3.1. Data Quality and Reduction
This section briefly describes the data selection and reduction performed to determine the radiometric fitting coefficients as well as the performance and stability metrics. The sensor specification defines the dynamic range over which VIIRS thermal bands must be calibrated [
8]. However, valid science data may exist for some bands outside the specified range, and in some cases data inside the specified range is of low quality. In this work the fitted range was modified to include all available measured data not contaminated at high temperature by saturation and for which the SNR was greater than 5 at low temperature. The dynamic ranges are shown in
Figure 1: the black lines represent the specified dynamic range; the red lines indicate the scene temperatures for which the SNR is equal to 5 (determined from Equation (9)) and the saturation temperatures; and the blue lines indicate the extent of the measurements used in the fitting. Note that the minimum and maximum BCS temperatures were 190 K and 345 K; these were in many cases limiting factors on modifying the fitting range. No band saturated below 345 K. However, the SNR fell below 5 inside the specified dynamic range for I4 (due to high noise, both 210 K and 230 K BCS levels were below the SNR threshold) and M14 (the large difference in RVS from SV to EV angle resulted in a negative offset corrected response at the lowest measured scene temperature). The lowest available measured levels that met the SNR threshold requirement for these two bands were 247 K (I4) and 210 K (M14). M13 low gain is not shown in
Figure 1, but all available M13 low gain data was used in the fitting (TMCBB at-aperture scene temperatures from 355 K to 644 K). M13 low gain was not observed to saturate in thermal vacuum testing and the SNR was well above 5 at the lowest measured temperature.
Figure 1.
Specified dynamic ranges shown for the VIIRS thermal bands (black lines). Maximum and minimum scene temperatures used in the fitting as shown with blue lines. Saturation and low temperature SNR threshold are also shown (red lines).
Figure 1.
Specified dynamic ranges shown for the VIIRS thermal bands (black lines). Maximum and minimum scene temperatures used in the fitting as shown with blue lines. Saturation and low temperature SNR threshold are also shown (red lines).
Data collections were recorded at various source temperature levels, during each of which the source temperatures were stable (excepting when the OBCBB was cooling down). For each data collection, 100 scans were recorded (50 for each HAM side). For each scan and detector, 48 moderate resolution pixels were recorded for both the OBCBB view and SV (96 pixels for the imaging bands). The BCS and TMCBB views subtended a subset of the EV pixels. Each view was averaged over all valid pixels for a given scan. The SV was used as a dark reference which was subtracted from the signal in all other views to produce the . The was then averaged over all valid scans in a given data collection. 1-sigma estimates were also determined for each , and the corresponding SNR was derived. Outlier rejection was performed at each stage of the calculation. The resulting , , and were inserted into Equation (6).
VIIRS telemetry data was also extracted and averaged over all scans in each data collection. The telemetry data largely consisted of temperature readings, for both internal and external sources as well as instrument component temperatures. In some cases, more than one thermistor was used for a source in the thermal model; in those cases the average value was used. Data was supplied on the external sources (BCS, TMCBB, and SVS) about every 10 s, from which collection averaged temperature values were determined, and then radiances derived (a Planck function of the temperature reading convolved with the band averaged spectral response functions [
13]) and inserted into the equations in
Section 2. The telemetry of interest for the thermal model were the RTA, CAV, SH, and HAM temperatures as described in
Section 2; these temperatures were also convolved with the band averaged spectral response functions and inserted into the equations in
Section 2. Note that there was no direct temperature reading from the RTA, so that the CAV temperature was used with an instrument temperature dependent offset (∼4 K). Temperature data from the FPAs, electronics (ELEC), and instrument (OMM) was also collected, and used in the stability assessment.
Some of the input parameters, such as
[
12],
, and
[
16], were measured in previous testing and their derivation will not be described in this work.
