Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study
Abstract
:1. Introduction
2. Materials and Methods
2.1. Sample Size Determination
2.2. Adhesive Preparation
2.3. Sample Preparation
2.4. Fabrication of Crown for Cyclic Loading
2.5. Cyclic Loading
2.6. Measuring the Removal Torque Value
2.7. Measuring the Rotational Misfit
2.8. Assessing the Biocompatibility
2.9. Statistical Analysis
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
References
- La Monaca, G.; Pranno, N.; Annibali, S.; Massimo, C.; Polimeni, A.; Patini, R.; Paola Cristalli, M. Survival and complication rates of tooth-implant versus freestanding implant supporting fixed partial prosthesis: A systematic review and meta-analysis. J. Prosthodont. Res. 2021, 65, 1–10. [Google Scholar] [CrossRef]
- González-Valls, G.; Roca-Millan, E.; Céspedes-Sánchez, J.M.; González-Navarro, B.; Torrejon-Moya, A.; López-López, J. Narrow diameter dental implants as an alternative treatment for atrophic alveolar ridges. Systematic review and meta-analysis. Materials 2021, 14, 3234. [Google Scholar] [CrossRef]
- Oh, S.L.; Shiau, H.J.; Reynolds, M.A. Survival of dental implants at sites after implant failure: A systematic review. J. Prosthet. Dent. 2020, 123, 54–60. [Google Scholar] [CrossRef] [Green Version]
- Delucchi, F.; De Giovanni, E.; Pesce, P.; Bagnasco, F.; Pera, F.; Baldi, D.; Menini, M. Framework Materials for Full-Arch Implant-Supported Rehabilitations: A Systematic Review of Clinical Studies. Materials 2021, 14, 3251. [Google Scholar] [CrossRef]
- Binon, P.P. Implants and components: Entering the new millennium. Int. J. Oral Maxillofac. Implant. 2000, 15, 76–94. [Google Scholar]
- Alnasser, A.H.; Wadhwani, C.P.K.; Schoenbaum, T.R.; Kattadiyil, M.T. Evaluation of implant abutment screw tightening protocols on reverse tightening values: An in vitro study. J. Prosthet. Dent. 2021, 125, 486–490. [Google Scholar] [CrossRef] [PubMed]
- Guda, T.; Ross, T.A.; Lang, L.A.; Millwater, H.R. Probabilistic analysis of preload in the abutment screw of a dental implant complex. J. Prosthet. Dent. 2008, 100, 183–193. [Google Scholar] [CrossRef]
- Bindl, A.; Mormann, W.H. Marginal and internal fit of all-ceramic CAD/CAM crown-copings on chamfer preparations. J. Oral Rehabil. 2005, 32, 441–447. [Google Scholar] [CrossRef]
- Winkler, S.; Ring, K.; Ring, J.D.; Boberick, K.G. Implant screw mechanics and the settling effect: Overview. J. Oral Implantol. 2003, 29, 242–245. [Google Scholar] [CrossRef]
- Huang, Y.; Wang, J. Mechanism of and factors associated with the loosening of the implant abutment screw: A review. J. Esthet. Restor. Dent. 2019, 31, 338–345. [Google Scholar] [CrossRef] [PubMed]
- Paepoemsin, T.; Reichart, P.A.; Chaijareenont, P.; Strietzel, F.P.; Khongkhunthian, P. Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading. Oral Implant. 2016, 9, 213–221. [Google Scholar]
- Norton, M.R. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J. Prosthet. Dent. 1999, 81, 159–166. [Google Scholar] [CrossRef]
- Sahin, C.; Ayyildiz, S. Correlation between microleakage and screw loosening at implant-abutment connection. J. Adv. Prosthodont. 2014, 6, 35–38. [Google Scholar] [CrossRef] [Green Version]
- Eckert, S.E.; Wollan, P.C. Retrospective review of 1170 endosseous implants placed in partially edentulous jaws. J. Prosthet. Dent. 1998, 79, 415–421. [Google Scholar] [CrossRef]
- Artzi, Z.; Dreiangel, A. A screw lock for single-tooth implant superstructures. J. Am. Dent. Assoc. 1999, 130, 677–682. [Google Scholar] [CrossRef]
- Cavazos, E.; Bell, F.A. Preventing loosening of implant abutment screws. J. Prosthet. Dent. 1996, 75, 566–569. [Google Scholar] [CrossRef]
- Arshad, M.; Shirani, G.; Refoua, S.; Rahimi Yeganeh, M. Comparative study of abutment screw loosening with or without adhesive material. J. Adv. Prosthodont. 2017, 9, 99–103. [Google Scholar] [CrossRef] [PubMed]
- Alikhasi, M.; Monzavi, A.; Bassir, S.H.; Naini, R.B.; Khosronedjad, N.; Keshavarz, S. A comparison of precision of fit, rotational freedom, and torque loss with copy-milled zirconia and prefabricated titanium abutments. Int. J. Oral Maxillofac. Implant. 2013, 28, 996–1002. [Google Scholar] [CrossRef]
- Binon, P.P. The effect of implant/abutment hexagonal misfit on screw joint stability. Int. J. Prosthodont. 1996, 9, 149–160. [Google Scholar] [PubMed]
- Binon, P.P. Evaluation of three slip fit hexagonal implants. Implant Dent. 1996, 5, 235–248. [Google Scholar] [CrossRef]
- Abdous, H.; Solati Hashjin, M.; Ataei, M.; Arshad, M. Design and Formulation af a Bio-Adhesive for Controlling of “Implant/Abutment Screw Loosening; Amirkabir University of Technology-Tehran Polytechnic: Tehran, Iran, 2017. [Google Scholar]
