Optimizing Low-Cost Gas Analysis with a 3D Printed Column and MiCS-6814 Sensor for Volatile Compound Detection
Abstract
:1. Introduction
2. Materials and Methods
2.1. Samples and Auxiliary Materials for Experiments
- Methanol—99.8% Methanol G.R., Lach-Ner, s.r.o., Neratovice, Czech Republic, CAS: 67-56-1, EINECS: 200-659-6;
- Ethanol—96% Ethanol, Ing. Petr Švec-PENTA s.r.o., Prague, Czech Republic, CAS: 64-17-5, EINECS: 200-578-6;
- Toluene—99% Toluene G.R., Lach-Ner, s.r.o., Neratovice, Czech Republic, CAS: 108-88-3, EINECS: 203-625-9.
- Drink “Vodka”—Vodka, GAS Familia, s.r.o., Stará Ľubovňa, Slovakia, alcohol content of 37.5% vol. Composition of the drink: very fine refined spirit of high quality and demineralized water (expected composition is 62.5% vol. water and 37.5% vol. ethanol).
- Drink “Tuzemsky”—Tuzemsky, GAS Familia, s.r.o., Stará Ľubovňa, Slovakia, alcohol content of 37.5% vol. Composition of the drink: ethanol, aroma, sugar, coloring, and plain caramel color.
- Gas from a cigarette lighter—Royce cigarette lighter (commonly reported as butane (97.5%) and propane (1.2%)).
2.2. Experimental Equipment
3. Introduction to Optimization and Measurement with Experimental Equipment
- Capillary—material, length, diameter, cross-sectional shape, capillary shape, capillary internal surface properties (so-called stationary phase type), etc.;
- Sensor (detector)—type, the number of types of analytes that the sensor detects, sensitivity, speed, temperature, etc.;
- Measuring chamber—location of the sensor in the chamber, shape of the chamber, dead volume, material, analyte flow, location, and size of the inlet/outlet, etc.;
- Mixing chamber—volume, shape, inlet/outlet;
- Other specific parameters and characteristics of the experimental apparatus.
- Sample—volume, concentration, time and rate of dosing, sample preparation and collection, etc.;
- Carrier gas—gas type, flow rate, purity, etc.;
- Temperature—Temperature of the various essential parts of the equipment;
- Analysis time—duration of measurement, optimization for peak resolution, etc.;
- Calibration—selection of standards used for qualitative and quantitative analysis;
- Result processing—selection of the order of mathematical operations to be performed on the measured raw data;
- Other specific parameters and properties of the measurement.
4. Results and Discussion
4.1. Selected Device Optimization Processes
- Minimizing the dead space (to prevent analyzed gas from becoming trapped in the measuring chamber and causing measurement errors).
- Creating a path for the gas to pass through the chamber (to bring the analyte from the capillary to the sensor quickly and to drain it sufficiently fast).
- Sealing the chamber properly.
- Clean chamber (Figure 3a)—default empty version.
- Chamber with U-channel and baffle (Figure 3b)—the U-channel was made by cutting out the wall of a 5 mL syringe. The baffle was made from a rubber tube and placed in the gas path upstream of the sensor. Even though the U channel was well flushed with the sensor circuit board, there was still a lot of dead space in the chamber. In addition, the U-channel was slightly deformed (found after opening the chamber).
- Chamber with a U-shaped channel and narrowing in the sensor area (Figure 3c)—a taper was inserted into the U-channel to direct the flow of analyte into the sensor (to the sensor elements through the protective metal grid) and back again. However, in practical measurements, the response was reduced—presumably, the metal grid of the sensor was clogged by the created constriction (the sensor was probably too small and the constriction too large).
- U-channel chamber with a taper in the sensor area and cutouts to improve analyte drainage (Figure 3d)—similar to the previous version with notches in the U-channel wall for faster analyte drainage from the sensor. The narrowing is shifted behind the sensor.
- A chamber with a layered green liner with a shaped channel containing rubber leak guard (Figure 3e)—currently, the final version has a channel shaped from individual layers of smooth PET film stacked on top of each other and two rubber bands sealing the channel area from the rest of the compartment.
4.2. Selected Measurement Optimization Processes
- Setting the data acquisition program and mechanism to the pre-start position.
- Manually flushing the capillary with clean air from a 165 mL syringe (often several times).
- Inserting a syringe containing a defined volume of clean air into the device.
- Adjusting the carrier gas mechanism to the start position with the carrier gas pusher mechanism against the syringe plunger (eliminating dead time).
- Removing the syringe sample and inserting it into the dispensing mechanism.