Once the component radiances were determined, the path difference radiances from Equations (4), (5), and (8) were computed. These path difference radiances and their corresponding
were then inserted into Equation (6), and fits of the radiometric calibration coefficients
were performed over all valid data within the fitting range defined in
Figure 1. These fit coefficients are next used to derive the retrieved radiance and performance metrics (non-linearity and uniformity). The SNR was also fit over the same range via Equation (9); this SNR fit was used to estimate the SNR and NEdT at
. Lastly, the uncertainty was propagated using Equation (15) and the various uncertainty contributors described in
Section 2.3.
To determine the saturation and gain transition scene temperatures, partial views as VIIRS scanned across the BCS and TMCBB sources were used. Here the scan averaged dark offset was subtracted from each pixel of the scan across the sources profiles; the largest was averaged over all scans in a data collection and then converted into a scene radiance using Equation (11), from which the scene temperature was determined via Planck’s law.
3.2. Radiometric Performance
The offset corrected detector response
versus path difference radiance using BCS data for the MWIR bands, fit using a quadratic polynomial as described in Equation (6), are shown in
Figure 2a for detector 9. The corresponding fitting residuals in % are shown in
Figure 2b. For bands I4, M12, and M13 high gain, a significant portion of the 12 bit analog-to-digital converter (ADC) range was used. The radiance residuals for these bands increase at lower temperatures (worst case of up to ∼8% at 230 K for M12); the 230 K scene temperature translates to very low radiances in this spectral region, and this places a significant constraint on the fitting. M13 low gain (not shown) also exhibits this behavior at the low end of its measured range (near 355 K). Similarly,
Figure 3a,b show the response
versus radiance curves along with their corresponding radiance residuals in % for the LWIR bands using BCS data. For these bands, most of the 12 bit ADC range was used. The radiance residuals for these bands was generally small (less than 1% for all detectors, and much less when the lowest temperature level was excluded). The exception was detector 5 in M16B, which showed a higher residual than other detectors due to higher than average noise. All measured cases (instrument plateaus, electronics sides, HAM sides, and FPA temperatures as listed in
Table 2) showed similar behavior in terms of fitting and residuals.
In general, the
coefficient is on the order of
or less and is roughly consistent over plateaus and electronic sides. I5, M14, and M15 had the largest offsets with absolute values up to ∼0.03 [
]. The detector dependence is fairly stable
versus instrument condition for the LWIR bands, while some variation with detector was observed for the MWIR bands. There is a good deal of detector variation in the
trends
versus instrument temperature. However, it should be noted that the variation of the offset with plateau is on the order of
or less. In the majority of cases, the 2-sigma error bars overlap for the different plateaus (electronics side dependent), indicating that
is generally consistent over instrument temperature conditions. M14 is the exception, in that there appears to be some instrument temperature dependence. In addition,
does not show any noticeable trend for the three FPA temperatures measured at nominal plateau, electronic side A (see
Table 2). For M13 low gain, the derived
was non-zero and positive (between 0 and 0.5 [
]).
Figure 2.
The offset corrected detector response (a) and the radiance residual in % (b) shown versus path difference radiance. The symbols in (a) represent measured data and the lines indicate quadratic fits. The measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using detector 9 (BCS data).
Figure 2.
The offset corrected detector response (a) and the radiance residual in % (b) shown versus path difference radiance. The symbols in (a) represent measured data and the lines indicate quadratic fits. The measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using detector 9 (BCS data).
The detector-to-detector patterns for the gains (
) are very consistent from instrument temperature plateau to plateau as well as across electronics side. In addition, there is a strong odd – even detector dependence observed in bands I5, M12, M13 (both high and low gain), and M15. The gains trend fairly linearly over instrument temperature plateau (see
Table 3), with bands I4, M12, and M13 (both high and low gain) roughly consistent over plateaus and LWIR bands decreasing with instrument temperature. In general, the LWIR bands decrease by about 2% or less and the MWIR bands change by less than 1% over the ∼20 K measured instrument temperature range. Here the temperature variation is outside the 2-sigma error bars in most cases for the LWIR bands, but consistent for the MWIR bands. In addition, the electronics sides are not consistent for bands I5 and M14–M16. The gains determined at the three measured FPA temperatures for nominal plateau, electronics side A are consistent for bands I4, M12, and M13 high gain; however, for the LWIR bands, the gains decrease with increasing FPA temperature by between 3%–8% for every 1.5 K (see
Table 3).
