tellurides possess lesser electron negativity, large atomic size, high conductivity, and less toxicity. Due to these peculiarities, transition-metal tellurides received considerable attention in the field of biosensors. For example, Ni3Te2 nanostructure was reported as a successful composite for non-enzymatic glucose sensing [103]. Similarly, CoTe2 nanosheets, which were grown over a 3D nickel foam was, also show excellent electrochemical activity and used for glucose sensing. Among the transition metal tellurides, CoTe2 exhibits exceptional electrochemical performance and high stability, making it an excellent candidate for glucose sensing fabrication [104].

3.4. Layered Double Hydroxides (LDHs)

Nanostructures, having a large surface area and a rough surface, may be capable of transmitting faraday currents for oxidation while reducing the impact of external interfering molecules, such as UA, AA, and AP. LDHs are a type of two-dimensional material with the structural formula [M1−x II M xIII(OH)2](An−)x/n.mH2O, where MII and MIII are divalent and trivalent metals, respectively, and An− signifies the anions between the interlayer spaces. The properties required for glucose sensing are shown in Table 3.

3.4.1. Enzymatic Glucose Sensors Based on LDH

Enzyme-based biohybrid materials, which are made by immobilising various oxidoreductases in the LDH host structure, have been used to detect target analytes such as hydrogen peroxide, phenol derivatives, glucose and other inhibitors or enzyme substrates utilizing amperometric transduction methods, primarily chronoamperometry. There have been very few publications on LDH-based enzymatic glucose sensors. Because of its stability, glucose oxidase (GOx) is frequently used as a model to examine various biosensor setups. The major findings from investigations using glucose biosensors based on the immobilization of GOx in LDH matrices are presented here. Physical adsorption over the electrode surfaces produced by gradual evaporation of colloidal suspensions of mixes of LDH and GOx has been the most extensively utilized method for GOx immobilization. The entrapped biomolecules are cross-linked with the help of glutaraldehyde in the vicinity of bovine serum albumin to prevent the enzyme from being released. Glutaraldehyde was not used since composite materials based on chitosan or alginate with Zn/Al-Cl LDH were used instead. This method of deposition allows a substantial and known quantity of enzyme to be entrapped in the clay layer, assuring great response sensitivity. In this regard, Colambari et al. synthesized Mg/Al LDHs containing ferrocene sulfonate or ferrocene carboxylate (Fc-COOH) as interlayer anions using the coprecipitation method, and they built glucose sensors based on the immobilization of GOx in a glutaraldehyde and BSA network. MnO2 nanoparticles were effectively integrated into the outer protective glutaraldehyde/BSA membrane to improve the selectivity of this GOx/Fc-COOH LDH/GC biosensor, allowing oxidizable interferents (e.g., acetaminophen, urate, and ascorbic acid) present in the sample solution to be analyzed to be peroxidised before reaching the electrode surface [105]. Shan and colleagues created a biosensor that is extremely sensitive to glucose utilizing Zn/Cr LDH. Ferricinium derivatives reoxidised FADH2 to FAD, which is followed by reoxidation of ferrocene (Fc) to Fc+ directly at the electrode, enabling glucose measurement under anaerobic circumstances in this biosensor [106]. The biosensor design reported by Shan et al. is based on a “ping-pong” mechanism for transmitting electrons from glucose to the mediator [107]. Mousty et al. developed another glucose sensor based on redox-active Zn/Cr LDH intercalated in the interlayer domain and contained (3-sulfopropyl) ferrocene carboxylate. Chronoamperometry at 0.5 V was used to examine the biosensor’s performance for glucose measurement under anaerobic conditions, and the sensitivity was 65 µA mM−1 cm−2 in the concentration range of 10–25 µM [108]. GOx was immobilized at the surface of the electrode by entrapment in inert LDH matrices (i.e., NiAl-NO3, ZnAl-Cl,) and ZnAl–Cl/biopolymers (chitosan or alginate) [109] composites with the goal of generating glucose amperometric electrochemical sensors. The majority of these LDH/GOx biosensors were made by solvent casting a predetermined amount of LDH and GOx combination onto a glassy carbon electrode or Pt surface. In these investigations, GOx was immobilized on LDH particles through adsorption, and then glutaraldehyde cross-linking was used to inhibit enzyme leaching. Tonelli’s group demonstrated yet another way of GOx immobilization by trapping GOx during the electrogeneration of NiAl LDH films over the Pt electrodes [110]. Farhat et al. recently published an amperometric biosensor based on Co3Mn–CO3 layered double hydroxide/GOx with biopolymer carrageenan. They used a coprecipitation approach to make this Co3Mn–CO3 LDH and then impregnated this LDH composite over a carbon felt to make this 3D porous sensor [110]. Zhang et al. revealed the first-ever direct electron transfer of GOx immobilized in Mg/Al hydrotalcite nanosheets, creating new avenues for the development of third-generation GOx glucose sensors employing LDHs as the immobilising matrix [111].

3.4.2. Non-Enzymatic Glucose Sensors Based on LDH

Through coprecipitation and electrodeposition, nickel and cobalt inculcated LDHs were produced and used for non-enzymatic glucose sensing [112]. Using the electro-oxidation processes of Ni ions, the LDH nanostructures outperformed GOD-based biosensors for glucose sensing even in the absence of GOD catalase. Furthermore, simple LDH materials were fabricated, such as porous structured NiFe LDH nanolayers grown over the nickel foam by the simple hydrothermal method and used for glucose detection [113]. Similarly, Zhao et al. fabricated a non-enzymatic glucose sensor using a multi-component sensing system where the electrode is made up of NiCo-LDH and cobalt copper carbonate hydroxide (CCCH) nanorods over a Cu foam. Here the large surface area and excellent conductivity of the Cu foam help in enhancing the electrocatalytic properties. This multicomponent hierarchical nanostructure features a large accessible surface area, as well as effective ion diffusion and electron transfer, thus enhancing the sensing properties of NiCo-LDH [114].
Hai et al. fabricated a NiAl LDH, loaded carbon microcylinder (CMC) composite using microelectromechanical systems (C-MEMS). When combined with electrocatalytic Ni centres, the suggested method demonstrated good performance due to improved conductivity [115]. Furthermore, several LDH nanoarchitectures intercalated with valuable noble metals, such as Au, NPs incorporating NiAl LDH SWCNTs–rGO hybrids exhibited stronger electrocatalytic capabilities for glucose oxidation than others. Good conductivity, generated from electron tunnelling junctions and 3D CNTs/rGO networks, is thought to be the source of these features [116].
When LDH-based catalysts are integrated with conductive carbon supports such as conductive polymers, CNTs, and graphene, the extraordinary properties of materials based on carbon, such as greater active sites, better consistency, and larger surface to volume ratio, can help improve the catalytic activity and stability of LDH layers. Inserting graphene between the LDH interlayers improves conductivity and glucose detection with high sensitivity while also boosting resistance to poisoning in chloride ion solutions. NiCo LDH-rGO [117] nanoribbons and NiFe LDH-rGO nanohybrids are used in this context [118] and were synthesized using electrochemical coprecipitation and deposition, respectively, and their catalytic properties were investigated. The electrodes with the nanostructures had a significant diffusion coefficient and excellent electrochemical catalytic activity towards glucose. Shahrokhian and group developed a core–shell structure to improve the electrocatalytic properties of CoNi-LDHs, notably in terms of shape, by growing CoNi-LDHs NS over a nanoporous tubular Cu(OH)2 utilizing a green, simple, and extremely controllable direct three-step in situ process over a GC. The step involves (1) copper film electrodeposition on GCE, (2) copper conversion to Cu(OH)2 nanotubes, and (3) CoNi-LDH electrodeposition over Cu(OH)2NTs/GCE. The rate-determining steps of the entire electrocatalytic reaction are represented by the equation below, which involves the diffusion of glucose molecules to the metal sites in the core–shell structure, where the OH travels from the alkaline solution and then permeates into the CoNi-LDH interlayers.
LDH (OH)− MIII + Glucose → LDH COIII + Gluconolactone
2CuOOH + 2e + Glucose → 2CuO + 2OH + Gluconolactone
This highly porous core-shell nanostructure adheres strongly to the conductive substrate, allowing for electrical and physical contact between the GCE surface and the active components, improving the overall stability and conductivity of the modified electrode. The commercialization capability of this sensor was also evaluated by growing this 3D hierarchical structure over a screen-printed electrode [119].

3.5. Other 2D Materials

Balasubramanian et al. synthesized a cobalt vacancy-rich Co(OH)2 ultrathin nanosheet over a screen-printed electrode and demonstrated its catalytic activity. A practical, simple, and in situ technique for obtaining cobalt hydroxide nanosheets with numerous cobalt vacancies is described here. The cobalt defects significantly enhance the charge transfer rates and increase the number of electroactive sites, resulting in outstanding glucose and L-cysteine oxidation performance. This VCo-Co(OH)2 electrode has a low detection limit of roughly 295 nM over a dynamic range of 0.4 M–8.23 mM [120]. Sahoo et al. studied the catalytic activity of non-faceted and faceted crystal cupric oxide (CuO) nanoribbons using a microwave and hydrothermal method. Here the non-enzymatic glucose sensor made-up of both non-faceted and faceted CuO crystals shows enhancement in amperometric oxidation current that is proportional to the glucose concentration. Their study shows that the glucose sensitivity of faceted CuO is higher than that of non-faceted crystal CuO due to the 2D thin and polygonal shape of this facet plane CuO. Here the facet plane of CuO, such as (100) and (110) planes, play a prominent role in improving the catalytic activity due to their higher surface energy. The faceted ends give more surface area and more electroactive species and are advantageous for electrolyte ion transferring and exchanging activities. The catalytic activity with glucose is as follows:
2CuO(OH)) + glucose → 2CuO + gluconolactone + H2O
Gluconolactone → Gluconic acid
The activation of these CuO nanostructures causes the amperometric current to increase with increasing glucose concentration. Here the hopping free-charges in the facet plane of CuO show a significantly lower detection limit of around 58 µM [121].
Other elements of group VA (also known as “pnictogens” or the “nitrogen” group) can likewise have a layered structure. These elements are similar to graphene: phosphorene, bismuthene, antimonene, and arsenene. Ling chia and group, for the first time, explored the use of pnictogens for non-enzymatic glucose detection. Because pnictogen nanosheets have strong carrier mobilities, they may aid in promoting fast electron transfer kinetics for glucose oxidation. In addition, pnictogens have excess lone-paired electrons on the surface and interact with glucose at high pH with good adsorption capabilities; thus, it could also aid in facilitating the adsorption mechanism by bringing glucose molecules into close vicinity with the electrocatalytic Au@Ag nanorods and enhance the sensing capabilities of both pnictogens and Au@AgNR. This study theoretically proves and opens up new vistas for the potential use of pnictogens-based composites in the future development of electrochemical sensors for the pharmaceutical, biomedical, and food industries [122].

4. Wearable and Flexible Electrochemical Glucose Sensors

Flexible and implanted glucose biosensors are a new technology for diabetes patients’ continuous blood glucose monitoring. It is necessary to develop flexible conductive substrates with a large active surface area in order to advance the technology. Zhao et al. prepared a free-standing non-enzymatic flexible electrochemical glucose sensor using a 3D monolithic nanoporous gold (Au) grown over a graphene paper which was further deposited with binary PtCo alloy nanoparticles using electrodeposition techniques. This flexible NPG/PtCo/GP electrode shows outstanding mechanical bending strength and can be integrated into a wearable biomedical or an implantable device. This flexible sensor offers great selectivity, repeatability, and stability, but it has a high manufacturing cost due to the many materials employed [125]. Similarly, another free-standing flexible sensor using graphene paper and binary PtCo alloy nanoparticles was reported by He et al. They formulated a conductive ink by combining 3D porous graphene-CNT assembly in ionic liquid (IL) and coating over this graphene paper. Due to the synergistic action of this 3D porous graphene-CNT assembly and PtAu alloy, this nanohybrid paper shows excellent sensing performances [126].
Abellán et al. created an enzymatic and non-enzymatic glucose sensor out of Pt-Graphite, which is regarded as the initial work on graphene-based wearable electrochemical glucose sensors [127]. Using simple, low-cost manufacturing procedures, Xuan et al. microfabricated and micro-patterned a reduced graphene oxide (rGO)-based nanohybrid working electrode on a flexible polyimide substrate. Here platinum and gold alloy nanoparticles were electrochemically coated over the microfabricated rGO surface containing glucose oxidase-chitosan. This sweat-based wearable sensor exhibits excellent amperometric response to glucose at a detection range of around 0–2.4 mM with a sensitivity of around 48 μA mM−1 cm2 and high linearity [128]. Cao et al. fabricated an electrochemical glucose sensor using a 3D paper microfluid that could prevent biomolecule immobilization from interfering with electrode modification. For glucose oxidase immobilization, an aldehyde-functionalized reference and counter-electrode hydrophilic zone were constructed, while rGO-TEPA/PB modified paper was used as a working electrode to quantify and detect H2O2 generated by coating Prussian blue over rGO-TEPA. Under ideal conditions, the proposed biosensor may be utilized to quantify glucose over a large linear range of 0.1 mM to 25 mM, with a detection limit of 25 µM [129]. Using a simple substrate-assisted electroless deposition (SAED) approach, a flexible non-enzymatic amperometric glucose sensor using laser-induced graphene decorated with a Cu nanoparticles (Cu NPs-LIG) composite has been successfully fabricated by Zhang et al. The Cu NPs-LIG sensor, as constructed, has a high glucose sensitivity of 495 µA mM−1 cm−2 with a low detection limit of 0.39 μM. Here electron transfers occur as a result of redox interactions between Cu NPs and glucose molecules; however, the Cu/graphene combination can speed up the electron transfer and improve the glucose-sensing efficiency. In this case, the continual addition of glucose leads to a stepwise current response. The response time of this sensor is quick, and it only takes 0.49 s to acquire a steady-state current. This composite’s large surface area aids in the fast diffusion of glucose molecules [130].
Similarly, another flexible non-enzymatic glucose sensor was fabricated by Lin et al. here, Cu-NPs were electrodeposited over a DVD-laser scribed graphene (LSG). This sensor shows more sensitivity than Zhang et al. reported with the Cu NPs-LIG sensor with a similar detection limit of around 0.35 μM. A significant limitation of reverse iontophoresis (RI)-based ISF extraction detection approaches is that ISF is significantly diluted prior to quantification due to its random extraction over a vast region of the skin, necessitating finger-stick validations. Lipani et al. recently devised a novel ISF extraction method that utilised an array of tiny pixels with graphene working electrodes decorated with Pt to produce a transdermal non-invasive glucose monitoring patch [131]. Park and colleagues created a wearable multifunctional sensor over a soft contact lens using hybrid architectures of 1D and 2D nanomaterials to provide wireless detection of intraocular pressure and tear glucose with improved resilience, flexibility, and conductivity. The main components were graphene and its hybrid with Ag nanowires, which were fabricated over a transparent and stretchy substrate to prevent any impact on the wearer’s eyesight ability. GOx was immobilized on the graphene channel via covalent bonding with the pyrene molecule via N-hydroxysuccinimide (NHS) chemistry. Both in vitro and in vivo real-time glucose detection were performed on a live bovine and a rabbit eyeball by tracking the increase in drain current as glucose concentrations increased [132]. A wearable diabetic patch made up of Au-doped graphene and a serpentine bilayer of Au mesh, as well as thermoresponsive microneedles, has demonstrated the feasibility of closed-loop wearable monitoring and feedback therapy on a single platform [133].
Even though these LIG-based electrochemical flexible/wearable sensors are popular still, their inherent electrical properties and low surface conductivity hinder their application in sensors [134]. To enhance the conductivity, Yoon et al. proposed a surface modification method using acetic acid. This simple treatment helps to increase the ratio of c–c bonds and thereby enhances the conductivity. Following this, a working electrode was fabricated by electroplating with Pt nanoparticles (PtNPs) and immobilising glucose oxidase using cyclic voltammetry. Finally, they used a real sample test to perform and evaluate the practical feasibility of an as-produced sweat glucose biosensor based on acetic acid-treated LIG electrodes [135].
Lei et al. developed a stretchy and wearable biosensor that uses MXene/Prussian blue (MXene/PB) to detect two metabolites in sweat, lactate, and glucose. The high conductivities and unusual structure of the MXene/PB composite significantly improved the electrochemical sensor’s performance over CNTs/PB and graphene/PB composites. One distinguishing aspect of this biosensor is that all active component of this sensor is self-contained and may be replaced at any time. The biosensor’s one-of-a-kind design ensures an adequate flow of O2 during the investigation, enhancing the long-term stability. Immobilizing lactate oxidase (LO) (for lactate) and GOx (for glucose) over porous and ultrathin CNTs/PB/Ti3C2Tx/CFMs electrodes separately resulted in the working electrode. The electrodes were then inserted into the sensor platform’s slot. The pH sensor is made up of Ag/AgCl as the reference electrode and PANI as the working electrode. This sensor, which had been modified with glucose oxidase, demonstrated a linear dynamic response to glucose in artificial sweat spanning the range of 10 × 10−6 M to 1.5 × 10−3 M, with a LOD of 0.33 × 10−6 M. Similarly, this sensor modified with lactate oxidase demonstrated a linear relationship with lactate concentration spanning 10 × 10−6 M to 22 × 10−3 M, with a LOD of 0.67 × 10−6 M. The biosensor’s practical application was proved by analyzing three components in sweat during strenuous cycling. This wristband biosensor based on MXene gathers enough sweat in less than 2 min and independently detects the concentrations of lactate, glucose, and pH value. The pH sensor is essential for monitoring local pH levels. The selectivity of a wearable sensing device is crucial due to the multiplicity of interfering components found in human perspiration. Interestingly, lactate, AA and UA were shown to have no effect on the glucose sensor. Similarly, AA, glucose and UA had no effect on the lactate sensor’s response. This study shows how to make a wearable and flexible MXene-based electrochemical sensor for detecting human biomarkers, which can be used for advanced personal care and disease detection [136]. M. et al. created a highly integrated sensing (HIS) paper that used foldable all paper substrates and Ti3C2/MB as the active material to make sweat analysis patches. This HIS sensor contained a signal processing platform for sweat analysis that could detect lactate and glucose in real-time. Because of a well-designed 3D diffusion path and the hydrophilic action of paper substrates, the functional sections of the HIS sensor were printed on paper substrates and folded into a three-dimensional structure. Human perspiration could be effectively gathered and promptly diffused from human-device interfaces in a vertical manner. According to the HIS paper, they created a dual-channel electrochemical sensor that could detect lactate and glucose simultaneously with a sensitivity of 0.49 µA of 2.4 nA µM−1 and mM−1, respectively. The HIS paper’s low cost and convenience make it a good candidate for wearable bioelectronic devices and non-invasive electrochemical sensors [137]. Myndrul et al. developed a stretchable and skin-attachable enzymatic glucose sensor using ZnO tetrapods (TPs) and MXene very recently. Because of the superior electrical conductivity of MXene and high surface area of ZnO TPs, this electrochemical glucose sensor shows enhanced sensitivity in sweat samples with LOD around 17 μM in a broad linear detection range (LDR = 0.05–0.7 mM) and enhanced mechanical stability (up to 30% stretching) [138]. The employment of MoS2 and flexible polymeric materials as electrodes at the same time generates a synergetic effect for fabricating a flexible biosensor platform. In a study by Yoon et al., they fabricated a MoS2/Au NPs glucose sensor over a polymer electrode, where they fabricated this sensor using spin coating. This sensor shows flexure extension values around 3.36 mm and 3.48 mm, which is larger than that of a rigid electrode [139]. NiSe2 is a prominent TMD because of its higher electrochemical activity due to the synergistic effects of the two metals and the active sites present in them. Aside from that, it has a higher efficiency in electron transport, making it a good option for use as a biosensor. Vishnu et al. built a disposable glucose sensor using a simple hydrothermal process for direct NiSe2 development on cellulose paper [140]. Liu and colleagues used a simple hydrothermal process and a post-annealing treatment to create a 2D copper cobaltite (CuCo2O4) nanosheet with flower-like morphology decorated on a flexible graphite paper. Spinel cobaltite’s have higher electrical conductivity and electrochemical properties than monometallic oxides because electrons with low activation energy flow across multiple transition metal cations, resulting in better storage efficiency. Graphene paper with extra active sites has great mechanical strength, cheap cost, durability, and is lightweight, making it perfect for flexible electrodes [141]. Wang et al. fabricated another non-enzymatic sensor based on a 3D array of NiCO LDH over a carbon cloth using a one-pot co-precipitation method. Because of the loosely-packed fiber structure of carbon cloth, its high flexibility and superior conductivity, which promotes glucose diffusion at the electrode interface, CC is chosen as the conductive substrate to deposit the NiCO LDH active materials. Furthermore, the large surface area of the NiCO LDH NS and interconnected structure allows easy access to glucose molecules [142]. Another flexible non-enzymatic sensor based on NiAl-LDH modified carbon cloth was also fabricated by Hai and Zou [143].
To realize the miniaturized wearable glucose sensors, the currently employed energy sources are not sufficient. This constraint becomes even more significant in the case of miniaturized implantable devices with extended operation times, as the main impediment to fabricating these devices has been the large sizes of co-implanted electronic circuit boards and the batteries that power them [11]. The development of enzymatic BFCs as attractive self-sustaining energy devices that eliminate the need for external power sources by harvesting bioenergy directly from metabolites present in body fluids has been an appealing way to meet this critically significant demand. Willner and Katz published the first study on self-powered glucose biosensors in 2001 [144]. Cho et al. reported a self-powered BFC-based non-invasive glucose biosensor, which was incorporated into a regular Band-Aid adhesive patch to detect glucose in human perspiration. This biosensor was built up of chitosan/GOx/graphene-based bioanode and activated carbon (AC)/Ni-based cathode and polystyrene sulfonate:poly(3,4-ethylenedioxythiophene) (PSS:PEDOT) as a conductive reservoir. This skin-connected technology can immediately absorb sweat by capillary pressure into a conductive reservoir, which converts chemical energy into electrical energy and detects glucose without the need for an external power source [145].