- Starting the measurement (0 s)—measuring without activating the motors to establish the system baseline.
- Automatic start of the push mechanism motors (10 s).
- Automatic sample pitch (40 s).
- Self-measurement.
- Manually terminating the measurement (often after the carrier gas dose in the syringe is depleted).
- Returning the pusher mechanism to the pre-start position.
- Manually flushing the capillary with clean air from the 165 mL syringe to remove any sample residue (often several times, often combined with step 2 before the next measurement).
- Data recording, storage, and processing.
- Carrier gas:
- Type of carrier gas—air (determined by minimum cost);
- Purity of carrier gas—determined by the simple filter used in line with the minimum cost.
- Temperature—the temperature of the individual basic components of the device was kept at room temperature because changing and controlling it would make the experimental device more expensive. In addition, the column block material could not be heated to the temperature typically used for gas chromatographs (generally around 200–250 °C).
- Calibration standards—These are determined by the substances to which the exact type of sensor responds, as specified by the manufacturer. In our case, ethanol, food-grade methanol, and toluene were used as calibration standards (see the section on materials used). In small-scale production or even domestic settings, this selection is limited. One of the most readily available standards for qualitative analysis appears to be an alcoholic disinfectant with an ethanol content of up to 90%, or an alcoholic beverage such as vodka, which typically contains around 38% ethanol. Another readily available solution is spirit vinegar, with an acetic acid content of approximately 8%. However, the sensor data sheet (MiCS-6814) does not specify its response to the vapors of this solution; moreover, the solution may contain other additives to which the sensor could react. In addition, other substances can be used as calibration standards and can be purchased in the general commerce. However, their sale may be restricted (e.g., for toluene diluent), or their exact concentration and composition may be unknown (e.g., butane from a cigarette lighter, CO2 produced by a siphon, or propane–butane from camping fuel canisters). Other gasses, such as helium for filling children’s balloons, can be commonly purchased, but the sensor used must be sensitive to them. Under standard laboratory conditions, it is then possible to work with other calibration standards.
- Sample volume and dosage—A syringe with a 2 mL capacity was selected for sample injection. This syringe is a commonly available small-volume model compatible with a standard needle providing a sufficiently long plunger to facilitate dispensing within an accessible system. The standard volume used in this research was 1 mL or 2 × 0.7 mL. Previously, the full range allowed by the dispensing device (approximately 1.5 mL) was utilized. However, as mentioned above, dosing affects the pressure in the measuring system, to which the sensor is sensitive. Therefore, it is anticipated that a larger analyte dose volume will result in more significant pressure changes and, consequently, greater interference in the signals detected by the system. On the other hand, a smaller dose (below 0.5 mL) may be insufficient to break through the separating thin polyethylene film and penetrate the measuring system at the appropriate time.
- Carrier gas flow rate—It has a direct impact on the duration of the measurement. It is one of the limiting factors for this device; a minor trade-off is that the carrier gas can be repeatedly obtained for free by manually drawing it into the syringe without the use of additional energy sources and materials. Since the carrier gas is obtained from a syringe of a certain volume, measurements can only be performed until that volume is depleted. Thus, the higher the flow rate, the shorter the measurement time. However, the device is equipped with the capability to accommodate two syringes, allowing for the combination of their volumes. Alternatively, it is possible to use a slow flow rate during the initial phase of the measurement and a higher flow rate during the later phase.In practice, it was possible to vary the flow rate of the gas as follows:
- ○
- By adjusting the gearbox as follows:
- ▪
- Quick gear—originally designed to return the mechanism to its base position. However, due to the impracticality of constantly loosening and tightening the screws, this idea was abandoned. During test measurements in the standard configuration with the motor set to 12 V, the system pressure was already so high that the teeth on the motor axle began to skip, threatening to destroy the gearbox. When the voltage was reduced to 6 V on the motor, the measurement took about 240 s. In this configuration, using two 165 mL syringes, the carrier gas flow rate was approximately 330 mL over 240 s, which equates to s = 1.375 mL/s.
- ▪
- Slow gear—currently used for both measuring and returning the mechanism to its base position.
- ○
- Voltage to the pushing mechanism motor—The voltage to the motor is primarily drawn from an external adjustable source (0–20 V) and can be varied from 5 V to 12 V for measurement purposes. At lower voltages (bellow 5 V), current is supplied directly from the microcomputer board and USB port, which poses a risk of overloading the microcomputer’s power supply. The operating voltage at the push–pull motor of the feed motor is specified by the manufacturer at 12 V. It can be temporarily increased to 13 V for short periods. Further motor control (e.g., by PWM modulation) has not yet been tested.