For the LWIR bands, the detector pattern of the nonlinear term () is roughly constant over plateau and electronics side; however, for the MWIR bands, the pattern is less well defined. In terms of magnitude, the nonlinear term is consistently on the order of or less. For most bands and electronics sides, the nonlinear term exhibits a small, roughly linear increase with instrument temperature. There is some noticeable detector variation in the trending for the MWIR bands. In the majority of cases, the 2-sigma error bars overlap. The most glaring exception is M14. for the MWIR bands shows almost no change with FPA temperature, while increases with increasing FPA temperature for the LWIR bands.
Figure 3.
The offset corrected detector response (a) and the radiance residual in % (b) shown versus path difference radiance. The symbols in (a) represent measured data and the lines indicate quadratic fits. The measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using detector 9 (BCS data).
Figure 3.
The offset corrected detector response (a) and the radiance residual in % (b) shown versus path difference radiance. The symbols in (a) represent measured data and the lines indicate quadratic fits. The measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using detector 9 (BCS data).
The calibration coefficients derived using the OBCBB were compared to those computed from BCS data using a truncated fitting range that approximates the measured OBCBB range from instrument ambient to 315 K. For both cases, 2-sigma uncertainties were also determined for each coefficient. Comparing the two sets of calibration coefficients showed that for the majority of cases all three coefficients were within the 2-sigma uncertainties. Some detectors were exceptions in the following cases: the linear coefficient for M12 at cold and nominal instrument plateaus, the linear coefficient for M13 at hot instrument plateau, and the nonlinear coefficient for M14 at cold instrument plateau.
All of the thermal bands were shown to saturate above their design specifications [
8] (see
Table 4). All bands digitally saturate first, including both M13 gain stages; two bands (I4 and M12) exhibit analog saturation at some higher radiance within the measured range. For these two bands, the digital response decreases to zero at the highest measured radiance levels. On-orbit, this results in two possible values for a given digital response; fortunately, scenes with temperatures above saturation for these bands are usually fires, and can be correlated to M13 for quality control. The LWIR band saturation temperatures are generally consistent between different instrument plateaus to within ∼ 1 K and between electronics sides to within 2–3 K; for the MWIR bands, the separation between electronics sides is less clear, but all instrument conditions are consistent to within 2–3 K. The detector dependence is largely the reflection of the detector gain dependence. Bands I4, M12, and M14 saturate between 4 K and 10 K above the design specification; all other bands saturate more than 10 K above their design specifications.
Table 3.
Averaged gains () over all detectors and HAM sides measured during radiometric performance testing for all instrument temperature plateaus, electronics sides (A or B), and focal plane temperatures using BCS data.
Table 3.
Averaged gains () over all detectors and HAM sides measured during radiometric performance testing for all instrument temperature plateaus, electronics sides (A or B), and focal plane temperatures using BCS data.