5. Photoelectrochemical Sensors

Photoelectrochemistry, which originated from electrochemistry, is a vibrant subject that investigates the action of light over photoactive materials, involving the conversion of sunlight into electricity and the interconversion of chemical and electric energy [146,147,148]. Following the transfer of photogenerated charge carriers to the electrodes, electrical signals are formed in the electrolyte, and the energy conversion is performed by the associated redox reactions. Photoelectro chemistry has been widely employed in pollution treatment and energy harvesting because of these properties. Notably, the coupling of the PEC process with sensing has ushered in a novel but promising technology, PEC sensing. The photoactive materials at the interface of the electrode that serve as the signal converter in PEC sensing, produce an electrical signal (current mode is usually employed in current PEC sensors) that is influenced by light-irradiated targets. A basic sensing system is made up of three essential components: an excitation light source unit, a detecting system containing a metal electrode having electrical and catalytic activity, an electrolyte, and a working electrode tuned with photoactive materials, and finally, a signal reading device. Biological materials (nucleic acids, antibodies, enzymes, and so on) for particular recognition are also required for PEC biosensors. The creation of an electrical signal in the PEC detecting system as a whole requires a number of chemical and physical processes. The usually accepted mechanism consists of four parallel processes: (1) absorption of photons, (2) separation of charges, (3) migration and recombination of these charges, and (4) charge usage (producing an electrical signal and engaging in a redox reaction at the solid-liquid interface). Unlike traditional electrochemical sensing, the excitation source in PEC sensing is light, and an electrical signal is employed for readout. Because of the different energy forms between the excitation source and the detection signal, PEC sensing has the potential to be more sensitive than classic chemiluminescent and electrochemical techniques. Furthermore, unlike electrochemical sensing, which frequently generates signals at certain potentials, PEC sensing, which exploits the strong redox characteristic of electron-hole pairs, reduces reliance on the applied potential. As a result, when compared to electrochemical sensing with identical parameters, PEC sensing frequently beats electrochemical sensing in terms of sensing performance. As per the fundamental principles of photoelectric conversion, recombination is another key factor impacting the conversion efficiency of photoactive materials. As a result, many techniques for suppressing charge carrier recombination have been developed. The most common and effective methods are designing and constructing semiconductor–semiconductor heterojunctions, multicomponent heterojunctions and semiconductor–carbon heterojunctions because photogenerated carrier separation and transfer are more successful at the interface region of different components [149,150]. Multiwalled and single-walled carbon nanotubes, graphene oxide, reduced graphene oxide (rGO), and graphene hydrogel are the most popular carbon materials employed to fabricate the semiconductor–carbon heterojunction.
Graphene has the highest specific surface area of any carbon material and excellent chemical stability and electron mobility. As a result, it is regarded as a superior medium for electron transmission. Indeed, improving sensing performance by combining semiconductors with graphene benefits from a number of factors. Furthermore, a Schottky barrier junction between the semiconductor and graphene can be formed, increasing charge separation. On the other hand, graphene can enhance stimulated electron transport while inhibiting surface recombination due to its greater electron mobility. For example, in order to reduce the recombination rate and improve the photocatalytic capabilities of tungsten oxide (WO3), Devadoss et al. used CVD followed by sputtering to introduce graphene beneath WO3, and photocatalytic activity was further enhanced by depositing AuNPs. AuNPs not only improve enzyme catalytic activity but also improve charge collection at WO3/Graphene surfaces by producing a Schottky junction. Here the WO3 nanoparticles, which are photoactive, often form electron-hole pairs when exposed to light. These photogenerated holes were then scavenged by a biological analyte (which acts as an electron donor), resulting in the oxidation of the biomolecules via an intermediary step involving the electron acceptor FAD/FADH2 shown in Figure 10. At NHE of around—0.4 V, glucose is oxidized to gluconic acid throughout this cycle. Concurrently, photogenerated electrons convert water to create H2 on the cathode (Pt). The sensing signal is the produced photocurrent travelling across the circuit. Thus, the efficiency of the PEC sensor is solely dependent on the effectiveness of photoactive materials in responding to biomolecules of interest and transduction efficiency [151]. CdS is another photocatalytic material that is widely investigated in PES systems. Similar toWO3, this semiconducting material has a low recombination rate. To increase photovoltaic conversion efficiency and charge separation, CdS is modified with graphene oxide by Zhang et al. using the facile hydrothermal method. This hybrid was demonstrated as a photoanode material in non-enzymatic glucose sensors. Here rGO helps to stabilize the quantum dots using the functional group present in the rGO, and this composite shows enhanced photoelectrochemical properties [152].
Later Ma et al. electrochemically coated CoOx over these graphene oxide–CdS nanocomposites to achieve both electrocatalytic and enhance the photoelectrocatalytic oxidation of glucose. These CoOx/graphene oxide-CdS composites demonstrated remarkable performance as an enzyme-free photoelectrochemical glucose sensor, with a broad linear concentration range (0.005–0.37 mM), high sensitivity (796.7 μA mM−1 cm−2), and an LOD of around 0.5 μM with strong selectivity [153], better than the graphene oxide-CdS nanohybrid prepared by Zhang et al. (LOD around 7 μmoL dm−3) [152]. Another ternary hybrid based on graphene-CdS nanocomposites was reported by Jafari et al. They modified graphene oxide–CdS nanocomposites using electropolymerised Nile blue (P-NB). The produced nanocomposites displayed good electrocatalytic activity toward the oxidation of NADH at a relatively low anodic potential due to the synergistic action of P-NB and the graphene oxide–CdS nanocomposite (ca. 50 mV). The P-NB utilized in this study not only acts as a redox mediator for the electrocatalytic oxidation of NADH but also as an electron donor for the neutralization of photogenerated holes in CdS QDs, resulting in a considerable increase in response photocurrent. By covalently attaching the GDH enzyme with this nanocomposite, a photoelectrochemical sensor for glucose sensor was fabricated [154]. Laser-induced graphene that possesses fast charge transfer and ion diffusion employed as electrode materials for electrochemical energy storage devices motivated Li et al. to fabricate a photoelectrochemical glucose sensing. They created a Ni and CdS hybrid in LIG via a one-step laser-induced solid-phase transition. This PEC sensor exhibits outstanding photoelectric catalytic activity towards glucose, which is owing not only to the synergistic impact of Ni and CdS but also to the superior conductivity of the LIG’s 3D macroporous architecture. The electrocatalytic performance towards glucose and oxidation of glucose can be summarized by the equation given below:
NiO(OH) + glucose → Ni(OH)2 + glucolactone
NiS(OH) + glucose → NiS + glucolactone
This laser-induced hybrid photoelectrode shows an LOD of around 0.4 μM with good selectivity, stability, and reproducibility [155]. Cuprous oxide (Cu2O) is a low-bandgap semiconductor with a tolerable bandgap value (2.0–2.2 eV) that opens up new avenues for increased solar light harvesting. It is also a low-cost, non-toxic semiconductor with improved stability. By combining it with n-type TiO2, this p-type semiconductor has been investigated for water photocatalysis. Using these p–n heterojunctions and the enhanced electron transport capabilities of rGO, Bekir et al. created a PEC glucose sensor based on Cu2O/rGO-coated TiO2 nanotubes (NTs) arranged titanium foil. The creation of p-n heterojunctions between TiO2 NTs and Cu2O NPs, as well as the good conductive rGO, can both improve charge separation efficiency and permit electron transmission in this case, resulting in significantly higher photocurrent [156]. Recently, a visible-light-driven PEC glucose biosensor was proposed by Zhao et al. by utilizing rGO/TiO2 nanotubes with GOD as the identification element. This constructed PEC sensor shows excellent sensitivity with an LOD of around 5 μM [157].
Aside from semiconducting materials, layered BiOXs (X = Br, Cl, and I) have garnered a lot of attention as a photocatalytic material. Among these, BiOCl is regarded as one of the most promising. BiOCl has a layered structure that is made up of [Bi2O2] 2p slabs separated by double slabs of chain atoms. BiOCl’s strong photocatalytic performance is attributable to its layered structure’s effective charge carrier separation capabilities, as well as the induced dipole formed inside its molecular structure. Combining the individual advantages of these BiOCl and graphene, Gopalan et al. used a hydrothermal method for the synthesis of 2D bismuth oxychloride–graphene nanohybrid sheets (BiOCl-G NHS). This nanohybrid formation takes place in three stages: (i) electrostatic adsorption of bismuth precursor onto carboxylic acid-functionalized graphene nanosheets, (ii) formation of BiOCl nanocrystals as nuclei on the surface of graphene nanosheets, and (iii) growth of BiOCl nanosheets on the surface of functionalized graphene sheets, resulting in stacked stacks of BiOCl and graphene nanosheets. The cyclic voltammetry and differential pulse voltammetry (DPV) studies demonstrated this nanohybrid is able to generate photocurrent for glucose when it is illuminated with a source of light having a wavelength of 365 nm. The photocurrents produced in the presence of glucose displayed a linear relationship on glucose concentration in the range of 0.5 to 10 mM at a bias potential of +0.50 V, with a detection limit of 0.22 mM [158]. Similar to BiOCl, another ternary bismuth semiconductor compound, BiOBr, was reported to fabricate a PEC sensor. A BiOBr-TiO2 nanotube array (TNTA) composite electrode was prepared using vacuum impregnation and the chemical precipitation method. Here the matched levels of BiOBr and TNTA enhance the separation of holes and photoelectrons effectively and significantly enhance the photoelectrochemical performance with LOD around 10 nM [159].
Graphitic carbon nitride (g-C3N4) is a relatively new form of carbon-based material that has just recently been synthesized. Due to its simple synthesis method, strong photo-stability, low cost, and adequate band gap, graphitic carbon nitride (g-C3N4) has attracted a lot of attention in PEC sensing. ZnIn2S4, a ternary chalcogenide semiconductor, has been widely used in photodegradation pollutants and photocatalytic hydrogen evolution. It should be noted that gC3N4 and ZnIn2S4 can create a heterojunction with an overlapping band gap structure, effectively separating photogenerated electrons and holes and increasing photoelectric conversion efficiency. Furthermore, g-C3N4 and ZnIn2S4 exhibit superior PEC response and photo-stability when exposed to visible light. As a result, g-C3N4/ZnIn2S4 composites for PEC sensing were fabricated by Zhang et al. Using Au nanoparticles; they co-assembled horseradish peroxidase (HRP) and glucose oxidase (GOx) over this hybrid. With the addition of glucose and 4-chloro-1-naphthol (4-CN), this electrode can effectively catalyze the oxidation of glucose into H2O2 and gluconic acid. Here HCP can speed up the oxidation of 4-CN and thereby reduce the precipitation reaction time. The deposit on the electrode surface has the potential to significantly diminish the photocurrent signal. Because of the signal magnification process, this bi-enzyme glucose sensor has a “signal-off” sensing mode for glucose detection, with a dynamic detection range of 1–10,000 μM and a low LOD of around 0.28 μM. In addition, the developed PEC bi-enzyme glucose sensor has enormous promise for quantitative glucose detection in practical applications [160]. In addition, Fe3O4/g-C3N4 [161] and WO3/g-C3N4 [162] heterojunction structures were also employed in the PEC sensing platform for glucose detection. Çakıroğlu et al. utilized g-C3N4, MnO2, and Au nanoparticles to modify the fundamental semiconducting properties of TiO2. p-n heterojunctions were formed between the interface of g-C3N4/TiO2 and MnO2, which inhibit the recombination of excited electrons and also facilitate charge transport. Here the major role of g-C3N4 is to amplify the photocurrent by enhancing photogenerated electron-hole pairs and decreasing recombination [163]. When exposed to visible light, electron-hole pairs develop on MnO2, and the photogenerated electrons are quickly pushed from the conduction band (CB) of MnO2 to the CB of g-C3N4 [163].
MXene’s huge surface area and exposed metal sites enable more active sites for semiconductor growth. The very first MXene-based PEC sensor for glucose detection used a Ti3C2 MXene/Cu2O heterostructure. Cu2O, as a p-type semiconductor, had a low cathode photocurrent when exposed to visible light. Anode photocurrent was observed in pure Ti3C2. The cathode photocurrent increased dramatically after the combination because of the multilayered MXene’s superior high carrier mobility and metallic conductivity, which boosted hole transport from copper oxide’s valence band to MXene and accelerated the separation of photogenerated electron-hole pairs. However, when glucose was added to the buffer, the photocurrent decreased due to the partial reduction of the species being consumed. The Ti3C2 MXene/Cu2O composite demonstrated good PEC performance and sensitive photoelectric response to glucose with a LOD of 0.17 nM. [164]. Chen, G. et al. constructed a Z-scheme TiO2/Ti3C2/Cu2O heterostructure and used it as a photocathode in a self-powered PEC glucose sensor. An artificial indirect “Z-scheme” system was created by connecting two semiconductors (two separate photosystems (PS)) using electron mediators. According to their findings, the TiO2 nanoparticles were generated in situ on the MXene nanosheets by ethanol heat treatment of the MXene, and the Z-scheme of this nanocomposite heterojunction was prepared by hybridizing with the narrow bandgap semiconductor Cu2O. The Z-scheme of this nanocomposite heterojunction outperformed TiO2/Ti3C2 and Cu2O in terms of PEC performance, resulting in a TiO2/Ti3C2/Cu2O PEC sensor for glucose with a surprising low LOD of 33.75 nM and a broad range of 100 nM to 10 M [165].
TMDs and their composites are widely used in PEC glucose-sensing due to their advantages, including high charge mobility, narrow band gap, large surface area, and visible-light activity. MoS2 is the most popular 2D TMD in the PEC sector due to its ease of manufacture, low price, facile exfoliation, abundant raw ingredients, small band gap (about 1.8 eV), and visible light activity. Bulk MoS2 nanomaterials, on the other hand, are rarely used in PEC tests due to the considerable recombination of photogenerated electron-hole pairs. Several methods for modifying MoS2 in the PEC field have been developed to address this shortcoming, such as (1) exfoliation of bulk MoS2 to prepare single- or few-layered MoS2 nanosheets, (2) combination with other semiconductors to form heterojunction structures, and (3) modification with metallic nanoparticles. Using a simple thermolytical approach and a C3N4 sacrificial template, a 3D porous network made of ultrathin MoS2 nanosheets with good PEC performance was generated here. The following is the reaction procedure. The FAD redox group in GOD adsorbed on the MoS2 nanosheet surface oxidizes the glucose in the solution and transforms it to the reduced FADH2 group in this case. When exposed to visible light, the photo holes generated on the surface and inside the MoS2 films may easily donate holes to FADH2 and refill the enzyme redox centre FAD, while O2 molecules are reduced to H2O2. Electrons are transferred swiftly to the ITO substrate while being subjected to increased electrical power during this technique [166,167].
Liu’s group investigated the effect of TiO2 nanorods on the photocatalytic activity of MoS2 nanosheets and discovered that combining them with semiconductors helps to improve the photocatalytic activity of MoS2. The photoactivity of the MoS2-TiO2 composite was found to be significantly larger than that of pristine TiO2 and MoS2 nanorods. In fact, the photocurrent for this nanohybrid was 4.8 times that of pristine MoS2. It was explained as the matching band gap of MoS2 and TiO2, which promoted photogenerated electron transport to the electrode surface while inhibiting photogenerated electron-hole pair recombination. It was used to build a PEC biosensor by immobilising glucose oxidase (GOx) on a MoS2-TiO2/ITO surface. The photocurrent of this sensor grew linearly as the concentration of glucose increased from 0 to 8 Mm [168,169].
The other class of two-dimensional materials that were recently explored for glucose sensing were transition metal nitride materials. They are of interest due to their great electrical conductivity, biocompatibility, thermal and chemical stability, and electronic structural similarities to valuable metals. Some of the other reported electrochemical glucose sensors based on transition metal nitride and other 2D materials are given in Table 4. [170,171,172,173,174,175,176,177]

6. Theoretical Perspective for Glucose Sensing by 2D Materials

Experimentally, the dearth of speedy and scrimping synthesis procedures and unavailability of pre-testing facilities to confirm its proper functioning makes it challenging to develop and devise glucose sensors based on novel 2D materials. At present, various synthesis approaches such as CVD, sputtering, hydrothermal, electrodeposition methods etc., are being employed to fabricate fine-quality 2D materials for biomolecule detection. Nevertheless, most of these methods are costlier and cannot be used for mass production. Indeed, such constraints can be overwhelmed if the sensor materials are designed, modelled, and simulated using ab-initio theoretical techniques [178]. The ab-initio, also known as the first principal methods, are quantum mechanical methods that compute the electronic structure of the system by solving the Schrodinger equations numerically. Density Functional Theory (DFT) [179], Monte Carlo [180], Hartree, Fock [181] etc., are the standard quantum methods. DFT is the most effective and widely used technique, which proffers quick and precise determination of ground-state electronic properties of various materials. Unlike the typical laboratory methods, theoretical simulations have an explicit gain in both time and endeavor. The computational simulations utilizing DFT could produce atomic level clarifications and comprehensions of the laboratory results or even act as a predictive aid for designing novel sensors.
In the glucose-sensing process, the interaction between the glucose molecule and the 2D substrate materials is purely weak. From the atomic level perspective, in addition to understanding the adsorption mechanism, the fundamental investigation of the interaction between the glucose molecules and the surfaces of the 2D materials also provides a theoretical insight into hindering the glucose adsorption, which has essential practical relevance and research value. After the adsorption process, the surface of the 2D substrate material varies significantly. It is challenging to control the formation of point defects and clarify the grassroots mechanism experimentally. Hence, the theoretical simulations and computational calculations provide intimate knowledge of glucose sensing that cannot be traced barely from the experiments, namely optimum adsorption configurations, active sites, binding energy, variations in the electronic and optical properties of the host material after adsorption, bonding process, charge transfer (both amount and direction) etc. As the theoretical investigations on glucose-sensing by 2D materials are very scarce, these prolific details will encourage the prediction of the glucose-sensing properties of novel 2D materials, in turn, unwrap new tracks for the experimentalists to develop functionalized sensor materials with high sensitivity and selectivity.