- Syringe volume change—For measurement, syringes with standard volume 20 mL (maximum volume 24 mL), 50 mL (maximum volume 60 mL), and 150 mL (maximum volume 165 mL) were used. The maximum volumes were utilized for the measurements. Syringes with a small volume but longer length are advantageous for achieving low flow rates. Furthermore, a dual syringe holder provides additional flexibility by combining different syringe sizes. This is particularly useful, for example, for increasing the flow rate at the end of the measurement and for capillary flushing (enabling signal control and time saving). The potential disadvantage is the sensor’s reaction to a change in flow rate (see Figure 9).
- Use of NO2 sensor signal for noise elimination—The NO2 sensor is mainly unresponsive to food samples at low concentrations (it primarily responds to NO and NO2) but significantly reflects the same noise as the other two sensors, as shown in Figure 11 below (e.g., at time 2025 s). Further calculation is than performed. For example, to eliminate the noise from the CO sensor signal, the following equation is used:
- Signal averaging—After noise elimination, centered moving averaging is used, most often with parameter m = 11; see Figure 12. This parameter was also used in previous circuits. Since the sampling rate in this version was four times slower, the averaging covered a section four times longer. Therefore, parameters similar to the earlier time span (m = 41 or m = 45) were also tested, as well as the left moving average, to avoid taking into account future samples that will be measured after the current count. However, this approach could potentially result in the loss of some detail or cause an undesirable time shift in the final characteristic.
4.3. Results Achieved with the Optimized System
- 0.7 mL of air + 0.7 mL of air (labeled Air, used for standardization);
- 0.7 mL of methanol + 0.7 mL of air (labeled Met);
- 0.7 mL of air + 0.7 mL of ethanol (labeled Eta);
- 0.7 mL of methanol + 0.7 mL of ethanol (labeled MaE).
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- Sahana, V.W.; Thampi, G.T. 3D Printing Technology in Industry. In Proceedings of the 2nd International Conference on Inventive Systems and Control (ICISC), Coimbatore, India, 19–20 January 2018. [Google Scholar]
- Janigova, N.; Gajdziok, J.; Vetchy, D. Biomedical Use of 3D Printing (orig. in Slovak, Biomedicínske použitie 3D tlače). Chem. Listy 2021, 115, 356–362. [Google Scholar]
- Noor, M.N.M.; Leow, M.L.; Lai, W.H.; Hon, Y.K.; Tiong, L.L.; Chern, P.M. Research landscape on 3D printing applications in healthcare within South-east Asian countries: A systematic scoping review protocol. BMJ Open 2022, 12, e065546. [Google Scholar] [CrossRef]
- Zhou, S.; Zhao, Y.; Xun, Y.; Wei, Z.; Yang, Y.; Yan, W.; Ding, J. Programmable and modularized gas sensor integrated by 3D printing. Chem. Rev. 2024, 124, 3608–3643. [Google Scholar] [CrossRef]
- Song, D.; Chen, X.; Wang, M.; Wu, Z.; Xiao, X. 3D-printed flexible sensors for food monitoring. Chem. Eng. J. 2023, 474, 146011. [Google Scholar] [CrossRef]
- Muhamad, O.; Khairunisak, R.; Nadhir, W.; Hairul, H. Recent progress of conductive 3D-printed electrodes based upon polymers/carbon nanomaterials using a fused deposition modelling (FDM) method as emerging electrochemical sensing devices. RSC Adv. 2021, 11, 16557–16571. [Google Scholar] [CrossRef]
- Moleirinho, M.G.; Feast, S.; Moreira, A.S.; Silva, R.J.S.; Alves, P.M.; Carrondo, M.J.T.; Huber, T.; Fee, C.; Peixoto, C. 3D-printed ordered bed structures for chromatographic purification of enveloped and non-enveloped viral particles. Sep. Purif. Technol. 2021, 254, 117681. [Google Scholar] [CrossRef]
- Lee, J.Y.; An, J.; Chua, C.K. Fundamentals and applications of 3D printing for novel materials. Appl. Mater. Today 2017, 7, 120–133. [Google Scholar] [CrossRef]