| Cold | Nominal | Nominal | Nominal | Hot | Hot | Cold | Nominal | Hot |
---|
Band | A | A | A | A | A | A | B | B | B |
---|
| 80.5 K | 80.5 K | 82 K | 83.5 K | 80.5 K | 82 K | 80.5 K | 80.5 K | 80.5 K |
---|
I4 | 1134.6 | 1134.2 | 1133.8 | 1133.7 | 1134.3 | 1133.5 | 1134.0 | 1133.4 | 1134.3 |
I5 | 175.9 | 174.9 | 164.3 | 153.0 | 174.7 | 164.1 | 174.5 | 173.7 | 174.0 |
M12 | 1137.5 | 1137.1 | 1137.7 | 1138.4 | 1136.9 | 1137.5 | 1135.5 | 1135.6 | 1135.4 |
M13 HG | 597.8 | 597.7 | 597.2 | 596.8 | 597.8 | 597.4 | 596.9 | 596.9 | 597.2 |
M13 LG | 6.9 | 6.9 | – | – | 6.9 | – | 6.9 | 6.9 | 6.9 |
M14 | 175.1 | 174.2 | 162.5 | 149.0 | 173.4 | 161.6 | 172.3 | 171.4 | 170.5 |
M15 | 184.1 | 183.3 | 178.0 | 171.9 | 183.0 | 177.7 | 182.7 | 182.0 | 181.9 |
M16A | 201.7 | 200.0 | 190.1 | 179.3 | 199.4 | 189.5 | 200.7 | 199.3 | 199.2 |
M16B | 201.0 | 199.3 | 189.5 | 178.7 | 198.7 | 188.9 | 199.9 | 198.5 | 198.4 |
Using Equation (9), the radiance at which the SNR was equal to 5 was calculated. The equivalent scene temperature was estimated, the band average of which is listed in
Table 4 using nominal plateau data. For most bands this is below the design specifications [
8]; the exceptions were I4 and M14. Estimates of the measured
are consistent over instrument conditions. Note that there is some uncertainty in this estimate for bands I5, M13 low gain, M15, and M16 due to extrapolation well below the measured range.
The gain transition temperatures for M13 were measured to be between 343 and 349 K. There was about a 2 K spread in transition temperature with detector (largely due to detector-to-detector gain differences) and about a 3 K variation over instrument temperature (resulting from increasing background emission at higher instrument temperatures). This is slightly larger than the design specification [
8], but has been deemed low risk to the science products.
The NEdT as a function of scene temperature is graphed in
Figure 4 for all high gain bands (detector 9). The NEdT increases as the scene temperature decreases for all bands due to the influence of the derivative of the Planck function in Equation (
10). For the LWIR bands, the NEdT is below ∼0.6 K, even at the lowest scene temperatures; for the MWIR bands, the NEdT increases to between 2.5–3.5 K for M12 and M13 at 210 K and roughly 3.5 K for I4 at 230 K. The NEdT is consistent across instrument conditions, except for the slight increase with instrument temperature due to increasing background emission. M13 low gain is not shown, but behaves similarly to the other bands; the NEdT is below ∼0.2 K for scene temperatures above about 400 K and increases to up to ∼0.5 K below 400 K. The elevated focal plane measurements had the effect of increasing the NEdT for the LWIR bands at all scene temperatures, with the greatest increases at low temperatures; in contrast, the MWIR bands showed very little difference in NEdT with increasing focal plane temperature.
The NEdT at
for nominal plateau, electronic side A are listed in
Table 4 (band maximum over detectors and HAM sides). The specified limit for NEdT at
is also listed; all bands are well below the limit for all conditions. The only exception is detector 5 in band M16B, which is close to, but still below the specified limit. In addition, detector 4 in band M15 was consistently out-of-family. The NEdT at
is very consistent over the range of instrument conditions tested, both in terms of magnitude and detector dependence. The NEdT generally increases with instrument temperature for all bands and both electronics sides (although there is a good deal of detector dependence). This is the result of the increasing background emission in the detectors which occurs at higher instrument temperatures. The NEdT at
increases about 10% over the ∼20 K range of measured instrument temperatures (except for M12 which increases roughly 30%).
Figure 4.
Plots show the measured NEdT as a function of scene temperature for the MWIR bands (a) and for the LWIR bands (b). Measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using detector 9.
Figure 4.
Plots show the measured NEdT as a function of scene temperature for the MWIR bands (a) and for the LWIR bands (b). Measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using detector 9.
Table 4.
Band maximum
, NEdT at
,
, and
measured during radiometric performance testing at the nominal instrument temperature plateau, electronics sides A, and 80.5 K focal plane temperature compared to their respective sensor specifications [
8].
Table 4.
Band maximum , NEdT at , , and measured during radiometric performance testing at the nominal instrument temperature plateau, electronics sides A, and 80.5 K focal plane temperature compared to their respective sensor specifications [8].