6.1. Theoretical Modelling

Modelling the glucose adsorption system is of prime importance in examining the glucose adsorption mechanism. Since the molecule determination is a surface phenomenon, a 2D model of the substrate material satisfying the planar periodic boundary condition must be considered. The lattice parameters, bond distances and angles of the modelled structure should be consistent with their actual values. The interlayer interactions can be eluded by inserting enough vacuum spacing vertically above the 2D substrate materials. The next step is to optimise the modelled 2D material to a minimal energy configuration. This can be achieved by choosing suitable energy thresholds and k-point mesh samples for Brillouin zone integration. For 2D materials, the K-point mesh should have a unit value in the Z direction, such as 2 × 2 × 1, 5 × 5 × 1, 11 × 11 × 1, etc. Figure 11a,b shows the supercell of the 2D novel graphyne and Penta–Octa–Penta (POP)-graphene. [182,183].
Similarly, the structure of the glucose molecule should also need to be modelled, and each atom in the glucose should be optimized to its ground state configuration with minimum residual forces between them. Suppose the software used for implementing the DFT calculation is periodic. In that case, the glucose molecule should be enclosed in a large unit cell with proper vacuum spacing before the optimizations to evade the effect of periodic interactions (see Figure 11c). The glucose molecule is then placed over the surface of the 2D materials at different positions in different orientations. The final structures are relaxed to minimum energy state configurations. The adsorption energy calculation decides the most stable or favorable adsorption configuration.

6.2. DFT Simulation of Glucose Sensing by 2D Materials

The literature presents various explications of the glucose-sensing behavior of 2D materials. From the theoretical stance, the glucose molecules are physisorbed on the surface of the 2D nanomaterials. In this case, the glucose interacts with a long bonding distance, weak binding energy, and negligible charge transfer. The molecule adsorption usually induces modifications in the optical and electronic properties of the host materials. Whereas during the physisorption, the electronic and optical properties remain nearly constant. This is because of the weak hydrogen bonding or van Der Waals interactions between the 2D materials and the glucose molecules.
Generally, physisorption occurs when the pristine 2D materials are adopted as the substrate material for molecule adsorption. The introduction of defects, doping with foreign materials, heterostructure formation, etc., create more active sites on the surface than its perfect structure [185,186,187,188]. In addition, the tuning can cause a sudden change in the electronic states of the 2D material, which promotes the sensing mechanism [189]. For example, the pure MoS2 monolayer is a semiconductor with a direct bandgap of about 1.86eV. The substitutional doping of Nb into MoS2 lowered the Fermi level down to the valence band, implying a substantial improvement in conductivity, as shown in Figure 12c–e [189]. In effect, the glucose molecule can form a strong chemical or covalent bond with the surface of the defected or functionalized 2D materials. Such a type of adsorption process is called chemisorption. Here, the glucose molecules bind to the surface with strong adsorption energy and large charge transfer. As a result, the electronic properties of the host 2D material drastically vary immediately after the adsorption process.
The glucose sensing performance of the 2D materials can be predicted theoretically by analyzing the following parameters:
i.
Adsorption sites and configuration
Depending on the geometry, 2D materials have different adsorption sites with which the glucose molecules can interact. For example, various adsorption sites of graphyne and POP-graphene are indicated in Figure 11a,b. The glucose molecule can adsorb at various adsorption sites in different configurations. Different configurations imply that certain atoms of the glucose molecule, such as O, C, and H, are placed adjacent to various adsorption sites. Secondly, the glucose molecule can be oriented both parallelly or perpendicularly to the surface of the 2D material. Before geometrical optimization, a reasonable distance between the glucose molecules and the 2D surface should be set for each adsorption configuration. The adsorption energy and bond distances between the 2D material and the glucose molecule decide the most favorable adsorption sites or stable configurations. The more negative adsorption energy, the higher the configuration’s stability will be. Therefore, the structure with the lowest adsorption energy is identified as the optimal configuration. Figure 13a,c and Figure 14a,b illustrate the geometries of glucose adsorption in a pristine 2D C2N system and metal-doped graphene, respectively.
ii.
Adsorption energy
The adsorption energy defines the strength of the interaction of the glucose molecule with the 2D substrate materials and can be determined by the following formula:
E Ad = E 2 D   substrate   +   glucose E 2 D   substrate + E glucose
Each term in Equation (17) represents the energy of the corresponding optimized structures before and after the adsorption. The glucose molecule adsorbs spontaneously on the 2D material surface if the predicted adsorption energy values become negative. On the other hand, the positive result indicates that the adsorption process is endothermic, implying that the 2D material under investigation lacks the glucose-adsorption property. The adsorption energy can vary with the adsorption sites. The adsorption strength is proportional to the magnitude of the adsorption energy value. However, a significantly higher value may impact the sensor’s utility and reusability. If the adsorption energy of glucose on pristine 2D materials is poor, numerous tuning techniques can be employed to improve the adsorption ability. The standard LDA and GGA exchange functionals exclude long-range interactions when computing the adsorption energy [190,191]. Therefore, specific dispersion corrections, such as Grimme’s DFT-D2, D3 etc., are included during the simulation [192,193]. The adsorption energies of glucose with various 2D materials are listed in Table 5.
iii.
Variations in the electronic properties
The electronic properties of the pristine 2D materials alter soon after the adsorption of glucose molecules. The electronic properties will change significantly if the glucose molecules are chemisorbed on the 2D substrate. These modifications can be explained by comparing the electronic band structure and the total density of states (DOS) plots of the 2D material before and after the glucose adsorption. Glucose adsorption may induce new energy states near the Fermi level. If the 2D material used as the substrate is semiconducting, the adsorption can cause a reduction in the bandgap. For example, compared to the DOS of the pristine C2N monolayer in Figure 13b, the energy states near the Fermi level are enhanced after the glucose adsorption [195]. However, the bandgap remained the same even after the adsorption (Figure 13d).
iv.
Charge transfer
The surface of the 2D material behaves either as charge donors or acceptors. When the glucose molecule interacts with the 2D material, they are adsorbed on the surface by transferring charges. Superior charge transfer indicates the high sensitivity of the 2D material. The DFT analysis aids in the interpretation of the binding mechanism by examining the charge transfer between the glucose molecule and the 2D material. The HOMO and LUMO plot of D-glucose in Figure 11d,e reveal the electron-accepting nature of D-glucose [184]. The quantitative measurement of the charge transfer can be determined by different tools such as Bader charge analysis, Mulliken charge analysis [196], Lowdin charge analysis [197], etc., and is expressed by equation 18.
q = q 2 D   material + glucose q 2 D   material q glucose
ρ = ρ 2 D   material + glucose ρ 2 D   material ρ ( glucose )
If ∆q is positive in equation 2, charge transfer occurs from the glucose molecule to the 2D material and vice versa. The amount of charge transferred between the glucose and various 2D material systems is listed in Table 5. The charge gained by the glucose molecule is due to its electron-withdrawing property, as indicated by the negative q in the table. Moreover, the charge flowing direction during the adsorption process can be assessed from the charge density difference (Cdd) plots. Equation (3) gives the mathematical description of the Cdd, where ρ is the density. The Cdd of the D-glucose adsorption on N-, O-, and Cl-doped HGr is shown in Figure 14d–f. The blue color enveloping the glucose molecule at the dopant site indicates that the molecule receives electrons from the doped hGr. Furthermore, the Cl-doped HGr interacts strongly with the glucose molecule leading to a large charge transfer of around −0.20|e| from the system to the D-glucose molecule [184].
v.
Orbital interactions
DFT investigations can be used to study the orbital level binding mechanism. As previously stated, charge transfer is a key factor in glucose sensing; the qualitative concept of charge transfer across atomic orbitals can be explored by analyzing the partial density of states (PDOS). The 2p orbital of the glucose molecule interacts with the metal d orbitals via p–d hybridization if the 2D material under discussion is TMDs, MXenes, or any other transition metal-based structure. The strong orbital hybridization between the C 2p and X 3p orbitals near the Fermi level in X-doped HGr (X = Cl, N and O) is the reason for the increased glucose adsorption (see the PDOS in Figure 15a–c). The Cl 3p orbitals have enhanced peaks and large overlap with the O 2p orbitals of glucose molecule near the Fermi level than the N and O 3p orbitals, explaining its large adsorption activity [182].

6.3. Theoretical Analysis of the Practical Feasibility of the Glucose Sensor

For the convenient operation of the proposed glucose sensor, the material should be stable, sensitive, selective, and reusable. This section details how to theoretically assess the potential of 2D material-based glucose sensors.
(i)
Thermal stability
Since the DFT calculations are performed at 0 K, it is essential to check the room temperature stability of the system comprising the sensing device. Ab initio Molecular Dynamic simulations (MD) at 300 K are undertaken to ensure the same [198,199,200]. In addition, the reuse of the sensor device requires that glucose molecules be readily absorbed from its surface, which often occurs at a temperature higher than room temperature. Yong et al. performed the first principal MD calculations at different temperatures to estimate the thermodynamic stability of the B6N6H6 monolayer [201]. They found that at 1000 K, the total energy of the monolayer shows a stable oscillation around a mean value (see Figure 16), but at 1200 K, the system starts to decompose and completely decomposes when the temperature reaches 1500 K. The MD snapshots of the geometric structure at these three temperatures are shown in Figure 15d–f. The excellent thermodynamic stability below 1000 K allows the B6N6H6 structures to act as potential room temperature sensors.
(ii)
Sensitivity of glucose adsorption analysis based on the band structure
As aforementioned, glucose adsorption alters the inherent bandgap inherent electrical bandgap (EG) of the semiconducting 2D material. The study of this variation of bandgap helps in analysing the changes in the electrical conductivity ( σ ) of the glucose adsorption system using the following equation:
σ exp E G 2 KT
Here K and T, respectively, denote the Boltzmann constant and operating temperature of the glucose sensor [200]. The higher the value of σ , the greater the sensitivity of the system. A significant reduction in the bandgap improves the conductivity of the material, which can be acquired as an electrical impulse via the external circuits.
The bandgap variation after the glucose adsorption can sometimes be very low; therefore, it should be computed precisely. The correctness of the DFT results is highly influenced by the exchange correlational functional chosen. The widely used LDA and GGA are insufficient to define the localised electronic states in TMs and other elements with higher atomic numbers. We need to incorporate some corrections to DFT to address the shortcomings of standard GGA and LDA methods such as DFT + U [202], hybrid functionals such as the HSE06, B3YLP, and GW approximations [203,204,205], etc.
(iii)
Recovery time
The time taken for the desorption of glucose molecules from the surface of the 2D material substrate is called the recovery time ( τ ); or in other words, it defines the time taken by the sensor material to retain its original state after the desorption process. A quick recovery of the sensor material is vital for better functionality; otherwise, the sensor will be termed disposable. It has a strong correlation with the temperature and the potential desorption barrier. The adsorbed system is frequently heated to high temperatures to accomplish recovery. A theoretical prediction of the recovery time can be made based on transition-state theory [206]:
τ = A 1   exp   ( Eb KT )
where A, Eb, K, and T illustrate the attempt frequency, desorption barrier, Boltzman constant and temperature. Since the desorption is assumed as the opposite process of adsorption, the Eb can be made equal to EAd. If the adsorption of the glucose molecule on the 2D surface is stronger, the more extensive will be the recovery time. The reusability of the glucose sensor may be hindered by prolonged recovery periods.

7. Future Perspectives

Two-dimensional materials emerge as the advanced and high-performance electrode for glucose sensing application, but several challenges still remain that need to be addressed in future.
(1)
Electron transfer between the current collector and the catalytic active sites is governed by the electrical conductivity of the catalyst. The use of conductive materials can promote interfacial contact by lowering interfacial charge resistance and overpotential. Because of their great charge mobility and strong electrical conductivity, 2D materials can be employed to accomplish this. High stability, little residual current, high capacitance, vast flexibility, and substantial robustness characterize 2D-material-based electrochemical glucose sensors.
(2)
Another important quality of a glucose sensor is judged by the stability, which is reliant on a number of parameters. Graphene and BN have the highest chemical and structural stability of any 2D material, whereas other 2D layered materials can react with water and oxygen. Black phosphorus (BP) has a high reaction rate when compared to other stacked 2D materials. As a result, publications based on BP for glucose detection are relatively rare. Gu et al. recently described an enzymatic biofuel cell based on BP and explored the BP mimicking enzyme properties [207]. Although certain physical and chemical modifications can be made to improve the stability of 2D materials, they nonetheless present numerous barriers and challenges. One major challenge is that existing solutions for enhancing the stability of 2D-material-based devices are still in the early stages of commercialization and manufacture [208].
(3)
Since glucose monitoring is connected with direct contact with real samples such as blood, urine, serum, and so on, the use of a biocompatible catalyst that inhibits biological reactions and cell damage is critical. Based on synthesis technique, structure, layer number and elemental composition, the compatibility of 2D materials may vary. For example, Tao et al. studied the biocompatibility of TMDs and found out that WSe2 shows high toxicity compared to WS2 and MoS2, which can be attributed to the toxic effects of Se. Similar studies were used to examine the role of the chalcogen atom in biocompatibility and found that VS2 is less toxic compared to VSe2 and VTe2. The biocompatibility of graphene and analogues has already been well-studied and find their place in glucose sensors. Aside from graphene and TMDs, the biocompatibility of additional 2D family members such as h-BN, MXenes, and black phosphorus is understudied. MXenes are quickly finding a place in healthcare applications; certain studies proved their biocompatibility. However, the biocompatibility of these electrochemical glucose sensors based on 2D materials should be further explored because in vivo devices must meet more stringent biocompatibility requirements than in vitro diagnostics.
(4)
The accurate detection of glucose in complicated physiological media (e.g., blood, interstitial fluid, or saliva) is critical for commercially viable biosensors, including those based on 2D materials. Because of the presence of non-target species in these media, there is an additional difficulty known as biofouling. When exposed to a biofluid, undesired molecules and other species (e.g., proteins, bacteria, platelets, etc.) might build on the sensor’s surface, interfering with and deteriorating its capabilities. Surprisingly, 2D materials can have antifouling capabilities by repelling hydrophilic interferents and so maintaining stability. In general, the electrooxidation of poisoning intermediates results in a large reduction in active surface area, which reduces sensitivity. Because of the huge surface area and edge-plane defect sites in 2D materials, this constraint can be efficiently eliminated. This needs a thorough assessment of the performance of these 2D-material-based biosensors in real-world matrices (beyond PBS or other buffers).
(5)
The use of 2D materials in wearable devices, particularly sweat monitoring platforms, has recently gained popularity. Such 2D-based platforms have demonstrated promising performance in sensing and self-powered energy harvesting devices. The use of cost-effective fabrication techniques (e.g., screen printing), improved reproducibility and accuracy of the measured glucose signal, system integration in body-compliant flexible wearable platforms, validating sensors with blood glucose levels, and the improved (stimulated) collection of non-invasive biofluids have all resulted in significant advances. Despite these exciting recent improvements, properly deploying non-invasive glucose monitoring systems requires additional tuning followed by large-scale validation studies to critically analyse the reliability and accuracy of these devices. These lengthy investigations will establish the potential of such non-invasive approaches to dependably monitor glucose and the association of these non-invasive generated data with gold-standard blood glucose readings. Other issues that need to be addressed in epidermal non-invasive sensing include irregular biofluid extraction, surface contamination, and the effect of some physiological factors (e.g., temperature, pH) on measurement accuracy. Despite these obstacles, the next decade will see a surge in activity in the development of non-invasive glucose monitoring systems based on 2D materials.
(6)
The incorporation of 2D materials into various semiconductor-based photoactive materials might increase their PEC sensing performance. Despite significant advances, there is still a long way to go in the research and implementation of 2D photoactive materials. Two-dimensional materials with a suitable band gap and strong optical absorption such as BP and 2D perovskites have not yet been reported. The lack of a suitable synthetic method also challenges the use of 2D materials in PEC sensing. However, the flexibility of ultrathin nanosheets, along with their strong photoelectric characteristics, makes 2D materials ideal for developing wearable PEC sensors for non-invasive health detection.
The future perspectives of these 2D-material-based glucose sensors are provided in Figure 16.