- Poologasundarampillai, G.; Nommeots-Nomm, A. Materials for 3D printing in medicine: Metals, polymers, ceramics, hy-drogel. In 3D Printing in Medicine; Kalaskar, D.M., Ed.; Woodhead Publishing: Sawston, UK, 2017; pp. 43–71. ISBN 978-0-08-100717-4. [Google Scholar] [CrossRef]
- Zhou, L.Y.; Fu, J.; He, Y. A Review of 3D Printing Technologies for Soft Polymer Materials. Adv. Funct. Mater. 2020, 30, 2000187. [Google Scholar] [CrossRef]
- Phyo, S.; Choi, S.; Jang, J.; Choi, S.; Lee, J. A 3D-printed metal column for micro gas chromatography. Lab Chip 2020, 20, 3435–3444. [Google Scholar] [CrossRef]
- Meng, H.; Wei, Y.; Feng, L. A microchip gas chromatography column assembly with a 3D metal printing micro column oven and a flexible stainless-steel column. J. Chromatogr. A 2024, 1729, 465036. [Google Scholar] [CrossRef]
- Nawada, S.; Budel, T. A Novel 3D-Printing Method to Create Liquid Chromatography Columns. LCGC Eur. 2021, 34, 381–386. [Google Scholar]
- Salmean, C.; Dimartino, S. 3D-Printed Stationary Phases with Ordered Morphology: State of the Art and Future Development in Liquid Chromatography. Chromatographia 2019, 82, 443–463. [Google Scholar] [CrossRef]
- Adamek, M.; Mlcek, J.; Skowronkova, N.; Zvonkova, M.; Jasso, M.; Adamkova, A.; Skacel, J.; Buresova, I.; Sebestikova, R.; Cernekova, M.; et al. 3D Printed Fused Deposition Modeling (FDM) Capillaries for Chemiresistive Gas Sensors. Sensors 2023, 23, 6817. [Google Scholar] [CrossRef]
- Zvonkova, M.; Adamek, M.; Skowronkova, N.; Dlabaja, S.; Matyas, J.; Jasso, M.; Adamkova, A.; Mlcek, J.; Salek, R.N.; Buran, M. Compact 3D-Printed Unit for Separation of Simple Gas Mixtures Combined with Chemiresistive Sensors. Sensors 2024, 24, 4391. [Google Scholar] [CrossRef]
- SGX Sensortech. MiCS-6814, 1143 Rev 8. Data Sheet. Available online: https://www.sgxsensortech.com/content/uploads/2015/02/1143_Datasheet-MiCS-6814-rev-8.pdf (accessed on 15 July 2024).
- van den Broek, J.; Abegg, S.; Pratsinis, S.E.; Güntner, A.T. Highly selective detection of methanol over ethanol by a handheld gas sensor. Nat. Commun. 2019, 10, 4220. [Google Scholar] [CrossRef]
- Che Harun, F.K.; Taylor, J.E.; Covington, J.A.; Gardner, J.W. An electronic nose employing dual-channel odour separation columns with large chemosensor arrays for advanced odour discrimination. Sens. Actuators B Chem. 2009, 141, 134–140. [Google Scholar] [CrossRef]
- Joseph, J.A.; Akkermans, S.; Van Impe, J.F.M. Processing Method for the Quantification of Methanol and Ethanol from Biore-actor Samples Using Gas Chromatography–Flame Ionization Detection. ACS Omega 2022, 7, 24121–24133. [Google Scholar] [CrossRef]
- Itterheimova, P.; Kuban, P. An open source 3D printed autosampler for capillary electrophoresis. Anal. Chim. Acta. 2023, 1279, 341832. [Google Scholar] [CrossRef]
- Radi; Ciptohadijoyo, S.; Litananda, W.S.; Rivai, M.; Purnomo, M.H. Electronic nose based on partition column integrated with gas sensor for fruit identification and classification. Comput. Electron. Agric. 2016, 121, 429–435. [Google Scholar] [CrossRef]
- Hayasaka, T.; Lin, A.; Copa, V.C.; Lopez, L.P., Jr.; Loberternos, R.A.; Ballesteros, L.I.M.; Kubota, Y.; Liu, Y.; Salvador, A.A.; Lin, L. An electronic nose using a single graphene FET and machine learning for water, methanol, and ethanol. Microsyst Nanoeng 2020, 6, 50. [Google Scholar] [CrossRef]
- Palkowitsh, J. Nitrogen Gas in the Food Process. Compressed Air Best Practices®, 2018. Available online: https://www.airbestpractices.com/industries/food/nitrogen-gas-food-process (accessed on 15 July 2024).
- Zeleny, P. Forty-Eight Dead, Dozens Accused. Ten Years Have Passed Since the Methanol Case (orig. in Czech, Osmactyricet mrtvych, desitky obvinenych. Od metanolove kauzy uplynulo deset let. CT24, 2022. Available online: https://ct24.ceskatelevize.cz/clanek/domaci/osmactyricet-mrtvych-desitky-obvinenych-od-metanolove-kauzy-uplynulo-deset-let-16710 (accessed on 15 July 2024).