- De Carvalho, B.A.; Vedovatto, E.; de Carvalho, P.S.; Mazaro, J.V.; Falcon-Antenucci, R.M. Effect of mechanical cycling on screw torque in external hexagon implants with and without platform switching. Braz. Dent. J. 2015, 26, 284–287. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Van Meerloo, J.; Kaspers, G.J.L.; Cloos, J. Cell sensitivity assays: The MTT assay. Methods Mol. Biol. 2011, 731, 237–245. [Google Scholar]
- Misch, C.E.; Resnik, R. Misch’s Avoiding Complications in Oral Implantology; Elsevier: St. Louis, MO, USA, 2017; pp. 39–76. [Google Scholar]
- Jung, R.E.; Pjetursson, B.E.; Glauser, R.; Zembic, A.; Zwahlen, M.; Lang, N.P. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin. Oral Implant. Res. 2008, 19, 119–130. [Google Scholar] [CrossRef] [PubMed]
- Pardal-Peláez, B.; Montero, J. Preload loss of abutment screws after dynamic fatigue in single implant-supported restorations. A systematic review. J. Clin. Exp. Dent. 2017, 9, e1355–e1361. [Google Scholar] [CrossRef]
- Lee, J.H.; Cha, H.S. Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading. J. Adv. Prosthodont. 2018, 10, 415–421. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Jank, S.; Schröderr, D.; Schaanen, F. Effects of lubrication on preload of dental implant abutment screws. Int. J. Oral Maxillofac. Surg. 2015, 44, 231–240. [Google Scholar] [CrossRef]
- Nithyapriya, S.; Ramesh, A.S.; Kirubakaran, A.; Mani, J.; Raghunathan, J. Systematic analysis of factors that cause loss of preload in dental implants. J. Indian Prosthodont. Soc. 2018, 18, 189–195. [Google Scholar] [CrossRef]
- Nigro, F.; Sendyk, C.L.; Francischone, C.E., Jr.; Francischone, C.E. Removal torque of zirconia abutment screws under dry and wet conditions. Braz. Dent. J. 2010, 21, 225–228. [Google Scholar] [CrossRef] [Green Version]
- Tzenakis, G.K.; Nagy, W.W.; Fournelle, R.A.; Dhuru, V.B. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J. Prosthet. Dent. 2002, 88, 183–191. [Google Scholar] [CrossRef]
- Gross, M.; Abramovich, I.; Weiss, E.I. Microleakage at the abutment-implant interface of osseointegrated implants: A comparative study. Int. J. Oral Maxillofac. Implant. 1999, 14, 94–100. [Google Scholar]
- Triveni, V.V.; Jagadeesh, K.N.; Parihar, A.S.; Alduwayhi, S.; Annapoorneshwari, S.; Khalid, F.M.; Babaji, P. Evaluation of microleakage and microgap of Two different internal implant-abutment connections: An in vitro study. J. Contemp. Dent. Pract. 2020, 21, 683–685. [Google Scholar] [PubMed]
- Koutouzis, T.; Wallet, S.; Calderon, N.; Lundgren, T. Bacterial colonization of the implant–abutment interface using an in vitro dynamic loading model. J. Periodontol. 2011, 82, 613–618. [Google Scholar] [CrossRef] [PubMed]
- De Oliveira, G.R.; Olate, S.; Pozzer, L.; Cavalieri-Pereira, L.; Rodrigues-Chessa, J.G.; Albergaría-Barbosa, J.R. Bacterial contamination along implant-abutment interface in external and internal-hex dental implants. Int. J. Clin. Exp. Med. 2014, 7, 580–585. [Google Scholar] [PubMed]
- Yüzügüllü, B.; Avci, M. The implant-abutment interface of alumina and zirconia abutments. Clin. Implant. Dent. Relat. Res. 2008, 10, 113–121. [Google Scholar] [CrossRef] [PubMed]
Variable | Groups | Mean | SD | p-Value |
---|---|---|---|---|
Removal torque value | Control (no adhesive) | 20.80 | 4.158 | 0.008 * |
Test (with adhesive) | 26.60 | 1.838 | ||
Rotational misfit | Control (no adhesive) | 5.293 | 0.148 | 0.001 * |
Test (with adhesive) | 4.842 | 0.414 | ||
Test (with adhesive) | 4.842 | 0.414 |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Arshad, M.; Khayat Zadeh, S.; Atai, M.; Shirani, G.; Romanos, G.E.; Bassir, S.H. Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study. Materials 2021, 14, 6832. https://doi.org/10.3390/ma14226832
Arshad M, Khayat Zadeh S, Atai M, Shirani G, Romanos GE, Bassir SH. Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study. Materials. 2021; 14(22):6832. https://doi.org/10.3390/ma14226832
Chicago/Turabian StyleArshad, Mahnaz, Sina Khayat Zadeh, Mohammad Atai, Gholamreza Shirani, Georgios E. Romanos, and Seyed Hossein Bassir. 2021. "Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study" Materials 14, no. 22: 6832. https://doi.org/10.3390/ma14226832
APA StyleArshad, M., Khayat Zadeh, S., Atai, M., Shirani, G., Romanos, G. E., & Bassir, S. H. (2021). Effect of Application of a Bio-Adhesive on the Removal Torque Value and Rotational Misfit at the Implant–Abutment Junction: An In Vitro Study. Materials, 14(22), 6832. https://doi.org/10.3390/ma14226832