8. Conclusions

In this review paper, we examined recent advancements in enzymatic and non-enzymatic glucose sensor applications. Both types of glucose sensors are already taking advantage of the fascinating features of 2D materials on a regular basis. Considering their high sensitivity and specificity, enzyme sensors are sensitive to a narrow range of temperatures, pH levels, and humidity levels, which has sparked interest in non-enzymatic electrochemical glucose sensors. The activity of the majority of non-enzymatic glucose sensors is a direct function of the electrode material on which the glucose is oxidized. According to the study, the majority of glucose sensors’ catalytic activity is mostly dependent on the electrode material, which serves as the site of glucose oxidation. Earlier studies concentrated on elements such as Ag, Au, Pd, Cu, Pt, and Ni, as well as metal oxides such as iron oxide, silver oxide, copper oxide, and nickel oxide, among others. Researchers were particularly interested in metal/metal oxide compounds/alloys and metal oxide composites during the time. Polymers have recently generated interest due to properties such as biocompatibility and excellent selectivity.

Author Contributions

Writing—Original draft preparation, S.R., S.L. and S.S.; Writing—Review and editing, S.R.; supervision N.K., B.C. and C.S.R. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

The authors gratefully acknowledge financial support from the Department of Science and Technology (DST)-SERB Early Career Research project (Grant No. ECR/2017/001850), DST-Nanomission (DST/NM/NT/2019/205(G)), and Karnataka Science and Technology Promotion Society (KSTePS/VGST-RGS-F/2018-19/GRD NO. 829/315). KNK acknowledges SERB: CRG (Grant No. CRG/2021/001506). BC thanks Nandini Garg, T. Sakuntala, S.M. Yusuf and A.K. Mohanty for the support.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Reddy, V.S.; Agarwal, B.; Ye, Z.; Zhang, C.; Roy, K.; Chinnappan, A.; Narayan, R.J.; Ramakrishna, S.; Ghosh, R. Recent Advancement in Biofluid-Based Glucose Sensors Using Invasive, Minimally Invasive, and Non-Invasive Technologies: A Review. Nanomaterials 2022, 12, 1082. [Google Scholar] [CrossRef] [PubMed]
  2. Sharma, P. Over 95,000 Children Suffer from Type-1 Diabetes: ICMR. Available online: https://www.livemint.com/science/health/over-95-000-children-suffer-from-type-1-diabetes-icmr-11654542643638.html (accessed on 16 June 2022).
  3. Witkowska Nery, E.; Kundys, M.; Jeleń, P.S.; Jönsson-Niedziółka, M. Electrochemical Glucose Sensing: Is There Still Room for Improvement? Anal. Chem. 2016, 88, 11271–11282. [Google Scholar] [CrossRef] [PubMed]
  4. Nava, C.; Modiano Hedenmalm, A.; Borys, F.; Hooft, L.; Bruschettini, M.; Jenniskens, K. Accuracy of Continuous Glucose Monitoring in Preterm Infants: A Systematic Review and Meta-Analysis. BMJ Open 2020, 10, e045335. [Google Scholar] [CrossRef] [PubMed]
  5. Reghunath, R.; devi, K.; Singh, K.K. Recent Advances in Graphene Based Electrochemical Glucose Sensor. Nano Struct. Nano Objects 2021, 26, 100750. [Google Scholar] [CrossRef]
  6. Danielsson, B.; Hedberg, U.; Rank, M.; Xie, B. Recent Investigations on Calorimetric Biosensors. Sens. Actuators B Chem. 1992, 6, 138–142. [Google Scholar] [CrossRef]
  7. Chen, Y.-T.; Kolhatkar, A.G.; Zenasni, O.; Xu, S.; Lee, T.R. Biosensing Using Magnetic Particle Detection Techniques. Sensors 2017, 17, 2300. [Google Scholar] [CrossRef]
  8. Damborský, P.; Švitel, J.; Katrlík, J. Optical Biosensors. Essays Biochem. 2016, 60, 91–100. [Google Scholar] [CrossRef] [Green Version]
  9. Pohanka, M. Overview of Piezoelectric Biosensors, Immunosensors and DNA Sensors and Their Applications. Materials 2018, 11, 448. [Google Scholar] [CrossRef] [Green Version]
  10. Zaidi, S.A.; Shin, J.H. Recent Developments in Nanostructure Based Electrochemical Glucose Sensors. Talanta 2016, 149, 30–42. [Google Scholar] [CrossRef]
  11. Teymourian, H.; Barfidokht, A.; Wang, J. Electrochemical Glucose Sensors in Diabetes Management: An Updated Review (2010–2020). Chem. Soc. Rev. 2020, 49, 7671–7709. [Google Scholar] [CrossRef]
  12. Castle, J.R.; Kenneth Ward, W. Amperometric Glucose Sensors: Sources of Error and Potential Benefit of Redundancy. J. Diabetes Sci. Technol. 2010, 4, 221–225. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  13. Soldatkin, A.P.; El’skaya, A.V.; Shul’ga, A.A.; Jdanova, A.S.; Dzyadevich, S.V.; Jaffrezic-Renault, N.; Martelet, C.; Clechet, P. Glucose Sensitive Conductometric Biosensor with Additional Nafion Membrane: Reduction of Influence of Buffer Capacity on the Sensor Response and Extension of Its Dynamic Range. Anal. Chim. Acta 1994, 288, 197–203. [Google Scholar] [CrossRef]
  14. Walker, N.L.; Roshkolaeva, A.B.; Chapoval, A.I.; Dick, J.E. Recent Advances in Potentiometric Biosensing. Curr. Opin. Electrochem. 2021, 28, 100735. [Google Scholar] [CrossRef]
  15. Khan, R.; Radoi, A.; Rashid, S.; Hayat, A.; Vasilescu, A.; Andreescu, S. Two-Dimensional Nanostructures for Electrochemical Biosensor. Sensors 2021, 21, 3369. [Google Scholar] [CrossRef] [PubMed]
  16. Heller, A.; Feldman, B. Electrochemical Glucose Sensors and Their Applications in Diabetes Management. Chem. Rev. 2008, 108, 2482–2505. [Google Scholar] [CrossRef] [Green Version]
  17. Wang, J. Glucose Biosensors: 40 Years of Advances and Challenges. Electroanalysis 2001, 13, 983–988. [Google Scholar] [CrossRef]
  18. Zhang, C.; Zhang, Z.; Yang, Q.; Chen, W. Graphene-Based Electrochemical Glucose Sensors: Fabrication and Sensing Properties. Electroanalysis 2018, 30, 2504–2524. [Google Scholar] [CrossRef]
  19. Wu, I.-C.; Weng, Y.-H.; Lu, M.-Y.; Jen, C.-P.; Fedorov, V.E.; Chen, W.C.; Wu, M.T.; Kuo, C.-T.; Wang, H.-C. Nano-Structure ZnO/Cu2O Photoelectrochemical and Self-Powered Biosensor for Esophageal Cancer Cell Detection. Opt. Express OE 2017, 25, 7689–7706. [Google Scholar] [CrossRef]
  20. Shu, J.; Tang, D. Recent Advances in Photoelectrochemical Sensing: From Engineered Photoactive Materials to Sensing Devices and Detection Modes. Anal. Chem. 2020, 92, 363–377. [Google Scholar] [CrossRef]
  21. Graphene Market & 2D Materials Assessment 2021–2031. 2020. Available online: https://www.idtechex.com/zh/research-report/graphene-market-and-2d-materials-assessment-2021-2031/789 (accessed on 11 May 2022).
  22. Team Develops Novel Chemical Glucose Sensing Method Based on Boronic Acids and Graphene Foam. Available online: https://phys.org/news/2022-02-team-chemical-glucose-method-based.html (accessed on 16 June 2022).
  23. Hassan, M.H.; Vyas, C.; Grieve, B.; Bartolo, P. Recent advances in enzymatic and non-enzymatic electrochemical glucose sensing. Sensors 2021, 21, 4672. [Google Scholar] [CrossRef]
  24. Yoo, E.-H.; Lee, S.-Y. Glucose Biosensors: An Overview of Use in Clinical Practice. Sensors 2010, 10, 4558–4576. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  25. Clark, L.C.; Lyons, C. Electrode Systems for Continuous Monitoring in Cardiovascular Surgery. Ann. N. Y. Acad. Sci. 1962, 102, 29–45. [Google Scholar] [CrossRef] [PubMed]
  26. D’Orazio, P. Biosensors in clinical chemistry. Clin. Chim. Acta 2003, 334, 41–69. [Google Scholar] [CrossRef]
  27. Newman, J.D.; Turner, A.P. Home blood glucose biosensors: A commercial perspective. Biosens. Bioelectron. 2005, 20, 2435–2453. [Google Scholar] [CrossRef] [Green Version]
  28. Juska, V.B.; Pemble, M.E. A Critical Review of Electrochemical Glucose Sensing: Evolution of Biosensor Platforms Based on Advanced Nanosystems. Sensors 2020, 20, 6013. [Google Scholar] [CrossRef]
  29. Matthews, D.R.; Holman, R.R.; Bown, E.; Steemson, J.; Watson, A.; Hughes, S.; Scott, D. Pen-sized digital 30-second blood glucose meter. Lancet 1987, 1, 778–779. [Google Scholar] [CrossRef]
  30. Ipekci, H.H.; Kazak, O.; Tor, A.; Uzunoglu, A. Tuning Active Sites of N-Doped Porous Carbon Catalysts Derived from Vinasse for High-Performance Electrochemical Sensing. Part. Sci. Technol. 2022, 1–12. [Google Scholar] [CrossRef]
  31. Quintero-Jaime, A.F.; Conzuelo, F.; Schuhmann, W.; Cazorla-Amorós, D.; Morallón, E. Multi-wall Carbon Nanotubes Electrochemically Modified with Phosphorus and Nitrogen Functionalities as a Basis for Bioelectrodes with Improved Performance. Electrochim. Acta 2021, 387, 138530. [Google Scholar] [CrossRef]
  32. Quintero-Jaime, A.F.; Conzuelo, F.; Cazorla-Amorós, D.; Morallón, E. Pyrroloquinoline Quinone-Dependent Glucose Dehydrogenase Bioelectrodes Based on One-Step Electrochemical Entrapment over Single-Wall Carbon Nanotubes. Talanta 2021, 232, 122386. [Google Scholar] [CrossRef]
  33. Jiang, D.; Chu, Z.; Peng, J.; Luo, J.; Mao, Y.; Yang, P.; Jin, W. One-Step Synthesis of Three-Dimensional Co(OH)2/RGO Nano-Flowers as Enzyme-Mimic Sensors for Glucose Detection. Electrochim. Acta 2018, 270, 147–155. [Google Scholar] [CrossRef]
  34. Xue, B.; Li, K.; Feng, L.; Lu, J.; Zhang, L. Graphene Wrapped Porous Co3O4/NiCo2O4 Double-Shelled Nanocages with Enhanced Electrocatalytic Performance for Glucose Sensor. Electrochim. Acta 2017, 239, 36–44. [Google Scholar] [CrossRef]
  35. Thatikayala, D.; Ponnamma, D.; Sadasivuni, K.K.; Cabibihan, J.J.; Al-Ali, A.K.; Malik, R.A.; Min, B. Progress of Advanced Nanomaterials in the Non-Enzymatic Electrochemical Sensing of Glucose and H2O2. Biosensors 2020, 10, 151. [Google Scholar] [CrossRef] [PubMed]
  36. Zhang, L.; Wang, J.; Liu, F.; Xiong, Y.; Liu, Z.; Jiang, D.; Pu, X. Rapid detection of blaNDM-1 in multidrug-resistant organisms using a novel electrochemical biosensor. RSC Adv. 2017, 7, 12576–12585. [Google Scholar] [CrossRef] [Green Version]
  37. Li, K.; Liu, W.; Ni, Y.; Li, D.; Lin, D.; Su, Z.; Wei, G. Technical Synthesis and Biomedical Applications of Graphene Quantum Dots. J. Mater. Chem. B 2017, 5, 4811–4826. [Google Scholar] [CrossRef]
  38. Mohammadniaei, M.; Yoon, J.; Lee, T.; Bharate, B.G.; Jo, J.; Lee, D.; Choi, J.W. Electrochemical Biosensor Composed of Silver Ion-Mediated DsDNA on Au-Encapsulated Bi2Se3 Nanoparticles for the Detection of H2O2 Released from Breast Cancer Cells. Small 2018, 14, 1703970. [Google Scholar] [CrossRef]
  39. Fu, Y.; An, Q.; Ni, R.; Zhang, Y.; Li, Y.; Ke, H. Preparation of Polyaniline-Encapsulated Carbon/Copper Composite Nanofibers for Detection of Polyphenol Pollutant. Colloids Surf. A Physicochem. Eng. Asp. 2018, 559, 289–296. [Google Scholar] [CrossRef]
  40. Wei, M.; Qiao, Y.; Zhao, H.; Liang, J.; Li, T.; Luo, Y.; Sun, X. Electrochemical non-enzymatic glucose sensors: Recent progress and perspectives. Chem. Commun. 2020, 56, 14553–14569. [Google Scholar] [CrossRef]
  41. Lee, C.W.; Suh, J.M.; Jang, H.W. Chemical Sensors Based on Two-Dimensional (2D) Materials for Selective Detection of Ions and Molecules in Liquid. Front. Chem. 2019, 7, 708. [Google Scholar] [CrossRef]
  42. Georgakilas, V.; Otyepka, M.; Bourlinos, A.B.; Chandra, V.; Kim, N.; Kemp, K.C.; Kim, K.S. Functionalization of Graphene: Covalent and Non-Covalent Approaches, Derivatives and Applications. Chem. Rev. 2012, 112, 6156–6214. [Google Scholar] [CrossRef]
  43. Moses, P.G.; Mortensen, J.J.; Lundqvist, B.I.; Nørskov, J.K. Density Functional Study of the Adsorption and van Der Waals Binding of Aromatic and Conjugated Compounds on the Basal Plane of MoS2. J. Chem. Phys. 2009, 130, 104709. [Google Scholar] [CrossRef] [Green Version]
  44. Berger, C.; Song, Z.; Li, X.; Wu, X.; Brown, N.; Naud, C.; Heer, W.A. Electronic Confinement and Coherence in Patterned Epitaxial Graphene. Science 2006, 312, 1191–1196. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  45. Wang, Y.; Shao, Y.; Matson, D.W.; Li, J.; Lin, Y. Nitrogen-doped graphene and its application in electrochemical biosensing. ACS Nano 2010, 4, 1790–1798. [Google Scholar] [CrossRef] [PubMed]
  46. Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X. A novel non-enzymatic glucose sensor based on Cu nanoparticle modified graphene sheets electrode. Analyt. Chim. Acta 2012, 709, 47–53. [Google Scholar] [CrossRef] [PubMed]
  47. Kwak, Y.H.; Choi, D.S.; Kim, Y.N.; Kim, H.; Yoon, D.H.; Ahn, S.-S.; Yang, J.-W.; Yang, W.S.; Seo, S. Flexible Glucose Sensor Using CVD-Grown Graphene-Based Field Effect Transistor. Biosens. Bioelectron. 2012, 37, 82–87. [Google Scholar] [CrossRef]
  48. Parlak, O.; ncel, A.; Uzun, L.; Turner, A.P.F.; Tiwari, A. Biosens. Bioelectron 2017, 89, 545–550. [Google Scholar] [CrossRef]
  49. Su, S.; Zhang, C.; Yuwen, L.; Liu, X.; Wang, L.; Fan, C.; Wang, L. Uniform Au@ Pt Core–Shell Nanodendrites Supported on Molybdenum Disulfide Nanosheets for the Methanol Oxidation Reaction. Nanoscale 2016, 8, 602–608. [Google Scholar] [CrossRef]
  50. Huang, J.; He, Y.; Jin, J.; Li, Y.; Dong, Z.; Li, R. A Novel Glucose Sensor Based on MoS2 Nanosheet Functionalized with Ni Nanoparticles. Electrochim. Acta 2014, 136, 41–46. [Google Scholar] [CrossRef]
  51. Lin, X.; Ni, Y.; Kokot, S. Electrochemical and Bio-Sensing Platform Based on a Novel 3D Cu Nano-Flowers/Layered MoS2 Composite. Biosens. Bioelectron. 2016, 79, 685–692. [Google Scholar] [CrossRef]
  52. Zhong, X.; Yang, H.; Guo, S.; Li, S.; Gou, G.; Niu, Z.; Ma, J. In Situ Growth of Ni–Fe Alloy on Graphene-like MoS2 for Catalysis of Hydrazine Oxidation. J. Mater. Chem. 2012, 22, 13925–13927. [Google Scholar] [CrossRef]
  53. Sun, Z.; Zhao, Q.; Zhang, G.; Li, Y.; Zhang, G.; Zhang, F.; Fan, X. Exfoliated MoS2 Supported Au–Pd Bimetallic Nanoparticles with Core–Shell Structures and Superior Peroxidase-like Activities. RSC Adv. 2015, 5, 10352–10357. [Google Scholar] [CrossRef]
  54. Li, J.; Yang, Z.; Tang, Y.; Zhang, Y.; Hu, X. Carbon nanotubes-nanoflake-like SnS2 nanocomposite for direct electrochemistry of glucose oxidase and glucose sensing. Biosens. Bioelectron. 2013, 41, 698–703. [Google Scholar] [CrossRef] [PubMed]
  55. Guo, C.X.; Li, C.M. Direct Electron Transfer of Glucose Oxidase and Biosensing of Glucose on Hollow Sphere-Nanostructured Conducting Polymer/Metal Oxide Composite. Phys. Chem. Chem. Phys. 2010, 12, 12153–12159. [Google Scholar] [CrossRef] [PubMed]
  56. Zhu, J.; Jiang, J.; Liu, J.; Ding, R.; Li, Y.; Ding, H.; Huang, X. CNT-Network Modified Ni Nanostructured Arrays for High Performance Non-Enzymatic Glucose Sensors. RSC Adv. 2011, 1, 1020–1025. [Google Scholar] [CrossRef]
  57. Ginsberg, B.H. Factors Affecting Blood Glucose Monitoring: Sources of Errors in Measurement. J. Diabetes Sci. Technol. 2009, 3, 903–913. [Google Scholar] [CrossRef] [Green Version]