- Kar, P.; Pradhan, N.C.; Adhikari, B. Application of sulfuric acid doped poly (m-aminophenol) as aliphatic alcohol vapor sensor material. Sens. Actuators B Chem. 2009, 140, 525–531. [Google Scholar] [CrossRef]
- King, H.S.; Bell, A.C. The detection of methanol in the presence of ethyl alcohol. Proc. Nova Scotian Inst. Sci. 1932, 18, 11–13. [Google Scholar]
- Liu, W.-H.; Yang, W.; Wu, X.-Q.; Lin, Z.-X. Direct determination of ethanol by laser Raman spectra with internal standard method. Fenxi Huaxue/Chin. J. Anal. Chem. 2007, 35, 416–418. [Google Scholar]
- Raman Spectrometer. Available online: https://www.labx.com/product/raman-spectrometer (accessed on 15 July 2024).
- Hashemi, S.A.; Bahrani, S.; Mousavi, S.M.; Omidifar, N.; Arjmand, M.; Lankarani, K.B.; Shokripour, M.; Ramakrishna, S. Differentiable detection of ethanol/methanol in biological fluids using prompt graphene-based electrochemical nanosensor coupled with catalytic complex of nickel oxide/8-hydroxyquinoline. Anal. Chim. Acta 2022, 1194, 339407. [Google Scholar] [CrossRef]
- Barrios, C.A. Scotch Tape Optical Vapor Sensor for Ethanol–Methanol Mixtures. Sensors 2019, 19, 5381. [Google Scholar] [CrossRef]
Transmission | Voltage on Motor/Time to Empty Syringe Volume | Number and Size of Syringe | Carrier Gas Flow Rate |
---|---|---|---|
Quick | 12 V/(Not determined) | 2 × 165 mL | Motor overload |
6 V/240 s | 2 × 165 mL | 1.375 mL/s | |
Slow | 12 V/775 s (cca 3100 samples) | 2 × 165 mL | 0.4258 mL/s |
12 V/775 s (cca 3100 samples) | 1 × 165 mL | 0.2129 mL/s | |
5 V/2525 s (cca 10,100 samples) | 2 × 165 mL | 0.1307 mL/s | |
5 V/2525 s (cca 10,100 samples) | 1 × 165 mL | 0.0653 mL/s | |
5 V/1800 s (cca 7200 samples) | 2 × 60 mL | 0.0666 mL/s | |
5 V/1800 s (cca 7200 samples) | 1 × 60 mL | 0.0333 mL/s | |
5 V/1 357.5 s (cca 5430 samples) | 2 × 24 mL | 0.0354 mL/s | |
5 V/1 357.5 s (cca 5430 samples) | 1 × 24 mL | 0.0177 mL/s |
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content. |
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Skowronkova, N.; Adamek, M.; Zvonkova, M.; Matyas, J.; Adamkova, A.; Dlabaja, S.; Buran, M.; Sevcikova, V.; Mlcek, J.; Volek, Z.; et al. Optimizing Low-Cost Gas Analysis with a 3D Printed Column and MiCS-6814 Sensor for Volatile Compound Detection. Sensors 2024, 24, 6594. https://doi.org/10.3390/s24206594
Skowronkova N, Adamek M, Zvonkova M, Matyas J, Adamkova A, Dlabaja S, Buran M, Sevcikova V, Mlcek J, Volek Z, et al. Optimizing Low-Cost Gas Analysis with a 3D Printed Column and MiCS-6814 Sensor for Volatile Compound Detection. Sensors. 2024; 24(20):6594. https://doi.org/10.3390/s24206594
Chicago/Turabian StyleSkowronkova, Nela, Martin Adamek, Magdalena Zvonkova, Jiri Matyas, Anna Adamkova, Stepan Dlabaja, Martin Buran, Veronika Sevcikova, Jiri Mlcek, Zdenek Volek, and et al. 2024. "Optimizing Low-Cost Gas Analysis with a 3D Printed Column and MiCS-6814 Sensor for Volatile Compound Detection" Sensors 24, no. 20: 6594. https://doi.org/10.3390/s24206594
APA StyleSkowronkova, N., Adamek, M., Zvonkova, M., Matyas, J., Adamkova, A., Dlabaja, S., Buran, M., Sevcikova, V., Mlcek, J., Volek, Z., & Cernekova, M. (2024). Optimizing Low-Cost Gas Analysis with a 3D Printed Column and MiCS-6814 Sensor for Volatile Compound Detection. Sensors, 24(20), 6594. https://doi.org/10.3390/s24206594