Band | [%] | NEdT at [K] | [K] | [K] |
---|
| Meas | Spec | Meas | Spec | Meas | Spec | Meas | Spec |
---|
I4 | 0.1 | 1.0 | 0.428 | 2.500 | 357 | 353 | 231 | 210 |
I5 | 0.2 | 1.0 | 0.524 | 1.500 | 369 | 340 | 167 | 190 |
M12 | 0.2 | 1.0 | 0.131 | 0.396 | 359 | 353 | 216 | 230 |
M13 HG | 0.2 | 1.0 | 0.046 | 0.107 | 363 | 343 | 213 | 230 |
M13 LG | 0.1 | 1.0 | 0.231 | 0.423 | 644 | 634 | 276 | 343 |
M14 | 0.5 | 1.0 | 0.060 | 0.091 | 348 | 336 | 195 | 190 |
M15 | 0.2 | 1.0 | 0.035 | 0.070 | 357 | 340 | 173 | 190 |
M16A | 0.2 | 1.0 | 0.045 | 0.072 | 366 | 340 | 154 | 190 |
M16B | 0.5 | 1.0 | 0.072 | 0.072 | 367 | 340 | 156 | 190 |
The non-linearity metric (NL) is generally consistent over instrument conditions and well below the specified limit (see
Table 4). Detector 5 in M16B is again an outlier, but is still well below the 1.0% limit. M14 exhibits the largest average non-linearity at around 0.5%.
The potential for detector-to-detector striping was measured by the uniformity metric, graphed in
Figure 5 for the worst case detector per band. A value greater then one indicates the potential for striping (dashed horizontal red line in the plots) [
8]. The uniformity metric generally increases with increasing scene temperature, indicating increasing potential for striping as the temperatures rises. The potential for striping exists for many bands at the highest scene temperatures (in particular bands M13, M14, and M15). At higher temperatures, the deviation of the retrieved radiance from the band average increases, but the measured NEdL levels off (see Equation (13)); the result is a steadily increasing uniformity metric. On-orbit, the possibility of observing striping with M13 low gain is very limited, given the sparsity of measurements available.
Figure 5.
Plots show the measured uniformity metric as a function of scene temperature for the MWIR bands (a) and for the LWIR bands (b). Measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using the worst case detector.
Figure 5.
Plots show the measured uniformity metric as a function of scene temperature for the MWIR bands (a) and for the LWIR bands (b). Measured data was taken from nominal plateau, HAM side A, electronics side A, FPA temperature 80.5 K using the worst case detector.
3.4. Uncertainty
The total uncertainties for the EV retrieved radiance (for a single, un-aggregated pixel) are shown as a function of scene temperature in
Figure 6 (worst case detector). For the MWIR bands, the results are between 0.1–0.2 K at high scene temperatures (above ∼290 K), but increase rapidly below scene temperatures of about 260–270 K (above about 1 K for I4 below 260 K and for M12 and M13 below 230 K). The LWIR bands uncertainty estimate is between 0.1–0.2 K above 250 K, below which the uncertainty rises slowly to between 0.3–0.5 K at 190 K. I5 shows a little higher uncertainty than the other LWIR bands, with 0.2–0.3 K above 250 K and slowly increasing up to about 0.8 K at 190 K. The calculated total uncertainties (in % radiance) are listed in
Table 5 at the scene temperatures required by the design specifications; for most cases, the modeled uncertainty was below the design specifications [
8]. The exceptions were bands M12 and M13 at both 230 and 270 K.
The uncertainties for the MWIR bands were dominated by contributions from two terms:
and
. The
uncertainty was roughly constant with scene temperature at about 0.3% and was dominant above ∼320 K for band I4 and about 290 K for the bands M12 and M13. The
uncertainty increased rapidly as the scene temperature decreased and dominated the uncertainty at all other scene temperatures. As this estimate was for a single, un-aggregated pixel and this uncertainty contributor is largely a statistical error, this contribution will decrease when large data sets are used. In those cases, the uncertainties listed in
Table 5 for M12 and M13 at both 230 and 270 K will fall below the design specifications. The
uncertainty will then dominate above about 260 K for the bands I4, M12, and M13; at lower temperatures the largest contributor is the
error, which increases rapidly as the scene temperature decreases. As a result, the reported uncertainties for these bands at low scene temperatures pose a low risk to the science products.