  58. Lee, Y.J.; Park, D.J.; Park, J.Y.; Kim, Y. Fabrication and Optimization of a Nanoporous Platinum Electrode and a Non-Enzymatic Glucose Micro-Sensor on Silicon. Sensors 2008, 8, 6154–6164. [Google Scholar] [CrossRef]
  59. Jeon, W.Y.; Choi, Y.B.; Kim, H.H. Disposable Non-Enzymatic Glucose Sensors Using Screen-Printed Nickel/Carbon Composites on Indium Tin Oxide Electrodes. Sensors 2015, 15, 31083–31091. [Google Scholar] [CrossRef]
  60. Yu, X.; Zhang, J.; Tang, X.; Wei, Y.; Kou, S.; Niu, J.; Yao, B. Preparation and Performance of Non-Enzymatic Glucose Sensor Electrode Based on Nanometer Cuprous Oxide. Nanomater. Nanotechnol. 2018, 8, 1847980418793526. [Google Scholar] [CrossRef] [Green Version]
  61. Gong, X.; Gu, Y.; Zhang, F.; Liu, Z.; Li, Y.; Chen, G.; Wang, B. High-Performance Non-Enzymatic Glucose Sensors Based on CoNiCu Alloy Nanotubes Arrays Prepared by Electrodeposition. Front. Mater. 2019, 6, 1–9. [Google Scholar] [CrossRef] [Green Version]
  62. Shabbir, S.A.; Tariq, S.; Ashiq, M.; Khan, W.A. Non-Enzymatic Glucose Sensor with Electrodeposited Silver/Carbon Nanotubes Composite Electrode. Biosci. Rep. 2019, 39, BSR20181983. [Google Scholar] [CrossRef] [Green Version]
  63. Amin, B.G.; Masud, J.; Nath, M. A Non-Enzymatic Glucose Sensor Based on a CoNi2Se4/RGO Nanocomposite with Ultrahigh Sensitivity at Low Working Potential. J. Mater. Chem. B 2019, 7, 2338–2348. [Google Scholar] [CrossRef]
  64. Hasanzadeh, M.; Hasanzadeh, Z.; Alizadeh, S.; Sayadi, M.; Nasiri Nezhad, M.; Sabzi, R.E.; Ahmadi, S. Copper-Nickel Oxide Nanofilm Modified Electrode for Non-Enzymatic Determination of Glucose. J. Electrochem. Sci. Eng. 2020, 10, 245–255. [Google Scholar] [CrossRef]
  65. More, K.D.; Dadge, J.W.; Khairnar, R.S.; Bogle, K.A. Development of Nano-TiO2/Al Electrode for Non-Enzymatic Glucose Bio-Sensing Application. AIP Conf. Proc. 2020, 2220, 020003. [Google Scholar]
  66. Espro, C.; Marini, S.; Giusi, D.; Ampelli, C.; Neri, G. Non-Enzymatic Screen Printed Sensor Based on Cu2O Nanocubes for Glucose Determination in Bio-Fermentation Processes. J. Electroanal. Chem. 2020, 873, 114354. [Google Scholar] [CrossRef]
  67. Baek, S.H.; Roh, J.; Park, C.Y.; Kim, M.W.; Shi, R.; Kailasa, S.K.; Park, T.J. Cu-Nanoflower Decorated Gold Nanoparticles-Graphene Oxide Nanofiber as Electrochemical Biosensor for Glucose Detection. Mater. Sci. Eng. C 2020, 107, 110273. [Google Scholar] [CrossRef] [PubMed]
  68. Mao, Q.; Jing, W.; Zhou, F.; Liu, S.; Gao, W.; Wei, Z.; Jiang, Z. Depositing Reduced Graphene Oxide on ZnO Nanorods to Improve the Performance of Enzymatic Glucose Sensors. Mater. Sci. Semicond. Process. 2021, 121, 105391. [Google Scholar] [CrossRef]
  69. Hossain, M.F.; Slaughter, G. PtNPs Decorated Chemically Derived Graphene and Carbon Nanotubes for Sensitive and Selective Glucose Biosensing. J. Electroanal. Chem. 2020, 861, 113990. [Google Scholar] [CrossRef]
  70. Sakr, M.A.; Elgammal, K.; Delin, A.; Serry, M. Performance-Enhanced Non-Enzymatic Glucose Sensor Based on Graphene-Heterostructure. Sensors 2020, 20, 145. [Google Scholar] [CrossRef] [Green Version]
  71. Scandurra, A.; Ruffino, F.; Sanzaro, S.; Grimaldi, M.G. Laser and Thermal Dewetting of Gold Layer onto Graphene Paper for Non-Enzymatic Electrochemical Detection of Glucose and Fructose. Sens. Actuators B Chem. 2019, 301, 127113. [Google Scholar] [CrossRef]
  72. Jothi, L.; Jayakumar, N.; Jaganathan, S.K.; Nageswaran, G. Ultrasensitive and Selective Non-Enzymatic Electrochemical Glucose Sensor Based on Hybrid Material of Graphene Nanosheets/Graphene Nanoribbons/Nickel Nanoparticle. Mater. Res. Bull. 2018, 98, 300–307. [Google Scholar] [CrossRef]
  73. Ren, Z.; Mao, H.; Luo, H.; Liu, Y. Glucose Sensor Based on Porous Ni by Using a Graphene Bottom Layer Combined with a Ni Middle Layer. Carbon 2019, 149, 609–617. [Google Scholar] [CrossRef]
  74. Deng, Z.-P.; Sun, Y.; Wang, Y.-C.; Gao, J.-D. A NiFe Alloy Reduced on Graphene Oxide for Electrochemical Nonenzymatic Glucose Sensing. Sensors 2018, 18, 3972. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  75. Ayranci, R.; Demirkan, B.; Sen, B.; Şavk, A.; Ak, M.; Şen, F. Use of the Monodisperse Pt/Ni@rGO Nanocomposite Synthesized by Ultrasonic Hydroxide Assisted Reduction Method in Electrochemical Nonenzymatic Glucose Detection. Mater. Sci. Eng. C 2019, 99, 951–956. [Google Scholar] [CrossRef] [PubMed]
  76. Yan, X.; Gu, Y.; Li, C.; Zheng, B.; Li, Y.; Zhang, T.; Zhang, Z.; Yang, M. A Non-Enzymatic Glucose Sensor Based on the CuS Nanoflakes–Reduced Graphene Oxide Nanocomposite. Anal. Methods 2018, 10, 381–388. [Google Scholar] [CrossRef]
  77. Terán-Alcocer, Á.; Bravo-Plascencia, F.; Cevallos-Morillo, C.; Palma-Cando, A. Electrochemical Sensors Based on Conducting Polymers for the Aqueous Detection of Biologically Relevant Molecules. Nanomaterials 2021, 11, 252. [Google Scholar] [CrossRef]
  78. Lei, W.; Si, W.; Xu, Y.; Gu, Z.; Hao, Q. Conducting Polymer Composites with Graphene for Use in Chemical Sensors and Biosensors. Microchim. Acta 2014, 181, 707–722. [Google Scholar] [CrossRef]
  79. Muthusankar, E.; Ragupathy, D. Graphene/Poly(Aniline-Co-Diphenylamine) Nanohybrid for Ultrasensitive Electrochemical Glucose Sensor. Nano Struct. Nano Objects 2019, 20, 100390. [Google Scholar] [CrossRef]
  80. Deshmukh, M.A.; Kang, B.-C.; Ha, T.-J. Non-Enzymatic Electrochemical Glucose Sensors Based on Polyaniline/Reduced-Graphene-Oxide Nanocomposites Functionalized with Silver Nanoparticles. J. Mater. Chem. C 2020, 8, 5112–5123. [Google Scholar] [CrossRef]
  81. Eryiğit, M.; Çepni, E.; Kurt Urhan, B.; Öztürk Doğan, H.; Öznülüer Özer, T. Nonenzymatic Glucose Sensor Based on Poly(3,4-Ethylene Dioxythiophene)/Electroreduced Graphene Oxide Modified Gold Electrode. Synth. Met. 2020, 268, 116488. [Google Scholar] [CrossRef]
  82. Sivasankarapillai, V.S.; Sharma, T.S.K.; Hwa, K.-Y.; Wabaidur, S.M.; Angaiah, S.; Dhanusuraman, R. MXene Based Sensing Materials: Current Status and Future Perspectives. ES Energy Environ. 2022, 15, 4–14. [Google Scholar] [CrossRef]
  83. Kalambate, P.K.; Gadhari, N.S.; Li, X.; Rao, Z.; Navale, S.T.; Shen, Y.; Patil, V.R.; Huang, Y. Recent Advances in MXene–Based Electrochemical Sensors and Biosensors. TrAC Trends Anal. Chem. 2019, 120, 115643. [Google Scholar] [CrossRef]
  84. Wu, X.; Ma, P.; Sun, Y.; Du, F.; Song, D.; Xu, G. Application of MXene in Electrochemical Sensors: A Review. Electroanalysis 2021, 33, 1827–1851. [Google Scholar] [CrossRef]
  85. Mathew, M.; Rout, C.S. Electrochemical Biosensors Based on Ti3C2Tx MXene: Future Perspectives for on-Site Analysis. Curr. Opin. Electrochem. 2021, 30, 100782. [Google Scholar] [CrossRef]
  86. Zhu, J.; Ha, E.; Zhao, G.; Zhou, Y.; Huang, D.; Yue, G.; Hu, L.; Sun, N.; Wang, Y.; Lee, L.Y.S.; et al. Recent Advance in MXenes: A Promising 2D Material for Catalysis, Sensor and Chemical Adsorption. Coord. Chem. Rev. 2017, 352, 306–327. [Google Scholar] [CrossRef]
  87. Rakhi, R.B.; Nayak, P.; Xia, C.; Alshareef, H.N. Novel Amperometric Glucose Biosensor Based on MXene Nanocomposite. Sci. Rep. 2016, 6, 36422. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  88. Gu, H.; Xing, Y.; Xiong, P.; Tang, H.; Li, C.; Chen, S.; Zeng, R.; Han, K.; Shi, G. Three-Dimensional Porous Ti3C2Tx MXene–Graphene Hybrid Films for Glucose Biosensing. ACS Appl. Nano Mater. 2019, 2, 6537–6545. [Google Scholar] [CrossRef]
  89. Wu, M.; Zhang, Q.; Fang, Y.; Deng, C.; Zhou, F.; Zhang, Y.; Wang, X.; Tang, Y.; Wang, Y. Polylysine-Modified MXene Nanosheets with Highly Loaded Glucose Oxidase as Cascade Nanoreactor for Glucose Decomposition and Electrochemical Sensing. J. Colloid Interface Sci. 2021, 586, 20–29. [Google Scholar] [CrossRef]
  90. Li, M.; Fang, L.; Zhou, H.; Wu, F.; Lu, Y.; Luo, H.; Zhang, Y.; Hu, B. Three-Dimensional Porous MXene/NiCo-LDH Composite for High Performance Non-Enzymatic Glucose Sensor. Appl. Surf. Sci. 2019, 495, 143554. [Google Scholar] [CrossRef]
  91. Gopal, T.S.; Jeong, S.K.; Alrebdi, T.A.; Pandiaraj, S.; Alodhayb, A.; Muthuramamoorthy, M. Andrews Nirmala Grace MXene-Based Composite Electrodes for Efficient Electrochemical Sensing of Glucose by Non-Enzymatic Method. Mater. Today Chem. 2022, 24, 100891. [Google Scholar] [CrossRef]
  92. Mphuthi, N.; Sikhwivhilu, L.; Ray, S.S. Functionalization of 2D MoS2 Nanosheets with Various Metal and Metal Oxide Nanostructures: Their Properties and Application in Electrochemical Sensors. Biosensors 2022, 12, 386. [Google Scholar] [CrossRef]
  93. SpringerLink. Direct Electrochemistry of Glucose Oxidase and a Biosensor for Glucose Based on a Glass Carbon Electrode Modified with MoS2 Nanosheets Decorated with Gold Nanoparticles. Available online: https://link.springer.com/article/10.1007/s00604-014-1178-9 (accessed on 23 April 2022).
  94. Jeong, J.-M.; Yang, M.; Kim, D.S.; Lee, T.J.; Choi, B.G.; Kim, D.H. High Performance Electrochemical Glucose Sensor Based on Three-Dimensional MoS2/Graphene Aerogel. J. Colloid Interface Sci. 2017, 506, 379–385. [Google Scholar] [CrossRef]
  95. Huang, J.; Dong, Z.; Li, Y.; Li, J.; Tang, W.; Yang, H.; Wang, J.; Bao, Y.; Jin, J.; Li, R. MoS2 nanosheet functionalized with Cu nanoparticles and its application for glucose detection. Mater. Res. Bull. 2013, 48, 4544–4547. [Google Scholar] [CrossRef]
  96. He, W.; Huang, Y.; Wu, J. Enzyme-Free Glucose Biosensors Based on MoS2 Nanocomposites. Nanoscale Res. Lett. 2020, 15, 60. [Google Scholar] [CrossRef] [PubMed]
  97. Geng, D.; Bo, X.; Guo, L. Ni-Doped Molybdenum Disulfide Nanoparticles Anchored on Reduced Graphene Oxide as Novel Electroactive Material for a Non-Enzymatic Glucose Sensor. Sens. Actuators B Chem. 2017, 244, 131–141. [Google Scholar] [CrossRef]
  98. Ji, S.; Yang, Z.; Zhang, C.; Miao, Y.-E.; Tjiu, W.W.; Pan, J.; Liu, T. Nonenzymatic Sensor for Glucose Based on a Glassy Carbon Electrode Modified with Ni(OH)2 Nanoparticles Grown on a Film of Molybdenum Sulfide. Microchim. Acta 2013, 180, 1127–1134. [Google Scholar] [CrossRef]
  99. Fang, L.; Wang, F.; Chen, Z.; Qiu, Y.; Zhai, T.; Hu, M.; Zhang, C.; Huang, K. Flower-like MoS2 Decorated with Cu2O Nanoparticles for Non-Enzymatic Amperometric Sensing of Glucose. Talanta 2017, 167, 593–599. [Google Scholar] [CrossRef]
  100. Wu, S.; Zeng, Z.; He, Q.; Wang, Z.; Wang, S.J.; Du, Y.; Yin, Z.; Sun, X.; Chen, W.; Zhang, H. Electrochemically Reduced Single-Layer MoS2 Nanosheets: Characterization, Properties, and Sensing Applications. Small 2012, 8, 2264–2270. [Google Scholar] [CrossRef]
  101. Scientific Reports. Fabrication of 2D-MoSe2 Incorporated NiO Nanorods Modified Electrode for Selective Detection of Glucose in Serum Samples. Available online: https://www.nature.com/articles/s41598-021-92620-2 (accessed on 24 April 2022).
  102. Mani, S.; Ramaraj, S.; Chen, S.-M.; Dinesh, B.; Chen, T.-W. Two-Dimensional Metal Chalcogenides Analogous NiSe2 Nanosheets and Its Efficient Electrocatalytic Performance towards Glucose Sensing. J. Colloid Interface Sci. 2017, 507, 378–385. [Google Scholar] [CrossRef]
  103. Wan, L.; Liu, J.; Huang, X.-J. Novel Magnetic Nickel Telluride Nanowires Decorated with Thorns: Synthesis and Their Intrinsic Peroxidase-like Activity for Detection of Glucose. Chem. Commun. 2014, 50, 13589–13591. [Google Scholar] [CrossRef]
  104. Farid, A.; Pan, L.; Usman, M.; Khan, I.A.; Khan, A.S.; ul Ahmad, A.; Javid, M. In-Situ Growth of Porous CoTe2 Nanosheets Array on 3D Nickel Foam for Highly Sensitive Binder-Free Non-Enzymatic Glucose Sensor. J. Alloys Compd. 2021, 861, 158642. [Google Scholar] [CrossRef]
  105. Colombari, M.; Ballarin, B.; Carpani, I.; Guadagnini, L.; Mignani, A.; Scavetta, E.; Tonelli, D. Glucose Biosensors Based on Electrodes Modified with Ferrocene Derivatives Intercalated into Mg/Al Layered Double Hydroxides. Electroanalysis 2007, 19, 2321–2327. [Google Scholar] [CrossRef]
  106. Shan, D.; Cosnier, S.; Mousty, C. HRP Wiring by Redox Active Layered Double Hydroxides: Application to the Mediated H2O2 Detection. Anal. Lett. 2003, 36, 909–922. [Google Scholar] [CrossRef]
  107. Shan, D.; Yao, W.; Xue, H. Electrochemical Study of Ferrocenemethanol-Modified Layered Double Hydroxides Composite Matrix: Application to Glucose Amperometric Biosensor. Biosens. Bioelectron. 2007, 23, 432–437. [Google Scholar] [CrossRef]
  108. Mousty, C.; Prevot, V. Hybrid and Biohybrid Layered Double Hydroxides for Electrochemical Analysis. Anal. Bioanal. Chem. 2013, 405, 3513–3524. [Google Scholar] [CrossRef] [PubMed]
  109. Shi, Q.; Han, E.; Shan, D.; Yao, W.; Xue, H. Development of a High Analytical Performance Amperometric Glucose Biosensor Based on Glucose Oxidase Immobilized in a Composite Matrix: Layered Double Hydroxides/Chitosan. Bioprocess Biosyst. Eng. 2008, 31, 519–526. [Google Scholar] [CrossRef] [PubMed]
  110. Tonelli, D.; Scavetta, E.; Giorgetti, M. Layered-Double-Hydroxide-Modified Electrodes: Electroanalytical Applications. Anal. Bioanal. Chem. 2013, 405, 603–615. [Google Scholar] [CrossRef]
  111. Zhang, Y.; Chen, X.; Wang, J.; Yang, W. The Direct Electrochemistry of Glucose Oxidase Based on Layered Double-Hydroxide Nanosheets. Electrochem. Solid State Lett. 2008, 11, F19. [Google Scholar] [CrossRef]
  112. Scavetta, E.; Ballarin, B.; Gazzano, M.; Tonelli, D. Electrochemical Behaviour of Thin Films of Co/Al Layered Double Hydroxide Prepared by Electrodeposition. Electrochim. Acta 2009, 54, 1027–1033. [Google Scholar] [CrossRef]
  113. Lu, Y.; Jiang, B.; Fang, L.; Fan, S.; Wu, F.; Hu, B.; Meng, F. Highly Sensitive Nonenzymatic Glucose Sensor Based on 3D Ultrathin NiFe Layered Double Hydroxide Nanosheets. Electroanalysis 2017, 29, 1755–1761. [Google Scholar] [CrossRef]
  114. Zhao, Z.; Sun, Y.; Song, J.; Li, Y.; Xie, Y.; Cui, H.; Gong, W.; Hu, J.; Chen, Y. Highly Sensitive Nonenzymetic Glucose Sensing Based on Multicomponent Hierarchical NiCo-LDH/CCCH/CuF Nanostructures. Sens. Actuators B Chem. 2021, 326, 128811. [Google Scholar] [CrossRef]
  115. Hai, B.; Zou, Y.-Q.; Guo, G.-B.; Wang, Z.-D.; Liu, Y.-Y.; Wang, X.-X.; Yan, H.; Ma, L.-T.; Bai, Y.-C. A Novel Strategy to Prepare LDH Networks Loaded Carbon Structure by C-MEMS Techniques for Glucose Detection. Chin. Chem. Lett. 2017, 28, 149–152. [Google Scholar] [CrossRef]
  116. Fu, S.; Fan, G.; Yang, L.; Li, F. Non-Enzymatic Glucose Sensor Based on Au Nanoparticles Decorated Ternary Ni-Al Layered Double Hydroxide/Single-Walled Carbon Nanotubes/Graphene Nanocomposite. Electrochim. Acta 2015, 152, 146–154. [Google Scholar] [CrossRef]
  117. Hai, B.; Wei, H. Facile Preparation of Three-Dimensional NiO Nanosheets Array on Carbon Sphere and Its High-Performance Nonenzymatic Glucose Detection. IOP Conf. Ser. Earth Environ. Sci. 2019, 300, 052037. [Google Scholar] [CrossRef]
