1. Introduction
Zirconia ceramics used in dental implants are mostly 3Y-TZPs, standing for tetragonal zirconia polycrystal, stabilized in the tetragonal phase by 3 mol.% Y
2O
3. This ceramic can exist at room temperature in two main phases: the stable monoclinic phase and the metastable tetragonal phase. It is important to preserve the metastable tetragonal zirconia phase after processing since this phase offers the highest mechanical properties (fracture strength up to 1.5 GPa as an order of magnitude). However, the tetragonal phase is prone to a tetragonal-to-monoclinic transformation in the presence of water. The martensitic tetragonal-to-monoclinic (t–m) phase transformation results in a 5–7% volume increase in the transformed grains [
1]. In the presence of water, the t–m transformation proceeds from the surface to the bulk, accompanied by roughening and microcracking, and is called hydrothermal ageing or low temperature degradation (LTD).
The effects of ageing on mechanical properties can be diverse. Indeed, the most expected effect would be a strength decrease due to the microcracked surface. This was indeed observed in some materials [
2]. The strength decrease can even reach a complete loss of integrity, as summarized in the reviews by Piconi [
3] and Ramesh (Table 7 of ref [
4]). The same microcracking may lead to delamination of the aged surface layer [
5]. However, the volume increase associated with the t–m transformation may also induce compressive stresses around the transformed zone that may increase the measured strength [
3,
4,
6,
7]. Whatever the stress state of the near-surface area, both elastic modulus and hardness of the aged surfaces may be decreased [
8,
9], which is due to the microcracking more than to the presence of the monoclinic phase itself [
10].
At room or body temperature, hydrothermal ageing is a rather slow phenomenon. To be conveniently observable, zirconia ageing kinetics need to be measured at high temperatures (typically above 100 °C). It thus becomes necessary to establish a correspondence between the ageing kinetics measured at high temperature and the ones expected at body temperature. Zirconia ageing kinetics may be described by fitting the evolution of the monoclinic fraction (
f) with time (
t) by a Mehl–Avrami–Johnson (MAJ) Equation:
where
fmax is the saturation monoclinic volume fraction,
fmin is the initial monoclinic volume fraction, and
n is a constant (the “MAJ exponent”).
b is a thermally activated parameter; in the 37–145 °C temperature range it follows an Arrhenius law:
where
b0 is a constant (pre-exponential term),
R is the gas constant,
Q is the apparent activation energy, and
T is the absolute temperature [
11].
Knowing the activation energy
Q enables one to establish a time–temperature equivalence for the tetragonal-to-monoclinic transformation due to hydrothermal ageing. As shown in Equation (3) (derived from Equations (1) and (2)), the same monoclinic fraction is reached at the (
t1,
T1) and (
t2,
T2) (time, temperature) couples if
This is particularly useful to establish predictive ageing kinetics at body temperature from high temperature data (between 70 and 150 °C). However, although numerous works have measured zirconia hydrothermal ageing kinetics, only few discuss the activation energy. Even the most recent reviews neglect this parameter, even though it has been shown that it may vary greatly from one material to another, depending not only on the starting powder, but also on the sintering schedule (for example, values from 41 to 115 kJ·mol
−1 were measured on samples made from the same powder) [
12]. In most cases, activation energies for hydrothermal ageing lie between 100 and 120 kJ·mol
−1, resulting in a correspondence of 1 h in an autoclave at 134 °C for 1 to 4 years in vivo at 37 °C.
Therefore, assessing the durability of zirconia dental implants related to hydrothermal ageing relies on several steps. First, the ageing kinetics have to be measured at several temperatures so that the constants related to the MAJ Equation, in particular the activation energy, can be deduced, to enable extrapolation of the ageing kinetics to body temperature (37 °C). Second, in order to ensure that no fracture of implant can occur due to regular ageing, an assessment of the mechanical properties (mainly strength) has to be conducted before and after ageing (for ageing times corresponding to relevant durations in vivo). Third, it is interesting to assess the evolution of surface properties in order to ensure that no delamination can occur, in particular in the parts of the implants in contact with bone.
In the present work, this procedure was applied to dental one-piece zirconia implants (Straumann® PURE Ceramic Implant, Institut Straumann AG, Basel, Switzerland). More specifically, the constants related to their ageing kinetics according to the MAJ Equation (Q, b0, and n) were determined in order to deduce their ageing kinetics at body temperature from accelerated ageing kinetics in the laboratory. Then, their load to fracture after ageing in conditions equivalent to up to 90 years at 37 °C were measured. Finally, scratch tests were conducted to detect potential surface degradation.
The results show that the activation energy of these implants is in the usual range (around 100 kJ·mol−1). No negative influence of hydrothermal ageing on the load to fracture of these zirconia implants nor on their surface mechanical properties were detected.
2. Materials and Methods
2.1. Microstructural Observations
Fractographic observations were conducted on fracture surfaces of as-received and aged samples using a scanning electron microscope (Supra 55VP, Carl Zeiss AG, Oberkochen, Germany) without coating the samples. Imaging was performed at a low acceleration voltage (1.5 or 2 kV) with an Everhart-Thornley detector for secondary electrons. Chemical composition was assessed by Energy Dispersive X-ray Spectroscopy (EDS) analyses, performed at 25 kV acceleration voltage in partial vacuum mode (20 torr), using an Oxford SDD spectrometer with Aztec software (Oxford Instruments, High Wycombe, UK).
2.2. Measurement of Ageing Kinetics
The ageing kinetics of 20 dental two-piece zirconia implants (10 mm length, 4.1 mm diameter) (Straumann® PURE Ceramic Implant, Institut Straumann AG, Basel, Switzerland) were evaluated at 5 different temperatures (70, 110, 134, 145 °C) and for different durations of up to 4657 h. The equipment used for ageing was chosen according to the necessary temperature and times (indeed, it was not possible to conduct long experiments inside autoclaves, due to limits in their programming capacities and availability). For ageing at 110, 134, and 145 °C, the samples were placed in open water vials inside an autoclave (SanoClav LA-MCS, Wolf, Geislingen/Steige, Germany). For ageing at 95 °C, the samples were placed in a closed water vial inside a thermo-regulated oven (VT5050EK, Heraeus, Hanau, Germany). For ageing at 70 °C, the samples were placed in glass dishes filled with deionized water inside a thermo-regulated water bath (Polytest 20, Fisher Scientific, Illkirch, France). In all cases, the samples were in contact with deionized water during the ageing process, so that all experiments were conducted in the same conditions with regard to contact with water. The temperature sensors of the equipment were checked before ageing.
The implants were periodically taken out, cooled, and dried so that their monoclinic fraction could be analyzed by X Ray Diffraction (XRD). The fraction of the monoclinic phase on the surface of the implants was measured by XRD, in Bragg–Brentano configuration with Cu Kα radiation in a θ–2θ mode (2 θ ∈ [26–33°]), with a scan speed of 0.5°/min and step size of 0.05° (Geigerflex diffractometer, Rigaku, Tokyo, Japan). The samples were positioned vertically in the diffractometer, so that the whole length of the implant was analyzed at once.
The monoclinic content was then calculated using Garvie and Nicholson’s Equation [
13]:
where
Xm is the integrated intensity ratio,
Im(hkl) is the area of the (h k l) peak of the monoclinic phase, and
It(hkl) is the area of the (h k l) peak of the tetragonal phase (peak intensities were measured using the EVA software (Bruker AXS, Karlsruhe, Germany)). The experimental volume fraction (
f) of the monoclinic phase (average over the penetration depth of the X-rays: for CuK
α radiations, 90% of the signal comes from the first 5 µm below the surface) was then determined by Toraya’s Equation [
14]:
2.3. Assessment of Fracture Loads
Eighty dental one-piece zirconia implants (14 mm length, 3.3 mm diameter) (Straumann® PURE Ceramic Implant, Institut Straumann AG, Basel, Switzerland) and 20 titanium implants (14 mm length, 3.3 mm diameter) (Standard Plus Implants, Institut Straumann AG, Basel, Switzerland) were supplied by Straumann for the assessment of their fracture load. Before fracture tests, some samples were aged in an autoclave at 134 °C for different durations to mimic in vivo ageing duration of 22.5, 45, and 90 years at 37 °C.
Before mechanical testing, the samples were embedded in 1 cm diameter × 1 cm height cylinders made of epoxy resin with a Young’s modulus of 5.8 GPa (RenCast CW 5156–1/Ren HY 5158, hardened at 55 °C during 15.5 h) (Huntsman Advanced Materials, Basel, Switzerland). The embedding depth was determined to ensure that the geometrical requirements of the ISO 14801 standard [
15] were met.
The fracture tests were carried out on an Instron 8872 universal testing machine (Instron, Elancourt, France) using a procedure based on ISO 14801. The experiments were conducted in air, at a crosshead speed of 1 mm/min (fast enough to avoid slow crack growth during loading), until fracture of the samples. The fracture load was then determined as the maximum of the load-displacement curve.