For most of the LWIR bands, there is no clearly dominant contributor at higher scene temperatures (above about 270 K). The exception is I5, where higher noise causes an increased contribution from the uncertainty at high scene temperatures. As with the MWIR bands, this contribution will decrease when large data sets are used. At lower scene temperatures, the main contributors are the and uncertainties, which all increase as the scene temperature decreases.
The impact of pixel aggregation and of scan angle was also investigated. For the bands in which the was the dominant contributor at some scene temperatures (I4, I5, M12, and M13 especially), pixel aggregation had the effect of reducing the uncertainty. In the bands for which the was a major contributor (bands M14 and M15), the uncertainty tended to be largest at the end of scan and lowest at the beginning of scan. Band M16 showed a mix of the two effects.
Figure 6.
Plots show the modeled total temperature uncertainty as a function of scene temperature for the MWIR bands (a) and for the LWIR bands (b).
Figure 6.
Plots show the modeled total temperature uncertainty as a function of scene temperature for the MWIR bands (a) and for the LWIR bands (b).
Table 5.
Estimated uncertainty [% radiance] compared to the sensor specifications [
8].
Table 5.
Estimated uncertainty [% radiance] compared to the sensor specifications [8].
Band | | 190 K | 230 K | 267 K | 270 K | 310 K | 340 K |
---|
I4 | Spec | – | – | 5.00 | – | – | – |
| Meas | – | – | 2.64 | – | – | – |
I5 | Spec | – | – | 2.50 | – | – | – |
| Meas | – | – | 0.48 | – | – | – |
M12 | Spec | – | 7.00 | – | 0.70 | 0.70 | 0.70 |
| Meas | – | 9.14 | – | 0.78 | 0.35 | 0.33 |
M13 | Spec | – | 5.70 | – | 0.70 | 0.70 | 0.70 |
| Meas | – | 7.81 | – | 0.78 | 0.29 | 0.27 |
M14 | Spec | 12.30 | 2.40 | – | 0.60 | 0.40 | 0.50 |
| Meas | 2.86 | 0.73 | – | 0.30 | 0.20 | 0.18 |
M15 | Spec | 2.10 | 0.60 | – | 0.40 | 0.40 | 0.40 |
| Meas | 1.35 | 0.41 | – | 0.20 | 0.15 | 0.13 |
M16A | Spec | 1.60 | 0.60 | – | 0.40 | 0.40 | 0.40 |
| Meas | 0.97 | 0.30 | – | 0.17 | 0.12 | 0.11 |
M16B | Spec | 1.60 | 0.60 | – | 0.40 | 0.40 | 0.40 |
| Meas | 1.46 | 0.41 | – | 0.20 | 0.14 | 0.12 |
A covariance terms between the radiometric coefficients and the other contributors at the retrieved radiance level were estimated by using the Schwarz inequality as outlined in Equation (
16). For all bands, the uncertainty only marginally increased at ∼292 K (where the calibration is tied to the OBCBB) and increased as the scene temperature increased or decreased. For the majority of modeled scene temperatures, the increase did not exceed 0.15%; the exceptions were the lowest scene temperatures for bands I4, M12, M13, and M14 as well as high temperature for band I4.
The impact of model validity was investigated by performing linear, quadratic, and cubic fits, and propagating the errors separately for each. In general, only marginal improvement in the uncertainty was derived by increasing the polynomial order from 2 to 3 (the biggest differences were at low temperature in the MWIR bands); in contrast, the quadratic model was a significant improvement over the linear (especially for M14). In addition, the differences between the EV retrieved radiance and BCS source radiance for all bands were minimized at 292 K by removing a bias of ∼27 mK from the OBCBB temperature.