  118. Asadian, E.; Shahrokhian, S.; Iraji Zad, A. Highly Sensitive Nonenzymetic Glucose Sensing Platform Based on MOF-Derived NiCo LDH Nanosheets/Graphene Nanoribbons Composite. J. Electroanal. Chem. 2018, 808, 114–123. [Google Scholar] [CrossRef]
  119. Shahrokhian, S.; Sanati, E.K.; Hosseini, H. Advanced On-Site Glucose Sensing Platform Based on a New Architecture of Free-Standing Hollow Cu(OH)2 Nanotubes Decorated with CoNi-LDH Nanosheets on Graphite Screen-Printed Electrode. Nanoscale 2019, 11, 12655–12671. [Google Scholar] [CrossRef] [PubMed]
  120. Balasubramanian, P.; He, S.-B.; Deng, H.-H.; Peng, H.-P.; Chen, W. Defects Engineered 2D Ultrathin Cobalt Hydroxide Nanosheets as Highly Efficient Electrocatalyst for Non-Enzymatic Electrochemical Sensing of Glucose and l-Cysteine. Sens. Actuators B Chem. 2020, 320, 128374. [Google Scholar] [CrossRef]
  121. Sahoo, R.K.; Das, A.; Samantaray, K.; Singh, S.K.; Mane, R.S.; Shin, H.-C.; Yun, J.M.; Kim, K.H. Electrochemical Glucose Sensing Characteristics of Two-Dimensional Faceted and Non-Faceted CuO Nanoribbons. CrystEngComm 2019, 21, 1607–1616. [Google Scholar] [CrossRef]
  122. Chia, H.L.; Mayorga-Martinez, C.C.; Gusmão, R.; Novotny, F.; Webster, R.D.; Pumera, M. A Highly Sensitive Enzyme-Less Glucose Sensor Based on Pnictogens and Silver Shell–Gold Core Nanorod Composites. Chem. Commun. 2020, 56, 7909–7912. [Google Scholar] [CrossRef]
  123. Aziz, A.; Asif, M.; Ashraf, G.; Azeem, M.; Majeed, I.; Ajmal, M.; Wang, J.; Liu, H. Advancements in Electrochemical Sensing of Hydrogen Peroxide, Glucose and Dopamine by Using 2D Nanoarchitectures of Layered Double Hydroxides or Metal Dichalcogenides. A Review. Microchim. Acta 2019, 186, 671. [Google Scholar] [CrossRef]
  124. Ping, J.; Fan, Z.; Sindoro, M.; Ying, Y.; Zhang, H. Recent Advances in Sensing Applications of Two-Dimensional Transition Metal Dichalcogenide Nanosheets and Their Composites. Adv. Funct. Mater. 2017, 27, 1605817. [Google Scholar] [CrossRef]
  125. Zhao, A.; Zhang, Z.; Zhang, P.; Xiao, S.; Wang, L.; Dong, Y.; Yuan, H.; Li, P.; Sun, Y.; Jiang, X.; et al. 3D Nanoporous Gold Scaffold Supported on Graphene Paper: Freestanding and Flexible Electrode with High Loading of Ultrafine PtCo Alloy Nanoparticles for Electrochemical Glucose Sensing. Anal. Chim. Acta 2016, 938, 63–71. [Google Scholar] [CrossRef]
  126. He, W.; Sun, Y.; Xi, J.; Abdurhman, A.A.M.; Ren, J.; Duan, H. Printing Graphene-Carbon Nanotube-Ionic Liquid Gel on Graphene Paper: Towards Flexible Electrodes with Efficient Loading of PtAu Alloy Nanoparticles for Electrochemical Sensing of Blood Glucose. Anal. Chim. Acta 2016, 903, 61–68. [Google Scholar] [CrossRef] [PubMed]
  127. Abellán-Llobregat, A.; Jeerapan, I.; Bandodkar, A.; Vidal, L.; Canals, A.; Wang, J.; Morallón, E. A Stretchable and Screen-Printed Electrochemical Sensor for Glucose Determination in Human Perspiration. Biosens. Bioelectron. 2017, 91, 885–891. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  128. Xuan, X.; Yoon, H.S.; Park, J.Y. A Wearable Electrochemical Glucose Sensor Based on Simple and Low-Cost Fabrication Supported Micro-Patterned Reduced Graphene Oxide Nanocomposite Electrode on Flexible Substrate. Biosens. Bioelectron. 2018, 109, 75–82. [Google Scholar] [CrossRef] [PubMed]
  129. Cao, L.; Han, G.-C.; Xiao, H.; Chen, Z.; Fang, C. A Novel 3D Paper-Based Microfluidic Electrochemical Glucose Biosensor Based on RGO-TEPA/PB Sensitive Film. Anal. Chim. Acta 2020, 1096, 34–43. [Google Scholar] [CrossRef]
  130. Zhang, Y.; Li, N.; Xiang, Y.; Wang, D.; Zhang, P.; Wang, Y.; Lu, S.; Xu, R.; Zhao, J. A Flexible Non-Enzymatic Glucose Sensor Based on Copper Nanoparticles Anchored on Laser-Induced Graphene. Carbon 2020, 156, 506–513. [Google Scholar] [CrossRef]
  131. Lipani, L.; Dupont, B.G.R.; Doungmene, F.; Marken, F.; Tyrrell, R.M.; Guy, R.H.; Ilie, A. Non-Invasive, Transdermal, Path-Selective and Specific Glucose Monitoring via a Graphene-Based Platform. Nat. Nanotechnol. 2018, 13, 504–511. [Google Scholar] [CrossRef]
  132. Kim, J.; Kim, M.; Lee, M.-S.; Kim, K.; Ji, S.; Kim, Y.-T.; Park, J.; Na, K.; Bae, K.-H.; Kyun Kim, H.; et al. Wearable Smart Sensor Systems Integrated on Soft Contact Lenses for Wireless Ocular Diagnostics. Nat. Commun. 2017, 8, 14997. [Google Scholar] [CrossRef] [Green Version]
  133. Lee, H.; Choi, T.K.; Lee, Y.B.; Cho, H.R.; Ghaffari, R.; Wang, L.; Choi, H.J.; Chung, T.D.; Lu, N.; Hyeon, T.; et al. A Graphene-Based Electrochemical Device with Thermoresponsive Microneedles for Diabetes Monitoring and Therapy. Nat. Nanotechnol. 2016, 11, 566–572. [Google Scholar] [CrossRef]
  134. Lin, S.; Feng, W.; Miao, X.; Zhang, X.; Chen, S.; Chen, Y.; Wang, W.; Zhang, Y. A Flexible and Highly Sensitive Nonenzymatic Glucose Sensor Based on DVD-Laser Scribed Graphene Substrate. Biosens. Bioelectron. 2018, 110, 89–96. [Google Scholar] [CrossRef]
  135. Yoon, H.; Nah, J.; Kim, H.; Ko, S.; Sharifuzzaman, M.; Barman, S.C.; Xuan, X.; Kim, J.; Park, J.Y. A Chemically Modified Laser-Induced Porous Graphene Based Flexible and Ultrasensitive Electrochemical Biosensor for Sweat Glucose Detection. Sens. Actuators B Chem. 2020, 311, 127866. [Google Scholar] [CrossRef]
  136. Lei, Y.; Zhao, W.; Zhang, Y.; Jiang, Q.; He, J.-H.; Baeumner, A.J.; Wolfbeis, O.S.; Wang, Z.L.; Salama, K.N.; Alshareef, H.N. A MXene-Based Wearable Biosensor System for High-Performance In Vitro Perspiration Analysis. Small 2019, 15, 1901190. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  137. Li, M.; Wang, L.; Liu, R.; Li, J.; Zhang, Q.; Shi, G.; Li, Y.; Hou, C.; Wang, H. A Highly Integrated Sensing Paper for Wearable Electrochemical Sweat Analysis. Biosens. Bioelectron. 2021, 174, 112828. [Google Scholar] [CrossRef] [PubMed]
  138. Myndrul, V.; Coy, E.; Babayevska, N.; Zahorodna, V.; Balitskyi, V.; Baginskiy, I.; Gogotsi, O.; Bechelany, M.; Giardi, M.T.; Iatsunskyi, I. MXene Nanoflakes Decorating ZnO Tetrapods for Enhanced Performance of Skin-Attachable Stretchable Enzymatic Electrochemical Glucose Sensor. Biosens. Bioelectron. 2022, 207, 114141. [Google Scholar] [CrossRef] [PubMed]
  139. Yoon, J.; Lee, S.N.; Shin, M.K.; Kim, H.-W.; Choi, H.K.; Lee, T.; Choi, J.-W. Flexible Electrochemical Glucose Biosensor Based on GOx/Gold/MoS2/Gold Nanofilm on the Polymer Electrode. Biosens. Bioelectron. 2019, 140, 111343. [Google Scholar] [CrossRef]
  140. Vishnu, N.; Sahatiya, P.; Kong, C.Y.; Badhulika, S. Large Area, One Step Synthesis of NiSe2 Films on Cellulose Paper for Glucose Monitoring in Bio-Mimicking Samples for Clinical Diagnostics. Nanotechnology 2019, 30, 355502. [Google Scholar] [CrossRef]
  141. Liu, S.; Hui, K.S.; Hui, K.N. Flower-like Copper Cobaltite Nanosheets on Graphite Paper as High-Performance Supercapacitor Electrodes and Enzymeless Glucose Sensors. ACS Appl. Mater. Interfaces 2016, 8, 3258–3267. [Google Scholar] [CrossRef]
  142. Wang, X.; Zheng, Y.; Yuan, J.; Shen, J.; Hu, J.; Wang, A.-J.; Wu, L.; Niu, L. Three-Dimensional NiCo Layered Double Hydroxide Nanosheets Array on Carbon Cloth, Facile Preparation and Its Application in Highly Sensitive Enzymeless Glucose Detection. Electrochim. Acta 2017, 224, 628–635. [Google Scholar] [CrossRef]
  143. Hai, B.; Zou, Y. Carbon Cloth Supported NiAl-Layered Double Hydroxides for Flexible Application and Highly Sensitive Electrochemical Sensors. Sens. Actuators B Chem. 2015, 208, 143–150. [Google Scholar] [CrossRef]
  144. Katz, E.; Bückmann, A.F.; Willner, I. Self-Powered Enzyme-Based Biosensors. J. Am. Chem. Soc. 2001, 123, 10752–10753. [Google Scholar] [CrossRef]
  145. Cho, E.; Mohammadifar, M.; Choi, S. A Single-Use, Self-Powered, Paper-Based Sensor Patch for Detection of Exercise-Induced Hypoglycemia. Micromachines 2017, 8, 265. [Google Scholar] [CrossRef] [Green Version]
  146. Qiu, Z.; Shu, J.; Tang, D. Near-Infrared-to-Ultraviolet Light-Mediated Photoelectrochemical Aptasensing Platform for Cancer Biomarker Based on Core–Shell NaYF4:Yb,Tm@TiO2 Upconversion Microrods. Anal. Chem. 2018, 90, 1021–1028. [Google Scholar] [CrossRef]
  147. Wang, S.; Li, S.; Wang, W.; Zhao, M.; Liu, J.; Feng, H.; Chen, Y.; Gu, Q.; Du, Y.; Hao, W. A Non-Enzymatic Photoelectrochemical Glucose Sensor Based on BiVO4 Electrode under Visible Light. Sens. Actuators B Chem. 2019, 291, 34–41. [Google Scholar] [CrossRef]
  148. Luo, Z.; Qi, Q.; Zhang, L.; Zeng, R.; Su, L.; Tang, D. Branched Polyethylenimine-Modified Upconversion Nanohybrid-Mediated Photoelectrochemical Immunoassay with Synergistic Effect of Dual-Purpose Copper Ions. Anal. Chem. 2019, 91, 4149–4156. [Google Scholar] [CrossRef] [PubMed]
  149. Wang, H.; Zhang, L.; Chen, Z.; Hu, J.; Li, S.; Wang, Z.; Liu, J.; Wang, X. Semiconductor Heterojunction Photocatalysts: Design, Construction, and Photocatalytic Performances. Chem. Soc. Rev. 2014, 43, 5234–5244. [Google Scholar] [CrossRef] [PubMed]
  150. Xu, H.; Chang, Y.; Shen, X.; Liu, Z.; Zhu, B.; Macharia, D.K.; Wang, Z.; Chen, Z.; Zhang, L. Construction of Ag/AgCl-CN Heterojunctions with Enhanced Photocatalytic Activities for Degrading Contaminants in Wastewater. J. Colloid Interface Sci. 2019, 543, 25–33. [Google Scholar] [CrossRef]
  151. Devadoss, A.; Sudhagar, P.; Das, S.; Lee, S.Y.; Terashima, C.; Nakata, K.; Fujishima, A.; Choi, W.; Kang, Y.S.; Paik, U. Synergistic Metal–Metal Oxide Nanoparticles Supported Electrocatalytic Graphene for Improved Photoelectrochemical Glucose Oxidation. ACS Appl. Mater. Interfaces 2014, 6, 4864–4871. [Google Scholar] [CrossRef]
  152. Zhang, X.; Xu, F.; Zhao, B.; Ji, X.; Yao, Y.; Wu, D.; Gao, Z.; Jiang, K. Synthesis of CdS Quantum Dots Decorated Graphene Nanosheets and Non-Enzymatic Photoelectrochemical Detection of Glucose. Electrochim. Acta 2014, 133, 615–622. [Google Scholar] [CrossRef]
  153. Ma, Z.; Li, S. A Novel Photoelectrochemical Glucose Sensor Based on Graphene-CdS Nanocomposites Decorated with CoOx Nanosheets. Int. J. Electrochem. Sci. 2019, 14, 11445–11455. [Google Scholar] [CrossRef]
  154. Jafari, F.; Salimi, A.; Navaee, A. Electrochemical and Photoelectrochemical Sensing of Dihydronicotinamide Adenine Dinucleotide and Glucose Based on Noncovalently Functionalized Reduced Graphene Oxide-Cadmium Sulfide Quantum Dots/Poly-Nile Blue Nanocomposite. Electroanalysis 2014, 26, 1782–1793. [Google Scholar] [CrossRef]
  155. Li, H.; Guo, C.; Liu, C.; Ge, L.; Li, F. Laser-Induced Graphene Hybrid Photoelectrode for Enhanced Photoelectrochemical Detection of Glucose. Analyst 2020, 145, 4041–4049. [Google Scholar] [CrossRef]
  156. Çakıroğlu, B.; Özacar, M. Photoelectrochemical and Non-Enzymatic Glucose Sensor Based on Modified Fehling’s Test by Using Ti/TiO2 NTs-RGO-Cu2O Electrode. J. Electrochem. Soc. 2019, 166, B728. [Google Scholar] [CrossRef]
  157. Zhao, C.; Jing, T.; Tian, J.; Guo, J.; Wu, M.; Shi, D.; Zhao, Z.; Guo, Z. Visible Light-Driven Photoelectrochemical Enzyme Biosensor Based on Reduced Graphene Oxide/Titania for Detection of Glucose. J. Nanostruct. Chem. 2022, 12, 193–205. [Google Scholar] [CrossRef]
  158. Gopalan, A.I.; Muthuchamy, N.; Lee, K.P. A Novel Bismuth Oxychloride-Graphene Hybrid Nanosheets Based Non-Enzymatic Photoelectrochemical Glucose Sensing Platform for High Performances. Biosens. Bioelectron. 2017, 89, 352–360. [Google Scholar] [CrossRef]
  159. Wu, Z.; Zhao, J.; Yin, Z.; Wang, X.; Li, Z.; Wang, X. Highly Sensitive Photoelectrochemical Detection of Glucose Based on BiOBr/TiO2 Nanotube Array p-n Heterojunction Nanocomposites. Sens. Actuators B Chem. 2020, 312, 127978. [Google Scholar] [CrossRef]
  160. Zhang, X.Y.; Liu, S.G.; Zhang, W.J.; Wang, X.H.; Han, L.; Ling, Y.; Li, N.B.; Luo, H.Q. Photoelectrochemical Platform for Glucose Sensing Based on G-C3N4/ZnIn2S4 Composites Coupled with Bi-Enzyme Cascade Catalytic in-Situ Precipitation. Sens. Actuators B Chem. 2019, 297, 126818. [Google Scholar] [CrossRef]
  161. Li, W.; Jiang, D.; Yan, P.; Dong, J.; Qian, J.; Chen, J.; Xu, L. Graphitic Carbon Nitride/α-Fe2O3 Heterostructures for Sensitive Photoelectrochemical Non-Enzymatic Glucose Sensor. Inorg. Chem. Commun. 2019, 106, 211–216. [Google Scholar] [CrossRef]
  162. Chen, D.; Jiang, D.; Du, X.; Zhou, L.; Huang, L.; Qian, J.; Liu, Q.; Hao, N.; Li, Y.; Wang, K. Engineering Efficient Charge Transfer Based on Ultrathin Graphite-like Carbon Nitride/WO3 Semiconductor Nanoheterostructures for Fabrication of High-Performances Non-Enzymatic Photoelectrochemical Glucose Sensor. Electrochim. Acta 2016, 215, 305–312. [Google Scholar] [CrossRef]
  163. Çakıroğlu, B.; Demirci, Y.C.; Gökgöz, E.; Özacar, M. A Photoelectrochemical Glucose and Lactose Biosensor Consisting of Gold Nanoparticles, MnO2 and g-C3N4 Decorated TiO2. Sens. Actuators B Chem. 2019, 282, 282–289. [Google Scholar] [CrossRef]
  164. Li, M.; Wang, H.; Wang, X.; Lu, Q.; Li, H.; Zhang, Y.; Yao, S. Ti3C2/Cu2O Heterostructure Based Signal-off Photoelectrochemical Sensor for High Sensitivity Detection of Glucose. Biosens. Bioelectron. 2019, 142, 111535. [Google Scholar] [CrossRef]
  165. Chen, G.; Wang, H.; Wei, X.; Wu, Y.; Gu, W.; Hu, L.; Xu, D.; Zhu, C. Efficient Z-Scheme Heterostructure Based on TiO2/Ti3C2Tx/Cu2O to Boost Photoelectrochemical Response for Ultrasensitive Biosensing. Sens. Actuators B Chem. 2020, 312, 127951. [Google Scholar] [CrossRef]
  166. Wu, S.; Huang, H.; Shang, M.; Du, C.; Wu, Y.; Song, W. High Visible Light Sensitive MoS2 Ultrathin Nanosheets for Photoelectrochemical Biosensing. Biosens. Bioelectron. 2017, 92, 646–653. [Google Scholar] [CrossRef] [PubMed]
  167. Fang, Y.; Cao, Y.; Chen, Q. Synthesis of an Ag2WO4/Ti3C2 Schottky composite by electrostatic traction and its photocatalytic activity. Ceram. Int. 2019, 45, 22298–22307, ISSN 0272-8842. [Google Scholar] [CrossRef]
  168. Lathe, A.; Ansari, A.; Badhe, R.; Palve, A.M.; Garje, S.S. Single-Step Production of a TiO2@MoS2 Heterostructure and Its Applications as a Supercapacitor Electrode and Photocatalyst for Reduction of Cr(VI) to Cr(III). ACS Omega 2021, 6, 13008–13014. [Google Scholar] [CrossRef] [PubMed]
  169. Samoson, K.; Thavarungkul, P.; Kanatharana, P.; Limbut, W. A Nonenzymatic Glucose Sensor Based on the Excellent Dispersion of a Graphene Oxide-Poly(Acrylic Acid)-Palladium Nanoparticle-Modified Screen-Printed Carbon Electrode. J. Electrochem. Soc. 2019, 166, B1079. [Google Scholar] [CrossRef]