2.4. Scratch Tests
Three zirconia cylinders machined in the same conditions as zirconia implants were used for a scratch test. The samples were aged similarly to the ones used for fracture load measurement (one sample each). Scratch tests were conducted in a self-built scratch tester. Briefly, an alumina pin with a spherical surface was pressed on the top generatrix of the cylindrical samples and dragged along this generator at a fixed speed (15 mm·s−1) and with a constant normal load (10 mN). The tangential load necessary to maintain the desired speed was measured, then the friction coefficient was deduced as the ratio between tangential and normal load. This experiment was repeated at least three times on each sample, each time rotating the sample to test a different generatrix.
4. Discussion
The microstructure shows a bimodal grain-size distribution, with few large grains in a matrix of fine tetragonal grains. This is most likely due to a partitioning of yttrium during sintering (following the yttria–zirconia phase diagram), which gave rise to yttrium-rich cubic grains (large grains) and (slightly) yttrium-depleted tetragonal grains. Although this depletion of yttrium of the tetragonal phase can be expected to be detrimental to ageing (tetragonal phase being less stable), this was not observed here.
On the contrary, the potentially lesser stability may explain the good mechanical properties obtained on zirconia PURE ceramic implants. For comparison of the strength of the implants analyzed here with other available results on similar implants (from competitors or from previous results on the same implants), it is necessary to take into account the geometry of the implants, particularly their diameters. Assuming that, during mechanical tests, the implants behave like cylinders submitted to flexion, a rough approximation of their strength (
) can be determined from their load to fracture (
F) by Equation (7):
In this equation,
D is the nominal diameter of the implant,
L the solicited length (11 mm, to respect ISO 14801), and
is the angle between the implant axis and the loading direction (also chosen in agreement with ISO 14801). This value of “approximate strength” is thus shown on the sixth column of
Table 3 for different implants. As in first approximation it is independent from the geometry of the implants, it should reflect only the quality of the material and of the machining.
Even though this comparison is limited (for example, it does not account for the depth and shape of the threads), it shows that the 3.3 mm diameter zirconia implants from Straumann have a similar strength to the 4.1 mm diameter ones from the same manufacturer. The results on zirconia PURE ceramic implants obtained in this study are very similar to the ones obtained internally by Straumann on other batches of implants [
18], which is a good sign for the reproducibility of the material and of the machining procedure.
The influence of ageing on the mechanical properties is not obvious. The ANOVA did not show any statistically significant difference between the loads to fracture of the zirconia implants aged at different temperatures. Even though the Weibull modulus seems higher at 22.5 years ageing, the significance of this difference is doubtful given the small number of samples (20 per ageing time). In addition, plotting all zirconia samples together in a single Weibull plot shows a monomodal distribution (m = 6.6). Thus, all analyses point to the conclusion that up to 100 years (or rather, 89.9 years—see
Table 2) hydrothermal ageing does not influence the fracture load of these dental implants. The same was observed for the friction coefficient, which remains the same whatever the ageing time.
In Y-TZP materials, the role of yttria is to prevent the spontaneous tetragonal-to-monoclinic phase transition by stabilizing the tetragonal phase. However, in the presence of water, hydrothermal ageing counteracts this stabilization. Several possible mechanisms have been identified. In the opinion of the authors, the most probable one is the occupation of oxygen vacancies by hydroxyl ions, which destabilizes the tetragonal phase and overstabilizes the monoclinic one, therefore leading to the t–m phase transformation [
19]. It is important to keep in mind, however, that whatever its mechanism, the tetragonal-to-monoclinic phase transformation in the presence of water always progresses from the surface exposed to water to the bulk, being controlled by the diffusion of water and hydroxyl ions into the material. The transformed layer may be seen as a defect that could negatively impact the strength of aged specimens.
However, the tetragonal-to-monoclinic phase transformation results in a large volume increase (5–7%), which imposes compressive stresses to the near-surface [
3,
4,
5,
6,
7]. These compressive stresses should strengthen the material.
As a result, the evolution of strength with hydrothermal ageing results from a competition between the depth of the transformed zone (usually rather small [
20]) and the presence of compressive stresses. Indeed, some studies demonstrate that an increase in strength with ageing time is possible [
5]. Here, no effect of the transformation on strength can be seen. The probable explanation is that an equilibrium was reached between the effects of transformation depth and compressive stresses.