  170. Ansari, S.; Cho, M. Simple and Large Scale Construction of MoS2-g-C3N4 Heterostructures Using Mechanochemistry for High Performance Electrochemical Supercapacitor and Visible Light Photocatalytic Applications. Sci. Rep. 2017, 7, 43055. [Google Scholar] [CrossRef]
  171. Fu, X.; Chen, Z.; Shen, S.; Xu, L.; Luo, Z. Highly Sensitive Nonenzymatic Glucose Sensor Based on Reduced Graphene Oxide/Ultrasmall Pt Nanowire Nanocomposites. Int. J. Electrochem. Sci. 2018, 13, 4817–4826. [Google Scholar] [CrossRef]
  172. Farhat, H.; Célier, J.; Forano, C.; Mousty, C. Evaluation of Hierarchical Glucose Oxidase/Co3Mn-CO3 LDH Modified Electrodes for Glucose Detection. Electrochim. Acta 2021, 376, 138050. [Google Scholar] [CrossRef]
  173. Çakıroğlu, B.; Özacar, M. A Photoelectrochemical Biosensor Fabricated using Hierarchically Structured Gold Nanoparticle and MoS2 on Tannic Acid Templated Mesoporous TiO2. Electroanalysis 2019, 32, 166–177. [Google Scholar] [CrossRef]
  174. Deepalakshmi, T.; Tran, D.T.; Kim, N.H.; Chong, K.T.; Lee, J.H. Nitrogen-Doped Graphene-Encapsulated Nickel Cobalt Nitride as a Highly Sensitive and Selective Electrode for Glucose and Hydrogen Peroxide Sensing Applications. ACS Appl. Mater. Interfaces 2018, 10, 35847–35858. [Google Scholar] [CrossRef]
  175. Xie, F.; Liu, T.; Xie, L.; Sun, X.; Luo, Y. Metallic Nickel Nitride Nanosheet: An Efficient Catalyst Electrode for Sensitive and Selective Non-Enzymatic Glucose Sensing. Sens. Actuators B Chem. 2018, 255, 2794–2799. [Google Scholar] [CrossRef]
  176. Chen, J.; Yin, H.; Zhou, J.; Gong, J.; Wang, L.; Zheng, Y.; Nie, Q. Non-Enzymatic Glucose Sensor Based on Nickel Nitride Decorated Nitrogen Doped Carbon Spheres (Ni3N/NCS) via Facile One Pot Nitridation Process. J. Alloys Compd. 2019, 797, 922–930. [Google Scholar] [CrossRef]
  177. Zheng, L.; Teng, F.; Ye, X.; Zheng, H.; Fang, X. Photo/Electrochemical Applications of Metal Sulfide/TiO2 Heterostructures. Adv. Energy Mater. 2019, 10, 1902355. [Google Scholar] [CrossRef]
  178. Makov, G.; Payne, M.C. Periodic Boundary Conditions in Ab Initio Calculations. Phys. Rev. B 1995, 51, 4014–4022. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  179. Parr, R.G. Density Functional Theory. Annu. Rev. Phys. Chem. 1983, 34, 631–656. [Google Scholar] [CrossRef]
  180. Mooney, C.Z. Monte Carlo Simulation; Sage Publications, Inc.: Thousand Oaks, CA, USA, 1997; Volume 8, p. 103. ISBN 978-0-8039-5943-9. [Google Scholar]
  181. Slater, J.C. A Simplification of the Hartree-Fock Method. Phys. Rev. 1951, 81, 385–390. [Google Scholar] [CrossRef]
  182. Zhang, X.; Fang, R.; Chen, D.; Zhang, G. Using Pd-Doped γ-Graphyne to Detect Dissolved Gases in Transformer Oil: A Density Functional Theory Investigation. Nanomaterials 2019, 9, 1490. [Google Scholar] [CrossRef] [Green Version]
  183. Gehringer, D.; Dengg, T.; Popov, M.; Holec, D. Interactions between a H2 Molecule and Carbon Nanostructures: A DFT Study. C 2020, 6, 16. [Google Scholar] [CrossRef] [Green Version]
  184. Opoku, F.; Osikoya, A.O.; Dikio, E.D.; Govender, P.P. One-Step Synthesized 2D Heteroatom Doped Graphene for High Throughput Electrochemical Biosensing: A Combined Experimental and Computational Studies. Diam. Relat. Mater. 2019, 100, 107592. [Google Scholar] [CrossRef]
  185. Huang, Z.; Liu, H.; Hu, R.; Qiao, H.; Wang, H.; Liu, Y.; Qi, X.; Zhang, H. Structures, Properties and Application of 2D Monoelemental Materials (Xenes) as Graphene Analogues under Defect Engineering. Nano Today 2020, 35, 100906. [Google Scholar] [CrossRef]
  186. Mostufa, S.; Paul, A.K.; Chakrabarti, K. Detection of Hemoglobin in Blood and Urine Glucose Level Samples Using a Graphene-Coated SPR Based Biosensor. OSA Contin. 2021, 4, 2164–2176. [Google Scholar] [CrossRef]
  187. Chia, H.L.; Mayorga-Martinez, C.C.; Sofer, Z.; Lazar, P.; Webster, R.D.; Pumera, M. Vanadium Dopants: A Boon or a Bane for Molybdenum Dichalcogenides-Based Electrocatalysis Applications. Adv. Funct. Mater. 2021, 31, 2009083. [Google Scholar] [CrossRef]
  188. Zhang, Y.-H.; Chen, Y.-B.; Zhou, K.-G.; Liu, C.-H.; Zeng, J.; Zhang, H.-L.; Peng, Y. Improving Gas Sensing Properties of Graphene by Introducing Dopants and Defects: A First-Principles Study. Nanotechnology 2009, 20, 185504. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  189. Li, M.; Yao, J.; Wu, X.; Zhang, S.; Xing, B.; Niu, X.; Yan, X.; Yu, Y.; Liu, Y.; Wang, Y. P-Type Doping in Large-Area Monolayer MoS2 by Chemical Vapor Deposition. ACS Appl. Mater. Interfaces 2020, 12, 6276–6282. [Google Scholar] [CrossRef] [PubMed]
  190. Grimme, S. Semiempirical GGA-type density functional constructed with a long-range dispersion correction. J. Comput. Chem. 2006, 27, 1787–1799. [Google Scholar] [CrossRef]
  191. Stampfl, C.; Van de Walle, C.G. Density-Functional Calculations for III-V Nitrides Using the Local-Density Approximation and the Generalized Gradient Approximation. Phys. Rev. B 1999, 59, 5521–5535. [Google Scholar] [CrossRef] [Green Version]
  192. Moellmann, J.; Grimme, S. DFT-D3 Study of Some Molecular Crystals. J. Phys. Chem. C 2014, 118, 7615–7621. [Google Scholar] [CrossRef]
  193. Ebadi, M.; Reisi-Vanani, A. Methanol and Carbon Monoxide Sensing and Capturing by Pristine and Ca-Decorated Graphdiyne: A DFT-D2 Study. Phys. E Low Dimens. Syst. Nanostruct. 2021, 125, 114425. [Google Scholar] [CrossRef]
  194. Caglar, A.; Düzenli, D.; Onal, I.; Tezsevin, I.; Sahin, O.; Kivrak, H. A Comparative Experimental and Density Functional Study of Glucose Adsorption and Electrooxidation on the Au-Graphene and Pt-Graphene Electrodes. Int. J. Hydrogen Energy 2020, 45, 490–500. [Google Scholar] [CrossRef]
  195. Panigrahi, P.; Sajjad, M.; Singh, D.; Hussain, T.; Andreas Larsson, J.; Ahuja, R.; Singh, N. Two-Dimensional Nitrogenated Holey Graphene (C2N) Monolayer Based Glucose Sensor for Diabetes Mellitus. Appl. Surf. Sci. 2022, 573, 151579. [Google Scholar] [CrossRef]
  196. Govindasamy, P.; Gunasekaran, S.; Srinivasan, S. Molecular Geometry, Conformational, Vibrational Spectroscopic, Molecular Orbital and Mulliken Charge Analysis of 2-Acetoxybenzoic Acid. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2014, 130, 329–336. [Google Scholar] [CrossRef]
  197. Ertural, C.; Steinberg, S.; Dronskowski, R. Development of a Robust Tool to Extract Mulliken and Löwdin Charges from Plane Waves and Its Application to Solid-State Materials. RSC Adv. 2019, 9, 29821–29830. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  198. Lakshmy, S.; Sanyal, G.; Vaidyanathan, A.; Joseph, S.; Kalarikkal, N.; Chakraborty, B. Catechol Detection in Pure and Transition Metal Decorated 2D MoS2: Acumens from Density Functional Theory Approaches. Appl. Surf. Sci. 2021, 562, 150216. [Google Scholar] [CrossRef]
  199. Vaidyanathan, A.; Lakshmy, S.; Sanyal, G.; Joseph, S.; Kalarikkal, N.; Chakraborty, B. Nitrobenzene Sensing in Pristine and Metal Doped 2D Dichalcogenide MoS2: Insights from Density Functional Theory Investigations. Appl. Surf. Sci. 2021, 550, 149395. [Google Scholar] [CrossRef]
  200. Sanyal, G.; Lakshmy, S.; Vaidyanathan, A.; Kalarikkal, N.; Chakraborty, B. Detection of Nitrobenzene in Pristine and Metal Decorated 2D Dichalcogenide VSe2: Perspectives from Density Functional Theory. Surf. Interfaces 2022, 29, 101816. [Google Scholar] [CrossRef]
  201. Yong, Y.; Gao, R.; Wang, X.; Yuan, X.; Hu, S.; Zhao, Z.; Li, X.; Kuang, Y. Highly sensitive and selective room-temperature gas sensors based on B6N6H6 monolayer for sensing SO2 and NH3: A first-principles study. Results Phys. 2022, 33, 105208. [Google Scholar] [CrossRef]
  202. Zhao, Z.; Yong, Y.; Hu, S.; Li, C.; Kuang, Y. Adsorption of Gas Molecules on a C3N Monolayer and the Implications for NO2 Sensors. AIP Adv. 2019, 9, 125308. [Google Scholar] [CrossRef] [Green Version]
  203. Ivády, V.; Armiento, R.; Szász, K.; Janzén, E.; Gali, A.; Abrikosov, I.A. Theoretical Unification of Hybrid-DFT and DFT + U Methods for the Treatment of Localized Orbitals. Phys. Rev. B 2014, 90, 035146. [Google Scholar] [CrossRef] [Green Version]
  204. Shishkin, M.; Marsman, M.; Kresse, G. Accurate Quasiparticle Spectra from Self-Consistent GW Calculations with Vertex Corrections. Phys. Rev. Lett. 2007, 99, 246403. [Google Scholar] [CrossRef] [Green Version]
  205. Deák, P.; Aradi, B.; Frauenheim, T.; Janzén, E.; Gali, A. Accurate Defect Levels Obtained from the HSE06 Range-Separated Hybrid Functional. Phys. Rev. B 2010, 81, 153203. [Google Scholar] [CrossRef] [Green Version]
  206. Yong, Y.; Cui, H.; Zhou, Q.; Su, X.; Kuang, Y.; Li, X. C2N Monolayer as NH3 and NO Sensors: A DFT Study. Appl. Surf. Sci. 2019, 487, 488–495. [Google Scholar] [CrossRef]
  207. ACS Sustainable Chemistry & Engineering. Glucose Dehydrogenase-like Nanozyme Based on Black Phosphorus Nanosheets for High-Performance Biofuel Cells. Available online: https://pubs.acs.org/doi/abs/10.1021/acssuschemeng.0c05743 (accessed on 9 May 2022).
  208. Wang, X.; Sun, Y.; Liu, K. Chemical and Structural Stability of 2D Layered Materials. 2D Mater. 2019, 6, 042001. [Google Scholar] [CrossRef]
Figure 1. Summary of recent advancement of two-dimensional materials for electrochemical glucose sensors.
Figure 1. Summary of recent advancement of two-dimensional materials for electrochemical glucose sensors.
Biosensors 12 00467 g001
Figure 2. (a) YSI 23A Bio-Glucose Sensor. Reprinted with permission from Ref. [26]. 2003, Elsevier (Amsterdam, The Netherlands). (b) two-membrane system for glucose detection. Reprinted with permission from Ref. [27]. 2005, Elsevier.
Figure 2. (a) YSI 23A Bio-Glucose Sensor. Reprinted with permission from Ref. [26]. 2003, Elsevier (Amsterdam, The Netherlands). (b) two-membrane system for glucose detection. Reprinted with permission from Ref. [27]. 2005, Elsevier.
Biosensors 12 00467 g002
Figure 3. Schematic illustration of chemical sensors based on 2D materials with advantages of 2D materials and their application into ion/molecule sensing. Reprinted with permission from Ref. [41]. 2019, Lee, Suh and Jang.
Figure 3. Schematic illustration of chemical sensors based on 2D materials with advantages of 2D materials and their application into ion/molecule sensing. Reprinted with permission from Ref. [41]. 2019, Lee, Suh and Jang.
Biosensors 12 00467 g003
Figure 4. Glucose detection performance with (A) CVD Graphene FET and (B) response curves of glucose to different gate voltages. Reprinted with permission from Ref. [47]. 2012, Elsevier.
Figure 4. Glucose detection performance with (A) CVD Graphene FET and (B) response curves of glucose to different gate voltages. Reprinted with permission from Ref. [47]. 2012, Elsevier.
Biosensors 12 00467 g004
Figure 5. Glucose sensors using transition metal dichalcogenides. (A) Illustration for basic glucose sensor using Au nanoparticle decorated MoS2. (B) Response comparison between pristine MoS2, MoS2/GOx, and MoS2/Au NP/GOx electrodes. Reprinted with permission from Ref. [48]. 2017, Elsevier. (C) Scanning electron microscope images of pristine SnS2 (left) and GOx/MWCNTs–SnS2 (right). (D) Response curves of glucose having selectivity at GOx/MWCNTs–SnS2. Figures (C,D) were (Reproduced from [50]. 2014, Elsevier).
Figure 5. Glucose sensors using transition metal dichalcogenides. (A) Illustration for basic glucose sensor using Au nanoparticle decorated MoS2. (B) Response comparison between pristine MoS2, MoS2/GOx, and MoS2/Au NP/GOx electrodes. Reprinted with permission from Ref. [48]. 2017, Elsevier. (C) Scanning electron microscope images of pristine SnS2 (left) and GOx/MWCNTs–SnS2 (right). (D) Response curves of glucose having selectivity at GOx/MWCNTs–SnS2. Figures (C,D) were (Reproduced from [50]. 2014, Elsevier).
Biosensors 12 00467 g005
Figure 6. Fabrication procedures of non-enzymatic glucose micro-sensor with nanoporous Pt WE (A) and nanoporous Pt electrode (B) on silicon substrate. Reprinted from Ref. [58].
Figure 6. Fabrication procedures of non-enzymatic glucose micro-sensor with nanoporous Pt WE (A) and nanoporous Pt electrode (B) on silicon substrate. Reprinted from Ref. [58].
Biosensors 12 00467 g006
Figure 7. Schematic illustration of the (a) fabricating of Ni(OH)2/NiOOH-DSPNCE; (b) analysis of Ni(OH)2/NiOOH-DSPNCE; and (c) the electrochemical test. Reprinted from Ref. [59].
Figure 7. Schematic illustration of the (a) fabricating of Ni(OH)2/NiOOH-DSPNCE; (b) analysis of Ni(OH)2/NiOOH-DSPNCE; and (c) the electrochemical test. Reprinted from Ref. [59].
Biosensors 12 00467 g007
Figure 8. Graphene (G)/platinum oxide (PtO)/n-silicon (Si) heterostructure based glucose sensor working principle. Reprinted Ref. [70].
Figure 8. Graphene (G)/platinum oxide (PtO)/n-silicon (Si) heterostructure based glucose sensor working principle. Reprinted Ref. [70].
Biosensors 12 00467 g008
Figure 9. (A) Steps involved in the preparation of electrochemically rMoS2 and chitosan modified MoS2 electrodes and its CV curve in 0.1 M PBS. (Source: Reprinted with permission from Ref [100] Small 2012, 8, 2264–2270, doi:10.1002/smll.201200044. from Elsevier) (B) Amperometric curve NiO/MoSe2/GCE with the successive additions of glucose and its oxidation mechanism. (Source: Reprinted with permission from Ref [101] Scientific Reports Available online: https://www.nature.com/articles/s41598-021-92620-2 (accessed on 24 April 2022) from Elsevier).
Figure 9. (A) Steps involved in the preparation of electrochemically rMoS2 and chitosan modified MoS2 electrodes and its CV curve in 0.1 M PBS. (Source: Reprinted with permission from Ref [100] Small 2012, 8, 2264–2270, doi:10.1002/smll.201200044. from Elsevier) (B) Amperometric curve NiO/MoSe2/GCE with the successive additions of glucose and its oxidation mechanism. (Source: Reprinted with permission from Ref [101] Scientific Reports Available online: https://www.nature.com/articles/s41598-021-92620-2 (accessed on 24 April 2022) from Elsevier).
Biosensors 12 00467 g009
Figure 10. Schematic diagram of Graphene-WO3-Au triplet junction and its energy levels. Reprinted with permission from Ref [152]. Electrochimica Acta.
Figure 10. Schematic diagram of Graphene-WO3-Au triplet junction and its energy levels. Reprinted with permission from Ref [152]. Electrochimica Acta.
Biosensors 12 00467 g010
Figure 11. DFT optimized structures of (a) Graphyne, Reprinted from Ref. [182]. (b) POP-graphene with various adsorption sites, Reprinted from Ref. [183]. (c) D-glucose molecule and (d,e) HOMO and LUMO plots of D-glucose molecule. Reprinted from Ref. [184].
Figure 11. DFT optimized structures of (a) Graphyne, Reprinted from Ref. [182]. (b) POP-graphene with various adsorption sites, Reprinted from Ref. [183]. (c) D-glucose molecule and (d,e) HOMO and LUMO plots of D-glucose molecule. Reprinted from Ref. [184].
Biosensors 12 00467 g011
Figure 12. Geometries of 7 × 7 supercell of Nb-doped MoS2 in its (a) top and (b) side view. (c) Comparison of the total density of states plot of Nb-doped (upper panel) and pristine MoS2 (lower panel). Band structure plots of (d) pristine MoS2 (e) Nb-MoS2, Reprinted from Ref. [189].
Figure 12. Geometries of 7 × 7 supercell of Nb-doped MoS2 in its (a) top and (b) side view. (c) Comparison of the total density of states plot of Nb-doped (upper panel) and pristine MoS2 (lower panel). Band structure plots of (d) pristine MoS2 (e) Nb-MoS2, Reprinted from Ref. [189].
Biosensors 12 00467 g012
Figure 13. Optimised geometries and DOS of (a,b) 2 × 2 × 1 C2N monolayer (c,d) Glucose adsorbed C2N. Reprinted from Ref. [195].
Figure 13. Optimised geometries and DOS of (a,b) 2 × 2 × 1 C2N monolayer (c,d) Glucose adsorbed C2N. Reprinted from Ref. [195].
Biosensors 12 00467 g013
Figure 14. DFT optimised structures of glucose interaction with (a) Ag- (b)Au-doped graphene (c) Pt/graphene heterostructure. Reprinted with permission from Ref. [194], 2020, Elsevier. Charge density difference plots of glucose adsorbed (d) N- (e) O- and (f) Cl- doped hydrogenated graphene (HGr). Yellow and blue surfaces represent the charge density reduction and gain, respectively, with an isovalue of 0.0075 e Å−3. Reprinted from Ref. [184].
Figure 14. DFT optimised structures of glucose interaction with (a) Ag- (b)Au-doped graphene (c) Pt/graphene heterostructure. Reprinted with permission from Ref. [194], 2020, Elsevier. Charge density difference plots of glucose adsorbed (d) N- (e) O- and (f) Cl- doped hydrogenated graphene (HGr). Yellow and blue surfaces represent the charge density reduction and gain, respectively, with an isovalue of 0.0075 e Å−3. Reprinted from Ref. [184].
Biosensors 12 00467 g014
Figure 15. PDOS plots of D -glucose molecule adsorption on the (a) N-, (b) O-, and (c) Cl- decorated HGr, Reprinted from Ref. [184]. (d) 1000 K, (e) 1200 K, and (f) 1500 K after five ps of the Molecular Dynamic simulations. (g) the total energy variations of the B6N6H6 monolayer during the MD simulations at T = 300, 500, 1000, 1200, and 1500 K, Reprinted with permission from Ref. [201].
Figure 15. PDOS plots of D -glucose molecule adsorption on the (a) N-, (b) O-, and (c) Cl- decorated HGr, Reprinted from Ref. [184]. (d) 1000 K, (e) 1200 K, and (f) 1500 K after five ps of the Molecular Dynamic simulations. (g) the total energy variations of the B6N6H6 monolayer during the MD simulations at T = 300, 500, 1000, 1200, and 1500 K, Reprinted with permission from Ref. [201].
Biosensors 12 00467 g015
Figure 16. Future prospects on electrochemical glucose sensors based on 2D material.
Figure 16. Future prospects on electrochemical glucose sensors based on 2D material.
Biosensors 12 00467 g016
Table 1. Advantage and disadvantages of electrochemical glucose sensors based on 2D materials.
Table 1. Advantage and disadvantages of electrochemical glucose sensors based on 2D materials.
ProsCons
Low power requirements, linear output, and good resolution
Excellent repeatability and accuracy
Less expensive
Fast response time with high sensitivity and low detection limit
Multianalyte detection [11]
Narrow or limited temperature range
Short or limited shelf life
Table 2. Different types of electrochemical sensors.
Table 2. Different types of electrochemical sensors.
AmperometryConductometryVoltammetryPotentiometry
Amperometry means the measurement of the current flow in a closed loop of cells using an excitation signal produced by the generator
Advantages are
  • Good sensitivity and low detection limit
  • Excellent stability makes them particularly suitable for long-term monitoring
Disadvantages are
  • Upon applying a high polarizing voltage (Eapp = 0.6–0.8 V), interfering molecules such as uric acid and ascorbic acid, which are commonly present in biological fluids, are also oxidized, leading to nonspecific signals [12]
Response of the conductometric enzyme biosensors is mainly due to protons generated by a biocatalytic reaction inside the layer of immobilized enzyme
Advantages are
  • Suitable for miniaturization
  • No light sensitivity
  • Do not need any reference electrode, and the driving voltage can be sufficiently small to decrease the sensor power consumption substantially and reduce safety problems when used in living organisms.
Disadvantages are
Not very specific (less selective) [13]
Here measurements are related to the recording of either the current–time or the current–voltage relationship by applying known potential varying between the WE and REF electrodes
Advantages are
Simplicity, sensitivity, speed, and low costs
In this technique, negligible bias current flows as the potential between a working electrode and a reference electrode is measured across some interface
Advantages are
  • No extra potential is required
  • Simple, compact
  • The negligible current flow means the technique should be more resistant to interferent effects and ohmic drop considerations
  • Miniaturization without loss of sensitivity
Disadvantages are
Not very specific (less selective) [14]
Table 3. Properties of 2D materials required for sensing.
Table 3. Properties of 2D materials required for sensing.
2D MaterialsProperties Required for Sensing
Graphene
(1)
Large surface area aids in the adsorption and diffusion of glucose over the electrode, a larger sensing signal and improved sensitivity.
(2)
GO as well as rGO can improve the interfacial contact between the electroactive material and the current collector and thereby preventing glucose oxidation
(3)
Because of the excellent mechanical and thermal stability, graphene can help improve catalyst stability.
(4)
Biocompatible
(5)
With its negative charge, hydrophilicity, and surface smoothness, graphene is a good antifouling material that can repel negatively charged and hydrophilic foulants and effectively avoid sensitivity loss.
MXene
(1)
Compared to other 2D materials, MXene offers a distinct advantage in terms of electrical conductivity, which is crucial for increasing the rate of heterogeneous electron transport. Its high conductivity is one of the primary reasons for its application in electrochemical glucose sensors.
(2)
MXene is an easy to synthesize compound with good stability and solution dispersibility. This is crucial for producing electrochemical glucose sensors since drop-casting is the most prevalent process for making modified electrodes, which needs the preparation of the appropriately dispersed coating solution in advance.
(3)
MXene is an effective substrate material for printing inks.
(4)
Because of its outstanding stretchability and biocompatibility, MXene can be a powerful substrate for the development of flexible conductive platforms that can be used to construct wearable electrochemical glucose sensors [84].
TMDs
(1)
With TMDs nanosheets and their noble metal nanocomposites, electrochemical sensors have improved capabilities due to lower charge transfer resistance and the availability of more reaction sites.
(2)
A higher surface-to-volume ratio and a smaller size [123]
(3)
Two-dimensional TMD nanosheets are appealing as electrochemical glucose sensors because of their superior conductivity, huge surface area, high signal/noise ratio and quick electron transfer kinetics and, most importantly, their ability to create composites [124].
LDH
(1)
Methods of synthesis are simple and inexpensive
(2)
Rough surface enhances the mobility of analyte
(3)
Layers can be effectively tailored to intercalate preferred anions
(4)
Provide surfaces where enzymes can maintain their activity
(5)
Allow for easy loading of catalysts
(6)
Surfaces can be used to immobilize some modifiers or biomolecules.
Table 4. Overview of electrochemical glucose sensors based on two-dimensional materials for glucose detection.
Table 4. Overview of electrochemical glucose sensors based on two-dimensional materials for glucose detection.
ElectrodeAnalayteLinear RangeStability with Response CurrentDetection LimitRef
Ni–MoS2Glucose0–2.4 mM96.6% after 4 weeks 0.31 μ M[50]
Cu-nanoflower@AuNPs-GO NFsGlucose0.001–0.1 mM91% after 20 days0.018 μM[67]
GOx/rGO/ZnO/Au/PETGlucose0.1 to 12 mM90% after 10 scan cycle301.9 μM[68]
Au/CG/C@MWCNTs/PtNPs/GOx/nafionGlucose0.5 to 13.5 mM91.5% after 4 weeks1.3 μM[69]
Graphene (G)/platinum oxide (PtO)/n-silicon (Si) heterostructureGlucose2–20 mM-Sensitivity around 30 μA/mM.cm2[70]
Graphene paper/AuNPs laser dewettedGlucose and fructose detection20 μM–8 mM-2.5 μM[71]
GS/GNR/NiGlucose 5 nM–5 mM91.3% after 30 days2.5 nM[72]
Cu/Ni-EG/pNiHuman blood serum0.0005–1.0 mM92% after 8 days6161 mA/mM−1 cm−2[73]
NiFe/GOGlucose0.05 to 5 mMRecovery value close to 100%9 μM[74]
Pt/Ni@rGOGlucose 0.02–5.0 mM98% after 10 weeks6.3 μM[75]
rGO/CuSNFs/GCEGlucose1 to 2000 µM97% after 30 days0.19 µM[76]
Gra-PANI-co-PDPA-MEGlucose1 to 10 µM94% after 20 days0.1 µM[79]
Ag-PANI/rGOOrganic fluids (orange juice, apple juice, mango juice, Coke, and milk)50 μM to 0.1 μM91.3 after 30 days0.79 μM[80]
Au–PEDOT–ERGOGlucose0.1–100 mM90% after 15 days0.12 μM[81]
GOx/Au/MXene/Nafion/GCEGlucose0.1 to 18 mM-5.9 μM[87]
Ti3C2Tx MXene–graphene (MG) hybridPBS--0.10 mM in PBS and 0.13 mM in O2- saturated PBS[88]
Ti3C2/poly-L-lysine (PLL)/glucose oxidase (GOx)Glucose4.0–20 µM and 0.02–1.1 mM85% after 100 cycles2.6 μM[89]
MXene/Nickel–Cobalt layered double hydroxide (NiCo-LDH)Glucose0.002–4.096 mM92.5% after 15 days0.53 μM[91]
MXene–Cu2OGlucose0.01–30 mM -2.83 μM[90]
GOx/3D MGA/GCGlucose2 to 20mM-0.29 mM[94]
Cu–MoS2 hybridGlucoseUp to 4 mM-sensitivity 1055 µA mM−1 cm−2[95]
Ni–MoS2/rGOGlucose0.005–8.2 mM90% after 5 days2.7 μM[97]
Ni(OH)2/MoSxGlucose10–1300 μM90% after 2 weeks5.8 μM[98]
Cu2O–MoS2/GCEGlucose0.01 to 4.0 mM90% after 10 weeks1.0 μM[99]
MoSe2/NiO nanorod Blood and serum 50 µM to 15.5 mM91%after 50 cycles.0.6 µM[101]
Ni3Te2Glucose0.1–0.5 µM-25 nM[103]
3D CoTe2Glucose-96% after one month0.59 μM[104]
Zn–Cr–ABTS LDHGlucose1.3 10−8 –6.3 10−7 M-10 nm[106]
LDHs/MeOHFc/GODGlucose6.7 × 10−6 to 3.86 × 10−4 M-2.25 mM[107]
LDHs/CHT/GODGlucose1 × 10−6 to 3 × 10−3 M70% after 60 days0.1 µM[109]
NiFe–LDH/NF hybridBlood and serumUp to 0.8 mM-0.59 μM[113]
NiAl–LDH/CMC Glucose0.2–18.6 mmol L−1-0.12 mmol L−1[115]
Au/LDH-CNTs-GGlucose10 μM to 6.1 mM95% after 30 days1.0 μM[116]
NiCo NSs/GNR-GCEGlucose5 μM–0.8 mM
1–10 mM
93% after 3 weeks0.6 μM[118]
NiCo-LDH/CCCH/CuFGlucose0.001–1.5 mM-0.68 μM[114]
Cu(OH)2@CoNi-LDH NT-NSs/GSPE Glucose0.002–3.2 3.2–7.7-0.6 µM[119]
VCo-Co(OH)2 Glucose0.4 μM–8.23 mM-295 nM[120]
2D CuO nanoribbonGlucoseUp to 2 mMSimilar current response after 30 days58 µM[121]
Wearable electrochemical sensors
Ag/Pt @rGOSweat 48 μA/mMcm2[128]
PtCo/NPG/GPBlood35 μM–30 mM88% after 30 days5 μM[125]
GOx/PtNP/acetic acid-treated LIGSweat2.1 mM-0.3 µM[135]
CNTs/Ti3C2Tx/PB/CFMsSweat10 × 10−6–22 × 10−3 M-GlucoseLactate[136]
0.33 × 10−6 M0.67 × 10-6 M
rGO-TEPA/PB
Human sweat and blood0.1–25 mM75.1% after 2 weeks25 μM[130]
Ti3C2Tx/MBSweat0.08–1.25 mM-GlucoseLactate[137]
17.05 μM3.73 μM
ZnO TPs/MXeneSweat0.05–0.7 mM-17.05 μM[138]
GOx/gold/MoS2/goldGlucose500–100 nm-10 nM[139]
NiCo LDH/CCGlucose1 μM to 1.5 mM95% after one month0.12 μM[142]
NiSe2Glucose0.1–1 mM-24.8 μM[140]
NiAl-LDH
Glucose1 to 329 µM94.7% after one month0.22 µM[143]
Photoelectrochemical sensor
Graphene–CdS hybridGlucose0.1∼4mmol dm−3-7 μmol dm-3[152]
CoOx/graphene-CdS/GCEUrine5–370 μM-0.5 μM[153]
rGO-CdS QDs/PBA/P-NB/GDHGlucoseUpto 200 μM 95% response to NADH oxidation<1mM[154]
LI-NiEC-CdS-G@ITOGlucose--0.4 μM[155]
Cu2O-rGO TiO2 NTs/Ti electrodeGlucose0.0007–20 mMPhotocurrent response retained upto 94% after 7 weeks0.21 μM[156]
rGO/TiO2 nanotubeGlucose --5 μM[157]
BiOCl-G NHSHuman serum 500 µM–10mM97% after 4 weeks0.22 mM[158]
BiOBr-TNTAHuman serum5 × 10−2–3 × 107 nM95% after 4 weeks10 nM[159]
g-C3N4/ZnIn2S4Glucose1–10,000 μM 0.28 μM[160]
g-C3N4/WO3Glucose0.01~7.12 mM89.5% after 30 days0.1 μM[162]
g-C3N4/Fe2O3Glucoseg 0.1 to 11.5 mg L−1-0.03 mg/L[161]
ITO/TiO2-Au NPs- g-C3N4 -MnO2Glucose and lactose 0.004–1.75 mM92.3% after 5 weeks0.12 µM[163]
Ti3C2 MXene/Cu2OHuman serum samples--0.17 nM[164]
TiO2/Ti3C2/Cu2OHuman serum samples--33.75 nM[165]
MoS2 ultrathin nanosheetsGlucose-98% 0.61 nmol/L[166]
Other glucose sensors
rGO-PtNWGlucose0.032–1.89 mmol/L-4.6 μmol/L[171]
GOx/Co3Mn-CO3/CFGlucose--0.02 mM[172]
NiCo2N/N-doped grapheneGlucose2.008 µM to 7.15 mM98.52%0.05 µM[174]
Ni3N nanosheetsGlucose0.2 μM to 1.5 mM91.3% after one month0.06 µM[175]
Ni3N/NCSGlucose1 μM–3000 μM and 3000 μM–7000 μM-0.1 μM and 0.35 μM[176]
Table 5. Binding energy, bond distances, and charge transfer of various glucose adsorption systems.
Table 5. Binding energy, bond distances, and charge transfer of various glucose adsorption systems.
SystemAdsorption Energy
(Ev)
Bond Distance
(Å)
Charge
Transfer ∆q
(E)
Ref.
Graphene + Au + glucose−0.71Au—O = 2.28−0.21[194]
Graphene + Pt + glucose−0.38Pt—O = 2.39−0.17[194]
C2N + glucose−0.93 (gas phase)
−1.32 (aq. phase)
−0.01[195]
HGr + Cl + glucose−1.86Cl—O = 2.850
H—C = 1.746
−0.20[184]
HGr + N + glucose−1.69N—O = 3.372
H—C = 2.642
−0.08[184]
HGr + O + glucose−1.73O—O = 2.926
H—C = 2.371
−0.12[184]
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Radhakrishnan, S.; Lakshmy, S.; Santhosh, S.; Kalarikkal, N.; Chakraborty, B.; Rout, C.S. Recent Developments and Future Perspective on Electrochemical Glucose Sensors Based on 2D Materials. Biosensors 2022, 12, 467. https://doi.org/10.3390/bios12070467

AMA Style

Radhakrishnan S, Lakshmy S, Santhosh S, Kalarikkal N, Chakraborty B, Rout CS. Recent Developments and Future Perspective on Electrochemical Glucose Sensors Based on 2D Materials. Biosensors. 2022; 12(7):467. https://doi.org/10.3390/bios12070467

Chicago/Turabian Style

Radhakrishnan, Sithara, Seetha Lakshmy, Shilpa Santhosh, Nandakumar Kalarikkal, Brahmananda Chakraborty, and Chandra Sekhar Rout. 2022. "Recent Developments and Future Perspective on Electrochemical Glucose Sensors Based on 2D Materials" Biosensors 12, no. 7: 467. https://doi.org/10.3390/bios12070467

APA Style

Radhakrishnan, S., Lakshmy, S., Santhosh, S., Kalarikkal, N., Chakraborty, B., & Rout, C. S. (2022). Recent Developments and Future Perspective on Electrochemical Glucose Sensors Based on 2D Materials. Biosensors, 12(7), 467. https://doi.org/10.3390/bios12070467

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop